T M LIM2010. Line Isolation Monitor (LIM) Technical Bulletin NAE /

Similar documents
Line Isolation Monitor (LIM) LIM2000plus

MK Series. Remote Indicators For Line Isolation Monitors and Fault Location. Technical Bulletin NAE /

isopv and AGH-PV Ground Fault Detector for Ungrounded Solar Arrays up to 1100 V And Isolation Tester Prior to Array Startup (Grounded and Ungrounded)

Isolated Power Systems Equipment

isopv425 and AGH420 Ground Fault Detector for Ungrounded Solar Arrays < 100 kw And Isolation Tester Prior to Array Startup (Grounded and Ungrounded)

T M. User Manual LIM2010. Line Isolation Monitor

CMD420 / CMD421 Series

GM420. Digital Ground Continuity Relay For AC Systems. Technical Bulletin NAE /

Dipl.-Ing. W. Bender GmbH & Co. KG Londorfer Str Grünberg Phone: Fax:

Remote alarm indicator and test combination MK2430

ISOSCAN EDS460-DG. Insulation fault locator for DC IT systems with high system leakage capacitances

ISOSCAN EDS460/490 EDS461/491. Insulation fault locators with control and display function for EDS systems (insulation fault location systems)

ISOMETER IRDH375. Insulation monitoring device for unearthed AC, AC/DC and DC systems (IT systems)

Insulation fault evaluator EDS460-DG

1.7. Insulation fault evaluators EDS460/490 EDS461/491

1.7. Insulation fault evaluators EDS460/490 EDS461/491

ISOMETER IRDH275BM-7 with coupling device AGH675-7 and AGH675-7MV15

T M. User Manual. MarinaGuard. Ground Fault Monitoring Panel For Marina Shore Power

ISOMETER isomed427p. Application Medical IT system in accordance with IEC , IEC , IEC and DIN VDE

Dipl.-Ing. W. Bender GmbH & Co. KG Londorfer Str Grünberg Tel.: Fax:

Multi-Gas-Controller MGC2

1.7 A-ISOMETER IRDH575. RoHS. Dipl.-Ing. W. Bender GmbH & Co. KG Londorfer Str Grünberg Tel.: Fax:

ISOMETER IR425. Insulation monitoring device for unearthed AC/DC control circuits (IT systems)

ISOMETER IR425. Insulation monitoring device for unearthed AC/DC control circuits (IT systems)

LINETRAXX RCMA420. Residual current monitor for monitoring AC, DC and pulsed DC currents in TN and TT systems

Residual current monitor RCM470LY/RCM475LY. for TN and TT systems (AC and pulsating DC currents)

ISOMETER IR1575. Insulation monitoring device for unearthed AC/3(N)AC systems up to 480 V and DC systems up to 480 V

LINETRAXX RCMS460-D/-L RCMS490-D/-L

IT system floor-standing distribution

LINETRAXX VME420. Multi-functional relay for overvoltage, undervoltage and frequency monitoring in AC/DC systems with external supply voltage

Two-Channel Gas Controller

LINETRAXX VMD420. Multi-functional voltage relay for 3(N)AC systems, frequency/overvoltage/undervoltage, phase, phase failure, asymmetry

MLHG6 / MLHG6-AU. Iso-Gard Line Isolation & Overload Monitor (LIOM) and Line Isolation Monitor (LIM) Installation and Reference Guide DANGER

DEB71. Earth Leakage Monitoring Relay. Benefits. Description

Digital Gas Transmitters with Auxiliary Inputs & Control. PolyGard 2 DC6

Nitrogen Dioxide (NO2) Single-Point Gas Detection System

MLHG6-AU. Iso-Gard Line Isolation Monitor (LIM) Setup Guide ! WARNING. Components

RCMA423-DM RCMA423-DM1C RCMA423-DM2C RCMA423-DM3C. Operating Manual

TM800. Remote alarm indicator and operator panels

Operating Manual RCM420

Oxygen (O2) Single-Point Gas Detection System

Line Isolation Monitor LIM2000plus Confirms the Integrity of Isolated Power Systems

Isolated Power Panels Line Isolation Monitor (LIM) Remote Indicators/Displays Accessories 15-6

Replaceable LED modules. Sleep or unattended mode. Auto-silence and auto-acknowledge

Danfoss gas detection units

RTA421. Technical Information. Limit alarm switch Alarm switch and power supply for monitoring current or voltage signals

M2500 Engine Controller Installation Manual

DPB52. True RMS 3-Phase, Multifunction monitoring relay. Benefits. Description

NGC-40 PANEL MOUNTED ADVANCED MODULAR HEAT-TRACING CONTROL SYSTEM HTC HTC3 PRODUCT OVERVIEW

LC1350 Level Controller

Model D12-IR. Digital Gas Transmitter...

SLATE. Base Module INSTALLATION INSTRUCTIONS R8001A1001

Danfoss gas detection units

NGC-40 PANEL MOUNTED ADVANCED MODULAR HEAT-TRACING CONTROL SYSTEM HTC HTC3

Process Alarm Module RTA 421

21-light Remote Annunciator. Owner s Manual

NGC-40 Advanced Heat-Tracing Control System

02/11/2015

D326A/D334A/D339A. Installation Instructions. Point Contact Modules

Public Safety DAS Annunciator Panel

02/11/2015

C-Bus Four Channel General Input Unit Installation Instructions

Product Data Sheet. Remote Terminals. Features:

PolyGard 2-channel Gas Controller MGC-03

SECTION ISOLATED POWER SYSTEMS

Operating Manual VMD420

PolyGard Single Point Controller SPC-X3-20XX for refrigerant gases and vapours

SCHMIDT LED Measured Value Display MD Instructions for Use

HTC-915-CONT. Heat-Trace Control system

GasGard 100 Control System

DGC-500 DIGITAL GENSET CONTROLLER

Carbon Monoxide Transmitter

ABB Instrumentation. Specifications. COMMANDER 150 Universal Process Indicator. the 1 /8 DIN indicator to match all your display requirements

NU2 Smart Power Supply (SPS) Manual

Models NFPA 1221-A, NFPA 1221-B Public Safety DAS Annunciator Panel. Revision E 61117

4 & 8-POINT ANNUNCIATORS Instruction Manual

Universal fault annunciator for panel mounting

MODEL FPT-130 SINGLE POINT FREEZE PROTECTION HEAT TRACE CONTROL

Gas Controller Module GC-06

DENVER PUBLIC SCHOOLS DESIGN AND CONSTRUCTION STANDARDS This Standard is for guidance only. SECTION IBAS LIGHTING CONTROL

PolyGard Single Point Controller SPC-X3-11XX for Toxic Gases and Oxygen

GG-2 2-CHANNEL GAS DETECTION CONTROL PANEL. Installation and Operation Manual

Monitoring Devices 11/1

Across-the-Line SERIES MP300 Combined Manual and Automatic

CommStat 6. Controller for Redundant HVAC Systems PRODUCT DATA SHEET

Operating & Maintenance Manual. Alert-4 Ethernet LCD Master Alarm

M1000 Alarm Annunciator

These instructions detail procedures to follow in order to avoid damage to equipment.

Soft Start Series MP700 Solid State, Reduced Voltage

Fire Extinguishing Control Panel INSTRUCTION MANUAL. Revision 8/ Instruction Manual Page 1 Revision 8/01.17 of 63

ACC-25/50ZST. Zoned System Voice with Telephone Evacuation Control Panel

Addressable NAC Output Module D327A

DTA. Motor thermistor relay. Benefits. Description

Table of Contents. T538 Instruction Manual Rev # 6 June R6T538 June rtf E L Switchgear and Control Inc

Panel Mounted Fault Annunciator Series

Cat. N. Cat. N (s) 1. DESCRIPTION - USE 3. OVERALL DIMENSIONS. Technology : 4. PREPARATION - CONNECTION 2. PRODUCT RANGE

GASGUARDIAN Channel Controller OPERATING & INSTALLATION MANUAL

BC3250. Blowdown Controller

Product data sheet AIRWIN additional modules Accessories AIRWIN

1S25. Arc Fault Monitor 4 Zones, 8 Sensors. Features. Introduction. ARC Fault Protection

Transcription:

2 T M LIM2010 Line Isolation Monitor (LIM) Technical Bulletin NAE2022011 / 04.2015

Line Isolation Monitor LIM2010 Description The LIM2010 Line Isolation Monitor (LIM) measures the Total Hazard Current (THC) in an isolated (ungrounded) AC system. The Total Hazard Current is calculated by measuring the system s leakage impedance to ground. Alarm indication is displayed on a seven-segment display, LED bar graph, and digital display. The LIM2010 operates on systems from 100 to 240 V, 50 or 60 Hz. No auxiliary supply voltage is required. Two separate ground connections are provided for ground connection monitoring. A break in either of these connections will activate the connection alarm. The LIM2010 features many different alarms, including: LIM2010 Features No interference with electrical equipment Special phase-locking circuitry for ultimate stability and repeatability 2 programmable voltage-free SPDT contacts for external alarms Provision for remote indicators Easy to clean rugged front foil Digital and analog bar graph displays Automatic self-calibration and self-check Audible alarm volume adjustable via menu Transformer load monitoring (optional) Transformer overtemperature monitoring (optional) RS-485 connections for compatibility with Bender's remote communication system Provision to control multiple line isolation monitors via one intelligent remote Additional communication devices available: Web browser based E-mail and SMS Interfaces with BENDER ground fault location system Total Hazard Current (THC) Transformer overload Transformer overtemperature Overvoltage and undervoltage Ground connection Insulation resistance and impedance Ground fault location (when used with BENDER EDS location system) The LIM2010 is ideal for retrofit applications regardless of the make or type of the existing product. No metal needs to be cut and simple instructions will generally facilitate a changeover in less than one hour. Operational information Generally, the impedance between each isolated conductor and ground is different, resulting in a different current flowing through a person making physical contact between any one of the isolated conductors and ground. The LIM calculates and displays the true maximum value of the Total Hazard Current (THC). The BENDER LIM performs this function using a patented measurement technique. The THC is shown on the seven-segment display and the LED bar graph. In the normal condition, the green SAFE LED is illuminated, the display shows a low leakage value (green) and the bar graph is in the non-alarm, or normal, green zone. THC levels will increase as additional loads are connected to the system, and/or when a line-to-ground fault has suddenly occurred or is slowly developing. A visual and audible alarm is generated when the THC exceeds the LIM setting of either 2 ma or 5 ma (red). Two programmable relay output contacts are available, which can be wired into a circuit to trigger an external alarm. The red HAZARD LED remains illuminated for the duration of the alarm condition. The audible alarm may be silenced by pushing the MUTE button at the discretion of personnel. When the MUTE button is activated, the built-in amber LED in the MUTE button is illuminated to indicate a muted condition. After the fault is removed, the LIM will automatically reset to the normal condition. Line isolation monitor indicators THC THC display Text display Safe LED Hazard LED Buzzer < 5 (2) ma value (green) SAFE ON OFF OFF 5 (2) ma value (red) HAZARD OFF flashing ON > 9.9 ma EF (red) HAZARD OFF flashing ON 2 The audible alarm volume level is adjustable via the configuration menu. Activate the test button to check the LIM operation. This test does not add to the hazard current of a system in use, nor does the test include the effect of the line-to-ground stray impedance of the system. The LIM has provisions to connect one or more remote indicators, with or without a digital meter. The remote indicators duplicate the audible and visible alarm signals of the LIM.

Additional features Bender's communication bus enables the LIM2010 to connect to several intelligent devices to provide a complete isolated power solution, including: Intelligent remote control stations monitoring multiple systems Fault location equipment, reducing time and cost finding faults Communication equipment to connect to industry standard communication networks, including Ethernet and Modbus Standards The BENDER LIM2010 Series LIM complies with UL 1022 and CSA-C22.2 No. 204-M1984. The intent is to include the LIM as part of an isolated power system that conforms with the applicable requirements of ANSI/NFPA 99, ANSI/NFPA 70, and CAN/CSA-C22.2 No. 29-M1989. Front display and operating elements Ordering information - LIM2010 Line isolation monitor 100-240 V / 1-phase LIM LIM2010 B 9207 5021 Ordering information - Accessories Refer to devices' respective datasheets for more information. Devices marked with an asterisk (*) have multiple ordering numbers based on model. Connector plate LIM + remote connections CP-LIM2010 B 5111 00001 3 1 4 5 7 2 Remote indicators Mute MK2000-G1 B 5213 00002 Mute + Test MK2000P-G1 B 5213 00188 SAFE MUTE ESC Mute + Overload MK2000C-G1 B 5213 00020 Mute + Test + Overload MK2000CP-G1 B 5213 00021 TEST Digital remote indicator MK2000CBM B 5213 00022 HAZARD RESET MENU Digital remote station MK2430 * Digital remote station MK800 * Load monitoring current transformers 6 8 1 - THC Set Point LED Markers (amber) 5 ma or 2 ma 2 - TEST Button Checks functions of the LIM 3 - SAFE LED (green) Lit unless LIM is in alarm mode 4 - Digital Display Displays THC in ma 5 - Analog LED Bar Graph Displays THC in ma 6 - HAZARD LED (red) Dark unless THC > Set Point 7 - MUTE Button w/built-in LED (amber) Silences alarm buzzer 8 - Digital text display for status and menu options Up to 100 A load current STW3 B 9802 1000 Up to 200 A load current STW4 B 9802 1001 Up to 100 A load current (split core) SWL-100A B 9802 1002 3

Wiring diagram: Standard wiring for connector plate 1 - Connector, to LIM2010 (Viewed from mating end) L1, L2 To secondary of isolation transformer A B 12 V DC Com. Common connection 1 5 A, B RS-485 (BMS) communication interface RI1 Test button source K1/NC Alarm relay K1, N/C K1/Common Alarm relay K1, common 2 1S1 1S2 4 Z1 / M+ Z2 / M- K2 / Com. K2 / NC K2 / NO 3 L1 L2 12VDC CM A B RI1 K1/NC K1/COM K1/NO SAFE HAZARD RI2 GND2 LIMGND TEST 1S1 Z1/M+ 1S2 Z2/M- K2/COM K2/NC K2/NO 6 9 8 7 K1/NO Safe Hazard Alarm relay K1, N/O Safe light connection for external remote indicator Hazard light connection for external remote indicator RI2 Local and system muting from LIM and remote indicator GND2, LIM GND Separate ground connections Test Connection for remote test 2 - Connector, to LIM2010 Z1/M+, Z2/M- K2/Common K2/NC K2/NO Connection for overtemperature sensor or external analog meter Alarm relay K2, common Alarm relay K2, N/C Alarm relay K2, N/O 3 - Connector Plate CP-LIM2010 4 - Wiring Harness 5 - MK2000 LIM Remote Indicator Terminals 6 - To L1 and L2 of isolated power system 7 - Connection for optional system muting 8 - Connection for optional remote test 9 - Separate connections to ground 4

Wiring diagram: Basic connection to LIM2010 SAFE MUTE ESC TEST HAZARD RESET 2 L1 L2 12VDC CM A B RI1 K1/NC K1/COM K1/NO SAFE HAZARD RI2 2a 2b MENU 1 GND2 LIMGND TEST 1S1 Z1/M+ 1S2 Z2/M- K2/COM K2/NC K2/NO 1 - LIM2010 2 - Connector plate CP-LIM2010 2a - Isolated power system connections 2b - Panel ground or ground bus 5

6 Technical data: LIM2010 Insulation coordination acc. to UL 1022 and IEC 60664-1 Rated insulation voltage AC 250 V Rated impulse voltage / pollution degree 2.5 kv / III Voltage test acc. to UL 1022 and IEC 61010-1 2.0 kv Supply Voltage Supply voltage U S = U n Power consumption < 22 VA Isolated Power System Monitored Nominal voltage U n AC 100-240 V Operating range of U n 85% - 110% Frequency range f n 50 / 60 Hz Operating range of f n ± 5% Insulation and THC monitoring Response value: THC 2 ma / 5 ma (5 ma)* Response Tolerance 1.8-2 ma / 4.5-5 ma Hysteresis 20% Response value Z 10-200 kω (off)* Response tolerance ±15% Hysteresis 25% Response value R 20-200 kω (off)* Response tolerance ± 15% Hysteresis 25% Response time t an < 4 s Measuring circuit Measuring voltage U m ± 48 V Measuring current I m (at Z F = 0 Ω) < 32 μa Internal resistance 1.5 MΩ Monitor hazard current MHC, 120 V / 240 V 60 μa / 95 μa When EDS mode is active: Monitor Hazard Current MHC < 950 μa Test cycle / idle time 2 s / 4 s Voltage monitoring Response value, undervoltage / undervoltage (<U / >U) 80-300 V (off)* Response tolerance ± 5% Hysteresis 5% Load current monitoring ( C option) Response value 10-200 A (off)* Response tolerance ± 5% Hysteresis 5% Temperature monitoring Response value (fixed) 4 kω Release value 1.6 kω PTC resistor acc. to DIN 44081 max. 6 connected in series Adjustable time delays (does not apply to THC alarm) Response delay t on 0-99 s (0 s)* Delay on release t off 0-99 s (0 s)* Displays, memory 14-segment display 8 digits, multi-functional Displayable value, THC 0.0-9.9 ma Operating uncertainty, THC + 7%, ± 0.1 ma Measured value, load current (as % of response value) 10-199 % Operating uncertainty, load current (as % of response value) ± 5%, ± 0.2 A Measured value, load current (in A) 0.5-250 A Operating uncertainty, load current (in A) ± 5%, ± 0.2 A Measured value, system voltage 10-300 V Operating uncertainty, system voltage ± 5%, ± 2 V Measured value, insulation impedance Z 0-1 MΩ Operating uncertainty, insulation impedance ± 5%, ± 1 kω Measured value, insulation resistance R 2 kω - 1 MΩ Operating uncertainty, Z ~ R ± 20%, ± 1 kω Measured value, leakage capacitance C 0-500 nf Operating uncertainty, Z ~ X C ± 20%, ± 5 nf Condition for separate readings of R and C Z 2 kω 7-segment display 2 digits, digital THC indication Bar graph indicator analog THC indication History memory 300 data records Data logger 300 data records Inputs / Outputs Analog current output M+ / M- 0-400 μa Operating uncertainty ± 10% Output RI1, 12 VDC common 12 V / 200 ma RI2, SAFE, HAZARD, TEST Maximum four (4) MK2000(C)(P) Cable length 32 ft Serial Interface Interface A-B / Protocol RS-485 / BMS bus Baud rate 9600 baud Cable length 3900 ft Recommended cabling Shielded, twisted pair, one end grounded Termination resistor 120 Ω (also activated via DIP switch) (off)* Assignable BMS bus addresses 1-90 (1)* Relays Number of switching elements 2 SPDT contacts Operating principle normally energized or de-energized operation (N/E)* Electrical service life, number of cycles 10,000 Contact data acc. to IEC 60947-5-1 Relay 1: Utilization category AC-13 AC-14 DC-12 DC-12 DC-12 Rated operational voltage 230 V 230 V 24 V 110 V 220 V Rated operational current 5 A 3 A 1 A 0.2 A 0.1 A Minimum contact load 1 ma at AC / DC 10 V Relay 2: Utilization category DC-12 DC-12 DC-12 Rated operational voltage 24 V 110 V 220 V Rated operational current 1.2 A 0.4 A 0.25 A Minimum contact load 1 ma at AC / DC 10 V Rated Contact Voltage AC 125 V / DC 30 V Environment / EMC EMC IEC 61326 Operating Temperature Range + 14 - + 122 ºF - 10 - + 50 ºC Storage Temperature - 13 - + 158 ºF - 25 - + 70 ºC Connection Connection type Molex plug 15-pin, type 03-09-2159 and 12-pin, type 43045-1215 General data Operating mode continuous operation Mounting position display-oriented Degree of protection, internal components (EN 60529) IP30 (NEMA 1) Enclosure material polycarbonate Flammability class UL94 V-0 Screw fixing Qty. 4, #4-40 oval head, black oxide finished Screw torque (2.6-3.5 lb-in) 0.3-0.4 N-m Weight 1.2 lb

Technical data: Connector plate LIM2010 Connector Plate CP-LIM2010 Cable Length 20 Terminal Strip 22 terminals Connector 15 pin Molex Conductor Size AWG 22-12 Tightening torque 8 in-lb Mounting Orientation any Weight Approx. 7 oz. Technical data: MK2000-G1 / MK2000-G2 / MK2000P-G1 / MK2000C-G1 / MK2000CP-G1 / MK2000CBM Operating voltage 12V DC or 12V AC Max. current 50 ma (MK2000CBM 100 ma) Operation class continuous operation Ambient temperature when operating +32º F to +122º F 0º C to +50º when stored -13º F to +158º F -25º C to +70º C Connection screw terminal block Conductor size AWG 30-12 Tightening torque 5-7 lb In. Mounting by screws Weight MK2000-G1 / MK2000P-G1 / MK2000C-G1 / MK2000CP-G1 0.25 lb MK2000-G2 / MK2000CBM 0.32 lb Technical data: STW3, STW4, STW-100A Insulation coordination according to IEC 60664-1: Rated voltage U m AC 720 V / AC 720 V / AC 600 V Rated impulse voltage U isol 4 kv / 4 kv / 2.2 kv Measuring circuit Max. rated primary current 100 A / 200 A / 100 A Min. rated primary current 1 A / 2 A / 0.1 A Nominal frequency 50-400 Hz General data Ambient temperature, during operation +32º F to +122º F 0º C to +50ºC Operating mode continuous operation Position any position Connection Faston plug 6.3 x 0.8 mm / screw terminals Type of connection to the measuring current transformer Single wires AWG 18 up to 3 ft Single wires, twisted AWG 18 up to 30 ft Screened cable AWG 20 (single-ended connection to ground) e.g. J-Y(St)Y 2 x 0.8 up to 130 ft Mounting Screw Mounting M3 / zip ties Flammability class UL94V-0 7

Dimensions: Load monitoring CT STW3/STW4 Dimensions in inches (mm) 0.25 (6.25) 1.26 (32) 1.96 (50) Ø 0.7 (17.5) Dimensions: LIM2010 Dimensions in inches (mm) The front plate provides four holes with a diameter of 1/8 (3.2 mm) for screw mounting. Use the provided #4-40 oval head, black oxide finished screws. Use minimum 2.6 lb-in (0.3 N-m), maximum 3.5 lb-in (0.4 N-m) torque.a 15-pin female connector and a 12-pin female Molex connector are built into the back of the LIM2010. 1.96 (50) 1.96 (50) 7 (177) 6.5 (164) 1.96 (50) 1.575 (40) 0.63 (16) 1.26 (32) Dimensions: Load monitoring CT SWL-100A Dimensions in inches (mm) 0.630" (16) 1.220" (31) 1.469" (37.5) 1.181" (30) 2.185" (56) 0.591" (15) 4.4 (112) 3.2 (82) 4 (101.5) 2.4 (62) 6.5 (165) 5.3 (134.5) 3.1 (79.5) 8

A complete isolated power solution for healthcare facilities - Monitor, locate, communicate Web interface Visualization Ground fault location while the system remains online TCP/IP/Ethernet > F Line Isolation Monitor LIM2010 Modbus/TCP network Ethernet Modbus/TCP COM460IP GF Location Module EDS Series L1 L2 Power Supply Class 2 Fast, automated location of ground faults while the system remains online Reduced maintenance costs and downtime Indication of faulty circuit shown on LCD display at panel (EDS461), at remote indicating station (MK2430 / MK800), and at remote station through web browser-based GUI or Modbus/TCP (COM460IP) Available built into panel, modular design also allows for simple retrofitting and upgrading Current transformers for fault location built into panel as option, simple landing terminals provided for branch wiring MK2000 Series Complete system including LIM, fault location, and communication MK2430 Advanced, fast remote communication to hospital staff Notifcation of faults to nurse station Multiple isolated power panels may be monitored at a single remote station indicator (MK800 / 2430) with customizable messages Connecting Bender system to COM460IP allows for viewing status of isolated power system via simple web browser based GUI COM460IP also connects to Modbus/TCP networks to integrate into existing communication networks Ordering Information: Fault Location Equipment Fault location equipment can be ordered integrated into an isolated power panel, or separately for retrofitting purposes. Fault location modules Fault location for 12 branches EDS461-D-2 B 9108 0006 Ordering Information: Communication Gateways Communication gateways Ethernet and Modbus/TCP COM460IP B 9506 1010 Modbus/RTU COM462RTU B 9506 1022 Current transformers for fault location module 0.78" (20 mm) opening W20-8000 B 9808 0009 1.37" (35 mm) opening W35-8000 B 9808 0017 2.36" (60 mm) opening W60-8000 B 9808 0027 Refer to devices' respective datasheets for more information. Refer to Isolated Power Panels datasheet for information on integrating fault location into isolated power panels. 9

Isolated Power Equipment and Accessories Isolated power panels Standard isolated power panel Listed to UL 1047 (standard for isolated power systems equipment) Meets requirements for NFPA 99 / CSA Z32 for isolated power systems Single-phase, low-leakage isolation transformer, with primary and secondary voltages configured at factory and rated to system requirements Primary circuit breaker Up to 16 branch circuit breakers (maximum allowed by UL) LIM2010 line isolation monitor Reference ground bus Available with built-in receptacles and ground jacks Available with circuit lockout control Optional onboard, integrated ground fault location system Optional transformer load and temperature monitoring MK2000P-G1 remote indicator MK2000 series remote indicators Visual and audible indication of line isolation monitor (LIM) status Green SAFE LED and red HAZARD LED Audible alarm with mute button and amber LED LED display for long life Uses low voltage 12 V wiring Mounts to standard electrical gang box Easy-to-clean Lexan front foil Available with remote test button Available with transformer overload LED Available in digital version, duplicating THC digital display of the LIM2010 Advanced remote monitoring stations (MK2430 / MK800) MK2430 (left) / MK800 (right) Centralized remote station for Bender equipment Full-featured remote indicating station for operating status, warnings, and customized alarm messages Compatible with both LIM2010 line isolation monitor and EDS fault location systems Connects to Bender equipment via RS-485 Internal and external RS-485 network connections: Internal bus connects up to 150 Bender devices; up to 99 internal buses can connect to the external bus Large, backlit LCD display Up to 1000 customizable alarm messages Memory with real-time clock to store up to 1000 timestamped alarm messages Configurable with PC software Flush and surface mounted models 10

ZT1590 series digital clocks / timers Dual display for 12/24 hour clock and elapsed timer Elapsed time in minutes/seconds, automatically carried over to hours/ minutes All devices features and setup carried out either via onboard pushbuttons or connected MK1550 clock remote Plugable connectors for simple installation Utilizes external Class 2 power supply Integrated power outage backup for at least 24 hours, no batteries required ZT1590 (left) and MK1550 (right) GPM series ground/power module GPM series ground and power modules Hospital grade outlet devices for power supply and grounding of portable equipment Configurable to contain combination of hospital grade power receptacles, hospital grade ground jacks, and/or aluminum or copper ground buses Available with ground bus only to serve as a collection point for room grounding conductors Customizable quantity of hospital grade power receptacles and/or ground jacks Available on standard size wall plate or built into stainless steel front trim with backbox XRM series x-ray / laser receptacle modules Hospital grade outlet for supply of power to x-ray and laser equipment Includes NEMA rated plug matched to configuration of equipment Option for built-in MK2000 series remote indicator for line isolation monitor Door contact with limit switch Designed for use with Bender isolated power panels with PLC control Built on custom front trim with backbox Flush- or surface-mounted backbox Integrated LIM remote indicator In-use indicating light XRM series x-ray receptacle module 11

USA Exton, PA 800-356-4266 / 610-383-9200 Mexico and Central America Tampa, FL +1 813-240-2858 info@bender.org www.bender.org Canada Mississauga, ON 800-243-2438 / 905-602-9990 info@bender-ca.com www.bender-ca.com T M Your local contact: Document NAE2022011 / 04.2015 / Bender Inc. South America Santiago de Chile +56 2 2933 4211 info@bender-latinamerica.com www.bender-latinamerica.com