Polytron 2 The smart gas detector
2-951-95 Polytron 2 Highlights One universal transmitter for all electrochemical sensors Patented Plug & Play Sensor Technology Self-diagnostics of sensor and transmitter Polytron 2's patented self-test routines continuously monitor the function of the transmitter and the sensor. Any incident like a malfunction or just an expired calibration interval will be immediately identified and displayed and/or transmitted to the central controller. Polytron 2 is a universal transmitter for monitoring toxic gases and oxygen. More than 20 different electrochemical sensors capable of detecting over 200 different gases can be connected to the same transmitter. This makes Polytron 2 suitable for almost any gas detection application from semiconductor production to offshore oil & gas exploration. 4-20 ma outputs and digital HART communication CENELEC UL and CSA approvals Polytron 2 in combination with Dräger's high-tech electrochemical sensors is designed for highest performance at the lowest possible Cost of Ownership. Highest Performance For an optimum performance even in the roughest environmental conditions we have made use of state of the art microprocessor technology. A temperature sensor and a memory chip are integrated into every electrochemical sensor. The memory chip contains sensor specific information such as gas type(s) a lookup table for temperature compensation and the date of the last calibration.
Gases detectable with Polytron 2 (selection) Gas Formula Gas Formula Gas Formula Acetaldehyde CH 3 CHO Acetylene C 2 H 2 Acrylic acid C 2 H 3 COOH Ammonia NH 3 Antimony-V-chloride SbCl 5 Arsine AsH 3 Boron trichloride BCl 3 Boron trifluoride BF 3 Bromine Br 2 Butadiene-1.3 (C 2 H 3 ) 2 Butyl acrylate C 2 H 3 COOC 4 H 9 Butylamine sec. C 4 H 9 NH 2 Butylmercaptan tert. C 4 H 9 SH Carbon monoxide CO Chlorine Cl 2 Chlorine dioxide ClO 2 Chlorine trifluoride CIF 3 Diborane B 2 H 6 Dichlorosilane SiH 2 Cl 2 Diethylamine (C 2 H 5 ) 2 NH Diethylethanolamine (C 2 H 5 ) 2 NC 2 H 4 OH Dimethylamine (CH 3 ) 2 NH Dimethylethylamine (CH 3 ) 2 C 2 H 5 N Dimethyl sulfide (CH 3 ) 2 S Epichlorohydrin C 2 H 3 OCH 2 Cl Ethanol C 2 H 5 OH Ethylacrylate C 2 H 3 COOC 2 H 5 Ethylene C 2 H 4 Ethylene oxide C 2 H 4 O Ethylmercaptan C 2 H 5 SH Fluorine F 2 Formaldehyde HCHO Germanium hydride GeH 4 Germanium tetrafluoride GeF 4 Hydrogen H 2 Hydrogen bromide HBr Hydrogen chloride HCl Hydrogen cyanide HCN Hydrogen fluoride HF Hydrogen peroxide H 2 O 2 Hydrogen sulphide H 2 S i-propanol (CH 3 ) 2 CHOH Isopropylamine (CH 3 ) 2 CHNH 2 Isopropylmercaptan (CH 3 ) 2 CHSH Methanol CH 3 OH Methylmercaptan CH 3 SH Methyl methacrylate C 2 H 2 (CH 3 )COOCH 3 Monomethylamine CH 3 NH 2 Morpholine C 4 H 8 ONH Nitrogen dioxide NO 2 Nitrogen monoxide NO Nitrogen trifluoride NF 3 Oxygen O 2 Phosgene COCl 2 Phosphorus trichloride PCl 3 Phosphine PH 3 Phosphoryl chloride POCl 3 Propylene C 3 H 6 Propylene oxide C 3 H 6 O n-propylmercaptan C 3 H 7 SH Selenium hydride H 2 Se Silane SiH 4 Silicon tetrachloride SiCI 4 Silicon tetrafluoride SiF 4 Sulphur dioxide SO 2 Stannic tetrachloride SnCI 4 TEOS (C 2 H 5 O) 4 Si Tetrahydrothiophene C 4 H 8 S Thionylchloride SOCl 2 Titanium tetrachloride TiCl 4 Trichlorosilane SiHCl 3 Triethylamine (C 2 H 5 ) 3 N Trimethylamine (CH 3 ) 3 N Trimethylborane B(CH 3 ) 3 Tungsten hexafluoride WF 6 Vinyl acetate CH 3 COOC 2 H 3 Vinyl chloride C 2 H 3 CI For gases not listed above please contact us or our nearest agent.
Technical Data Signal transmission to controller Analog Digital 4 ma to 20 ma plus status information HART Explosion Protection Cenelec: Power supply 16.5 V DC to 30 V DC for -40 to 40 C: EEx ia IIC T6 for -40 to 65 C: EEx ia IIC T4 UL: Electrical connection Dimensions 2-wire (plus screen) 130 x 210 x 92 mm / 5.12 x 8.27 x 3.62 inch (Underwriters Laboratories) Class I; Div. 1; Groups A B C D Class II; Div. 2; Groups E F G for -40 to 40 C: Temp. Code T6 for -40 to 65 C: Temp. Code T4 CSA Weight Ambient conditions (without sensor) approx. 1.8 kg / 4.0 lb. -40 to 65 C / -40 to 150 F 700 to 1300 hpa / 20.7 to 38.4" Hg at 32 F 0 to 100 % R.H. non condensating Ingress Protection (Canadian Standards Association) Class I; Div. 1; Groups A B C D for -40 to 40 C: Temp. Code T6 for -40 to 65 C: Temp. Code T4 IP 67 according to EN 60 529 RFI/EMC CE marked electromagnetic compatibility (Directive 89/336/EEC) MTBF > 20 years Order Information Description Polytron 2 Polytron 2 Non Display Polytron 2 Hand-held Terminal Part No. 83 14 400 83 14 500 83 13 602 Gas and measuring range see individual sensor data sheets HART is a registered trademark of the HART Communication Foundation
For easy operation Polytron 2 provides display information in English German French or Spanish language. Sensor installation and replacement is a matter of seconds thanks to Dräger's innovative Plug & Play Sensor Technology. When a pre-calibrated sensor gets connected to a Polytron 2 transmitter the data in the sensor memory is downloaded and the transmitter configures itself automatically to the specific sensor - no potentiometers need to be adjusted. You will however enjoy the simplicity of this feature only every three to five years. That's how long Dräger's electrochemical sensors typically last. Networking Capabilities For easy networking Polytron 2 supports 4-20 ma and the HART fieldbus. HART allows 4-20 ma signal transmission and digital communication at the same time via the same single twisted pair connection. This makes it possible to transmit the time-critical measuring signal via 4-20 ma and perform all maintenance work simultaneously via bi-directional digital communication. Alternatively up to eight transmitters can be daisychained for operation on a single twisted pair connection in fully digital mode (up to seven transmitters on an intrinsically safe line). World-wide Approvals Authorities around the world such as the Underwriters Laboratories (UL) the Canadian Standards Association (CSA) or the BVS (CENELEC) have approved the installation of Polytron 2 in hazardous areas. FIELD COMMUNICATIONS PROTOCOL
Dräger Sicherheitstechnik GmbH Revalstrasse 1 D - 23560 Luebeck Germany Phone: + 49 451 8 82-27 94 Fax: + 49 451 8 82-49 91 Draeger Ltd. Ullswater Close Kitty Brewster Industrial Estate Blyth Northumberland NE24 4RG United Kingdom Phone: + 44 1670 352891 Fax: + 44 1670 356266 Draeger Industrie S.A. 3c Route de la Fédération F - 67025 Strasbourg Cedex France Phone: + 33 388 407676 Fax: + 33 388 407667 Draeger Safety Inc. P.O. Box 120 Pittsburgh PA 15230 U.S.A. Phone: + 1 412 787 8383 Fax: + 1 412 787 2207 DrägerService Dräger equipment must always be fully operational and safe. DrägerService offers a regular inspection service and is quickly available in the event of malfunction. Your equipment will be repaired and back in operation as quickly as possible. DrägerService facilitates the vital flow of information between client and manufacturer. Dräger Expertise Since 1889 Dräger has had an outstanding reputation for solving problems in the field of human breathing. Dräger has been deeply involved in the handling of gases in particular hazard protection and the saving of life in medical and industrial emergencies. Many of the company's 8500 employees are active in research and innovation to ensure that the latest techniques and scientific advances are fully tested before their inclusion in new equipment. Draeger South East Asia Pte Ltd. 67 Ayer Rajah Crescent #06-03 SGP-139950 Singapore Singapore Phone: + 65 872 9288 Fax: + 65 773 2033 Beijing Fortune Draeger Safety Equipment Co. Ltd. Jixiang Lu. B Area Beijing Tianzhu Airport Industrial Zone Houshayu Shunyi County Beijing 101300 P.R. China Phone: + 86 10 80498000 Fax: + 86 10 80498005 http://www.draeger.com/gds Dräger Quality The Dräger Sicherheitstechnik Quality Management System has been certified i.a. according to ISO 9001 and is recognized by numerous German and international organizations. Dräger Worldwide More than 25 subsidiaries and associated companies in over 100 countries ensure that Dräger is always within easy reach of its clients and in close contact with all important markets. Dräger's ever increasing market share demonstrates the company's international competitiveness and strength. Represented by: 90 44121 / KG 4676.000e/ 048 /RG-LP-ED / Market communications ST / Printed in Germany / chlorinefree-ecological / Subject to modifications