University of Pennsylvania. Environmental Health and Radiation Safety (EHRS) X-ray Diffraction User Guide

Similar documents
Appendix C X-Ray Diffraction User s Guide and Approval Form

Carnegie Mellon University Radiation Safety Committee

25 TEXAS ADMINISTRATIVE CODE Radiation Safety Requirements for Analytical and Other Industrial Radiation Machines

Analytical X-Ray Safety Program

The RSO will submit a copy of the revised manual to the RSC for approval when the revised manual has substantial changes.

California State University Bakersfield Radiation Safety Program Revised November 2017

Minnesota Rules, Chapter 4732 X-Ray Revision

RADIATION PRODUCING MACHINE SAFETY MANUAL

AUBURN UNIVERSITY X-RAY SAFETY STANDARD OPERATING PROCEDURES

Minnesota Rules, Chapter 4732 X-Ray Revision

X-Ray Protection Program

Safe Operating Procedure Analytical Radiation-Generating Equipment

Chapter 8 Radiation Producing Machines for Non-Medical Use

Radiation Emitting Devices Research X-ray Safety Program

OHIO UNIVERSITY RADIATION SAFETY REVIEW

PURPOSE OF THE PROCEDURE

University of Maryland Baltimore. Radiation Safety Procedure

Harvard University X-ray Safety Manual

Table of Contents for the REGULATION (EU) 2017/745 OF THE EUROPEAN PARLIAMENT AND OF THE COUNCIL of 5 April 2017

CHAPTER 3 RADIATION PRODUCING DEVICES I. AUTHORIZATION TO USE RADIATION PRODUCING DEVICES

Use of control and analytical X-ray apparatus

Approved by Principal Investigator: Date: Approved by Laser Safety Officer: Date:

X-Ray Safety Manual. Texas State University

PART V: WORKING WITH LASERS

Installation and Operation of Model CS X-ray Cabinet. 11 Emery Avenue Randolph, NJ Tel.:

RADIATION CONTROL GUIDE

Approved by Principal Investigator: Date: Approved by Laser Safety Officer: Date:

ALPHA Radiation Safety Manual

Laser Safety Citations

Stony Brook University Hospital Environmental Health & Safety Policy & Procedure Manual. EH&S 6-8 Original : 1/00 Revision Date: 3/16/15 Pages 6

Laser Safety. This Laser Safety Program applies to the acquisition, use, transfer, and disposal of all Class 3B and 4 lasers at UNC Asheville.

Emergency Evacuation Plan and Implementation Guidelines The Pennsylvania State University

ELEVATOR RESCUE PROCEDURE

Title: Laser Safety Effective Date: 9/91 Revision: 2/97 Number of Pages: 7

Laser Use at Subaru Telescope (NAOJ) Laser Safety Policy

Rutgers Environmental Health and Safety (REHS)

Radiation Protection Program Updates

North Dakota State University Laser Safety

THE REPUBLIC OF UGANDA ATOMIC ENERGY ACT, 2008 SCHEDULE 1 FORM 2A

AEC RADIATION SAFETY GUIDE SAFE USE OF INDUSTRIAl DEVICES CONTAINING RADIOACTIVE SOURCES

UC IRVINE LASER SAFETY PROGRAM

Emergency Evacuation Plan and Implementation Guidelines The Pennsylvania State University

Emergency Evacuation Plan and Implementation Guidelines The Pennsylvania State University BLUE BAND BUILDING ( )

LASER SAFETY. Class 2 and 2a - low power, low risk Class 2 and 2a lasers have low power and emit visible light. They will cause harm if viewed

Emergency Evacuation Plan and Implementation Guidelines The Pennsylvania State University

REGULATORY GUIDE 4.4 TEXAS DEPARTMENT OF STATE HEALTH SERVICES RADIATION SAFETY LICENSING BRANCH (RSLB) P.O. Box Austin, Texas

Radiation Safety Guidelines for Non-Users

HAZARDOUS MATERIALS/WASTE EMERGENCY CONTINGENCY PLAN University of Southern Maine Portland Campus

Emergency Response Plan

University of North Florida Laser Safety Manual. Environmental Health and Safety Department

Emergency Evacuation Plan and Implementation Guidelines The Pennsylvania State University ARTS COTTAGE ( )

Co-60 irradiator for therapy level calibrations at SSDLs

Radiation Safety Manual

Laser Safety Program Manual

ALABAMA. Downloaded

Safety Best Practices Manual

HAZARDOUS MATERIALS/WASTE EMERGENCY CONTINGENCY PLAN University of Southern Maine Portland Campus

REPUBLIC OF NAMIBIA MINISTRY OF HEALTH AND SOCIAL SERVICES REPORT BY: MS VERA UUSHONA TITLE: INSPECTION OF CONVENTIONAL DIAGNOSTIC X-RAY FACILITIES

Laser Safety Self-Audit Checklist

LASER SAFETY PROGRAM

S a n F r a n c i s c o F i r e D e p a r t m e n t B u r e a u o f F i r e P r e v e n t i o n & I n v e s t i g a t i o n 1 of 8 P a g e

A. General Safety LAB/ROOM: INSPECTED BY: DATE:

Corporate Standard Radiation Safety HB#

HEALTH AND SAFETY MANUAL

Laser Safety Program

SYNRAD Laser Safety 101

West Virginia University Laser Safety Program

F&S Control of Hazardous Energy Program

Laser Safety Orientation. Laser Safety Administration

Emergency Evacuation Plan. Kettering Labs. School of Engineering

Dosimetric verification of intensity modulated radiation therapy

Latest update: Ordinance of the Ministry of Internal Affairs and Communications No. 25 of March 27, 2013

DUQUESNE UNIVERSITY. Laser Safety Program

Required Materials For complete material(s) information, refer to

MANUFACTURING TECHNICAL INSTRUCTIONS - SAFETY REVIEW DATE: 10/29/2002

UNIVERSITY OF ROCHESTER ENVIRONMENTAL HEALTH & SAFETY

Emergency Evacuation Plan South Patterson Building (SPB)

HEALTH AND SAFETY PROGRAM 406 IONIZING/NONIONIZING RADIATION PROTECTION

SAN DIEGO COMMUNITY COLLEGE DISTRICT MIRAMAR COLLEGE ASSOCIATE DEGREE COURSE OUTLINE

THE UNIVERSITY OF TEXAS RIO GRANDE VALLEY LASER SAFETY MANUAL. August 2016

C Installation and Operation Manual

FSSG Training Guide for NYC DOB Elevator Inspectors

A. Administrative. B. Labeling and Posting. C. Control Measures. Laser Safety Self-Audit Checklist. Building Room Principal Investigator Date

FIRE EMERGENCY ACTION PLAN (EAP)

OPERATING MANUAL TORREY PINES SCIENTIFIC, INC. DIGITAL HPLC COLUMN HEATER MODEL CO20 DOCUMENT NUMBER CO20-00

AGENDA UNIVERSITY OF NEW HAMPSHIRE RADIATION SAFETY COMMITTEE 4th Quarter Meeting 2016 December 8, :30-2:30 pm Rudman G04

OPERATING MANUAL. EchoTherm PROGRAMMABLE DIGITAL CHILLING INCUBATOR MODELS IN35, IN45 and IN55-12VDC. DOCUMENT NUMBER IN35-03 Revised 15 May 2015

General PSS principles an introductory guide

USER'S MANUAL PGE Single Package Rooftop

REGISTRY OF RADIOACTIVE SEALED SOURCES AND DEVICES SAFETY EVALUATION OF DEVICE (Amends In Its Entirety)

Real-Time Remote Monitoring System Utilizing New Electronic Personal Dosimeter

Radiation Training for Ancillary Personnel. Introduction

OPERATING MANUAL MOLECULAR DIMENSIONS, INC. PROGRAMMABLE DIGITAL CHILLING INCUBATOR MODELS MD5-601 AND MD5-603

MAINTENANCE AND SERVICE OF AUTOMATIC FIRE EXTINGUISHING SYSTEMS

Title: Electrical Safety Effective Date: 11/28/2014 Control Number: THG_0038 Revision Number: 1 Date: 10/22/2015 Annual Review Completed: 5/13/2015

Chapter 18 Welding Table of Contents

Minor Change Office of Radiological Safety Number: Revision 00 By: Georgia Institute of Technology. Approved 04/21/2016 Date: / / Page 1 of 6

MINI-RADIAC INTRODUCTION

Laser Safety Part III Administration

MANUAL OF PROCEDURE. Keys and Access Control (Standard and Electronic)

Transcription:

University of Pennsylvania Environmental Health and Radiation Safety (EHRS) X-ray Diffraction User Guide (Reviewed: September 2012) I. INTRODUCTION II. III. IV. PA DEPARTMENT OF ENVIRONENTAL PROTECTION ENVIRONMENTAL HEALTH AND RADIATION SAFETY (EHRS) X-RAY DIFFRACTION UNITS A. UNIT REGISTRATION B. UNIT ACQUISITION C. UNIT RELOCATION, DISPOSAL, OR TRANSFER V. WORKER RESPONSIBILITIES A. WORKER TRAINING VI. VII. VIII. IX. RADIATION SURVEY SAFETY DEVICES AND SIGNS OPERATING REQUIREMENTS OVERRIDING SAFETY DEVICES X. PERSONNEL EXPOSURE MONITORING XI. XII. EMERGENCY PROCEDURES IMPORTANT PHONE NUMBERS

I. INTRODUCTION Analytical X-ray machines are defined as an assembly of components that utilize x-rays to determine elemental or chemical composition, or to examine the microstructure of a material through the use of x- ray diffraction or fluorescence. II. PA DEPARTMENT OF ENVIRONENTAL PROTECTION Use of x-ray equipment in Pennsylvania is regulated by the PA Department of Environmental Protection (DEP). The DEP has established regulations which must be followed by all individuals using energized (xray) equipment. These regulations are found in Title 25 Chapters 219, 220, and 227 of the Pennsylvania Code and are available for review in the Environmental Health and Radiation Safety (EHRS) office, or on the DEP website at http://www.pacode.com/index.html. III. ENVIRONMENTAL HEALTH AND RADIATION SAFETY (EHRS) EHRS is responsible for ensuring that radiation and radioactive material is used safely at Penn and associated facilities (e.g., Wistar Institute, CHOP). EHRS staff performs a safety check of each diffraction x-ray unit at installation and annually (at a minimum). If service is provided on the unit that would effect the arrangement, number, or type of components in the system contact EHRS for an additional safety check prior to use. [PA 227.12a] IV. X-RAY DIFFRACTION UNITS A. Unit Registration All analytical energized equipment must be registered with EHRS prior to use. This can be done by contacting EHRS, and must be done at the time of equipment installation. [PA 219.131-2] B. Unit Acquisition It is the responsibility of the clinical personnel to notify EHRS upon acquisition of any new x-ray equipment. Authorized EHRS personnel will conduct a radiation safety survey on all new units prior to use. C. Unit Relocation, Disposal, or Transfer EHRS must be notified prior to relocation, disposal, or transfer of ownership of cabinet x-ray equipment. Relocation includes moving equipment to a different room within the same building. EHRS will ensure that proper notifications to State Agencies are made. [PA 219.131-2]

V. WORKER RESPONSIBILITIES A. Worker Training [PA 227.14] 1. Radiation Safety Training All persons using x-ray diffraction equipment must attend the radiation safety training specific to this type of equipment. This training is provided on the EHRS website at http://www.ehrs.upenn.edu/programs/radiation/radiationsafetytraining/ 2. Operational Training Additionally, all the persons operating the equipment must receive the following instructions from the primary researcher and demonstrate competence in the following areas: a. Significance of the various radiation warning and safety devices incorporated into the equipment, or the reason they have not been installed on certain pieces of equipment, and the extra precautions necessary if the devices are absent or bypassed. b. Operating and emergency procedures. VI. RADIATION SURVEY All x-ray diffraction equipment must be surveyed at the following times [PA227.11a]: 1. Upon installation; 2. Annually; 3. Whenever the following occurs: a. When there is a change in the initial arrangement, number, or type of local components in the analytical x-ray system. b. Following maintenance requiring the disassembly or removal of a local component. c. During the performance of maintenance and alignment procedures if the procedures require the presence of a primary x-ray beam when a local component in the system is disassembled or removed. d. When a visual inspection of the local component in the system reveals an abnormal condition. e. When the machine is operated in a manner other than the routine manner specified in the written operating procedures. VII. SAFETY DEVICES AND SIGNS 1. On equipment with an open beam configuration manufactured and installed after December 19, 1987, each port on the radiation source housing must be equipped with a shutter that cannot be opened unless a collimator or coupling has been connected to the port. [PA 227.11a.(c)] Note: An open beam configuration system is an analytical x-ray system in which the beam is not enclosed or shielded so any portion of an individual's body, including fingers, could accidentally be placed in the beam path during normal operation.[pa 227.2] 2. Unused ports on radiation housings shall be secured in the closed position in a manner which will prevent accidental opening. [PA 227.11a.(e)]

3. An open-beam configuration unit must have a device which either prevents the entry of any portion of body into primary beam path or causes the beam to be terminated when a part of the body approaches the beam. [PA 227.11a.(a)] 4. A warning label must be posted on the x-ray source housing which states, "Caution - High Intensity X-ray Beam", and another label which states, "Caution Radiation - This Equipment Produces Radiation When Energized", near any switch that energizes the X-ray tube.[pa 227.11a.(f)] Note: These labels are available from EHRS. 5. An easily visible warning light which does the following [PA 227.11a.(d)]: a. Illuminates when the X-ray tube is energized and labeled with the words "X-Ray ON." b. Illuminates in each port on radiation source housing when the shutter is open (this applies to open beam configurations). VIII. OPERATING REQUIREMENTS Operating procedures must be written and available to the analytical x-ray equipment operators. Procedures must include instructions for the following [PA 227.13a]: 1. Sample insertion and manipulation 2. Equipment alignment 3. Routine maintenance and data recording procedures An individual may not operate analytical x-ray equipment in a manner other than that specified in the operating procedures unless that individual has obtained written approval from the EHRS Radiation Safety Officer (RSO). Except when written approval is given by the RSO to override safety devices, operations involving removal of covers, shielding materials or tube housings, or modifications to shutters, collimators or beam stops may not be performed without ascertaining that the tube is off and will remain off until safe conditions have been restored. Interlocks may not be used for routine shutdown in preparation for repairs (i.e. use the main shutdown switch). IX. OVERRIDING SAFETY DEVICES An individual may not bypass or otherwise circumvent a safety device unless that person has received prior written approval from the RSO. An approval request form can be found on the EHRS website at http://www.ehrs.upenn.edu/programs/radiation/equipment/. [PA 227.13a(b)] In order to receive written approval, the following will be needed: 1. Whole-body and extremity dosimeters. To request dosimeters, use the same EHRS webpage listed above. 2. Calibrated survey meter that measures milliroentgen per hour (mr/hr).

3. Controls and procedures to assure the safety of individuals during the override. 4. A readily discernible sign bearing the words, "SAFETY DEVICE NOT WORKING" to be placed on the source housing. X. PERSONNEL EXPOSURE MONITORING Exposure to scattered radiation from analytical x-ray equipment is extremely low. The radiation levels are measured by EHRS, and must not exceed 0.5 mr/hr at 5 centimeters from the surface of the unit for it to be routinely operable. Therefore personnel dosimetry is not required during routine use. [PA 227.32] However, personnel dosimetry is required for persons performing maintenance on x-ray diffraction units when a local component in the system is disassembled or removed, or when safety devices are disabled. Contact EHRS for dosimeters prior to such maintenance. XI. EMERGENCY PROCEDURES If anyone thinks they may have been exposed to the x-ray beam, please contact EHRS immediately. EHRS is required to report suspected overexposures to the DEP within 5 days of the incident, so immediate action is required by anyone suspecting that they have been exposed to the x-ray beam. [PA 227.14(2)(d)] XII. IMPORTANT PHONE NUMBERS University of Pennsylvania, Environmental Health and Radiation Safety: Monday Friday during business hours (215) 898-7187 On-Call Physicist for assistance after hours (215) 573-6626