Nellcor Bedside SpO 2 Patient Monitoring System, PM100N

Similar documents
SOFTWARE VERSION 8.5 FAQs

Nellcor Puritan Bennett Inc. is an affiliate of Tyco Healthcare. Nellcor, Oxiband, Durasensor, OxiCliq, Dura-Y, MAX-FAST, and OXIMAX are trademarks

N-600. Pulse Oximeter Operator s Manual

Capnostream 20p Bedside Capnography Monitor. The only integrated, complete picture of oxygenation and ventilation 1

Model 900. Pulse Oximeter User s Manual

N-600. Pulse Oximeter Home Use Guide

OPERATOR S MANUAL. Patient Monitor M20. EU representative TECNOMED 2000 S.L. Valencia, Madrid Spain

Nellcor with OxiMax Technology Product Ordering Guide A Full Line of Oximetry Sensors and Monitors

N-595. Pulse Oximeter Service Manual

OxiMax N-600x Pulse Oximeter with Alarm Management System

Connex Integrated Wall System 2.2 Software Version

Quick start guide. English

Passport 12m and Passport 17m BEDSIDE MONITOR GUIDELINE 5.21 Software Version NAME: HOSPITAL: DATE: VALIDATOR: Performed

P/N: Release Date: January 2012 Copyright EDAN INSTRUMENTS, INC All rights reserved.

OPERATOR S MANUAL. NPB-4000/4000C Patient Monitor

BeneVision Central Monitoring System. Quick Reference Guide

HUG.ai 1/24/06 9:33:34 AM HUG.ai 1/24/06 9:33:34 AM C M Y CM MY CY CMY K

T1 BEDSIDE MONITOR GUIDELINE

User Guide. Cardell Insight Veterinary Monitor. For Models: 8014 Blood Pressure Blood Pressure, Pulse Oximetry

HBP Patient Monitor. Operator s Manual. EU representative OBELIS S.A Bd. Général Wahis, Brussels, Belgium.

Puritan Bennett 980 Ventilator System Planning for Training Checklist

CLEO VERSATILITY IN VITAL SIGNS

Quick setup guide. English

Propaq LT Outline Train the Trainer

OMNI Express VS VERSATILITY IN VITAL SIGNS

Track the truth. Give no false assurance. Do no harm. Nellcor SpO 2. Module for use with Philips IntelliVue * Patient Monitors

SPECIFICATION FOR MULTIPARAMETER MONITOR

Operator s Manual. Avant Digital Pulse Oximeter. English

About this Manual. Statement. P/N: Release Date: August 2010 Copyright EDAN INSTRUMENTS, INC All rights reserved.

Cl inician Training Packet

Patient Monitoring System YM5500. Service Manual

General Skills. yes. Device controls, buttons, indicators, and connectors

PATIENT MONITORING SYSTEMS

Veterinary Portable Multi-parameter Patient Monitor. Operation Manual

signal extraction pulse oximeter OPERATOR S MANUAL

Guide, Quick Start, Short, X-100 System

Adult and paediatric ventilators. Quick Setup Guide. English

GE Healthcare. Quick Guide. Dash 2500 Patient Monitor

OMNI TOUCH SCREEN PATIENT MONITOR

Vital Signs Monitor. Operator s Manual

SureSigns VS4. Vital Signs Monitor Technical Data Sheet

740 SELECT. Multi-Parameter Monitor

qube mini Overview

SERVICE MANUAL. NPB-295 Pulse Oximeter

Operator s Manual. Avant Digital Pulse Oximeter. English

CNAP MONITOR 500 HD QUICK GUIDE. This Quick Guide DOES NOT REPLACE the user manual!

Clinical Information Center Pro v5 (CIC Pro)

SureSigns VS4 Vital Signs Monitor

SECTION L MONITORING EQUIPMENT

PULSE OXIMETER PACKAGE

TidalGuard HD. Handheld CO 2 / SpO 2 and CO 2 monitors. Operating Manual Rev. H (8/15)

Vital Signs Monitor 300 Series

Atlas Monitor Outline Train the Trainer

Omni (K) PATIENT MONITOR

Infinity M300 Patient Worn Monitoring

PM-9000 Express Portable Patient Monitor

MODEL 3X Series User s Manual

User Manual. December 2006 M Rev.A

with color display signal extraction pulse co-oximeter with Rainbow Technology OPERATOR S MANUAL

SERVICE MANUAL YM6000 Patient Monitor

HANDHELD PULSE OXIMETER

DPM 4 Patient Monitor. Operator s Manual

CLINICAL MANUAL REF JH 6/7/10 CLINICAL MANUAL

V6 Vital Signs Monitor User s Manual

Dryer Controller M720

Accutorr 7. Vital Signs Monitor. Operator s Manual

Servo i. CRC 335 University of South Alabama Cardiorespiratory Care

Operator s Manual. LifeSense LS1-9R. Vital Signs Monitor. Capnography/Pulse Oximeter

User Manual. Dryer Controller M720

ARGUS VCM Vital Compact Monitor

SERVICE MANUAL. Patient Monitor M30. EU representative TECNOMED 2000 S.L. Valencia, Madrid Spain

Instructions for Use Vista 120

Transport Pro Patient Monitor Operator s Manual Software Version 2.1

901-M / 901-ML. Pulse Oximetry System. Technical Information and Service Manual

Ultraview SL2400 With Wireless-LAN-Option

Multiparameter. er Monitor DM

TruSat Pulse Oximeter

DINAMAP ProCare Monitor. Operation Manual

5/ 5i/ 5c/ 7/ 7i/ 7c/ 5 elite/ 5i elite/ 5c elite/ 7 elite/ 7i elite/ 7c elite

Follett Performance Plus

Vital Signs Monitor 300 Series

Welch Allyn Connex Devices

DPM4 Bedside Monitor Default Settings

QAM.L L550 series. Patient Monitor Manual QAM.L

OctiveTech 300AH HANDHELD OXIMETER OPERATING INSTRUCTIONS

TPCA TM PATIENT CONTROLLED ANALGESIA SYRINGE PUMP

signal extraction pulse oximeters OPERATOR S MANUAL

IQvitals Zone TM. For Models: User Guide

End-Tidal CO2 User Guide CAPNO3 T NIBP. SpO2. Ver Charmcare Co., Ltd.

D Moving beyond telemetry INFINITY M300

(Place Logo from Lab here)

Ultraview SL Command Module Parameter Set ECG NIBP TEMP IP CO SpO2 Options SL Command Module: A B Our goals are simple:

Tissue Oximeter. Operator s Manual

Welch Allyn Connex Devices

TORONTEK-B400 Pulse Oximeter. User Instructions

Syringe Pump. User Manual

The information in this document is subject to change without notice.

Model 7500 Model 7500

User s Guide. Cardell Touch Veterinary Vital Signs Monitor. For Models:

Welch Allyn Vital Signs Monitor 6000 Series. Directions for use

Transcription:

Nellcor Bedside SpO 2 Patient Monitoring System, PM100N In-Service Presentation Nellcor Bedside SpO 2 Patient Monitoring System is CE marked and commercially av ailable in European Union countries. Nellcor Bedside SpO 2 Patient Monitoring System not available in all markets. Please ref er to local product labelling. COVIDIEN, COVIDIEN with logo, Covidien logo and positive results for life are U.S. and internationally registered trademarks of Covidien AG. Other brands are trademarks of a Covidien company. 2014 Covidien. EU-14-0860-2

The Nellcor Bedside SpO 2 Patient Monitoring System, PM100N, is intended for prescription use only for continuous monitoring of functional arterial oxygen saturation and pulse rate of adult, pediatric and neonatal patients, who are well or poorly perfused, during motion and no motion conditions. It is intended to be used in hospital, hospital type facilities, and intra-hospital transport as well as in the homecare environments. 2

New Homecare and Sleep Study Modes Nellcor Bedside SpO 2 Patient Monitoring System, PM100N is the only bedside monitor with standard, homecare, and sleep study modes. Standard Mode: Advanced functionality to monitor inhospital patients with SpO 2, pulse rate, and SatSeconds alarm management. Homecare Mode: Password protected, simplified monitoring for non-hospital settings. Prevents lay users from modifying alarm settings. Sleep Study Mode: Password protected, simplified monitoring for sleep studies. Reduces screen backlighting and eliminates alarms to reduce disturbances to sleeping patients. 3

Rear Panel 1. Nurse Call Port 2. AC Power Connector 3. Battery Cover 4

Front & Side Panel Audio Pause Silence audible alarms for set time Home Button Return to this main screen Power Button Press and HOLD to turn on and off USB Ports For data download and upgrades SpO 2 Connector Insert interface cable here Knob Rotate and press, to control system functions 5

Display Right Side AC Power Indicator Battery Charge Indicator Interference Indicator Sensor Disconnect Indicator Sensor Off Indicator Sensor Message Indicator 6

Hands on Review Attach Power Cord to device and plug into AC receptacle Power on device and determine when it is ready for use Connect cable and sensor to device Create the various sensor indicator messages Interference indicator Sensor disconnect indicator Sensor off indicator Sensor message indicator (articulate when this occurs) Power off the device 7

Standard Monitoring Display Main View SpO 2 : Upper/Lower Alarm Limits SpO 2 : Real Time Value Pulse Rate: Real Time Value Battery Indicator Plethysmographic Waveform (Pleth) Blip Bar Informative Message Area Pulse Rate: Upper/Lower Alarm Limits 8

Standard Monitoring Display Bottom Informative Message Area Patient Mode Alarm Limits Menu Options Menu 9

Standard Monitoring Alarm Limits Menu To access the Alarm Limits Menu, rotate the knob to highlight the alarm limits display area and press. Alarm Paused Icon Adjust Up/Down SpO 2, PR, SatSeconds Alarm Icon Status Alarm Active Alarm Audio Paused Alarm audio deactivated in Service Menu Alarm Audio icon in Sleep Study Mode 10

Standard Monitoring Quick Access Menus For quick access to alarm settings, rotate the knob to highlight one of two areas and press. SpO 2 Alarm Settings SatSeconds Settings Pulse Rate Alarm Settings 11

Hands on Review Attach the sensor to a patient Determine the patient s real time SpO 2 and PR reading Set parameter alarm limits SpO 2 target range: 85% to 100% Pulse rate target range: 80 bpm to 110 bpm Once alarm limits are set Discuss alarm status What do you see? What do you hear? How would you respond to the alarm(s)? 12

Standard Monitoring Patient Mode Menu To access the Patient Mode Menu, turn the knob to highlight the patient mode display area and press the knob. Patient Mode Menu 13

Standard Monitoring Options Menu To access the Options Menu, turn the knob to highlight the Options Icon and press the knob. Press as a shortcut to access the Options Menu Volume Adjustment Alarm Key Beep Pulse Patient Response Mode Normal Fast Service Menu 1. Power On Settings 2. Alarm Audio Paused 3. Alarm Audio Reminder 4. Permission to Mute Alarm 5. Language 6. Date/Time Setting 7. System Information 8. System Test 9. Trend Data Download Settings Clear Trend Data 10. Homecare Mode Settings 11. Communication settings 12. Alarm Priorities 13. Password Settings 14

Hands on Review Your next patient is a 1 year old child. Set the pulse oximeter to the appropriate patient mode. It is very loud in the patient s room. Set the alarm volume to the loudest setting for SpO 2 and PR. 15

Standard Monitoring SpO 2 Waveform Menu To access the SpO 2 Waveform Menu, turn the knob to highlight the waveform or trend display area and press the knob. Sweep Speed 6.25mm/s 12.5mm/s 25mm/s Trend Data 88,000 data events 96 hours Collected every 4 seconds 16

Standard Monitoring Sweep Speed Overview Allows the clinician to set the speed to where the SpO 2 waveform trace moves across the screen. Sweep Speed = 25.0mm/s (i.e. Zoom In ) The higher the sweep speed value, the more granular the data appears on the screen. Sweep Speed = 6.25mm/s (i.e. Zoom Out ) 17

Standard Monitoring Tabular Trend View To access the Tabular Trend View, turn the knob to highlight the waveform or trend display area and press the knob. Select Tabular Trend and press the knob. Date / Time Scroll Interval SpO 2 Value Pulse Rate Value Monitor errors 18

Standard Monitoring Tabular Trend Scroll To scroll through the Tabular Trend View, rotate the knob to the left for historical data and to the right to get back to present time Scroll Interval Options 1 (every 4 sec) 5 (every 20 sec) 100 (~6.5 min) 500 (~33.5 min) Turn knob to the left for historic data Note: To change Scroll highlight Interval Options, press the knob. 19

Standard Monitoring Trend Data Download To access the Trend Data Download Menu, turn the knob to highlight the Options Menu and press. Mini-USB Port 1. Connect mini-usb port to a PC 2. Select Options 3. Select Trend Data Download 4. Select Start 20

Standard Monitoring Graphical Trend View To access the Graphical Trend View, turn the knob to highlight the waveform or trend display area and press the knob. Select Graphical Trend and press the knob. SpO 2 Trend Line (blue line) SpO 2 Scale Pulse Rate Scale Pulse Rate Trend Line (green line) 21

Hands on Review Access tabular trend data for your patient How have your patient s SpO 2 values changed in the last hour? How have your patient s PR values changed in the last hour? Use the scroll interval options to review your patient s data from 3 hours ago. Download your patient s data using the data download capability. How would you use graphical trend? 22

Standard Monitoring SatSeconds Alarm Management Setting SatSeconds Setting SatSeconds Icon 23

Standard Monitoring Alarm Limits Menu To access SatSeconds Alarm Limits Menu, turn the knob to highlight the alarm limits display area and press the knob. SatSeconds Value Can be adjusted to off, 10, 25, 50, 100 24

Hands on Review Activate SatSeconds Alarm Management If you want to minimize clinically insignificant alarms that occur with your patient, set SatSeconds to 100. How would you respond to a SatSeconds Alarm? 25

Homecare Mode To access Homecare Mode, turn the knob to highlight and select the Options Menu. Within menu, select the Change Patient Mode display area and press the knob. Turn knob to highlight Change Patient Mode and press the knob 26

Homecare Mode After selecting Change Patient Mode display area, highlight Switch to Homecare and press the knob. Turn the knob to highlight Switch to Homecare and the press knob. 27

Homecare Mode After selecting Switch to Homecare display area, enter passcode. Entering Passcode: Turn the knob to highlight the number in passcode and press the knob. Once number is highlighted in yellow, turn knob to change number then press the knob to select. Once number is selected, repeat process for remaining passcode numbers. 28

Homecare Mode After passcode is entered, a confirmation screen will pop up. Select OK and then you will be in Homecare Mode. Select OK and press the knob Note: The Homecare mode monitoring screen does not appear until the user presses the knob to select OK 29

Sleep Study Mode To access Sleep Study Mode, turn the knob to highlight the Options Menu and select the Change Patient Mode display area and press the knob. Turn the knob to highlight Change Patient Mode and press the knob 30

Sleep Study Mode After selecting the Change Patient Mode, then highlight Switch to Sleep Study and press the knob. Sleep Study Mode 31