Dental Unit User's Guide MDD 93/42/EEC May 1999 8200590 Quint-7000 User's Guide We reserve the right to make changes. Manufactured by: FINNDENT OY Teollisuustie 5 FIN-07230 MONNINKYLÄ Tel. +358 19 5221500 Fax +358 19 643457 I
II
Contents 1. Introduction... 1 Scope of this manual...1 Associated documents...1 Warnings and precautions...2 2. The main parts... 3 Quint 7000 A1 Dental Unit...3 Quint 7000 A2 Dental Unit...4 Quint 7000 E2 Dental Unit...4 Dentist's Control Panel...5 Assistant's Control Panel...5 The foot control...6 3. Using the dental unit... 7 Switching the dental unit on...7 Switching the dental unit off...7 Rotating the cuspidor...8 Positioning the instrument bridge...8 Positioning the instrument cart...9 Using the instruments...10 The Micromotor...10 The Turbine...12 The Scaler...13 The Polymerisation light (Satelec)...15 The Polymerisation light (Lysta)...15 The Syringe - 3-way...16 The Syringe - 6-way...16 Dentist's control panel...17 Water bottle...18 Tray holder...18 The Suction Arm...18 Positioning the suction arm...19 Using the suction nozzles...19 The Syringe...20 Suction arm controls...21 The Dental Light...22 III
4. Cleaning, disinfecting and sterilising the unit... 23 Suction system...23 Suction hoses...23 Instruments and suction nozzles...24 Unit surfaces...24 5. Maintenance... 27 Oil collector...27 Tightening instrument holders...27 Instrument water adjustment...28 6. Service... 29 7. Approved accessories... 30 8. Technical Specifications... 32 Symbols and abbreviations...33 Dimensions...34 IV
1. Introduction Scope of this manual This manual describes how to use and look after all dental units in the FD 7000 range. The range includes the following versions: - 7000A1 Chair mounted cuspidor, whip arm instruments - 7000A2 Chair mounted cuspidor, hanging tube instruments - 7000B1 Floor mounted cuspidor, whip arm instruments - 7000B2 Floor mounted cuspidor, hanging tube instruments - 7000C2 No cuspidor, instrument cart with hanging tube instruments - 7000D2 Chair mounted cuspidor, instrument cart with hanging tube instruments - 7000E2 Floor mounted cuspidor, instrument cart with hanging tube instruments Operating differences between the different versions of the unit are described where appropriate. Quint 7000 Dental Unit 1
Associated documents User's manual supplied with the amalgam separating system that is installed in the dental unit. User's manuals for the instruments that are supplied with the dental unit. Warnings and precautions Read this user's manual and the associated documentation before using the dental unit or any of its instruments or accessories. Make sure that you know how to use the dental unit before starting to use it. Do not let children or unauthorized people play with or use the unit or any of its instruments or accessories. Only use instruments and accessories listed in this manual with the unit. Only carry out the maintenance procedures listed in this manual. All other maintenance MUST be carried out by authorized service personnel. For the dental unit to function efficiently it must be regularly serviced. If the quality of the water being used with the unit is poor, an additional water filter may have to be installed. (Refer to the installation manual) The compressed air being used with the unit must be dry and oil free. The air compressor must be fitted with an air dryer and/or condensation valve. (Refer to the installation manual) 2 User's Manual
2. The main parts Quint 7000 A1 Dental Unit Chair mounted cuspidor with whip arm instruments Dental light Instrument arm Instrument bridge and dentist's control panel Instrument tray Water bottle (Option) Suction arm and assistant's control panel Foot control Floor box Cuspidor NOTE With the dental units Quint 7000 BI, 7000 B2 and 7000 E2 the cuspidor is attached to the floor box via a metal tube. Quint 7000 Dental Unit 3
Quint 7000 A2 Dental Unit Chair mounted cuspidor with hanging tube instruments Hanging tube instruments Quint 7000 E2 Dental Unit Floor mounted cuspidor with instrument cart Instrument cart with hanging tube instruments Metal tube 4 User's Manual
Dentist's Control Panel Reverse rotation indicator Scaler power indicators Micromotor spray indicator Turbine spray indicator Accessory key Scaler power key Door lock/nurse call key Turbine spray key Glass fill key Bowl rinse key Assistant's Control Panel Glass fill key Bowl rinse key Door lock/nurse call key Quint 7000 Dental Unit 5
The foot control Option switch (For example - nurse call) Lifting handle Micromotor spray switch Reverse direction switch Foot lever Instrument on/off switch Instrument speed/power control Chip blow switch 6 User's Manual
3. Using the dental unit Switching the dental unit on Switch the dental unit on with the ON/OFF switch that is located on the side of the floor box. When you switch the dental unit on you will hear an audible signal and the switch light will come on. On Off Note If the dental chair is connected to the power supply via the dental unit (i.e. the dental chair does not have its own power supply), the dental chair will automatically come on when the unit is switched on. Note that the chair's ON/OFF switch must be in the ON position for this to happen. Switching the dental unit off Switch the dental unit off with the same switch. When the unit is switched off the switch light will go out. ALWAYS switch the unit off at the end of each work day and at weekends. Quint 7000 Dental Unit 7
Rotating the cuspidor The cuspidor can be rotated approximately 80 away from the chair. This may be necessary if the assistant needs additional working space when special dental procedures are being carried out. Cuspidor Positioning the instrument bridge Move the instrument bridge in the required position with the handles on the sides of the bridge. The height of the instrument bridge can be adjusted by lifting it up or pushing it down. If you wish to lock the instrument bridge at a particular height turn the height locking knob clockwise. To unlock the bridge turn the knob anticlockwise. Height locking knob Handle 8 User's Manual
If the instrument bridge "drifts" out of position, the friction of the arm joints needs to be increased. Contact your service technician for assistance. Positioning the instrument cart Move the instrument cart in the required position with the handles on the side of the instrument bridge. Handle Instrument cart Quint 7000 Dental Unit 9
Using the instruments When an instrument is lifted from its holder it becomes the "active" instrument and no other instrument, except the syringe and polymerisation light, will operate while the active instrument is in use. The Micromotor Switching it on and off Lift the micromotor from its holder and press and hold down the foot lever. The micromotor will start. To stop the micromotor, remove your foot from the foot lever. Adjusting the speed Slide the foot lever to the right to increase the speed of the micrometer. Slide the foot lever to the left to decrease the speed. 10 User's Manual
Activating the spray Press the right-hand foot switch once to switch the instrument spray on. When the spray is on the green M light on the dentist's display will come on. Press the right-hand foot switch a second time to switch the spray off. The green light will go out. Chip blower Double press and hold down the foot lever. Chip blow will start. To stop chip blow remove your foot from the foot switch. Reversing the rotation Press the left-hand foot switch down once to reverse the direction of rotation of the micromotor (anticlockwise). The red M light on the dentist's display will come on when the micromotor is set to rotate in the reverse direction. Press the foot switch a second time to switch the direction of rotation of the micromotor back to the normal (clockwise) direction. The red M light will go out. Quint 7000 Dental Unit 11
The Turbine Switching it on and off Lift the turbine from its holder and press and hold down the foot lever. The T light will come on and the turbine and spray will start. To stop the turbine remove your foot from the foot lever. The T light will go out. Note that there is no speed control with the turbine. Chip blower Double press and hold down the foot lever. Chip blow will start. To stop chip blow remove your foot from the foot switch. 12 User's Manual
The Scaler Switching it on and off Lift the scaler from its holder and press and hold down the foot lever. The scaler will start. To stop the scaler, remove your foot from the foot lever. Selecting the power level The maximum power level of the scaler can be set at either 100%, 50% or 30%. The lights next to the SP key indicate what power level has been selected. To change the power levels press the SP key repeatedly until the light combination for the power level you require appears. The power level/ light combinations are as follows: - 100% power = no lights on - 50% power = top light on - 30% power = bottom light on No lights Top light Bottom light 100% 50% 30% Note that the scaler power level will automatically return to the 100% level, i.e. no lights on, whenever the unit is switched off and then on again. Quint 7000 Dental Unit 13
Adjusting the power level To increase the power of the scaler slide the foot lever to the right to decrease the power slide the foot lever to the left. Note that when the foot control is moved to the maximum power position, the maximum power level will be the level selected using the SP key. For example, if you selected the 50% power level the power level can adjusted from 0% to a maximum of 50%. 14 User's Manual
The Polymerisation light (Satelec) Lift the polymerisation light from its holder and press the button on the side of the instrument. The light will come on. The light will automatically go out after 30 seconds. During the time the light is on you will hear a "beep" every ten seconds. If you wish to turn the light out before the 30 seconds elapses, press the button a second time. For more information on how to operate and service the Satelec polymerisation light refer to the documentation supplied with the light. Activation switch The Polymerisation light (Lysta) Lift the polymerisation light from its holder and press the button on the top of the instrument. The light will come on. The light will automatically go out after 35 seconds. After seventeen seconds you will hear a beep. If you wish to turn the light out before the 35 seconds elapses, press the button a second time. For more information on how to operate and service the Lysta polymerisation light refer to the documentation supplied with the light. Quint 7000 Dental Unit 15
The Syringe - 3-way Spray Press the left-hand button to activate the water spray. Press the right-hand button to activate the air. Press both buttons together to activate the air/ water spray. Water Air The Syringe - 6-way Spray Water Air There is a heater in this syringe that allows you to have either hot air and water or cold. To switch the heater on twist the collar at the end of the handle anticlockwise - the indicator spot will be green. To switch the heater off twist the collar at the end of the handle clockwise - the indicator spot will be black. Press the left-hand button to activate the water spray. Press the right-hand button to activate the air. Press both buttons together to activate the air/ water spray. Heater off Heater on 16 User's Manual
Dentist's control panel Bowl rinse - on/off Briefly press the bowl rinse key on the dentist's control panel to switch the bowl rinse on. Rinsing will automatically stop after the preset rinse duration has elapsed. If you wish to stop the bowl rinse before the preset rinse time has elapsed, press the bowl rinse key a second time. Bowl rinse - setting the rinse duration Press and hold down the bowl rinse key for the amount of time that you wish the bowl rinse to operate (minimum one second, maximum ten seconds). Now, when the bowl rinse key is pressed to rinse the bowl it will remain on for the same amount of time as the bowl rinse key was held down. Glass fill - on/off Briefly press the glass fill key on the dentist's control panel to switch the glass fill on. Filling will automatically stop after the preset fill duration has elapsed. If you wish to stop the glass fill before the preset fill duration has elapsed, press the glass fill key a second time. Glass fill - setting the fill duration Press and hold down the glass fill key for the amount of time that you wish the glass fill to operate (minimum one second, maximum ten seconds). Now, when the glass fill key is pressed to fill the glass it will remain on for the same amount of time as the glass fill key was held down. Door lock (Option) If a remote door lock is in use and has been connected to the dental unit this key is used to activate it. Auxiliary (Aux) key If any other equipment is being used with the unit this key can be used to activate it. Quint 7000 Dental Unit 17
Water bottle To remove and refill the water bottle unscrew it. Water bottle Tray holder The tray holder is attached to a tray holder arm which is attached to the underside of the instrument bridge on either the left- or the right-hand side. The arm is angled allowing the tray to be positioned either high or low depending on which way round the arm is attached. To position the tray holder loosen the two tray arm knobs, move the tray holder to the required position and then tighten the two knobs. Instrument bridge Tray holder arm Attaching knobs Tray holder The tray holder will hold one standard size tray or two half size trays. A double size tray holder is available as an option. 18 User's Manual
The Suction Arm Positioning the suction arm The height of the suction arm can be adjusted by squeezing the locking handle, on the underside of the top part of the suction arm, and pulling the suction arm up or pushing it down. When it is at the required height, release the locking handle. Suction arm Locking handle Warning Do not leave the suction arm under the chair backrest as the arm could be damaged if you lower the chair. Using the suction nozzles Starting and stopping the suction When a suction nozzle is lifted from its holder suction will automatically start. When the suction nozzle is replaced suction will stop. Quint 7000 Dental Unit 19
Increase Adjusting the suction Slide the knob on the side of the nozzle upwards to increase the suction and downwards to decrease the suction. Suction nozzle Decrease Knob Suction tip To attach the suction tip push it into the opening at the end of the suction nozzle. The Syringe Spray Press the green button to activate the water spray. Press the blue button to activate the air. Press both buttons together to activate the air/water spray. Water Air 20 User's Manual
Suction arm controls Bowl rinse - on/off Briefly press the bowl rinse key on the assistant's control panel to switch the bowl rinse on. Rinsing will automatically stop after the preset rinse duration has elapsed. If you wish to stop the bowl rinse before the preset rinse time has elapsed, press the bowl rinse key a second time. Bowl rinse - setting the rinse duration Press and hold down the bowl rinse key for the amount of time that you wish the bowl rinse to operate (minimum one second, maximum ten seconds). Now, when the bowl rinse key is pressed to rinse the bowl it will remain on for the same amount of time as the bowl rinse key was held down. Glass fill - on/off Briefly press the glass fill key on the assistant's control panel to switch the glass fill on. Filling will automatically stop after the preset fill duration has elapsed. If you wish to stop the glass fill before the preset fill duration has elapsed, press the glass fill key a second time. Glass fill - setting the fill duration Press and hold down the glass fill key for the amount of time that you wish the glass fill to operate (minimum one second, maximum ten seconds). Now, when the glass fill key is pressed to fill the glass it will remain on for the same amount of time as the glass fill key was held down. Door lock (Option) If a remote door lock is in use and has been connected to the dental unit this key is used to activate it. Quint 7000 Dental Unit 21
The Dental Light Switching it on and off The switch on the underside of the arm that holds the dental light is used to switch the light on and off. Adjusting the light intensity. The switch on the underside of the dental light is used to adjust the intensity of the light. On/off switch Intensity switch For more information on using and maintaining the dental light, refer to the documentation supplied with the light. 22 User's Manual
4. Cleaning, disinfecting and sterilising the unit Suction system Suction hoses For information on how to clean the suction system, refer to the documentation supplied with the system After every patient allow one glass of water to be sucked through each suction hose. Twice a day, midday and evening, allow one glass of suction hose cleaning agent to be sucked through each hose. Make sure that you prepare and use it in accordance with the manufacturer's instructions. Twice a day, midday and evening, empty and rinse the suction hose filters. The suction hose filters are located inside the connectors that connect the suction tubes to the cuspidor. To empty the filter, pull the suction hose connector from the cuspidor. Lift the filter out and rinse it. Also empty any debris that is inside the connector. Replace the filter and reconnect the suction hose connector to the cuspidor. Cuspidor Suction hose filter Connector Suction hoses Quint 7000 Dental Unit 23
Instruments and suction nozzles Unit surfaces Instrument bridge For information on how to clean and sterilize the instruments, refer to the documentation supplied with them. Caution Switch the dental unit off before you remove any instruments. Press the syringe air and water buttons to relieve any pressure before removing the syringe. Suction nozzles can be disconnected from the hoses and autoclaved. Unit surfaces must cleaned and disinfected at least once a day with a suitable cleaning and disinfecting agents. Make sure that you prepare and use them in accordance with the manufactures instructions. Do not uses abrasive cleaning agents. Note that the rubber instrument support can be removed from the instrument bridge and autoclaved. Instrument support 24 User's Manual
The rollers and holders at the end of the whip arms can be removed so that the instrument tubes can be cleaned. To remove the roller and holder slide the holder up slightly until the roller can be removed, and then remove the holder. Roller When replacing the roller and holder note that the two flats on the roller axle must engage in the slot in the holder, so that the holes line up, before the roller and holder can be slide back onto the whip arm. Holder Roller axle Flat Whip arms Flat Slot On units that are fitted with hanging tube instruments the instrument holders can be removed. To do this tilt the instrument holder forward as far as it will go, and then lift the holder up slightly until it disengages from the holding pin. The holder can then be removed. Instrument holder Tilt forward Lift and remove Quint 7000 Dental Unit 25
To replace the instrument holder place the instrument holder so that the holding pin in the instrument bride goes into the hole in the holder bracket. Push the instrument holder down to lock it in position. Instrument holder Holding pin Holder bracket Hole 26 User's Manual
5. Maintenance Oil collector The cotton pad in the turbine oil collector must be replace monthly or more frequently if free oil starts to appear. The oil collector on the underside of the instrument bridge. To remove and empty the oil collector unscrew it. Instrument bridge Oil collector Tightening instrument holders On units that are fitted with hanging tube instruments the instrument holders can tilted to make it easier to remove and replace instruments. If a holder becomes loose it can be tightened with screws on the side of the holder. Instrument holder Screws Quint 7000 Dental Unit 27
Instrument water adjustment The supply of water to the instruments can be adjusted. This is done by turning the regulator screws located on the side of the instrument bridge with a small screwdriver. The side on which the regulators are located will depend on whether the unit is left- or right-handed. Water adjustment regulators Micromotor 1 water Syringe water Micromotor 2 water Scaler water Turbine water 28 User's Manual
6. Service Every six months a FINNDENT approved service technician must carry out the following inspection and service tasks: check the output pressures of the air and water control valves check the air pressure of the instruments check to see if the filters in the air and water control valves need replacing check the operation of the instruments and if necessary replace and instrument O-rings or seals Caution Switch the dental unit off before you remove any instruments. Press the syringe air and water buttons to relieve any pressure before removing the syringe. Quint 7000 Dental Unit 29
7. Approved accessories Micromotor drills Turbine drills Air micromotors Light probes The following list contains instruments that can be used with the Quint 7000 dental unit. Some of the listed instruments are not part of the Finndent product range. Kavo Intramatic Lux - Models 190, 192 and 196 Satelec Servotorque - Models IML, IS, ES, AV, S Quint (Satelec Servotorque IML) Bien Air - Isolite 100, 300, MC40IR and MC20IR Kavo - 630 630B, 640B and 650B Bien Air - Black Pearl, Ondine, Gyro Long Life and Prestige S30L W&H Toplight - all models Kavo - all models Bien Air - all models W&H - all models Kavo - Dia-Lux 2300L W&H - LS 1 Diagnostic Light Bien - Air Translite 30 User's Manual
Tartar scalers Satelec Suprasson SP 4055 Amdent 930 Quint (manufactured by Amdent) Polymerisation lights Lysta LC-85 Satelec Servolite SL 4055 Satelec Acta Syringes Quint 3-way syringe Water heaters Luzzani 6-way syringe Amalgam separators Dental lights SIE-Dental - water heating element 50/100W, 24V Cattani - Suction separator Metasys Multisystem Type MST 1 Metasys Multisystem MST 1 ECO Faro - S90, S2000 and SC2000 Forma (230 V or 24 V) Quint 7000 Dental Unit 31
8. Technical Specifications PROTECTION CLASS: I PROTECTION GRADE: B PROTECTION AGAINST FLUIDS: IPX0 OPERATIONAL SAFETY GRADE: - Standard: The unit is not suitable for use in an environ ment with a mixture of inflammable anaesthesia gas and air, or a mixture of inflammable anaesthesia gas and oxygen or laughing gas (nitrous oxide). MODE OF OPERATION: Continuous operation with short-term load. ALLOWED OPERATING CONDITIONS: - Temperature: 15 o C - 40 o C - Relative humidity: 45% - 75% ELECTRICAL FEATURES: - Power supply: 230 V, 50 Hz - Maximum power consumption: - Main fuse: 6.3 AT. - Primary fuse of transformer: 2AT 1200 W (includes power consumption of possible accessories connected to the unit) WATER CONNECTION: COMPRESSED AIR CONNECTION: WEIGHT : MANUFACTURER: 3 bar - 6 bar particle size < 10 um 5.5 bar - 8 bar 78 kg without accessories 90 kg with accessories 125 kg in package with accessories FINNDENT OY TEOLLISUUSTIE 5 07230 MONNINKYLÄ FINLAND TEL. +358 19-5221500 FAX +358 19-643457 32 User's Manual
Symbols and abbreviations v Voltage A W Ampere Watt ~ Alternating current FI Approved by Fimko Symbol for protection class B European Community approval mark AT Amperes, slow blow fuse Control symbols are described where appropriate in this guide. Quint 7000 Dental Unit 33
Dimensions 34 User's Manual
Quint 7000 Dental Unit 35
36 User's Manual
Quint 7000 Dental Unit 37