IBC Position on Bunsen Burner Use in Biosafety Cabinets

Similar documents
VP, Facilities and Construction Management

Adv. 8.4 Guidelines for Biosafety Cabinets Rev

Safe Use of a Class II Biological Safety Cabinet

Appendix C IBC Safe Use of Biosafety Cabinets SOP

Safe Operating Procedure (Revised 2/10)

Biological Safety Cabinets. Biological Safety Cabinets

Florida International University. Biosafety Cabinet Program

Animal Tissue Culture SQG 3242 Equipment and Lab design. Dr. SITI PAULIENA MOHD BOHARI

SECTION LABORATORY EQUIPMENT CONSTRUCTION STANDARD PART 1: GENERAL Scope of Standard

Class II Biosafety Cabinets

BIOLOGICAL SAFETY CABINET

Biological Safety Cabinets. Presented by Mr. Rene Soetens President

Environmental Health and Safety Biological Safety Program Guidance Document

5.0 Installation Instructions

PART III. Laboratory equipment

Section 1.0: LAB VENTILATION TABLE OF CONTENTS. C. General Ventilation Considerations 3. D. Pressurization 7. E. Supply Air Arrangements 8

Clean Air Device. Yale University. Office of Environmental Health & Safety. Primary Containment Device

Research Document Type: PROCEDURE. Effective Date: January 01,2015 Author Steven Hayes

Biological Safety Cabinets or Fume Hoods Which one to use?

Fume Hood Safety TYPES OF FUME HOODS

Primary Containment for Biohazards: Selection, Installation and Use of Biological Safety Cabinets

THE IMPACT OF FACILITY HVAC SYSTEMS ON BIOLOGICAL SAFETY CABINETS

LABORATORY EQUIPMENT. A. Section Accessibility for People with Disabilities. B. Section Plumbing Basic Materials and Methods

CHEMICAL FUME HOOD PROGRAM

Biosafety Cabinets and Fume Hoods Supplemental Material (OHS_BIO304)

OPERATION AND MAINTENANCE MANUAL

Section 5.0: HAZARDOUS MATEIRAL STORAGE TABLE OF CONTENTS. A. Regulations, Standards and References 2. B. Scope 3. C. Approvals and Listings 3

AES Environmental. Operation and maintenance manual. VWS Series. Laminar Vertical Flow clean work station

University of Maryland Baltimore Environmental Health and Safety updated 1/3/2018 General Lab Inspection Common Findings and Corrective Actions

SECTION BIOLOGICAL SAFETY CABINETS AND LAMINAR AIRFLOW WORK STATIONS

Smoke detection in hazardous areas:

ChBE Safety Training. July Emily Edwards Lab Coordinator Chemical & Biological Engineering Tufts University 4 Colby St., Room

LSU Inter-Institutional Biological and Recombinant DNA Safety Committee (IBRDSC)

Forma 1400 Series Class II, Type A2 Biological Safety Cabinet. Superior Ergonomics, Uncompromising Safety

ESCO. A Guide To Biosafety & Biosafety Cabinets

GUIDANCE NOTE 3: SAFE USE OF EQUIPMENT

NEW! Thermo Scientific 1300 Series B2 Class II, Type B2 Biological Safety Cabinet

INCREASE YOUR LAB EFFICIENCY

6/19/2015 BIOLOGICAL SAFETY CABINET. Types of Classes. Types of BSC.

Clean Bench Animal Transfer Stations

Laboratory Safety Audit

INFRARED RADIANT HEATER RSCA SERIES CERAMIC HEATERS NATURAL GAS PROPANE GAS DATE: PROJECT: ARCHITECT/ENGINEER: CONTRACTOR: SUBMITTED BY:

SECTION SPECIAL EXHAUST SYSTEMS. A. Laboratory Ventilation and Hazardous Exhaust Systems. 3. Chemical Fume Hoods and Storage Cabinets

INCREASE YOUR LAB EFFICIENCY

Appendix A.4.1. Safe Fume Hood User Guide

LABORATORY SAFETY SERIES: Flammables & Explosives

Personnel & Product Protection: AGuideTo Biosafety Enclosures

MIT Emergency Response Guide

Thermo Scientific 1300 Series A2

Laboratory Safety Audit

Connection Instructions: Welch 2042 Vacuum Pump To TVO-2 Vacuum Oven

USER S INFORMATION MANUAL

This document provides general instructions on the use of chemical fume hoods to protect lab personnel against exposure to hazardous materials.

Lessons Learned. Bunsen Burner Fires

Beakers hold solids or liquids that will not release gases when reacted or are unlikely to splatter if stirred or heated. Beaker

Chapter 10: Laboratory Equipment

TG-10 Wall Mount Installation and Operation Manual. High Output Ozone Generator

IVF/ART Series Laminar Flow Cabinets

OPERATOR S MANUAL. Horizontal Laminar Flow Clean Bench

Beakers hold solids or liquids that will not release gases when reacted or are unlikely to splatter if stirred or heated. Beaker

Faster UltraSafe Class II Microbiological Safety Cabinets. Protection, safety, reliability. And more.

Microbiological Safety Cabinets

Fire Prevention Plan

Laboratory Control and Safety Solutions Application Guide

CBC s Inspection Findings and Possible Remedies Thereto

USER S INFORMATION MANUAL

Familiarize yourself with the following:

Control of Biological Hazards

Major Hazards in Laboratory

BSC CLASS II, TYPE A2 BIOLOGICAL SAFETY CABINET SERIES. Optimum protection of personnel, product and environment. BSC Class

A look into our complete product collection.

LABGARD NU-602 CLASS II TYPE A2 ANIMAL HANDLING BIOLOGICAL SAFETY CABINET

UC DAVIS Campus Standards & Design Guide DIVISION 11 - EQUIPMENT

Beakers hold solids or liquids that will not release gases when reacted or are unlikely to splatter if stirred or heated. Beaker

LABORATORY INSPECTION CHECKLIST

INDIAN INSTITUTE OF TECHNOLOGY BOMBAY

Quick Reference Guide to Cabinet types

USER S INFORMATION MANUAL

LABORATORIES, CHEMICAL AND BIOLOGICAL

QUARTERLY LABORATORY INSPECTION FORM

Labgard Class II, Laminar Flow Biological Safety Cabinet. Operation Manual

Lab Safety and Equipment

Aseptic Technique A.4D.1 APPENDIX 4D. Tomasz Bykowski 1 and Brian Stevenson 2 ABSTRACT

Safety In Chemistry. Identified Risks and Hazards

Infrared Sterilizer Operations Manual.

WORKPLACE INSPECTION CHECKLIST FOR LABORATORY ENVIRONMENTS

Biological Safety Cabinet Failure Response Procedure

Department of Civil and Environmental Engineering University of Waterloo

Steam Fundamentals. (in heating & processing applications)

Products Solutions Services CT287. Hazardous Area Classifications and Protection Systems

LABORATORY SAFETY DESIGN GUIDE

STANDARDS FOR VAPOUR EMISSION CONTROL SYSTEMS ( MARPOL reg VI/15 )

Fire Prevention Plan

Commissioning of a Mobile CL3 Laboratory: Lessons Learned

Environmental Health and Safety Iowa State University Services and Fees March 2017 (Updated February 2018)

Laboratory Equipment. Section 1L. Chemistry I & II

Laboratory Design, Accommodation & Environment. Presenter : Sonam Bansal Head -Nestlé Quality Assurance Centre Nestlé India Limited

Labconco XPress Laboratory Equipment Where You Need It, When You Need It

Environmental Health and Safety Iowa State University Services and Fees March 2017 (Updated May 2018)

VACUSAFE. The aspiration system for safe disposal of liquids

Transcription:

Background: IBC Position on Bunsen Burner Use in Biosafety Cabinets The Bunsen burner was once a part of every microbiology laboratory and is still widely used today. In the past, microbiologists had to rely on the use of open flames to ensure sterility while performing certain procedures. As modern technology has advanced with the introduction of biological safety cabinets and sterile disposable items, the need for an open flame has become virtually obsolete. The Center for Disease Control and Prevention (CDC) in the Biosafety in Microbiological and Biomedical Laboratories (BMBL) 5 th edition states that open flames are not required in the near microbe free environment of a biological safety cabinet and create turbulence which disrupts the pattern of air supplied to the work surface jeopardizing the sterility of the work area and safety of the worker. In addition to the CDC, many institutions across the country, the major biosafety cabinet manufacturers and the World Health Organization (WHO) also strongly discourage the practice. The use of Bunsen burners inside of a biological safety cabinet is not recommended because it: Disrupts airflow, compromising the protection of the worker and the product. The Class II BSC maintains product protection through delivery of laminar flow (air volumes traveling in a single direction at a constant speed without turbulence) down over the work area of the cabinet. The heating of air from the Bunsen burner causes up flow of air that mixes with the down flowing airstreams to produce turbulence and recirculation within the working area. The notion of laminar flow may be completely destroyed and any aerosols generated beneath the burner may be carried to other parts of the cabinet, jeopardizing the product and personnel working within the cabinet. Causes excessive heat build up within the cabinet. As most Class II BSCs recirculate the majority of the air within the cabinet, heat from the Bunsen burner builds up over time. The excessive heat can inactivate and degrade components in media such as vitamins, amino acids and growth factors, possibly below the threshold for finicky cell lines. In addition, the excessive heat may make it an uncomfortable environment for the worker, leading to errors and mistakes. May damage the HEPA filter or melt the adhesive holding the filter together, compromising the cabinet s integrity. An open flame has the capacity for melting the bonding agent that holds the HEPA filter media to its frame. This destroys the HEPA filters effectiveness, leading to loss of containment in the positive pressure plenum. Technical Safety Services (TSS) will charge $181 for decontamination of the cabinet, $250 for filter replacement labor, $250 $1000 for the filter, and $130 for recertification each time the HEPA filter needs to be replaced.

Labconco Corporation image of a biosafety cabinet HEPA filter damaged from use of an open flame inside the cabinet. Presents a potential fire or explosion within the cabinet. The cabinets are not constructed to be explosion proof. If the flame was to go out, there was a leak, or the valve was not shut off completely, flammable gas would be introduced to the cabinet at a steady rate. In the case of a Class II A2, where 70% of the air in the BSC is recirculated, concentrations of the flammable gas could reach explosive potential and pose a serious risk to not only the cabinet, but to the user and the laboratory it is occupied in. Electrical components like the fan motor, lights, or electrical outlets could ignite a flash fire with a spark in this volatile environment. Manufacturers often post their cabinets with warning labels stating that flammable materials should not be used in the cabinet. BSC fire. Source: Stanford University, Use of open flames in Cabinets/Tissue Culture Hoods (May 29, 2003).

Inactivates manufacturer s warranties on the cabinet. Biological safety cabinet manufactures are opposed to the practice and will assume no liability in the event of fire, explosion or worker exposure due to the use of a flammable gas in their cabinet. Automatically voids UL approval. Underwriters Laboratories Inc. (UL) is an OSHA approved independent product safety certification organization that develops standards and test procedures for products, materials, components, assemblies, tools and equipment, chiefly dealing with product safety. The use of a Bunsen burner in the cabinet will void UL approval of that piece of equipment. Requires hook up of central gas source. Some laboratories may not be fitted with gas lines and will require costly room renovations for retrofitting. Facilities will also need to install plumbing from the house lines to the cabinet. If the cabinet needs to be moved, this will incur additional costs from Facilities. In addition, gas connectors are generally not supplied with new biosafety cabinets without customer insistence and at an additional cost. Cited Quotes from Agencies and Manufacturers: NIH/CDC: National Institutes of Health and the Centers for Disease Control and Prevention: Open flames are not required in the near microbe free environment of a biological safety cabinet. On an open bench, flaming the neck of a culture vessel will create an upward air current which prevents microorganisms from falling into the tube or flask. An open flame in a BSC, however, creates turbulence which disrupts the pattern of HEPA filtered air supplied to the work surface. WHO: World Health Organization s Laboratory Biosafety Manual: Open flames should be avoided in the near microbe free environment created inside the BSC. They disrupt the airflow patterns and can be dangerous when volatile, flammable substances are also used. To sterilize bacteriological loops, micro burners or electric furnaces are available and are preferable to open flames. Public Health Agency of Canada; The Laboratory Biosafety Guidelines: The provision of natural gas to BSC s is not recommended. Open flames in the BSC create turbulence, disrupt airflow patterns and can damage the HEPA filter. When suitable alternatives (e.g., disposable sterile loops, micro incinerators) are not possible, touch plate micro burners that have a pilot light to provide a flame on demand may be used. NSF/ANSI Standard 49 2009 published by NSF International, Annex G; Section G.3.3.1: Service valves allow inert gases, air, or vacuum lines to be plumbed into the BSC. Although many users connect gas to a service valve in the cabinet, this practice should be avoided because open flames in a Class II BSC disrupts the airflow, and there is the possibility of a buildup of flammable gas in BSC s that recirculate their air.

The Baker Company (BSC manufacturer): The Baker Company does not endorse the use of flammable gasses within BSC s under any conditions. There are alternatives to open flames such as small electrical incinerators, use of disposables, and proper aseptic technique. NuAire (BSC manufacturer): NuAire doesn t recommend the use of natural gas within the BSC and assumes no liability for its use. USE AT YOUR OWN RISK. The Bunsen burner flame within the BSC not only contributes to heat build up; it also disrupts the laminar air stream, which must be maintained for maximum efficiency. If the procedure demands use of a flame, a Bunsen burner with on demand ignition is strongly recommended. Do not use constant flame gas burners. During use, the Bunsen burner should be placed to the rear of the workspace where the resulting air turbulence will have a minimal effect. Recommendations: Use disposable sterile loops and lab supplies. Autoclave utensils and equipment. Replace Bunsen burners with electrical incinerators for sterilization. If a flame is absolutely necessary, use a touch plate microburner to provide flame on demand. Alternatives to Continuous Flame Bunsen Burners: Safety Lab Gas Burner Safety enhanced laboratory gas burner with "Touch Free" IR Senso r and button function. (Approx. cost $532)

Bacti cinerator Utilizes infrared heat to incinerate organic material deep within the ceramic funnel. (Approx. cost $100 used $350 new) Touch O Matic Bunsen Burner (Approx. cost $85, free with EH&S trade in program) Glass Bead Sterilizer The glass beads in the well are maintained at 250 C for complete destruction of microorganisms and spores in just seconds. (Approx. cost $388) FLAMEBOY Portable flame sterilizer (Approx. cost $500)