Columns XK 16, XK 26, XK 50 Adaptors AK 16, AK 26, AK 50 INSTRUCTIONS

Similar documents
GE Healthcare. Mixer M-925. Instructions

ReadyKart mobile processing station

G Trap FliQ FPLC Columns

Wide-Bore Heat Exchanger Installation Kit ( )

1 L Rotary Vacuum Evaporator Instruction Manual

VACUUM VALVES AND HOSES

INSTALLATION INSTRUCTIONS

UVC 500 Ultraviolet Crosslinker

Instruction Manual GP110 GelPump

EDGEMERE DUAL CONTROL WIDESPREAD LAVATORY FAUCET INSTALLATION INSTRUCTIONS

I N ST R UC T I ON MODEL 2612 & 2712 SLICERS MODELS 2612 ML ML FORM (12-98) 2612 SLICER

Rhino AB Upper Check Valve Replacement

Aspirator with trap flask FTA-1

Lavatory Faucet INSTALLATION INSTRUCTIONS P24490-CR, P24490-LV, P24491-CR, P24491-LV, P24492-CR, P24492-LV, P24700, P24705, P24706, P24736, P24800

CryoFit. Handling of Flow Cell and Capillaries. User Guide. Version 002

NMR SAMPLING ACCESSORIES CLEANING NMR SAMPLE TUBES

SUPPOSITORY MOLD RR 2159 Instruction Manual

I NS T RUC T ION MODEL 4346 MIXER-GRINDER MODEL

Corning Storage Bottles Selection and Use Guide

Eldex Column Heater. Operator s Manual

Methods of solid phase extraction. Methods of solid phase extraction

Pos-D Positive-Displacement Pipette for Problem Liquids

Regulux. Radiator lockshield with drain-off facility. To be precise.

PLATE HEAT EXCHANGERS. Operation and Maintenance Manual

Agilent Vac Elut SPS 24 Manifold User Guide

Series 1140 and 1141 Temperature Regulators

PENGUIN SERIES HSC PUMPS INTRODUCTION

Agilent InfinityLab Online SPE Solution

Corning Storage Bottles Selection and Use Guide

Oblo Basin and Bidet Monobloc Installation and User Guide

13-1. Temperature Regulator

CONDENSERS 119. Allihn / Cold Finger / Dewar. Allihn Micro Condenser. Accessories. Allihn Condensers with Full Length Joints. Cold Finger Condensers

LABORATORY ULTRA ZERO AIR GENERATOR. ZAC-ULT Series USER MANUAL

Operating Instructions

AUC Cell Washer. User Manual

Recirculating Cooler SRC14

FTA-1 Aspirator with trap flask

Retractable ph/orp Sensor

HOT LIPS TUBE SEALER TM

Vacuum Pump Kit User Manual

FORENSICS GLASS BLOCK MANIFOLD INNOVATION THROUGH CHEMISTRY

DRAIN WATER TEMPERING INSTALL

Model 4146 Chopper. This manual replaces and should be used instead of form (9-84) FORM (September 2012)

Mariner Microspray and Low-Flow Electrospray Fused Silica Spray Tip and Capillary Kit. Operating Instructions. 1 Product Description.

Husky Diaphragm Pump Suction Kits

Vertical Gel Systems-Dual CAT NO: EPS-V0011 & EPS-V0012

REVE ORDERING INFORMATION NOTES T01-A INSTALLATION INSTRUCTIONS 现代型分体座便器 CONTEMPORARY TWO-PIECE TOILET K-17178T-S/K-17178T-SP

A HEATER KIT With Diesel Engine Connection Fittings

Installation & Operating Instructions

Spare parts list. Hydraulic compactor RC

MODEL 4246 MIXER-GRINDER

INSTALLATION OF DURAVENT POLYPRO SINGLE WALL POLYPROPYLENE VENTING SYSTEMS

Operation of heat exchangers

OWNER S MANUAL. How to install, operate and maintain your. EcoWater Systems Automatic Sediment Filter. EcoWater Systems Poland

INSTALLATION INSTRUCTIONS

MODEL SK2750. COMPRESSOR SERVICE KIT For use on 2750 Series Compressors

For more information contact: Baker Water Systems Phone:

Instructions for CO 2 Cold Trap System. Part No

MODEL COMPRESSOR SERVICE KIT For use on Midmark Power Air Oil-less Compressor Models P21, P22 & P32

52 CEILING FAN READ AND SAVE THESE INSTRUCTIONS FAN RATING AC 120V.

SALES SPECIFICATION. E3000 and E3100 Critical Point Drying Apparatus

PIPE DREAMS 96. Aeroponic Garden IMPORTANT:

Quick Start Guide LIQ-QSG-Hx438, Rev A June Rosemount Hx438. Amperometric Steam Sterilizable Dissolved Oxygen Sensors

PRODUCT GUIDE. High performance heating system filter Improves energy efficiency Captures magnetic and nonmagnetic.

ROTARY VANE VACUUM PUMPS Exploded View DuraVane Rotary Vane Pumps Oil-Lubricated RVL301HH

Rapidly process up to 20 samples without the need for repeated centrifugation steps.

Disposal Unit (BKSDU) Burger King Shortening. Operation, Service & Parts Manual BURGER KING BKSDU

TUB AND SHOWER FAUCET

Liquid Samplers. Powders. Tel: +44 (0)

THE NEXT GENERATION CARBOY

Processing Instructions. Cleaning so easy.

ALH SYSTEMS LTD ALH SYSTEM 1 SERVICE VALVE AND DRILLING MACHINE OPERATING MANUAL

Models: GW/R-T GW/R-2-T GW/R-3-T

SPECIFICATIONS CWC - 2P / 4P CWC 020 CWC 030 CWC 040 CWC 050 CWC 070 CWC

Medium main line filter

INSTALLATION INSTRUCTION

E.2 series high efficiency circulator. Installation and operating instructions

VPMmini. Operator's Manual

TECHNICAL INSTRUCTIONS

Harvey Chemiclave EC Series Sterilizers

INSTALLATION OF CENTROTHERM INNOFLUE SINGLE WALL POLYPROPYLENE VENTING SYSTEMS

Instruction Manual - Anti-Siphon Ejector Chlorine & Sulfur Dioxide 500 PPD (10 kg/h) Maximum Capacity

I N ST R UC T I ON. MODEL HAB500 AroMatic TM BREAD SLICERS MODEL FORM (4-99) AROMATIC BREAD SLICER 701 S. RIDGE AVENUE TROY, OHIO

ACQUITY UPLC HT Column Heater Instructions

Owner s Guide and Installation Manual

INSTALLATION GUIDE. Yellow Jacket Misting System. For help, call BIG-FANS or visit

OWNER S MANUAL. Models: Herkules Equipment Corporation 2760 Ridgeway Court Walled Lake, MI USA Fax

Platen Heater Pad Replacement

HP-500 Plus, XP-1500 Plus & OMNI

FCHM Hydromatic Pump Control Module

Drinking Water Purification Systems

Agilent 1290 Infinity II Online SPE Solution

DIRTCAL - DIRTMAG dirt separator

Trooper Portable Carpet Extractor. Operator and Parts Manual. Model No.: Can Rev. 00 (09-99)

Explorer CleaningPartsDirect.com Carpet Extractor. Operator and Parts Manual. Home Find... Go To.. Model No.: Can.

Installation manual. Tumble dryers T4130. Compass Control

1500kN Concrete Machine. Impact Test Equipment Ltd & User Guide. User Guide

Xpert HD Xpert HD 7140, ,

original your automatic choice Socorex The Socorex original models Twin syringe = extended programme

SPRINGHILL 44 CEILING FAN

Transcription:

Columns XK 6, XK 26, XK 0 Adaptors AK 6, AK 26, AK 0 INSTRUCTIONS XK columns, designed for Standard Liquid Chromatography of macromolecules. They are jacketed and available in the following dimensions: length: 20 cm 30 cm 40 cm 60 cm 70 cm 00 cm i.d. 6 mm XK 6/20 XK 6/40 XK 6/70 XK 6/00 i.d. 26 mm XK 26/20 XK 26/40 XK 26/70 XK 26/00 i.d. 0 mm XK 0/20 XK 0/30 XK 0/60 XK 0/00 AK adaptors are for use with XK columns. One adaptor is fitted in each column on delivery. By fitting an additional adaptor at the bottom end of the column a larger range of bed heights is obtained (Table ). Unpacking XK columns are delivered complete with thermostat jacket, one flow adaptor and flanged tubing at both ends for direct connection to valves, pumps and UVmonitors. Unpack the XK column or adaptor carefully and check the contents against the packing list supplied. Inspect for any missing components or damage that may have occured during transit. Report any damage immediately to the local Amersham Biosciences representative and to the transport company concerned. Description Please refer to the exploded diagram (Fig. 3) for details. Chromatographic tube (6). The highprecision bore, borosilicate glass tube has a length of 200, 300, 400, 600, 700±0.2 or 000± mm with an inner diameter of 6±0.02 mm, 26±0.02 mm or 0±.00 mm. Thermostat jacket (7). The jacket is an acrylic plastic tube with threaded ends. Sealing is provided by an Oring and a sealing ring. Column end pieces (). The red end pieces, made of reinforced acetal plastic, hold the chromatographic and jacket tubes in position. Each end piece houses an O ring, sealing ring, washer and locking ring. The end piece also carries a tubing connector through which thermostat fluid enters or leaves the column. Column bottom piece (). The bottom piece screws onto the column end piece. The gel bed is supported by a 0 µm nylon net ring () on top of the plunger (3). The bottom piece seals against the chromatographic tube (6) via the Oring (2). Adaptor (8). The adaptor consists of a top piece that fits the column end piece, a plunger (3) that carries the gel bed protection and support screen (4, ), and adjusting mechanism. The adjusting mechanism allows the plunger to be set in position in the chromatographic tube, and forms the mechanism for sealing against the column wall. Tubing. XK columns are supplied with flanged M6type capillary tubing. The adaptor tubing is 0 cm long. The bottom piece tubing is 30 cm long. A cm protection tubing is threaded over the capillary tubing in the inlet of the adaptor and column bottom piece to prevent squeezing of the capillary tubing during transport and storage. Materials Under normal operating conditions, the only materials in contact with the effluent are: glassfibre reinforced polypropylene, polyamide, acrylic plastic, borosilicate glass, fluororubber and Tefzel capillary tubing. Resistance The columns can be used in aqueous solutions and nearly all organic solvents commonly used in liquid chromatography of macromolecules with the following few exceptions: chlorinated hydrocarbons, acetone and other ketones, aliphatic esters and phenol. Solutions containing more than 0 % NaOH, 0 % HCl, other strong mineral acids or % acetic acid should not be used. 63082 Edition AH

The thermostat jacket should only be used with water as cooling fluid. Acrylic plastic is not resistant to alcohols, ethylene glycol or other organic solvents. The column and thermostat jacket may be used at temperatures up to 60 C and at pressures up to 0. MPa ( bar) for XK 6 and XK 26 columns and 0.3 MPa (3 bar) for XK 0 columns. The jacket can take 0.2 MPa (2 bar). Cleaning Suitable cleaning agents are soapy water or laboratory detergents. Enzyme detergents are recommended for removing proteinaceous contaminants. The column is autoclavable without the thermostat jacket, tubing and net ring. Alternatively, the complete column may be sterilized with ethylene oxide. Instructions Important! Always loosen the black adjusting knob (9) before inserting or removing the bottom piece and adaptor of the column. Connecting capillary tubing to XK columns (Fig. ). Dismantle the bottom piece ( ) or adaptor (8) as described below. 2. Take a piece of capillary tubing (0.8 mm i.d. for XK 6 and XK 26,.2 mm i.d. for XK 0) and make a flanged tubing connection with an M6 connector at one end. (See Tubing connection and Fig. 2 for detalils). Pass the unflanged end of the tubing down through the adaptor or up through bottom piece. 3. Slide a nipple over the tubing and flange using a flanging tool. See also Tubing connection. 4. Fingertighten the plunger (3).. Reassemble the bottom piece or adaptor. Fig.. Connection of flanged capillary tubing with nipple. 3 Tubing connection (Fig. 2) In the XK column system, the tubing connection is of M6 flangedtube type. For flanging 0.0.8 mm i.d. capillary Tefzel tubing, the Flanging/Start Up Kit 20 V ( 90790 ) or 220 V ( 90900 ) is necessary. For.2 mm tubing the Flanging Tip Kit.2 mm (84970 ) is required. Instructions for flanging capillary tubing are included in these kits. Fig. 2. Flanged tubing connection. Dismantling the column XK 6 and XK 26. Unscrew the bottom piece () and adaptor (8). 2. Use the dismantling tool (0) to unscrew the locking rings and to free the chromatographic tube (6). 3. Remove the sealing ring, washer and chromatographic tube (6). 4. Unscrew the column end pieces from the thermostat jacket (7). The Oring may now be removed. Dismantling the column XK 0. Unscrew the bottom piece () and the adaptor (8). 2. Unscrew the column end pieces from the thermostat jacket (7). The Orings may now be removed. Note: The XK 0 column chromatographic glass tube is held in position by builtin sealing rings in the end piece. Washer and locking rings are not required. Important! Since the XK 0/00 chromatographic glass tube is very heavy it can easily slide out from its hold in the builtin sealing rings. To keep the tube in position when washing the column, screw the conical adaptor (supplied with Packing Reservoir RK 0) to one of the end pieces of the column and hold that end of the column downwards during washing. After washing, screw the bottom end piece into the end piece before mounting the column vertically.

Dismantling the adaptor or bottom piece. Loosen the black adjusting knob (9). 2. Unscrew the plunger (3). 3. The Oring and expander ring (only XK 0 adaptor) (2) may now be removed. 4. The net ring () can be removed from the plunger with a sharp tool. The support screen (4) may now be removed. 8 Replacing the net ring and bed support (Fig. 4). Remove the net ring () with a sharp tool. 2. Warm the new net ring in hot water 060 C. 2 3 4 3. Lay the new support screen (4) and net ring () in place. 4. Press on the net ring. Reassembling the column XK 6 and XK 26. Put the chromatographic glass tube (6) into the thermostat jacket (7). 2. Install one of the end pieces by screwing in the thermostat jacket (7), ensuring that the Oring is in place. 6 3. Put a sealing ring onto the end of the glass tube followed by a washer. 7 4. Screw in the locking ring with the dismantling tool (0).. Repeat steps 24 for the other end of the column. Reassembling the column XK 0. Screw the two end pieces onto the thermostat jacket (7). Ensure that the metal springrings are seated in the builtin sealing rings and that the Orings are in place in the end pieces. 4 3 2. Wet the builtin sealing rings with water. Slide in one end of the chromatographic glass tube through the the builtin sealing ring, the thermostat jacket and then through the builtin sealing ring of the other end piece. Push the tube towards the second sealing ring. Use a small screwdriver or a similar blunt tool to carefully ease the sealing ring over the tube while at the same time pushing the tube through the sealing ring. 2 0 3. Important! Ensure that the distance from the sealing ring to the end of the chromatographic glass tube is equal at the both ends. 9 4. Screw the bottom piece () and the adaptor (8) into the end piece. Important! Tighten properly Fig. 3.

Twist the adaptor 3 mm down into the gel. Fingertighten the adjusting knob to obtain a good seal. Do not overtighten. Fig. 4. Replacing the support screen and net ring. Reassembling the adaptor. Connect the capillary tubing to the plunger as described above. 2. Finger tighten the inner adjusting tube in the plunger. 3. Adjust the Oring on the plunger against the outer adjusting tube using the adjusting knob (9). 4. The adaptor is now ready for use. Reassembling the bottom piece. Connect the capillary tubing to the plunger as described above. 2. Fingertighten the inner adjusting rod in the plunger. 3. Adjust the Oring on the plunger against the bottom piece using the adjusting knob. 4. Screw the bottom piece into the column end piece. Finger tighten the adjusting knob (9) to obtain a good seal. Do not overtighten. Inserting and adjusting the adaptor. Close the column outlet. 2. Adjust the level of liquid in the column to within 2 mm of the end of the chromatographic tube. 3. Slacken the tightening mechanism and insert the plunger at an angle so that no air is trapped under the net. 4. Adjust the black knob to give a sliding seal between the column wall and the Oring (2).. Slide the plunger slowly down the column so that the air in the adaptor above the net and in the capillary tubing is displaced by the eluent. 6. Screw the top piece into the column end piece. 7. Adjust the adaptor at the bed surface by pressing the outer adjusting tube down through the top piece. The adaptor moves down one thread at a time and is automatically locked in the desired position by QuickLock. 4 Adjusting the QuickLock adaptor in a packed column When the column is operating at high pressures 0.30. MPa (3 bar) it is not possible to press the QuickLock as it is selflocking. To readjust the position of the adaptor in the column follow the description below:. Stop the pump. Put the flow stopper pin into the outlet. 2. Disconnect the column from the pump. 3. Loosen the adjusting knob. 4. Press in QuickLock and adjust the adaptor at the bed level.. Fingertighten the adjusting knob to obtain a good seal. Do not overtighten. 6. Connect the column to the pump. 7. Disconnect the flow stopper pin. 8. Twist the adaptor 3 mm down into the gel, fingertighten the adjusting knob. Note: When using organic solvents, the liquid level in the column can be lowered to mm above the gel bed surface. The adaptor may then be inserted directly into the column and lowered to the liquid surface. The air in the adaptor is displaced and the position of the adaptor secured as described above. When the adaptor is in position and the gel bed has been equilibrated at the operating pressure, there should be no liquid space between the gel bed surface and the adaptor net. The position of the adaptor may be readjusted after equilibrating the gel bed, but before the adaptor is moved the column outlet must be closed.

Table. The maximum bed heights (cm) and bed volumes (ml) possible using one adaptor or two adaptors in the various XK columns. With one adaptor With two adaptors Column Volume (ml) Bed height (cm) Volume (ml) Bed height (cm) XK 6/20 XK 6/40 XK 6/70 XK 6/00 XK 26/20 XK 26/40 XK 26/70 XK 26/00 XK 0/20 XK 0/30 XK 0/60 XK 0/00 3 470 030 690 66 2286 28344 44004 0274 269 794088 88862 2.. 22.3 2.6 82.9 2. 233 36 839 04 428 406 89 03 670 7630 3690 066 486 204344 36304 0274 09 00088 274862 0. 83 386 689 02. 8.3 38.6 68.9 04 028 266 69 Ordering information Column Adaptor* Packing reservoir XK 6/20 XK 6/40 XK 6/70 XK 6/00 887730 887740 88770 887760 AK 6 887780 RK 6/26 887930 XK 26/20 XK 26/40 XK 26/70 XK 26/00 800072 887680 887690 887700 AK 26 887660 RK 6/26 887930 XK 0/20 XK 0/30 XK 0/60 XK 0/00 80007 8870 88720 88730 AK 0 AK 0FF XK 0FF Kit** 88740 800073 800070 RK 0 887900 * Each XK column is delivered with one AK adaptor ** XK0 FF Kit containing one AK0 FF and one XK0 FF end piece Accessories Sample Applicators Manual Valves Tubing Connectors SA SA0 Superloop 0 ml Superloop 0 ml 92460 92470 9780 97800 SRV SRV3 SRV4 V7 V8 9240 920980 920990 97000 97760 SRTC2 SRTC3 Luer lock female/ M6 male Luer lock female/ M6 male M6 male M6 female 92430 92440 802762 80272 80798 800368 Spare parts The exploded diagram (Fig. 3) shows the positions of the different parts of the XK columns. For replacements please order according to the spare parts list below using the appropriate code numbers.

Item No. Designation XK 6 XK 26 XK0 Material No. per pack 2 3 4 6 7 8 Bottom piece complete Oring Plunger Support screen Net ring (0 µm) Glass tube for column length 20 cm 30 cm 40 cm 60 cm 70 cm 00 cm Thermostat jacket for column length 20 cm 30 cm 40 cm 60 cm 70 cm 00 cm AK adaptor (complete) End piece complete 80347 90630 80380 9060 88760 9030 9030 9040 900 80072 80073 80074 8007 887780 864880 80348 906880 8038 893770 887600 800084 9040 90460 90470 800082 80067 80068 88790 887660 864890 80349 90680 88780 906640 88790 80008 93260 9020 90090 800083 80078 80077 88760 88740 887970 A E B D B,C F G Available but not illustrated Flanging/Startup Kit 220 V 20 V Flanging Tip Kit i.d..2 mm Capillary tubing.2 mm i.d.,.8 mm o.d. Capillary tubing 0.8 mm i.d.,.8 mm o.d. 90900 90790 84970 943700 97430 2 m 2 m Materials A = Reinforced acetal plastic B = Polypropylene (reinforced glassfibre) C = Polyamide (Nylon) D= Polypropylene Trademarks Superloop is a trademark of Amersham Biosciences Limited. Amersham and Amersham Biosciences are trademarks of Amersham plc. TEFZEL is a trademark of DuPont. Amersham Biosciences AB 2002 All rights reserved All goods and services are sold subject to the terms and conditions of sale of the company within the Amersham Biosciences group which supplies them. A copy of these terms and conditions is available on request. E = Ethylenepropylenedione polymer (EPDM) F = Borosilicate glass G= Acrylic plastic Amersham Biosciences AB, Björkgatan 30 SE7 84 Uppsala Sweden Amersham Biosciences UK Limited Amersham Place, Little Chalfont Buckinghamshire England HP7 9NA Amersham Biosciences Corporation 800 Centennial Avenue Piscataway, NJ 088 USA Amersham Biosciences Europe GmbH Postfach 480 D7902 Freiburg Germany Amersham Biosciences K.K. Sanken Building, 32 Shinjukuku, Tokyo 690073 Japan Printed in Sweden by TK i Uppsala AB, January 2003