Guide, Quick Start, Short, X-100 System

Similar documents
Operator s Manual. Model X-100. Universal Oximetry System English

Nellcor Bedside SpO 2 Patient Monitoring System, PM100N

Quick start guide. English

Model Operator s Manual. 4-Channel Regional Oximeter with Equanox Technology and Bluetooth Wireless Technology. English

LifeSense Vet LS1-10R

User Guide. HTM Nurse Call. In the event of requiring assistance, please contact your Installer/Supplier.

PalmSAT Model 2500A VET

N-600. Pulse Oximeter Home Use Guide

Quick Reference Guide *

Propaq LT Outline Train the Trainer

MODEL 3X Series User s Manual

CER-1 & CER-2 Installation & Testing Instructions

(Place Logo from Lab here)

SOFTWARE VERSION 8.5 FAQs

Operator s Manual. LifeSense LS1-9R. Vital Signs Monitor. Capnography/Pulse Oximeter

Quick setup guide. English

Operator s Manual. Avant Pulse Oximeter and Noninvasive Blood Pressure (NIBP) Monitor English

BeneVision TM80 Telemetry Monitor

Quick start guide for Altrix Precision Temperature Management System

i.c³ User Guide For Helmer i.series Ultra-Low Freezers A/A

Tissue Oximeter. Operator s Manual

Model 900. Pulse Oximeter User s Manual

Quick-Start Guide Digi-Sense TC9000 Advanced PID and On/Off Temperature Controller with Thermocouple Input Models and

Protégé EliteSuite Eclipse LED Keypad Installation Manual ELT-KLES

MOBILE CALL GSM Alarm System User s Manual

HUG.ai 1/24/06 9:33:34 AM HUG.ai 1/24/06 9:33:34 AM C M Y CM MY CY CMY K

Syringe Pump. User Manual

CircuFlow Quick Setup Guide

PULSE OXIMETER PACKAGE

Follett Performance Plus

User Guide V3 4/13 Rev. C

U ser's Guide PC6010

N-600. Pulse Oximeter Operator s Manual

PiSector GSM Cellular Wireless Alarm System

Digital Marine Exhaust Temperature Alarm

Operator s Manual. Model 7500FO. Model 7500FO. Digital Pulse Oximeter. English

Operator s Manual. Avant Pulse Oximeter and NIBP Monitor. English

About this Manual. Statement. P/N: Release Date: August 2010 Copyright EDAN INSTRUMENTS, INC All rights reserved.

Model 7500 Model 7500

Nellcor Puritan Bennett Inc. is an affiliate of Tyco Healthcare. Nellcor, Oxiband, Durasensor, OxiCliq, Dura-Y, MAX-FAST, and OXIMAX are trademarks

Pella Insynctive Product Guide

CNAP MONITOR 500 HD QUICK GUIDE. This Quick Guide DOES NOT REPLACE the user manual!

User Manual. December 2006 M Rev.A

IntelliVue Cableless Measurements

Table of Contents. Appendix A Special Characters 31

Operating & Maintenance Manual. Alert-4 Ethernet LCD Master Alarm

GSM RFID VOICE Alarm System

Security GSM Alarm System

Watchguard WGAP864 User Manual

NX-148 LCD CODE PAD TABLE OF CONTENTS

i.c³ User Guide for Compartmental Access Refrigerators

Clinical Information Center Pro v5 (CIC Pro)

Operator s Manual. Avant Pulse Oximeter and NIBP Monitor. English

Operator s Manual. Avant Avant 4000 Digital Pulse Oximetry System with Bluetooth Wireless Technology. English

SmarTemp Control fx 2.0

Operator s Manual. Avant Digital Pulse Oximeter. English

KFP-CF Series Operation Manual

INFOV.A.C. USER MANUAL THERAPY SYSTEM. Rx Only

Important information. Government Regulations and Safety information. Your Warranty. Replacement remote controls

User Guide K /05 TN LIC#: C-0183 C-0180 C-0685 AL LIC#: GA LIC#: LVA ARMED READY MESSAGE FUNC PANIC BACK

Nellcor with OxiMax Technology Product Ordering Guide A Full Line of Oximetry Sensors and Monitors

User Guide V1 7/12 Rev. A

Entero Voice Alarm system.

About this Manual. Statement. Release Date: January, 2014

Dryer Controller M720

A1UL PERS. Personal Emergency Response System. For Technical Support Please Contact Your Service Provider Or Distributor

Code Alert Series 30 Software User Guide

ACCESSORIES ISLEEP VIVO Z1

Fire Command Keypad. XR5 User s Guide

Model: Edge-HC. 1 edge-hc

User Manual. Dryer Controller M720

Area Systems While the alarm bell or siren is sounding, enter your user code. The system silences the alarm bell or siren.

Interactive Technologies Inc North 2nd Street North St. Paul, MN Technical Manuals Online! -

BeneVision Central Monitoring System. Quick Reference Guide

Atlas Monitor Outline Train the Trainer

901-M / 901-ML. Pulse Oximetry System. Technical Information and Service Manual

TS-STS Sump Test System Quick Start Guide 1

Quick Reference Guide STAN S31 fetal heart monitor. Power ON/OFF. ST Analysis (Fetal ECG analysis)

Welch Allyn Vital Signs Monitor 6000 Series. Directions for use

FLX 9050 Pager. Operating Guide. Contents

Spa Touch Control Panel with BP2100, BP6013 spa controllers. (Spa Owner s Manual insert)

SILENCING AN ALARM When the alarm bell or siren is sounding, enter your user code or present your SecuraProx fob to your keypad.

T1 BEDSIDE MONITOR GUIDELINE

WIRELESS ALARM SYSTEM WITH TELEPHONE AUTO DIALER

Connex Integrated Wall System 2.2 Software Version

SECURITY SYSTEM NOTES SPECIAL CODES ENTRY / EXIT DELAY TIMES ARM / DISARM CODES ZONE DESCRIPTIONS

Tomorrow s technology for today s security needs.

IQ PANEL USER GUIDE SMARTHOME.

Spa Touch Control Panel with 2000, 2100 controllers. (Spa Owner s Manual insert)

Passport 12m and Passport 17m BEDSIDE MONITOR GUIDELINE 5.21 Software Version NAME: HOSPITAL: DATE: VALIDATOR: Performed

Operator s Manual. Avant Digital Pulse Oximeter. English

Directions for Use. Alaris System. Alaris Products. Supports Guardrails Suite MX (v8) August RATE (ml/h) RATE (ml/h) CHANNEL SELECT PAUSE

Cl inician Training Packet

80 CHANNELS WIRELESS RECEIVER WITH LCD DISPLAY M1.1.1-Hx.x-F1.1-ENG [AN] [SPV] MADE IN ITALY INSTALLATION AND USE MANUAL

Transport Pro Patient Monitor Operator s Manual Software Version 2.1

FiRe mobile-2 Operation Manual

bibag System User s Guide

ALARM SYSTEM USER S MANUAL Rev

VPAP COPD. H5i. Welcome Guide. Making quality of care easy. Respiratory Care Solutions. English NONINVASIVE VENTILATOR HEATED HUMIDIFIER

NexSysLink. 2 CAN Display Operation Manual. CAN Instruments Product Family

Transcription:

Rev. CR/CO Date Orig. A 9950 2/15/16 DH Guide, Quick Start, Short, X-100 System P/N LANGUAGE NONIN ELECTRONIC FILE FINALSIZE 10885-001 ENGLISH 10885-001.01.INDD 5.5 X 8.5 NOTE 1: a. MATERIAL: 100lb cover with 5 MIL lamination b. COLOR: 4/Color c. PRINT: 5 pages 2/sided. OPERATION: ¼ Punch hole in upper left corner. BINDING: #3 Size (3/32 Diameter) stainless steel ball key chain, length 6. d. Sheet 2 of 2 is the guide. Title: MEDICAL, INC. MINNEAPOLIS, MN GUIDE, QUICK START, SHORT, X-100 SYSTEM Prepared By Date DH 2/10/2016 Part Number Checked By Date BAK 2/15/16 10885-XXX Rev A Sheet 1 of 2

Model X-100 Universal Oximetry (rso 2 /SpO 2 ) System Quick Start Guide

4CA SenSmart System Components Refer to the SenSmart Model X-100 Universal Oximetry System Operator s Manual for complete instructions for use to fully understand the device, its operation, and the potential risks to user and patient. Monitor Hub Cable 1 Hub 2 3 INT-100 Intermediate Cable Signal Processor Cables 4 Signal Processors 4CA 4CA 8100SM SpO 2 Sensor SenSmart rso 2 Sensors Nonin Medical, Inc. 13700 1st Avenue North Plymouth, MN 55441-5443 U.S.A. Tel: +1.763.553.9968 1.800.356.8874 Fax: +1.763.577.5521 E-mail: info@nonin.com Nonin Medical B.V. Prins Hendriklaan 26 1075 BD Amsterdam Netherlands Tel: +31 (0)13-79 99 040 Fax: +31 (0)13-79 99 042 E-mail: infointl@nonin.com sensmart.com nonin.com 2016 Nonin Medical, Inc. All trademarks are the property of Nonin Medical, Inc. unless otherwise noted. P/N 10885-001-01

Monitor Features and Patient Monitoring Graphs Channel Values Navigation Buttons Speaker 1 2 3 4 5 6 AC Power LED Power Cable Input (Bottom) Monitor Connector Port Nurse Call Input (Bottom) 1 2 3 4 5 6 ALARM SILENCE: Press once to silence audible alarms for 2 minutes, press once to reactivate. EVENT MARK: Press once to mark an event; press and hold for 2 seconds to open Event Mark table. MENU: Press to open Settings menu. Allows access to System menu which changes device settings. SELECT: Select item to edit using navigation buttons; press Select to save changed values. BASELINE: Press twice to quickly set all baseline(s) to the current rso 2 values. ON: Press once to power on. STANDBY (OFF): Press and hold for at least 1 second to power off. SenSmart Signal Processor Model X-100SP-1 Channel 1 X-100SP-2 Channel 2 X-100SP-3 Channel 3 X-100SP-4 Channel 4 X-100SP-5 Channel 5 X-100SP-6 Channel 6 Color Channel Blue Orange White Purple Green Pink The measurement from the Signal Processor will display on the Monitor screen in the same color and channel location as shown on the Signal Processor label. 3

Monitor Setup MONITOR SETUP: 1. Connect the SenSmart X-100 components as shown here. Up to six channels of rso 2 and SpO 2 monitoring can be utilized. 2. Press the On/Standby button to power on the Monitor. The Monitor will display the opening screen allowing the user to select a Preset. 3. Select a Preset. Verify the alarm limits are correct. 4. Press Menu to set the display to the monitoring screen. 1 2 3 4 To connect the Hub Cable to the Monitor port: Align the arrow on the Hub Cable connector with the small triangle on the Monitor connector port. Push connector straight into the port to click and lock into place. To disconnect: Grasp the sleeve on the Hub Cable connector, retract the sleeve to unlock, and pull the connector straight back to detach. To place the Hub into the Hub Holster Align the Hub and Hub Cable to the Hub Holster, push firmly into the Hub Holster; clip to linens or pole. To remove: Pull the clip back on the tip of the Hub Holster and pull out of holster. If not being used, the hub port covers should remain closed to eliminate the potential for environmental contamination. To connect a Signal Processor to the Hub: Align the arrow on the Signal Processor cable connector with the arrow on the Hub port or the Monitor connector port and push until it clicks and locks into port. To disconnect: Grasp the sleeve on the Signal Processor cable connector, retract to unlock and pull the Signal Processor connector straight back. To connect an Intermediate Cable or Sensor to the Signal Processor: (Refer to pages 5-6 for detailed instructions.) Note: The INT-100 intermediate cable is required to connect (patients >40 kg) sensors to the signal processor. The 8004CB and 8004CB-NA sensors (patients <40 kg) and 8100S(X) pulse oximetry soft sensors connect directly to the signal processor. Flip the clear lock on the signal processor back to expose the connection port, align the white arrow on the INT-100 connector or sensor connector with the white arrow on the signal processor port, and push until the connector is firmly in place. Flip the lock over the connector and click into place. If using an INT-100, connect a compatible sensor to the end of this intermediate cable. To disconnect: Flip the lock back to disengage the lock from the connector. Grasp the connector and remove it from the signal processor. NOTE: Do not discard the INT-100 intermediate cable. 4

Proper Connection and Disconnection User s Guide to Proper Connection and Disconnection of Sensors and INT-100 Intermediate Cable from the SenSmart Signal Processor To ensure proper sensor or intermediate cable function and to avoid damage to SenSmart Signal Processors, follow these guidelines. Note that in all steps little effort is required. If resistance is encountered, refrain until you have checked proper procedure. For proper connection, see the images below Ensure the sensor lock is completely open. Firmly grasp the Signal Processor and Sensor or Intermediate Cable with the white Nonin logo facing you. Ensure the white alignment arrow on the connector lines up directly across from the white alignment arrow on the Signal Processor. Insert the sensor connector firmly into the Signal Processor port until it stops and gently close the sensor lock. CAUTION! Do not attempt to connect at an angle. Sensor Lock Alignment Arrows Finished connection with closed sensor lock 5

Proper Connection and Disconnection For proper disconnection, see the image below Firmly grasp the Signal Processor and open the sensor lock. Firmly grasp the sensor connector and extract the sensor connector. Gently close the lock and properly clean and stow the Signal Processor. CAUTION! Do not attempt to disconnect at an angle. 6

4CA 4CA Sensor Application WARNING! Refer to the sensor IFU for complete instructions for use, including potential risks to patient and user. Cerebral/Somatic Sensor Application 4CA 1. Cerebral Site. Select the site(s) on the patient s forehead lateral of the superior sagittal sinus, superior to the eyebrow and inferior to the hairline (see cerebral examples above). The area(s) should be free of hair or surface blemishes such as moles or freckles. Avoid placing the sensor(s) over nevi, sinus cavities, hematomas, or arteriovenous malformations. 2. Somatic Site(s). Select the site(s) that provides optimal access to desired tissue. 3. Skin Preparation. Gently cleanse the patient s skin to remove oils, makeup, or soil that might interfere with adhesive or block light. Ensure the skin is thoroughly dried. 4. Sensor Placement. Remove the protective backing from the sensor pad and gently, but firmly, place the sensor(s) on the desired site(s) (see examples above). Ensure sensor surface adheres to the skin to prevent emitted light from traveling along the surface between emitters/detectors, or ambient light from entering. 8100SS, 8100SM and 8100SL Reusable Pulse Oximeter Sensor Application 1. Insert the selected digit into the sensor as shown. The patient s digit must reach the end of the sensor. 2. Direct the cable along the patient s digit, parallel to the arm or leg. 3. Secure the sensor cable with medical tape so the cable does not become caught on nearby equipment. Ensure that the tape securing the cable does not restrict blood flow or pull the sensor out of position. 7

System Configuration Screens and Menus The System has four (4) operating menus to configure system settings accessed via the Menu button and navigated using the Navigation buttons on the left side of the Monitor. Up / Down and Left / Right : In menus, used to navigate between items. Select P to Select item to edit and Save entered or changed values. Monitoring Screen Navigation: Up / Down : Change rso 2 trend line timescale. Left / Right : Scroll back/forward in time in the current case. SENSOR SITE: To select, customize, or clear a Sensor Site name. rso 2 BASELINE: To set manually as a specific value. 1. Select the Baseline field. 2. Press Up/Down to manually set Baseline value, and press Select to save and exit to Settings menu. SET HIGH/LOW ALARM LIMITS 1. Connect a sensor to the Signal Processor. 2. Press Left/Right and Up/Down to move to and Select the desired channel and the alarm limit setting; press Select to save the setting. GRAPH POSITION 1. Set location of individual channel trend lines (locations A F, or Off) or plethysmograms (On or Off). PRESETS: To activate, create, or rename a Preset: 1. Use Settings and System menus to set all limits and settings to desired values. 2. Select Presets. 3. Highlight Save Current Settings as New Preset. 4. Press Select to display Presets 1-10 in pop-up menu. 5. Press Up/Down to navigate to desired Preset and Select to overwrite. NOTE: If the Preset is locked, the message Cannot overwrite locked preset! displays. 6. Enter Preset name (maximum of 11 alphanumeric characters) using the alphanumeric keyboard. 7. Highlight Save, and Select to save and activate Preset. 8

System Configuration Screens and Menus START NEW CASE and PATIENT ID: 1. Highlight Start new case and press Select; Start new case? pop-up displays with No highlighted. 2. Highlight Yes and press Select. 3. If system is set up to enter a patient ID at the start of a new case: Enter ID using alphanumeric keyboard and press Select. 4. Monitor returns to monitoring screen and all baselines from the previous case are cleared. 5. If system is not set up to enter a patient ID at the start of a new case: Starting new case... displays. 6. Monitor returns to monitoring screen and all baselines from the previous case are cleared. The case will not have a patient ID. All System settings can be changed and saved in this menu. PRESET SYSTEM SETTINGS: These System settings can be included in a PRESET by changing the settings before creating the PRESET: Brightness, Alarm Volume, rso 2 Low Alarm Mode, Pulse Tone Volume and Source, and Data Output Modes. CLEAR MEMORY: Deletes all patient data recordings from Monitor. 9

High and Medium Priority Alarms The System has audible and visual alarm indicators to alert the user when immediate patient attention may be required or an equipment alarm occurs. WARNING! Ensure all alarm volumes are set appropriately and are audible in all situations. Do not cover or otherwise block any speaker opening. HIGH PRIORITY ALARMS: Require immediate attention and include high and low rso 2 alarms. MEDIUM PRIORITY ALARMS: Signal potential problems with the equipment or other non-life-threatening situations. HIGH Priority Alarm Visual Indicator Audible Indicator rso 2 Limit (rso 2 high alarm limit OR rso 2 low alarm limit) SpO 2 Limit (SpO 2 high alarm limit OR SpO 2 low alarm limit) Pulse Limit SpO 2 channel only (Pulse is high alarm limit OR pulse low alarm limit) Critical Low Battery Low Perfusion displays when the system detects low perfusion at the SpO 2 sensor site. rso 2 background flashes in RED at 2 times per second. Channel text becomes white. SpO 2 portion of channel background flashes RED 2 times per second; SpO 2 value becomes white. Pulse portion of channel background flashes RED 2 times per second; Pulse rate value becomes white. Battery indicator second. 10 flashes RED twice every Channel background flashes RED 2 times per second. Channel text becomes white. 3 beeps, pause, 2 beeps, pause, 3 beeps, pause, 2 beeps, and a 6-second pause. This cycle repeats until silenced or the alarm condition is cleared. MEDIUM Priority Alarm Visual Indicator Audible Indicator rso 2 Warning (rso 2 5% or less above low rso 2 alarm limit) Low Battery Sensor Fault Poor Signal Signal Processor Communication Error Error Codes (E01, E02, E03, E04, E06, E08) rso 2 background flashes YELLOW once every 2 seconds when the rso 2 value is within 5 points of the lower alarm limit. Channel text turns gray. Battery indicator flashes YELLOW once every 2 seconds. Sensor Fault indicator flashes YELLOW once every 2 seconds when a sensor is disconnected, sensor has failed, or is not compatible with the Monitor. ACTION: Check sensor site and connection. Replace sensor if necessary. Poor Signal indicator flashes YELLOW once every 2 seconds when there has been a sustained period of poor patient signals from the sensor. ACTION: Check the sensor site and reposition or replace the sensor if necessary. Communication Lost indicator flashes YELLOW once every 2 seconds and X-100SP not connected displays when the respective Signal Processor has stopped communicating with the monitor. ACTION: Check the signal processor connections or replace the signal processor to correct the issue. Indicated by error code across the screen. ACTION: Contact Nonin Technical Service at 1-800-356-8874, +1 (763) 553-9968 or +31 (0)13-79 99 040. 3 beeps followed by a 20-second pause. This cycle repeats until silenced or the alarm condition is cleared. Loud, two-tone steadily beeping signal.