SERVICE PROCEDURES. Mattress 2 Pipe Overlay or Replacement

Similar documents
DIRECTHEALTHCAREGROUP.COM. Air Suresse. Service Manual

SERVICE PROCEDURES. Control Unit 220DIO 4 pipe

Service Manual DIRECTHEALTHCARESERVICES.CO.UK

User Manual standard for Medical Electrical. Control Number 95CG. Classified by. Underwriter s Laboratory Inc. for UL

USER INSTRUCTION MANUAL SAPPHIRE SERIES 800

Dyna-Form Air. Alternating Pressure Relief Overlay System

USER INSTRUCTION MANUAL SAPPHIRE SERIES 1100 & 1100EC

Anti-Decubitus Air Alternating Pressure Mattress System Operation Manual

AMBIENCE Low Air Loss Mattress Replacement System

User Instruction Manual

PRELUDE. User Manual. PA3500 Mattress Overlay System. Guarantees and Warranty. 1.1 General Information 1.2 Intended Use. 3.1 Unpacking 3.

English 1 AIR ALTERNATING PRESSURE MATTRESS REPLACEMENT SYSTEM

Iron - low. - medium - high Do not iron. Bleach - chlorine. Do not bleach. Tumble dry - low - medium - high Do not tumble dry

User Instruction Manual

User Instruction Manual

Baros Comfort Turn Mattress

DELUXE ALTERNATING PRESSURE SYSTEM USER GUIDE

Med-Aire 8 Alternating Pressure Mattress Replacement System With Low Air Loss OPERATOR S MANUAL

Operating Instructions. Transair Alternating Pressure Care Overlay System

Delta Plus. PA8900 Bariatric Mattress Replacement System. User Manual

PUMP. General User/ Safety Guide HARVEST 1 PUMP. E

User Instruction Manual

OPAL ACTIVE OVERLAY SYSTEM

User Instruction Manual

User s Manual. Alternating Air Mattress Model No.: MA-6100

WINDSOR ACTIVE REPLACEMENT

Enhance DX. Mattress Replacement System. User Manual

User Manual. Invacare APM Model CG9900 Mattress

AIR EXPRESS LAL. P1500 LOW AIR LOSS THERAPY SYSTEM REF C1500 Control Unit REF M1500 Mattress. Operator s Manual P/N /03

Operating Instructions. Transair Alternating Pressure Care Mattress Replacement System

USER MANUAL. Safety Precautions. Unit Specifications and Features. Operating Instructions. Installation Instructions.

Med-Aire Assure 5 Air + 3 Foam Base Alternating Pressure and Low Air Loss Mattress System

APH166 Morpheus Air USER MANUAL. Issue 3-24/05/2017. Apollo Healthcare Technologies Ltd. Holme Street, Liversedge, West Yorkshire, WF15 6JF.

MULTI-FUNCTIONAL PORTABLE ROTATION SYSTEM P4000 P4001 C4000 Series Control Unit M4000 Series Mattress. Operator s Manual P/N Rev B 10/10

CRISTAL Built-in Dish Sterilizer Instruction Manual

GDDEM10 DEHUMIDIFIER

- 2 - IMPORTANT SAFETY REMINDERS IT IS IMPORTANT THAT YOU:

INSTRUCTIONS FOR SAFE USE INSTRUCTIONS FOR SAFE USE

Bosch Split-Type Ductless Air Conditioner / Heat Pump

PRESSURE WASHER MODEL NO: JETSTAR 1750 OPERATION & MAINTENANCE INSTRUCTIONS. WARNING Read the instructions before using the machine PART NO:

Midi Dehumidifi er IMPORTANT! KEEP FOR FUTURE REFERENCE. Instruction Manual. Please register your product on-line at

Enteral Pump Instructions for Use To Assemble When the low

User Manual CuroCell S.A.M. PRO Static Antidecubitus Mattress. Item number user manual: EN0000

Mobile Air Conditioner Instruction Manual Model TC-N9KM

G-7s. Instruction Manual. G-Series Cooler COUNTERTOP COOLER. Part No.11IPA

Table of Contents. English

Important Safeguards. Topic 2 SAVE THESE INSTRUCTIONS - FOR HOUSEHOLD USE ONLY SAVE THESE INSTRUCTIONS - FOR HOUSEHOLD USE ONLY 3

User Manual CuroCell AUTO420 Alternating pressure mattress. Item no. user manual:

Use r manual Edition MG Tube Master Anti-bedsore alternating pressure mattress of tubular structure VCM502D

USER MANUAL. FairWind FWPC6L Pressure Cooker. Contact Us:

Multi Function Food Processor

IMPORTANT: RETAIN FOR FUTURE REFERENCE

User Manual HEAT PUMP DRYER

High-Pressure Washer BY02-VBP-WTR BY02-VBP-WTH. CAUTION : Do not use this device before reading the instructions.

ELECTRIC BLANKET INSTRUCTIONS

One-Touch Dispense. Multi-temp selections. 208 F For tea, coffee, instant noodle. 194 F Keep warm around 194 F. 176 F Keep warm around 176 F.

LAKELAND LUXURY FLEECE HEATED MATTRESS PROTECTOR

LED Bottle Cooler. Maintenance for. I1457 Iss. 4 Jan 2014 Copyright Glen Dimplex Professional Appliances

10L Dehumidifier. Model Number: LDH V AC 50/60Hz 290W

2 000W CYCLONIC BAGLESS VACUUM

INSTRUCTIONS Français 2 Nederlands 15 English 28 Italien

Mason Air AS8800/ AS8800-RR/AS Instruction Manual. Low air loss/alternating pressure mattress

ASPIRE Laboratory Aspirator

Utopian Split A/C. Thank you for purchasing this quality Split A/C system!

MARSHALLTOWN MEDICAL AND SURGICAL CENTER ENVIRONMENTAL SERVICES. SUBJECT: Room Cleaning Occupied Room (All Areas) PROCEDURE NO: 032

Washing Machine. User manual WW9*K5*****/WW8*K5*****/WW7*K5***** Untitled :17:40

Infection Control Manual Residential Care Part 3 Infection Control Standard IC9: Environmental

#8 STAINLESS STEEL ELECTRIC MEAT GRINDER OWNER S MANUAL

Local air conditioner

DEHUMIDIFIER OWNER S MANUAL. Read this manual carefully before operating the appliance and retain it for future reference.

Instructions for Installation and Use

SPECIFICATION, INSTALLATION AND OPERATION MANUAL. Heat Lamp Assemblies. Models: W.HLA900, W.HLA1200 & W.HLA1500

Steam Station. Instruction Manual THE SCIENCE BEHIND CLEAN.

NEXXOS INVERTER USER MANUAL

INSTRUCTION BOOKLET. Suits all KFA837 models

1/2011 V1. Puzer AINO USER MANUAL. Read this instruction before using the vacuum cleaner! Keep the instructions near the vacuum cleaner for later use!

User manual CuroCell S.A.M. Static Antidecubitus Mattress. User manual item number:

Convertible Powered/ Non-Powered Dynamic Mattress System. Operation Manual. Distributed by:

Table of Contents SAFETY FIRST. Owner s Manual. Safety Precautions Unit Specifications and Features Manual Operation (Without Remote)...

Compression therapy device & accessories Model 701A User manual

Owner s Manual SPLIT-TYPE ROOM AIR CONDITIONER

20L Dehumidifier. Model Number: LDH V AC 50/60Hz 440W

AKRON BOBATH Models 4411, 4412, 4421, 4422 USER MANUAL

VARIABLE SPEED GRINDER OWNER S MANUAL R MANUAL VERSION SERIAL #

Suits all KPF849 models

Oscillating Belt/Spindle Sander 230V

Owner s Manual. Fairwind Series All Model Numbers SPLIT-TYPE ROOM AIR CONDITIONER CS

Conception Wall Fire. Instruction Manual THE SCIENCE BEHIND CLEAN.

Unstirred Baths FLD Series. Unstirred DuoBaths. Note:

OVATION 56 EGG INCUBATORS CLEANING / SERVICE GUIDE

User Guide. Please read carefully before use. Your Hoover Guarantee. Your Hoover Guarantee. Contents. IMPORTANT Safety Reminders

12L DEHUMIDIFIER OPERATING INSTRUCTIONS

ULTRASONIC HUMIDIFIER

Simulator Intravenous Arm (Two arm and circulation pump set)

WARNING: Warns of health hazards and identifies possible risks of injury. CAUTION: Indicates possible dangers to the machine or other objects.

DIGITAL ULTRASONIC HUMIDIFIER WITH HUMIDISTAT

PS-2/ES Automated pack & tag machine with IndES fastening system

Platinum Series Electronic Panel Heaters

Outdoor Beverage Center

YOUR SAFETY & INSTRUCTION MANUAL PLEASE READ CAREFULLY HAIR DRYER MODEL C80002

Transcription:

Mattress 2 Pipe Overlay or Replacement

Contents 2 pipe Mattress Mattress 3 Air Leak Test 4 Air Cell Replacement 5 Air Pipe Connector Replacement 6 Troubleshooting 7 Safety 9 Care & Cleaning 10 Definition of Symbols Used Caution: Warning of possible hazard to system, patient or staff. Warning of possible electrical hazard. Important operational information. 2

2 pipe Mattress Unless the problem is known to be a fault with the Mattress Overlay or Mattress Replacement, begin the service on the Control Unit. Should the system continue to lose pressure once all Control Unit service procedures have been successfully completed, follow these procedures to check the Mattress Overlay or Mattress Replacement for potential damage and make the relevant repair. Always complete Basic System Checks before commencing any Mattress service procedure. It is recommended that any returned Mattress be thoroughly cleaned before commencing service (refer to Care and Cleaning) and that protective gloves and gown are worn throughout the procedure. Any fault or damage to the product that is not discussed in these Service Procedures is deemed a non serviceable issue. The faulty product must be returned to the Manufacturer for repair or replacement. Please note that this is a generic service manual which is applicable for several different models. Thus the products used in the pictures may differ from the product shown below. 3

SERVICE PROCEDURES Air Leak Test 2 pipe Mattress Testing for a leak in the Air Cells Insufficient system pressure will be identified by the visible and audible Alarm functions and flashing Pressure Setting LEDs. One of the Mattress Air Cells may be punctured or damage, causing air to leak from the system. Follow the steps below to determine if an Air Cell is leaking. The following tools are required: Control Unit Plastic sheeting (or other protective material) 1 2 3 5 6 7 Ensure the Mattress has been correctly set up following the procedures outlined in the User Manual provided with the system. Sand bags (or other weighted device) 4 Turn off the Control Unit. Disconnect the Air Pipes from the Control Unit and quickly connect the male and female Air Pipe connectors together to seal air inside the Mattress (i.e. Transport Mode). Replace the Mattress Top Cover to protect the individual Air Cells from damage. Cover the Mattress with plastic sheeting or other protective material to protect the Top Cover from soil or damage during testing. Remove the Mattress Top Cover. Check all CPR Tag sealing connectors are pushed firmly on to the Mattress air pipes. Check each Air Cell is firmly connected to each internal air pipe connector. Using sand bags or other soft weighted device, spread sufficient weight evenly along the surface of the Mattress to simulate a patient lying on the system. Do not exceed the maximum weight limit of the System. Connect the Control Unit and wait the prescribed time for complete Mattress inflation. When initial inflation is complete, the system automatically defaults to Alternating Mode. Switch to Static Mode and wait a further 10 minutes to ensure all cells are inflated to maximum pressure. Leave the weight on the Mattress for a minimum of one hour (in case of slow leak). 8 After one hour, remove the protective sheeting and remove the Mattress Top Cover. Manually feel each Air Cell to check inflation. Any partially inflated cell should be removed and replaced. Go to Air Cell Replacement procedure for next steps. 4

Air Cell Replacement 2 pipe Mattress Replacing a leaking Air Cell Having completed the Air Leak test detailed previously (steps 1 through 8), replace all suspected leaking or damaged Air Cell/s. No tools are required for this procedure. The following replacement parts are required: Air Cell/s 9 Deflate the Mattress by pulling the CPR Tag and applying pressure on the Air Cells to release the air. 10 To disconnect a faulty Air Cell, detach the press stud (connecting the individual Air Cell to the Mattress base). Please note, Mattress Replacements have two press studs at each end of the cell. 11 Gently pull and twist the internal air pipe to disconnect it from the cell. Some strength may be required to free the air pipe from the connector. Please note, Mattress Replacements have two internal air pipes, one connected to the upper alternating cell and one to the lower static cell. Both air pipes need to be removed. 12 Carefully remove the Air Cell by pulling it through the material positioning straps. 13 Unpack the replacement Air Cell and feed back through the material positioning straps in the same alignment as before. 14 Reconnect the Air Cell air pipe by pushing firmly until the pipe fully covers the connector. Reconnect the press stud fastening the cell to the base. For the Mattress Replacement, replace both air pipes and both press studs. 15 Repeat for all suspected leaking or damaged Air Cells. Repeat the Air Leak Test to ensure all Air Cells are maintaining maximum pressure and there are no leaks or loose air pipe connections. 5

Air Pipe Connector Replacement 2 pipe Mattress Replacing a damaged or faulty Air Pipe connector On occasion, the male and/or female connectors that attach the dual Air Pipes to the Control Unit will be damaged or will break. These plastic connectors should be replaced by following the steps below. As part of any routine Mattress service, always visually inspect the dual Air Pipes that connect the Mattress to the Control Unit for any splits, breaks or cuts. If either Air Pipe is damaged, the Mattress must be returned to the Manufacturer for further repair or replacement. The following tools are required: Cutter and/or sharp knife Heat gun The following replacement parts are required: Male pipe connector Female pipe connector Cable tie Heat shrink tubing (30 mm) 1 Firmly hold the end of the damaged Air Pipe connector just below the black sheath covering the cable tie that secures the connector and air pipe. 2 Using Wire Clippers, cut through the sheath to severe the cable tie, at the visible protrusion of the cable tie grip. Completely remove the sheath and the cable tie (carefully using the clippers where necessary) and discard. 3 Remove the damaged connector from the air pipe and discard. 4 Unpack the replacement sheath and connector (male or female as required). Thread the sheath over the air pipe and hold it away from the pipe end while inserting the new connector. Press firmly until the air pipe sits firm against the central ridge of the connector. 5 Fit a replacement cable tie over the end of the black pipe (position approximately 5mm from the edge). Thread the tie and pull firmly before removing excess length. 6 Pull the sheath up and over the cable tie, bringing it flush with the end of the pipe. Use a Heat Gun to shrink the plastic sheath and secure it firmly in place. Do not hold the Heat Gun too close to the air pipe. 7 Repeat the process for the alternative Air Pipe connector as required. Connect and disconnect the dual Air Pipes to the Control Unit to ensure the connectors are secure and the correct male and/or female connectors have been installed. 6

Troubleshooting The following section outlines a list of common troubleshooting issues, together with recommended initial checks and subsequent Technical Service steps and procedures to follow if there is a suspected fault in the System. If a repair seems unmanageable, the issue falls outside the scope of serviceable repairs or the problem persists following the recommended service checks and repair procedures, please return the faulty System (Control Unit; Mattress; or both) to the Manufacturer. Problem Cause Initial Checks Technical Service Procedure Control Unit does not operate; no display lights illuminate. Control Unit operational but red Alarm LED illuminates (Mattress fails to reach or drops below accepted minimum level). Mattress fails to inflate. The Control Unit is grinding or making an unusual noise. The Mattress is inflated but static head cells are still deflated. The Control Unit may not be attached to a power source. The power cord may be damaged. A fuse may need replacing. The system may have an air leak. The Air Filters may need cleaning or replacing. The Synchronous Motor (timing motor) may be damaged. System may need more time to inflate air pressure must build in alternating section before static head cells inflate. CPR Tag is not fully connected. Valves to static head cells may be damaged or cracked. 1. Check the Control Unit is connected to mains power outlet with the correct voltage. 2. Check the Control Unit is switched on. Switch off and disconnect the unit before restarting. 3. Check the Control Unit with a spare power cord (in known working condition). 1. Check both Air Pipes are securely attached to the Control Unit. 2. Check the CPR Tag is intact, ensuring all sealing connectors are fully fitted to the Mattress air pipes. 3. Check all pipes along the inside of the Mattress each should be firmly connected. 4. Check each Air Cell is securely attached to the connecting air pipe. 5. Check all cells, pipes and pipes for any air leakage. 6. Check air filter covers are secure and air filters are clean. 1. Check system has had sufficient time for initial inflation (40 to 50 minutes). 2. Check the CPR Tag is intact, ensuring all sealing connectors are fully fitted to the Mattress air pipes. 3. Visually check the internal Mattress air pipes that run from the CPR Tag to the static head cells for splits, cracks or other damage. Control Unit Fuse Check and Replacement Control Unit Printed Circuit Board Check and Replacement Control Unit Air Filter Maintenance Control Unit Compressor Check and Replacement Control Unit Synchronous Motor Replacement Return System to Manufacturer for further repairs or replacement. 7

Troubleshooting Problem Cause Initial Checks Technical Service Procedure The Mattress is not alternating. Patient is sinking or bottoming out while lying flat on Mattress. Static or Max Firm modes may be activated. The pressure may be set too low for the patient s weight. The system may be losing pressure. 1. Switch the Control Unit off and disconnect the power before restarting. Ensure Alternating Mode is selected. 1. Increase Pressure Setting an incremental increase is usually sufficient but wait 12 minutes (one full cycle) before checking. 2. Perform bottoming out test: a. With patient lying supine, fold back one side of the Mattress Top Cover just past sacral region (lower spine). b. Slide hand underneath the patient and feel for a deflated cell under the patient s lower spine. Any inner static cell will remain inflated however your hand should easily slide between patient and base. c. If patient is adequately suspended, pressure setting can be lowered however this test should be repeated after approximately 20 minutes. Synchronous Motor Replacement Mattress Service Procedures 8

SAFETY Precautions General Safety Precautions Do not use this equipment in the presence of flammable anesthetics. Explosions could result. The air intakes for the Control Units are at the sides of the unit. To avoid the risk of blocking these intakes, it is recommended that the unit be mounted on the footboard of a bed frame. Should the bed not have a suitable footboard, it is recommended that the unit be placed on a solid surface such as a table, or on the floor. Protection Against Hazards Fluids Avoid spilling fluids on any part of the Control Unit. If spills do occur: Disconnect the unit from the mains wall socket. Clean fluids from the case. Ensure that there is no moisture in or near the power inlet, power switch and power plug before reconnecting the power supply. Check the operation of controls and other components in the area of the spill. Perform applicable checkout procedures. Liquids remaining on the electronic controls can cause corrosion that may cause the electronic components to fail. Component failures may cause the unit to operate erratically, possibly producing hazards to patient and staff. Disposal Dispose of the Air Filter, Air Cells, Foam inserts and/or Mattress Top Cover according to local procedures and regulations. At the end of useful life, dispose of waste according to the European Union Waste Electrical and Electronic Equipment (WEEE) Directive and in compliance with relevant local regulations. Power Cord The system should never be operated with a worn or damaged power cord. Should the power cord be found to be worn or damaged, replace immediately and dispose of the damaged cord. Interference Although this equipment conforms to the intent of the Directive IEC 60601-1-2 in relation to Electromagnetic Compatibility, all electrical equipment may produce interference. If interference is suspected, move equipment from sensitive devices or contact the Manufacturer. (IEC 60601-1-2. Medical Electrical Equipment Part 1: General Requirements for Safety, Amendment No. 2. Collateral Standard: Electromagnetic Compatibility Requirements and Tests). 9

CARE & CLEANING Recommendations General Recommendations Follow the Manufacturer s instructions as detailed below and in the applicable User Manual. Failure to do so may result in cross contamination or equipment damage. To prevent cross contamination, the Manufacturer recommends that the system be cleaned and laundered between patients according to the instructions below. Always disconnect the Control Unit from mains power before cleaning. Do not spray disinfectant directly on to the unit, or immerse the unit in any type of liquid. Failure do so could result in equipment damage and / or electric shock. Do not use high temperature autoclave steam cleaning devices or phenolic based cleaning products. Use of either of these items could result in damage to equipment and/or loss of waterproof qualities of the Top Cover. The System may be cleaned according to local protocols and regulations/procedures for blood borne pathogens provided the Manufacturer s instructions are followed. Refer to the Top Cover wash tag for detailed cleaning instructions. Cleaning While System in Use 1 2 3 4 5 6 The System should be cleaned every two weeks if in constant use. Allow all parts to dry thoroughly before returning to use. Remove patient from the System. Unplug the Control Unit from mains power. Remove the Top Cover. Using a well wrung out cloth dipped in warm soapy water, wipe the surface of the Top Cover. Rinse with plain water and allow to dry. Pull the CPR Tag and disconnect the Air Pipes from the Control Unit to allow the Mattress to deflate completely. Where applicable, remove foam inserts from the underside pocket. Inspect the interior of the Mattress for any signs of fluid contamination. 7 8 9 If the interior of the Mattress shows signs of fluid contamination, inspect the Top Cover for punctures or damage. If the Top Cover is damaged it should be replaced before returning the system to use. If necessary for inspection or cleaning, the Air Cells may be removed by carefully disconnecting the air pipes and unsnapping the cell. Using a well wrung out cloth dipped in warm soapy water, wipe the Mattress base, cells and plastic pockets housing the foam inserts (where applicable). Rinse with plain water and allow to dry. Using a well wrung out cloth dipped in warm soapy water, wipe the Control Unit and Air Pipes. Do not allow fluid to penetrate the Control Unit. If the Top Cover or Air Cells have become excessively soiled they may be laundered in a washing machine at up to 95 C using normal domestic washing powder. Do not add bleach to the wash cycle. Rinse well using plain water and dry thoroughly before use. Do not dry the Top Cover using the heat cycle or a dryer. Air dry or select a low of non-heat dry cycle. 10

Cleaning and Maintenance Between Patients Inspect all parts for damage and replace as necessary before returning to service. Do not return the Control Unit to service without replacing air filters. Do not allow fluid to penetrate the Control Unit. 1 2 3 4 CARE & CLEANING Recommendations The Mattress should be dismantled according to the instructions on the previous page. Launder the Top Cover and cells as described on the previous page. Using warm soapy water (or sodium hypochlorite solution at 10,000ppm) and a well wrung cloth, clean the Mattress, Control Unit, Air Pipes and plastic pockets housing the foam inserts (where applicable). Rinse with plain water and allow to dry thoroughly. Replace air filters (see the relevant Air Filter Maintenance procedure for details). 11

CONTACT INFORMATION Aeda Healthcare Limited 15th Floor 100 Queen s Road Central Central Hong Kong Tel: + 852 6832 5199 Fax: + 852 6832 5373 Email: aeda@aeda.com http://www.aeda.com 12