Easidew PRO I.S. Process Dewpoint Transmitter User s Manual

Similar documents
DB5 Intrinsically Safe Sounder Type DB-5

INSTRUCTION MANUAL IS-mB1 Minialite Intrinsically Safe Round LED Beacon

Rosemount 3490 Series 4 20 ma + HART Compatible Controller

Model 144H. Product Data Sheet , Rev DB February Content

4) Intrinsic Safety Certification

Dew-Point Transmitters and Instruments

VERSAFLOW VORTEX Supplementary instructions

OilSET Installation and Operating Instructions. Oil Separator Alarm Device with SET/DM3AL sensor

Technical Data. General specifications Switching element function Rated operating distance s n 8 mm

Mobrey MSP900SH Level and MSP900FH Flow Ultrasonic Transmitters

RADAR LEVEL GAUGE SPECIAL SAFETY INSTRUCTION

Intrinsically Safe, NEMA 4, Type 4X, IP67 Loop-Powered Meter

Technical Data. General specifications Switching element function Rated operating distance s n 5 mm

Moisture and Hydrocarbon Dew-Point Measurement

E2S Datasheet v10a. The IS-mA1 is suitable for all intrinsically safe signalling applications including fire, security and process control.

Inductive slot sensor

Installation Guide. EXGP Oxygen Analyzer System (ATEX) Interface Electronics Unit

Model 144H. Product Data Sheet , Rev DB February Content

Promet I.S. Promet I.S. Process Moisture Analyser

Benefits & Technology for Decentralizing a -40 F Dew Point

Annubar Flowmeter Series

MICROpeL 75 Combustible Gas Sensor Part Number: PM

SITRANS T. SITRANS T Explosion protected temperature sensor. General information. Marking of the degree of protection. Range of uses.

Rosemount 4500 Hygienic Pressure Transmitter. Start

INSTRUCTION MANUAL (ATEX)

DSF xx15.xx xhv Ex-Atex Ex-Atex Certified Two wire Hall Effect Speed Sensor

ExDetector HC-100. Data Sheet. Gas Measuring and Alarm Systems

Certification Supplement Manual

805QS Pressure Switch-Transmitter

Technical Data. Dimensions

Rosemount 148 Temperature Transmitter. Quick Start Guide , Rev GA September 2016

IR1xxx Series 1, IR2xxx Series 1 Miniature Infrared Gas Sensors for Hazardous Areas and Intrinsic Safety in Mining

PCMini52 Digital RH & Temperature Mini Probe User s Manual

INSTRUCTION MANUAL (ATEX / IECEx)

RT FLOW RATE TOTALISER

RTK Instruments Limited St James Business Park, Knaresborough, North Yorkshire, England. HG5 8PJ

INSTRUCTION MANUAL. SIL 3 - SIL 2 IIB Group Power Supply for Hazardous Area Equipment DIN-Rail Model PSD1001C PSD1001C

INSTRUCTION & SERVICE MANUAL

Hazardous Area Zirconia Oxygen Systems

Rosemount 2051T In-Line Pressure Transmitter

IECEx Certificate of Conformity

HT-1300 Series. Temperature and Humidity Duct Sensor. Product Bulletin. Power Supply VDC / 24 VAC. Humidity Accuracy 2.5% RH from 10 to 95% RH

INSTRUCTION MANUAL. E2xB05 & E2xB10 Xenon Beacons For use in Hazardous Locations

Technical Data. General specifications Switching element function DC Dual NC Rated operating distance s n 3 mm

Tank Gauging System Special Safety Instruction

Rosemount 3490 Series

Rosemount 4500 Hygienic Pressure Transmitter

SET Installation and Operating Instructions. Level switch for one sensor

standards: standards:

Rosemount 4500 Hygienic Pressure Transmitter Demonstrated best-in-class performance during SIP/CIP for process temperatures up to 400 F (204 C)

A Characteristics for Electrical Equipment integrated Into SLR Enclosure - Intrinsically Safe Reference Document

Intrinsically Safe Pressure Transmitters for installation in hazardous locations Models IS-20-S, IS-21-S, IS-20-F, IS-21-F WIKA Datasheet IS-20

Rosemount 248 Temperature Transmitter. Quick Start Guide , Rev. EA April 2013

Safe Operation Manual. Honeywell IQ Force Gas Detector

Lamp Supply (Recommended)

Installationsmanual GWSL1100/2100, GWML3100

POINTEK CLS 100 OINTEK CLS 100 CAPACITANCE LIQUIDS/SOLIDS. Instruction Manual March 2001

MAINS KLARGESTER SEPARATOR ALARM INSTALLATION & OPERATIONS MANUAL

Barrier. ED /07/09 Release to production R.P. A ECO N DATE DESCRIPTION OF CHANGE CHG By ISSUE TITLE TC842B1007 AE PRODUCT SPECIFICATION

Ar-Mite TM Low Flow Armored Flowmeter

Rosemount 3100 Series Ultrasonic Level Transmitters

Technical Data. General specifications Switching element function Rated operating distance s n 4 mm

FUNCTIONAL SAFETY MANUAL. Gassonic Observer-H and Observer-i Ultrasonic Gas Leak Detector

1510 Side Mounted Level Switch

Global Series. Installation and Maintenance Instructions. Model G-STR Strobe Beacon. For Use in Hazardous Locations

OPTIWAVE X500 Supplementary Instructions

Intrinsically Safe Transmitters, Option Code 2n

Additional Operating Instructions SITRANS F. Vortex flowmeters. SITRANS FX330 Ex-i.

Druck DPI 705/DPI 705 IS Series

Dependable Solutions for Gas Detection

MANUAL Oil Level Sensor

IR15TT-R Miniature Infrared Gas Sensor for Monitoring Carbon Dioxide and Methane up to 100% Vol.

Certificate of Compliance

FUNCTIONAL SAFETY MANUAL

OPTIBAR DP 7060 Supplementary Instructions

Rosemount 3900/3900VP

Using the DewTrak II Dew Point/ Humidity Transmitter

SmartPower Solutions [ SmartPower Solutions. Quick Installation Guide , Rev AC January 2012

Additional Operating Instructions SITRANS F. Vortex flowmeters. SITRANS FX330 Ex-nA.

Technical Data. General specifications Switching element function Rated operating distance s n 5 mm

CE110I. Piezoelectric accelerometer with integrated electronics. IECEx

YTA Series Temperature Transmitters [Style: S3] EN :2006, -11:2007, -26:2007. Supply circuit: Ui=30 V, Ii=165 ma,

Using the Model 1500 Portable Chilled Mirror Hygrometer

Operating Instructions

14300 MAINS SEPARATOR ALARM MANUAL

Certification Supplement Manual

E2S Datasheet v10a

Introduction Portable Meters Transmitter Probes RELATIVE HUMIDITY

SET-2000 Oil/Sludge 12 VDC

Hygro and Hygrothermo Transducers (capacitive) for climatic applications

Pressure sensors. Ex ia I / IIC T6 acc. to ATEX. Features. Description. Measuring ranges. Applications

Technical Data. Dimensions

icountms Range Fluid Condition Monitoring Moisture Sensors

MANUAL Overflow sensor NVF-104/34-PF

Certificate of Compliance

OPTIBAR P 1010/2010 C Supplementary instructions

Dräger Polytron 7000 Fixed gas detector

Software Version 2.01 LEVEL MONITOR MODEL 220

Intrinsically Safe Flame Detector Installation Guide

Hygro/hygrothermal transducers (hygrometric)

Transcription:

Easidew PRO I.S. Process Dewpoint Transmitter User s Manual KAHN INSTRUMENTS, INC. 885 Wells Road, Wethersfield, CT 06109 Phone: 860-529-8643 Fax: 860-529-1895 E-mail: info@kahn.com www.kahn.com

Inside front cover (blank)

Kahn Instruments, Inc

TABLE OF CONTENTS Safety...3 Pressure Safety...3 Toxic materials...3 Repair and Maintenance...3 Calibration...3 Abbreviations...4 1. INTRODUCTION... 5 2. ABOUT EASIDEW PRO I.S. PROCESS DEWPOINT TRANSMITTER... 5 3. FACTORY CALIBRATION... 5 4. PREPARATION FOR USE... 6 5. TRANSMITTER CABLE... 7 5.1 Preparation of the Sensor Cable... 7 6. INSTALLATION IN HAZARDOUS AREAS... 8 7. OPERATION... 9 7.1 Measurement Range and Analog Output... 9 7.2 Dry Down... 9 7.3 Sampling Hints... 9 8. MAINTENANCE... 10 9. SINTERED GUARD... 10 10. HAZARDOUS AREA CERTIFICATION... 11 10.1 Terminal Parameters...11 10.2 Special Conditions of Use... 11 11. EASIDEW PRO I.S. DIMENSIONAL DRAWING... 12 12. CERTIFICATES... 13 12.1 Baseefa Approved System Drawing... 14 12.2 Baseefa Certificates... 14 12.3 IECEx Certificates... 25 12.4 FM Approved System Drawing... 29 12.5 FM Approvals Certificates...30 12.6 CSA Approved System Drawing... 33 12.7 CSA Certifi cate of Compliance... 34 12.8 EC Declaration of Conformity... 37 13. TECHNICAL SPECIFICATIONS... 38 13.1 Fault Conditions... 39 Issue 12, January 2013 1

LIST OF FIGURES Figure 4.1 Wall-mount bracket...6 Figure 5.1 Bare wires...7 Figure 5.2 Crimped wires...7 Figure 5.3 Cut to 5mm...7 Figure 5.4 Connection to Easidew PRO I.S....7 Figure 5.5 Hazardous Area Connection....8 Figure 7.1 Installation location...9 Figure 7.2 Indication of dead space...9 Issue 12, January 2013 2

CAUTION Use of any electrical device in a Class 1 Division 1 environment requires proper use to minimize risk of explosion and personnel injury or death. Use of this Easidew PRO I.S. transmitter in a Class 1 Division 1 environment requires that the user follow all instructions herein as well as all applicable National Electrical Code and National Fire Protection Association (NFPA) requirements for this and any other device installed in the same environment.

SAFETY The manufacturer has designed this equipment to be safe when operated using the procedures detailed in this manual. The user must not use this equipment for any other purpose than that stated. Do not apply values greater than the maximum value stated. This manual contains operating and safety instructions, which must be followed to ensure the safe operation and to maintain the equipment in a safe condition. The safety instructions are either warnings or cautions issued to protect the user and the equipment from injury or damage. Use qualified personnel and good engineering practice for all procedures in this manual. Pressure Safety DO NOT permit pressures greater than the safe working pressure to be applied to the instrument. The specified safe working pressure is 5000 PSIG. Refer to the Specification section of this Manual. Toxic Materials The use of hazardous materials in the construction of this instrument has been minimized. During normal operation, it is not possible for the user to come into contact with any hazardous substance, which might be employed in the construction of the instrument. Care should, however, be exercised during maintenance and the disposal of certain parts. Repair and Maintenance The instrument must be maintained either by the manufacturer or an accredited service agent. Calibration The recommended calibration interval for the Easidew PRO I.S. is 12 months. The instrument should be returned to Kahn Instruments or one of their accredited service agents for re-calibration. Issue 12, January 2013 3

ABBREVIATIONS bara bar absolute barg bar gauge C degrees Celsuis F degrees Fahrenheit DC direct current μm micro-metre ma milli Ampere Mpa megapascal mw milli Watts nf nano-farad Nl/min normal litres per minute Nm Newton metre ppm(v) parts per million by volume RH relative humidity V volts Note: Abbreviations are not pluralized, they remain the same in both the singular and plural. Issue 12, January 2013 4

1. INTRODUCTION Thank you for choosing the EASIDEW PRO I.S. Process Dew-point Transmitter from Kahn Instruments. Your EASIDEW PRO I.S. has been manufactured, tested and calibrated to the highest available standards and should come to you in perfect working order, ready for installation into your gas measurement application. If you have any questions about the instrument or how to install and operate it, please contact Kahn Instruments, Inc. at: KAHN INSTRUMENTS, INC 885 Wells Road Wethersfield, CT 06109 Phone: 860-529-8643 Fax: 860-529-1895 Email: hygros@kahn.com Web: www.kahn.com 2. ABOUT EASIDEW PRO I.S. PROCESS DEWPOINT TRANSMITTER The EASIDEW PRO I.S. is a continuous, on-line, 4-20 ma transmitter for the measurement of moisture content in air and other non-corrosive gases, and is designed specifically for use within Class I, Divisions 1 and 2 (FM and CSA Certification). Its key features are: Class I, Divisions 1 and 2, group, A,B,C and D (FM and CSA certification) Zone 0, ATEX and IECEx certification Rugged weatherproof housing to NEMA 4/IP66 2-wire connection / Linear 4 20mA signal Operating pressure range - up to 45 Ched Operating range - 148 F to +68 F dew point Powered by any DC source from 12 to 28 V Output configurable for dew point or ppm(v) 3. FACTORY CALIBRATION The EASIDEW PRO I.S. is fully factory-tested and calibrated prior to delivery and is supplied with its own calibration certificate, providing direct traceability to both US National Institute of Standards and Technology (NIST) and UK National Physical Laboratory (NPL) Humidity Standards. The sensor is certified at thirteen dewpoint levels across its operating range against a certified reference hygrometer using a mass-flow humidity generator system as a source of reference calibration gas. Periodic re-calibration is recommended in order to maintain the highest quality of measurement in your application. Kahn Instruments recommends that you have your EASIDEW PRO I.S. re-calibrated annually unless it is used in a mission-critical application or in a dirty or contaminated environment, in which case the calibration interval should be reduced accordingly. Kahn Instruments can offer a variety of re-calibration and exchange transmitter programs to suit your specific needs. Kahn Instruments will be pleased to provide detailed, custom advice. Issue 12, January 2013 5

4. PREPARATION FOR USE On delivery, please check that all the following standard components are present in the packing box: EASIDEW PRO I.S. Transmitter Bonded Seal Certificate of Calibration Quantity (3) cable crimps Mounting Bracket (optional) The Easidew PRO I.S. is protected within the shipping box by the inclusion of a small red cap covering the transmitter connector and a small desiccant capsule installed inside the plastic protective transit cover. Neither of these items is required for the operation of the Easidew PRO I.S. Just prior to installation of the Easidew PRO I.S., unscrew and remove the plastic protective transit cover and retain for future use. Take care to prevent any contamination of the transmitter before installation - do not handle the sintered guard. The Easidew PRO I.S. Transmitter can be mounted either in a flow-through sample block (optional) or directly inserted into a pipe or duct. It can be operated at pressures up to 5000 PSIG when used with the bonded seal provided. NOTE: Pass the bonded seal over the 5/8-18 UNF mounting thread and assemble into the sampling location by hand, using the wrench flats only. DO NOT grip and twist the Easidew PRO I.S. cover when installing the transmitter. When installed, fully tighten using a wrench until the bonded seal is fully compressed to a torque of 22.5 foot pounds to 24.0 foot pounds. The recommended gas flow rate, when mounted in the optional sampling block, is 2 to 10 scfh/min. However, for direct insertion applications, gas flow can be from static to 30 ft/sec. The Easidew PRO I.S. can be supplied with an optional wall-mounting bracket. This allows the customer to support the transmitter physically insuring that the stress on the mounting flange is kept to a minimum. The optional bracket needs to be attached to the Easidew PRO I.S. (see Figure 4.1) using the hex screws provided. It can be installed either horizontally or vertically and can then be attached to a wall or plate to provide support for the transmitter. Figure 4.1 Wall-mount bracket Issue 12, January 2013 6

5. TRANSMITTER CABLE Cable connection to the Easidew PRO I.S. is made via the internal terminal block. 5.1 Preparation of the Sensor Cable NOTE: To comply with Hazardous Area Certification of the product, it is essential that the crimps supplied be installed onto any cable which, in turn, is installed into the connector. Figures 5.1 and 5.3, shown below, should be followed in detail. The crimps should be applied such that there is no possibility of a conductor strand of a core becoming free. Figure 5.1 Bare wires Figure 5.2 Crimped wires When the crimp is made it should have a minimum of 2 positions of crimping. After the crimp is made it should be trimmed to a length of 0.2 inches (see Figure 5.3). When the crimps are installed into the connector terminal block, insure they are fully inserted, as shown in Figure 5.4, before tightening the terminal clamping screw. Figure 5.3 Cut to 0.2 inches Figure 5.4 Connection to Easidew PRO I.S. Issue 12, January 2013 7

When all wire connections are made, insure that there is a minimum clearance distance and a minimum creepage distance in air of 0.08 inches between each terminal. CAUTION: Always connect the 4-20 ma return signal to a suitable load (see Figure 5.5) before the power is applied. Without this connection, the transmitter may be damaged if allowed to operate for prolonged periods. Electrical Connection 2 4 3 Max Load 250 Ohm @ 12V 500 Ohm @ 24V + Supply 12V Min 28V Max GALVANIC ISOLATION INTERFACE HAZARDOUS AREA SAFE AREA DEWPOINT TRANSMITTER CERTIFICATION NO S Baseefa 06ATEX0330X IECEX B AS 06.0090X FM 3030328 CSA 2013218 TRANSMITTER VERSION TERMINAL NUMBER EASIDEW I.S. 3 1 EASIDEW PRO I.S. 2 4 3 (+) (RETURN) (+) (-) INTERFACE BARRIER REFER TO TABLE ABOVE + 4/20mA LOAD +VS (20 to 35 VDC) VS - Figure 5.5 Hazardous Area Connection 6. INSTALLATION IN HAZARDOUS AREAS The EASIDEW PRO I.S. is certifi ed intrinsically safe for use in hazardous areas. For Hazardous Area Certification, see Section 10 of this manual. Before using the EASIDEW PRO I.S. in any hazardous environment, ensure that you are fully familiar with the above FM and CSA standards relating to the certifi cation of this instrument; and up-to-date codes of practice in the country of installation. The EASIDEW PRO I.S. must be installed using a specifi ed GALVANICALLY ISOLATED INTERFACE unit as shown in the system drawings on page 13 (Baseefa), page 29 (IECEx) and page 33 (CSA). NOTE: Installation of the EASIDEW PRO I.S. MUST be as per the system drawings in order to comply with the Intrinsic Safety Certification. Issue 12, January 2013 8

7. OPERATION Operation is very simple assuming the following installation techniques are adhered to: 7.1 Measurement Range and Analog Output The standard Easidew PRO I.S. is delivered with the 4 20 ma output signal set to cover the range 148 F to +68 F dew point. Please note, however, that the transmitter output range can easily be changed to cover any dew-point range or to provide an output in terms of moisture content in the range 0-3000 ppm(v). Non-standard ranges can either be programmed at the factory, or by the customer, using the Easidew Communications Kit (EA2-CK) and Easidew application software, both of which can be purchased from Kahn Instruments. The application software help file will provide detailed instructions on how to perform this task. 7.2 Dry Down If the Easidew PRO I.S. is installed into a new application, then the time taken for the sensor to dry down from ambient conditions to the operational dew-point level of the process will normally be shorter than the time taken to dry down the process itself. However, if the Easidew PRO I.S. is installed into an application which has previously been purged with dry gas, then there may be a significant time required for sensor dry down. The exact time taken will depend on a number of factors such as target dew point, construction of the sampling system (see below) and flow rate of the gas. 7.3 Sampling Hints Be Sure the Sample is Representative of the Gas Under Test The sample point should be as close to the critical measurement point as possible. Also, never sample from the bottom of a pipe (see Figure 7.1), as entrained liquids may be drawn into the sensing element. Minimize Dead Space in Sample Lines Figure 7.1 Installation location Dead space (see Figure 7.2) causes moisture entrapment points, increased system response times and measurement errors, as a result of the trapped moisture being released into the passing sample gas and causing an increase in partial vapor pressure. Figure 7.2 Indication of dead space Issue 12, January 2013 9

Remove Any Particulate Matter or Oil from the Gas Sample Particulate matter at high velocity can damage the sensing element. At low velocity, particulate matter may blind the sensing element by coating it and thereby reduce its response speed. If particulate, such as degraded desiccant, scale or rust is present in the sample gas, use an in-line filter. Use High Quality Sample Tube and Fittings Kahn Instruments recommends that, wherever possible, stainless steel tubing and fittings should be used. This is particularly important at low dew points since other materials have hygroscopic characteristics and adsorb moisture on the tube walls, slowing down response and, in extreme circumstances, giving false readings. For temporary applications, or where stainless steel tubing is not practical, use high quality thick walled PTFE tubing. 8. MAINTENANCE Routine maintenance of the Easidew PRO I.S. is confined to regular re-calibration by exposure of the Easidew PRO I.S. to sample gases of known moisture content to insure that the stated accuracy of the Easidew is maintained. Calibration services traceable to the National Institute of Standards and Technology (USA) and National Physical Laboratory (UK) are provided by Kahn Instruments. 9. SINTERED GUARD The sintered guard provides protection to the dew-point sensor, and should be replaced at regular intervals, depending on the contamination in the gas being measured. When replacing the HDPE guard, care should be taken to handle it on the lower part only to prevent contamination. Issue 12, January 2013 10

Easidew PRO I.S. User s Manual 10 Hazardous Area Certification The Easidew PRO I.S is certified compliant to the ATEX Directive (94/9/EC), and IECEx for safe use within a hazardous area and has been assessed so by Baseefa Ltd. This product conforms to the Standards: EN 60079-0:2012 EN60079-11:2012 IEC60079-0:2011 IEC60079-11:2011 and is attributed with a product certification code: II 1 G Ex ia IIC T4 Ga (-20 C Ta +70 C) ATEX Certificate Number : Baseefa 06ATEX0330X/3 ATEX System Certificate Number: Baseefa 07Y0027 IECEx Certificate Number: IECEx BAS 06.0090X The Easidew PRO I.S is also certified for use in Hazardous Areas by FM Approvals and CSA, with certification code: IS / I / 1 / ABCD / T4 Ta = +70 C, Entity Ex90385, IP66 FM Certificate Number: 3030238 CSA Certificate Number: 2013218 This product also holds GOST-R and GOST-K certificates. 10.1 Terminal Parameters Ui = 28 V li = 93 ma Pi = 651 mw Ci = 37 nf Li = 0 10.2 Special Conditions of Use The wiring connections to the free socket must be made via crimped connectors in such a way that all the strands of the wire used are held securely by the crimp. The plastic plug and socket create a potential for electrostatic discharge so must not be rubbed with a dry cloth or cleaned with solvents. The Easidew PRO I.S Dew-Point Transmitter does not withstand the 500 V AC insulation test to frame. This must be taken into account when installing the equipment. Issue 12, January 2013 11

11. EASIDEW PRO I.S. DIMENSIONAL DRAWING Issue 12, January 2013 12

12. CERTIFICATES 12.1 Baseefa Approved System Drawing Issue 12, January 2013 13

12.2 Baseefa Certificates Issue 12, January 2013 14

Issue 12, January 2013 15

Issue 12, January 2013 16

Issue 12, January 2013 17

Issue 12, January 2013 18

Issue 12, January 2013 19

Issue 12, January 2013 20

Issue 12, January 2013 21

Issue 12, January 2013 22

Issue 12, January 2013 23

Issue 12, January 2013 24

12.3 IECEx Certificates Issue 12, January 2013 25

Issue 12, January 2013 26

Issue 12, January 2013 27

Issue 12, January 2013 28

12.4 FM Approved System Drawing THE CAPACITANCE AND EITHER THE INDUCTANCE OR THE INDUCTANCE TO RESISTANCE RATIO (L/R) OF THE CABLE MUST NOT EXCEED THE FOLLOWING VALUES: GROUP CAPACITANCE INDUCTANCE OR L/R RATIO ( F) (mh) ( H/ohm) D C AB 2.11 F 613 nf 46 nf 33mH 12.6 mh 4.2mH 435 H/ 217 H/ 54 H/ THE ISOLATION OF THE SIGNAL WIRES WITH THE EASIDEW DISCONNECTED, MUST BE ABLE TO WITHSTAND A 500V AC INSULATION TEST. THE INSTALLATION MUST COMPLY WITH THE INSTALLATION PRACTICES OF THE COUNTRY OF USE, i.e. ANSI/ISARP12.6(INSTALLATION OF INTRINSICALLY SAFE SYSTEMS FOR HAZARDOUS [CLASSIFIED] LOCATIONS) AND THE NATIONAL ELECTRICAL CODE ANSI/NFPA 70. THE CAPACITANCE AND THE INDUCTANCE OF TEH HAZARDOUS AREA CABLES MUST NOT EXCEED THE VALUES GIVEN IN TABLE 1. DEWPOINT TRANSMITTER CERTIFICATION No's: Baseefa06ATEX0330X IECEX BAS 06.0090X HAZARDOUS LOCATION CLASS 1, DIVISION 1, GROUPS A,B,C, & D TRANSMITTER VERSION TERMINAL NUMBER EASIDEW I.S. 3 1 EASIDEW PRO I.S. 2 4 3 (+) (RETURN) NON-HAZARDOUS LOCATION (+) (-) KFD2-CR-Ex1.20200 BAS 00 ATEX 7164 000 000 000 000 000 KFD2-CR-Ex1.30200 BAS 00 ATEX 7164 KFD0-CS-Ex1.50P BAS 98 ATEX 7343 MTL5041 BAS 01 ATEX 7155 MTL5040 BAS 98 ATEX 2227 KFDO-CS-Ex2.50P BAS98ATEX7343 + - 4/20mA LOAD +VS (20 TO 35V DC) VS - 3rd ANGLE PROJECTION TOLERANCES: UNLESS OTHERWISE STATED 0 DEC. PLACE: ± 0.5 +0.1 HOLE Ø: DIMENSIONS: 1 DEC. PLACE: ± 0.2-0.0 2 DEC. PLACE: ± 0.1 ANGLES: ±0.5 DRAWING UNITS mm SCALE NTS THIS DOCUMENT IS THE PROPERTY OF MICHELL INSTRUMENTS LTD. AND MUST NOT BE COPIED NOR DISCLOSED TO A THIRD PARTY WITHOUT THE CONSENT OF MICHELL INSTRUMENTS. MATERIAL FINISH MICHELL INSTRUMENTS LTD. 01/11/05 DOF03 100mm 4 Inches DRAWN DATE MSB 10/03/06 CHECKED DATE APPROVED DATE TITLE USED ON EASIDEW I.S. DEWPOINT TRANSMITTER FM SYSTEM DRAWING MICHELL INSTRUMENTS LTD. CAMBRIDGE 05 11081 06/04/11 IMA 04 CERT ISS 24/03/09 IMA 03 CERT ISS 21/01/09 IMA 02 CERT ISS 23/12/08 IMA 01 CERT ISS 16/07/07 IMA ISSUE MOD. No. DATE SIGN DRAWING NUMBER Ex90385 SHEET 1 OF 1 A3 Issue 12, January 2013 29

Easidew PRO I.S. Dewpoint Transmitter User s Manual 12.5 FM Approvals Certificates Issue 12, January 2013 30

Issue 12, January 2013 31

Issue 12, January 2013 32

12.6 CSA Approved System Drawing Issue 12, January 2013 33

12.7 CSA Certificate of Compliance Issue 12, January 2013 34

Issue 12, January 2013 35

Issue 12, January 2013 36

Easidew PRO I.S. User s Manual EC Declaration of Conformity Issue 12, January 2013 37

13. TECHNICAL SPECIFICATIONS Performance Measurement Range (dew point) Accuracy (dew point) Response Time Repeatability -148 F to +68 F dew point ±3.6 F dew point 5 mins to T95 (dry to wet) 1.0 F dew point Electrical Output/Input Output Signal Supply Voltage Load Resistance Current Consumption Supply Voltage Information Output signal 4 20 ma (2-wire) current source, configurable over the entire range Dew point -148 to +68 F -100 to + 20 C 0 3000 ppm(v) ppm(v) output or non-standard dew-point range must be specified at time of order 12-28 VDC Max 250 Ω @ 12 V 500 Ω @ 24 V 20 ma ±0.005% RH/V Operating Conditions Operating Humidity Operating Temperature Operating Pressure Flow Rate Temperature Coefficient 0 100% RH -40 to +158 F 5000 PSIG max 2-10 SCFH mounted in standard sampling block 0 to 30 ft/sec direct insertion Temperature compensated across operating temperature range Mechanical Specification FM Certificate Ingress Protection Housing Material Filter Weight Electrical Connections Interchangeability Class I, Division 1, groups A, B, C and D IP66 in accordance with standard BS EN 60529:1992, and NEMA 4 in protection accordance with standard NEMA 250-2003 Stainless steel 80 μm sintered guard (optional HDPE Guard <10 μm) 1.7 pounds Screw terminal Fully interchangeable transmitters Issue 12, January 2013 38

13.1 Fault Conditions Condition Output Sensor fault 23 ma } Under-range dew point 4 ma Factory default setting Over-range dew point 20 ma User selectable via software NOTE: The current output range and the fault conditions are user programmable. Re-ranging or changing the fault conditions of the Easidew PRO I.S. Transmitter requires the use of a Communications Kit and Configuration Software. Contact Kahn Instruments Customer Service Department for details. 14. CONTACTS For advice on this, or any other Kahn Instruments product, please feel free to contact us: KAHN INSTRUMENTS, INC. 885 Wells Road Wethersfield, CT 06109 USA Phone: 860-529-8643 Fax: 860-529-1895 Email: hygros@kahn.com Web: www.kahn.com Issue 12, January 2013 39