VARIABLE RATE INFUSION PUMP

Similar documents
SINGLE SPEED SYRINGE PUMP

WATER MAZE POOL HEATER

PULSE OXIMETER PACKAGE

Alaris Syringe Module v9

TK-1801, TK-1811 TK-1801, TKR-1811

Alaris Syringe Module v9

ACME DIAGNOSTICS STANDARD OPERATING PROCEDURE

Administration of AMBISOME Using the GemStar Pump

Syringe Pump. User Manual

TK-1001, 1101, 1201, 1301, 1601

Code 542 Peltier Rack User's Guide

Exercise 8. Controlling a Batch Mixing Process EXERCISE OBJECTIVE

Rocket Quick Guide. Rocket PSU Portable Suction Unit

HOMEPUMP ECLIPSE* Ambulatory Infusion System PATIENT INFORMATION

CONTENTS. III. OPERATING INSTRUCTIONS... pg 7 LOADING DISPOSABLE SET (Figure 4)... pg 9. IV. INDICATORS AND ALARMS... pg 10

Operation and Maintenance Manual Chamber Conditioning Unit, CCU 06. Document Number: EC OM 6847 Revision: 3

FACSCalibur Startup Procedures

Freestyle 2000 Cleaning

Advanced Laundry Chemical Dosing System for Single and Multi-Washers

Aquamax KF Oil Evaporator. User Manual

Service Data Vickers. Piston Pumps. PVH74/81 Variable Displacement Piston Pump - 11 Design M-2207-S. Revised 11/01/97

Alaris Syringe module FAQs

Apex E Quick Installation Guide

Reliability in Laboratory Centrifuges Since Operator s Manual. Models 614B, 614L, 614U/UL and 614V/VL Laboratory Centrifuge. Rev. 5.

WATER/WASTE SYSTEM - SERVICING

Service Data Vickers. Piston Pumps. PVH98/106 Variable Displacement Piston Pump - 11 Design M-2208-S. Revised 05/01/97

Mitocell: Instructions for Use

Instructions for Use. A-dec 461 Cuspidor with Support Center

The door can be opened with a RF card. The door can be opened with a transit RF card. Up to 20 cards be registered.

ultra quiet fan USER MANUAL INSTALLATION OPERATION MAINTENANCE

ESSEX ENGINEERING CORPORATION

Oral Syringe Product Stability &

Eldex Column Heater. Operator s Manual

Operating Manual "Mark II" Solid-State Digital Automatic Controlled Sterilizers

Installation S19-324D1. Class I, Division 1, Groups C & D Explosion Proof. Emergency Signaling System

OWNERS MANUAL. To Order Parts Call SPEED AIR MOVER

THERMAPHASE INSTALLATION AND OPERATING INSTRUCTIONS

HH80C HYDRAULIC CRIMP TOOL

TECHNICAL INFORMATION Touchtronic Clothes Dryers

Multi-Mode Electrode pro

DWYER INSTRUMENTS, INC. P.O. BOX 373 MICHIGAN CITY, INDIANA 46360, U.S.A. Series 670 Hood Monitor. Bulletin AV-670

G.I. Procedures. Handbook

MODEL A-316 CARBON MONOXIDE (CO) MONITOR AND ALARM FOR COMPRESSED AIR TESTING

60" Hatteras BayTM. Patio. Instruction Manual. Includes our new Wall Control System. A Kichler Décor ceiling fan

Models Goldfisch Fat Extraction Apparatus

Sanitise Syrup Lines & Valves. Taylor PH61 Cleaning every 14 days

INSTALLATION AND OPERATION MANUAL ISM-4081 INTEGRATED SHUTDOWN MODULE

918 Clock Gauge Installation, Operation, and Maintenance Instructions

These pumps offer unique capabilities which are either not available or costly to acquire with other pumps.

Nilfisk Inc Winnetka Avenue North Minneapolis, MN REV.03( ) VF80189

ASTRO FF14 FRICTION FEEDER SERVICE MANUAL

S/601 Controller Installation Manual

H uline.com 80" AUTOMATIC STRETCH WRAP DISPENSER FILM DELIVERY SYSTEM MACHINE DIMENSIONS ELECTRICAL SPECIFICATIONS

User Manual. Dryer Controller M720

Cleaning the internal optics surfaces

TFD-1 Installation & User s Guide

RTE-5/5B & RTE-9/9B Refrigerated Bath Circulators. NESLAB Manual P/N Installation, Operation and Service Manual

Portable Roller Oven. Features

Technical Update TAA.TU rev. 1

Marley X7 Fan USER MANUAL DESIGNED FOR INDUCED-DRAFT COOLING TOWER APPLICATIONS

1.2.3 Returning Instrument for Repair.

SmarTemp Control fx 2.0

52 DorsetTM. Instruction Manual. Basic Function Wall Control System Included. A Kichler Decor ceiling fan

Perfect Draft Beer Tap HD3600/20

Dryer Controller M720

Operating Manual Espresso EP System

Waters 2767 Sample Manager, Injector, and Collector Installation and Maintenance Guide

Fully Automatic Water Filters

Owner s Manual RD432-0 Chemical Controller

Exercise 2-4. Heat Exchangers (Optional Exercise) EXERCISE OBJECTIVE DISCUSSION OUTLINE. Description of a brazed plate heat exchanger DISCUSSION

Marley Ultra Quiet Fan

PORTABLE SMART WASHER

MAINTENANCE MANUAL TAIYO SEIKI CO., LTD.

USER GUIDE. Software Version 1.11

Sierra Series 900 Single & Dual Channel System. Instruction Manual

Soldering Station INSTRUCTION MANUAL

INSTRUCTION MANUAL and DETAILED PRODUCT SPECIFICATION TRIPLEX PUMP CONTROL SYSTEM MODEL NUMBER CPC-3. M336 Rev F. November 21, 2000.

ALLERGEN MIXING ASSISTANT

Coolant Heater Thermo Top (TTC) Heavy-duty Operating Instructions Installation Instructions

Installation Manual PS-200 & PS-201

Vickers. Service Data. Piston Pumps. PVH57/63 Variable Displacement Piston Pump - 11 Design M-2206-S. Revised 11/01/97

OPR Power Series AC to DC POWER SUPPLY SERIES WITH REMOTE MANAGEMENT AND ALARM SYSTEMS. Model Nos. OPR200-24S / OPR200-24R. Manual

Clinician Information

ACSI MODEL 1410 POWER SUPPLY INSTALLATION INSTRUCTIONS

Tube Feeding by Pump UHN. Information for patients and families. Read this book to learn:

Sprite III Warewash Dispensing System

Model Series ELECTRIC WATER SYSTEM PUMP. PumpAgents.com - buy pumps and parts online. Model Series. Automatic Multi-Outlet FEATURES

Installation Manual PS-225 & PS-275

Global Water Instrumentation, Inc.

WSK300 Multifunction Wall Control

Simple to Set Timer (SST) Frequently Asked Questions

Owners Manual. SM-HP A SM-HP8-10 Meter. Manual D

ENGINE MONITOR VOTT MODEL: YEAR WARRANTY

DUST CAP. 2mm MICROPHONE CABLE FROM THE SERVOREELER REED SWITCH OUTPUT PLUGS INTO SERVOREELER NON-MAGNETIC CLAMP

KD-CLN-LP200 Ultrasonic Vinyl Record Cleaner User s Manual English v1.2 Printed in Korea

Global Water Instrumentation, Inc.

OPERATION AND MAINTENANCE INSTRUCTIONS MODEL 530 VIBRATORY LOSS-IN-WEIGHT FEEDER

Fig. 1. Bottle Height---Top of bottle MUST be at this height or slightly higher.

CEILING FAN OWNER'S MANUAL

VACUUM OVEN. Model: VPX9-2. Installation and Operation Manual

Transcription:

instrumentation and software for research VARIABLE RATE INFUSION PUMP PHM-100VS-AC USER S MANUAL DOC-302 Rev. 1.0 Copyright 2016 All Rights Reserved P.O. Box 319 St. Albans, Vermont 05478 Phone: 802.527.2343 Fax: 802.527.5095 www.med-associates.com

This page intentionally left blank

Table of Contents Chapter 1 Introduction... 1 Chapter 2 Types of Control... 1 Chapter 3 Syringe Placement... 2 Glass Syringes... 3 Chapter 4 Operating Instructions... 4 Manual Operation... 4 Automatic Operation... 5 Flow Rates and Pump Speeds... 6 Chapter 5 Automatic Shut-Off Switch... 7 Appendix A Contact Information... 8 - iii - DOC-302 Rev. 1.0 Copyright 2016

Notes Diagrams - iv - DOC-302 Rev. 1.0 Copyright 2016

CHAPTER 1 INTRODUCTION The PHM-100VS-AC provides infusions lasting from 0.1 to 99.9 seconds at rates from 0.5 to 20 RPM. The unit can be operated with a MED interface output bit, a TTL control line, or a switch closure. This pump also has the benefit of overdose safety protection, since the internal circuit limits the infusion time to the preset value. The PHM-100VS-AC requires 110 volts AC. CHAPTER 2 TYPES OF CONTROL Figure 2.1 Connections on the Back of the PHM-100VS-AC Controller Connections for the three methods of operation are explained below: MED Control Using the cable provided with the order, connect the MED Control 3-pin Molex connector to any available Output on a Standard MED Connection Panel. TTL Control Connect the TTL ground to the (-) screw terminal and the TTL operate signal to the (+) screw terminal. A TTL high to low transition on the operate line activates the pump. CAUTION VOLTAGE LEVELS ABOVE 5.2 VOLTS DC AT THE TTL TERMINALS CAN DAMAGE THE UNIT. Switch Closure Control Simply connect a switch to the (+) and (-) screw terminals. Closing the switch will operate the pump. The switch must be opened and closed again to activate the pump for another cycle. I/O This port is not used on this model. - 1 - DOC-302 Rev. 1.0 Copyright 2016

MED ASSOCIATES, INC. PHM-100VS-AC VARIABLE RATE INFUSION PUMP CHAPTER 3 SYRINGE PLACEMENT Care should be exercised that the hub of the syringe barrel is positioned adjacent to the syringe clamp when loading the syringe pump. If a gap exists between the hub and the clamp, accurate flow rates cannot be assured, as the entire syringe (both barrel and plunger) may move forward. A visual check by observing the plunger move in relation to the barrel by rotating the front knob of the pump is advised. Figure 3.1 shows the procedure for loading a syringe into the pump. Figure 3.1 - Loading a Syringe 1. Move the slide to rear (toward motor section) by squeezing the jaws. 2. Insert syringe by lifting clamp cover and placing the syringe body into the clamp. 3. Move the slide to the end of the syringe. 4. Rotate the knob so that the slide makes contact with the plunger and the syringe contacts the clamp. Once the syringe is installed, operate the pump at its fastest setting ( 00 ) until liquid drips out of the syringe or tubing. This ensures that fluid will be infused properly when the Pump is activated. -2- DOC-302 Rev. 1.0 Copyright 2016

Glass Syringes Extra caution is needed when using glass syringes with a ground glass plunger. These syringes exhibit almost no sliding friction and thus can cause an uncontrolled infusion in the following two ways: 1. The weight of the plunger may be sufficient to push the fluid out of the syringe if the syringe is held with the plunger above the syringe. 2. The weight of the fluid in the tubing may be sufficient to siphon the fluid out of the syringe if the catheter infusion site is below the height of the syringe. 3. To test for these two conditions, it is suggested that the syringe be connected to the tubing and held vertically at the height of the pump. If no motion occurs, the syringe can then be placed in the pump. The following may reduce the danger of an uncontrolled infusion: 1. Lower the relative height of the infusion pump in relation to the infusion site. With the pump below the infusion site, the instrument will pump the fluid to the higher elevation. 2. Use a smaller bore catheter, which will reduce the weight of the fluid in the tubing and increase the friction on the flowing fluid. 3. Position the pump so that the syringe is vertical (plunger below), thus the weight of the syringe plunger will be acting against the weight of the fluid. 4. Use a syringe with a rubber seal on the plunger, i.e. an O-ring sealed or plastic syringe. - 3 - DOC-302 Rev. 1.0 Copyright 2016

CHAPTER 4 OPERATING INSTRUCTIONS Manual Operation The pump can be operated manually to prime the line. There are two ways to operate the pump manually. The first is by setting the Flush switch in the down position for as long as required. The settings of the pushbutton switches are not recognized when the Flush switch is on. The pump will operate at 20 RPM and will only operate when a timed infusion is not in progress. The second method of operating the pump manually is to set the desired infusion time and rate using the pushbuttons switches, then pressing the Operate button. Figure 4.1 Front of the PHM-100VS-AC Controller - 4 - DOC-302 Rev. 1.0 Copyright 2016

Automatic Operation For automatic operation, the desired infusion time and RPM are set using the pushbutton switches on the front of the PHM-100VS-AC Controller. The three leftmost switches are used to set infusion times from 00.0 to 99.9 seconds, and the rightmost switch is used to select the pump RPMs as indicated by the table on the pump (see Figure 4.2 and Table 4.1). For example, a setting of 2 would produce a speed of 1 RPM. Note that 0 is not used and has no effect. Upon receiving an operate signal, the pump will operate for the time and RPM selected and then it will stop. The operate signal must be toggled off between infusions. Any changes in the thumbwheel settings will not be acknowledged until the cycle is complete. In the example shown in Figure 5, the infusion time has been set to 35.0 seconds and the rate is set to 2.0 RPM. If the end of the syringe is reached during an infusion the Automatic Shut-Off Switch will be activated and the infusion will be stopped. Refer to the Automatic Shut-Off Switch Adjustment section of this manual for further information. Figure 4.2 Pushbutton Switches Figure 4.3 Example Pushbutton Switch Settings - 5 - DOC-302 Rev. 1.0 Copyright 2016

Table 4.1 RATE Pushbutton Switch Settings RATE Pushbutton Setting Corresponding RPMs 0 N/A 1 0.5 2 1.0 3 1.5 4 2.0 5 3.33 6 5.0 7 10.0 8 15.0 9 20.0 Flow Rates and Pump Speeds The following table shows the flow rates in milliliters per minute of several B-D Plastipak syringes. Flow Rate ML/MIN B-D PLASTIPAK SYRINGES Push Button Setting RPM 1 ml 2.5 ml 5 ml 10 ml 20 ml 30 ml 60 ml 1.5 0.017 0.056 0.110 0.160 0.278 0.358 0.543 2 1 0.034 0.113 0.221 0.319 0.557 0.715 1.086 3 1.5 0.051 0.169 0.331 0.479 0.835 1.073 1.628 4 2 0.068 0.226 0.441 0.639 1.114 1.431 2.171 5 3.33 0.113 0.376 0.735 1.064 1.854 2.383 3.615 6 5 0.169 0.565 1.103 1.597 2.784 3.577 5.428 7 10 0.338 1.129 2.206 3.194 5.568 7.155 10.855 8 15 0.507 1.694 3.309 4.792 8.352 10.732 16.283 9 20 0.676 2.259 4.412 6.389 11.137 14.310 21.711 NOTE: Flow Rates are shown for B-D Plastipak Syringes. Flow Rate can be found using the following equation: Flow Rate (ml/min) =.19538 x RPM x Syringe Cross Sectional Area (cm 2 ) - 6 - DOC-302 Rev. 1.0 Copyright 2016

MED ASSOCIATES, INC. PHM-100VS-AC VARIABLE RATE INFUSION PUMP CHAPTER 5 AUTOMATIC SHUT-OFF SWITCH The automatic shut-off switch is activated when the pump reaches the end of the syringe, and it causes the pump motor to cease operation. The automatic shut-off switch uses an internal micro-switch that detects when the slide assembly reaches a certain point. This point is adjustable so that the pump can accommodate various sizes and brands of syringes. To adjust the automatic shut-off switch, insert an empty syringe with the plunger at the desired shut-off position. Move the slide assembly to the rear of the plunger (Figure 5.1A). With the pump turned on, switch the Flush switch on and rotate the adjustment screw (Figure 5.1B) until the pump stops operating. Turning the screw clockwise will allow the pump to operate longer before the automatic shut-off switch activates. Conversely, turning the screw counterclockwise will cause the automatic shutoff switch to activate sooner. In order to reset this switch and resume pump operation the slider must be moved back until the Automatic Shut-Off Switch is deactivated. Figure 5.1 - Automatic Shut-Off Switch Adjustment -7- DOC-302 Rev. 1.0 Copyright 2016

APPENDIX A CONTACT INFORMATION Please contact for information regarding any of our products. For Technical questions, email support@med-associates.com. For Sales questions, email sales@med-associates.com. Visit our website at www.med-associates.com. - 8 - DOC-302 Rev. 1.0 Copyright 2016