Compression therapy device & accessories Model 701A User manual

Similar documents
Multi-Flo Model #1545-KT. Multi-Flo Model #1545-F

SEQUENTIAL CIRCULATOR

SEQUENTIAL CIRCULATOR MODEL 3008

SEQUENTIAL CIRCULATOR MODEL 3008

TributeNightTM. user guide

SEQUENTIAL CIRCULATOR

Med-Aire 8 Alternating Pressure Mattress Replacement System With Low Air Loss OPERATOR S MANUAL

DELUXE ALTERNATING PRESSURE SYSTEM USER GUIDE

Model IC-1545-DL Multi-Flo DVT Combo

Instructions For Use SC-2004-OC, SC-2004FC-OC and SC-2008-OC. Sequential Circulators

Anti-Decubitus Air Alternating Pressure Mattress System Operation Manual

Bio Compression s SC-3004-DL

SC-3008-DL SEQUENTIAL CIRCULATOR

User Instruction Manual

USER INSTRUCTION MANUAL SAPPHIRE SERIES 800

Dyna-Form Air. Alternating Pressure Relief Overlay System

Use r manual Edition MG Tube Master Anti-bedsore alternating pressure mattress of tubular structure VCM502D

Glass Door Refrigerators

Thank you for purchasing Pegasus automatic unit.

STEAM WALLPAPER STRIPPER MODEL HTW5

Instructions for Installation and Use

Convertible Powered/ Non-Powered Dynamic Mattress System. Operation Manual. Distributed by:

USER INSTRUCTION MANUAL SAPPHIRE SERIES 1100 & 1100EC

User Instruction Manual

Med-Aire Assure 5 Air + 3 Foam Base Alternating Pressure and Low Air Loss Mattress System

BIOVAC 500 SMOKE EVACUATOR

User Instruction Manual

SERVICE PROCEDURES. Mattress 2 Pipe Overlay or Replacement

OGCF2002A User Manual

The FAWSsit USER CARE GUIDE

English 1 AIR ALTERNATING PRESSURE MATTRESS REPLACEMENT SYSTEM

Active Sport Massager

AMBIENCE Low Air Loss Mattress Replacement System

Installation Instructions

AIR EXPRESS LAL. P1500 LOW AIR LOSS THERAPY SYSTEM REF C1500 Control Unit REF M1500 Mattress. Operator s Manual P/N /03

CircuFlow Quick Setup Guide

BEFORE USE, PLEASE READ AND FOLLOW ALL SAFETY RULES AND OPERATING INSTRUCTIONS.

11,000 BTU Portable Air Conditioner with dehumidifier & Fan PE4-11R-03 Operating Instructions. Model No. PE4-11R Soleus Air International

General Care and Safety Guide

MULTI-FUNCTIONAL PORTABLE ROTATION SYSTEM P4000 P4001 C4000 Series Control Unit M4000 Series Mattress. Operator s Manual P/N Rev B 10/10

APH166 Morpheus Air USER MANUAL. Issue 3-24/05/2017. Apollo Healthcare Technologies Ltd. Holme Street, Liversedge, West Yorkshire, WF15 6JF.

Model Letter Opener Revision 3 - Brush Transport

CONTENTS QUICK REFERENCE GUIDE INSTRUCTIONS FOR USE PROGRAMME CHART BOOKLET

Enhance DX. Mattress Replacement System. User Manual

Flowtron ACS900. Instructions For Use. ...with people in mind INT1 Rev A - 05/2015

User s Manual. Alternating Air Mattress Model No.: MA-6100

CHEST FREEZER INSTRUCTION MANUAL. Model No.: EWCF5WBX EWCF7WBX

5) Do not start or stop the unit by inserting or pulling out the power plug.

Digital Fabric Steam Press

Air Paper Jogger AJ-700 Operation Manual. Keep this manual at hand where it can be referenced at any time.

System One Heated Humidifier USER MANUAL

BL Series INSTRUCTIONS

UL U TR LTRASONIC S ONIC SCALE ALER PIEZO MINI

Owner s Manual Window Air Conditioner

12,000 BTU Evaporative Portable Air Conditioner

MW001 Integrated Dishwasher. Manual for Installation, Use and Maintenance

Built-In Dishwasher. Installation Instructions. BEFORE YOU BEGIN Read these instructions completely and carefully. IMPORTANT The dishwasher MUST be

FROST FREE REFRIGERATOR INSTRUCTION MANUAL

HOUSEHOLD FREEZER 048-GM-48303/048-GM-48304/048-GM-48305

PORTABLE HAND POWER THREADER 1/2-2

Portable Air Conditioner 6,000 BTU 8,000 BTU 10,000 BTU

Installation Instructions

Tornado Operations & Maintenance Manual

Model No.: PS08-01 PS10-01 Ref: KY80 KY100

PS-2/ES Automated pack & tag machine with IndES fastening system

G-7s. Instruction Manual. G-Series Cooler COUNTERTOP COOLER. Part No.11IPA

General Care and Safety Guide

TECHNICAL INFORMATION Touchtronic Clothes Dryers

User s Manual UPRIGHT FREEZER WHS-507FWEW1 / WHS-507FWESS1 WHS-625FWEW1 / WHS-625FWESS1 WHS-772FWEW1 / WHS-772FWESS1

SOLO OWNERS MANUAL, OPERATING INSTRUCTIONS & INSTALLATION INSTRUCTIONS

Technical data. Bathroom/shower room installations. Water services (WRAS approved)

OFFICE PRO 36 and OFFICE PRO 60 READ THIS MANUAL CAREFULLY FOR INSTRUCTIONS ON CORRECT INSTALLATION AND USAGE AND READ ALL SAFEGUARDS

CONSTANT SUCTION UNIT

REFRIGERATOR SAFETY. Your safety and the safety of others are very important.

DISHWASHER INSTALLATION GUIDE SPECIFICATIONS, INSTALLATION, AND MORE

User Manual. Invacare APM Model CG9900 Mattress

MODEL 7000 SUCTION UNIT

Operating Instructions

CONTENTS QUICK REFERENCE GUIDE PROGRAMME CHART

USER MANUAL. Safety Precautions. Unit Specifications and Features. Operating Instructions. Installation Instructions.

INSTRUCTIONS FOR USE PORTABLE VACUUM SYSTEM LEI Part # s / , , , IMPORTANT INFORMATION

OPERATION MANUAL. CLASSIC 10 and CLASSIC 18 SERIAL NUMBER FROM JANUARY 2009 (0109) TO PRESENT

Removing shipping locks

ImPress IRONING STATION

CONTENTS QUICK REFERENCE GUIDE INSTRUCTIONS FOR USE BOOKLET PROGRAMME CHART

home use only in accordance with the instructions provided in this manual. Please read these instructions carefully and retain for future use

SERVICE MANUAL SPARE PARTS CATALOG AND EQUATOR COMBINATION WASHER-DRYER. First Edition - October Advanced Appliances.

Instruction Manual. Head & Eye Massager with Acupressure and Heat Compression.

This page is intentionally left blank

CEILING SPEAKER SYSTEMS F-2352C

Installation. 324 Series Built-In Dishwashers U L. Viking Range, LLC 111 Front Street Greenwood, Mississippi USA (662)

ELECTRIC UNDER BLANKET QUEEN FLEECY FITTED MODEL NO: TT-1101XQF INSTRUCTION MANUAL

Portable Air Conditioner with Heat Pump Technology Operating Instructions. Model No.: HCB-P13HP-D. Reference No.: BPD13HP V140217

3-In-1 Heated Wrap WarmtH WHere you Need It lk rints:1/1,b _P rap eated W utput:100% _O _3-in-1 H S x4.75 H 97_IN 43 ize:5 W 73 S

OWNER S MANUAL FOR AIR ADMIRAL VACUUM/PRESSURE STATION MODELS:

18 INCHES, OSCILLATING STAND FAN

User Manual standard for Medical Electrical. Control Number 95CG. Classified by. Underwriter s Laboratory Inc. for UL

Installation Instructions Built-In Dishwasher

Operating & Installation Instructions Fully Automatic Washing Machine (Domestic use)

Portable Air Conditioner with Heat Pump Technology PH4-10R-01, PH4-12R-01, & PH4-14R-01 Operating Instructions

Transcription:

Compression therapy device & accessories Model 701A User manual April 2008 L98101A6 Ver. 01

Table of Contents 1 GENERAL... 1 2 MAIN COMPONENTS... 4 3 OPERATING THE PETITE BASIC SYSTEM... 6 4 MAINTENANCE...10 5 MANUFACTURER DETAILS AND WARRANTY...12 6 REPRESENTATIVE IN USA...12 Table of Contents

1 General The information listed in this user manual will assist you in using the Petite Basic System (PBS) device and accessories to obtain maximum benefits from the treatment. Should you have any questions or require further information, please do not hesitate to contact your local dealer. The Petite Basic System is a non-invasive intermittent sequential external pneumatic compression system that aids in the reduction and management of peripheral edema, such as lymphedema and ulceration associated with venous insufficiency. 1.1 Indications Venous stasis ulcers Reduce and manage peripheral edema Venous insufficiency Lymphedema 1.2 Contraindications Compression is not recommended in the following conditions. Known or suspected deep venous thrombosis During the inflammatory phlebitis process During episodes of pulmonary embolism Pulmonary edema or congestive heart failure Severe arteriosclerosis or other ischemic vascular disease Any circumstance where increased venous and lymphatic return is undesirable Warning: Federal law (U.S.A.) restricts this device to sale by, or on the order of, a physician. 1.3 The Petite Basic System cycle The PBS is an intermittent sequential external pneumatic compression system that aids in the reduction and management of peripheral edema, such as lymphedema and ulceration associated with venous insufficiency. The treatment cycle consists of inflation and pause time. During inflation, the cells inside the compression sleeve are inflated one after the other from distal to proximal. The time required for inflation will depend on the pressure level selected and on the size of the treatment sleeve. When all four cells have reached the required pressure, all of the cells will deflate, followed by the pause time after which the inflation cycle repeats. The pressure and the pause between cycles are adjustable. 1

1.4 Specifications and Dimensions Specifications Dimensions Model 701A Height 3.9 inches Catalog No. L10000A6 Length 10.2 inches Class II (double isolation) Width 5.1 inches Pressure range 20~80 mmhg Net Wt. 5.1 lb. Voltage 115V ~ 50/60 Hz Power consumption 11W Fuses rating 2A Note: Although the device is Class II, it is equipped with a functional grounding (three-pin plug), for safety purposes. 1.5 Symbols on devices Symbol Explanation Location Attention! On bottom of Refer to manual air pump Consult instructions before use Level of protection - type BF equipment Temperature limitations for storage and/or transport of the device Date of manufacture Accompanied by the name and the address of the manufacturer On sleeve's tag On bottom of air pump On carton box On bottom of air pump On bottom of unit and on sleeve label Classified by Underwriters Laboratories Inc. with respect to electric shock, fire and mechanical hazards only in accordance with UL 60601-1 and CAN/CSA C22.2 No. 601.1. On back of unit 2

1.6 Safety features 1.6.1 In case of power failure In the event of power failure, the pneumatic control unit is designed to deflate all the air from the sleeves automatically. 1.6.2 One-way hose connection In order to prevent wrong connection of the air hose bundle to the device, which will cause improper inflation order (sequence) of the sleeves, the connector is designed for one-way connection. Proper connection can only be done when the connector plug is inserted in the proper direction, with the connector's "The Petite Basic System" logo facing up. The hose end fittings (on the sleeve side of the hose bundle) are numbered from 1 to 4, indicating the order in which they are to be connected to the sleeve, from distal to proximal, corresponding to the numbers 1 through 4, next to the air inlets on the sleeve. Each number also has a specific color, both on the hoses' end and on the sleeve's air inlets. 1.6.3 Overload Protection Fuses The device is equipped with two overload fuses, on each of the power lines ("~" and "0"). 1.6.4 "Lock" button The "Lock" button fixes the pressure adjustment knob in place, preventing the pressure knob from being accidentally reset to an inappropriate pressure. 3

2 Main components 2.1 Device Unit This device is powered by electricity, provided by a wall outlet with a grounding connection. It supplies compressed air in fixed cycles at regulated pressures. The power unit is able to simultaneously operate two sleeves with 4 air compartments each. 2.2 Hose Bundle The hose assembly that transfers compressed air from the device unit to the sleeves. It includes 4 hoses, a connector that attaches to the corresponding receptacle of the device unit and 4 numbered and color-coded end fittings that attach to the sleeve. 2.3 Pronged Plug Plug used to seal off the unused outlet of the device unit, when only one limb is treated. 2.4 Sleeves Each zipper sleeve has 4 air compartments and uses cyclic air pressure. Different sizes of sleeves are available, 2.5 Sleeve Expanders If necessary, a zipper expander may be added between the two sides of the zipper to increase the sleeve s circumference. 2.6 Inner Cotton Hygienic Sleeve A tubular cotton inner sleeve, worn under the treatment sleeve for hygienic purposes. The inner cotton sleeve is a one-piece tubular bandage, much like an open-ended stocking, that fits over the limb. 4

Figure 1 the Petite Basic System components Pause knob "Lock" button Pressure knob Main switch a. Device unit b. Hose bundle c. Prong plug d. Cotton inner hygienic sleeve e. Leg sleeve f. Arm sleeve g. Expander 5

3 Operating the Petite Basic System 3.1 Setting up the Petite Basic System device Place the Petite Basic System on a flat table surface or suspend it on the end of a bed with the attached hooks. The device should be easily reached by the patient. Be sure the device is securely positioned to ensure that the device will not slip or fall. If the device is hung from the bed rail, it is recommended to center the device in the middle of the rail. 3.2 Attaching the air hoses to the sleeves Attach the air hoses to the sleeve by inserting each hose into its appropriate air inlet, matching the numbers and colors on both the end fittings of the hoses and the sleeve s inlets. Air inlet number 1 is always the farthest one from the body. 3.3 Putting on the sleeve First cover the limb(s) with the inner cotton sleeve(s), then put on the Petite Basic System sleeve(s) as follows: a. For leg sleeves: Insert foot into the boot. Pull the sleeve towards knee, extending it to its maximum length, close the zipper, starting from the bottom. b. For arm sleeves: Zip the sleeve closed at least part of the way up before inserting your arm. Make sure the zipper is facing up, and that the cut-out area is under your armpit. Zip the sleeve closed. Support your arm on an armrest or a pillow during treatment. If the sleeve tends to slide down during treatment, use the optional belt. To use the belt, attach one end of the belt to the snap closure on the sleeve, at the back, then pull on the sleeve, zip it closed, pull the belt around your body under the arm not being treated and fasten it to the other snap closure at the front. Correctly fastened, the belt should cross from the shoulder of the arm being treated, across the chest and under the arm not being treated. Tighten the belt so that it fits just enough to keep the sleeve from slipping down. c. During treatment, the zipper should be closed and left closed with the hoses attached. The patient should be able to slip sleeve off and pull on again for each use without unzipping the sleeve each time. (Note: After use, the sleeve will deflate within a short time, enabling it to be removed easily). Note: Do not use any sleeve other than the Petite Basic System sleeves, made especially for the Petite Basic System device. 6

3.4 Connecting the air hoses to the device Securely insert the hose bundle connectors - with the Petite Basic System logo facing up - into one of the air sockets on the Device unit (it does not matter which one). If only one sleeve is to be used, remove the Pronged Plug from its storage socket and place it into the unused air outlet. This will prevent air from escaping through the unused socket. Make sure the air hoses are not bent, kinked or pinched. Note: When only one sleeve is used, the second outlet must be plugged. Otherwise the device will stop working and the alarm buzzer will beep. 3.5 Operating instructions (See Figure 2) Note: When using the device, do not use any pressure other than the pressure prescribed by your physician. Treatment time should not exceed that prescribed by your physician. If you have any questions, contact your physician. In the case of a power failure, turn the device OFF and remove the electrical cord from the wall outlet. Figure 2 the Petite Basic System - general view of front panel Pause knob "Lock" button Air outlets Pressure knob 7

a. Connect the power cord to an electrical outlet. b. The first time you use the device, set the Pressure and the Pause knob to their minimum settings. c. Turn the device ON and wait for a few cycles to become accustomed to the treatment sensations. Note that the first inflation cycle may take a longer time to fill, especially if you are using two sleeves. d. Next, set the pressure and pause time according to your physician's instructions. Then lock the pressure knob by turning the "Lock" knob clockwise. This will prevent accidental tampering with the pressure knob. If you ever need to change, the pressure, simply loosen the "Lock" knob, change the pressure settings, and re-tighten the "Lock" knob. e. Continue the treatment for the time ordered by your physician. Note:! If your dealer has preset and locked the pressure settings according to your physician's instructions, please omit steps b through d in the instructions above. Simply turn the device on and continue treatment for the time ordered by your physician. Note that the first inflation cycle may take longer to complete, especially if you are treating two limbs at the same time. The subsequent cycles will be shorter.! Note: For larger sleeves, a longer pause time will be necessary to ensure sufficient deflation of the sleeves between cycles. The device will not inflate the sleeve unless the air from the previous cycle has adequately emptied. If the pause time has been set too short for adequate deflation, the device will automatically double the pause time. To prevent this, increase the pause time slightly and try again. 8

3.6 End of treatment At the end of the treatment, shut the unit OFF. Remove the sleeve(s) from the limb(s). You do not have to disconnect the hoses from the sleeves or unzip the sleeves after each use. (If you travel with the device, the hoses should be disconnected from the sleeves). If you wish to disconnect the hoses from the sleeve, gently work the end fitting free from the air inlet using a twisting / pulling motion. Note:! 1. The treatment should be pleasant and relaxing. 2. If the device stops working and the warning signal sounds, it means that the time needed to fill up the sleeve is too long. Usually this is due to air leakage because the sleeves or hoses are not properly connected. If this happens, turn the device OFF, and a. Check that all the hose ends are correctly inserted into the air inlets. b. Check that the hose connectors are properly connected to the air outlets on the device. c. If using only one sleeve, make sure that the unused outlet is plugged with the prong plug. If all the connections are OK and the problem persists, contact your dealer. 3. In the case of a power failure, turn the device OFF, take off the sleeve, and remove the electrical cord from the wall outlet. 9

4 Maintenance! Warnings: 1. Only an authorized technician should open the device. 2. Before cleaning the device, disconnect the power cord from the electrical wall outlet. 4.1 Cleaning the device Gently wipe the outside of the device using a damp, nonabrasive cloth. Do not spill any liquids on the device. 4.2 Cleaning the sleeves Gently wipe the sleeve using a soft cloth, warm water (not exceeding 40 C (100 F)) and a non-allergenic cleaning liquid. Do not allow water to enter the air inlets of the sleeve! Wipe dry using a soft towel only. When dry, the sleeve may be wiped down using a hospital-grade antiseptic or antimicrobial sanitizing lotion. Warnings: Do not dry clean, hand or power wring, iron, tumble or force heat dry. Do not use bleach! Clean only according to instructions. 4.3 Storage 1 Store in a dry, shaded place at temperatures between -20 C to +70 C. 2 The device should be stored after rolling up the power cord. Do not bend the cord. 3 Since the sleeve is inflated by compressed air, it must remain airtight. Therefore, avoid contact with pins, needles and any other sharp objects or instruments. If one of the cells is damaged, the sleeve must be replaced. 4 Store the sleeve flat, without folding. 5 The hoses should be kept untwisted and unfolded. 4.4 Transport Transport in original packaging, at temperatures between -20 C to +70 C. 10

4.5 Troubleshooting Note: Before continuing, check all accessories visually for any defects. Table 1 - Troubleshooting guide Symptom Possible cause Corrective action The device is not No electricity. Check the electrical wall outlet. working. Power cord. Check the power cord visually for any defects. Fuses. Check the fuses and replace if necessary. If they burn out again, contact your dealer. The device starts working and stops immediately. The air cannot move through the hose bundle. Check hose bundles for kinks, twists and folds. One sleeve inflates but the second one does not. The device stops working and buzzer beeps. The device works at a very low pressure, regardless of the pressure set by the user. An irregular noise. The second sleeve does not receive air. Hose bundle is not connected properly to sleeve or device, or prong plug is not inserted into unused air outlet. Defective sleeve. Check its hose bundle for kinks, twists and folds. Check and fasten all air connections. When treating only one limb, always plug the unused air outlet with the prong plug provided with the device. If all air connections are OK and problem persists, contact your dealer. Replace sleeve and check again. An internal problem. Contact your dealer. Transferring of Make sure the device is standing on vibrations to the table all four of its bumpers, or is hung or the patient s bed. properly on its hooks while supported by the two rear bumpers. An internal problem. Contact your dealer. 11

5 Manufacturer details and warranty 5.1 Manufacturer address Mego Afek AC Ltd. Kibbutz Afek, 30042 Israel 5.2 Warranty a. Mego Afek AC Ltd. warrants the Petite Basic System and its sleeves to be free of defects in materials and workmanship. b. This warranty applies as follows: For devices: for a period of twelve (12) months from date of purchase. For sleeves: for a period of six (6) months from date of purchase. c. This warranty does not include or cover malfunctions caused by unreasonable use, non-compliance with user and maintenance instructions, or damage caused by unauthorized or unqualified repairs. 6 Representative in USA LYMPHA PRESS USA, Ltd. 232 Park Avenue Manalapan NJ 07726 Tel.: 1-888-596-7421 Fax: 1-732-792-9745 12