LaserPoint srl - Via Burona, Vimodrone (Milano) - Italy Phone Telefax

Similar documents
2 Ensure that all personnel in the area are wearing the appropriate protective eyewear.

Customers must fill in and mail the warranty card in order to activate the warranty.

I/R Thermometer Gun 12:1

64 Max IR Thermometer

Dry Bath Heating Systems

Operator's Manual. LIGHTNING 2 Laser Detector

PL400 Patient Lift Scale Indicator Owner s Manual

METEOROLOGICAL INSTRUMENTS

Wide Range InfraRed (IR) Thermometer with Type K input and Laser Pointer

Infrared Thermometer

User Manual. Humidity-Temperature Chart Recorder. Model RH520

MODEL WIND TRACKER

Filter Oven with Controller 101FRDC00-50 / 101FRDC00-CTRL

MODEL MARINE WIND TRACKER

64 Max IR Thermometer

3R, 3G 5R, 5G Point Laser Levels

Dual Laser InfraRed (IR) Thermometer

Precision Infrared Thermometer. Users Manual

GMH 285 / GMH 285-BNC

62 MAX/62 MAX + Infrared Thermometer

MICROtrac MICROPROCESSOR BASED WATER TREATMENT CONTROLLER. Installation and Operation Manual REV. L

Thermo-/Hygrometer with Mould Alert Model WS 1500 Operating Manual

On Electromagnetic Compatibility

Model PT-05 Model PT-06

VS2000V Veterinary Vital Signs Monitor Operation Manual

Adam Equipment. NCSA 20plus. (P.N Revision B - September 2012)

Miniature Thermoelectric Cooler Controller MTCC-01 user's guide

Version 1.6. Operating instructions HYDROMETTE BL COMPACT TF 2. Hydromette BL Compact TF 2

Infrared Thermometer. Getting Started

KM300 Carbon Monoxide Detection System Installation Manual

Preamplifier Power Supply PPS-02 user's guide

6R, 6G 180R, 180G. Users Manual. Point and Line Lasers. Line Laser Levels

SEC Signature Process Gas Analyzer

Kinsa Smart Ear ThermometerTM Instructions For Use

Model 1916-C/-R Hand-held Optical Power Meter

esense CO2 Transmitter and Alarm

Wöhler CM 220. Carbon Monoxide Meter. The Measure of Technology. Contents. 1. Important information Specification... 13

Model: LD5 Dual Mode Refrigerant Gas Leak Detector. Operating Instructions

User Manual. FieldMate Laser Power Meter

Corning LSE Digital Dry Bath Heater

In accordance with the Radio Waves Act Chapter 58 Article 8(3), the document below is a

INSTRUCTION MANUAL REMOTE CONTROL DEVICE EASYREMOTE

User Guide InfraRed (IR) Thermometer with Wireless Datalogger, Type K input & Laser Pointer MODEL 42560

VISION H. Instruction manual

USER MANUAL AND TECHNICAL MANUAL

Laser Safety Products. Edition 20

CALIBRATION HOW IT WORKS

ANALOX 101 D2 PORTABLE OXYGEN MONITOR USER MANUAL

Owner & Operator s Manual. Table of Contents

MGC Dock User s Manual

High Temperature InfraRed Thermometer with Laser Pointer

MT4 MAX/MT4 MAX + Infrared Thermometer

Health Check Ear and Forehead Thermometer Model No.: Instruction Manual

User Manual. FieldMate Laser Power Meter

CALIBRATION HOW IT WORKS

User's Guide. Pinless Moisture Meter. Model MO257

METEOROLOGICAL INSTRUMENTS

Instruction Manual. Model AccuPlate Hotplate AccuPlate Stirrers AccuPlate Hotplate Stirrer 12/10 CLS-EQ-LHP01REV

Software Version 2.01 LEVEL MONITOR MODEL 220

KERN MPK Version GB

Three Phase Prepayment Meter Liberty-3P

ANNEALING FURNACE MODEL 414 User Maintenance Manual/Handbook

100 Ft. LASER DISTANCE METER

User's Guide. Pinless Moisture/Humidity Meter with Memory + IR Thermometer. Model MO295

818E Series Energy Detector. User s Manual

OI-2400-DOCK Multi Gas Docking Station. Operation Manual trevision 2.2w

Instruction Manual. For All Corning Hot Plates, Stirrers, Stirrer/Hot Plates with Digital Displays, and External Temperature Controller (6795PR)

HIn3550 Magnetic Field Monitor User's Manual

HXL. Operating instructions

Instruction manual EN

User's Guide. Pinless Moisture/Humidity Meter + IR. Model MO290

USERS MANUAL FOR GAS BOILERS

SMART EVO 2 - User Manual ELECTRICAL PANEL FOR 2 MOTORS

Mark 25 Ultrapure Water Conductivity Analyzer

VX SERIES Wireless Thermostat with Occupancy Sensor

Our photo detectors are offered for both power or energy measurements. Measure as low as a few femtojoules in energy or a few picowatts in power.

MicrOTDR. Optical Fault Locator. User Manual

Analox 1000 Series. User Manual. Analox Sensor Technology Ltd. 15 Ellerbeck Court, Stokesley Business Park North Yorkshire, TS9 5PT, UK

Digital Heat Block User Manual

IFS Fiber Module Installation & Operation Instructions

62 MAX/62 MAX + Infrared Thermometer. Users Manual

R E N HI ITI CUSTOM AN LO +12v ADJUSTMENT

FS-200. Model. Instruction Manual. Infrared Food Thermometer. reedinstruments. www. com

OPERATING MANUAL TORREY PINES SCIENTIFIC, INC. DIGITAL HPLC COLUMN HEATER MODEL CO20 DOCUMENT NUMBER CO20-00

Thermometer with Probe model 00891A

Non-Touch Thermometer. User Manual. ARC Devices Ltd.

Polar WearLink + Coded Transmitter 31 Coded Transmitter W.I.N.D. User Manual

Wireless Humidity/ Temperature Monitor Remote Wireless Sensor

Our photo detectors are offered for both power or energy measurements. Measure as low as a few femtojoules in energy or a few picowatts in power.

INSTALLATION GUIDE. AXIS A1001 Network Door Controller

PROCESSING UNIT II V Hz V Hz. Table of Contents

(1135/2016) Translation from Finnish Legally binding only in Finnish and Swedish Ministry of Economic Affairs and Employment, Finland

Matrix2000. User Guide - CPD Advanced Analog Fire Alarm Control Panel 4 24 Zones. Version: 2.0 Revision: 1

safety & regulatory information

KFP-CF Series Operation Manual

Important Safety Notice

This instruction manual explains the use of three models listed on the left.

Macurco Single-Gas XL Series Monitor, CM-1XL Carbon Monoxide (CO), HS-1XL Hydrogen Sulfide (H2S) User Instructions

LogTag TICT - is0 Tag

h is ngl E ASSISTANCE CENTRE AUTHORIZED RETAILER

Transcription:

Cronos Instruction manual (Rev11- Dec 2011; Firm. vers.4) WARNING! The user of this instrument must be trained to the use of power lasers and associated risks. LaserPoint is in no way liable for any damage resulting from improper, careless or above rated limits use of this instrument. LaserPoint srl - Via Burona, 51-20090 Vimodrone (Milano) - Italy Phone +39 02 27 40 02 36 - Telefax +39 02 25 02 91 61 www.laserpoint.it LaserPoint s.r.l. Cronos Instruction Manual All Rights are Reserved 1

Declaration of Conformity Application of Council Directive(s): 89/336/EEC The EMC Directive Manufacturer s Name: LaserPoint s.r.l. Manufacturer s Address: Via Burona 51, 20090 Vimodrone (MI) - Italia Type of Equipment: Laser Power / Energy Meter Model No: CRONOS Years of test & manufacture: 2014 Standard(s) to which Conformity is Declared: EN 50081-1:1992 Emission generic standard - Part 1: Residential, commercial and light industrial EN 50082-2:1995 Immunity generic standard - Part 1: Industrial Standard Description Performance Criteria EN 55022:1994 Limits and methods of measurement of Class B amendment 1: 1995 radio interference characteristics of amendment 2: 1997 information technology equipment EN 61000-4-2:1995 Electromagnetic compatibility (EMC) Class B Part 4: Testing and measurements techniques- Section 4.2: Electrostatic discharge. I the undersigned, hereby declare that the equipment specified above conforms to the above Directive(s) and Standard(s). Place: Vimodrone - Italia Date: Dr. Luigi Argenti (Technical Director) LaserPoint s.r.l. Cronos Instruction Manual All Rights are Reserved 2

Calibration Model Reference Power (1064nm) Reference Power (10600nm) CRONOS LP W W Serial Number: Measured Power: Measured Power: W W Date of calibration: 1) In Models LP10 and LP5.0 measured values at wavelengths of CO 2 and Yag are the same, within the specified measurement accuracy. LaserPoint certifies that, at the time of calibration, the above listed instrument meets or exceeds all technical specifications defined at page 8 of this Instruction Manual. The above listed instrument has been calibrated using standard references traceable to the National Institute of Standards and Technology (NIST) or Physikalisch Technische Bundesanstalt (PTB). This is not a Calibration Certificate for traceability purpose; such a Calibration Certificate can be quoted on demand. Important - For Customer Use Only The reference setting of this instrument can be matched by user to a custom own reference by means of the instructions reported at Paragraph n.11 of the Operation Section of this Manual. LaserPoint s.r.l. Cronos Instruction Manual All Rights are Reserved 3

Operation Section 1-Safety Take all the required safety procedures needed when working with power laser beams and wear protection glasses all the time! The laser beam must hit ONLY the sensor head front side. Avoid to incidentally expose the protected stainless steel stem to the laser beam as dangerous back-reflections can be produced! Never touch the sensor head as this part can reach 70 C of temperature! In the case the laser power density is close to the damage threshold of the head absorber coating, damages like coating discoloration or underlying beginning of metal melting could occur. In both cases consult LaserPoint 2-Starting the Instrument To switch on the instrument, keep the ON/Zero button pushed until the LCD shows the serial number and the model of your instrument (LP.X, where X = 1.5, 5.0 or 10), then release it (this step takes approximately 2 seconds). The instrument will also display the selected laser wavelength; two options are possible: CO 2 (10.6 m) or YAG (1060nm). The wavelength displayed the first time is random; to change it refer to Paragraph 10 of this section. 3- Auto-check Cycle When the ON/Zero button is released, the instrument starts an auto-check cycle during which both circuit and display are controlled and reset. During this cycle, the LCD shows all available segments of alphanumeric characters and the bar graph at full scale. Upon completion of this cycle, the LCD displays the last recorded value of measured power.. 4-Zeroing/Resetting Before starting the first measurement or a new one, the instrument needs to be zeroed; to do so, just push and soon release the ON/Zero button. The LCD displays CAL and the LED on the right side of the LCS, turns steady green for approximately 5 seconds. At this point, any previously recorded value is zeroed and the instrument is ready to measure. LaserPoint s.r.l. Cronos Instruction Manual All Rights are Reserved 4

5-Measurement Cycle Check the laser shutter is closed and the laser is on; align the probe to the laser beam path and start the measurement cycle by opening the laser shutter. The measurement starts automatically and the measurement time is shown by a count down displayed on the LCD starting by Run 8 down to Run 0. This step takes approximately 8 seconds. During the acquisition time the LED blinks green. The bar graph indicates the head temperature on an arbitrary scale. Avoid measurements while the laser is warming up or when it shows instabilities on a few seconds period, as they can be meaningless or misreading. Always keep the instrument steady and the beam well centred, in particular on conical probes; do not move it during the readings to avoid measurement errors. 6-Measurement Display As soon as the measurement is over the LED turns into steady red. Remove the probe from the laser beam: the display shows the actual power, and the remaining thermal capacity. If the bar graph is below full scale, one or more measurements can be carried out before the sensor head needs to be cooled. The following measurement can be executed after the LED automatic shuts off and this takes between 10 and 20 sec depending on probe model. To perform a new measurement go back to Paragraph 4. 7-Cooling the Head Should the probe reach its limit temperature, the LCD shows COOL and the LED blinks red. Put the absorber head (only the head!) into water to cool it down. When the temperature falls below its maximum allowable limit, the LCD displays the last measured value and when the bar graph shows few bars, accurately dry the absorber by dry and clean air; the probe is ready for new measurements. 8-Auto Off /Battery Warning As the battery pack is about getting exhausted, a battery symbol pops up on the LCD: it warns you have approx. 10 hours of residual power. LaserPoint s.r.l. Cronos Instruction Manual All Rights are Reserved 5

Note: to increase the battery life, the instrument automatically switches off after 5 minutes of stand by operation. However the last recorded value is not lost; to retrieve it, go back to Paragraph 2 and follow the instructions. After the internal auto check cycle, the display will show the last measured value. To intentionally shut the instrument off, keep the ON/Zero button pushed until the LCD shuts off (this takes approximately 4 seconds). 9-Out of Range Notice When the head is exposed to a source exceeding its full scale, the LCD shows ERR instead of a numeric value. 10-Wavelength Selection To select the wavelength of operation (from CO2 to YAG or viceversa), go back to Paragraph 2 (shut the instrument off), but keep the ON/Zero button pushed until the wavelength change appears on the display (this step takes approximately 7 seconds). At this point release the button and the instrument is ready for a new measurement.. The displayed value is the last recorded power measurement carried out at this wavelength setting. 11-Customer s calibration This instrument can be re-calibrated by the user to match a custom reference. Remove the plastic cover of the adjustment switch located at the right side of the instrument body. Zero the instrument as indicated in Paragraph 2 of this Operation Section and afterwards carry out a measurement of a known laser source. Shift the switch on + or if you need to increase or decrease the reading displayed on the LCD to match your reference. This is done by limited pushes on the ON/Zero key, until the desired power value is achieved on the display. Reset the switch to its middle position CAL and re-cap. The instrument is ready for new measurements. LaserPoint s.r.l. Cronos Instruction Manual All Rights are Reserved 6

Notice to European Union Customers: European Union Waste Electrical and Electronic Equipment (WEEE) Directive 2002/96/EC The European Parliament will enforce new Directives in 2006 concerning the disposal of electrical and electronic equipment such as the Monitoring and Control Instruments. The regulations place responsibilities on the instruments supplier and the purchaser/user. One of the actions required by suppliers is to inform users of their obligations. The present instrument has been assessed in accordance with the European Parliament Directive2002/96/EC on Waste Electrical and Electronic Equipment, usually referred to as the WEEE Directive. The WEEE Directive requires that the instrument is disposed of at the end of its useful life in an environmentally responsible manner. Parts and materials that can be re-used and/or re-cycled have been identified in order that the use of new resources and the amount of waste going for landfill can be minimised. The WEEE Directive requires that if you are replacing the instrument with a new equivalent product from the original supplier, then that supplier can collect the old item without cost to yourself. It is emphasised that this no-cost return option is only available when you purchase a new product of equivalent type that fulfils the same function. Please inform your supplier of your wish to have the old instrument collected when ordering the replacement. If you wish to dispose of the instrument without replacing it then the appliance must not be mixed with unsorted municipal waste. The crossed-out wheeled bin symbol on the unit label or unit packaging, as shown in the table below, indicates this requirement. You must ensure that the instrument is disposed of at an authorised treatment facility; details can be obtained from your local council. Your rôle is critical and will help to ensure that the Earth s resources are maintained and that as much re-usable and re-cyclable material as possible is processed. It will also ensure that landfill volume requirements are kept at a minimum and that hazardous materials are not buried thereby providing potential future problems for the environment and human health. Wheeled bin symbol and description The symbol apply to all LaserPoint electronic products placed on the European Market after August 13, 2005. LaserPoint will affix the wheeled bin symbol with a bar to appropriate products. The bar indicates that the product was placed on the market after August 13, 2005. The WEEE Directive comes into force in January 2006. LaserPoint s.r.l. Cronos Instruction Manual All Rights are Reserved 7

Technical Specifications (*) : Parameter LP1.5 LP5.0 LP10 Maximum Measurable Power (W) 1500 5000 10000 Minimum Measurable Power @ ±4% Accuracy (W) (1) 150 (1) 500 (1) 1000 Absolute Minimum Measurable Power (W) 30 100 200 Power density damage threshold @ full scale at CO 2 wavelength (W/cm 2 ) 2600 1900 1500 Repeatability ±2% ±5% ±5% Measurement Accuracy ±4% Resolution (W) 1 Maximum allowable absorber temp. ( C) 150 Time to measure and display data at full scale power (s) 10 Time to measure and display data at min. scale power (s) 15 Power consumption in On status (mw) 26 Power consumption in Off status ( W) 25 Power supply (2x AA Batteries) (V) 3 Continuous operation without battery replacement (h) 200 Operating temperature range ( C) 10 to 40 Storage temperature range ( C) -10 to 60 Weight (g) 480 950 1300 Length with handle (mm) 300 310 320 (*) Specifications subject to change without notice. (1) Individually compared against a NIST or PTB reference. Warranty All Laser Point instruments are warranted against defect in material and workmanship for one year after invoice date. During the warranty period Laser Point will repair, or at his sole option, replace at no charge components that prove to be defective, provided the parts are returned, shipping prepaid, to Laser Point Customer Service. This warranty does not apply if the instrument or any accessory has been damaged by accident or misuse, or as a result of service or modification by other body than Laser Point. No warranty is given by Laser Point on coatings and battery pack. LaserPoint s.r.l. Cronos Instruction Manual All Rights are Reserved 8