Integrity 270 Integrity 270VP INTEGRITY STEAM STERILIZER INTRODUCTION

Similar documents
AMSCO CENTURY SMALL STERILIZERS DOMESTIC

AMSCO EVOLUTION STEAM STERILIZERS - NORTH AMERICA

AMSCO C SERIES SMALL STEAM STERILIZERS DOMESTIC

AMSCO CENTURY MEDIUM STEAM STERILIZER 26 X 37.5" NORTH AMERICA

Sterilizer Model LS2020 Series Beta Star Life Science Products of R-V Industries

AZTECA A LITRE STERILISER

Castle. Model 233HC Vacuum and Gravity Steam Sterilizer for Healthcare Applications. data and specifications

AMSCO CENTURY MEDIUM STEAM STERILIZER 26 X 26" (SCIENTIFIC)

AMSCO REMANUFACTURED MEDIUM STERILIZERS

AMSCO CENTURY MEDIUM STEAM STERILIZER 26" X 37.5" (SCIENTIFIC)

OPERATOR MANUAL Amsco Century TM Medium Steam Sterilizer 26" x 26" (660 x 660 mm) Scientific. (07/16/07) P Rev. 0

Control Operating Costs. Gain Process Efficiency.

Operating Manual "Mark II" Solid-State Digital Automatic Controlled Sterilizers

Open the Door to Increased Productivity and Cost Efficiency

AMSCO C Series Small Steam Sterilizers 16 x 16" (406 x 406mm) 20 x 20" (508 x 508 mm)

STERITITE INSTRUCTIONS FOR USE

SMALL STERILIZER SERIES 26 x 26 MODEL

SMALL STERILIZER SERIES LSII MODEL (20 x 20 )

AUTOMATIC STERILIZERS

LSII /51 STERILIZER MACHINE SPECIFICATION A REV DATE REVISION BY CHK APP 0 5/18/2016 RELEASED FOR INFORMATION JPK GBJ KWH

BASIL 6000 TUNNEL CAGE WASHER

AMSCO V-PRO MAX LOW TEMPERATURE STERILIZATION SYSTEM (For Use Outside U.S.A. Only)

Best Practice Guidelines For Cleaning, Disinfection and Sterilization of Critical and Semi-critical Medical Devices

Commercial Semi-Instantaneous Water Heaters

cgmp STEAM STERILIZERS

Operations Manual. Part Number: For Research Use Only

AMSCO STERILIZATION CONTAINER SYSTEM User s Guide

LTE Scientific Ltd. Touchclave-PL Porous Load Steam Sterilizer

Consolidated Sterilizers Designed to Transform Your Laboratory

VS LD series vertical steam sterilizers

OPERATOR MANUAL Amsco Eagle 3000 Series FLJ Floorloading Vacamatic Sterilizers. (10/05/01) P Rev. 7

BASIL 6700 TUNNEL CAGE WASHER

Available from. M3 ULTRAf AST. Phone (502) Toll-free (800) Online

M9 M /M3 11/ Automatic Sterilizers

OPERATION & MAINTENANCE MANUAL

RELIANCE 400 LABORATORY GLASSWARE WASHER

VHP ARD BIODECONTAMINATION SYSTEM

Instrument Processing. Midmark Animal Health Products

ESTEEM AUTOCLAVE SLIDING SINGLE DOOR:-

OPERATION & MAINTENANCE MANUAL

Across-the-Line SERIES MP300 Combined Manual and Automatic

Operation and Maintenance Manual

APPENDIX 4. Thank you for choosing the Biomega BioClave steam sterilizer. HYDRAULIC DRAWING

FLASH PAK STERILIZATION CONTAINER SYSTEM

TECHNICAL SPECIFICATION 44 and 55 Compact Sterilizer Series

Steam sterilizer AHS-75-DRY

MARK-COSTELLO WASTE STERILIZER SPECIFICATION SHEET

CLINOCLAV steam sterilizer AH-21 L

2364 Leicester Road, P.O. Box 175, Leicester, NY Phone (585) Fax (585)

MIDMARK INSTRUMENT PROCESSING

Consolidated Sterilizers Designed to Transform Your Laboratory

Soft Start Series MP700 Solid State, Reduced Voltage

Large Steam Sterilizers for Life Sciences

Consolidated Sterilizers Designed to Transform Your Laboratory

PRE AND POST VACUUM CLASS B AUTOCLAVE

Variable Frequency Drive SERIES MP800 VFD

M4.9. NO LONGER IN PRODUCTION Some service parts may not be available for this product- Service and Parts Manual M4.9-0

NC SERIES. Horizontal Steam Sterilizers

Advance Autoclave. Designed to give you more for your money with a high capacity for faster throughput: More features and benefits...

LTE Scientific Ltd. Touchclave-R. The dynamic new range of cylindrical-section autoclaves from LTE Scientific

COMMERCIAL WASHERS MODELS T-300/350/400/450/600/650/750/900/950/1200/1450 VENDED C-SERIES CONTROL

RELIABLE. The Midmark M9 and M11 UltraClave Automatic Sterilizers - make sterilization reliable and efficient.

AUTOCLAVE STEAM STERILIZERS (BENCH TOP)

RELIANCE VISION SINGLE-CHAMBER WASHER/DISINFECTOR

Operation and Maintenance Manual

EZPlus Fully Automatic Sterilizers

OPERATION MANUAL. Move it. Hang it. Rack it. Stack it. MaxPower Corporation 230 Yuma Street Denver, CO

Consolidated Sterilizers Designed to Transform Your Laboratory

DIVISION FINNED WATER-TUBE BOILERS

MST sterilizer overview

Autoclave Operations Manual

CENTRAL STERILISATION. Innovative, hardworking Sterilisation & Infection Control products perfect for your Central Sterile Supply Department

Tabletop 2340 EA-D Class S Autoclave 19 Litres

OT 300/430/570 STEAM STERILIZERS

WATER-TUBE POOL HEATERS

FN 032/055/120 DRY HEAT STERILIZERS / OVENS FN 300/400/500 DRY HEAT STERILIZERS / OVENS KD 200/400/700 OVENS EV 018 VACUUM OVEN

DIVISION FINNED WATER-TUBE BOILERS

OT 300/430/570/710/850 STEAM STERILIZERS

Process Water Heater For Industrial Applications All Voltages, Single or Three Phase Up To 88 KW

OPERATION & MAINTENANCE MANUAL

Product Data. Features/Benefits. 19XB Positive Pressure Storage System

: 1.5kw/AC220V.50Hz (AC110V 60Hz) : 340*200*30mm (3piece) 400*200*30mm (3 piece)

WATER-TUBE POOL HEATERS

XTHERM, TYPE WH - MODELS 1005A- 2005A DIVISION COMMERCIAL, WATER-TUBE, CONDENSING GAS DOMESTIC WATER HEATERS

Kronos Class B Sterilizer

HWD-2105 Hot Water Dispenser. NOTICE TO INSTALLER: Please leave this book with the machine.

Title: YALE OFFICE OF FACILITIES PROCEDURE MANUAL Chapter: 01 - Yale Design Standard Division: HVAC Standards

Refrigerated Incubator Model and Operating Instructions

DIVISION COMMERCIAL, GRID-TYPE, FINNED-TUBE, GAS DOMESTIC WATER HEATERS

DIVISION FINNED WATER-TUBE BOILERS

INDIRECT FIRED, SEMI-INSTANTANEOUS STEAM & BOILER WATER HEATER. Compact Design. High Recovery. Energy Efficient. Stainless Alloy Construction

Consolidated Sterilizers Designed to Transform Your Laboratory

LabClave.Com OPERATION & MAINTENANCE MANUAL. Horizontal Autoclave Model: Z-LABCLAVE-01-1 S/N

HWD-2110 Hot Water Dispenser

Advance Autoclave. Designed to give you more for your money with a high capacity for faster throughput: More features and benefits...

Series 151 and AV5. A wide range of process thermostatic steam traps is available to match almost any application need.

SECTION ROTARY-SCREW WATER CHILLERS

A WORLD OF QUALITY, SAFETY AND HYGIENE STERILCLAVE. 18 litres-bhd/b/s

FINNED WATER-TUBE POOL HEATERS

Operation and Maintenance Manual

Transcription:

5085 Corporate Exchange Blvd. S.E. Grand Rapids, MI 49512 616.656.2900 Integrity 270 Integrity 270VP INTEGRITY STEAM STERILIZER INTRODUCTION The SKYTRON Integrity 270 sterilizers utilize gravity displacement and pressure/vacuum pulse cycles for sterilization. Each Integrity sterilizer is equipped with prevacuum, gravity, immediate use and express prevacuum cycles together with leak test, self-diagnostic cycle test, warm up and Bowie-Dick test cycles. Liquid cycles are available as an option. The Integrity Series sterilizers are equipped with 12 factory programmed cycles and 12 fully customizable cycles which can be added, modified and renamed with password access. All sterilization parameters are monitored for deviation of temperature, pressure and time by the control system. APPLICATION The Integrity 270 sterilizers incorporate high-pressure steam to sterilize heat and moisture stable devices and materials used in the health care industry. Various sterilization cycles are capable of processing non-porous and porous items wrapped or unwrapped, including towel, gown or other linen packs. Optional liquid cycles are capable of sterilizing liquids in borosilicate containers with vented closures. FEATURES Spacious 26.5 x 37.5 x 53 (675mm x 955mm x 1350mm) sterilization chamber designed to accommodate a carriage and loading cart with a wide range of trays, containers, packs and devices of various sizes. Standard water recirculation system is equipped on all Integrity Series sterilizers within the installation footprint. The integrated water recirculation system drastically reduces the amount of water consumed per cycle up to 33% compared to existing units without affecting cycle times. Additional water conservation can be achieved by lengthening overall cycle times. The Integrity 270 has been validated to process up to sixteen (16), 25 pound (11.36 kg) instrument trays. Please see section titled Processing Cycles for specific load recommendations for each cycle. The Integrity 270 has been validated to process a maximum of three (3) lumen instruments per cycle load, with an inside diameter of 2 mm (minimum) and 400 mm (maximum) length with Immediate Use 2, Prevacuum 270, Prevacuum 275, and Express Prevacuum cycles only (3 mm inside diameter lumens may be processed according to current ANSI/AAMI ST8:2008). TEC-F-0521 REV1 11/18/2015

Page 2 A 5.7 (145 mm) touch panel screen displays easy to use operator controls, operating conditions, service interface panel and troubleshooting information. Standard electronic cycle data recording to CF card in addition to traditional ink-to-paper printed receipt. Steam actuated door gasket reduces the potential for incoming air to the chamber. Bürkert Solenoid Valves with LED indicators are used to assist with service troubleshooting and eliminate the need for compressed air. The Integrity 270 features a dual walled, fully jacketed, all welded rectangular chamber to reduce the potential for cool spots inside the chamber compared to partially jacketed sterilizers. 12 open slots to create custom sterilization cycles. 12 factory programmed cycles are capable of modification with password access. Password protected access to supervisor and service modes. Auto run timer, daily timer and operator identification functionality. Standard electric-powered door locking/unlocking system functions with only four moving parts Fully integrated water recirculation system reduces water usage, stabilizes water pressure while also accounting for drain temperature requirements OPTIONS - Vacuum pump/voltage requirements Electrical connection required for vacuum pump: 208V 3Ph, 15 Amp, 60 Hz 480V 3Ph, 7 Amp, 60 Hz - Install location Recessed sterilizer is mounted inside a wall opening with service accessibility provided behind finished wall. Double Door sterilizer is mounted between one or two walls with pass through capability. Service accessibility provided on one side or between the two walls. Freestanding/cabinet stainless steel side panels and frame provided to allow stand alone room installation. Seismic tie-down kit/anchorage assembly Seismic anchoring requirements per California Building Code (CBC). Liquid cycle Optional liquid cycle and liquid temp load probe. STANDARDS ASME Code, Section VIII, Division 1 for unfired pressure vessels, shell and door are constructed to withstand working pressure of 45 psig. USA UL 61010-1, IEC61010-2-040:2005 as certified by Met Laboratories Inc. CSA C22.2 No. 61010-1 Factory programmed cycles have been performance validated to ANSI/AAMI ST-8:2008 Vacuum pump option offers high efficiency evacuation of air and steam while further reducing water consumption. Engineering Data Model Chamber Size inch (mm) Heating Total Weight lbs (kg) Freestanding Heat Loss BTU/hr at 86 F (30 C) Operator Side Recessed Service Side Unload Side Integrity 270 26.5 x 37.5 x 53 House 3223 (1465) 7032 2563 4470 N/A Integrity 270VP (675 x 955 x 1350) Steam 3685 (1675) N/A 2563 4470 2691

SKYTRON Integrity 270 Order Configuration Page 3 Model Control Power Vacuum High Voltage Door Installation Integrity 270 Integrity 270DD Integrity 270VP Integrity 270VPDD 120V, 60Hz, 10A Venturi Pump 120V, 60Hz, 3A Vacuum Pump N/A 208V, 3pH, 60Hz, 15A 480V, 3 ph, 60Hz, 7A 208V, 3pH, 60Hz, 15A 480V, 3 ph, 60Hz, 7A Single Double Single Double Recessed Freestanding Recessed Recessed Freestanding Recessed Freestanding Recessed Door Swing * * * * * * Mechanical Cart Equipment Number ASSR-AO12-RRSR ASSR-AO12-RRSL ASSR-AO12-RLSR ASSR-AO12-RLSL ASSR-AO12-FRSR ASSR-AO12-FRSL ASSR-AO12-FLSR ASSR-AO12-FLSL ASSR-AO12W-RRSR ASSR-AO12W-RRSL ASSR-AO12W-RLSR ASSR-AO12W-RLSL ASSR-AO12P-208R-RSR ASSR-AO12P-208R-RSL ASSR-AO12P-208R-LSR ASSR-AO12P-208R-LSL ASSR-AO12P-208F-RSR ASSR-AO12P-208F-RSL ASSR-AO12P-208F-LSR ASSR-AO12P-208F-LSL ASSR-AO12P-480R-RSR ASSR-AO12P-480R-RSL ASSR-AO12P-480R-LSR ASSR-AO12P-480R-LSL ASSR-AO12P-480F-RSR ASSR-AO12P-480F-RSL ASSR-AO12P-480F-LSR ASSR-AO12P-480F-LSL ASSR-AO12PW-208R-RSR ASSR-AO12PW-208R-RSL ASSR-AO12PW-208R-LSR ASSR-AO12PW-208R-LSL ASSR-AO12PW-480R-RSR ASSR-AO12PW-480R-RSL ASSR-AO12PW-480R-LSR ASSR-AO12PW-480R-LSL * refers to door swing on operational side, opposite for unload side

Page 4 PROCESSING CYCLES Immediate Use 1 (3 minute exposure time) The Immediate Use 1 cycle program will sterilize one (1) unwrapped, non-porous instrument no heavier than 0.22 lbs (100 grams) This program provides quick sterilization for accidently dropped instruments for immediate use. Porous, lumened or cannulated items cannot be processed using the Immediate Use 1 cycle. Exposure temperature is 270 F (132 C) for 3 minutes, 1 minute dry time. This cycle has been validated to AAMI ST-8:2008. Immediate Use 2 (4 minute exposure time) The Immediate Use 2 cycle program is recommended for processing sixteen (16), 25 pound (11.36 kg) unwrapped instrument trays and up to three (3) lumen instruments (2 mm ID minimum by 400 mm long maximum). This program provides quick sterilization for immediate use. Exposure temperature is 270 F (132 C) for 4 minutes, 1 minute dry time. This cycle has been validated to AAMI ST-8:2008. Express Prevacuum (Immediate Use 3) The Express Prevacuum cycle program is recommended for processing sixteen (16), single wrapped, 25 pound (11.36 kg) instrument trays and up to three (3) lumen instruments (2 mm ID minimum by 400 mm long maximum). Peel pouches are not acceptable for sterilization with an Express cycle program. Sterilized items are intended for immediate use. The single wrapper used with the Express cycle program protects the sterilized item from contaminants en route from the Integrity 270 to the point of use. Exposure temperature is 270 F (132 C) for 4 minutes, 3 minute dry time. This cycle has been validated to AAMI ST-8:2008. Prevacuum 270 The Prevacuum 270 cycle program is recommended for processing sixteen (16), double-wrapped, 25 pound (11.36 kg) instrument trays and up to three (3) lumen instruments (2 mm ID minimum by 400 mm long maximum) Exposure temperature is 270 F (132 C) for 4 minutes, 30 minute dry time. This cycle has been validated to AAMI ST-8:2008. Prevacuum 275 The Prevacuum 275 cycle program is recommended for processing sixteen (16), double-wrapped, 25 pound (11.36 kg) instrument trays and up to three (3) lumen instruments (2 mm ID minimum by 400 mm long maximum) Exposure temperature is 275 F (135 C) for 3 minutes, 30 minute dry time. This cycle has been validated to AAMI ST-8:2008. Gravity 1 The Gravity 1 cycle will sterilize a maximum of thirty two (32) fabric packs. Exposure temperature is 250 F (121 C) for 30 minutes, 15 minute dry time. This cycle has been validated to AAMI ST-8:2008. Gravity 2 The Gravity 2 cycle is recommended for processing sixteen (16), 25 pound (11.36 kg) double-wrapped instrument trays. Exposure temperature is 270 F (132 C) for 15 minutes, 30 minute dry time. This cycle has been validated to AAMI ST-8:2008. Custom cycles The custom cycle programs allow the customer to modify the parameter values to their unique specifications. The operator/supervisor/ service technician may create twelve custom programs using any of the eight (8) program icons. It is the customer s responsibility to validate all custom cycle programs for sterilization assurance. The range for minimum steam exposure time will be determined by the selected steam exposure temperature. Liquid cycle (optional) The Liquid cycle program will sterilize a maximum of three (3) borosilicate containers with vented closures. Exposure temperature is 250 F (121 C) for 45 minutes, dry time is not applicable. This cycle has been validated to AAMI ST-8:2008. Liquid cycles are not intended for the sterilization of materials that come into direct contact with patients.

Page 5 TEST CYCLES Warm up The Warm up cycle program warms the sterilizer chamber before daily operation. Exposure temperature is 270 F (132 C) for 4 minutes, 3 minute dry time. Do not use this program to sterilize materials. BD test The BD Test cycle is utilized to perform the Bowie-Dick Test on the sterilizer. Do not use this program to sterilize materials. Exposure temperature is 273 F (133.8 C) for 3 ½ minutes, 2 minute dry time. This cycle has been validated to AAMI ST-8:2008. Cycle test The Cycle Test program verifies the repeatability of the operation cycles. The Cycle Test is a sterilizer-performance test conducted by authorized service technicians. Exposure temperature is 270 F (132 C) for 15 minutes, 15 minute dry time. Do not use this program to sterilize materials. Leak test The Leak Test cycle program verifies the vacuum integrity of the sterilizer. The Leak Test cycle is a sterilizer-performance test conducted by authorized service technicians. The chamber must be empty for this test. Exposure temperature is 270 F (132 C) for 4 minutes, 3 minute vacuum time, 5 minute keep time and 15 minute test time. This cycle has been validated to AAMI ST-8:2008. CONTROLS LCD touch screen display The 5.7 (145 mm) color liquid crystal touch panel displays easy to use operator controls, operating conditions, service interface panel and troubleshooting information. Countdown timer and progression bar During cycle operation, the LCD touch screen displays a large countdown timer in addition to a progression bar to indicate the remaining cycle time during operation. Operator/Supervisor/Service Mode The Integrity 270 has the ability to restrict access to custom cycle creation, sterilizer settings and service menus by utilizing the password identification feature. Operator ID The Integrity 270 has the ability to enable the Operator ID feature that requires an up to 8 digit, alpha-numeric login when starting a sterilization or test cycle program. The operator ID will then appear on the cycle program printout and operation history. Auto run timer Allows the operator/supervisor/ service technician to set the starting time for an automatic sterilization or test cycle. Daily start up/shut down timer The operator/ supervisor/service technician has the ability to program the daily start-up (power-up) and daily turn-off (power-down) times for each day of the week. The power-down mode functions as an Energy Saver by closing the incoming steam supply to conserve utilities. Printed cycle reports Standard ink to paper impact printer equipped on all Integrity sterilizers to provide a hard copy cycle report. Printed cycle reports provide the start time, date, cycle count, sterilizer number, cycle type, temperature and pressure information during the cycle. Printed cycle reports also provide the operator ID and/or cycle error information when applicable. Electronic data recording The Integrity Sterilizer Compact Flash (CF) interface captures cycle report data electronically. The data is saved as a.csv file that can be moved to a personal computer for analysis and conversion to a graph in Microsoft Excel. Skytron recommends retrieving the cycle report files and saving to a PC on a monthly basis to avoid losing any recorded cycles. On screen troubleshooting The 5.7 (145 mm) LCD touch screen display features an error warning and on-screen troubleshooting system that can help identify a possible fault and reduce the amount of downtime for repair. Accessibility to this menu can be set for password protection. CONSTRUCTION Pressure vessel and steam jacket The Integrity 270 sterilizer features a rectangular, fully jacketed, pressure vessel with a maximum effective chamber capacity of 32.3 ft 3 (916 L). The sterilization chamber, flange and door plate are constructed of SA-240 grade 316L stainless steel. The full steam jacket and reinforcement material are constructed of SA-240 grade 304L Stainless Steel. The vessel socket is manufactured with SA-479 Type 316L Stainless Steel. Maximum working pressure for the chamber and jacket is 45 psi (310 kpa). Piping A combination of Stainless Steel SUS304TP screwed joint and copper tubing C1220T ring joints are used throughout the steam and water piping system.

Page 6 Integrity 270 Single Door, Venturi Pump, Recessed Installation Door Swing & Mechanical Cart shown Door Swing & Mechanical Cart available, see Order Configuration Chart (For Door Swing & Mechanical Cart mirror all Plan View dimensions from the center line and the wall opening) 24 (610 mm) 24 (610mm) 25 1 /4 (642 mm) 31 1 /2 (800 mm) 49 1 /4 (1251mm) 17 11 /16 (449mm) 18 7 /8 (479mm) STERILIZER Center Line Chamber Door Swing 29 (736 mm) 30 1 /4 (768mm) 38 (965 mm) 104 3 /4 (2661mm) Drawings are not to scale. Integrity 270 Sterilizers require a model-specific detailed installation drawing package provided by SKYTRON. Minnimum clearances shown are recommended for installing and servicing the sterilizer. Access to recessed area from the control end of the sterilizer is recommended. UTILITIES COLD WATER SUPPLY 3 / 4 NPT Fitting 24 (610 mm) Venturi Pump Mechanical Cart 55 1 /2 (1410 mm) 36 1 /2 (768mm) STEAM SUPPLY RECIRCULATION TANK MANUAL DRAIN 3 /8 ODT RECIRCULATION TANK DRAIN 1 1 /2 NPT Fitting JACKET SAFETY VALVE PLAN VIEW 79 1 /16 (2008 mm) 6 3 /4 (171 mm) 61 3 /4 (1569 mm) 4 3 /16 (106mm) 44 (1118 mm) Front Panel Swing 55 (1397 mm) 82 11 /16 (2100 mm) 74 1 /4 (1886 mm) 78 7 /16 (1992 mm) SIDE VIEW FRONT VIEW

Integrity 270 Single Door, Vacuum Pump, Recessed Installation Door Swing & Mechanical Cart Shown Door Swing & Mechanical Cart Available, See Order Configuration Chart (For Door Swing & Mechanical Cart mirror all Plan View dimensions from the center line and the wall opening) Page 7 24 (610 mm) 24 (610 mm) 37 3 /8 (948 mm) 26 5 /8 (677 mm) Vacuum Pump Mechanical Cart 61 3 /8 (1558 mm) 18 7 /8 (479mm) 17 11 /16 (449mm) 29 (736 mm) Chamber Door Swing STERILIZER Center Line 42 3 /8 (1076 mm) 38 (965 mm) 112 (2844 mm) Drawings are not to scale. Integrity 270 model sterilizers require a detailed installation drawing package provided by SKYTRON. Clearances shown are recommended for installing and servicing the equipment. Access to the recessed area from the control end of the sterilizer is recommended. UTILITIES COLD WATER SUPPLY 3 / 4 NPT Fitting STEAM SUPPLY RECIRCULATION TANK MANUAL DRAIN 3 /8 ODT x 8 L Tube 24 (610 mm) 50 5 /8 (1287 mm) 31 5 /8 (803 mm) RECIRCULATION TANK DRAIN 1 1 /2 NPT Fitting JACKET SAFETY VALVE PLAN VIEW 70 3 /4 (1798mm) 6 3 /4 (171 mm) 61 3 /4 (1569 mm) 4 3 /16 (106mm) 44 (1118 mm) Front Panel Swing 55 (1397 mm) 82 11 /16 (2100 mm) 74 1 /4 (1886 mm) 78 7 /16 (1992 mm) SIDE VIEW FRONT VIEW

Page 8 Integrity 270 Double Door, Venturi Pump, Recessed Installation / Door Swing & Mechanical Cart Shown (/ Door Swing & Mechanical Cart Available, See Order Configuration Chart) (For / Door Swing & Mechanical Cart mirror all Plan View dimensions from the center line and the wall openings) 30 1 /4 (768mm) 24 (610mm) 30 1 /4 (768mm) Drawings are not to scale. Integrity 270 model sterilizers require a detailed installation drawing package provided by SKYTRON. Clearances shown are the minimum for installing and servicing the equipment. 104 3 /4 (2662mm) 38 (965mm) 36 1 /2 (927mm) 18 7 /8 (479mm) 17 11 /16 (449mm) 25 1 /4 (642mm) 31 1 /2 (800mm) 24 (610mm) Venturi Pump Mechanical Cart 49 1 /4 (1252mm) 55 1 /2 (1410mm) 18 7 /8 (479mm) 17 11 /16 (449mm) 29 (736mm) Chamber Door Swing STERILIZER Center Line 38 (965mm) 36 1 /2 (927mm) 104 3 /4 (2662mm) Access to the recessed area from the control end of the sterilizer is recommended. UTILITIES COLD WATER SUPPLY 3 / 4 NPT Fitting STEAM SUPPLY RECIRCULATION TANK MANUAL DRAIN 3 /8 ODT x 8 L Tube RECIRCULATION TANK DRAIN 1 1 /2 NPT Fitting JACKET SAFETY VALVE PLAN VIEW 6 3 /8 (161mm) 68 13 /16 (1748mm) 56 1 /16 (1322mm) 6 3 /8 (161mm) 4 3 /16 (106mm) 44 (1118mm) Front Panel Swing 35 3 /16 (894 mm) 82 11 /16 (2100mm) 82 11 /16 (2100mm) 74 1 /4 (1886mm) 74 1 /4 (1886mm) 78 7 /16 (1992mm) SIDE VIEW FRONT VIEW

Integrity 270 Double Door, Vacuum Pump, Recessed Installation / Door Swing & Mechanical Cart Shown (/ Door Swing & Mechanical Cart Available, See Order Configuration Chart) (For / Door Swing & Mechanical Cart mirror all Plan View dimensions from the center line and the wall opening) Page 9 24 (610 mm) Drawings are not to scale. Integrity 270 model sterilizers require a detailed installation drawing package provided by SKYTRON. 61 3 /8 (1558mm) Vacuum Pump Mechanical Cart 30 1 /4 (768mm) Clearances shown are recommended for installing and servicing the equipment. Access to the recessed area from the control end of the sterilizer is recommended. 112 (2844mm) 17 11 /16 (449mm) 18 7 /8 (479mm) 37 3 /8 (948 mm) 26 5 /8 (677mm) 61 3 /8 (1558mm) 50 5 /8 (1287mm) 17 11 /16 (449mm) 18 7 /8 (479mm) STERILIZER Center Line Chamber Door Swing 29 (736 mm) 38 (965 mm) UTILITIES COLD WATER SUPPLY 3 / 4 NPT Fitting STEAM SUPPLY RECIRCULATION TANK MANUAL DRAIN 3 /8 ODT x 8 L Tube 50 5 /8 (1287mm) 24 (610mm) 36 1 /2 (927mm) RECIRCULATION TANK DRAIN 1 1 /2 NPT Fitting JACKET SAFETY VALVE PLAN VIEW 68 13 /16 (1748mm) 6 3 /8 (161mm) 56 1 /16 (1322 mm) 6 3 /8 (161mm) 4 3 /16 (106mm) 44 (1118 mm) Front Panel Swing 35 3 /16 (894 mm) 82 11 /16 (2100 mm) 74 1 /4 (1886 mm) 74 1 /4 (1886 mm) 82 11 /16 (2100 mm) 78 7 /16 (1992 mm) SIDE VIEW FRONT VIEW

Page 10 Integrity 270 Single Door, Freestanding Installation Door Swing & Mechanical Cart Shown ( Door Swing & Mechanical Cart Available, See Order Configuration Chart) (For Door Swing & Mechanical Cart mirror all Plan View dimensions from the center line and the wall openings) 24 (610 mm) 52 5 /16 (1329 mm) Clearances shown are the minimum for installing and servicing the equipment. Drawings are not to scale. Integrity 270 model sterilizers require a detailed installation drawing package provided by SKYTRON. 28 5 /16 (719 mm) UTILITIES COLD WATER SUPPLY 3 / 4 NPT Fitting Vacuum Pump Mechanical Cart STERILIZER Center Line 29 (736 mm) Chamber Door Swing 40 5 /16 (1329 mm) 64 1 /8 (1630 mm) 116 1 /2 (2959 mm) STEAM SUPPLY RECIRCULATION TANK MANUAL DRAIN 3 /8 ODT x 8 L Tube RECIRCULATION TANK DRAIN 1 1 /2 NPT Fitting JACKET SAFETY VALVE 24 (610 mm) PLAN VIEW 68 1 /2 (1740 mm) 63 3 /4 (1620 mm) 17 1 /16 (433mm) 44 (1118 mm) Front Panel Swing 6 5 /8 (168mm) 74 1 /2 (1892 mm) SIDE VIEW FRONT VIEW

Page 11 SAFETY FEATURES Alarms An error detected by the sterilizer control system will result in a related error message on the touch screen display. An audio alarm buzzer will sound and any cycle operating at the time will be aborted. Safety valve(s) ASME Section I/Section VIII capacity certified pressure release valves are in place to prevent the sterilizer from exceeding the rated pressure limit of the vessel. Door safety The door safety switch prevents the steam supply from entering the chamber unless the door is closed and sealed/locked. The door interlock also prevents the door from unlocking if the chamber pressure is not at atmospheric pressure. Emergency E-stop knob Pressing the emergency E-stop knob will terminate the operation of the sterilizer and return the chamber to atmospheric pressure. Circuit protector Electrical current is cut-off in the event of a short circuit or over-current condition in the control circuit of the sterilizer. A circuit breaker activation interrupts the power supply if the current exceeds the rated amperage to the circulation pump motor, door motor and optional vacuum pump motor. Preventative maintenance Annual maintenance and service agreements are available to help assure the reliability and low cost of ownership associated with SKYTRON products. UTILITY REQUIREMENTS Steam Supply 1" NPT connection provided. 1 facility piping, 50-80 PSIG (0.34-0.55 MPa) dynamic pressure, 220 lb/hour (100 kg/hour) flow rate. Recirculation Tank Overflow Drain 1½ NPT connection provided. Floor sink with minimum 2 ID drain to accommodate recirculation tank overflow recommended. Water drain temperature may exceed 140 F (60 C) if instructions are not followed. Recirculation Tank Manual Drain ⅜ ODT. Cold Water Supply ¾ NPT connection provided. 30 to 50 PSIG dynamic (0.21-0.34 MPa), 50º to 70º F (10 to 25 ºC) Total hardness as CaCO², 50-120 mg/l and max 171 mg/l, Total dissolved solids 100-200 mg/l and max 500 mg/l, ph 6.8-7.5 and max 6.5-8.5. Steam Pressure Relief Valve 1" NPT outlet for jacket. System Electrical Power 120 VAC, 60 Hz, single phase, 15A power cord with 3 prong grounded plug provided, dedicated outlet recommended. Vacuum Pump (Optional) Electrical Power 208 VAC, 60 Hz, 3 phase, 15A 480 VAC, 60 Hz, 3 phase, 7A Utilities Consumption Water Steam Model Chamber Size inch (mm) W x H x D Heating Peak gpm (lpm) Maximum Consumption gal/cycle (l/cycle) Average Consumption gal/cycle (l/cycle) Per cycle lb/cycle (kg/cycle) Integrity 270 276 (1046) 184 (696) 26.5 x 37.5 x 53 House 10.6 (40) (675 x 955 x 1350) Steam Integrity 270VP 214 (809) 138 (524) 45.2 (20.5) The base language for this document is ENGLISH. Any translations must be from the base language document. Printed copies are not controlled documents.

REVISION HISTORY Date Revision Revision History 5/2014 0 Initial Release 11/18/2015 1 Pg 5 - Removed the 9,999 statement from the electronic data recordings section The base language for this document is ENGLISH. Any translations must be from the base language document. Printed copies are not controlled documents.