User Guide. ROSS Process Combination ph Electrode

Similar documents
User Guide. Double Junction Reference Electrode

Thermo Scientific Orion COD165 Thermoreactor. User Guide

Thermo Scientific Orion COD125 Thermoreactor. User Guide

Quick Start Guide LIQ-QSG-Hx338, Rev M July Rosemount Hx338. Steam Sterilizable and Autoclavable ph Sensor

ExStik DO600 USER GUIDE. Dissolved Oxygen Meter. 99 Washington Street Melrose, MA Phone Toll Free

User's Guide. ExStik DO600. Dissolved Oxygen Meter RECALL. Dissolved Oxygen DO600

Retractable ph/orp Sensor

Quick Start Guide LIQ-QSG-Hx438, Rev A June Rosemount Hx438. Amperometric Steam Sterilizable Dissolved Oxygen Sensors

Rosemount 372. HF Resistant ph sensor. Quick Start Guide LIQ-QSG-372 Rev. C June 2017

HydroMaster USER S GUIDE HM-500

ACQUITY UPLC HT Column Heater Instructions

ph Measuring System Start-up and maintenance manual e Contents

Tracer DO PockeTester TM Code 1761

Orion ionplus 2111LL low-level sodium monitor. The next generation in sodium protection. Analyze Detect Measure Control

FACSCalibur Startup Procedures

Table of Contents 1. OVERVIEW SYSTEM LAYOUT SPECIFICATIONS FUNCTION... 11

Durafet Non-Glass ph Electrodes Specifications

HP 101A Series Humidity Temperature Probes INSTRUCTION MANUAL

Dissolved Oxygen Sensors

Models TX-KE and TX-KP Toxic Gas Transmitters

Guide for Thin Film Evaporation Edited 7/22/13 by Stephen Albright, BJAM & Steven Palefsky

Digital Salt / TDS Meter. 390a. The Value Leader TM

Durafet II and Meredian II Combination ph/orp Electrodes for Honeywell 7777 Mountings Series 7777 Specifications

Mitocell: Instructions for Use

G Trap FliQ FPLC Columns

K-COMPACT Use & Care Guide. Get the most from your new Keurig brewer

Chanson Water USA, INC. Installation Guide for SCF-3000/SCF-6000

User Manual. Version 3.0

JENCO INSTRUMENTS, INC. JENCO ELECTRONICS, LTD ARJONS DR., SUITE C P.O. BOX LINKOU-117

K-SELECT Use & Care Guide. Get the most from your new Keurig brewer

CryoFit. Handling of Flow Cell and Capillaries. User Guide. Version 002

OPERATION MANUAL. THE JENCO MODEL 6230 MICROCOMPUTER BASED ph/mv/temp PORTABLE METER JENCO ELECTRONICS, LTD. MANUFACTURER OF PRECISION INSTRUMENTS

Use & Care Guide. Get the most from your new Keurig brewer

MP 601A Humidity Probe INSTRUCTION MANUAL

HydroFlux Welder Torch System T0-102 / T0-103

Model Q45P/R. ph/orp Monitor...

User Manual. Version 2.0

Up to 60 Pint Commercial Grade DEHUMIDIFIER

CLEAN MIST DUAL TANK SMART HUMIDIFIER. Instruction Manual and Warranty Information READ AND SAVE THESE INSTRUCTIONS IM0036A

ACROBAT GT OPERATING AND TROUBLESHOOTING GUIDE. Automatic Pool Cleaner. We hope you enjoy your ACROBAT GT automatic pool cleaner!

HTO-45 HUMIDITY TEMPERATURE PROBE INSTRUCTION MANUAL

EVAPORATIVE COOL MIST HUMIDIFIER Instruction Manual

MiniAmp Thermal Cycler and MiniAmp Plus Thermal Cycler Installation and Operation

matt GRD: GRD 28/02/08 9:17 Page 1.

Instruments. for hydroponics & horticulture 2016

JENCO ELECTRONICS, LTD.

Concise catalogue Water-Industry

SimpleVIS Viscometer Reference Guide P

JENCO ELECTRONICS, LTD. MANUFACTURER OF PRECISION INSTRUMENTS

Introduction Portable Meters Transmitter Probes RELATIVE HUMIDITY

Mariner Microspray and Low-Flow Electrospray Fused Silica Spray Tip and Capillary Kit. Operating Instructions. 1 Product Description.

FCHM Hydromatic Pump Control Module

Product waste disposal - Protection of the environment:

Model # VWK-2 INSTRUCTION MANUAL

SECTION Free/Available Chlorine Analysis System

Model Q45P/R. ph/orp Monitor...

CLEAN MIST SMART HUMIDIFIER TABLETOP & FLOOR STANDING WITH REMOTE CONTROL. Instruction Manual and Warranty Information

SKY Stainless Steel Compact Dishwasher. Instructions

CHAMPION PUMP OWNER S MANUAL

I N ST ML (3-PAN, SELF-SERVICE) ML (4-PAN, SELF-SERVICE)

SLUSH PUPPiE MACHINE Instructions For Use

ULTRASONIC DIGITAL HUMIDIFIER Instruction Manual and Warranty Information IM0002A READ AND SAVE THESE INSTRUCTIONS

INSTALLATION & OPERATING INSTRUCTIONS

Operation and Maintenance Manual for the SX-90 & SX-80 Sodr-X-Tractor Handpieces Rev F

FL-1000 FLEC Air Control

CLEAN MIST ULTRASONIC HUMIDIFIER. Instruction Manual and Warranty Information READ AND SAVE THESE INSTRUCTIONS IM0036A

AD XT. Air Dryer Service Manual. Innovation with Integrity. Version 003 NMR

Ion OneTouch Template Kit Publication Part Number Revision E Date 31 October 2011

Assembly of an F12 Hydroponic Tower

OPERATING AND MAINTENANCE MANUAL

CATALOGUE.PDF BLUELAB DOSETRONIC ATC PH PROBE. Code Product Unit RRP

Elite Primer Baldor Series External Pond Pump

MAYFIELD CEILING FAN LISTED E ITEM # MODEL #BTH44ABZC5C BTH44BNK5C Español p. 20 ATTACH YOUR RECEIPT HERE.

Reliable Detection of Transparent Objects

ECONO FLO 2.7HP VARIABLE SPEED PUMP OWNER S MANUAL

SMF PUMP OWNER S MANUAL

SEA FROST BD 12 OR 24-VOLT D.C. AIR-COOLED SYSTEM

Sanitise Syrup Lines & Valves. Taylor PH61 Cleaning every 14 days

Installation and Operation Manual For Hunter Ceiling Fans /16/2004

DL424/425 DirectLine Sensor Module for Dissolved Oxygen Probes Series DL5000 Specifications

Connector Maintenance

95-Pint Dehumidifier with Built-in Pump USER GUIDE IVALDH95PWWP

OPERATING MANUAL BFS-MP SERIES FREEZING STAGES

SERVICING INFORMATION

Electron Microscopy Sciences INSTRUCTIONAL MANUAL CAT A. P.O. Box Industry Road Hatfield, PA 19440

Cleaning the internal optics surfaces

30-pint Dehumidifier User Guide

ENVAIR lab CO2 Tri-Gas Incubator

HFG 12 and 24 Hydrogen Generators Installation Instructions

F AN-ASPIRATED RADIATION SHIELD

Model Super TJE. Ultra Precision Pressure Transducer DESCRIPTION FEATURES

AP Air Ion Counter Instructions

7758 Pipeline Electrode Mountings for Industrial ph and ORP Measurements. Specification. Overview. Description

kéìíê~=iqd=qìåöëíéå=dêáåçéê= lééê~íáçåë=j~åì~ä=

ECONO FLO VSA 165 VARIABLE SPEED PUMP OWNER S MANUAL

MINIMED 670G SYSTEM CHANGING THE MINIMED SILHOUETTE INFUSION SET QUICK REFERENCE GUIDE

Ultrasonic Humidifier

Vertical Gel Systems-Dual CAT NO: EPS-V0011 & EPS-V0012

Parasitic Cysts - Cryptosporidium and Giardia, Chemicals, Heavy Metals, Dissolved Salts, Offensive Tastes and Odours, Sediment, Dirt,

CALIBRATABLE 3-FUNCTION THERMO- HYGROMETER USER S MANUAL

Transcription:

User Guide ROSS Process Combination ph Electrode

ROSS and the COIL trade dress are trademarks of Thermo Fisher Scientific Inc. AQUAfast, Cahn, ionplus, KNIpHE, No Cal, ORION, perphect, PerpHecT, PerpHecTion, phisa, phuture, Pure Water, Sage, Sensing the Future, SensorLink, ROSS, ROSS Ultra, Sure-Flow, Titrator PLUS and TURBO2 are registered trademarks of Thermo Fisher. 1-888-pHAX-ION, A+, All in One, Aplus, AQUAsnap, AssuredAccuracy, AUTO-BAR, AUTO-CAL, AUTO DISPENSER, Auto-ID, AUTO-LOG, AUTO-READ, AUTO-STIR, Auto-Test, BOD AutoEZ, Cable-Free, CERTI-CAL, CISA, DataCOLLECT, DataPLUS, digital LogR, DirectCal, DuraProbe, Environmental Product Authority, Extra Easy/Extra Value, FAST QC, GAP, GLPcal, GLPcheck, GLPdoc, ISEasy, KAP, LabConnect, LogR, Low Maintenance Triode, Minimum Stir Requirement, MSR, NISS, One-Touch, One-Touch Calibration, One-Touch Measurement, Optimum Results, Orion Star, Pentrode, phuture MMS, phuture Pentrode, phuture Quatrode, phuture Triode, Quatrode, QuiKcheK, rf link, ROSS Resolution, SAOB, SMART AVERAGING, Smart CheK, SMART STABILITY, Stacked, Star Navigator 21, Stat Face, The Enhanced Lab, ThermaSense, Triode, TRIUMpH, Unbreakable ph, Universal Access are trademarks of Thermo Fisher. 2008 Thermo Fisher Scientific Inc. All rights reserved. All trademarks are the property of Thermo Fisher Scientific Inc. and its subsidiaries. The specifications, descriptions, drawings, ordering information and part numbers within this document are subject to change without notice. This publication supersedes all previous publications on this subject.

Introduction General Information The ROSS process ph electrode, Cat. No. 2001SC, is a combination electrode designed for online applications. The ph sensing electrode and reference electrode are combined into a single electrode. This electrode is designed to be used in the Orion flow cell, Cat. No. 2001FC. This flow cell is designed to hold and position the electrode for optimal results. The 2001SC ROSS combination ph electrode is a researchgrade electrode designed specifically for use in online ph measurement applications. It is designed with a sidearm, which allows it to be attached to a reservoir of specially formulated ROSS reference filling solution. This ensures that the outer reference chamber is always full, allowing the ceramic liquid junction to flow freely. Figure 1 ROSS Process Combination ph Electrode Screw Cap Side Arm Reference Coil Reference Junction Sensing Element

Theory of Operation With its unique internal element system, the 2001SC ROSS process combination ph electrode provides far better stability, faster response and greater accuracy than can be obtained from any conventional electrode with silver chloride or calomel internal systems. The electrode response is fast; even in samples varying by 50 C or more in temperature. Drift is less than 0.02 ph units per day, which eliminates the need for frequent standardization. The ROSS internal system is designed to have virtually zero temperature coefficient, that is, the potential difference between the elements is zero regardless of the difference in the temperature of the elements. This is of great importance since the internal elements of combination ph electrodes are housed in an electrode body that, in use, is partially immersed in the sample and partially exposed to the ambient air. This means that the internal elements may be at different temperatures. As a consequence, conventional ph electrodes will drift and give inaccurate readings when being used to measure solutions with varying temperatures. On the other hand, the ROSS electrode gives stable readings and accurate results regardless of differences in sample and ambient temperatures. The silver chloride or calomel used in conventional ph electrodes tends to dissolve in the filling solution (usually KCl), eventually precipitating in the ceramic frit that forms the liquid junction. This, in turn, results in electrode failure. Differences in solubility of the silver chloride or calomel with temperature also gives rise to thermal hysteresis problems in conventional electrodes, so that re-standardization is usually necessary after temperature cycling. The ROSS system eliminates these problems as well.

Using the Electrode Electrode Preparation Caution: Wear gloves and protective clothing. ROSS reference filling solution can stain clothing and skin. 1. Unpack the electrode and reference filling solution tubing assembly from the shipping box. Carefully remove the protective shipping caps from the electrode sensing element and sidearm and save the caps for storage. 2. Clean any salt deposits from the electrode exterior by rinsing the electrode with distilled water. With a damp cloth, gently wipe around the sidearm area. Do not get water into the sidearm. 3. Take the tubing assembly and pass the smaller 1/8 inch tubing into the electrode sidearm, while sliding the larger 1/4 inch tubing over the sidearm. The outside tubing should extend 1/4 to 1/2 inch over the sidearm. Figure 2 Electrode with Attached Reference Filling Solution Reference Filling Solution Bottle Filling Solution Tubing Assembly Side Arm Sensing Element Reference Filling Solution

4. Remove cap and cap plug from the ROSS reference filling solution bottle (Cat. No. 2001FS). Hold the bottle in the upright position. Then connect the cap end of the tubing assembly to the bottle. The 1/8- inch smaller tubing should extend into the bottle. 5. Hold the filling solution bottle above the electrode in an inverted position until the electrode is completely filled with solution. Gently shake the electrode to allow any trapped air bubbles to rise into the bottle. For proper operation, the electrode should not have any air in it. 6. Dry off the ceramic frit reference junction by gently dabbing it with lint free tissue paper. To reduce errors due to polarization, do not rub or wipe the electrode sensing element. Squeeze the filling solution bottle for a few seconds. A small amount of filling solution should bead up on the surface of the reference junction, indicating good electrolyte flow. If no moisture is visible, the electrode may be clogged and should be cleaned.

Electrode Installation into the Flow Cell, Cat. No. 2001FC The flow cell housing needs to be cleaned and installed into an appropriate area with its mounting bracket. The reference filling solution bracket and clip should be assembled on the flow cell cap before installing the electrode. See the flow cell instruction sheet for details. Figure 3 ROSS Electrode with Electrode and O-ring Screw Cap Electrode Screw Side Arm O-Ring Reference Junction Sensing Element Screw Electrode O-Ring

1. Take the electrode holder and slide the holder screw onto it with the thread side down. The screw should be just below the notch on the holder. See Figure 3. 2. Take the electrode holder and screw and slide it all the way onto the electrode. The top of the holder should touch the bottom of the electrode cap, and the sidearm should fit into the notched area. See Figure 3. 3. Slide the large black O-ring onto the electrode up to the bottom of the holder. See Figure 3. 4. Insert the holder with electrode, into the flow cell cap. See Figure 4. Figure 4 Assembled Flow Cell Cap with Sensors Reference Filling Solution Reference Filling Solution Bracket Detachable Electrode Cable Filling Solution Tubing Assembly Electrode Screw Temperature Probe (ATC) ATC Flow Cell Cap Sensing Element ATC Probe

5. Align the sidearm 180 degrees away from the bracket holding the reference filling solution to insure correct positioning of the reference junction. 6. Hold the electrode in position while screwing the holder into the flow cell cap until finger tight. Do not overtighten. 7. Invert the filling solution bottle and insert it into the bottle holder. Use the supplied pushpin to puncture three air vents into the inverted bottom of the bottle. Venting the bottle will help to avoid noisy and drifting output signals. 8. Mark the level of the solution in the bottle and note the date. The solution level should be checked regularly and replaced as required. 9. For inserting an optional ATC probe, take the ATC holder and slide onto the ATC probe until it is under the ATC cap. Take the smaller black O-ring and slide onto the ATC probe until it fits snugly against the ATC holder. Place this ATC assembly into the flow cell cap and turn the screw on the holder until finger tight. 10. If no ATC probe is to be used, place the smaller black O-ring into the ATC probe well on the flow cell cap. Take the ATC holder and screw on until finger tight. 11. Take the flow cell cap assembly and gently place it straight into the flow cell housing. The assembled cap needs to be firmly placed straight into the cell. Do not tap the sensing electrode against the flow cell housing. 12. Twist the cap so that the lock screw is aligned with the notch on the cap and tighten the lock screw to position the cap properly on the flow cell housing. Flow Cell Cautions When removing the cap from the flow cell, always unscrew the lock screw first. Slowly pull the cap straight up from the housing. Do not hit the sensing element against the housing. When removing the electrode or ATC from the flow cell cap, unscrew the holder screw or ATC holder first. Do not pull the electrode or ATC out of the holder without first removing the O-rings. Be careful not to lose the O-rings. When placing the cap with assembled electrode into the flow cell, place it in straight. Do not hit the sensing element against the housing.

Figure 5 Assembled Flow Cell with Sensors Reference Filling Solution Reference Filling Solution Bracket Detachable Electrode Cable Filling Solution Tubing Assembly Electrode Screw Lock Screw (to Position Flow Cell Cap) Flow Cell Temperature Probe (ATC) ATC Flow Cell Cap Sample Outlet (3/8 NPT) Mounting Bracket Sample Inlet (1/8 NPT)

Attaching the Electrode to a Monitor Attach the 2001SC ROSS process ph electrode to the monitor or amplifier using the interface cable, Cat. No. 2001EC. 1. Attach the screw cap end of the cable to the electrode. 2. Attach the stripped, unterminated ends to the monitor. Refer to the monitor user guide for more detail. The stripped, unterminated ends are easily identified and can be inserted to any monitor s terminal strip. Figure 6 Electrode Connector (Stripped Unterminated Ends) Transparent Wire (ph) Black Wire (Ref) Assistance After troubleshooting all components of your measurement system, contact Technical Support. Within the United States call 1.800.225.1480 and outside the United States call 978.232.6000 or fax 978.232.6031. In Europe, the Middle East and Africa, contact your local authorized dealer. For the most current contact information, visit www.thermo.com/contactwater. Warranty For the most current warranty information, visit www.thermo.com/water.

For Best Measurement Results To maintain optimal electrode performance, use the following guidelines. Check electrode slope by performing two buffer calibration. Slope should be 92 to 102%. Always use fresh ph buffers when calibrating. Choose buffers that are no more than three ph units apart. For detailed calibration and temperature compensation procedures, consult the monitor user guide. Use only ROSS reference filling solution, Cat. No. 2001FS. Use of other filling solutions may result in electrode failure. Inspect the electrode for scratches, cracks, salts or other deposits on the sensing element or reference junction. Rinse off any precipitate by rinsing the electrode with distilled water. To reduce errors due to polarization, do not rub or wipe the electrode sensing element. Best results will be achieved if the electrode is used with the Orion flow cell, Cat. No. 2001FC. The first time a new electrode is installed, allow the electrode to equilibrate for 1 hour up to 24 hours. Change the reference filling solution tubing assembly once every three to six months. Make sure that ROSS reference filling solution bottle is properly vented. Periodically check that there is sufficient reference filling solution and replace bottle as needed. Keep electrode and electrode holder areas clean. 10

Ordering Information Cat. No. 2001SC 2001FC 2001EC 20015M 2001XM 2001CP 2001FS Description ROSS process combination ph electrode with screw cap and sidearm Flow cell for ROSS process ph electrode, includes stainless steel housing with mounting bracket, PVC holders and cap, and solution mounting bracket Detachable 1 meter cable for ROSS process ph electrode, coaxial cable with screw cap attachment for electrode and tinned, stripped, unterminated ends for attachment to monitor Detachable 5 meter cable for ROSS process ph electrode Detachable 10 meter cable for ROSS process ph electrode 6 month consumables pack for ROSS process ph electrode, includes 5 x 2 oz (60 ml) bottles of ROSS filling solution and 1 pint (475 ml) bottle each of ph 4.01, 7.00 and 10.01 buffers ROSS reference filling solution for ROSS process ph electrode, 5 x 2 oz (60 ml) bottles 910104 ph 4.01 buffer, 1 pint (475 ml) bottle 910107 ph 7.00 buffer, 1 pint (475 ml) bottle 910110 ph 10.01 buffer, 1 pint (475 ml) bottle 2001TP 2001XT 2001LN 2001TM 2001EK Process ATC temperature probe (30 kohm) with 1 meter cable Process ATC temperature probe (30 kohm) with 10 meter cable Process ATC temperature probe (8.5 kohm) with 5 meter cable Process ATC temperature probe (1 kohm) with 5 meter cable ROSS process ph electrode kit, includes ROSS electrode (2001SC), flow cell (2001FC), ATC probe (2001XT) and cable (20015M) 11

Specifications ph Range 0 to 14 ph Temperature Range 0 to 100 C Drift Less than 0.5 ph units for 30 days Slope 92 to 102% (54.4 to 60.3 mv/dec) Isopotential Point ph 7 Junction Ceramic frit Internal Reference ROSS Size Electrode Diameter: 12 mm Electrode Length: 125 mm Electrode Cap Diameter: 16 mm Electrode Cap Length: 30 mm Detachable Cable Coaxial Cable Length 1 meter Filling Solution ROSS reference filling solution * Specifications are subject to change without notice 12