METASYS. Amalgam Separator COMPACT A8 / A16. Installation, operation and maintenance

Similar documents
ECO II. Economy System Typ 2. Amalgamabscheider ECO II/ECO II Tandem ECO II D/ECO II Tandem D Gerätedokument Einbau, Betrieb und Wartung

ECO II. Economy System Typ 2. Amalgamabscheider ECO II/ECO II Tandem ECO II D/ECO II Tandem D Gerätedokument Einbau, Betrieb und Wartung

INSTALLATION PROOF REGISTER YOUR PRODUCT. for warranty claims STAMP. Dealer. Date. here. Please complete and return. For the dental office:

ECO MST 1. Separierautomatik MULTI SYSTEM TYP 1 ECO Gerätedokument Einbau, Betrieb und Wartung

METASYS. Central Suction System EXCOM hybrid 2/5. Equipment Logbook Assembly, operation and maintenance

REGISTER YOUR PRODUCT. for warranty claims INSTALLATION PROOF. Please complete and return

METASYS. Amalgam separator ECONOMY SYSTEM TYPE 2 ECO II. Installation, operation and maintenance

CENTRAL SUCTION SYSTEMS. Dry / semi-wet suction For 1-15 dental units Underpressure of 180 mbar

Silent TS. Nr / Ideas for dental technology / A. Made in Germany

APPENDIX 4. Thank you for choosing the Biomega BioClave steam sterilizer. HYDRAULIC DRAWING

MODEL A5-2 SOLVENT & WATER RECOVERY SYSTEMS (EXPLOSION PROOF UNITS)

GB Installation instructions DÜRR Combi-Separator 7110

Instruction Manual. Alarm Unit For Low Gas Level # Read manual before use! Observe all safety information! Keep manual for future use!

Bedienungsanleitung. Ideas for dental technology. Made in Germany C

Operating and Maintenance Instructions

SILENT compact /

MODEL A18 SOLVENT RECOVERY SYSTEMS (EXPLOSION PROOF UNITS)

Electronically controlled instantaneous water heater. MCX: 27300, and models. Installation instructions

Operator Guide. Publication No. OI /97-Eng. KODAK Automixer II Plus

NEPTUN I Sole cleaning unit Art.-No.:

Operating Manual Bathroom Balance

Cleaning unit for coolant. :_decftez`_>r_fr] Book No.: V2

TECHNICAL INFORMATION Touchtronic Clothes Dryers

INSTALLATION INSTRUCTIONS DÜRR MOTORSEPA S 300

Flameproof Manual Call Points

Automatic Mixer Operating Instructions

1.1 Setup Installation

Service Information. WNes 2956 appliance documentation. Service Information no. 27/2004 LHG/TKD-Fe/June SI

Moisture inside the drum is due to final testing.

SMART EVO 2 - User Manual ELECTRICAL PANEL FOR 2 MOTORS

Super Elite Watch cleaning machine

DÜRR SUCTION SYSTEMS. High demands reliably exceeded DÜRR AMALGAM SEPARATORS

Millo / Millo pro. Nr x000 / 1805-x000. Ideen für die Dentaltechnik A

BWT ECO. Fitting and operating instructions. Ultra Violet Water Disinfection System. ECO Range: LC 8 UV LC19 UV LC36 UV LC51 UV

Operating manual BDC-654 BDC-654. EN - Version 1.1. Gebruikershandleiding. Operating manual.

Refrigerator BRFB1920SS BRFB1900FBI BRFB1920FBI

Operating Manual for Profident / Profident PLUS

MW001 Integrated Dishwasher. Manual for Installation, Use and Maintenance

Moisture inside the drum is due to final testing.

smardy - pure water Reverse osmosis system YR100-A 75 GPD

Removing shipping locks

TECHNICAL INFORMATION T 15xx Dryers

Operating Instructions. Keep for future reference! Maschinenbau Werkzeuge Laborgeräte. Machine no.:... Hausanschrift:

Operating Instructions for Reka Hot-Melt Glue Applicators

Hand Dryer Installation & User Manual

Table of Contents. English

Installation Instructions

Silent TS. Nr / Content Introduction...1 Symbology...1 Information for Operators...2. Introduction.

ST1. Mod. USE AND MAINTENANCE MANUAL

USER GUIDE. DRENA 2 - User Manual ELECTRICAL PANEL FOR 2 MOTORS - WASTE WATER -

Installation and Operating Instructions DÜRR Regeneration Unit for X-ray developers XR 24, XR24 II, XR 24 Nova, XR 24 Pro

Dyna-Form Air. Alternating Pressure Relief Overlay System

Operating Manual. Reverse Osmosis Water Purifier. MELAdem 55

Servicing manual. 600 Series - 11S / 19S / 24S / 24C. Wall-mounted condensing gas boiler. For trade use

AirStar Clinical Air Compressor AS60, AS90 and AS120. Installation & Operating Instructions

Servicing manual. Wall-mounted condensing gas boiler 600 Series - 11S / 19S / 24S / 24C /2002 GB(EN) For trade use

Instructions for use. 301 plus

Instructions For Use ENG. LED Magnifier Lamp

Operational Instructions. Glue Applicator MS Translation from German to English

A few simple steps to ensure perfect water, every time.

WHA 3000P Hot Air Station

Engineers manual. Under counter machines

HYDROCYCLONE ISO 18 INDEX. Page

OPERATING INSTRUCTIONS

Dental Unit User's Guide

Electronically controlled instantaneous water heater. MCX: 27300, and models. Instructions for the user

User, Installation and Servicing Instructions. Silverlink 600 Gas Griddles GS4 and GS7 IS212 ECN3592

BRITA. Water Filter & Chiller II. Chilled water filter. Instruction Booklet WF6100

1. Take the hoses and elbow out of the holders, depending on model. max. 100 cm. min. * cm

Inspection Light LED 6149/2. Operating instructions EN EN EN EN EN EN EN EN EN EN EN EN EN EN EN EN EN EN EN EN EN EN EN EN

INSTALLATION MANUAL GUTHD2. Universal Two Way Digital Thermostatic Valve for Shower Systems

User s Guide ENGLISH. Other languages available on our website:

Instruction Manual. Electronic gas and smoke detector GRM

Dry Bath Heating Systems

Amalgam separator CA 4

CE Unit is suitable for material transport. The transported gas may be up to 80 C.

AS Medizintechnik GmbH Sattlerstrasse 15, Tuttlingen, Germany Tel 07461/ Fax 07461/

B Saneo. Saneo. Saneo BS Saneo CS Saneo BW Saneo CW. C Saneo BLP Saneo CLP. Dometic Dometic. Dometic Dometic. Dometic Dometic.

Service: to SLK Laboratory Hot Plates glass made of ideas

POLYMIX PX-IG 2000 Operating Instructions

Manual for installation, use and maintenance MW200. Integrated Slimline Dishwasher. year. parts + 1 year labour guarantee.

Hydraulic crimping tool. HO-16 Art QZERTIFIKAT DIN EN ISO convincing solutions

The Professional s Choice

SECTION TRANSFER SYSTEM STS. for ROTARY MICROTOMES INSTRUCTION MANUAL. Section Transfer System

Installation & Operators Manual

Illustration shows item no Z Original operating manual. Operating manual. BlueMobil 60. Item no: Z-3285

Product Manual Glue Applicators MS 80 and MS 200

L / -014L / -015L

SPA DE USO PRIVADO SPA FOR PRIVATE USE SPA POUR UN USAGE PRIVÉ SPA FÜR PRIVATE NUTZUNG SPA PER USO PRIVATO SPA PARA USO PRIVADO SPA VOOR PRIVE-GEBRUIK

Instruction Manual for FORMALDEHYDE GAS DETECTOR MODEL FP-30/FP-40

SOLDERING GUN PLP 100 A1

Twister / Twister venturi

Operating Instructions

Cooker Hood Instruction Manual

SERIES 4000 AUTO GATE MANUAL

MW402 Integrated Dishwasher

Original operating instructions Safety switch with guard locking AC901S AC902S

SAFETY AND OPERATING MANUAL

Vacuum Cleaner User's manual

MODEL 7000 SUCTION UNIT

Transcription:

METASYS Amalgam Separator COMPACT A8 / A16 Installation, operation and maintenance

List of contents Explanation of the pictograms The groups of people for whom certain types of activities are relevant can be found in the footer. 1. Table of contents Page 1. Table of contents 2 2. Explanation of the pictograms 2 3. General Instructions 3 4. Use 4 5. Construction 4 6. Explanation of the type plate 4 7. Technical data 5 8. Functional description 5 9. Hose connections 6 10. Electrical connections 6 11. Explanation of the operating components 7 12. Exchange of collection container 8 13. Disposal of the full collection container 9 14. Care, cleaning and disinfection 10 15. Service mode 11 16. The annual inspection 12 17. The 5-year inspection 13 18. Connection COMPACT A8 / A16 to EXCOM Z2 / Z5 14 19. Disposal 14 2. Explanation of the pictograms means that if the instructions are disregarded danger for people, operational problems or physical damage to the equipment or its immediate surroundings can occur. means that the attention of operating or maintenance personnel is particularly drawn to important items in this section. Seite / Page

General Instructions 3. General Instructions: The safety, reliability and performance of the equipment can only guaranteed by METASYS if the following instructions are followed: q Installation, alterations or repairs must only be carried out by authorised specialists who guarantee compliance with standard IEC 601-1 (International Standard for Medical Electrical Equipment, particularly Part 1: General Safety Requirements). q The electrical installation must correspond with IEC (International Electrotechnical Commission) requirements. q The equipment must only be used in accordance with the installation, operation and maintenance instructions. q Only genuine METASYS spare parts must be used for repairs or replacement. After commissioniong of the amalgam separator, the installation proof found in the Equipment Logbook must be completed and returned to METASYS in order to determine the guarantee period. q Every service and inspection must be recorded in the Equipment Logbook. q The disposal certificates for the amalgam waste must be retained in compliance with local legislation. q If requested by a qualified technician, METASYS declares to provide all prepared documentation likely to be of use to the technically qualified person for service and repairs of the equipment's components. q METASYS takes no responsibility for damage arising from external influences (inaccurate installation), incorrect information, use of the equipment for purposes other than the one intended, or for improper repairs. q If the whole amalgam separator should be removed, it is to be returned to the manufacturer for proper disposal. Seite / Page

Use, construction, explanation of the type plate 4. Use 1 COMPACT A8 / 2 COMPACT A16 The METASYS COMPACT A8 / A16 is a two-stage amalgam separator for the installation after a central suction system with air/water separation. 5. Construction 1 The COMPACT A8 / A16 amalgam separator is composed of 3 modules: Module 1 is the central fixing unit. Electrical and waste water connections are made here. Module 1 contains the wall holder and the control electronics. Module 2 is the centrifuge (the dynamic stage) and the sedimentation stage of the amalgam separation. Module 2 contains the filter housing with the filter drawer, the centrifuge and the collection container. Module 3 is the wall mounting bracket with transformer, external display and main switch. 2 6. Explanation of the type plate 3 See illustration 3.4 COMPACT- A 8 P max = 100 VA V max = 8 l/min U N = 24 V~ 50/60 Hz 3.1 3.2 The type plate is to be found on Module 1 and is visible when Module 2 is removed (to do this, lift the yellow locking bracket out of its fixing towards you. A8-00000 Z-64.1-19 3.3 3.1 Equipment designation 3.2 Connection data 3.6 3.5 METASYS Medizintechnik GmbH A-6063 Rum bei Innsbruck 3.8 3.7 3.3 Approval Number from the German Institute for Construction Technology 3.4 Serial number 3.5 Manufacturer s address 3.6 CE mark 3.7 UZVC conformity mark 3.8 Protection Class 3 Seite / Page 4

Technical data Functional description 7. Technical data COMPACT A8 COMPACT A16 350 mm Supply voltage: 230 V AC 230 V AC Frequency: 50/60 Hz 50/60 Hz Maximum power consumption: 4 A 2 x 4 A Equipment fuse: 6,3 A T 2 x 6,3 A T max. power loading: 100 VA 2 x 100 VA Separation rate: 95% 95% Collection container: 300 cm 3 2 x 300 cm 3 Maximum ambient temperature: 40 C 40 C 390 mm Maximum water flow volume: 8 l / min 16 l / min (via the air/water separator) Dimensions 4 in mm (H x W x D): 390 x 350 x 170 390 x 350 x 190 Possible suction systems: dry vacuum engine with air/water separation, e.g. METASYS EXCOM central suction system 170 mm 4 8. Functional description 5.1 5 See illustration 5.5 The waste water from the air/water separation is inserted through the water inlet directly into the centrifuge via the filter with the filter drawer and the entry funnel 5.1. The motor 5.2, which drives both the centrifuge 5.3 and the pump 5.4 is started by a sensor contact 5.5 from the incoming water. The coarse particles are retained on the walls of the two interleaving centrifuge chambers. 5.3 After purification, the water continuously runs into the outflow 5.6. As soon as no more water enters the air/water separator via the filter, the motor is switched off and the centrifuge stops abruptly after a brief follow-up time. 5.6 5.2 The water column, which continues to rotate, has a self-cleaning effect on the centrifuge, which washes the coarse particles into the collection container 5.7. 5.4 As soon as the liquid level in the collection container reaches the electrical area of the capacitve sensor, the motor starts once again, the water is pumped to the filter housing and the cycle begins once more. 5 5.7 Seite / Page

Hose and electrical connections 9. Hose Connections 1.1 1 See illustration 1.1 Water inlet connection 1.2 Water outlet connection 1.2 10. Electrical connections Power supply conection: 1 Before connecting with the mains, the nominal voltage on the type plate on the equipment must be compared with the mains voltage. In order to allow self-diagnosis, the COMPACT A8 / A16 must be switched off at least once per working day. 2 See illustration 2.1 D14 LED network (illuminated when current present) 2.13 2.2 J1 connected: sensor response normal removed: sensor response sensitive 2.3 S1 Fuse T 6.3 A (only replace with the same type) 2.4 X3 24 V AC supply 2.5 X2 Motor connection (internal) 2.2 2.6 X6 No function for COMPACT A8 2.7 X8 No function for COMPACT A8 2.1 2.8 X7 No function for COMPACT A8 2.3 2.9 Connection for fill height measurement X7 D14 2.10 Connection for the capacitve sensor 2.4 2.5 24 V X3 MOT X2 X8 X4 X5 X6 LIS X1 TS 2.11 X4 Connection for external display COMPACT A8 2.12 X1 Only for COMPACT A8 Connection for inlet sensor 2.13 Connection for processor 2.7 2.8 2.11 2.10 2.12 2.6 2.9 2 Seite / Page Practice Technicians staff / technicians 6

Operating components 11. Explanation of operating components 3 See illustration 3.1 3.2 3.3 3.1 Signal 1: ready for operation q illuminated green: mains current present 3.2 Signal 2: Centrifuge malfunction q flashing red, buzzer sounding: malfunction! See illustration 4 - Malfunction displays 3.3 Signal 3: Collection container level display q illuminated yellow, buzzer can be switched off by pressing the button: the collection container is 95% full. It is recommended to change the collection container now. However, it is possible to continue working until the indicator shows the collection container is 100% full. The indicator light continues to be illuminated as a reminder. The buzzer sounds again each time the main switch is switched on. q Illuminated yellow, buzzer cannot be switched off by pressing the button: the collection container is 100% full. The collection container must be changed. Further operation is not possible. 3.4 Alarm button q The buzzer can be turned off by pressing on the red surface if the collection container is 95% full or in the case of an indicated malfunction. 3 = leuchten = blinken = permanent = flashing = allumé = intermittent = rimane accesa = lampeggiare = encendido = intermitente 3.4 4 Malfunction displays (see illustration) 4.1 Signal 1 3.1 1 illuminated green, signal 2 3.2 flashing red, buzzer sounding: electronic control malfunction! The buzzer is turned off by pressing the button. In this case turn the device off with the main switch and turn it on again after a short pause (approx. 5 seconds). If the indicator lights are illuminated again after a short time, please inform your service technician. 4.2 Signal 2 3.2 flashes red, buzzer sounds: The collection container is not positioned correctly! The buzzer cannot be turned off. No other signal is illuminated. Please follow the instructions in Point 12 in order to position the collection container correctly. 4.1 4.2 4 Seite / Page

Replacing the collection container 12. Replacing the collection container 1.1 q q Switch off the COMPACT A8 / A16 at the main switch q Put on protective gloves Get a new collection container ready and remove the disinfection pouch from inside 1 Disconnect the water connection 2 Pull the amalgam separator out so that it is com- 1.2 1.1, holding the amalgam separator whilst doing so. 1.2 Lift the locking bracket on the amalgam separator and pull the amalgam separator out of its fixing towards you. 2 1 pletely free from its fixing and place it on a level, non-slip surface. 3 Hold the collection container firmly and turn the upper part anti-clockwise. Lift the upper part off. 2 4 If the pump filter is soiled, remove the filter and clean it over an amalgam collection container. Replace the filter on the pump inlet housing. 1 5 3 Place the upper part onto a new collection container Hold the collection container firmly and turn the upper part clock-wise until its stop. Visually check that all stops on the collection container are properly positionioned. 4 6 1 7 Grease sealing surfaces' holder with Vaseline. Push the amalgam separator carefully back into the 5 wall mounting bracket 8.1 and close the locking bracket 8.2. Re-connect the water inlets removed in Point 1. 2 q VASELINE If the collection container has been properly inserted the amalgam separator starts and runs for a short period; the external display signals ready for use (signal 1 is illuminated green). 6 5 Switch on the equipment s main switch. If the container has been inserted incorrectly, signal 2 on the external display flashes and a beep can be heard. If this occurs, switch off the main switch and carefully repeat the procedure described above in Points 6-8. 8.1 8.2 7 Seite / Page

Disposal of the full collection container by Dental ECO Service 13. Disposal of the full collection container Wear protective gloves and possibly a face mask. Avoid contact with collection container's contents. For technical and hygiene reasons, the collection container is designed to be used only once! Re-use of the collection container can lead to operational problems and is in breach of the guarantee conditions. The full collection container can be dispatched to our own disposal company, Dental ECO Service GmbH METASYS Group. Certificates of recycling must be retained in accordance with local legislation. If the amalgam separator is to be removed at the end of its working life, it should be returned to the manufacturer for proper disposal. The simplest way disposal by ECOTransform. 9 See illustration To disinfect the container before sending it off, open the disinfection sachet (enclosed with the new collection container) and pour its contents into the full collection container. 9 10 10 See illustration Close the full collection container using the green lid (enclosed with the new collection container) by turning it clock-wise. Check that it is properly sealed by turning it upside down. Re-fasten the lid if necessary. 11 11 See illustration Place the properly closed collection container into the two halves of the Styrofoam shells of the postal packing. Close the package as in the attached instructions. 12 See illustration Prepare the package for despatch by sticking the enclosed label onto it and return to Dental ECO Service GmbH, or its national collection point. 12 13 See illustration Dispose off the amalgam residues from the filters in the suitable container, e.g. ECOCenter. Do not use the suction system to remove the residues from the filters. 13 Seite / Page 9

Care, cleaning and disinfection with Green&Clean M2 14. Care, cleaning and disinfection 1 See illustration Shortly flush spittoon bowl after each treatment. 2 See illustration Suck some water through all suction hoses after every treatment. 1 2 3 See illustration After having sucked some water, suck Green&Clean M2, the disinfectant suitable for the amalgam separator, twice a day. Ideally, disinfectant should also be used after the equipment has remained unused for a while (e.g. lunch break, holidays). 4 See illustration Also flush the spittoon bowl twice per day using Green&Clean M2, suitable for the amalgam separator. 3 4 Cleaning the filter 5 See illustration Empty and clean the filter drawer 5.1 at least once a week. 5.2 Operation of the amalgam separator without the filter is not permitted. 5 5.1 6 See illustration Collect the residues containing amalgam from the primary filter in the suitable container EcoCenter and dispose of them properly with EcoTransform. If the amalgam separator is used for lengthy periods and does not switch itself off automatically, even though there is no water flow, the sensors are probably short-circuited due to dirt. 6 10 Seite / Page

Service mode 15. Service mode The COMPACT A8 / A16 amalgam separator electronics allow the equipment s functions to be checked using a service mode. IMPORTANT! In order to enter normal operation or service mode after the equipment has been switched off, wait at least 5 seconds before switching it on again! 8 Entering the service mode 8.1 When switching the main switch on, depress the button on the operating component for at least 5 seconds. Signal 1 (ready for operation) flashes at twice its normal frequency (25 Hz) 8.2 As soon as signal 1 flashes, release the button Signal 1 flashes very rapidly (25 Hz) q Press the button briefly to start the motor q The motor is stopped by pressing once again In case of a malfunction of the motor monitoring system (shortcircuit monitoring and monitoring of the running of the motor) the electronic systems will not allow the motor to start. q The motor can be switched on and off at will by briefly pressing the button q Pressing the button for approximately 5 seconds will change the service mode from motor monitoring to testing the inlets and outlets. (Release the button after hearing a beep) 8 8.1 8.2 PUSH 5sec = leuchten = blinken = permanent = flashing = allumé = intermittent = rimane accesa = lampeggiare = encendido = intermitente 9 10 q Signal 1 flashes; signals 2 and 3 are steadily illuminated Procedure for testing the inlets and outlets see Point 16 of Annual Inspection. q Press the button once again for approximately 4 seconds to end Service Mode Signal 3 flashes (25 Hz) The equipment switches from Service Mode to normal operation Signal 1 is permanently on const. 11 12 const. 9-14 Please contact your service engineer if the following malfunction messages appear: 9 Malfunction in the short-circuit monitoring 10 Malfunction in the monitoring of the running of the motor 11 Motor malfunction on starting 12 Current consumption of the motor when running is too high 13 Power supply to motor interrupted 14 Short-circuit of motor power supply const. 13 14 2,5Hz Seite / Page 11

Annual inspection 16. The Annual inspection According to the German Institute for Building Technology, the functioning of displays of amalgam separators should be checked at least once a year. q Enter the service mode, as described in Point 15 q Start the motor at least once q Switch to the checks for the inlet and outlet Signal 1 flashes; signals 2 and 3 are continuously illuminated. q Remove the separator from the wall fixing 1 1.2 1.1 1.3 2.1 Display simulation 1 See illustration 1.1 Cover only the lower light barrier of the liquid level indicator. Signals 1 and 3 flash. Signal 2 is permanently illuminated 1.2 Cover only the upper light-barrier of the liquid level indicator. Signals 1 and 2 flash, signal 3 remains permanently lit 1.3 Cover both the light-barriers of the liquid level indicator. Signals 1, 2 and 3 flash q Place the separator back into the wall mounting fixing. 2 See illustration. 2.1 Withdraw the inlet sensor and short circuit it with a wet cloth The buzzer sounds q Replace the inlet sensor q Press the button once again for approximately 4 seconds to end Service Mode. Signal 3 flashes Signal 3 flashes (25 Hz) The equipment changes from Service Mode to normal operation Signal 1 is permanently illuminated As a preventive maintenance measure we recommend the replacement of the following parts by an authorised service engineer during the annual inspection: all seals, pump filter, filter drawer. The annual inspection must be recorded in the Equipment Logbook! 2 12 Seite / Page Practice Technicians staff / technicians

5-year inspection 17. The 5-year inspection: Following the Waste Water Regulations, Appendix 50, amalgam separators must be checked in accordance with local legislation to ensure their good condition in intervals not exceeding 5 years. 3.3 3.1 q Carry out the annual inspection as described in Point 16. q Check that the installation and connection of the amalgam separator is in good condition and meet the installation guidelines. q Flush the suction hoses and the spittoon bowl using at least 1 litre of clean water, and with Green&Clean M2. 3.5 3.2 3 See illustration Wear protective gloves. Ensure that the main switch is OFF. q Remove the separator from the wall fixing (as described in Point 12, Sections 1 4) 3.1 Remove the inlet sensor 3.2 Carefully remove the electrical contact for the inlet sensor using a suitable tool 3.3 Disconnect the pump hose 3.4 Remove the three screws on the upper side of the equipment using a suitable tool (Philips screw-driver) 3.5 Remove the filter housing 3.6 Remove the four screws on the upper part of the centrifuge using a suitable tool (Torx 20) 3.7 Remove the upper part of the centrifuge 3.8 Remove the sealing sleeve from the Z-clamps using a screwdriver 3.9 Remove the Philips head screw with the spring collar on the upper side of the Z-clamp with a suitable tool (Imbus). IMPORTANT: it has a left hand thread! 3.10 Remove the Z-clamp and the centrifuge chambers by pulling them upwards Check the centrifuge chamber optically for dirt. Centrifuge chambers containing thick linings residues must be renewed. q Re-assemble the equipment by following the above instructions, in the reverse order. 3.4 3.6 3.9 3.4 3.6 3.7 3.8 3.10 3.11 IMPORTANT! When re-assembling the amalgam separator make sure that the 4 lugs 3.11 on the centrifuge chamber's bottom are positioned in their intended holes. Care must also be taken that all screws are tightened up to their heads. The 5-year inspection is to be entered in the Equipment Logbook! 3 Practice Technicians staff / technicians Seite / Page 13

Connection COMPACT A8 / A16 to EXCOM Z2 / Z5 Disposal 18. Connection COMPACT A8 / A16 to EXCOM Z2 / Z5: 1 Hose connections min. 2% max. 0,4 m 2 Electrical connections (connection operator's control/ display 2.1 Mainboard COMPACT A8/2 2.2 Mainboard COMPACT A8/1 2.3 Operating part A8/1 1 2.4 Operating part A8/2 2.1 2.2 2.3 2.5 Intermediate circuit board of COMPACT A8/A16 for connections 2.6 Mainboard EXCOM, cable 3-pole 2.7 Display EXCOM, cable 6-pole 2.8 External display, cable 6-pole 2.5 2.4 19. Disposal Parts of the COMPACT A8 / A16 can be contaminated and amalgam can adhere to it. Therefore, these parts need to be disposed of in accordance with national requirements. 2 2.6 2.7 2.8 The electronic components of the COMPACT A8 / A16 are classified as electronic waste and need to be disposed of accordingly. 14 Seite / Page

Notes Practice staff Notes / technicians Seite / Page 15

METASYS Austria METASYS Medizintechnik GmbH Florianistraße 3 A-6063 Rum bei Innsbruck 1++43/(0)512 /20 54 20-0 5 ++43/(0)512/20 54 20-7 Germany METASYS Medizintechnik GmbH Ahornstraße 19 D-85614 Kirchseeon 1++49/(0)89 / 613874-0 5 ++49/(0)89 / 6135829 France METASYS France S.a.r.l. 9, bd E. Michelet F-69008 Lyon 1++33/(0)4/37 90 22 15 5 ++33/(0)4/37 90 22 47 e-mail: info@metasys.fr www.metasys.fr email: info@metasys.com www.metasys.com Italy Dental Eco Service Italia S.r.l. Florianistraße 3 A-6063 Rum bei Innsbruck 1 +39/045/981 4477 5 +39/045/981 4475 e-mail: desitalia@metasys.com Ihr METASYS Berater:/Your METASYS agent: EBW Compact A8 / A16, 21.07.2008 ; 70100310 ZK-55.212/02 Subject to technical changes, printing and setting errors! TK 2005-6