Instruction Manual. gray Mounted Detectors. for. greenteg AG Technoparkstrasse Zürich, Switzerland T: F:

Similar documents
Instruction Manual. gray Housed Detectors. for. greenteg AG Technoparkstrasse Zürich, Switzerland T: F:

gray Laser Power Detectors For fast and accurate measurements

Key Features. General Description

APPLICATION NOTE. 1. Introduction Application Considerations

Technical Information OUSAF21

PM-200 Power Module Operating Instructions

Technical Information OUSAF44

Table of Contents. Page Quick Set-up Instructions 1. Theory of Operation 2. Specifications 3. TM 1003W Controls and Connections 4

Dated 15-Dec-2000; J. Schilz, subject to change

Temperature dependent Components within the Pyroelectric Detector

Understanding total measurement uncertainty in power meters and detectors

Flame Monitor for Multiple Burner Furnaces D-LE 603 D-UG 660

COMPACTNESS 34.1 PERFORMANCE ANTICIPATION

Model OEM-2 INSTRUCTIONS FOR USE

Technical Information OUSAF22

MODEL-BASED OPTIMIZATION OF AN INFRARED GAS SENSOR

Chapter 1. Introduction

Laser Distance Sensor Type M Analog-Output

154BC SPECIFICATIONS FEATURES APPLICATIONS STANDARD RANGES

PUV, PIR and PFO Series BROCHURE

154BC SPECIFICATIONS. Please refer to the 154N uncompensated and constant voltage datasheets for more information on different features of the 154.

Broadening IR applications through using spectral filters

Standard SPECIFICATIONS FEATURES APPLICATIONS

SIMRAD GD10 IR GAS DETECTOR

2 Ensure that all personnel in the area are wearing the appropriate protective eyewear.

COMPACTNESS 34.1 PERFORMANCE ANTICIPATION

An FT-NIR Primer. NR800-A-006

Class 1 laser beam sensor safe for your eyes

Miniature Thermoelectric Cooler Controller MTCC-01 user's guide

TSI AEROTRAK PORTABLE PARTICLE COUNTER MODEL 9110

Spectrum Detector, Inc.

818 Series. Low-Power Detectors ABSOLUTE SPECTRAL RESPONSIVITY (A/W)

Class 1 laser beam sensor safe for your eyes

DESIGN MANUAL MODEL GIR900 INFRARED COMBUSTIBLE GAS SENSOR

86 Compensated SPECIFICATIONS. 316L SS Pressure Sensor Small Profile 0-100mV Output Absolute and Gage Temperature Compensated

Sub psi SPECIFICATIONS

Digital Heat Block User Manual

760nm Single mode VCSEL with Peltier Element and Thermistor in TO510

Pyroelectric Infrared Sensor PIR

DESIGN MANUAL. The information in this doucment is applicable to sensors built in September, 2009 or later (Firmware 3.0)

Class 1 laser beam sensor safe for your eyes

Model 5190-E Temperature/Humidity Probe

SIMRAD GD10P - IR GAS DETECTOR OPERATING MANUAL

IR detection module VPAC-1000F user's guide

Low-Power Detectors. Key Features. 818 Series. Power levels from pw to 2W. Wavelengths from 190 nm to 1800 nm. Proprietary detector optics

SPECTRAL INSTRUMENTATION

86 Compensated SPECIFICATIONS. 316L SS Pressure Sensor Small Profile 0-100mV Output Absolute and Gage Temperature Compensated

THE NEXT GENERATION IN VISIBILITY SENSORS OUTPERFORM BOTH TRADITIONAL TRANSMISSOMETERS AND FORWARD SCATTER SENSORS

Models S310 & S310D Laser Power and Energy Meters Setup and Operating Procedures

FABRICATION SOLAR REFRIGERATION SYSTEM BY PELTIER EFFECT

85 Compensated SPECIFICATIONS. 316L SS Pressure Sensor Small Profile 0-100mV Output Absolute and Gage Temperature Compensated

SEC SIGNATURE ETO MONITOR TRAINING SENSOR ELECTRONICS CORPORATION

HM1500LF Relative Humidity Module

SIMRAD GD10PE IR GAS DETECTOR

Erbium-Doped Fiber Amplifier (EDFA) - MW Series

MST, INC. MODEL 2002 OXYGEN MONITOR (AMBIENT-TYPE) OWNER S MANUAL

86 Compensated SPECIFICATIONS. 316L SS Pressure Sensor Small Profile 0-100mV Output Absolute and Gage Temperature Compensated

Utilizing Finite Element Analysis (FEA) for Flexible Heater Designs

OPERATION MANUAL MODEL 3038 ACCELEROMETER

818 Series Low-Power Detectors

to get there. together. Infrared Sources Swiss made

Laser Analyzers for NH3 Safety Monitoring and Leak Detection

SAFEVIEW-MINI2 TRANSILLUMINATOR

LEARNING SPECIAL HAZARD DETECTION TYPES

Appearance and sensing characteristic Specification (temperature range) Model 10 to 70 C ES1B. 2 mm 20 mm 40 mm 60 mm. 40 dia. 60 dia.

Identify the key elements of the laser source, regardless of the type of measurement or sensor required;

Fiber Optics and Liquid Level Sensors Line Guide

MS4525 SPECIFICATIONS

Technical Information OUSAF12

NTC THERMISTOR SENSOR PERFORMANCE

The J22 and J23 series are high performance InGaAs detectors operating over the spectral range

1040 Gas Monitor INSTALLATION AND OPERATING INSTRUCTIONS AMC-1040 WITH INTEGRAL ELECTROCHEMICAL SENSOR

Model 5190-F Temperature/Humidity Probe

HTF3000LF PVH-3.3 TEMPERATURE AND HUMIDITY MODULE

Measuring THz Radiation choose a Pyroelectric Detector or Golay Cell?

A Cost Effective Multi-Spectral Scanner for Natural Gas Detection

Laser Safety Goggle. UV Visible

WHITE PAPER FIBER OPTIC SENSING. Summary. Index. Introduction. About Fischer Connectors

Development of a variable output LED-365 irradiance standard

DSSLF2X04H 2 IN 1. Intelligent Lead Free Soldering Station. OPERATION and MAINTENANCE MANUAL

Omnisens DITEST TM FIBER OPTIC DISTRIBUTED TEMPERATURE & STRAIN SENSING TECHNIQUE

HM1520LF Relative Humidity Module

S304 and S305 Dust Emission Monitors. User Manual. Distributor

Date: Wednesday, January 31, 2018 Time: 8:30 am 5:00 pm Location: Moscone West (Room 2018, Level 2)

POWER METER INSTRUCTION MANUAL

Sintrol Snifter A1+ Manual. Version 1.2.2

Adelaide Homes Design Guide 4 - Winter warming

Issue C. Installation Instructions for the Basic Amplified Board Mount Pressure Sensors, ABP Series

TOCOS. Guidelines & Precautions. For Using Trimmer Potentiometers. Guidelines for Circuit Design. Guidelines for Production Processes

optris TM Temperature Monitor for Infrared Sensors Operators manual

MST, INC. MODEL 2002 CARBON MONOXIDE MONITOR (AMBIENT-TYPE) OWNER S MANUAL

Chemistry Instrumental Analysis Lecture 14. Chem 4631

MP 601A Humidity Probe INSTRUCTION MANUAL

Technical Information OUSBT66

Force Sensors Line Guide

Miniature Thermoelectric Cooler Controller MTCC-01 user's guide

mach 5 100kHz Digital Joulemeter: M5-SJ, M5-IJ and M5-PJ Joulemeter Probes

ML562G84 APPLICATION. Display system. Symbol Parameter Conditions Ratings Unit. Pulse (Duty 40%)

CATALYTIC EMITTER - An old idea in new clothes

February 12, PHY357 Lecture 11. Experimental Methods. Accelerators. Particle Interactions. Particle Detectors. Full experiment (eg.

Transcription:

gray Instruction Manual for gray Mounted Detectors F: +41 44 633 13 68

2 / 12 gray Mounted Detectors: Instruction Manual CONTENT 1. SHORT USER GUIDE... 4 2. THERMAL INTEGRATION OF THE DETECTOR... 5 2.1. Mounting the detector on a heat sink... 5 2.2. Active cooling with water... 6 3. DATA ANALYSIS... 6 3.1. Laser measurement... 6 3.2. Temperature corrected sensitivity... 6 3.3. Wavelength corrected sensitivity... 7 4. USE WITH gray AMPLIFIER (AMP 2040, AMP 2050 or AMP 2060)... 8 5. MAINTENANCE OF THE DETECTOR... 9 5.1. Cleaning the detector... 9 5.2. Storage... 9 5.3. Recalibration... 9 6. GENERAL CONSIDERATIONS... 9 6.1. Electromagnetic interference... 9 6.2. Application in temperatures outside of the operating temperature range... 9 7. TECHINAL DRAWINGS... 10 7.1. B01-SMC... 10 7.2. B05-SMC... 10 7.3. B05-MC... 10 7.4. C50-MC... 11

3 / 12 gray Mounted Detectors: Instruction Manual Preface All gray laser power detectors are of high quality. To maximize benefit from their outstanding performance, some precautions must be taken during storage, assembly and packaging. Therefore, please read the following instructions carefully. Applicability This document is applicable to all gray mounted detectors supplied by greenteg AG. This document is specifically applicable to gray B01-SMC, gray B05-SMC, gray B05-MC and gray C50-MC. Precautions gray laser power detectors have long lifetimes when used under normal operating conditions. greenteg will repair or replace at its discretion, any gray detector which proves to be defective within a one year period or purchase, except in the case of product misuse. Any unauthorized alteration or repair of the product is not covered. greenteg is not liable for consequential damages of any kind. The user must avoid any misuse that could cause damage to the detector. Misuse includes, but is not limited to, laser exposure outside greenteg s published specifications, high voltage exposure outside greenteg s specifications, physical damage due to improper handling and exposure to harsh environments. Harsh environments include, but are not limited to, excessive temperature, vibration, humidity, chemicals or surface contaminants, exposure to flame, solvents or water, and connection to improper electrical voltage. greenteg products are not authorized for use as critical components in life support devices/systems or in any military application without the express written approval of a board member at greenteg. For support regarding any of the above points, please contact us at: greenteg AG Technoparkstr. 1 8005 Zurich Switzerland Email: Tel: +41 44 632 04 20

4 / 12 gray Mounted Detectors: Instruction Manual 1. SHORT USER GUIDE About the gray mounted detectors The gray mounted detectors are specifically designed to be integrated into systems where space is limited and electronic signal processing is handled remotely. Their mechanical integration allows for easy mounting onto a mechanical support, which may act as heat sink simultaneously. All mounted detectors contain gray thermopile detectors. Thermal detection of radiation power permits the detection of light across a broad wavelength spectrum (UV to MIR). gray mounted detectors are ideal for integration into power meters or laser systems for industrial or medical applications. As an option, the output signal can be amplified and normalized by a voltage amplifier unit, facilitating the straightforward conversion from voltage to power. B01-SMC & B05-SMC B05-MC C50-MC Figure 1: gray mounted detectors. The detectors are mechanically integrated for simpler handling. The electrical connection is done through solder pads. Prepare your measurement Connect the detector electrically by soldering wires to the solder pads of the detector (see Fig. 2). For detector B05-MC, the signals of the two sensors can either be read-out individually or differentially. For the differential signal, a solder bridge needs to be formed on the board at position SB in Fig. 2. Mount the detector onto a post or similar holder. To ensure sufficient cooling, an additional heat sink is recommended for both detectors. Procure a voltmeter as a read-out device. Optional: When used with temperature sensors, a suitable read-out device needs to be procured. Optional: When used with a voltage amplifier, procure a voltage source for the amplifier supply voltage. B01-SMC & B05-SMC B05-MC C50-MC Figure 2: Schematics of gray mounted detectors. The respective solder pads are indicated for the two detector models. How to record the detector signal Contact the detector electrically by connecting the two wires to your voltmeter and record the analog voltage signal U while the detector is illuminated. Never apply a power density higher than 1.5 kw/cm 2 to the detector coating as this will damage the absorber!

5 / 12 gray Mounted Detectors: Instruction Manual How to calculate the incident power The power of the laser incident to the detector surface is proportional to the voltage output of the detector. Its unit is W, and it is calculated using the following formula: Ф = U / Z [W] where U is the detector output voltage, in V; and Z is the radiant sensitivity of the detector, in mv/w. 2. THERMAL INTEGRATION OF THE DETECTOR This section describes the recommended mounting methods for the gray detectors. When measuring with a thermal detector, any thermal influence from the environment will cause background signals besides the actual radiation signal. If the sensor temperature is considerably different than the room temperature, conductive heat flow occurs, inducing an additional signal at the sensor. Further, the sensitivity Z of the sensor depends on the detector temperature (see Fig. 3). The change of Z needs to be taken into account when converting the output voltage into power. Therefore, in order to obtain reliable results, heating of the detector module needs to be avoided. This is achieved by providing passive or active cooling to keep the module temperature close to room temperature and stable throughout the measurement. The method of thermal stabilization depends on the incident power expected during the measurement. Use the following table as guideline for selecting the thermal stabilizer. Incident power on detector Thermal stabilizer (example) <1 W Heat sink (e.g. aluminum block) of 25 cm 3 <5 W Passive heat sink e.g. 5cm x 5cm x 2.5cm (thermal resistance <5 K/W) <20 W Passive heat sink e.g. 10cm x 10cm x 4cm (thermal resistance <2.5 K/W) <50 W Water cooling through supplied cooling circuit 2.1. Mounting the detector on a heat sink 1. If an element of the laser system, where the detector will be integrated into, has features or thermal resistance as stated in the table above, you can mount the detector directly onto it. Please note that these values are guidelines other important factors as the room temperature or air convection need to be taken into account for each system individually. 2. Ensure that the surface of the stabilizer is flat, dry, and free of dust and grease. Clean the backside of the detector surface with ethanol or isopropanol and use the screw holes to tighten the detector on the thermal stabilizer. Do not use acids or bases for cleaning the detector! When mounting the detector onto a heat sink, it is recommended to apply a heat conductive paste between the detector and the heat sink to ensure good thermal coupling. Air gaps are thermally insulating and

6 / 12 gray Mounted Detectors: Instruction Manual greatly distort the cooling efficiency of the heat sink. 2.2. Active cooling with water 1. Use any water/liquid cooling circuit of your liking and mount the detector onto it as described in Section 2.1. 2. The water/liquid temperature must be maintained at temperatures between 15 and 25 C to ensure accurate measurement results based on the calibration at 20 C. Temperature changes on the time scale of minutes can be misinterpreted as power fluctuations and need to be avoided. 3. Maintain a constant water/liquid flow rate. This can be controlled by a valve or the pump settings. Notice that flow rates that fluctuate or are too high may cause noise on the output signal of the detector. 3. DATA ANALYSIS This section contains the basic analysis methods needed to interpret data from the gray laser power detectors. 3.1. Laser measurement The power of the laser, incident to the detector surface, is proportional to the voltage output of the detector. Its unit is W, and it is calculated using the following formula: Ф = U / Z [W] where U is the detector output voltage, in V; and Z is the radiant sensitivity of the detector, in mv/w. Alternatively, the value of Z can be provided as a calibration constant with the detector. For gray mounted detectors, typical values are: B05-MC: C50-MC: Z = 100 mv/w Z = 0.5 mv/w 3.2. Temperature corrected sensitivity The sensitivity of the gray laser power detectors depend on the temperature level at which they are used. The temperature-corrected sensitivity of the detector is calculated using the following formula: Z = Zo + (T To) Zc [mv/w] where Zo is the radiant sensitivity at calibration temperature and calibration wavelength, in mv/w; Zc is the linear correction factor for the radiant sensitivity, in (mv/w)/ C; To is the calibration temperature (typically 20 C), in C; and T is the heat sink temperature level, in C. Values Zo, Zc, and To are detector specific calibration values. T can either be measured with the integrated T-Sensor (B05-MC) or a thermocouple mounted onto the aluminum plate on which the sensor is mounted (C50-MC).

7 / 12 gray Mounted Detectors: Instruction Manual Relative sensitivity change 8.0% 6.0% 4.0% 2.0% 0.0% 0 10 20 30 40 Temperature change from To [K] Figure 3: Dependence of sensitivity Z on detector temperature T. Plotted is the relative change of Z as a function of the temperature difference to the calibration temperature To=20 C. Figure 3 shows the temperature dependence of the sensitivity Z for the type C detector. Typically, the type C gray detectors show a highly linear increase of Z by 0.175 % per K. The type B detectors show a linear dependence with an increase of 0.125 % per K. 3.3. Wavelength corrected sensitivity The coating on the detector surface determines the percentage of incoming light that is absorbed. The gray absorber coating is an inorganic absorber designed for high damage threshold and broad band absorption characteristics. Although the spectral absorption is designed to be flat throughout the wavelength range from UV to MIR, a small variation of absorptivity cannot be excluded completely. All gray detectors are calibrated at 1064 nm. If a different measurement wavelength is used, the sensitivity should be corrected according to the wavelength specific absorption coefficient. The absorption spectrum is plotted in Figure 4 as absolute data. Figure 5 shows the same data but normalized to the absorption value at 1064 nm. 100% 90% 80% 70% Absorption 60% 50% 40% 30% 20% 10% 0% 100 1000 10000 Wavelength [nm] Figure 4: Absorption spectrum of gray broad band absorber coating. In order to obtain the exact value for Z at a specific wavelength, the provided calibration value for Zo needs to be corrected by the respective correction factor C. The latter can be read from Figure 5. As an example, the sensitivity of the C05-HC is Zo = 2 V/W at 1064 nm. If the measurement is carried out at 10.6 um, the following correction needs to be carried out: Z = Zo C = 2 V/W 89.3% = 1.79 V/W. If you need exact values for a specific wavelength, contact us.

8 / 12 gray Mounted Detectors: Instruction Manual Normalized absorption 110% 100% 90% 80% 70% 60% 50% 40% 30% 20% 10% 0% 100 1000 10000 Wavelength [nm] Figure 5: Absorption spectrum of gray broad band absorber normalized to absorption value at the calibration wavelength of 1064 nm. 4. USE WITH gray AMPLIFIER (AMP 2040, AMP 2050 or AMP 2060) We optionally supply a voltage amplifier with our gray detectors. Depending on the power range you intend to read-out, the basic gain will be adjusted by us. Fine tuning needs to be done inside your system, when the calibration is carried out. Figure 6 shows, how the amplifier is wired. The two wires on the top of the image are the connections to the detector (red: Out +, black: Out -). The read-out requires three wires for the voltage supply of the amplifier and the amplified signal. Solder the wires to the pads in the center of the circuit board (see Fig. 5). The color code is as follows: Red wire: Black wire: White wire: DC voltage supply 12-24 V Ground (shared between voltage supply and signal out) Detector signal out (0-10V) Provide the amplifier with the supply voltage and record the analog voltage signal U with your read-out device while illuminating the detector. It is convenient to adjust the amplifier gain to such a value, that the conversion from voltage to power gets straightforward e.g. 1 V/W. You therefore might want to adjust the amplifier gain accordingly while illuminating the detector with a known power. Turn the screw of the trimmer until the desired conversion factor is reached. For robust integration into your system, you can mount the amplifier with two screws. Use the two holes at the sides of the amplifier. Figure 6: gray AMP with electrical connections. By turning the screw of the trimmer, the amplifier gain can be adjusted.

9 / 12 gray Mounted Detectors: Instruction Manual 5. MAINTENANCE OF THE DETECTOR 5.1. Cleaning the detector Cleaning is only necessary before mounting the detector onto a heat sink. Clean the detector backside surface with ethanol or isopropanol. Once the detector is mounted, no further cleaning is necessary. Never touch the absorber surface and do not bring it in contact with any chemical agents. 5.2. Storage Store the unused gray laser power detectors at ambient temperatures in a clean and dry place. In order to protect the absorber surface, put it back into the shipping package (wrap the detector in the polymer foam). 5.3. Recalibration greenteg recommends a yearly recalibration to ensure accurate measurement results. This calibration can be carried out at greenteg or at any certified calibration institute. 6. GENERAL CONSIDERATIONS 6.1. Electromagnetic interference Due to the low electrical resistance of the detector and its aluminum packaging, electromagnetic interference is usually not of concern. If electromagnetic interference is observed within an application, the typical countermeasures (e.g. shielded cables, proper grounding) should be taken. 6.2. Application in temperatures outside of the operating temperature range The operation temperature range of the gray laser power detectors is stated in the respective data sheets. Outside of this range, thermal effects might be non-linear and too large to be compensated for.

10 / 12 gray Mounted Detectors: Instruction Manual 7. TECHINAL DRAWINGS 7.1. B01-SMC 7.2. B05-SMC 7.3. B05-MC

11 / 12 gray Mounted Detectors: Instruction Manual 7.4. C50-MC

12 / 12 gray Mounted Detectors: Instruction Manual LIST OF SYMBOLS Name Symbol Unit Laser power Φ W Detector output voltage (measured) U V Temperature corrected radiant sensitivity Z mv/w Radiant sensitivity at calibration temperature Zo mv/w Temperature correction factor for radiant sensitivity Zc (mv/w)/ C Spectral correction factor for radiant sensitivity C % Absolute thermal resistance K K/W Electrical resistance R Ohm Temperature of the cold side Tc C Calibration temperature To C Detector temperature T C Detector area A m 2 Detector thickness d µm Disclaimer The above given restrictions, recommendations, materials, etc. do not cover all possible cases and items. This document is not to be considered to be complete and it is subject to change without prior notice. Revision History Date Revision Changes 09. March 2015 1.0 Initial revision