Corning LSE Benchtop Shaking Incubator

Similar documents
Corning LSE 6L Digital Water Bath

Corning LSE Digital Dry Bath Heater

POLYMIX PX-IG 2000 Operating Instructions

Instruction Manual. For All Corning Hot Plates, Stirrers, Stirrer/Hot Plates with Digital Displays, and External Temperature Controller (6795PR)

Digital Heat Block User Manual

Instruction Manual. Model AccuPlate Hotplate AccuPlate Stirrers AccuPlate Hotplate Stirrer 12/10 CLS-EQ-LHP01REV

ES-20 Orbital Shaker-Incubator

i-cube Microplate Incubator E10030

INSTRUCTION MANUAL. Laboratory Equipment Pty Ltd Ph: Fax:

MD-MINI Dry Bath. Instruction Manual

Reacti-Therm I and III Heating/Stirring Modules

INSTRUCTIONS AND WARRANTY

Reacti-Therm I and III Heating Modules

Instruction Manual for the Revolutionary Science RS-IF-203 Incufridge

Package Contents. Dry Block Heater Power Cord Instruction manual Warranty Card. Serial Number: Date of Purchase: Supplier: Table of Contents

Dry Bath Heating Systems

StableTemp Hotplates, Stirrers and Stirring Hotplates

Decon Scrubber Owner s Manual

User Manual Watt Drop-In Induction Range. Model: DC /2017. Please read and keep these instructions. Indoor use only.

FastGene Plate Centrifuge. Instruction Manual. Catalog No. NG

Revolutionary Science RS-IF-202

Operating Instructions. Oil Column Heater. Models: GOC155, GOC157, GOC1511

TTT Digital Incubator

Wine Cellar. Instruction Booklet. Model: WC-2038

BEFORE FIRST USE: TABLE OF CONTENTS

Operator s Manual. Model G32-S Model G32-E Disinfection Soak Stations

8L DEHUMIDIFIER OPERATING INSTRUCTIONS

Table of Contents DIGITAL DRYING OVEN. Features. Setup. Cautions. Description of the parts. Temperature Controls. Maintenance and Storage

Operation Manual Vortex mixer unitexer 1

Model S-530 Coin Sorter / Value Counter Operating Manual

12L DEHUMIDIFIER OPERATING INSTRUCTIONS

TS-100C. Thermo-Shaker with cooling for microtubes and PCR plates. Operating Manual Certificate. Medical Biological Research & Technologies

Operator s Manual. The Quietek

WHYNTER PORTABLE ICE MAKER 49 LB CAPACITY - STAINLESS STEEL

General Care and Safety Guide

CR-1730R User s Manual

INSTRUCTIONS AND WARRANTY

Whynter Portable Ice Maker 33 lb capacity - White

WiseBath WSB Shaking Water Bath Operation Manual (Version 1.4.0)

MODEL NO RUSSELL HOBBS APOLLO 1.8L DIGITAL FILTER COFFEE MAKER

StableTemp Digital Hotplates, Stirrers and Stirring Hotplates

INSTALLATION AND OPERATION MANUAL

General Care and Safety Guide

PROCESSING UNIT II V Hz V Hz. Table of Contents

3800 North Mill Road Vineland, NJ USA Tel: Fax: Web: Rev. A

Note To The User: This product has updated features not yet mentioned in the product manual.

FAN-FORCED UTILITY HEATER

Health Check Ear and Forehead Thermometer Model No.: Instruction Manual

30-pint Dehumidifier User Guide

Models: SV1 (110V) SV1EU ( V) User s Manual. Register your product and get support at. v

Refrigerated Incubator Model and Operating Instructions

User s Manual and Operating Instructions

Instruction Manual for the Revolutionary Science RS-PB-200 Dual Poly Pro Bath

WHYNTER PORTABLE ICE MAKER 49 LB CAPACITY - STAINLESS STEEL

Tissue-Tek TEC Plus Cryo Console OPERATING MANUAL

De Longhi Evaporative Cooler

OWNER S MANUAL ELECTRIC FIREPLACE INSERT. Model No.: WESTERN (EF42D-45D) WARNING CAUTION

Operation Manual. (Version 1.5.1)

General Care and Safety Guide

Tower Fan NSTF-06MI IMPORTANT! KEEP FOR FUTURE REFERENCE. Instruction Manual. Please register your product on-line at

By Silver King Refrigeration, Inc.

PERSONAL 1500W CERAMIC HEATER Instruction Manual

BAR FRIDGE 118 & 228 LITRE

INSTRUCTIONS AND WARRANTY

EDR Series. Model. Instruction Manual. Rev 5.0. Thank you for purchasing the EDR Auto Smart.

MD 25/40/48. 3, 4, 5 level INSTRUCTIONS IMPORTANT! Installation, operation and care of: MD Series Multideck Hot Food Display Cases

Owner & Operator s Manual. Table of Contents

PORTABLE AIR CONDITIONER OWNER S MANUAL

Portable Heated Black Glass Shelves

Cavitator Ultrasonic Cleaner Instruction Manual Model 20G

Incubating Shaker with Cooling

Operating Instructions

BRITA. Water Filter & Chiller II. Chilled water filter. Instruction Booklet WF6100

BEFORE USE, PLEASE READ AND FOLLOW ALL SAFETY RULES AND OPERATING INSTRUCTIONS.

Wine Cabinet Manual. We know you don t want to read through this manual, but trust us, there is good information in here.

Digi-Sense TC9000 Advanced PID and On/Off Temperature Controller with Thermocouple Input

OPERATING INSTRUCTIONS

HI COD REACTOR and Test Tube Heater 2008 Series

User Manual. TN6900 H&E Slide Stainer. Read Prior to Operation! Always keep this manual near the unit! WP

INSTRUCTION MANUAL FOR Freeze Dry Systems PLEASE READ THIS MANUAL CAREFULLY BEFORE OPERATION

INSTRUCTIONS AND WARRANTY

MS-H-Pro-T Magnetic Hotplate-Stirrer with Timer

Read these instructions before using and keep them available at all times!

User Manual. TN1500 Embedding Center. Read Prior to Operation! Always keep this manual near the unit! WP

Read these instructions before using and keep them available at all times!

Store Name: Location: Date Purchased: Serial Number (located on back of unit):

INSTRUCTIONS AND WARRANTY

Type SP87300 SAFE-T SHP9 Explosion Proof Stirring Hot Plate

PORTABLE AIR CONDITIONER

AIRGOCLEAN 10 E OPERATING MANUAL AIR CLEANER TRT-BA-AIRGOCLEAN10E-TC-001-EN

GB Operating instructions

CERAMIC FAN HEATER URBAN

Stand Fan. English. Owner s Guide. Model F R

User Manual. Hot Water Dispenser. Models: 177HWD15G 12/2018. Please read and keep these instructions. Indoor use only.

Operator s Manual. IP-100 Immersion Probe Cooler

WCD-200W Water Dispenser

HR 601 C A AUS. ENGLISH Instructions for use Page 2

THERMOFILM CONVECTIVE PANEL HEATER INSTALLATION, OPERATION AND MAINTENANCE MANUAL. Models: CP 2400 CP 2000 CP 1500 CP 1000 TABLE OF CONTENTS

MV 4 MV 4 Premium. Operating Instructions. Before first use of the unit read these operating instructions and act in accordance with them.

ELECTRIC STOVE HEATER Models: FS2213B

Transcription:

Corning LSE Benchtop Shaking Incubator INSTRUCTION MANUAL Models 6790 (120V US) 6791 (230V EU)

Table of Contents 1.0. GENERAL INFORMATION GENERAL INFORMATION... 1 1.1 Definition... 1 1.2 Key features... 1 2.0. TECHNICAL DESCRIPTION... 2 2.1 Construction... 2 3.0 INSTALLATION... 2 3.1 Unpacking... 2 3.2 Selecting the right location... 2 3.3 Attaching the power cord... 2 3.4 Attaching shaker platforms... 2 3.5 Starting up... 3 3.5.1 Starting up checklist... 3 3.6 General safety recommendations... 3 4.0 INSTRUCTIONS FOR USE... 4 4.1 Introduction... 4 4.2 Basic operations... 4 4.3 Changing parameters during operation... 7 4.4 Shaking without heating... 7 4.5 Heating without shaking... 7 5.0 TROUBLESHOOTING... 8 5.1 Errors... 8 5.2 Description of possible errors... 8 6.0 TECHNICAL SPECIFICATION... 9 6.1 Technical data... 9 7.0 TEMPERATURE RE-ADJUSTMENT... 9 8.0 MAINTENANCE AND CLEANING INSTRUCTIONS... 10 8.1 Cleaning of the working chamber (painted stainless steel)... 10 8.2 Maintenance and cleaning of the PMMA (acrylic) lid... 10 9.0 POSITIONING OF ATTACHMENTS ON THE SHAKING PLATFORM... 11 10.0 COPYRIGHT AND TRADEMARK INFORMATION... 12 11.0 WARRANTY... 12 12.0 WASTE ELECTRICAL AND ELECTRONIC EQUIPMENT (WEEE) DIRECTIVE... 12

1.0 GENERAL INFORMATION 1.1 Definition Corning LSE Benchtop shaking incubator is a combination product of a shaker and a temperature chamber, designed for simultaneous shaking and heat treatment of samples, saving time and laboratory space. It is suitable for daily use in biochemistry, microbiology, and many other laboratories. 1.2 Key Features High temperature working range. Large environment temperature range suitable for cold room use. Digital readout of all parameters during run. Ability to change parameters during run (except timer). Last parameter memory, even if the device was switched off. Temperature accuracy +/- 0.5 C of absolute deviation from the set temperature in device working area. Incorporated HOLD and RUN time function with 1 minute resolution. Special parallel temperature protection for your safety. Easy interchangeable platforms. Lid sensor. 2.0 TECHNICAL DESCRIPTION 2.1 Construction The casing of the benchtop shaking incubator is made of steel plate varnished with polyurethane lacquer. Temperature chamber lid is made of Polymethyl methacrylate which is resistant to temperatures up to 90 C. Equipment consists of an independent shaking device and temperature chamber. Both sections are driven and regulated via microcontroller driven electronics, which also control all necessary sensors for motor speed, temperature, lid, temperature safety fuse, time, and self-diagnostics. 1

3.0 INSTALLATION 3.1 Unpacking Before starting the installation, carefully examine the delivery for possible damage or missing parts: Open the box and lift the device, together with the shock absorbers, out of the box. Remove the shock absorbers and check that the machine has not been visibly damaged during the transportation. Keep the packing material until you are sure that the machine functions properly. Check the information on the name plate on the rear side of the machine: - Type - Serial number - Article number Check that the main cord has a pin configuration that fits into a wall socket according to the local standard. If any kind of damage occurred during transportation, immediately make a complaint to the carrier. Any incorrect delivery or missing parts should be reported to the distributor. 3.2 Selecting the right location When selecting the right place for device, please consider the following: Put the device on a smooth, stable surface. Leave enough space on the back side of the device for normal air circulation. Leave enough working space around the device. Avoid areas with fast temperature and humidity changes, are in direct sunlight, are nearby heat producing devices, or where there is a possibility of shocks and/or vibrations. Do not use the device in a flammable or explosive atmosphere! 3.3 Attaching power cord Fit the main cord (included), into the main receptacle. Connect the other end of the cord to a grounded wall socket. 3.4 Attaching shaking platforms Platforms of many sizes are available to suit a wide range of applications. These platforms mount to the shakers via four mounting posts which easily insert into four rubber mounting points on the shaking platform. 2

3.5 Starting up 3.5.1 Starting up checklist Unpack and place the device as specified. Open the lid and place samples. Close the lid cover. Switch on the device by the ON/OFF switch on the front panel (indicator light will illuminate). Refer to section 4.2 for instructions for setting parameters. 3.6 General safety recommendations Before any installation read this User s Manual. Before cleaning, disconnect the device from the main power supply. Follow cleaning instructions in section 8. Do not use the device near water sources. Assure that no water drips in the device during cleaning or otherwise. Do not use device in explosive or flammable environments. NOTE: If the device is not working properly, follow only commands and procedures outlined in this Instruction Manual. Use of any other commands, procedures, or adjustments could result in device destruction or longer service repair time. If the equipment is used in a manner not specified by the manufacturer, the protection provided by the equipment may be voided. Not for use with flammable or explosive samples! 3

4.0 INSTRUCTIONS FOR USE 4.1 Introduction POWER 1. POWER switch: Switch to ON (power switch illuminates) or OFF. 2. RUN: Green light is on when incubator is running. 3. TEMPERATURE: Yellow light is on when the incubator is set to temperature. 4. SPEED: Yellow light is on when the incubator is set to speed. 5. TIME: Yellow light is on when the incubator is set on time. 6. ENCODER: Rotating the Encoder knob clockwise (+) or counter-clockwise (-) to change the Time, Speed, and Temperature settings. Push the Encoder knob to change between Time, Speed, and Temperature settings. If you rotate the Encoder knob fast, the value on display goes up or down quickly. 7. START/STOP key: START or STOP operating. 8. TEMPERATURE display. 9. SPEED/TIME display. 4.2 Basic Operation Press the POWER switch on the control panel. Shaking incubator automatically detects power supply frequency F50 or F60 and shows it on the display. Incubator will automatically be set to the last saved parameters. 4

To set Temperature: Press and hold the Encoder knob for more than 2 seconds. The Temperature indicator light will blink. Rotate knob clockwise (+) or counterclockwise (-) to set Temperature to desired value: 37.5 37.5 C If you want to use the shaking incubator without temperature regulation, set the display to OFF. The OFF function will display when you rotate the Encoder knob below 0.0 C or above 70.0 C. To set Speed: Press Encoder knob until the Speed light blinks. Rotate the Encoder knob clockwise (+) or counter-clockwise (-) to set Rotating Speed to desired value: 34 34 rpm 120 120 rpm If you want to use the shaking incubator without shaking, set OFF on display. The OFF function will display when you rotate the Encoder knob under 20 rpm or above 300 rpm. 5

To set Time: Press the Encoder knob until the Time indicator light blinks. Rotate the Encoder knob clockwise (+) or counter-clockwise (-) to set Time to desired value between 1 minute to 99 hours: 70.3 70 hr 30 min 1.55 1 hr 55 min 0.43 43 min 0.01 1 min To set Time to hold (continued operation) - set HLd on the display. The HLd function will display when you rotate the Encoder knob to under 0.01 minute or above 99.0 hours. Press START/STOP key once to exit the changing of parameters. You are now in starting mode. The light from the parameter that was last changed will remain on. Note: To exit the changing of parameters, always press the START/STOP key once. Press the START/STOP key to start the operation. The time will begin counting down and the Run indicator light will be illuminated. Time and Speed can be alternately viewed in the bottom display by pressing the Encoder knob during the run. 6

4.3 Changing parameters during operation Note: You cannot modify time during operation. If you want to change the Temperature during operation, press and hold the Encoder knob for more than 2 seconds. Temperature indicator light blinks. Rotate the Encoder knob clockwise (+) or counter-clockwise (-) to set the desired value. The Temperature indicator light will continue to blink until the desired value is set. To set the desired value, press START/STOP key only ONCE. The shaking incubator goes back to working parameters and actual temperature will show on the display. Holding the Encoder knob for more than 2 seconds will toggle the display back to the temperature. To change Speed or Temperature during operation, press and hold the Encoder knob until the Temperature or Speed indicator light blinks. To switch between setting Temperature and Speed, press the Encoder knob one time until the desired indicator light is illuminated. When finished press the START/STOP key one time to exit. When the run is completed, End appears on the display and the Run indicator light pulses. The Run indicator light will only turn off when the shaking incubator has come to a complete stop. The last used values for Time, Speed, and Temperature will remain for the next run. If you lift the lid during operation the shaking incubator will immediately stop and the message Opn Lid appears on displays. When you close the lid again, the shaking incubator will resume operation. ` In the event of a power outage, the shaking incubator will restart automatically, as soon as the electricity returns. A warning will be displayed indicating the power had been off. Pressing the Encoder knob will turn this off. 4.4 Shaking without heating If you want to shake the samples without heating them, set Temperature to OFF. 4.5 Heating without shaking If you want to heat the samples without shaking them, set Speed to OFF. 7

5.0 TROUBLESHOOTING 5.1 Errors Device will constantly check functions vital for safety and reliable operation. An error message will appear if there is a problem detected. Sample of Error display: 5.2 Description of possible errors Error Code Description Solution E 11 Something is wrong with the motor Contact Corning Customer Service. regulation (PWM regulator, pulse generator, motor). Shaking incubator is automatically stopped. E 12 The motor does not reach set speed in Contact Corning Customer Service. 30 seconds. Shaking incubator is automatically stopped. E 13 Set speed oscillates for more than 100 Contact Corning Customer Service. rpm in 2 seconds. Shaking incubator is automatically stopped. E 21 Temperature sensor does not work properly. Shaking incubator is automatically stopped. Check temperature sensor for any visible damage or material obstructions. Contact Corning Customer Service. E 22 Temperature sensor notices 5 C higher temperature than set temperature. Shaking incubator is automatically stopped. E 23 Set temperature is not reached in 2 hours. Shaking incubator is automatically stopped. Verify set temperature is not set more than +/- 5 degrees below chamber temperature. Contact Corning Customer Service. Verify that nothing is obstructing the lid from properly closing. Check that the circulating fan on the back panel is functioning. Contact Corning Customer Service. 8

6.0 TECHNICAL SPECIFICATIONS 6.1 Technical Data Heater power 700W Fan power 17W Shaking motor power 35W Fuse 2 x 3.15A 250V 2 x 6.3A 120V Environment temperature 4 C to 65 C Relative humidity Up to 85%, non-condensing Speed regulation Digital, load independent, from 20 to 300 rpm in 1 rpm steps Shaking orbit 19 mm Temperature operating range 5 C above room temperature to 70.0 C Temperature sensor PT100 Temperature accuracy ±0.5 C Temperature stability ±0.5 C Timer 1 min to 99.0 hours, timer HOLD (Hld) function Capacity 16 x 125 ml, 9 x 250 ml, 5 x 500 ml, 4 x 1,000 ml Dimensions (W x D x H) 370 x 530 x 400 mm (14.6 x 20.9 x 15.7 inches) Weight 20.2 kg (44.4 lbs.) Overvoltage category Installation Category I 7.0 TEMPERATURE RE-ADJUSTMENT Note: Temperature re-adjustment procedure is only for qualified technical persons. Measure the temperature in the middle of the incubator or in a sample (preferred) for a minimum of 2 hours. A calibrated digital thermometer with precision of 0.1 C or more should be used. After 2 hours or more, read the temperature on the thermometer and compare it with the temperature on the display of the shaking incubator. The difference between the temperatures on the thermometer and the display is the value to be entered. Example 1: Temperature on the thermometer is 37.9 C. Temperature on the display is 37 C. Difference is 37.9-37 = 0.9 Value is 0.9 Example 2: Temperature on the thermometer is 36.2 C. Temperature on the display is 37 C. Difference is 36.2-37 = -0.8 Value is -0.8 Procedure for temperature calibration: Press and hold the Encoder knob for more than 5 seconds message Cor (correction) appears on display. Press the Encoder knob once, then turn the Encoder knob until the value calculated above is displayed paying attention to the plus or minus sign. When you set the value, press the Encoder knob again. Press START/STOP key to finish the temperature re-adjustment. 9

Note: Temperature re-adjustment procedure is only for qualified technical persons. 8.0 MAINTENANCE AND CLEANING INSTRUCTIONS 8.1 Cleaning of the working chamber (painted stainless steel) The working chamber should be cleaned regularly with mild cleaning solutions. Abrasive cleaners should not be used. Spills should be cleaned up immediately. Use cleaners with neutral ph for decontamination. Stainless steel platform can be decontaminated by autoclaving (120 C). 8.2 Maintenance and cleaning of the PMMA (acrylic) lid Maintenance Acrylic glass is sensitive to scratching. Use only a soft cloth and non-abrasive cleaners. Grease and oil that may have accidentally spilled can be cleaned with hexane, petroleum, or other chemicals in aromatic base. Cleaning Use: It is best to use 2% aqueous solution of non-aggressive cleaning products (dishwashing detergent). Wipe the surface with a lightly damp soft cloth. Remove dust and dirt with dry soft cloth. Do not use: Do not use acetone, benzene, paint thinner, carbon tetrachloride, or other aggressive liquids, because they can damage the surface of the acrylic lid. Do not use cleaners containing ketones or esters. Do not use cleaners with solid and abrasive particles. Do not use polishes. Never clean the surface with products containing acetone or other solvents or with cleaning products for chemical cleaning, paint thinner, gasoline, and similar. Avoid any contact with these chemicals. Do not use detergents with high levels of alcohol. Device can be very hot after operating, especially if used at temperatures higher than 60 C. Do not touch the platform until it is cooled. Before you begin cleaning the device, unplug the main cord from the power supply socket. Check with the manufacturer before using any other cleaning or decontamination methods to assure that the equipment is will not be damaged. 10

9.0 POSITIONING OF ATTACHMENTS ON THE SHAKING PLATFORM Capacity of platform is: 16 x 125 ml flasks, 9 x 250 ml flasks, 5 x 500 ml flasks, 4 x 1000 ml flasks. Use the following diagram as a guideline for positioning different flasks and accessories. 11

10. COPYRIGHT AND TRADEMARK INFORMATION All goods and services are sold subject to the terms and conditions of sale of the company that supplies them. The company reserves the right, subject to any regulatory and contractual approval, if required, to make changes in specifications and features shown herein, or discontinue the product described at any time without notice or obligation. Corning Incorporated Life Sciences 836 North St. Building 300, Suite 3401 Tewksbury, MA 01876 Tel: 800.492.1110 Tel: 978.442.2200 Fax: 978.442.2476 www.corning.com/lifesciences For more specific information on claims, visit the Certificates page at www.corning.com/lifesciences. 11. WARRANTY The Corning products are warranted against defects in materials and workmanship for two years unless otherwise outlined on your sales order. If any defect occurs in the instrument during this warranty period, Corning will repair or replace the defective parts at its discretion without charge. However, the following defects are specifically excluded: Defects caused by improper operation. Repair or modification done by anyone other than a Corning or an authorized Corning agent Use of spare parts supplied by anyone other than Corning Damage caused by accident or misuse Damage caused by disaster Corrosion caused by improper solvents or samples Warranty/Disclaimer: Unless otherwise specified, all products are for research use only. Not intended for use in diagnostic or therapeutic procedures. Corning Life Sciences makes no claims regarding the performance of these products for clinical or diagnostic applications. 12. PRODUCT DISPOSAL INFORMATION FOR CORNING CUSTOMERS AND RECYCLERS Corning Benchtop Shaking Incubator models 6790 and 6791 According to Directive 2012/19/EU of the European Parliament and Council of 4 th July 2012 on waste and electronic equipment (WEEE) as amended, Corning Benchtop Shaking Incubator is marked with the crossed-out wheeled bin and must not be disposed of with domestic waste. Consequently, the buyer shall follow the instructions for reuse and recycling of waste electronic and electrical equipment (WEEE) provided with the products and available at the following link: www.corning.com/weee. 12

For additional information, visit www.corning.com/lifesciences or call 1.800.492.1110. Outside the United States, call 978.442.2200. For Corning technical information, e-mail us at: ScientificSupport@corning.com or call 1.800.492.1110. Outside the United States, call 978.442.2200. For more specific information on claims, visit the Certificates page at www.corning.com/lifesciences. Warranty/Disclaimer: Unless otherwise specified, all products are for research use only. Not intended for use in diagnostic or therapeutic procedures. Corning Life Sciences makes no claims regarding the performance of these products for clinical or diagnostic applications. 13

Corning Incorporated Life Sciences 836 North St. Building 300, Suite 3401 Tewksbury, MA 01876 t 800.492.1110 t 978.442.2200 f 978.442.2476 www.corning.com/lifesciences Worldwide Support Offices ASIA/PACIFIC Australia/New Zealand t 61 427286832 China t 86 21 3338 4338 f 86 21 3338 4300 India t 91 124 4604000 f 91 124 4604099 Japan t 81 3-3586 1996 f 81 3-3586 1291 Korea t 82 2-796-9500 f 82 2-796-9300 Singapore t 65 6572-9740 f 65 6861-2913 Taiwan t 886 2-2716-0338 f 886 2-2516-7500 EUROPE France t 0800 916 882 f 0800 918 636 Germany t 0800 101 1153 f 0800 101 2427 The Netherlands t 31 20 655 79 28 f 31 20 659 76 73 United Kingdom t 0800 376 8660 f 0800 279 1117 All Other European Countries t 31 (0) 20 659 60 51 f 31 (0) 20 659 76 73 LATIN AMERICA grupola@corning.com Brasil t (55-11) 3089-7400 f (55-11) 3167-0700 Mexico t (52-81) 8158-8400 f (52-81) 8313-8589 For a listing of trademarks, visit us at www.corning.com/clstrademarks. All other trademarks are the property of their respective owners. Corning Incorporated. All rights reserved. Printed in USA 9/15 CLS-EQ-108 14