OX100, OX102. Limiting-current type Oxygen Analyzer OX100, OX102. Bulletin 11M10A01-03E

Similar documents
OX100/OX102. Limiting-current type Oxygen Analyzer. Bulletin 11M10A01-03E

Intelligent ph Converter PH400G. PH400G Intelligent ph Converter. Bulletin 12B7C1-01E.

Personal ph, ph/orp, Conductivity Meters

Personal ph, ph/orp, Conductivity Meters

New Features TDLS 200. The Process TDLS Solution. Tunable Diode Laser Spectroscopy (TDLS) Analyzer.

Personal ph, ph/orp, Conductivity Meters PH71/PH72/SC72

General Specifications

General Specifications

User s Manual. AXF Series Magnetic Flowmeter Installation Manual ATEX and IECEx Explosion Proof Type IM 01E20A01-11EN IM 01E20A01-11EN.

General Specifications

SPECIFICATION SHEET. SLC-1615 (for low concentration) SLC-1625 (for high concentration) SILICA ANALYZER

FAST/TOOLS. Alarm System Performance Analysis (ASPA) Bulletin 50A01A00-02EN

Series 3300 TELEDYNE ANALYTICAL INSTRUMENTS. The 3300 series of low cost, microprocessorbased

General Specifications

General Specifications

+ A B V B NO NC V COM PV PV

MEMS Carbon Monoxide Gas Sensor. (Model No.:GM-702B) Manual. Version: 1.1. Valid from: Zhengzhou Winsen Electronics Technology Co.

pre-wired E3FA-TP11-F2 2M E3FB-TP11-F2 2M

Process Gas Chromatograph

ROYCE SUSPENDED SOLIDS SYSTEM MODEL 7011A AND SENSORS MODELS 72A, 72P, 73B, 74A, 75A & 76A

General Specifications

Two global companies, one solution.

User s Manual. Model VJRK IM 77J01R21-01E 1. CAUTIONARY NOTES FOR SAFE USE OF THE PRODUCT 4. MODEL AND SUFFIX CODES 5.

DF-340E PROCESS ANALYZERS APPLICATIONS FEATURES

SPECIFICATION SHEET ORGANIC POLLUTION MONITOR. Model: OPM-410A - Organic Pollution Monitor - Issue: OPM-410A-0107-R0 Page 1 ISO ISO-9001

PCMini52 Digital RH & Temperature Mini Probe User s Manual

STT 3000 Series STT250 /150 Surge Protection Device Specifications

air.iq Moisture Analyzer Packaged Solution GE Measurement & Control Applications Features

OIL FILM DETECTOR (Flame-proof Version) SODL-1600

General Specifications

MODEL A-316 CARBON MONOXIDE (CO) MONITOR AND ALARM FOR COMPRESSED AIR TESTING

IN-SITU ZIRCONIA OXYGEN ANALYZER <HART communication>

M5200 Industrial Pressure Transducer

Sensepoint. Fit For Purpose Flammable, toxic and oxygen versions available High performance, low cost Suitable for new and retro fit applications

PRODUCT INFORMATION. GM901 Carbon Monoxide Gas Analyzers CO MEASUREMENT FOR EMISSION MONITORING AND PROCESS CONTROL. In-situ Gas Analyzers

Type 2 safety light curtain Type 2 according to the IEC/EN parts 1 and 2 Compact profile: 31 x 32 mm 30 mm resolution for hand detection Wide r

Dräger PointGard 2100 Detection of toxic gases and vapours

IWAKI HI-CERA PUMPS V

IN-SITU ZIRCONIA OXYGEN ANALYZER

Gemini. MAG500, MPG500 Vacuum Gauge. Wide Range Heavy Worker The compact solution for vacuum measurement in challenging environments

Standard SPECIFICATIONS FEATURES APPLICATIONS

User's Manual. Model VJQ8 Pulse to Analog Converter. (Multi-function) MODEL AND SUFFIX CODES CAUTIONARY NOTES FOR SAFE USE OF THE PRODUCT

Compact Ultra Low Temperature Chamber

FUTURE AIR SOLUTIONS OIL FREE SCROLL COMPRESSORS

85 Uncompensated SPECIFICATIONS FEATURES APPLICATIONS STANDARD RANGES. 316L SS Pressure Sensor Small Profile 0-100mV Output Absolute and Gage

U5300 Industrial Pressure Transducer

Harrier + Smart Controller & Chemical Management System

Pulp & Paper Solution. Pulp & Paper Solution. Instruments and Solution for Pulp & Paper Industry LF3BUSS03-00EN.

CER-1 & CER-2 Site Requirements

CER OPTIMA. Endoscope Reprocessing System CER-1 & CER-2 Site Requirements. Introduction. Space Requirements

86 Compensated SPECIFICATIONS. 316L SS Pressure Sensor Small Profile 0-100mV Output Absolute and Gage Temperature Compensated

Mark 25 Ultrapure Water Conductivity Analyzer

Model OEM-2 INSTRUCTIONS FOR USE

ESD Safe UL Approved INSTRUCTIONAL MANUAL EX V EX V TEMPERATURE CONTROLLED SOLDERING STATION

Dew-Point Transmitters and Instruments

ER-TF SERIES. Wide-area Ionizer Steady-state DC Method. Slim in shape, Wide in charge removal area, An evolutionary form in expression.

Sieger Sensepoint. Easy to Use Factory set operation Simple to replace Minimal training required

Using the Model 1500 Portable Chilled Mirror Hygrometer

THE YOKOGAWA PHILOSOPHY

FLEXIBLE, AFFORDABLE, HIGH SPECIFICATION OXYGEN GAS ANALYSIS DESIGNED FOR DEMANDING HAZARDOUS AREA PROCESS CONTROL APPLICATIONS.

Dräger Bump Test Station Calibration and Bump Testing

Accuracy by Design Products for the HVAC Professional

CLARITY II TURBIDIMETER

IN-SITU ZIRCONIA OXYGEN ANALYZER

SHAKING WATER BATHS ENGLISH

OPERATING INSTRUCTIONS AND REPLACEMENT PARTS

Economical Lead Free Reflow Oven Series

Site Requirements. MODEL 2.0 Australia

SIL 2. Ultima X Series Gas Monitors CERTIFIED

85 Compensated SPECIFICATIONS. 316L SS Pressure Sensor Small Profile 0-100mV Output Absolute and Gage Temperature Compensated

89BSD Digital Output

Customized Industrial Products

GMS800 FIDOR TOTAL HYDROCARBON ANALYZER CONTINUOUS MONITORING OF HYDROCARBON EMISSIONS

SOLDER POT POT-100C / 103C POT-200C / 203C

SHIMADEN DIGITAL INDICATORS

As a company, our goal is to contribute to society through broad-ranging activities in the areas of measurement, control, and information.

Zirconia Oxygen Gas Analyzers

U86B. Urea Dosing SPECIFICATIONS FEATURES APPLICATIONS

Sensepoint. Fit For Purpose Flammable, toxic and Oxygen versions available High performance, low cost Suitable for new and retro fit applications

MS1471 SPECIFICATIONS. PC Board Mountable Pressure Sensor 0-60 mv Output Gage and Absolute Low Cost

5New features & EXtreme design, now with HART Protocol and DuraSource Technology. Ultima X Series Gas Monitors

Catalytic Flammable Gas Sensor. (Model: MC105B) User s Manual. Version: 1.3. Valid from: Zhengzhou Winsen Electronics Technology Co.

General Specifications

Model C-380. Instruction Manual. Refrigerant Leak Detector. reedinstruments

154BC SPECIFICATIONS FEATURES APPLICATIONS STANDARD RANGES

Smoke Sensor. (Model:MQ-2) Manual. Version: 1.4. Valid from: Zhengzhou Winsen Electronics Technology Co., Ltd

Repair System. Instruction Manual

General Specifications

APPLIED. OMA-300 Hydrogen Sulfide Analyzer H 2. Ex p. Ex d. The world s safest continuous H 2. S analyzer. The OMA continuously measures H 2

POT-100C / POT-102C POT-200C /POT-202C

YTA Series Temperature Transmitters [Style: S3] EN :2006, -11:2007, -26:2007. Supply circuit: Ui=30 V, Ii=165 ma,

Models TX-KE and TX-KP Toxic Gas Transmitters

SAFETY MANUAL. Intelligent Sensors for H 2 S Gas Applications

Safe Operation Manual. Honeywell IQ Force Gas Detector

Air-Quality Detection Module. User s Manual

Catalytic Gas Sensor. (Model: MC113/ MC113C) Manual. Version: 1.3. Valid from: Zhengzhou Winsen Electronics Technology Co.

Improve Performance and Reduce Maintenance Costs. Model Q45D. Dissolved Oxygen Monitor...

OMA-300 Process Analyzer

BA2702-P Linear Actuator

316L SS Pressure Sensor High Pressure 0-100mV Output Absolute

Transcription:

Limiting-current type Oxygen Analyzer OX00, OX02 OX00, OX02 Limiting-current type Oxygen Analyzer One Instrument measures 00 ppm to 25% Oxygen Simple "One Touch" Calibration Self Diagnostics Assure High Reliability "One Touch" Replacement Compact Size Accessories to Assure Accuracy Dedicated Sampling Unit (Separate from Oxygen Analyzer) (Removes Dust, Mist and Organic Solvents) Calibration Kit Simple On-Site Calibration (Using liter Gas Cylinder) Bulletin M0A0-03E www.yokogawa.com/an/

Low range 0 to 00 ppm Autoranging in four ranges from 0-00 ppm to 0-25% High and Low Alarms Spans 000 ppm to 25% Autoranging in two ranges, 0-000 ppm and 0-25% High Alarm or Low Alarm (depends on specification) Oxygen concentration FAIL lamp High, Low Alarm lamps Alarm Fixed-range 4-20 ma / -5 V, Autorange 4-20 ma, keys Calibration keys holder: 7 mm (* ) dia. or G3/8 Plastic sensor holder (Specify Option code /J or L98LA (* 3) ) plus cable: 5 m (OX00), 5, 0 or 30 m (OX02) : When mount in 7 mm hole in panel, panel thickness can be up to 4 mm. 2 : Within 5 cm of sensor tip, can reach 250 C. 3 : Specify Option code /S, or Model K9424DB, for metal nipple. (* 2) OX00 and OX02 sensors are not interchangeable Lit when range is.0 to 25% between 0 to.0% and 0 to 25% other than 0 to 0,000ppm Lit when range is 0 to 0,000 ppm Validates parameter entry High alarm or Low alarm lamp (depends on specification) Oxygen concentration Calibration air flow lamp Lit during alarm setting Alarm setting key Mode switching key Main applications Globe box (processing crystal parts) Globe box (in research lab.) N2 reflow furnace / auto soldering Clean room Aging machine Monitoring for deficiency of O2 Foodstuffs Measurement cautions Usage cautions Box is to be filled with inert gas If organic solvents in box, use sampling unit Use sampling unit, remove alcohol / flux from furnace Use sampling unit if organic solvents present High temperatures, so use metal nipple K9424DB Use OX02 with Hi/Lo Alarms, or OX00-3 (Lo) Use sampling unit if disinfectant is alcohol-based * : should not be exposed to silicone vapor, silicon gas. * 2 : Corrosive or flammable gases, halogens, Freon, or organic solvents may adversely affect sensor. Fail lamp Units (% or ppm) Lit when range is 0 to 0,000 ppm not when range is 0 to 25% Lit during 0 to 000 ppm calibration Entry setting key Example: N2 reflow furnace Temperature : 20-30 C Gas flow : m/sec. Environment : Alcohol / Rosin / Flux, Oxygen 00 ppm Compared with oxygen meter it replaced, Response was faster, so changes were quickly detected Easy maintenance for Sampling unit (easy replacement of filter, activated carbon) 000 500 Oxygen ppm Approx. 6 sec. Fast response A OX02 oxygen analyzer B Installed oxygen analyzer A Difference Example: Measuring oxygen concentration in welding furnace Temperature : Room temperature, atmospheric pressure Gas flow : 30 Nl/min. (N2) Environment : Oxygen about 500 ppm in N2 Low oxygen concentration in furnace: safety precautions required OX00 can detect atmospheric air leakage into furnace which causes oxygen to rise to 200 ppm, so can help with welding quality assurance. 200ppm Oxygen ppm 000 500 Air leak 20mm/hr B A A OX00 oxygen analyzer B Dew point measurement 0 Solder flux flow start B 0

OX00 EXTERNAL DIMENSIONS Convertor 96 26.2 Bracket (Parts : B9877BA) 3 00 9.6 ALM FAIL 9 2 0 3 % 9 OXYGEN ANALYZER SNSR 4-20mA 2 0 ppm 4 2 MODEL OX00 3 HTR ALM 4 2 SUFFIX -5J 96 5 3 5 3 9.6 No FAIL 6 4 PART No K9424AA 7 5 00VAC SET AIR 000ppm 4-20mA 8 6 6 4 7 Made in Japan 7 5 MODE ENT. 8 6 7 23 ~0 Panel thickness Weight : Approx 0.6kg 89 + 0 Panel cut dimension 89 + 0 Other instrument 55 00 Min. space for multiple installation 00 * * : Allow the minimum clearance of 5mm for removable installation.

OX02 EXTERNAL DIMENSIONS Convertor 88 03 96 (22) 0 80 22 H.ALM L.ALM % ppm FAIL PASS 96 88 7 + 4-20 5 + 2 ma -5V - 6 3 + - RNG 7 NC 4-8 NC 9 ALM 3 NC 0 4 00-240 FAIL 5 VAC 2 6 G 7 ALM 4-20mA % ppm 7 SNSR CAL ENT Weight : Approx.2kg 89 + 0 89 + 0 Other instrument 55 00 00 * * : Allow a minimum distance of 5mm for ease of removal. Panel cut dimension Min. space for multiple installation (90) 8 8 0.5 (50) parameter Table NO. XXXX Gain.05 A.043 Gain2 B -65 A.002 Gain3 B -25 A 0.990 Gain4 B -9 A.429 Gain5 B -7 4 50 Temp. 250 C Max. Temp. 50 C Max. Temp. 70 C Max. Max temp. 70 C when plastic nipple (option code "/J" or P/N L98LA) is used. External Dimension L(m) D-sub 5 pin Male Model Code OX02- L (m) 4 OX02-2 9 OX02-3 29

Sampling unit (K9424GA) Measurement gas inlet Measurement gas outlet Filter Pump 2 3 Activated charcoal, 4 Flow membrane filter meter Needle valve connector Pressure meter (flow adjustment) Gas outlet cock (K9424JC, JD) 000 ppm O2/N2 Quantity 6, one-liter Calibration gas cylinder (K9424JP) Sampling unit (K9424GA) 25 x 272 x 40 (mm), approx 4.8kg Membrane filter: removes dust / mist / flux 2 Pump 00-240 V AC,.2A 3 Activated charcoal, membrane filter: remove organic solvents 4 Flow meter: 0. to liters / min. Operating conditions: Atmospheric pressure, gas temperature 0 to 40 C Mounting Rc /4 Power 00-240 V AC Calibration kit (K9424JA, JB) Gas outlet cock + Qty. 6, one Calibration gas cylinder - 20 x 75 x 335 mm, approx 800 g. Sampling Unit (K9424GA) Sample inlet (Rc /4) Sample outlet (Rc /4) Power line connector Unit:mm Activated carbon filter (220) ( 240 ) ( 36 ) ( 46 ) ( 290 ) Needle valve Flowmeter Power line switch Connector to joint sensor (ø8) Calibration Kit (K9424JA, JB) Standard Gas ( 6 cylinders ) OX00/OX02 (30) Gas vent tube Pressure gage Approx. 80 (90) Approx. 300 K9424JC NEEDLE VALVE ASSY With Pressure gage and Capillary /8 Rc Piping connection for K9242JD Standard gas cylinder (255) ø75 K9424JD NEEDLE VALVE ASSY Approx. 340

OX02 Standard Specifications Model Meas. principle Sampling Sample gas condition Measured gas Mounting Meas. ranges Repeatability Calibration Fine tune mode Analog output Contact output Alarm output Self-diagnostics Storage environment Converter size OX02 Limiting-current type Natural diffusion Temperature 0-70 C, Humidity 0-80% RH (dew free), Pressure 03 ± 40 hpa (Abs) O2 in N2 or mixed gases; cannot be used with silicone vapor, combustible gases or organic solvents Indoor, panel mounting 0-00 ppm / 0-000 ppm / 0-0000 ppm / 0-25%, automatic / fixed switching mode ±5 ppm (Meas. range : 20-00 ppm), ±25 ppm (Meas. range : 00-000 ppm), ±0.3%of reading (Meas. range : 000-5000 ppm), ±7.5% of reading (Meas.range : 5000-0000 ppm), ±0.25% (Meas.range : -25%). Atmospheric pressure calibration (between 5 to 25% O2) 2. 00 ppm calibration (between 80 and 20 ppm) 3. 000 ppm calibration (between 900 and 00 ppm) 4. 9000 ppm calibration (between 8500 and 9500 ppm) 5. Atmospheric pressure, 20, 00, 500, 000 and 9000 ppm calibration The measuring value can be adjusted to true analyzed value.. 4-20 ma DC isolated, fixed range (0-00, 0-000, 0-0000 ppm, 0-25% selectable),or 4-20 ma DC autoranging. 2. -5 V DC, fixed range (0-00, 0-000, 0-0000 ppm, 0-25% selectable) 3. Zone signal output for 4-20 ma DC autoranging Two outputs, rated 0 V AC, 24 V DC, A Hi or Lo alarm, sensor disconnected alarm Hi or Lo alarm. setting range : ppm to 24.9%. disconnected and converter temperature 0 to 70 C (dew free) 00-240V AC ±0%, 50/60Hz 5VA 96 x 96 x 90 mm OX00 Standard Specifications Model Meas. principle Sampling Sample gas condition Measured gas Mounting Meas. ranges Repeatability Calibration Natural diffusion Temperature 0-70 C, Humidity 0-80% RH (dew free), Pressure 03 ± 40 hpa (Abs) O2 in N2 or mixed gases; cannot be used with silicone vapor, combustible gases or organic solvents Indoor, panel mounting 0 to 000 ppm, 0 to 25% (measuring range autoswitching) ±5% of span (for 0-000 ppm range, with O2 being in the range 0 to 00 ppm). ±2% of span (for 0-000 ppm range, with O2 being in the range 00 to 000 ppm). ±2% of span (for 0 to 25% range).. Atmospheric pressure calibration (between 5 to 25% O2) 2. 000 ppm standard gas calibration (950±50 ppm) Analog output 4-20 ma DC isolated, fixed range (0-000 ppm or 0-25%), (for 0-25%, 4 ma DC corresponds to 500 ppm or less). Contact output Alarm output Self-diagnostics Storage environment Converter size OX00 Limiting-current type Two outputs, rated 250 V AC, A Hi or Lo alarm, sensor disconnected alarm High alarm contact output if concentration abnormal Hi or Lo alarm (depending on specification), setting range : ppm to 24.9% disconnected and converter temperature 0 to 70 C (dew free) 00 V AC ±0%, 50/60 Hz, 5 VA 96 x 96 x 90 mm OX00 measuring principles, measured gas, and system configuration are same as for OX02 If you want an oxygen (concentration) control range of 500 to 4500ppm, select the OX02. Restrictions Can't be used in flammable or exhaust gas, corrosive gas, or organic solvent environment To use in an environment with organic silicone, flow furnace, or N2 reflow furnace: () Use a sampling unit(k9424ga), Can't use the OX00-. (2) The maintenance requirements (filter and activated charcoal) depend on the measured gas. Be sure to read the manual and follow the safety precautions in it.

OX02 Limiting-current type Oxygen Analyzer Model Suffix Code Option Code OX02 Measuring range Manual - 2 3 N OX00 Limiting-current type Oxygen Analyzer Model Suffix Code Option Code OX00 Application - -2-3 Measuring Range Part No. K9424GA K9424JA K9424JB 5 Manual J E L98LA plastic resin sensor holder is provided as standard accessory. * For measurement in the range of 500 to 4500 ppm, use the OX02. N J E Option code* /S /J *Standard Product does not have a sensor holder. To get the sensor holder, specify " /S " or " /J". (If using Sampling Unit, the sensor holder is not required.) Sampling unit, Calibration kit Enhanced Oxygen analyzer 0-00/000/0,000 ppm / 0-25% O2 With sensor (5 m; sensor assembly. m + cable 4 m) With sensor (0 m; sensor assembly. m + cable 9 m) With sensor (30 m; sensor assembly. m + cable 29 m) Without sensor assembly Japanese English Metal sensor holder Plastic resin sensor holder General purpose Oxygen Analyzer Oxygen Analyzer (Can't use an activated carbon filter)* For (N2) reflow furnace (Specify in case of using an activated carbon filter)* Monitoring for deficiency of Oxygen with low alarm, Measuring range 0-25% 0-000 ppm / 0-25% O2 00 V AC ±0%, 50/60 Hz assy. cable 3m + extension 2m Without sensor assembly Japanese English Sampling unit Calibration kit: Six -liter gas cylinder, gas outlet cock (with flow adjuster) Calibration kit: Six -liter gas cylinder, gas outlet cock (without flow adjuster) Accessories Part No. K9424GY K9424JC K9424JD K9424JP K9424BA K9424CA B9877BA K9424LA K9424LB K9424LC K9424LD K9424DB L98LA Activated charcoal element, five per pack: for sampling unit Gas outlet cock (with flow adjuster): qty. one Gas outlet cock (without flow adjuster): qty. one Standard gas cylinder ( 000 ppm O2/N2 ) : qty. six assy. cable 3 m for OX00 only Extension cable, 2 m: for OX00 only Bracket for mounting OX00, two per set. assy. One per set : for OX02 only Extension cable, 4 m: for OX02 only Extension cable, 9 m: for OX02 only Extension cable, 29 m: for OX02 only holder (metal) for OX00 or OX02 holder (plastic) for OX00 or OX02 YOKOGAWA ELECTRIC CORPORATION World Headquarters 9-32 Nakacho 2-chome, Musashino-shi, Tokyo 80-8750, JAPAN Tel: +8-422-52-636 Fax: +8-422-52-669 http://www.yokogawa.com/an/ North America YOKOGAWA CORPORATION OF AMERICA Georgia, U.S.A http://www.yokogawa.com/us/ South America YOKOGAWA AMERICA DO SUL LTDA. BRAZIL http://www.yokogawa.com.br/ Europe YOKOGAWA EUROPE B.V. European Headquarters THE NETHERLANDS http://www.yokogawa.com/eu/ YOKOGAWA ELECTRIC CIS LTD. RUSSIAN FEDERATION http://www.yokogawa.ru/ Middle East YOKOGAWA MIDDLE EAST B.S.C.(C) http://www.yokogawa.com/bh/ Singapore YOKOGAWA ENGINEERING ASIA PTE. LTD. http://www.yokogawa.com/sg/ Korea YOKOGAWA ELECTRIC KOREA CO., LTD. http://www.yokogawa.com/kr/ China YOKOGAWA CHINA CO., LTD. http://www.yokogawa.com/cn/ YOKOGAWA SHANGHAI INSTRUMENTATION CO., LTD. http://www.ysi.com.cn/ YOKOGAWA SICHUAN INSTRUMENT CO., LTD. http://www.cys.com.cn/ Taiwan YOKOGAWA TAIWAN CORPORATION http://www.yokogawa.com.tw/ India YOKOGAWA INDIA LTD. http://www.yokogawa.com/in/ Australia YOKOGAWA AUSTRALIA PTY. LTD. http://www.yokogawa.com/au/ Represented by: Vig-PMK-G NL-0E Printed in Japan, 3(KP) [Ed : 05/b] Subject to change without notice. All Rights Reserved, Copyright c 200, Yokogawa Electric Corporation.