FA compressed air plants (refrigeration/adsorption driers) - granulate driers - medical gases - non-corrosive gases, e. g.

Similar documents
DP 300 mobile dew point instrument

Digital Display DS 52 in wall housing. Order no

Compressed air quality meas. according to ISO 8573 Residual oil content - particles - moisture

EN - english. Instructions for installation and operation. Mobile humidity meter DP 207

WIO200 Water In Oil measuring instrument

graphic indication of the dew point curve/ the switching behavior of the dryer over a longer period of time.

Dew point set DS 400

Dew point. Special features: Dew point monitoring DS 400. Technical data DS 400. Technical data FA 510

DP 300 mobile dew point instrument For measuring pressure dew point and atmospheric dew point in different applications:

English. Instruction and operation manual S 530. Leak detector for pneumatic systems

Duct Humidity and Temperature Transmitter. Features. IP65 protection. Ordering

English. Instruction and operation manual S 530. Leak detector for pneumatic systems

Installation and operation manual DPC200 - DIFFERENTIAL PRESSURE CONTROLLER. Low pressure with PI-control-mode

Warranty Registration

EN-ENGLISH INSTRUCTION MANUAL. METPOINT LKD 300 Leak Detector (LD109) For leak detection in pneumatic systems.

WIO200 Water in oil sensor

Snifter ATEX22 VERSION. User Manual. Distributor

Duct Humidity and Temperature Sensors 0-10 Vdc or 4-20 ma Outputs Passive sensor PT1000, PT100 NTC, NI1000 etc. Features. IP65 protection.

Version 1.6. Operating instructions HYDROMETTE BL COMPACT TF 2. Hydromette BL Compact TF 2

M-Sens 2. Online moisture measurement for solids. Operating Instructions. SWR engineering Messtechnik GmbH PART OF THE ENVEA GROUP

Hygro Transducer, Hygrothermal Transducer, and CO 2 Measuring Probe for Building Management and Climate Monitoring

Danfoss gas detection units

Hygro and Hygrothermal Transducers (Capacitive) for Air Conditioning Applications

Danfoss gas detection units

Measurement Solutions for Refrigeration Technology.

M-Sens 2. Online-Moisture Meter for Solids. Operating Instructions. SWR engineering Messtechnik GmbH

Chlorine gas leak detector

Specifications: ACCURACY: RH: ±1% Temperature: ±0.5 C. Dew point: Typically, +/-2.0C above 0C; +/-3.0C 0 to -60C OPERATING RANGE:

portable gas analyser

Hygro and hygrothermo transducers (capacitive) for HVAC monitoring

Snifter. User Manual. Distributor

Nitrogen Dioxide (NO2) Single-Point Gas Detection System

Hygro Transducer / Hygrothermal Transducer and CO 2 Measuring Probe for Climate Monitoring

Constant climate chambers

User Manual. GPD 3000 e. gas detector

Constant climate chambers

GMH 285 / GMH 285-BNC

gas monitoring system

TS400. Operating Manual. Test Station for Microtector II Series (G450/G460)

SCHMIDT LED Measured Value Display MD Instructions for Use

Operating Manual Hand-held Measuring Device GMH 3690 GL. for Gaseous Oxygen and Temperature with Alarm Function

Original operating instructions Safety switch with guard locking AC901S AC902S

ST Wiring diagram. Product description. Standard temperature controller. Order number

Duct and Rough Service Carbon Monoxide Sensor

Warranty Registration

Flameproof Manual Call Points

Oxygen (O2) Single-Point Gas Detection System

S304 and S305 Dust Emission Monitors. User Manual. Distributor

Operation Manual for Drying Storage Cabinet with Drying Unit U-5002

Constant climate chamber with humidity

Copying contents of this booklet or handing over to third party is strictly forbidden without written permission from Detector Oy.

Dixell Installing and Operating Instructions release 7.0

MSD Cabinets. Modular Dry Cabinet. Outstanding performance

Constant climate chamber with ICH-compliant illumination

Pressure measuring instruments

DEW-POINT SENSOR TRAINING. Moisture Sensors DP & Rh The basic low cost solution With a wide range

V A C U U M A N D P R E S S U R E T E C H N O L O G Y

INSTRUCTION MANUAL T119 1MN0110 REV. 0. operates with ISO9001 certified quality system

Service: to SLK Laboratory Hot Plates glass made of ideas

EasyLog Data Logger Series

EasyLog Data Logger Series

Entwicklung + Fertigung + Service: Heerweg 15 D, Denkendorf Tel.: / Fax.: /

XWA11V-KIT Walk-In Temp / Door /Alarm / Light Module with Mounting Box and Wiring

Dräger Polytron 5200 CAT Detection of flammable gases and vapours

R9900. Model. Instruction Manual. Indoor Air Quality Meter. reedinstruments. www. com

4-20mA CYBER Cyber Transmitter for flammable, toxic and IR gas detection Cyber Head Increased security in ATEX certified head

CMF, CMX & CMA series

Introduction Portable Meters Transmitter Probes RELATIVE HUMIDITY

Table of Contents SECTION PAGE SECTION 1 INTRODUCTION Description Features Models... SECTION 2 RS-232 COMMUNICATIONS...

testo 315-1/2 Instruction manual 99 Washington Street Melrose, MA Phone Toll Free

Danfoss gas detection unit Type GD Heavy Duty

METPOINT THE COMPLETE RANGE OF PRECISION CONTROL DEVICES FOR COMPRESSED AIR DEW POINT, FLOW, AND LEAKAGE

Table of Contents 1. OVERVIEW SYSTEM LAYOUT SPECIFICATIONS FUNCTION... 11

KM300 Carbon Monoxide Detection System Installation Manual

PVL1300 INSTRUCTIONS FOR USE

User Manual. Humidity-Temperature Chart Recorder. Model RH520

Advantages / Characteristics

Electronic Temperature Controller 70304,

User manual. Humimeter AW3 Meter

Sintrol Snifter A2 EX. Manual. Version 1.2.3

Refrigerated air dryers

Using the Model 1500 Portable Chilled Mirror Hygrometer

MAKING MODERN LIVING POSSIBLE INT INT EFIT 550. User Guide and Installation Manual. Danfoss Heating

EST User Manual. (Please read this manual carefully before using)

GMA200-MT. Operation Manual. Gas detection controller for mounting rail assembly

973 Dew Point Mirror. Portable Industrial Chilled Mirror Hygrometer. Typical applications: REFLECTING YOUR STANDARDS

Dräger Polytron 5700 IR Detection of flammable gases and vapours

WHA 3000P Hot Air Station

Measurement Solutions for Refrigeration Technology.

WHA 3000V Hot Air Station

Visit us at testo 835 Infrared and contact thermometer

HUM-M2 Humidity Sensor Module Description

Heated Gas Sampling Probe PSG Basic

It offers easy installation and the possibility for fast exchange of the sensor unit for service purposes.

GreenCon On/Off Room Thermostat

Technical Description. Laboratory Climate Test Chamber. System Weiss. Mat. No.: Type WKL 34/40

GreenCon On/Off Room Thermostat

GMA200-MW16. Operation Manual. Gas detection controller for wall mounting

Constant climate chambers

1 measurement signal input 4-20 ma for connection to an external gas sensor 5 freely programmable alarm switching points per measuring point

Transcription:

INSTRUCTION MANUAL FA 400 Humidity measuring instruments with display and alarm for measurement of the pressure dew point and the atmospheric dew point in different applications: - compressed air plants (refrigeration/adsorption driers) - granulate driers - medical gases - non-corrosive gases, e. g. nitrogen V4-02-2011

FUNCTIONS Page Introduction 2 Safety instructions 3 Description 3 Technical data 4 Diagram of instrument / Dimensions of instrument 5 Display operation 6-8 Calibration / Adjustment 9 Warranty 9 Ordering data 10 Notes 11 Contact 12 INTRODUCTION Dear customer, Thousands of customers buy our high standard products every year. There are a few good reasons for doing so: The cost-performance ratio - reliable quality at a fair price. We have the ideal solutions for your measuring tasks based on our expert experience gained over 20 years. Our high quality standard. Of course, our instruments carry the CE symbol required by the EU. We issue calibration certificates and hold seminars. Our service guarantees fast help. Measuring instrument conforms with DIN EN 61326-1 and 61010-1 2 V4-02-2011

SAFETY INSTRUCTIONS! Please read carefully before starting the device! Attention: Do not exceed a pressure range of > 50 bar with standard version. With special versions up to 350 bar. Observe measuring ranges of the sensor! The probes will be damages if they are overheated. Observe max. storage and transport temperature as well as max. operating temperature (e. g. protect measuring instrument from direct sunlight). The manufacturer cannot be held liable for any damage which occurs as a result of nonobservance or non-compliance with these instructions. Should the device be tampered with in any matter other than a procedure which is described and specified in the manual, the warranty is cancelled and the manufacturer is exempt from liability. The device is destined exclusively for the described application. CS Instruments GmbH offers no guarantee for the suitability for any other purpose and is not liable for errors which may have slipped into this operation manual. CS Instruments GmbH is also not liable for consequential damage resulting from the delivery, capability or use of this device. We offer you to take back the instruments of the instruments family FA 400 which you would like to dispose of. Adjustments and calibrations should only be carried out by qualified employees from the measurement and control technology branch. Important: Before installation briefly bleed the compressed air in order to remove condensate and particles. This prevents soiling of FA 400. Standing air leads to long measuring times. BESCHREIBUNG The FA 400 (from -80 bis 20 Ctd) is the ideal dew point meter with integrated display and alarm relay for refrigeration, membrane and adsorption driers. Special features: - -80 to 20 Ctd - Integrated display - Alarm adjustable via keyboar (60 V, 0.5 A) - Pressure-tight up to 350 bar (special version) - Extremely long-term stable langzeitstabil - Fast response time - 4...20 ma analogue output - 2 versions, for refrigeration and for adsortpion driers Programming via SFA Software. - Analogue output 4 20 ma scalable, switching for alarm relay adjustable - Calibration and ajustment including certificate - Switching between Ctd, % RH, g/m³ and so on - Read out of service data 3 V4-02-2011

Measuring range Type 0699.0401, FA 400 Type 0699.0402, FA 400 Type 0699.0403, FA 400 TECHNICAL DATA -80 50 C pressure dew point resp. dew point in Ctd 0 100 % RH -30 70 C -20 50 Ctd 4 20 ma -80 20 Ctd 4 20 ma -60 30 Ctd 4 20 ma Accuracy: typical ± 2 Ctd from -80-40 Ctd ± 1,5 Ctd from -40 0 Ctd ± 1 Ctd from 0 20 Ctd 3,0 2,5 Accuracy Genauigkeit FA 400 FA 410 Genauigkeit [ C] Accuracy ( C) 2,0 1,5 1,0 0,5 Pressure range: -1 50 bar standard, high pressure version up to 350 bar Power supply: 24 VDC smoothed (16...30 VDC/60 ma) Output: 4 20 ma 3-Draht-Technik Protection class: IP 65 EMV: DIN EN 61326 Operating temperature: 0 50 C Storage temperature: -40 80 C Burden for analogue output: < 500 Ohm Screw-in thread: G 1/2" stainless steel Material housing: Zinc alloy, PC, ABS Sensor protection: Connection: Response time t95: 0,0-80 -60-40 -20 0 20 Taupunkt [ C] Dew point ( C) Sintered filter 50 μm stainless steel M12, 5-pole < 30 seconds (dry) < 10 seconds (humid) Display: Alarm relay: 4-digit LCD max. 60 V, 0.5 A, NO relay opens in case of alarm and power failure, alarm value adjustable via keyboard, please see page 9 4 V4-02-2011

INSTALLATION DESCRIPTION Connector plug A (signal and supply) Connector plug B (alarm) FA 400 Pin 1 Pin 2 Pin 3 Pin 4 Pin 5 Connector plug A SDI -VB +VB +I NC 4...20 ma Connection cable A 0554.0104 (5 m) 0554.0105 (10 m) brown white blue black grey Connector plug B NC NC NC REL REL Connection cable B 0554.0106 (5 m) 0554.0107 (10 m) NC NC NC black grey SDI -VB Digital Signal (internal data transfer) Negative supply voltage 0 V +VB Positive supply voltage 16...30 VDC smoothed +I Positive 4...20 ma signal NC Not connected REL Relay output M12 Connector plug A/B FA 400: Relay NO (60 V, 0.5 A) Relay opens in case of alarm and power failure Connector plug A : If no connection cable ( 0553 0104, 0553 0105 ) is ordered the sensor will be supplied with a M12 connector plug. the user can connect the supply and signal cables as indicated in the connection diagram. Connector plug B : Connector plug 5 V4-02-2011

INSTALLATION Please note : Advantage : CS Instruments recommends the indirect installation with measuring chamber Easy mounting and dismounting of the probe without interruption of the line. Quick response time due to quick coupling. Optimum sensor protection. Indirectly in the compressed air system Connect probe with measuring chamber to the compessed air pipe by means of a quick coupling. In case of compressed air containing oil and dirt particles a pre -filter should be installed in front of the measuring chamber. Compressed air flows continuously (at 7 bar approx. 1 l/min expanded) in the capillary pipe of the measuring chamber. The reaction times for the humidity reading are shorter than in case of a direct mounting. Directly in the compressed air system Screw in probe with G 1/2" thread pressure-tight in the center or at the top of the compressed air pipe. Take care that measurement is effected close to the compressed air flow. U-bend pipes or non-flowing compressed air result in very slow reaction times for the moisture reading. Measurable gases: In general, humidity can be measured in all non-corrosive gases. In case of measurements incorrosive gases please consult CS Instruments GmbH. 6 V4-02-2011

DISPLAYFUNCTION Display FA 400 can display up to 3 measured values. This are temperature, rel. humidity and dew point. For better reading only one measured value is displayed. The measured values are toggled during normal operation mode every 3 seconds. The following changes an be made: - Alarm value - Display of measured value (temperature, rel. humidity and dew point) - Display contrast - Display upside down Settings ex work: - Dew point in Ctd Menu buttons On the top of FA 400 are the capacitive key buttons to operate the display menu. SPEZIAL INDICATIONS FA 400 is configured to be connected to DS 300. In this mode, alarm function in FA 400 is disabled and will be handled by DS 300 Indicates that an alarm is reached. The relay will be switched and the measurement value is blinking 7 V4-02-2011

Connection FA 400 DISPLAYFUNCTION After power on, the display will go through an initialisation procedure and will finally show the actual online value (ex work dew point). Via the configuration menu up to 3 measuring values can be configured for online display. FA 400 will toggle between the measuring values every three seconds. Display current set alarm value A short click on the Enter -button shows the current set alarm value. Changing of the alarm value see next page. Slave mode Changing the settings is only possible with DS 300 Master mode Changing the settings with pusbuttons FA 400 8 V4-02-2011

DISPLAYFUNCTION Alarm setting In order to change the alarm value and the configuration, keep the Enter button pressed for 3 seconds. After input of the unlock code the menu begins with alarm setting. The first digit will start blinking and can be changed with the Up key. Every digit has to be confirmed with the Enter button. Menu You can exit the configuration process by pressing and keeping the Enter key for 3 sec or no key event for 20sec. The configuration before this picture will be save and effected late. 9 V4-02-2011

CALIBRATION/ADJUSTMENT According to DIN ISO certification of the measuring instruments we recommend to calibrate and if applicable to adjust the instruments regularly from the manufacturer. The calibration intervals should comply with your internal specificaton. According to DIN ISO we recommend a calibration interval of one year for the instrument FA 400. WARRANTY If you have reason for complaint we will of course repair any faults free of charge if it can be proven that they are manufacturing faults. The fault should be reported immediately after it has been found and within the warranty time guaranteed by us. Excluded from this warranty is damage caused by improper use and non adherence to the instruction manual. The warranty is also cancelled once the instrument has been opened - as far as this has not been mentioned in the instruction manual for maintenance purposes - or if the serial number in the instrument has been changed, damaged or removed. The warranty time for the FA 400 is 12 months. If no other definitions are given the accessory parts have a warranty time of 6 months. Warranty services do not extend the warranty time. If in addition to the warranty service necessary repairs, adjustments or similar are carried out the warranty services are free of charge but there is a charge for other services such as transport and packaging costs. Other claims, especially those for damage occurring outside the instrument, are not included unless responsibility is legally binding. After sales service after the warranty time has elapsed We are of course there for you even after the warranty time has elapsed. In case of malfunctions please send us the instrument with a short-form description of the fault. Please do not forget to indicate your telephone number so that we can call you in case of any questions. 10 V4-02-2011

ORDERING DATA Order no. Description 0699.0401 FA 400 dew point sensor for refrigeration driers, -20...50 Ctd 0699.0402 FA 400 dew point sensor for adsorption driers, -80...20 Ctd 0699.0403 FA 400 dew point sensor, -60...30 Ctd 0533 0112 Cable FA 200 to FA 400 0553.0104 Connection cable, length 5 m 0553.0105 Connection cable, length 10 m 0553.0106 Alarm cable, length 5 m 0553.0107 Alarm cable, length 10 m 0699.3390 Standard measuring chamber for compressed air up to 16 bar 0699.3590 High pressure measuring chamber up to 350 bar * 0699.3690 Measuring chamber for atmospheric dew point 0699.3790 Measuring chamber for respiratory air bottles up to 350 bar 0699.4004 Special scaling, output in ppm, V/V, % RH, mg/m³ 0699.3495 Mains unit in wall housing 230 VAC/24 VDC with alarm processing 0554.0002 Control and calibration set 11.3 % RH 0554.0004 Control and calibration set 33 % RH 0554.0005 Control and calibration set 75.3 % RH 0554.2005 CS Service Software for VA/FA 400 sensors incl. PC connection set, USB connection and interface adapter to the sensor as well as CS Soft Professional incl. USB-cable A-A 3200.0003 Precision calibration at 0 Ctd and 10 Ctd incl. ISO certificate 0699.3396 Precision calibration at -40 Ctd incl. ISO certificate 0699.4003 * Special version FA 400 for 350 bar 11 V4-02-2011

DECLARATION OF CONFORMITY 12 V4-02-2011

13 V4-02-2011 NOTES

14 V4-02-2011 NOTES

15 V4-02-2011 NOTES

CONTACT Sales Office SOUTH Zindelsteiner Str. 15 D-78052 Villingen-Schwenningen Phone +49 (0) 7705 97 89 9-0 Fax +49 (0) 7705 97 89 9-20 info@cs-instruments.com www.cs-instruments.com Sales office NORTH Am Oxer 28c D-24955 Harrislee Phone +49 (0) 461 700 20 25 Fax +49 (0) 461 700 20 26 info@cs-instruments.com www.cs-instruments.com 16 V4-02-2011