Regulatory Information and Specifications

Similar documents
Performer Cuspidor. Installation Instructions. (For Performer III Chair & International Performer I Chair) You Will Need. About This Install

MaxiCompressor. Limited Warranty. High Performance 50 PSI Compressor 501-S

InstructIons for use A-dec 561 Cuspidor

Instructions for Use. A-dec 461 Cuspidor with Support Center

Instructions for use / Alarm unit, Fiber optic cable & Sensor patch ENGLISH

CARESTREAM DRX-1 System Detector CARESTREAM DRX-1C System Detector CARESTREAM DRX 2530C Detector Model DRX User Guide

Instructions for Use A-dec Self-Contained Water System

Veterinary Patient Warming System Controller User Manual

Enhance DX. Mattress Replacement System. User Manual

INTERNATIONAL STANDARD

INSTALLATION GUIDE. AXIS A1001 Network Door Controller

CARESTREAM DRX-1 System

A-dec 300 Support Systems

SAVE THESE INSTRUCTIONS

Classic Twelve O clock Cabinet Installation Instructions

5. Instrument and accessories discharging must be done following current law regulations in every country of use.

ASPIRE Laboratory Aspirator

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005

Veterinary Patient Warming System Service Manual. For information on operating the Hot Dog Patient Warming System, refer to the Instructions for Use

IEC60601_1K. Medical electrical equipment Part 1: General requirements for basic safety and essential performance

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005

LED CEILING LIGHT WITH MOTION SENSOR AND REMOTE. ITM. / ART Model: LM56123 CARE & USE INSTRUCTIONS

Certificate of Compliance

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005

safety & regulatory information

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005

Instruction Manual M6627 / Version 7.4. Slide Drying Bench MH6616 MH6616X1

PEAK Platelet Rich Plasma System

Certificate of Compliance

HP Power Distribution Unit with Power Monitoring (Models S124 and S132) User Guide

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005

CERTIFICATE OF ACCREDITATION

CL1000 Ceiling Light OWNER S MANUAL. Marus Fruehauf Drive Charlotte, NC USA

AUGUST 2016 CONTENTS:

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005. F2 LABS Peters Road Middlefield, OH Wendy Fuster Phone: Ext.

REVISION HISTORY Rev Description Author. Approved by Date A Initial release per ECO Dean King Dean King November 16, 2012

Transcend Heated Humidifier User Manual

UL TEST REPORT AND PROCEDURE

Operator s Manual. WarmTouch. Model WT-5300A Patient Warming System

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2017

Installation Instructions Outdoor Refrigerator

Installation Instructions Remote Blowers

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005

USER MANUAL SAVE THESE INSTRUCTIONS

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005

Operation and Maintenance Manual

Instruction Manual. CAT 100 & CAT 200 Continuous Analyzer Transmitter Addendum for Hazardous Area Applications. 3 rd Edition 05/2007

OPERATORS MANUAL FOR Mi-T-M CORONA DISCHARGE OZONE GENERATORS

General safety precautions English

ANSI-ASQ National Accreditation Board/ACLASS

General safety precautions English

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005

GETTING STARTED GUIDE NI TC, ±78 mv, 24 Bit, 75 S/s Aggregate

2364 Leicester Road, P.O. Box 175, Leicester, NY Phone (585) Fax (585)

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005

Power Conditioners User s Manual. For use with Standard Power Conditioners, Medical Power Conditioners and Ground Guard Power Conditioners

IEC60601_1J. Medical electrical equipment Part 1: General requirements for basic safety and essential performance

LC3000 Level Controller Installation and Maintenance Instructions

Installation and Operation Manual. 445 Procedures Cart. Page 3. Page 5. Page 8. Page 10. Important Information Page 2. Installation.

Test Report. Appliance EMF Testing

CLASSIC. SIDE CABINET Installation Instructions

Dry Bath Heating Systems

Schedule of Accreditation issued by United Kingdom Accreditation Service 2 Pine Trees, Chertsey Lane, Staines-upon-Thames, TW18 3HR, UK

Thermo-Sealer. Manual Heat Sealer. User Manual

SINAMICS G130. Cabinet design and EMC. Operating instructions 03/2011 SINAMICS

Cascade Comfort Bathing Systems with Aqua-Aire Safe Operation & Daily Maintenance Instructions

Installation Instructions Built-In Wine Coolers, Beverage Centers

Operator s Manual. WarmTouch. Model WT-5900 Patient Warming System

Declaration of Conformity

Approvals, Safety Instructions, and Statutory Information

General safety precautions English

ULTRASONIC AIR SALINIZER. Owner s Manual. Saltair Home Salt Therapy

CLETEF-786. Infrared Thermometer Instruction Manual

Agilent N4011A MIMO/Multi-port Adapter

Table of Contents. English

Certificate of Compliance

GETTING STARTED GUIDE NI RTD, 0 Ω to 400 Ω, 24 Bit, 400 S/s Aggregate, PT100

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005

A m e r i c a s D e n t a l E q u i p m e n t C o m p a n y. Rev 2/15. D a n s e r e a u

CONTENTS IMPORTANT SAFEGUARDS INTRODUCTION PRODUCT DESCRIPTION INSTALLATION OPERATIONS CLEANING...

Certificate of Compliance

CERTIFICATE OF ACCREDITATION

Recent Topics in IEC TC 62 (Electrical Equipment in Medical Practice) and its Subcommittees

Titan Series Blanket Warmer

ANNEALING FURNACE MODEL 414 User Maintenance Manual/Handbook

COOLFLOW IRRIGATION PUMP USER MANUAL

Companion 5. Oxygen Concentrator PROVIDER TECHNICAL MANUAL

General safety precautions English

General safety precautions English

General safety precautions English

Operator s Manual. Model G32-S Model G32-E Disinfection Soak Stations

Installation Instructions Induction Cooktop

CPS V Cabinet Power System

Square D Clipsal Outdoor Light-Level Sensor

FV-11VQL6. Thank you for purchasing this Panasonic product Back cover FV-15VQL6

Cisco TelePresence MCU 4200 Series safety and compliance information

Transcription:

Regulatory Information and Specifications Introduction This document contains the required regulatory information and specifications for A-dec products. Products Requiring Agency Information Specific regulatory product information must accompany some products. The types of equipment that require this information include: Dental chair Delivery system and assistant s instrumentation Cuspidor Dental light Lab equipment Monitor Mounts Equipment Alterations Policy Certain modifications or alterations of A-dec equipment which expand the use of A-dec equipment beyond its design and intent, or which override any safety features of A-dec equipment may jeopardize doctor, patient, or staff safety. Field modifications that alter the electrical and/or mechanical safety of A-dec dental devices are in conflict with Underwriters Laboratory (UL) construction file requirements and are not sanctioned by A-dec. Examples of field modifications that diminish safety design include, but are not limited to: rendering access to the line voltage without the use of tools, modification of supporting elements that increase or shift loading characteristics, and the addition of any powered device that exceeds the design limits of the dental system. The use of accessory equipment not complying with the equivalent safety requirements of A-dec equipment may lead to a reduced level of safety of the resulting system. It is the responsibility of the equipment distributor and the installer to assure that the installation complies with all building code requirements. The responsibility to determine whether a modification or alteration of A-dec equipment falls within these constraints is with the person(s) who initiates, approves and/or performs such modification or alteration. A-dec will not respond to inquiries on an individual basis. This person(s) will be deemed to have assumed all associated risks with such alteration or modification and will hold A-dec harmless from resulting claims, including product liability claims. Additionally, such modification or alteration voids A-dec s warranty and may invalidate UL or other regulatory agency approval. 86.0221.00 Rev G

Regulatory Information and Specifications Serial Number When you inquire about service, please provide the serial number, which is the S/N code on the product serial tag. Figure 1. Serial Number Label Examples Model Number Products Manufactured before June 2011 Serial Number Products Manufactured June 2011 and after arcode and Software The first letter of the serial number indicates the month the product was manufactured. For products manufactured before June 2011, the first digit of the serial number indicates the year of manufacture (for example, L0 = December 2010). For products manufactured June 2011 and after, the first two digits indicate the year of manufacture (for example, 11F = June 2011). Table 1. Key to Month of Manufacture Letter Month Letter Month A January G July February H August C March I September D April J October E May K November F June L December When calling A-dec Customer Service about a circuit board issue, please have the assembly revision available. The assembly revision is located on the barcode label on each circuit board that contains software. Figure 2. Example of Circuit oard arcode Label oard Type Serial Number Assembly Revision Date Software is not field upgradeable. Consult A-dec for questions about compatibility, upgradability, or software revision (which is derived from the assembly revision shown on the barcode label). NOTE The software revision of the A-dec 311 chair circuit board is 1.nnnn where nnnn denotes the number (0000 through 9999). 2 86.0221.00 Rev G

Safety Considerations for Accessory Equipment Universal Cautions The use of accessory equipment not complying with the equivalent safety requirements of this equipment may lead to a reduced level of safety of the resulting system. Consideration relating to the use of accessory equipment shall include evidence that safety certification of the accessory equipment has been performed in accordance to the IEC 60601-1 standard along with any national deviations. Low voltage communication cables (US, Ethernet, etc) either supplied by A-dec or installed in the field, shall be routed away from single insulated or non-insulated mains voltage (100-240 VAC). Electrical connections to the A-dec Dental System are not permitted unless the combination of the accessory and the A-dec Dental System has been evaluated to the IEC 60601-1 standard along with any national deviations. All configurations shall comply with the system standard IEC 60601-1-1 or IEC 60601-1:2005. Anyone connecting additional equipment to the signal input part or signal output part is configuring a medical system, and is, therefore, responsible to ensure that the system complies with the requirements of the Medical Electrical Systems standard IEC 60601-1-1 or IEC 60601-1:2005. If you have questions or concerns, contact A-dec Customer Service or your local authorized A-dec dealer or distributor. CAUTION Local regulation requires licensed plumbers and electricians to install utilities. All plumbing and utilities must conform to prevailing local codes. CAUTION The manner and method for accessing utilities within the wall is the responsibility of the dental dealer, architectural services, and/or contractors. Utilities must be accessible without the use of tools. Preventative Inspection Over time, wear and tear can affect the performance of the equipment. Periodically inspect the water and air lines for any visible cracking or cuts in the tubing which can lead to leaks. Inspect the fittings and check for any loose screws. To prevent problems from occurring, replace the tubing and tighten the screws and fittings as necessary. Transporting the Dental System When transporting the dental system: Chair base should be fully down, and the chair back should be fully up. Chair body should be secured to the chair baseplate. Delivery system should be over the seat. Upholstery should be detached, and the light and the upholstery should be centered and secured above the chair. Delivery system and light should be secured to prevent movement. Dental system should be secured to the transporting vehicle. 86.0221.00 Rev G 3

Regulatory Information and Specifications Identification of Symbols These symbols appear on the actual product where appropriate. These same symbols are used in documentation to alert the user/reader of cautions, warnings, hazards, or tips. Symbol Description Recognized by Underwriters Laboratories Inc. with respect to electric shock, fire and mechanical hazards only in accordance with UL 60601-1 (2601-1) and CAN/CSA C22.2, No. 601.1 and/or ANSI/AAMI ES60601-1 and CAN/CSA C22.2 No. 60601-1. Classified by Underwriters Laboratories Inc. with respect to electric shock, fire and mechanical hazards only in accordance with UL 60601-1 (2601-1) and CAN/CSA C22.2, No. 601.1 and/or ANSI/AAMI ES60601-1 and CAN/CSA C22.2 No. 60601-1. UL listed to UL 61010A-1, S EN 61010-2-010 and Canadian (CAN/CSA C22.2, No. 1010.1-92) safety standards. Refer to accompanying documents. Conforms to applicable European Directives (refer to Declaration of Conformity). Conforms to MDD 93/42/EEC for Class IIa devices. 0843 Protective earth (ground). Functional earth (ground). Type applied part. Class II equipment. Caution: Metal surfaces can be hot during and following the dry cycle. General mandatory action sign. Important to follow instructions. Not a caution. i.e., NOTE: Assemble parts as shown. Caution. Failure to follow instructions could result in damage to product or minor injury. i.e., CAUTION: Do not over tighten the adjustment screw. Over tightening could break the screw. iohazard. Potential infection if instructions are not properly followed. i.e., IOHAZARD: iohazard from dental equipment can result in life-threatening diseases or infection. DANGER. Failure to turn off the power before you begin this procedure can lead to electrical shock. i.e., DANGER: Disconnect the main power or shut off the main power before servicing. Failure to turn off the power before you begin this procedure can lead to electrical shock. DANGER: Line voltage. Only a licensed electrician should remove the cover. WARNING. Failure to turn off the power before you begin this procedure can lead to product damage and result in serious injury or death. i.e., WARNING: Turn off the power before removing the pump cover. Failure to turn off the power before you begin this procedure can lead to product damage and result in serious injury or death. 4 86.0221.00 Rev G

Environmental Specifications Temperature/Humidity Specification Storage/Transportation Temperature -29 C to 50 C (-20 F to 122 F) - Relative humidity: 80% for up to 31 C, decreasing linearly to 50% at 40 C. Operating Temperature 10 C to 40 C (40 F to 104 F) - Relative humidity: 80% for up to 31 C, decreasing linearly to 50% at 40 C. Indoor Use Altitude up to 2,000 m (6,563'), installation category II, pollution degree 2. Classification of Equipment (IEC-60601-1) Type/Mode Types of Shock Protection Degree of Shock Protection Degree of Protection Against Water Ingress Mode of Operation Flammable Gasses Classification CLASS I EQUIPMENT: Dental chairs, dental lights, power supplies, and monitor mounts. CLASS II EQUIPMENT: Chair, wall, and cart-mounted delivery systems. TYPE APPLIED PART: Delivery systems only. ORDINARY EQUIPMENT: All products. CONTINUOUS OPERATION: All models except dental chairs. CONTINUOUS OPERATION WITH INTERMITTENT LOADING: Dental chairs - 5% duty cycle. Not suitable for use in the presence of a flammable anesthetic mixture with air, oxygen, or nitrous oxide, where such gasses may accumulate in concentration (closed space). Classification of Equipment (IEC-61010-1) Type/Mode Types of Shock Protection Classification CLASS I EQUIPMENT: (Earthed) Simulator, Preference ICC, ICV, lab control. NOTE Allowable mains voltage fluctuations ± 10% of rated voltage. Electrical Rating A-dec Product Frequency (Hz) Voltage Range (VAC) Maximum Current (Amps) Dental Chairs A-dec 200, A-dec 311, A-dec 511, 1040, 1221, & 8000 50-60 100/110-120/220-240 10/10/10 Delivery Systems, Assistant s Instrumentation, and Cuspidors Systems with 300W Power Supply 50-60 100/110-120/220-240 3.1/2.8/1.4 Dental Lights (Mains Voltage) Ceiling, Post, Wall, Single Track, & Preference Mount Lights 50-60 100/110-120/220-240 3.1/2.8/1.4 Dual Track Light 50-60 110-120/220-240 5.6/2.8 86.0221.00 Rev G 5

Regulatory Information and Specifications Electrical Rating (continued) A-dec Product Frequency (Hz) Voltage Range (VAC) Maximum Current (Amps) Dental Lights (Chair-Mounted) Performer 8000 50-60 10.8/12.1 4.4 Cascade 6300 50-60 12.1/16/17 5.5 A-dec 571/572 50-60 12.1/16/17 5.5 A-dec 371/372 50-60 12.1/17 5.5 A-dec 200 50/60 12.1/17 5.5 Power Supplies 25W Power Supply 50-60 100/110-120/220-240 0.3/0.3/0.15 80W Power Supply 50-60 100/110-120/220-240 0.9/0.8/0.4 300W Power Supply 50-60 100/110-120/220-240 3.1/2.8/1.4 Dental Furniture Preference Collection 50-60 110-120 20 Preference ICC 50-60 110-120 20 ICV 50-60 110-120/220-240 0.5/0.5 Miscellaneous Simulator 41L, 42L, 4810, 4820, M66AA 50-60 100/110-120/220-240 10/10/5 Lab Control with 300W Power Supply 50-60 100/110-120/220-240 3.1/2.8/1.4 itewing X-Ray Viewer 50-60 24 0.5 International Junction ox 50-60 100-240 15 Monitor Mounts 50-60 100-240 10 NOTE For products that are permanently connected to fixed wiring (no power cord plug), a switch or circuit breaker shall be used to disconnect the product from mains power. Mains connections shall be made by qualified personnel in compliance with local building and electrical codes. NOTE Countries using a mains plug other than the North American plug (such as Australia, Denmark, etc.) shall use a plug that is rated appropriately for the voltage and current of the product. Electromagnetic Compatibility This equipment has been tested and found to comply with the limits for medical devices in IEC 60601-1-2. These limits are designed to provide reasonable protection against harmful interference in a typical medical installation. In the event of interference, power devices from separate mains supplies and/or increase physical distance between devices. Contact A-dec Customer Service if you have any questions. 6 86.0221.00 Rev G

Electromagnetic Emissions Emissions Test Compliance Electromagnetic Environment - Guidance RF Emissions CISPR 11 Harmonic emissions IEC 61000-3-2 Voltage Fluctuations/ Flicker Emissions IEC 61000-3-3 Class A Class A Complies The dental equipment is suitable for use in all establishments other than domestic, and may be used in domestic establishments and those directly connected to the public low-voltage power supply network that supplies buildings used for domestic purposes, provided the following caution is heeded (see CAUTION below). CAUTION This equipment/system is intended for use by health care professionals only. The equipment/system may cause radio interference or may disrupt the operation of nearby equipment. It may be necessary to take mitigation measures, such as re-orienting or relocating the dental equipment or shielding the location. Electromagnetic Immunity Immunity Test Electrostatic Discharge (ESD) IEC 61000-4-2 Electrical Fast Transient/urst IEC 61000-4-4 Surge IEC 61000-4-5 Voltage Dips, Short Interruptions and Voltage Variations on Power Supply Input Lines IEC61000-4-11 Power Frequency (50-60 Hz) Magnetic Field IEC 61000-4-8 IEC 60601 Test Level ±6 kv contact ±8 kv air ±2 kv for power supply lines ±1 kv for input/output lines ±1 kv line(s) to line(s) 2± kv line(s) to earth <5% U T (>95% dip in U T ) for 0.5 cycle 40% U T (60% dip in U T ) for 5 cycles 70% U T (30% dip in U T ) for 25 cycles <5% U T (>95% dip in U T ) for 5 sec Compliance Level C Electromagnetic Environment Guidance Floors should be wood, concrete, or ceramic tile. If floors are covered with synthetic material, the relative humidity should be at least 30%. Mains power quality should be that of a typical commercial or hospital environment. Mains power quality should be that of a typical commercial or hospital environment. Mains power quality should be that of a typical commercial or hospital environment. If the user of the dental equipment requires continued operation during power mains interruptions, it is recommended that the dental equipment be powered from an uninterruptable power supply or a battery. 3 A/m A Power frequency magnetic fields should be at levels characteristic of a typical commercial or hospital environment. NOTE U T is the AC mains voltage prior to application of the test level. 86.0221.00 Rev G 7

Regulatory Information and Specifications Decommissioning and Disposal of A-dec Equipment Contact Information A-dec dental equipment removed from service should be decommissioned in accordance with local regulatory requirements. Circuit boards and electrical cabling should be recycled as electrical salvage. Aluminum and steel components should be recycled as metal salvage. Molded plastic components include mold marks indicating the type of plastic and should be recycled accordingly. The cuspidor, waste lines from the cuspidor, and extraction lines should be treated as biologically contaminated materials and handled with appropriate precautions during dismantling. Any material unsuitable for recycling should be disposed of appropriately. For specific questions regarding material type, please contact A-dec Customer Service. For decommissioning information on associated equipment from other manufacturers, refer to the documentation from the manufacturer. If you have a question that is not addressed in this document, please contact A-dec Customer Service at one of the following phone numbers: 1.800.547.1883 (within USA and Canada) 1.503.538.7478 (outside USA and Canada) Customer service is available Monday through Friday, from 5 a.m. to 5 p.m. Pacific Standard Time (PST). A-dec Headquarters 2601 Crestview Drive Newberg, OR 97132 USA Tel: 1.800.547.1883 Within USA/Canada Tel: 1.503.538.7478 Outside USA/Canada Fax: 1.503.538.0276 www.a-dec.com / www.a-dec.biz International Distribution Centers A-dec United Kingdom EU Authorized Representative Austin House, 11 Liberty Way Nuneaton, Warwickshire CV11 6RZ England Tel: 0800 ADECUK (233285) Within UK Tel: +44 (0) 24 7635 0901 Outside UK www.a-dec.co.uk A-dec Australia Unit 8 5-9 Ricketty Street Mascot, NSW 2020 Australia Tel: 1.800.225.010 Within Australia Tel: +61 (0)2 8332 4000 Outside Australia www.a-dec.com.au A-dec Inc. makes no warranty of any kind with regard to the content in this document including, but not limited to, the implied warranties of merchantability and fitness for a particular purpose. ÍvÈ.Ç"5È.005Î 86.0221.00 Rev G Copyright 2011 A-dec Inc. All rights reserved.