MST MODEL S1 AIRLINE "CO" MONITORING AND LOW PRESSURE WARNING SYSTEM MANUAL

Similar documents
MST, INC. MODEL 2002 CARBON MONOXIDE MONITOR OWNER S MANUAL

MST, INC. MODEL 2002 CARBON MONOXIDE MONITOR (AMBIENT-TYPE) OWNER S MANUAL

MST, INC. MODEL 2002 OXYGEN MONITOR (AMBIENT-TYPE) OWNER S MANUAL

INSTALLATION. and INSTRUCTION MANUAL. for QUALITY AIR BREATHING SYSTEMS. Model ABM - 700

INSTALLATION. and INSTRUCTION MANUAL. for QUALITY AIR BREATHING SYSTEMS. Model ABM - 715

Carbon Monoxide Monitor and Retrofit Carbon Monoxide Monitor Kit W-2808/37027

OPERATING INSTRUCTIONS AND REPLACEMENT PARTS

MODEL A-316 CARBON MONOXIDE (CO) MONITOR AND ALARM FOR COMPRESSED AIR TESTING

Carbon Dioxide Sample-Draw Detector Operator s Manual

RK-02 Multi Point Detector Operator s Manual

Models: CO-91IS and CO-91ISLA

Operating Manual. Models BB30-COAA, BB50-COAA & BB100-COAA Manual No. AABOX011 (Rev 2 June 2005)

RK/ RK Carbon Monoxide Detector Operator s Manual

RK CO 2 Transmitter Operator s Manual

RK-05 Carbon Monoxide Detector Operator s Manual

Operating Manual Manual de instrucciones Manuel d utilisation

RK/ RK Oxygen Detector Operator s Manual

MODEL 5800A CARBON MONOXIDE (CO) MONITOR AND ALARM SAMPLE-DRAWING TYPE

RK / RK CO 2 Detector Operator s Manual

RAM 50 Series Operation Manual

RK Multi Point Detector Operator s Manual

RK Carbon Monoxide Transmitter Operator s Manual

RK/ RK Oxygen Detector Operator s Manual

Carbon Monoxide Sample-Draw Detector Operator s Manual

QAS-101 SERIES SINGLE CHANNEL CONTROLLERS

RK Hydrogen Sulfide Transmitter Operator s Manual

Combustible Gas Sample-Draw Detector Operator s Manual

RK-05 Carbon Monoxide Transmitter Operator s Manual

RK/ RK Combustible Gas Detector Operator s Manual

RK-OXY Oxygen Sample-Draw Detector Operator s Manual Part Number: RK Revision: P1 Released: 7/18/02

RK-05 Combustible Gas Transmitter Operator s Manual

Chargebuster Ion Gun Installation, Operation and Maintenance

RK Carbon Monoxide Transmitter Operator s Manual

RK / RK Methane Detector Operator s Manual

RK Carbon Monoxide Transmitter Operator s Manual

RAM 50 Series Operation Manual

Chargebuster Ion Gun Installation, Operation and Maintenance

Warranty Registration

Rev. None AUGUST 2012 MDM SERIES MODULAR MEMBRANE AIR DRYERS OPERATOR MANUAL MDM1 MDM4 MDM9 MDM13 MDM24 MDM49 MDM66 MDM106 MDM138

CT-7 Series Toxic Gas Transmitter Operator s Manual

GG-CO-EXP CARBON MONOXIDE SENSOR. Installation and Operation Manual

RK-03 Hydrogen Detector Operator s Manual

H-10PM. Refrigerant Leak Detector

QAS SERIES PLUG-IN CONTROLLERS c/w SOLID STATE SENSOR

Read all instructions & warnings before using this product. Keep this manual for future reference. Warnings. Parts & Accessories

Single Gas Monitor 777

GAS MONITOR Model Model

Combustible Gas/Carbon Monoxide Sample-Draw Detector Operator s Manual

RK Combustible Gas Transmitter Operator s Manual

Combustible Gas/Hydrogen Sulfide Sample-Draw Detector Operator s Manual

1040 Gas Monitor INSTALLATION AND OPERATING INSTRUCTIONS AMC-1040 WITH INTEGRAL ELECTROCHEMICAL SENSOR

Sentry LIQUID LEVEL GAUGE MODEL 200 or 200C OWNER MANUAL REV 1.7 SEPT08 PAGE 1 OF 12

TG-Series. Installation & Operations Manual. Model: TG-75, TG-150, TG-300, TG-450, TG-600. Water Cooled Ozone Generator

OWNERS MANUAL MODEL 451 (-101 & -201)

esense CO2 Transmitter and Alarm

LED LAMP: To turn on the LED Lamp, press the LIGHT BUTTON located on the front. Press again to turn off.

Carbon Monoxide (CO) Detecting Ventilation Fan Controller Model 120VC Single Relay (100/25 PPM) (200/35 PPM)

Tox-COMB/ANA Combustible Gas Transmitter Operator s Manual

RK-CH4-4 Combustible Gas Transmitter Operator s Manual

ANALOX 101 D2 PORTABLE OXYGEN MONITOR USER MANUAL

Bench Top Ionizer Installation, Operation and Maintenance

OPEARATING GUIDELINES FOR AERO-252 (Pages 1 to 3 contain general information on CO Guardian. Pages 4 to 6 contain information specific to Aero 252)

Controllers. Instruction Manual WARNING

Reacti-Therm I and III Heating Modules

! "! # Please read these installation and operating instructions completely and carefully before starting.

Macurco Single-Gas XL Series Monitor, CM-1XL Carbon Monoxide (CO), HS-1XL Hydrogen Sulfide (H2S) User Instructions

Integra Family. Operations Manual. Models: Integra TX SS, Integra HD, Integra DS, Integra DS LA, Integra UHS, Machine Name: Serial Number:

GASGUARD NH 3. Ammonia Sensor OPERATING & INSTALLATION MANUAL

MODEL 7000 SUCTION UNIT

AMC-CD2. Human Occupancy Detector INSTRUCTION MANUAL. Please read these instructions completely and carefully before starting.

ANALOX 5001 Carbon Dioxide Monitor. User Manual ANALOX Analox 5001 Carbon Dioxide Monitor User Manual

3 Macurco Carbon Monoxide Detector CM-16A

Guardian CR2000 Overhead Ionizing Air Blower. User s Manual

GG-NO2 NITROGEN DIOXIDE GAS SENSOR. Installation and Operation Manual

GASGUARD SO2 Sulfur Dioxide Sensor OPERATING & INSTALLATION MANUAL

Installation & Service Guide

GG-NH3 AMMONIA GAS SENSOR. Installation and Operation Manual

PORTER INSTRUMENT COMPANY, INC. 245 TOWNSHIP LINE RD. P.O. BOX 907 HATFIELD, PA USA [215] / FAX [215]

Static Eliminators. Overhead Ionizing Blower Model IN5130. Instruction Manual

Instruction Sheet LTR - Low Water Cutoff for Hot Water Boilers

ISA-RAL-M ISA-RAH-M Operation and Maintenance Manual

Model 1104 Troubleshooting Guide

Chargebuster Ion Gun Installation, Operation and Maintenance

RAM3 Remote Alarm Module

CA5000 Liquid CO 2 Freezer Backup System

USER S MANUAL BUCKET FAN SERIES BUCKET FAN WHISPER SERIES

Air Dryers & After Coolers Dry Land

Because you re not like everyone else. MISTING FAN MF24B OWNER S MANUAL

For technical support call:

BST-MG07 Multi-gas Detecting Alarm Manual Instruction

TG-10 Wall Mount Installation and Operation Manual. High Output Ozone Generator

LABORATORY ULTRA ZERO AIR GENERATOR. ZAC-ULT Series USER MANUAL

DRY AIR SYSTEMS, INC Metro Boulevard Maryland Heights, Missouri (314) fax (314)

Instruction Manual WARNING

WHOLE HOUSE FILTER &

Technical Data. Installation

GASGUARD Cl 2. Chlorine Sensor OPERATING & INSTALLATION MANUAL

NATURAL GAS DETECTOR. Confined Space Monitor. Leak Investigation. Line Purge Testing. Bar Hole Sampling

Operation Manual. Rev. E GEM-MRI. Oxygen Sampling System.

QUALITY ASSURANCE/QUALITY CONTROL DOCUMENTATION SERIES OPTEC NGN-2 NEPHELOMETER AUDIT PROCEDURES (IMPROVE PROTOCOL) STANDARD OPERATING PROCEDURE

Transcription:

MST MODEL 5700 - S1 AIRLINE "CO" MONITORING AND LOW PRESSURE WARNING SYSTEM MANUAL WARNING: Do not attempt to operate this equipment without first reading and understanding the service manual enclosed with this device. 06/09

2 CONTENTS SPECIFICATIONS...3 GENERAL DESCRIPTION...4 GENERAL OPERATIONS...4 MST'S AIRLINE "CO" SYSTEM INITIAL INSTALLATION AND START-UP...7 MST'S AIRLINE "CO" SYSTEM GENERAL OPERATION AND MAINTENANCE...9 RECORD KEEPING...10 SERVICE RECORD...11

3 SPECIFICATIONS MST AIRLINE "CO" MONITORING SYSTEM MST MODEL 5700 - S1 SIZE: WEIGHT: OPERATING PRESSURE (MAX.): SENSOR TYPE: RANGE: ACCURACY: 16"W X 16"H X ½" THICK (406mm X 406mm X 13mm) 6.5 LBS (2.9 KG) 100 PSI (6.9 BAR) ELECTROCHEMICAL 0-199 PPM CO +/-5% OF READING ALARM-TYPE: MONITOR PIEZOELECTRIC - 85 db(a) AT (1) FT. (8008403 REMOTE ALARM 119 db (A) AT (10) FT.) ALARM LEVEL SETTING: LOW PRESSURE: USER ADJUSTABLE FACTORY SET AT 20 PSI MONITOR FACTORY SETTING: 10 PPM (US) 5 PPM (CANADA) AMBIENT OPERATING TEMPERATURE RANGE: 32-104 o F (0-40 o C) OTHER FEATURES ADJUSTABLE LOW PRESSURE WARNING SYSTEM BETWEEN 10 AND 100 PSI ADJUSTABLE SAMPLE FLOW METER RED LED ALARM INDICATOR 120 VAC (12 VDC OUTPUT) POWER ADAPTER

4 GENERAL DESCRIPTION The MST Airline Carbon Monoxide and Low Pressure Warning System is designed to take a continuous air source sample and monitor for levels of Carbon Monoxide and to alarm if either carbon monoxide, above the preset maximum permissible level is encountered, or if the airline pressure falls below a preset low pressure set point. The system is mounted on a panel for wall mounting. GENERAL OPERATION (Refer to Figure No. 1) Initially the 120 VAC continuous power adapter (A) must be plugged into a dedicated 120 VAC outlet to provide power to the system. NOTE: If monitor s switch (B) is turned on before the minimum pressure is applied at (C), the low pressure warning system will activate the alarm system. The air source to be monitored will enter through the system s inlet at (C). After air source has been connected to the system s inlet, adjust the black flow control knob of the Flowmeter (D) to meter the correct sample air to the CO Monitor s (E) electrochemical CO sensor (F). If the CO content in the sample air exceeds the monitor s CO alarm setting, the monitor s internal buzzer (G), red indicator LED (H) and the Remote Audible Alarm (I) will be energized. During normal operation, should a low pressure situation occur, only the Remote Audible Alarm (I) will sound. FIGURE NO. 1

5 INSTALLATION AND ADJUSTMENT Locate a suitable place to mount or install the Model 5700-S1 and secure the panel utilizing the mounting holes on the corners and attach the air supply (100 psi max.) to the 3/8" FPT inlet located at C in Figure No. 1. DO NOT APPLY EXCESS PRESSURE to the aluminum manifold unless supporting the manifold with another wrench. Should the factory preset level (20 psi) be satisfactory, the unit is ready to operate by pressurizing the system and turning the ON-OFF Switch on the 5700 CO Monitor to the ON position. The CO Monitor must now be calibrated to assure proper operation (see Monitor s Manual). To prevent the audible external alarm from sounding when calibrating the monitor, unplug the alarm from the monitor. Be sure to reattach the alarm when calibration is completed. Should you need to adjust the switch to some other setting, loosen the four (4) screws located at the four corners of the housing and carefully lift off front cover to expose the pressure switch and internal wiring (see Figure No. 2). Locate the white adjustment dial located at the base of the pressure switch. Note: The numbers located on the adjustment dial are for approximation only - DO NOT assume the setting noted on the dial is the alarm set pressure. Adjust the system pressure to alarm set point level desired and turn the adjustment dial until the unit just activates the alarm. Slowly increase the system pressure above this setting until the alarm stops. Slowly lower the system pressure until alarm activates. Note the system pressure, and if this is the desired alarm set point, the unit is ready for operation. Readjust if necessary and replace front cover. Warning: The Alarm Set Point should be verified prior to each use to assure the unit is operating properly and no one has tampered with the device.

6 MST, Inc. strongly recommends that a complete safety program be instated to ensure that the respiratory air is in compliance with all OSHA/CSA standards and other applicable laws regulating the use of supplied air respiratory systems. MST, Inc. recommends that the air quality be tested upon installation and periodically re-tested to ensure that the minimum requirements for breathing air are maintained. MST, Inc. will not assume any liability for accidents or personal injury resulting from the improper use of this equipment. Service on this equipment should only be performed by qualified personnel. This system is to be used only by trained qualified personnel in accordance with a respiratory program as outlined in OSHA Regulation 29 CFR 1910.134 (b).

7 MST'S AIRLINE "CO" MONITORING SYSTEM INITIAL INSTALLATION AND START-UP (Refer to Figure No. 1) 1) SUPPLIED AIR HOOK-UP - MST Unit s inlet is supplied with 3/8" NPT female threads for customer hook-up (C). MST suggests the supplied air be regulated to a maximum pressure of 100 psi and free of water and/or oil and dirt particles. 2) POWER MONITOR AND CALIBRATE -Plug in the 120 VAC continuous power adapter (A) into a dedicated 120 VAC outlet. At this time, to avoid hearing the loud Remote Audible Alarm (I), the alarm can be disconnected during the calibration procedure by disconnecting alarm at (J). WARNING: Be sure to reconnect the Remote Audible Alarm when calibration is complete. After a five (5) minute warm-up period, calibrate the monitor according to the MST Monitor Manual. 3) CALIBRATION GAS REQUIREMENTS - Zero Gas: Nitrogen or Air, free of CO. Span Gas: 95 PPM of CO concentration in air. NOTE: DO NOT USE GAS THAT IS INTENDED TO CALIBRATE MULTIPLE SENSOR GAS MONITORS, EVEN IF THE CO CONCENTRATION IS CORRECT. Calibration gas flow to monitor should be 1.0 SCFH (472 ml/min). 4) EXTREME TEMPERATURE CHANGES - Avoid; the MST 5700 CO Monitor best performs at an ambient temperature range of 32-104 F (0-40 C). Always calibrate the monitor after it has stabilized in the surrounding temperature where the system is to be used. 5) LOW PRESSURE WARNING SYSTEM AND MONITOR SAMPLE FLOW ADJUSTMENTS - After the monitor has been calibrated, turn off the monitor and reconnect the Remote Audible Alarm (I). Turn the black flow control knob of flowmeter (D) clockwise to make sure it is turned off. Connect your compressed air to the inlet (C). Now carefully turn the black flow control knob counter-clockwise to open the valve. Open just enough to keep the black ball floating about midrange in the green safe zone area of the flow meter (refer to Figure No. 2).

8 WARNING : SERIOUS INJURY could result if the AIR SAMPLE METERING VALVE is not properly adjusted. Proper sample air flow to CO monitor is required for monitor to give correct CO level readout. To adjust the low pressure setting, an external regulator will be required to pressurize the system to the low pressure set point. For example, if the lowest pressure allowed for your respiratory system is 50 psi, then adjust the external regulator to 50 psi and using the procedure outlined on page 5 of this manual, adjust the new set point to 50 psi. Recheck pressure setting by turning the external regulator above the set point to deactivate the alarm, and slowly reduce pressure until alarm activates. Note if correct pressure was reached when alarm activated and readjust if required. FIGURE NO. 2

9 MST AIRLINE "CO" MONITORING SYSTEM GENERAL OPERATION AND MAINTENANCE WARNING: If the Monitor's alarm should sound, remove mask or hood immediately and move to a safe breathable atmosphere. Have a proper qualified personnel examine air system and make the appropriate corrections before using again. 1) MONITOR ALARM CONDITION - The monitor will alarm due to one or more of the following conditions: a) Monitor is out of calibration. The Monitor should be calibrated monthly if used continuously and prior to use if used on a non-continuous basis. Calibrate Monitor as outlined in the MST Monitor Manual. b) If the Monitor can be and is calibrated, but the alarm still sounds, excessive CARBON MONOXIDE IS PRESENT IN YOUR AIR LINE. REMOVE MASK OR HOOD IMMEDIATELY AND MOVE TO A SAFE BREATHABLE ATMOSPHERE. CHECK AIR SYSTEM TO CORRECT PROBLEM BEFORE USING AGAIN. c) If the Monitor can not be calibrated, the Carbon Monoxide Sensor may require replacement. See MST MONITOR MANUAL for replacement instructions and other trouble shooting information. The MST MONITOR has a (1) year warranty. All warranty work must be performed at factory. d) If the Monitor was calibrated in a surrounding temperature other than where the system was being used and the temperature difference was 36 o F (20 o C) or greater, the Monitor may give a false alarm due to its characteristics. Always calibrate the Monitor in the temperature conditions where the monitor is to be used in. Monitor best performs at temperature range of 32 to 104 o F (0 to 40 o C). e) Inlet air pressure below low pressure set point is encountered. NOTE: Only the external alarm will sound. Increase inlet pressure to required pressure for your respirators (or readjust low pressure set point to the correct pressure requirement of your respiratory system). 2) MONITOR "NORMAL" MODE - The Monitor's "NORMAL GREEN LED" light will be on full bright from 0-9 PPM of carbon monoxide (or 0-5 PPM in Canda) and the "ALARM RED LED" light will be faintly blinking while the Low Battery Amber LED is off.

10 3) MONITOR "ALARM" MODE - If carbon monoxide concentrations exceeds the alarm point (factory set at 10 PPM or 5 PPM in Canada), the "ALARM RED LED" light will come on full bright, the "NORMAL GREEN LED" light will be off and the audible alarm will sound. Also the Remote Audible Alarm will sound. 4) CONTAMINATED AIR SUPPLY - If water and/or oil is visibly seen in the flow meter and/or sample tube to MST Monitor, excessive water and/or oil is present in the supplied air. These contaminates should be coalesced out prior to inlet of MST's "CO" Monitoring System to eliminate clogging of MST's Flow Meter and or "CO" Sensor. RECORD KEEPING Record all periodical air quality checks, monitor calibration date and any other service performed on the MST "CO" Monitoring System. MST, INC. SHALL NOT BE LIABLE FOR ANY INJURY LOSS OF DAMAGE, (DIRECT OR CONSEQUENTIAL), ARISING OUT OF THE USE OF OR THE INABILITY TO USE THIS PRODUCT, BEYOND THE REPLACEMENT OF THE DEFECTIVE MATERIALS OR WORKMANSHIP. USER OF SUPPLIED AIR RESPIRATORS SHOULD EVALUATE THEIR OWN PARTICULAR APPLICATION AND PERFORM THEIR OWN TESTS FOR AIR QUALITY TO DETERMINE THE SUITABILITY FOR USE OF THIS PRODUCT. For further information, or questions about service or maintenance care of this unit, contact your local distributor or MST, Inc. @ (800) 542-6646.

11 SERVICE RECORD MST AIRLINE "CO" MONITORING SYSTEM MODEL 5700 - S1 DATE OF SERVICE SERVICE PERFORMED