SECTION BIOLOGICAL SAFETY CABINETS AND LAMINAR AIRFLOW WORK STATIONS

Similar documents
SECTION COMMERCIAL-KITCHEN HOODS

SECTION MORTUARY EQUIPMENT

A. Standards: Except where modified or exceeded by the requirements of this Specification, conform to the following standards:

SECTION CUSTOM FABRICATED LABORATORY EQUIPMENT. This section specifies necropsy // dissecting tables // and // animalcage

B. Design meeting requirements of the particular class and type of cabinet as follows:

SECTION CUSTOM FABRICATED FOODSERVICE EQUIPMENT

Adv. 8.4 Guidelines for Biosafety Cabinets Rev

Biological Safety Cabinets. Presented by Mr. Rene Soetens President

SECTION AIR COILS

5.0 Installation Instructions

LABGARD NU-602 CLASS II TYPE A2 ANIMAL HANDLING BIOLOGICAL SAFETY CABINET

SECTION GAS-FIRED RADIANT HEATERS

SECTION ISOLATED POWER SYSTEMS

THE IMPACT OF FACILITY HVAC SYSTEMS ON BIOLOGICAL SAFETY CABINETS

XRRS Specification Fire Ready Hood

SECTION LABORATORY EQUIPMENT CONSTRUCTION STANDARD PART 1: GENERAL Scope of Standard

BSC CLASS II, TYPE A2 BIOLOGICAL SAFETY CABINET SERIES. Optimum protection of personnel, product and environment. BSC Class

SECTION AIR DUCT ACCESSORIES

ELECTRIC COOKTOP INSTALLATION INSTRUCTIONS

A. Hydraulic calculations shall be prepared in accordance with Chapter 23 of NFPA 13 with the following exceptions:

SECTION AUTOMATIC TRANSFER SWITCHES

University of Houston Master Construction Specifications Insert Project Name SECTION DRY-PIPE FIRE SPRINKLER SYSTEM PART 1 - GENERAL

UNIFIED FACILITIES GUIDE SPECIFICATIONS

UC DAVIS Campus Design Guide DIVISION 11 - EQUIPMENT

KINGS COUNTY JAIL EXPANSION PHASE III COUNTY OF KINGS SECTION

Environmental Health and Safety Biological Safety Program Guidance Document

Safe Operating Procedure (Revised 2/10)

Gas Station & C Store SECTION CWM Job No

SECTION AIR DOORS

UC DAVIS Campus Standards & Design Guide DIVISION 11 - EQUIPMENT

SECTION ELECTRIC-DRIVEN, FIRE PUMPS

A. AMERICAN ARCHITECTURAL MANUFACTURERS ASSOCIATION (AAMA) 101: Appendix Dissimilar Materials.

SPECIFICATION DELETED SECTION CENTRIFUGAL HVAC FANS

SECTION GENERATOR PARALLELING CONTROLS

Lab Design shall comply with the UC Lab Safety Design Manual located at: LABORATORY FUME HOODS

A. Product Data: Include rated capacities, operating characteristics, furnished specialties, and accessories.

Research Document Type: PROCEDURE. Effective Date: January 01,2015 Author Steven Hayes

LABORATORY AIR COMPRESSORS AND VACUUM PUMPING SYSTEMS

SECTION COMPRESSED AIR SYSTEM. A. Pipe and pipe fittings, including valves, unions and couplings.

Innovation in air purification unmatched air quality

Clean Bench Animal Transfer Stations

A. American National Standards Institute (ANSI): Establishes requirements applicable to certifying direct gas-fired heaters.

Construction Standards Page Exclusively published and distributed by Architectural Computer Services, Inc. (ARCOM) for the AIA

SECTION FUEL-FIRED DOMESTIC WATER HEATERS

Direct Gas-Fired Make-Up Air

SECTION (15486) - FUEL-FIRED, DOMESTIC WATER HEATERS

A. General: Refer to paragraph entitled "SUBMITTAL" in Section Include the following data:

1.1 This section applies to air handling units for HVAC Systems.

A. Product Data: Include rated capacities, furnished specialties and accessories.

Biological Safety Cabinets or Fume Hoods Which one to use?

PhiloWilke Partnership Addendum No. 2 A/E Project No April 2016

NEW! Thermo Scientific 1300 Series B2 Class II, Type B2 Biological Safety Cabinet

High Plume Blowers Suggested Specification Section (Master Format 1996) Suggested Specification Section (Master Format 2004)

Second Revision No. 3 implements the text of Committee Input No. 36 as proposed in the First Draft Report, along with the following amendments:

Guide Spec Summary. Option List. Date: 05/21/2001. EarthWise VAV Terminal Units Full Spec. Prepared by: Phone Number: Prepared for:

UNIVERSITY SERVICES ANNEX James Madison University Harrisonburg, Virginia State Project Code: Architect s Project Number:

FIRE SUPPRESSION SPRINKLER SYSTEM

Indirect Gas-Fired Make-Up Air

SECTION AIR COILS

A. American National Standards Institute (ANSI) : Establishes requirements applicable to certifying direct gas-fired heaters.

SECTION INTERIOR LIGHTING

SECTION WATER-SOURCE UNITARY HEAT PUMPS

A. Product Data: Include rated capacities, operating characteristics, furnished specialties, and accessories.

SECTION STEAM CONDENSATE PUMPS

Design and Construction Standards SECTION PLUMBING EQUIPMENT

Class II, Type B2 (Total Exhaust) Biological Safety Cabinets The Safety Solution for Life Science Laboratories

DeCA DESIGN CRITERIA HANDBOOK PRODUCT REFRIGERATION SYSTEM

A. This Section includes the following emergency plumbing fixtures:

SECTION WET-PIPE SPRINKLER SYSTEMS

Mercy Hospital of Buffalo Catholic Health System 2 nd Floor Patient Holding Center 3 rd Floor Locker Rooms Shaflucas Architects, PC Project No.

CubeSmart Gulf Breeze Storage AUTOMATIC FIRE SPRINKLER SYSTEM

SECTION (15833) - HVAC POWER VENTILATORS

AMERICAN AIR & WATER, INC.

A. Section Common Motor Requirements for HVAC Equipment: Fan motors.

SECTION (13900) - FIRE SUPPRESSION SYSTEMS

C. ASSE 1013: Performance Requirements for Reduced Pressure Principle Backflow Preventers.

SECTION FIRE-SUPPRESSION STANDPIPES

SECTION AIR HANDLING UNIT

ANIMAL RESEARCH PRODUCTS. Biological Safety Cabinets Animal Transfer Stations

SECTION Fire Pumps

BIOLOGICAL SAFETY CABINET

SECTION COOLING TOWERS

Installation Instructions

D. Reference Standards (Utilize latest editions available):

Brown University Revised August 3, 2012 Facilities Design & Construction Standards SECTION AIR HANDLING UNITS

SECTION CONVECTION HEATING AND COOLING UNITS

Steel Casework Acid Storage

Design & Construction Standards, Revised January

REFRIGERANT PIPING SYSTEM

SETION EMERGENCY LIGHTING NATIONAL FIRE PROTECTION ASSOCIATION (NEPA) (2006) Life Safety Code, 2006 Edition

A. Section Instrumentation and Control Devices for HVAC. B. Section Building Automation Systems(BAS) for HVAC

SECTION LOW-VOLTAGE ELECTRICAL POWER CONDUCTORS AND CABLES (600 VOLTS AND BELOW)

SECTION DRY PIPE SPRINKLER SYSTEMS MODEL G SERIES DRY VALVE

6/19/2015 BIOLOGICAL SAFETY CABINET. Types of Classes. Types of BSC.

SECTION AIR HANDLING UNITS WITH COILS. 1. Central Station Air Handling Units. 1. Section Basic Mechanical Requirements.

SECTION WET-PIPE SPRINKLER SYSTEMS EASY RISER CHECK VALVE

SECTION DRY PIPE SPRINKLER SYSTEMS MODEL F SERIES DRY VALVE

SECTION PLUMBING EQUIPMENT

TYPICAL CPA SPECIFICATIONS

Forma 1400 Series Class II, Type A2 Biological Safety Cabinet. Superior Ergonomics, Uncompromising Safety

Transcription:

SECTION 11 53 53 BIOLOGICAL SAFETY CABINETS AND LAMINAR AIRFLOW WORK STATIONS SPEC WRITER NOTES: 1. Delete between // // if not applicable to project. 2. Delete other items or paragraphs in the section that are not applicable and renumber the paragraphs. PART 1 - GENERAL 1.1 DESCRIPTION This section specifies Biological Safety Cabinets: // Class I // Class II, Type A1 // Class II, Type A2 // Class II, Type B1 // Class II, Type B2 // and Laminar Airflow Work Stations (LAFW): // Vertical // Horizontal //. 1.2 DEFINITIONS A. Class I Biological Safety Cabinet: A ventilated cabinet for exposure protection of personnel and the environment, suitable for work involving low to moderate risk agents (BSL 1,2, and 3). Cabinet air is exhausted through a HEPA filter, either into the laboratory or to the outside. B. Class II Biological Safety Cabinet: A ventilated cabinet for exposure protection of personnel, product and the environment, suitable for work involving low to moderate risk agents (BSL 1,2, and 3). Cabinet air is exhausted through a HEPA filter either into the laboratory or to the outside. Class II cabinets are available as two types (A and B) based on construction, air flow velocities and patterns, and exhaust systems. Refer to Table 1. C. Laminar Airflow Work Station(LAFW): A self-contained, positive pressure cabinet that provides an ISO 5/Class 100 clean air environment, designed to provide product protection when handling non-hazardous products. Streams of unidirectional air move in parallel lines through the cabinet, then through a HEPA filter, and into the laboratory. Both horizontal and vertical laminar flow work stations are available. 1.3 RELATED WORK A. Section 22 15 00, GENERAL SERVICE COMPRESSED-AIR SYSTEMS: Connections to Compressed Air System. B. Section 22 11 00, FACILITY WATER DISTRIBUTION: Plumbing Connections. C. Section 22 13 00, FACILITY SANITARY SEWERAGE: Plumbing Connections. 11 53 53-1

D. Section 22 62 00, VACUUM SYSTEMS FOR LABORATORY and HEALTHCARE FACILITIES: Connections to Gas and Vacuum Systems. E. Section 22 63 00, GAS SYSTEMS FOR LABORATORY and HEALTHCARE FACILITIES: Connections to Gas and Vacuum Systems. F. Section 23 09 23, DIRECT-DIGITAL CONTROL SYSTEM FOR HVAC: Pressure Switches. G. Section 23 31 00, HVAC DUCTS and CASINGS: Ductwork. H. Section 23 36 00, AIR TERMINAL UNITS: Airflow Control Valves. I. Section 23 40 00, HVAC AIR CLEANING DEVICES: HEPA Filters. J. Section 26 05 21, LOW-VOLTAGE ELECTRICAL POWER CONDUCTORS and CABLES (600 VOLTS AND BELOW): Electrical Connections. K. Section 26 27 26, WIRING DEVICES: Electrical Devices. 1.4 QUALITY CONTROL A. Mechanical, electrical and associated systems shall be safe, reliable, efficient, durable, easily and safely operable, maintainable, and accessible. B. Standard Products: Material and equipment shall be the standard products of the selected manufacturer, and they should be regularly engaged in the manufacture of such products for at least 3 years. C. All items furnished shall be free from defects that would adversely affect the performance, maintainability and appearance of individual components and overall assembly. D. NSF Compliance: Equipment bears NSF (National Sanitation Foundation) Certification Mark indicating compliance with NSF 49. This certification applies only to Class 1 and 2 Biological Safety Cabinets, not LAFW. E. Electrical Components and Devices: UL listed and labeled for intended use. 1.5 SUBMITTALS A. Submit in accordance with specification Section 01 33 23, SHOP DRAWINGS, PRODUCT DATA, AND SAMPLES. B. Manufacturer's Literature and Data: Include the following: 1. Illustrations and descriptions of the unit and factory-installed devices associated with it. 2. Catalog or model numbers for each item incorporated into the work. 3. Utility requirements. 11 53 53-2

C. Shop Drawings: Show details of fabrication, installation, adjoining construction, coordination with mechanical and electrical work, anchorage, and other work required for complete installation. D. Factory Testing: Provide manufacturer s QC checklist or other reports that indicate comprehensive factory testing has been performed, and the results of these tests. E. Field Test Reports: Provide certification reports from accredited service technicians or installers. F. Operating Instructions: Comply with requirements in specification Section 01 00 00, GENERAL REQUIREMENTS. SPEC WRITER NOTE: Edit the following requirements to coordinate with project sustainability goals. G. LEED Information: 1. LEED (v 3.0) MR Credit 4, Recycled Content: Product data indicating percentages, by weight of post-consumer and post-industrial recycled content for products having recycled content: a. Include statement indicating costs for each product having recycled content. 2. LEED (V 3.0) MR Credit 5, Regional Materials: Manufacturer s data identifying point of origin for products procured within 500 mile radius of the project: a. Include statement indicating costs for each product submitted. 1.6 APPLICABLE PUBLICATIONS A. The publications listed below form a part of this specification to the extent referenced. The publications are referenced in the text by the basic designation only. B. American National Standards Institute / National Electrical Manufacturers Association (ANSI/NEMA): WD 6-2002 (R2008)...Wiring Devices--Dimensional Specifications C. National Sanitation Foundation International / American National Standards Institute (NSF/ANSI): 49-2009...Biosafety Cabinetry: Design, Construction, Performance and Field Certification D. Scientific Equipment and Furniture Association (SEFA): 2-1999...Recommended Practices for Installation 7-2007...Recommended Practices for Fixtures E. National Fire Protection Association (NFPA): 11 53 53-3

45-2011...Standard on Fire Protection for Laboratories PART 2 - PRODUCTS 2.1 BIOLOGICAL SAFETY CABINETS using Chemicals SPEC WRITER NOTE: Edit the following requirements to coordinate with the equipment shown on drawings. A. General Specification for Biological Safety Cabinets SPEC WRITER NOTE: The following specifications apply to both Class I and Class II Biological Safety Cabinets. Furnish and install biological safety cabinets that have the following characteristics: 1. Cabinet Exterior: a. Class I: // Epoxy-coated aluminum // polycarbonate and steel // steel //. b. Class II: // Reinforced cold-rolled steel with acid-resistant painted finish // stainless steel //. 2. Cabinet Interior: a. Class I: // Epoxy-coated aluminum // polycarbonate and steel // steel //. b. Class II: // Reinforced cold-rolled steel with acid-resistant painted finish // stainless steel //. 3. View Screen: Hinged, 6-mm- (1/4-inch-) minimum thick, laminated safety glass. 4. Motor/Blower System: To circulate filtered air into and/or through cabinet. 5. Duct: Stainless-steel duct to direct air from the recirculation blower cabinet to the exterior. Refer to Section 23 31 00, HVAC DUCTS and CASINGS. 6. HEPA Filtering: 99.99 percent effective on 0.3 microns for both recirculated and exhausted air. Refer to Section 23 40 00, HVAC AIR CLEANING DEVICES. 7. Lighting: Fluorescent lights producing a minimum of 1076 lux (100 fc) of non-glare illumination in the work area, and ultraviolet lamp electrically interlocked to be inoperable while fluorescent lighting is "on." 8. Magnehelic Gauge: Located on the front of the cabinet to indicate cabinet interior pressure. Provide DDC sensor with remote indicator. Refer to Section 23 09 23, DIRECT-DIGITAL CONTROL SYSTEM FOR HVAC. 11 53 53-4

SPEC WRITER NOTE: Show types, quantities, and locations of wiring devices and service fixtures on the drawings. 9. Equip with listed hospital-grade duplex receptacles having drip proof covers, and control switch. Refer to Section 26 27 26, WIRING DEVICES for additional requirements 10. Equip with gas, vacuum, water, and air valves, as needed. 11. Equip with the following controls: a. On/off switch and circuit breaker with pilot "on" light for blowers. b. On/off switch for // fluorescent // and ultraviolet lamps //. c. Circuit breaker and on/off switch for duplex outlets. d. Removable stainless steel exhaust duct transition unit with airtight damper // On type A2 cabinets, provide an indirect canopy ( thimble ) connection by the cabinet manufacturer //. 12. Locate HEPA filters and blower so that they are removable from the front without entry into workspace. 13. Equip with drain spillage trough in each unit. Provide ball valve and cap on trough drain outlet. 14. Equip with optional Intravenous (IV) bar, suspended across the top of the work area. 15. Equip with a supportive work surface or base stand. SPEC WRITER NOTE: Edit the following requirements to coordinate with the equipment shown on drawings. B. Class II Biological Safety Cabinet Types: Furnish and install Class II biological safety cabinets, according to the following table: Table 1 TYPE AIRFLOW MAXIMUM FACE VELOCITY A1 30 percent indirect 0.38 m/s exhaust / 70 percent (75 fpm) recirculated A2 B1 B2 30 percent indirect exhaust / 70 percent recirculated 70 percent direct exhaust / 30 percent recirculated 100 percent direct exhaust 0.51 m/s (100 fpm) 0.51 m/s (100 fpm) 0.51 m/s (100 fpm) Biosafety Level COMMENTS 1,2,3 Contaminated positive-pressure plenums permitted 1,2,3 Contaminated plenums must be surrounded by negative pressure. 1,2,3 Contaminated plenums must be surrounded by negative pressure. 1,2,3 All ducts and plenums must be under negative pressure. 11 53 53-5

2.2 LAMINAR AIRFLOW WORK STATIONS (LAFW) A. Specifications for Laminar Airflow Work Stations (LAFW) 1. Cabinet Exterior: // Reinforced cold-rolled steel with acidresistant painted finish // stainless steel //. 2. Cabinet Interior: // Reinforced cold-rolled steel with acid resistant painted finish // stainless steel //. 3. Motor/Blower System: To circulate filtered air into and/or through cabinet. 4. HEPA filtering: 99.99 percent effective on 0.3 microns for exhausted air. 5. Duct: Self-contained unit; no ducting required. 6. Lighting: A minimum of 1076 lux (100 fc). 7. Magnehelic Gauge: +/- 2% accuracy of full scale with a 0.35 bar (15 psi) pressure rating. 8. Equip with hospital-grade duplex receptacles having drip proof covers, and control switch. Refer to Section 26 27 26, WIRING DEVICES for additional requirements. 9. Equip with gas, vacuum, water and air valves as indicated on drawings. 10. Equip with the following controls: a. On/off switch and circuit breaker with pilot on light for blowers. b. On/off switch for // fluorescent // and ultraviolet lamps //. c. Circuit breaker and on/off switch for duplex outlets. 11. Locate HEPA filters and blower so that they are removable from the front without entry into workspace (Vertical Laminar Airflow Work Stations only). 12. Equip with drain spillage trough in each unit. Provide ball valve and cap on trough drain outlet. 13. Equip with optional Intravenous (IV) bar, suspended across the top of the work area. 14. Equip with a supportive work surface or base stand. 2.3 MECHANICAL SERVICE FIXTURES A. Valves, General Requirements: 1. Comply with requirements in SEFA 7. 2. Cast red brass alloy bodies with copper content not less than 81 percent, or drop forged brass alloy with high density and no porosity. 11 53 53-6

3. Locate valves so that they are accessible for maintenance and repair of internal working parts. 4. Equip valves with four-arm handles. 5. Design valves to withstand 689 kpa (100 psig) without leakage. B. Gas, Air, and Vacuum Valves: 1. Provide floating needle valves with a replaceable cone and a replaceable valve seat. 2. Provide bonnet with exterior packing nut and packing gland designed for valve to be repacked while under pressure. C. Outlet Fittings: Fit each outlet with a 10 serrated hose connector. D. Electrical System: 115 V, 1 phase, 60 Hz. E. Identification: Code valves with full-view plastic index buttons as follows: Table 3 SERVICE BUTTON COLOR CODE LETTER COLORS Air Orange AIR Black Gas Dark Blue GAS White Vacuum Yellow VAC Black F. Finish: 1. Fixtures, Handles, and Escutcheons: Polished chrome plate. 2. Fixtures Inside Hoods: Acid- and solvent-resistant coating applied by fixture manufacturer. G. Electrical Receptacles: Hospital-grade; ANSI/NEMA WD 6 Configuration 5-20R; duplex; with chrome-plated brass or stainless-steel cover plates; minimum 120 V, 20 A. PART 3 EXECUTION 3.1 PREPARATION Install equipment after installation of finish flooring in rooms to receive cabinets has been completed. 3.2 INSTALLATION A. General: 1. Install biohazard safety cabinets and LAFWs according to manufacturer's written instructions 2. Coordinate installation with related mechanical and electrical work. Provide cutouts and openings for plumbing and electrical work as indicated or as required by trades involved. 3. Install level, plumb, true, and straight without distortion. 11 53 53-7

a. Shim cabinets using concealed shims. 4. Adjust hardware so that doors and drawers operate smoothly without warp or bind. Lubricate operating hardware as recommended in writing by manufacturer. 5. Locate unit away from fans, heating and air conditioning registers, laboratory hoods, high traffic areas and doors that could interfere with airflow patterns. 3.3 TESTING A. Biological Safety Cabinets: After installation, an independent accredited service technician must test the biological safety cabinet according to NSF 49 and permanently attach certificate of compliance to equipment. B. Laminar Airflow Work Stations: After installation, an independent accredited service technician must test the laminar flow work station and provide a written report certifying that it provides an ISO 5/Class 100 clean air environment. 3.4 PROTECTING AND CLEANING A. Protect equipment from dirt, water, and chemical or mechanical injury during the remainder of the construction period. B. At the completion of work, clean equipment as required to produce ready-for-use condition. 3.5 INSTRUCTIONS Instruct personnel and transmit operating instructions in accordance with requirements in specification. - - - E N D - - - 11 53 53-8