Rigid Reusable & Single use Sigmoidoscopes, Anoscopes, and Accessories ENGLISH 1. Perform reprocessing immediately after each use.

Similar documents
ON TRACK Reprocessing In-service / Competency for CF/PCF/GIF Endoscopes (EVIS, EXERA, EXERA II)

Attendance Sheet. (List each model reviewed during this session) Print Name Signature Title. Print Name Signature Title.

Operator s Manual. Video Laryngoscope. ( Instructions for use ) Doc no: Revision 1.1 8th January 2013

2015 Myco Industries, Inc Fax:

CONVENTUS CAGE TM - DR Procedure Instruments Instrument Tray. Equipment Cleaning and Sterilization

CONVENTUS CAGE TM PH Procedure Instruments Instrument Tray Cleaning and Sterilization

CONVENTUS CAGE TM PH Procedure Instruments Instrument Tray Cleaning and Sterilization

CLEANING INSTRUCTIONS

Made in USA. Sunoptic Technologies 6018 Bowdendale Avenue Jacksonville, FL 32216, USA (904)

Reusable Laryngoscope Blades and Handles

Caution: Federal U.S. law restricts this device to sale by or on the order of a physician.

INSTRUCTIONS EVIS EXERA GASTROINTESTINAL VIDEOSCOPE

Integra Salto Talaris and Integra XT Non-Sterile Device Reprocessing. Instructions for Use. 4. Manual Cleaning Instructions

HANDS-ON MODULE 5. CLEANING FLOW-i V3.0

Manual Pre-disinfection/Cleaning and Sterilization Instructions for SATELEC Tips and Files

Operating / User Manual

MODEL 7000 SUCTION UNIT

Cleaning & Sterilization

I.F.U Re/Processing Reusable Medical Devices

Instructions for Use: Fiberscope and LED Light Source

Manual Pre-disinfection/Cleaning and Sterilization Instructions for SATELEC Tips and Files

Schuco byallied MEDICAL ASPIRATOR. USER'S MANUAL A Caution

H5i disinfection guide

EMAX 2 PLUS AND XMAX CLEANING AND STERILIZATION BASKET. Mechanical/Automated Cleaning and Sterilization Guide in a Variety of Applications

*smith&nephew VERSAJET HYDROSURGERY SYSTEM

Instructions for Use Spinal Punches IFU

SKLAR COATED SURGICAL INSTRUMENTS RECOMMENDED CARE & CLEANING INSTRUCTIONS

المملكة العربية السعودية وزارة الصحة اإلدارة العامة لمكافحة عدوى المنشئات الصحية

Instructions for care, maintenance, cleaning and sterilization of Smith & Nephew orthopaedic devices

INSTRUCTIONS FOR CLEANING AND STERILIZATION OF ALL B.BRAUN AESCULAP HAND HELD SURGICAL INSTRUMENTS AND POWER PRODUCTS (UNITED KINGDOM REQUIREMENTS)

CLEANING AND HANDLING OF WRIGHT INSTRUMENTS

Manual Pre-disinfection/Cleaning and Sterilization Instructions for SATELEC wrenches

When you sterilize an instrument set, you know what the standards are.

Product Catalogue 2016

Accessories. Maybachstraße. Symmetry Fax: Revised 10/15

INSTRUCTIONS EVIS EXERA II DUODENOVIDEOSCOPE

4/8/2014 GOAL PATIENT EQUIPMENT

Aesculap Implant Systems

Flexible Nasopharyngoscope. Operator Manual. Manufactured by Henke-Sass, Wolf Distributed by Anthony Products, Inc.

INSTRUCTIONS EVIS EXERA II DUODENOVIDEOSCOPE

Operator s Manual. Model G32-S Model G32-E Disinfection Soak Stations

V. Mueller. V. Mueller Reusable Devices: Cleaning and Sterilization Guide USA. Rx Only

Christina Bradley Laboratory Manager Hospital Infection Research Laboratory Queen Elizabeth Hospital Birmingham, UK

IMPORTANT INFORMATION ON THE DEPUY SYNTHES

HOMES AND SENIORS SERVICES. APPROVAL DATE: April 2004 REVISION DATE: March 2016 REVISION DATE: March 2017 REVIEW DATE: November 2018

INSTRUCTIONS FOR CLEANING AND STERILISING HUMECA MEDICAL EQUIPMENT FOR BOTH ELECTRIC AND MANUAL DEVICES

Visao HIGH-SPEED OTOLOGIC DRILL CATALOG

FLASH PAK STERILIZATION CONTAINER SYSTEM

AS Medizintechnik GmbH Sattlerstrasse 15, Tuttlingen, Germany Tel 07461/ Fax 07461/

Aesculap Implant Systems

ASPIRE Laboratory Aspirator

EG-2990i, EG-2790i EG-1690K, EG-2490K EG-2790K, EG-2990K EG-3490K, EG-3890TK EG27-i10, EG29-i10

User Guide for Reprocessing of the SUPER VIEW Wide Angle Viewing System Sterilizable Components and Accessories

OPERATING INSTRUCTIONS AND MAINTENANCE MANUAL

SUCTION PUMPS CareLinc Home Medical Equipment and Supply, LLC SS170CE00101

PROFESSIONAL USER GUIDELINES FOR PRACTICE INFECTION CONTROL AND HYGIENE IN COMPANION ANIMAL VETERINARY CLINICS/HOSPITALS

Cavitron Systems, Sterilizable Handpieces and Reservoir Bottles INFECTION CONTROL PROCEDURES

PROFESSIONAL USER GUIDELINES FOR PRACTICE INFECTION CONTROL AND HYGIENE IN COMPANION ANIMAL VETERINARY CLINICS/HOSPITALS

CORRECT ENDOSCOPE REPROCESSING

OPERATION MANUAL. (Cleaning, Disinfection and Storage) Endoscope ED-530XT

DIRECTIONS FOR USE. Keeping you insafehands to avoid needlestick injuries. see for demo video

AS Medizintechnik GmbH Sattlerstrasse 15, Tuttlingen, Germany Tel 07461/ Fax 07461/

Processing (cleaning, disinfection and sterilization) of instruments from Smith & Nephew Orthopaedics AG

CER-1 / CER-2 (MV-1 / MV-2) (CLM-5001X / CLM-5002VX)

Operator s Guide. Autoclavable Internal Handles with Integrated Paddles Rev. M

Clinical Use Only. ResMed HumidAire 2i. Disinfection and Sterilisation Guide. English

idrive Ultra Powered Stapling System

Classification of Instruments -

Operator s Manual. Model G17-EU (International) Disinfection Soak Station for Hysteroscopes, Cystoscopes and ENT scopes

Table of Contents. English

Additionally, consider the legal provisions valid for your country as well as to the hygienic instructions of the doctor s practice or hospital.

Best Practice Guidelines For Cleaning, Disinfection and Sterilization of Critical and Semi-critical Medical Devices

Enteral Pump Instructions for Use To Assemble When the low

The Do's and Don ts in the Instrument Sterilisation Process

The GAK-800 Heated Exercise Gas Analyzer Accessory Kit

Cavitator Ultrasonic Cleaner Instruction Manual Model 20G

VET /2018-E VETPUMP 2. Precision Suction & Irrigation System

SterilContainer S for use in the Steris V-PRO 60 Low Temperature Sterilization System

RECOMMENDATIONS FOR CLEANING DEVICES

Cleaning and Sterilisation for Burs, accessories and filling material

STERITITE INSTRUCTIONS FOR USE

2. Push the drain connector into the disinfectant removal port until it clicks.

Cleaning and disinfecting

AS Medizintechnik GmbH Sattlerstrasse 15, Tuttlingen, Germany Tel 07461/ Fax 07461/

CLEANING OF INSTRUMENTS

ENDOSCOPE REPROCESSOR OER-Pro Operation Manual

Environmental Cleaning Tool Kit

BIOVAC 500 SMOKE EVACUATOR

Rapid Heat Sterilizers with 6, 8, and 12 Minute Sterilization Cycle Times

PICNet CONFERENCE 2013

AS Medizintechnik GmbH Sattlerstrasse 15, Tuttlingen, Germany Tel 07461/ Fax 07461/

DRAFT. Module 4. Decontamination & Environmental Cleaning

User manual. DreamStation Heated Humidifier

INSTRUCTIONS FOR USE

RAP-hip Hip stem extraction set

CryoPen Cryosurgical System OPERATOR S MANUAL

OPERATION MANUAL KLEENSCREEN FILTRATION SYSTEM SUPPLEMENT INSTRUCTION MANUAL FOR: Gas Fryer Models. Electric Fryer Models

PENTAX VIDEO DUODENOSCOPE ED-3490TK

PROFESSIONAL USER GUIDELINES FOR PRACTICE INFECTION CONTROL AND HYGIENE IN VETERINARY EQUINE PRACTICE

Cleaning and Sterilization for Instruments & Posts

Transcription:

Rigid Reusable & Single use Sigmoidoscopes, Anoscopes, and Accessories ENGLISH 1 ENGLISH Intended Use The Welch Allyn Rigid Reusable & Single Use Sigmoidoscopes and Anoscopes are used to illuminate and facilitate the examination of the rectal and anal cavity. The instrument also provides a means of entry for surgical instruments into the rectal and anal cavity. Endoscopes and the insufflation system are categorized as semi-critical devices requiring high level disinfection and the illumination system is categorized as a non-critical device requiring low-level disinfection according to the Multi-society guideline for reprocessing flexible gastrointestinal endoscopes published by Gastrointestinal Endoscopy (volume 58, No. 1, 2003) and simultaneously in Infection Control and Hospital Epidemiology. Optional higher level processes have been included in these instructions. Refer to ASTM F1518-00, Standard Practice for Cleaning and Disinfection of Flexible Fiberoptic and Video Endoscopes Used in the Examination of the Hollow Viscera, for additional guidance on reprocessing of these devices/accessories. Although this standard is for flexible endoscopes, it contains information that can be applied to rigid endoscopes as well. U.S. federal law restricts this device to sale by or on the order of a physician or licensed healthcare practitioner. This device should only be used by trained personnel. Perform reprocessing immediately after each use. Consult your facility's procedures and the reprocessing agents manufacturer's instructions for the agents reuse, recommended Personal Protective Equipment and other safety precautions such as ventilation, etc. Examine surfaces of all components for wear or damage after cleaning, disinfection, and/or sterilization. For cleaning and/or disinfection, use brushes appropriate for the size of the endoscope channel parts, connectors, and orifices (e.g. bristles should contact all surfaces). Cleaning/disinfection tools (brushes, etc.) should be single use or thoroughly cleaned, disinfected, and/or sterilized between uses. Unless otherwise specified, perform all reprocessing procedures at ambient (room) temperature. Clean and disinfect all devices or accessories per these instructions prior to returning to Welch Allyn (e.g. for repair and service). Users are responsible to qualify any deviations from these procedures.

2 ENGLISH Cleaning, Disinfection, and Sterilization Instructions Rigid Sigmoidoscopes, Anoscopes, and Light Carrier 37023 32820 36019 Anoscope Sigmoidoscope Light Carrier 37019 38614 32010 32820 36019 37023 38619 32020 32830 38700 37027 38622 32410 33220 Suction Tube* 38108 38900 32420 33620 38114 39614 32810 33830 30130 38119 39619 30140 38122 39622 *Suction Tubes not illustrated Note For all steps, remove obturator for separate cleaning and open viewing window. Welch Allyn recommends the use of the Single Use Insufflation Bulb Filter ( 30210) for all applicable endoscopic procedures. The filter does not eliminate the need to properly process the insufflation bulb with each use, but restricts internal contamination and aspiration of fluids. Cleaning 1. Prepare an enzymatic detergent formulated for endoscopic instruments according to the manufacturer s instructions.these instructions were validated using ENZOL / CIDEZYME. 2. Submerge the item in the cleaning solution. 3. Using a soft bristle brush, scrub the item submerged in the cleaning solution for a minimum of 5 minutes then rinse with sterile water for 5 minutes. High Level Disinfection 2. Using CIDEX OPA (0.55% OPA, 14-day maximum reuse), soak the item for a minimum of 12 minutes. 3. Perform 3 sequential rinses of the item for a minimum of 5 minutes each with sterile water.

Rigid Reusable & Single use Sigmoidoscopes, Anoscopes, and Accessories ENGLISH 3 Sterilization 4. Allow the item to dry for 24 hours or until no moisture remains in the device. Any moisture in the device could occlude the insufflation bulb filter. 2. Perform one autoclave cycle that includes 6 minutes at 270 F (132 C) and a cool down period of at least 15 minutes. These instructions were validated using a gravity autoclave. Validate any other sterilization equipment and load configurations. Light Handle Assembly and Handle Adapter Part Numbers 73210 and 73211 (including 07800-U lamp) Part Number 73500 (including 03100-U lamp) WARNING Allow the lamp to cool for at least 5 minutes before handling. Note Leave the lamp installed while reprocessing 73210, 73211 and 73500. Assembly and Disassembly Instructions for 73210 and 73211 1. Disconnect the light handle from the power supply cord and the device. 2. Unscrew the outer sleeve of the handle by turning it counterclockwise. Rectal Light Handle Body and Cord CCW Rectal Light HandleOuter Sleeve 3. Follow one of the reprocessing methods described below. CW Do not immerse the rectal light handle body and cord in any solution. Doing so may damage the rectal light handle body and cord and void the warranty.

4 ENGLISH Cleaning, Disinfection, and Sterilization Instructions 4. After performing a reprocessing procedure, wipe the lamp with a soft, dry, lint free cloth. 5. Re-attach the clean sleeve to the handle by screwing in a clockwise direction. 6. Re-connect the power supply cord to the light handle and verify that the unit works by plugging it in. If the lamp fails to illuminate follow the troubleshooting instructions located in light handle manual ( 700791) supplied with your handle. Cleaning (73210, 73211 and 73500) 1. Prepare an enzymatic detergent formulated for endoscopic instruments according to the manufacturer s instructions.these instructions were validated using ENZOL/ CIDEZYME. 2. Submerge the rectal light handle outer sleeve in the cleaning solution. 3. Using a soft bristle brush, scrub the item submerged in the cleaning solution for a minimum of 5 minutes then rinse with sterile water for 5 minutes. 4. Wet the rectal light handle body and cord with the cleaning solution. 5. Using a soft bristle brush, scrub the cord with the cleaning solution for a minimum of 5 minutes then rinse the cord with sterile water for 5 minutes. Intermediate Level Disinfection (73210, 73211 and 73500) 2. Using CaviWipes, wipe all item surfaces and keep wet with CaviWipe solution for a minimum of 5 minutes. 3. Allow the item surface to dry for a minimum of 10 minutes. Sterilization (73211 ONLY) Autoclaving the light handle may reduce the product s functional life. 2. Perform one autoclave cycle that includes 6 minutes at 270 F (132 C) and a cool down period of at least 15 minutes. These instructions were validated using a gravity autoclave. Validate any other sterilization equipment and load configurations.

Rigid Reusable & Single use Sigmoidoscopes, Anoscopes, and Accessories ENGLISH 5 Insufflation System Do not disassemble this connection Part Number: 30200 Welch Allyn recommends the use of the Single Use Insufflation Bulb Filter ( 30210) for all applicable endoscopic procedures. The filter does not eliminate the need to properly process the insufflation bulb with each use, but restricts internal contamination and aspiration of fluids. Do not disassemble the insufflation bulb. The insufflation bulb may discolor after disinfection or sterilization without effecting performance. Do not autoclave insufflation system. Damage will occur.

6 ENGLISH Cleaning, Disinfection, and Sterilization Instructions Cleaning 1. Prepare an enzymatic detergent formulated for endoscopic instruments according to the manufacturer s instructions.these instructions were validated using ENZOL/ CIDEZYME. 2. Submerge the item in the cleaning solution. 3. Using a soft bristle brush, scrub the item submerged in the cleaning solution for a minimum of 5 minutes then rinse with sterile water for 5 minutes. 4. Using a 60 cc syringe, deliver 60 cc of the cleaning solution to the inside of the bulb through the hose-end. 5. Lightly agitate the solution inside of the bulb for a minimum of 5 minutes. 6. Empty the cleaning solution from the bulb. 7. Using a 60 cc syringe, deliver 60 cc of sterile water into the bulb through the hoseend. 8. Lightly agitate the water inside of the bulb for a minimum of 5 minutes. 9. Empty the rinse water from the bulb, and repeat rinse with sterile water 5 more times. 10. The bulb must be dried sufficiently. Any moisture inside the device could occlude the insufflation bulb filter (if used) or dilute disinfection or sterilization agents. One way to achieve this is to hang the bulb with the tube down for a minimum of 24 hours. High Level Disinfection 1. Follow steps 1 through 9 of the Cleaning instructions. 2. Using CIDEX OPA (0.55% OPA, 14-day maximum reuse), soak the item for a minimum of 12 minutes. 3. Perform 3 sequential rinses of the bulb for a minimum of 5 minutes each with sterile water. 4. Using a clean 60 cc syringe, deliver 60 cc of the CIDEX OPA solution to the inside of the bulb through the hose-end. 5. Lightly agitate the solution inside of the bulb for a minimum of 12 minutes. 6. Empty the solution from the bulb. 7. Using a clean 60 cc syringe, deliver 60 cc of sterile water into the bulb through the hose-end. 8. Lightly agitate the water inside of the bulb for a minimum of 5 minutes. 9. Empty the rinse water from the bulb, and repeat rinse with sterile water 5 more times. 10. The bulb must be dried sufficiently. Any moisture inside the device could occlude the insufflation bulb filter (if used) or dilute disinfection or sterilization agents. One way to achieve this is to hang the bulb with the tube down for a minimum of 24 hours.

Rigid Reusable & Single use Sigmoidoscopes, Anoscopes, and Accessories ENGLISH 7 Sterilization 1. Follow steps 1 through 10 of the Cleaning instructions. 2. Sterilize using a STERRAD 100S for one 55 minute cycle. Validate any other sterilization equipment and load configurations. Single Use Sigmoidoscopes and Anoscopes 53110 53130 Single Use Anoscope Single Use Sigmoidoscope 53110 53130 53110-B 53130-B 53110-JP 53130-F 53130-JP Sterilization Welch Allyn recommends the use of the Single Use Insufflation Bulb Filter ( 30210) for all applicable endoscopic procedures. The filter does not eliminate the need to properly process the insufflation bulb with each use, but restricts internal contamination and aspiration of fluids. Sterilization process is not valid for 53130-L and 53130-LF. These single use devices are lubricated and cannot be sterilized. These devices are not reusable, but can be sterilized one time prior to single use. 1. Remove the device from the plastic bag before sterilizing. 2. Sterilize using a STERRAD 100S for one 55 minute cycle. Validate any other sterilization equipment.

8 ENGLISH Cleaning, Disinfection, and Sterilization Instructions Power Supply 73305, 73322, 73324, 73326 The following cleaning instructions apply to the Power Cord ONLY: 1. Disconnect the power supply cord from the light handle. 2. Disconnect the power supply from its wall receptacle. Cleaning the Power Cord Do not immerse the power supply in any solution or sterilize. Doing so may damage the power supply and void the warranty. The power supply can only be wiped with a soft, dry, lint free cloth or scrubbed with a dry brush. 1. Prepare an enzymatic detergent formulated for endoscopic instruments according to the manufacturer s instructions. These instructions were validated using ENZOL/ CIDEZYME. 2. Wet the cord with the cleaning solution. 3. Using a soft bristle brush, scrub the cord with the cleaning solution for a minimum of 5 minutes then rinse the cord with sterile water for 5 minutes. Intermediate Level Disinfection of the Power Cord 2. Using CaviWipes, wipe the cord and keep wet with CaviWipe solution for a minimum of 5 minutes. 3. Allow the cord surface to dry for a minimum of 10 minutes.

Rigid Reusable & Single use Sigmoidoscopes, Anoscopes, and Accessories ENGLISH 9 Rechargeable Handle Rheostat (Brightness Control) Part Number 71000-A Cleaning Do not saturate the rheostat area. Excess solution entering the assembly may damageinternal components in the handle. 1. Using CaviWipes, wipe the surface to remove all gross debris. 2. Using a soft bristle brush, scrub the handle for a minimum of 5 minutes then repeat wipe with a second CaviWipe. 3. Allow the handle surface to dry for a minimum of 10 minutes. Intermediate Level Disinfection 2. Using CaviWipes, wipe the handle and keep wet with CaviWipe solution for a minimum of 5 minutes. 3. Allow the handle surface to dry for a minimum of 10 minutes.