Integra Ultrasonic Tooth Scaler

Similar documents
ULTRASONIC SCALER INSTRUCTIONS FOR USE STOCK NO. D680

Table of Contents. English

ART-SP1 Scaler /Polisher Combo Unit. Instruction Manual

UL U TR LTRASONIC S ONIC SCALE ALER PIEZO MINI

MODEL 7000 SUCTION UNIT

X99 Series. Dental Handpiece. Instruction For Use

*smith&nephew VERSAJET HYDROSURGERY SYSTEM

User Manual J841Q. Magnetostrictive Scaler / Polisher Combo Unit. Jorgensen Laboratories Tel: Fax:

ASPIRE Laboratory Aspirator

Kinsa Smart Ear ThermometerTM Instructions For Use

SERVICE PROCEDURES. Mattress 2 Pipe Overlay or Replacement

USER MANUAL SAVE THESE INSTRUCTIONS

PRESSURE WASHER MODEL NO: JETSTAR 1750 OPERATION & MAINTENANCE INSTRUCTIONS. WARNING Read the instructions before using the machine PART NO:

Schuco byallied MEDICAL ASPIRATOR. USER'S MANUAL A Caution

Procenter Delivery Systems. [Includes: Delivery Units and Assistant s Units (All Styles)] (no serial numbers on Assistant s Units)

AS Medizintechnik GmbH Sattlerstrasse 15, Tuttlingen, Germany Tel 07461/ Fax 07461/

SAVE THESE INSTRUCTIONS

OPERATION and MAINTENANCE INSTRUCTION MANUAL. AEU-425 Transport II Portable Dental System

Contents. English. English

A few simple steps to ensure perfect water, every time.

CLEAN MIST DUAL TANK SMART HUMIDIFIER. Instruction Manual and Warranty Information READ AND SAVE THESE INSTRUCTIONS IM0036A

EL400 OPERATION MANUAL

Installation & Operating Guide

OPERATION and MAINTENANCE INSTRUCTION MANUAL. ADU-17A Express Air Portable Dental System

OPERATION and MAINTENANCE INSTRUCTION MANUAL. ADU-17X2 Express Plus II Portable Dental System

Garment Steamer Model SC620

ULTRASONIC HUMIDIFIER

INSTRUCTION MANUAL 4HP WET / DRY VACUUM VC V 60 Hz 7A. Gal

ULTRASONIC HUMIDIFIER With Night Light and Fragrance Diffuser

Operating Manual for Profident / Profident PLUS

Installation & Operating Guide

Enteral Pump Instructions for Use To Assemble When the low

MARSHALLTOWN MEDICAL AND SURGICAL CENTER ENVIRONMENTAL SERVICES. SUBJECT: Room Cleaning Occupied Room (All Areas) PROCEDURE NO: 032

Operator s Manual. 6 Gallon Model No. VQ607SFD FOR YOUR SAFETY. Read and understand this manual before use Keep this manual for future reference

Manual Pre-disinfection/Cleaning and Sterilization Instructions for SATELEC Tips and Files

Turbo Stick Vac. OWNER S MANUAL Model EP V., 60Hz., 7.5 Amps.

Manual Pre-disinfection/Cleaning and Sterilization Instructions for SATELEC Tips and Files

Models Product Information. Important Safety Instructions REV

CLEAN MIST SMART HUMIDIFIER TABLETOP & FLOOR STANDING WITH REMOTE CONTROL. Instruction Manual and Warranty Information

System One Heated Humidifier USER MANUAL

Vax Careline: Instruction Manual. Vax Model Number V-029 Vax Model Number V-029Q

OPERATION and MAINTENANCE INSTRUCTION MANUAL. ADU-20B CompriCart

OPERATION MANUAL. Multi Function Ultrasonic Scaler. Varios 970. Please read this Operation Manual carefully before use, and file for future reference.

Instructions for use. BA Optima. Air motors BA602 (BA640081) BA604 (BA640060)

Model # VWK-2 INSTRUCTION MANUAL

ULTRASONIC DIGITAL HUMIDIFIER Instruction Manual and Warranty Information IM0002A READ AND SAVE THESE INSTRUCTIONS

MaxiCompressor. Limited Warranty. High Performance 50 PSI Compressor 501-S

CLEAN MIST SMART HUMIDIFIER Instruction Manual and Warranty Information IM0031B READ AND SAVE THESE INSTRUCTIONS

Operator s Manual. Wet/Dry Vacuum with Detachable Blower 12 Gallon Model No. VBV1210 FOR YOUR SAFETY

OPERATOR S MANUAL FOR YOUR SAFETY. 4 Gallon / Liter Part No. PB408SV

HOT WASHER MODEL NO: KING150

PRO 2000i LOW LEVEL DEPOSITOR

OPERATION AND MAINTENANCE MANUAL

Ultrasonic Humidifier

Bagless upright vacuum

Model HD 3501 Part No High Pressure Washer Operator Manual

PRESSURE WASHER. OWNER'S MANUAL Model VPW49HA 120V., 60 Hz., 13 Amps.

HydroFlux Welder Torch System T0-102 / T0-103

CLEAN MIST SMART HUMIDIFIER Instruction Manual and Warranty Information IM0002E READ AND SAVE THESE INSTRUCTIONS

30 Misting/Oscillating Fan Model Number 01030

EXCELSIOR ENGLER HIGH SPEED VETERINARY DENTAL AIR UNIT INSTRUCTION MANUAL

Ultrasonic Humidifier

CLEAN MIST ULTRASONIC HUMIDIFIER. Instruction Manual and Warranty Information READ AND SAVE THESE INSTRUCTIONS IM0040A

Get Cleaning... What s your Vax s model number? User Guide. Carpet Washer Vax Careline: AU: NZ:

VIVA ace Quick Operation Guide

CLEAN MIST ULTRASONIC HUMIDIFIER. Instruction Manual and Warranty Information READ AND SAVE THESE INSTRUCTIONS IM0036A

User Manual standard for Medical Electrical. Control Number 95CG. Classified by. Underwriter s Laboratory Inc. for UL

Electric Digital Kettle

Dental Unit User's Guide

1300W Pressure Washer RACHP133A

CP-200 Push Button. CP Warewash Pump Instruction & Operation Manual Push Button , Rev1.0,

Electric Pressure Washer NOTE: REFER TO PRODUCT MANUAL FOR ALL SAFETY INSTRUCTIONS

- website: Vax Ltd., Kingswood Road, Hampton Lovett, Droitwich, Worcestershire WR9 0QH, UK

Pet. Easy Guide. VAX Ltd., Kingswood Road, Hampton Lovett, Droitwich, Worcestershire, WR9 OQH, UK - website:

professional FABRIC STEAMER

CONTENTS. III. OPERATING INSTRUCTIONS... pg 7 LOADING DISPOSABLE SET (Figure 4)... pg 9. IV. INDICATORS AND ALARMS... pg 10

DIGITAL ULTRASONIC HUMIDIFIER WITH HUMIDISTAT

OPERATION and MAINTENANCE INSTRUCTION MANUAL. ADU-10 AseptiMini Pneumatic Portable Dental System

Dyna-Form Air. Alternating Pressure Relief Overlay System

Natural Mist Humidifier. Vicks. (Model V3500) IMPORTANT! READ AND SAVE THESE INSTRUCTIONS. Use and Care Manual

BIOVAC 500 SMOKE EVACUATOR

Triple Action Humidifier Model: AW600

Owner s Manual READ AND SAVE THESE INSTRUCTIONS

User Guide. Please read carefully before use. Your Hoover Guarantee. Your Hoover Guarantee. Contents. IMPORTANT Safety Reminders

Owners Manual. Model MINI-CLASSIC CT. Pure Water. Water Distiller

Operator's Manual. PCP October 2017

Operating Instructions

ULTRASONIC HUMIDIFIER WITH ANTIBACTERIAL TANK & BASE Instruction Manual and Warranty Information IM0012A READ AND SAVE THESE INSTRUCTIONS

FT982. Electric Professional Fogger. CAUTION: Read and follow all instructions

CF1400 DO NOT RETURN PRODUCT TO PLACE OF PURCHASE. IF YOU HAVE ANY QUESTIONS, CALL OUR TOLL FREE HOTLINE:

SM088 STEAM MOP INSTRUCTION MANUAL. For Authentic Monster Replacement Parts Call Model #: SM088 G-B-R

Canister Vacuum. OWNER S MANUAL Models CV520HB/HR/HY 120V., 60Hz., 10 Amps

DANGER Any appliance is electrically. WARNING To reduce the risk of IMPORTANT SAFETY INSTRUCTIONS

1600 Watt Professional

User Manual J0841X. Piezo Electric Ultrasonic Scaler / Polisher Combo Unit. Jorgensen Laboratories, Inc.

MODEL MC-UL592 MC-UL594 POWER SOURCE. 230V - 240V ~ 50 Hz MAX INPUT 1400 W 1500 W NOMINAL INPUT DIMENSIONS (W x L x H)

A few simple steps to ensure perfect water, every time.

OPERATOR'S MANUAL. IMPORTANT: READ OPERATOR'S MANUAL CAREFULLY Please fill out & return your warranty card! DP80405

IMPORTANT SAFETY INSTRUCTIONS

INSTANT HOT WATER DISPENSER

Transcription:

Caution: Federal Law restricts this device to sale by or on the order of a properly licensed practitioner. Integra Ultrasonic Tooth Scaler Model D680 Device Description: The Integra is an autotune ultrasonic tooth scaler that operates at either 25KHz or 30KHz. The Integra automatically detects whether the handpiece contains a 25KHz or 30KHz insert, and switches operating frequency to match. A dramatically expanded low-power range improves comfort during debridement. Intended use/indications: For removal of calculus and plaque during dental prophylaxis. Contraindications: Not for use if the patient or operator is wearing a pacemaker, or any other internal electronic device. Precautions: Caution: Water flow during use must be sufficient to cool handpiece and insert. Do not run this device dry. IMPORTANT: The water supply line to the scaler should always be turned off whenever the device is being connected, disconnected, or when not in use. Caution: Do not allow prolonged contact of tip with lips, cheeks, tongue or other soft tissues. Caution: Take precautions to prevent transmission of contagious diseases from ultrasonic aerosols. Caution: Use the sides of the tip to remove deposits from the tooth. Do not gouge the tooth with the point. Caution: Precautions should be taken to guard against entrance of water from any source into the transformer. The transformer should be located away from any source of water. Individualization of treatment: If patient or operator is pregnant, consult physician concerning advisability of using ultrasonic scalers. Like all scalers of this type, the handpiece produces a small magnetic field. It has been reported that this can cause some sensitive patient monitoring devices such as pulse oximeters to give inconsistent or inaccurate readings. Conformance to Standards: The Parkell Integra conforms to IEC60601-1 and CAN/CSA C22.2 No. 601.1. The device is CE marked certified to European Medical Device Directive (93/42/EEC). Parkell s quality system is certified to ISO9001/13485. Included in your unit are: 1-Scaler Unit with an attached handpiece, and 2-autoclavable handpiece sheaths 2-fluid bottles, 1-water line with male quick-connect coupler, and 1-detachable foot controller 2-In-line Water Filters (Stock No. D419) 1-Transformer with cable (Stock No. D408-120V, D403-230V) Instructions and Warranty registration card.

Dip Tube Easy-to-clean touch panel resists all standard disinfectants Digital fl uid and power controls mean fast, precise settings (Use the Smart Foot Pedal for hands-free power adjustment.) Easily-removable dual reservoirs give you a choice of irrigants Automatically adjusts operating frequency to match either 25KHz or 30KHz inserts Autoclavable handpiece sheaths Handpiece-holder mounts on left or right Ultrasonic inserts are not included with the Parkell Integra. However, this scaler will accept any scaler insert manufactured by Parkell, Cavitron/Dentsply, Hu-Friedy, or other Cavitron-compatibles. Specifications: Operating frequencies - 25KHz and 30KHz Operates at room temperature between 50 F and 95 F (10 C - 35 C) Weight: 35 oz. / Dimensions: 5 W x 6 H x 11.5 D Power: l20vac model - 60Hz, 120VA; 230VAC model - 50Hz, 120VA; 100VAC model - 50/60Hz, 120VA; Locating your Integra Scaler: Locate the device where the control panel will be easy to reach during scaling procedures. The scaler requires a grounded electrical outlet and optionally, a source of drinking quality, sediment free water at 20-50 psi. We suggest that an external water filter be used to minimize the frequent changes of the device s in-line water filter. Change the in-line filter when it becomes visibly discolored. The scaler and its separate transformer generate a minimal amount of heat. Avoid covering them, as this will prevent normal cooling by the air. Connecting your Integra to the Water Supply: Before plugging in the device, connect the scaler s water line to a supply of drinking quality water (20-50psi). The male quick-connect that comes on the end of the water hose is the same as that currently fitted to the Cavitron scaler. If you are replacing an older scaler that uses a different type of connection, remove the fitting from the old unit and attach it to the water line. Figure 1 External water source Rear bottle Foot Pedal Power External water fl ow regulator Internal bottle Front bottle External water connection

IMPORTANT: The water line to the scaler should be turned off whenever the device is not in use. If you are using the internal water bottles to supply fluid to the unit, fill them with the appropriate fluid such as water-diluted Peridex or mouthwash. Do not overfill the bottles. Take care to avoid spills when placing the bottles into the scaler. Connecting your Integra to the Electrical Connection: (See figure 1) The scaler must be powered via the Transformer Assembly supplied with the scaler. 1) Plug the transformer cable into the back of the scaler. 2) Plug the transformer into the nearest grounded outlet. 3) Plug the foot pedal into the back of the scaler, to allow it to control the device. Scaling Inserts: The Integra accepts any Parkell ultrasonic scaling insert as well as 25K or 30K inserts designed for Cavitron scalers. The scaler automatically adjusts the operating frequency to match the frequency of the insert in the handpiece (either 25KHz or 30KHz) without operator intervention. Brand new inserts may fit very tightly in the handpiece. Using a slight twisting motion when inserting and lubricating the O ring with water will make insertion easier. Old, worn or blunt inserts will provide poor scaling action and excess heat, and should be replaced. Turning the scaler on and off: Press any button on the touch panel. The display will light to indicate the scaler is on. To turn the scaler off, press the water minus ( ) button and while holding it down press the power minus ( ) button. The display will go out showing that the unit is off. The unit will remember any settings you made when it is turned on again. When you adjust the power using either the touch panel or the foot pedal, color-coded lights indicate the operating power. To conserve electricity the Integra will go to sleep if it is not used for about 30 minutes. The display will go out and the unit will not respond to the foot pedal. To wake up the Integra, simply press any of the buttons on the touch panel. Controlling the power: The Integra gives you three ways to adjust the scaling power: The Power (+ and ) buttons on the touch panel at the front of the unit, a button on the foot pedal and a Turbo mode. Use buttons on the touch panel to set the initial power for the scaling procedure at hand. Alternatively you can adjust the power by pressing the button on the top of the foot pedal. If you press and release (tap) the button, the power will increase by one step (i.e. from 4 to 5 as shown by the display). If you press the button and hold it down, the power will decrease in steps. In either case the color coded LEDs will show any change in the current power setting. (Note: The foot pedal power adjustment feature can be disabled. There is a small switch on the rear of the foot pedal to allow you to turn this on and off). A tip powered by the Integra will begin oscillating at much lower amplitude (power)

than a tip in a traditional scaler. This permits more comfortable debridement of sensitive patients. For short-term increases in power during the procedure, simply increase foot-pressure on the pedal. (The Turbo light will illuminate.) This will instantly boost scaling power to a point midway between the current setting and the scaler s maximum power. (In other words, increasing footpressure when the scaler is operating at a low setting will cause a major boost in power. Increasing foot-pressure when the scaler is already operating at maximum power will produce no significant increase.) To return to normal scaling power, lighten foot pressure. (The Turbo light will go out.) The Turbo mode allows you to scale at lower power than usual because you do not have to allow for occasional pieces of residual calculus. If you encounter hard-to-remove accretions during debridement, simply use the Turbo feature to eliminate them. (The Turbo mode does not change the water flow, so extended procedures may require a water adjustment.) Selecting the water source: The water source is selected using the two toggle switches on the front of the unit. One switch selects the source, either external or from the internal bottles. The other switch allows the user to choose which bottle is used as a source. Controlling the water flow: If you are using an external water source, turn the water control on the back of the unit counter-clockwise to increase the flow. If you are using the internal bottles as a source, the flow is controlled by using the Water (+ and ) buttons on the touch panel at the front of the unit. When you press either button the display will start to flash to indicate that it is now showing the water flow rate (1 is the lowest 7 the highest). As you hold the button the flow will change as shown by the display. When you release the button the display will return to showing the power level after a few seconds. If water does not flow through the insert when the foot pedal is depressed, the water passage in the scaling insert may be clogged or the bottle(s) may be empty. Infection control: When removing the autoclavable handpiece sheath, hold the handpiece at the base with one hand and turn the top counter-clockwise with the other, taking care not to twist the hose. The sheath will unscrew and slip off. Sheaths can be sterilized in any conventional steam autoclave, following manufactures standard procedures. DO NOT USE DRY HEAT OR CHEMCLAVE. (Note: The sheath must be washed in soap and water and rinsed in running water prior to autoclaving. Autoclaving does not remove debris that has accumulated on the sheath. Failure to adequately remove debris may result in inadequate sterilization.) Before replacing the sheath, lubricate the O-ring on the handpiece with petroleum jelly to assure a water-tight seal. Slide the sheath over the handpiece carefully to avoid damaging or dislodging this O-ring, and screw the sheath in place. (Do not overtighten.) The exterior case of the scaler may be wiped with a cloth dampened by any standard surface spray cleaner as desired. How to replace the filter disk (time required: 10 seconds.) The water filter disk should be replaced when it becomes very dark or so clogged with debris that water flow is impeded.

1) Disconnect the scaler from the water supply. 2) Remove the old filter disk by holding it in one hand and twisting off the hose connectors on either side. (They rotate in opposite directions.) 3) Install the replacement filter disk, and firmly tighten the connectors using hand pressure. 4) Reconnect the scaler to the water supply and check for leaks. (Note: If the fi lter becomes clogged and you do not have a replacement disk, the hose connectors can be locked together without the fi lter disk to allow emergency operation.) If you have any questions, please contact Parkell Service Dept. at (1-800-243-7446). Replacement Filters (10 per package): Stock No. D419 General comments on scaling: Ultrasonic scalers should be used only by trained, licensed professionals. Inserts and the handpiece sheath should be sterilized before each use. Do not operate unit without water flowing. Use of face mask is recommended when operating this device, to avoid inhalation of contaminated aerosol by the operator. High-volume suction to remove this aerosol from the dental environment (as close to the source as possible) is also recommended. Use the lowest effective scaling power for the case at hand. This keeps heat generation to a minimum. Before using the insert in the oral cavity, adjust the water spray over sink or cuspidor until desired mist is obtained. For maximum patient comfort, use a good flow of water which acts as coolant, lubricant and lavage to flush out debris. If patient is new to ultrasonic scaling, tell him/her what to expect. If procedure becomes uncomfortable, adjust power and water controls accordingly. Grasp handpiece in a comfortable manner, like holding a pen. To remove deposits from teeth, always use a light brushing stroke with the side of the insert tip in contact with the tooth surface. Excess pressure will not improve scaling action, but may cause heating and pain to the patient. Keep the scaler insert in motion at all times and use repeated gentle strokes to remove all tenacious deposits. Exercise caution near porcelain and resin restorations as they can fracture. Occasionally turn unit off to evaluate deposit removal with an explorer. If you test any scaler tip on your fingers while operating, it will feel uncomfortably hot. This is not a true test of how it feels on a tooth. To operate, adjust the power control to the minimum setting for the case at hand (start at zero). With a good water flow, the patient will not experience any uncomfortable heat. Common operator errors: Failure to keep insert tip parallel with the long axis of the tooth surface. Excessive use of hand pressure. Failure to use the lowest effective power setting for the case at hand. Insufficient water flow. Scaling with the point of the insert instead of the side. These failures in procedure can result in ineffective or incomplete removal of deposits and possible damage to the pulp or tooth structure. TROUBLESHOOTING Unit does not operate ( power on indicator does not light) Transformer cable disconnected from wall outlet or scaler Wall outlet not live Unit fuse has failed (Contact Parkell) Insert does not vibrate properly Faulty, damaged or worn insert Insert not correctly seated in handpiece Power control not correctly adjusted Excess heat at handpiece or tip of insert Faulty, damaged or worn insert Too little water flow (insert blocked)

Power too high for the amount of water Excessive hand pressure applied No water spray Water line blocked or kinked Water filter clogged Water canal in insert blocked No water supply Water bottles empty Insert does not go into handpiece Lubricate O ring with water and use twisting motion to seat insert Insert falls out of handpiece or water leaks from front of handpiece O ring damaged or worn. Replace O ring on insert. Additional Specifications: Protection Against Electric Shock Class 1, Type B applied part Equipment not suitable for use in the presence of a flammable anesthetic mixture (when used with air or with oxygen). Protection Against Ingress of Liquids Foot Switch & Scaler - IPX1 (drip-proof), Power Supply - IPX0 (Ordinary) Mode of Operation of Equipment Intermittent: 66% duty cycle, 10 minutes on, 5 minutes off. Operating conditions: 15-35 C, 10-80% RH (non-condensing) Transport and Storage conditions: 10-40 C, 10-80% RH (non-condensing) As an additional safety feature, the microprocessor will reduce the output power in the unlikely event the device is left running, other than in normal usage. This equipment produces electromagnetic energy and may cause interference with other electronic devices. Should this occur, changing the positioning or location of the device may be necessary. The power supply line cord is the AC power disconnect device for the scaler. The symbol on the scaler indicates the Protective Earth connection. The symbol or is intended to alert the user to the presence of important operating and maintenance (servicing) instructions in the literature accompanying the scaler. Service: The handpiece sheath, power supply, foot pedal, water filter, dip tubes, bottles and ultrasonic inserts are all detachable and may be replaced by the user. Contact Parkell for a list of authorized inserts. Otherwise, there are No User Serviceable Items in this scaler, including the foot controller. For service, return the unit freight prepaid and carefully packed to Parkell Repair Dept., 300 Executive Drive, Edgewood, NY 11717, with an explanation of problem. During the warranty period, the unit will be repaired and returned to you. After the warranty period, we will call with an estimate and request approval. If you have any questions or a problem with the installation or use of your scaler, please call our Toll-Free Technical Support Service at 1-800-243-7446 from 9am to 5:30pm Eastern time (outside USA and Canada, call 631-249-1134). European Authorized Representative (Not a dealer/distributor): EMERGO EUROPE Molenstraat 15, 2513 BH, The Hague, The Netherlands Tel: +31 (0)70 345 8570 Fax: +31 (0)70 346 7299 300 Executive Drive, Edgewood, NY 11717 Toll-Free: 1-800-243-7446 Phone: 631-249-1134 Fax: 631-249-1242 E-mail: info@parkell.com www.parkell.com Made in U.S.A. A00545revB0313