REPUBLIC OF NAMIBIA MINISTRY OF HEALTH AND SOCIAL SERVICES REPORT BY: MS VERA UUSHONA TITLE: INSPECTION OF CONVENTIONAL DIAGNOSTIC X-RAY FACILITIES

Similar documents
Victoreen & 660-8

25 TEXAS ADMINISTRATIVE CODE Radiation Safety Requirements for Analytical and Other Industrial Radiation Machines

RaySafe Xi Classic. A complete system for X-ray measurements

Effect of Different Types of Fabricated Anti-scatter Grids in Reducing Backscattered Radiation

RAMSES: THE LHC RADIATION MONITORING SYSTEM FOR THE ENVIRONMENT AND SAFETY

Design, Testing and Performance of Exposure Devices RD/GD-352

APPLICATION FOR AUTHORISATION - PART 1

Co-60 irradiator for therapy level calibrations at SSDLs

Carnegie Mellon University Radiation Safety Committee

Guidance Document Design, Testing and Performance of Exposure Devices RD/GD-352

Integration of safety and security for effective control and protection of radioactive sources

Use of control and analytical X-ray apparatus

REGULATORY GUIDE 4.4 TEXAS DEPARTMENT OF STATE HEALTH SERVICES RADIATION SAFETY LICENSING BRANCH (RSLB) P.O. Box Austin, Texas

CALIBRATION OF A GAMMAMED 12i 192 Ir HIGH DOSE RATE SOURCE

Regulatory Challenges While Harnessing Societal Benefits of Radiation Sources in India

Appendix C X-Ray Diffraction User s Guide and Approval Form

The RSO will submit a copy of the revised manual to the RSC for approval when the revised manual has substantial changes.

Environmental Protection Agency, Office of Radiation Protection and Environmental Monitoring.

Measuring Couch Motion Accuracy with X-ray Beam Fluence

University of Pennsylvania. Environmental Health and Radiation Safety (EHRS) X-ray Diffraction User Guide

X-Ray Protection Program

Situation on Radon Assessment in Latvia

Document Preparation Profile (DPP)

User Manual X-Check FLU L981319

Minnesota Rules, Chapter 4732 X-Ray Revision

QUALITY ASSURANCE AND SAFETY AT A TID RADIATION TEST LABORATORY

DEVELOPMENT OF LARGE AREA PLASTIC SCINTILLATION DETECTOR FOR RADIOACTIVE CONTAMINATION MONITOR

Understanding total measurement uncertainty in power meters and detectors

Radiation Control and Monitoring System on the HTTR

CURRENT INDOOR RADON SITUATION IN LITHUANIA

California State University Bakersfield Radiation Safety Program Revised November 2017

INTERNATIONAL STANDARD

Montanez, Boutet, Fox, Van Campen, Hays, Woods

Forward-scattered radiation from the compression paddle should be considered when average (or mean) glandular dose is estimated

NZQA registered unit standard version 1 Page 1 of 5. Install fixed gas detection apparatus in hazardous areas

Installation and Operation of Model CS X-ray Cabinet. 11 Emery Avenue Randolph, NJ Tel.:

EVALUATION OF A NEUTRON-PHOTON SHIELD FOR TRANSURANIC (TRU) WASTE CONTAINERS

The Regulatory Perspective on Radiation Protection in Canadian Uranium Mines

Protecting our Borders while Ensuring Radiation Safety

SIEMENS ULTRAMAT 6 IR CARBON MONOXIDE ANALYZER METHOD VALIDATION FOR TESTING CARBON MONOXIDE IN NITROGEN, NF

TOWARD ENSURING SAFETY AND SECURITY IN RADIOGRAPHY WORK ENVIRONMENT FOR WORKERS AND PUBLICS

INTERNATIONAL STANDARD

When Precision is not good enough

Minnesota Rules, Chapter 4732 X-Ray Revision

Industrial Radiography (IR) General overview. R. Berden

DOWNLOAD OR READ : RADIATION PROTECTION PDF EBOOK EPUB MOBI

FEATURE ARTICLE. The PLCA-800 Series Inline Particle Sensor for Chemical Solutions. Yoshihito Yuhara. Abstract

Using the GammaRAE II Responder Personal Radiation Detector and Dosimeter in One. Firmware Version 3.00 D August 2006

HEALTH AND SAFETY PROGRAM 406 IONIZING/NONIONIZING RADIATION PROTECTION

A new fully digital system for RT inspection of metal tube to tube sheet joints of heat exchangers

Competency: Critically evaluate the function of the ionisation chamber in the Linear Accelerator and its importance for correct treatment delivery

Radiation Protection Program Updates

Radiation Safety Guidelines for Non-Users

RADIATION PRODUCING MACHINE SAFETY MANUAL

Analytical X-Ray Safety Program

VERIFICATION OF IONISING RADIATION MONITORS FOR PRACTICAL USE IN HOSPITALS. C. Montes, F. Sáez, C. Martín, P. Collado, C. J. Sanz, E.

EC - RADON Latest strategies and draft regulations

Qualification Testing of Nuclear High Voltage Splice (NHVS) Title Pages: 49. Report Number: EDR-5391 Rev. 0 06/15/05. Approved by: Signature: Date:

Radiation Emitting Devices Research X-ray Safety Program

Radioactive Contamination Monitor

ISO INTERNATIONAL STANDARD. Coating powders Part 13: Particle size analysis by laser diffraction

A Review of EU Legislation on EP&R

Atomic Energy Licensing Board

RADON SURVEY IN KALAMATA (GREECE) Medicine, London SW7 2AZ U.K. Greece. Republic

A. General Safety LAB/ROOM: INSPECTED BY: DATE:

Review of options for cabin dosimetry & operation. Helen Amatiello & Colin Martin Health Physics Service Glasgow

RESEARCH PAPERS FACULTY OF MATERIALS SCIENCE AND TECHNOLOGY IN TRNAVA, SLOVAK UNIVERSITY OF TECHNOLOGY IN BRATISLAVA, 2016 Volume 24, Number 39

X-ray Inspection System Discover What You've Been Missing... DETECTION PERFECTION

Corporate Standard Radiation Safety HB#

Freiburger Materialforschungszentrum SG Material Characterization & Detector Technology. Albert-Ludwigs-Universität Freiburg

Dosimetry issues related to the SSRT project at the ESRF

Efficiency at Hand. Digital Radiography GF50 Series CT-GF EN. Tube 300kHU (210kJ) 150kHU (105kJ)

AeroDR X30. Digital Mobile X-ray System

Model : LPSA-40A Please read this manual carefully before using the instrument Labnics Equipment

TSI AEROTRAK PORTABLE PARTICLE COUNTER MODEL 9110

ESSENTIAL REQUIREMENTS CHECK LIST

Quality Assurance of Radiation Protection Monitoring Instruments in India

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005

LIGO Caltech 40 Meter Laboratory Laser Safety Plan

Continuous and Point Level Measurement. Radiometric Sensors for Liquid and Solid Level Measurement

THEORY OR OPERATION 2 SENSOR UNIT - MECHANICAL 6 SENSOR UNIT - ELECTRICAL 8 CONTROL UNIT - MECHANICAL 9 CONTROL UNIT - ELECTRICAL 9 OPTIONS 11

818 Series. Low-Power Detectors ABSOLUTE SPECTRAL RESPONSIVITY (A/W)

NOTE: The color of the actual product may differ from the color pictured in this catalog due to printing limitation.

NATIONAL COMPETENCY STANDARD

HEAT REMOVAL TESTS FOR THE HIGH PERFORMANCE VAULT STORAGE SYSTEM OF SPENT NUCLEAR FUEL

Radon in thermal waters and radon risk in chosen thermal water spas in V4 countries - preliminary results

LHCb Rich Detectors Control and High Voltage Systems

Fundamentals is subdivided into Sources of Radiation, Biological Effects, Mathematics, Chemistry, Physics, and Units and Terminology.

Infrared Thermometer with 12:1 Distance-to-Sight Ratio and Temperature Alarm

THE REPUBLIC OF UGANDA ATOMIC ENERGY ACT, 2008 SCHEDULE 1 FORM 2A

Safety Manual Section 15 Radiation Safety

AM-806. Waste/Door Radiation Monitor OPERATIONS MANUAL

Small Controlled Area Radiography Fundamentals and Technology Advancements. J. R. Lapinskas, PhD and R. L. Kelly

NATIONAL INSTITUTES of HEALTH FUME HOOD TESTING PROTOCOL. For CONSTANT VOLUME FUME HOODS

Hygrometry. The art or science of humidity observation

Beam Loss Position Monitoring with Optical Fibres at DELTA

Instruction Manual. gray Mounted Detectors. for. greenteg AG Technoparkstrasse Zürich, Switzerland T: F:

American Nuclear Society (ANS)

It only takes an. instant to go DR. DR Retrofit: no limits to retrofitting your X-ray department with DR AED detectors

AEC RADIATION SAFETY GUIDE SAFE USE OF INDUSTRIAl DEVICES CONTAINING RADIOACTIVE SOURCES

Transcription:

REPUBLIC OF NAMIBIA MINISTRY OF HEALTH AND SOCIAL SERVICES REPORT BY: MS VERA UUSHONA TITLE: INSPECTION OF CONVENTIONAL DIAGNOSTIC X-RAY FACILITIES

ABSTRACT TITLE: Inspection of conventional diagnostic x-ray facilities in Namibia Inspectors: Ms NV Uushona and Mrs E Embumbulu National Radiation Protection Services, Ministry of Health and Social Services. Introduction The National Radiation Protection Services (NRPS) is the competent unit responsible for the regulation of all activities involving ionising radiation. The enabling legal framework for this function is the Hazardous Substance Ordinance of 1974 and the National Radiation Protection Policy of 1994. The enabling legislation is expected to be replaced by the Atomic Energy and National Radiation Protection Act in the 2009/10 financial year. The inspection mainly consists of the x-ray-unit optimal performance and the general radiation safety provisions. Method The inspection is performed using a checklist which lists the general radiation provision such as radiation warning light, shielding, protective clothing etc. Then different tests for the x-ray unit were performed which are: kvp accuracy and reproducible test, the beam quality, beam collimation and alignment and the radiation dose output. The RMI 241 meter was used to measure the kvp and the Rad check was used to measure the radiation output and the halfvalue layer. The beam alignment and collimation was done using the collimator and beam alignment tools. Results For the purpose of compliance and enforcement 53 out of 57 conventional diagnostic x-ray facilities were inspected during March 2008 and April 2009, which mount to 92% coverage. 26% of the facilities failed the kvp test, while radiation warning lights followed with 20%. In most cases the radiation leakage and half value layer were within the recommend limits. Conclusion As a result of the inspection performed recommendations were issued to the management of the facility in order to implement corrective measures and 6 facilities that were inspected and not authorized carried out the corrective actions recommended. The division is thus recommended to improve its enforcement mechanism. i

Table of content 1. Introduction 2 2. Method and apparatus 2 3. Results 5 4. Conclusion 6 1

1. Introduction The National Radiation Protection Services (NRPS) is the competent unit responsible for the regulation of all activities involving ionising radiation. The enabling legal framework for this function is the Hazardous Substance Ordinance of 1974 and the National Radiation Protection Policy of 1994. The primary role of the NRPS is to ensure exposure to ionising radiation is controlled and monitored in workplaces, environment and living places. This is achieved mainly through regular inspection and authorization of facilities utilising or producing sources of ionising radiation and radioactive material. The enabling legislation is expected to be replaced by the Atomic Energy and National Radiation Protection Act in the 2009/10 financial year. The inspection mainly consists of the x-ray-unit optimal performance and the general radiation safety provisions. 2. Method and apparatus The inspection complies of the following test. 2.1 The Kilovoltage (kvp) accuracy test Image quality and patient dose are dependent on any variation in the generator kilovoltage (kv) of the x-ray set. Therefore an accurate kv calibration is required. A non-invasive tube voltage check over the whole used kv-range is performed with the RMI 240 multi-function meter (Figure 2.1). Firstly the RMI meter was set at 100 cm from x-ray tube focus, and was centered using the laser. 20 mas was set on the machine control panel. The kvp was then measured from 50-125 kvp in increment of 10. At every set kvp, the measured kvp was recorded. The percentage difference between the set and measured kvp was calculated and should be within ± 5%. Figure 1: RMI Multi functional meter 2

2.2 The kvp reproducibility test This test checks that the kvp will stay the same from exposure to exposure. The set up is the same as above, but at a fixed kvp of 70. Five exposures were recorded and the coefficient of variation (CoV) which is the standard deviation/mean of the 5 readings was recorded. The CoV must be less than 2%. 2.3 Tube Output Measurement This test checks that the radiation output [mgy/mas] remains constant as the ma is varied. For this test the dose Victoreen Rad check meter was used (Figure 2.2). Figure 2.2: Victoreen Rad check meter The Victoreen meter was placed at distance of 75cm from the tube focus. Some lead vinyl was placed under the detector to standardize backscatter. The kvp was fixed at 70, while the mas was varied from 20-100 mas, in increments of 10. At every mas setting the radiation dose was recorded. For each exposure, the radiation dose / mas was calculated. The maximum and minimum value of dose/mas was identified and the following was calculated (max-min)/ (max+min). This should be less then 0.1 2.4 Half Value Layer (HVL) Measurement The half-value layer measures the quality of the x-ray beam. The apparatus used are the Aluminum (Al) attenuator set, Victoreen meter and lead vinyl. The meter was placed at 75cm from the tube focus on the lead vinyl to standardize back scatter. The kvp was set at 80 and fixed mas of 50mAs. The x-ray beam was collimated to the size of the meter. Three exposures were recorded taking the dose. After a 1mm of Al was added to the beam, and the dose was measured. Al attenuator ware added and the dose recorded until the dose has fallen to below 50% of the initial unattenuated value. 3

A graph with the dose against the aluminum attenuator thickness was plotted. From the graph, the thickness of Al required to reduce the unattenuatted dose by 50% is the Half Value layer. 2.5 Beam Collimation The following apparatus were used: beam alignment tool (Figure 2.3), collimator tool (Figure 2.4), cassette and film. Figure 2.3 Beam alignment tool Figure 2.4: Collimator tool The collimator tool was placed at a distance of 100cm from the tube focus with a cassette underneath it. The beam alignment tool was then placed at the center on the collimator tool. The collimator shutters were adjusted so that the edges of the light field coincide with the rectangular outline on the collimator tool. Then the cassette was exposed. If the x-ray field falls just within the image of the rectangular frame there is good alignment. If an edge of the x-ray field falls on the first spot, ± 1cm, on either side of the line it shows that the edges of the x-ray field are misaligned by 1%. 2.6 Other inspection parameters Apart from the equipment performance an inspection checklist is used to check other component of radiation protection, these include: Leaking at the entrance door and control panel Radiation warning light Local rules of the department including radiation warning notices Protective clothing Monitoring of the radiation workers And the departments quality assurance program Public protection Medical exposure 4

3 Results For the purpose of compliance and enforcement 53 out of 57 diagnostic x-ray facilities were inspected. Figure 3.1 (below) indicates the total number of facilities inspected and authorized. Total conventional diagnostic facilities 120 100 100 92.98 80 60 59.65 Series1 40 20 0 Total conventional diagnostic facilities Total conventional dignostic facilities inspected total facilities licensed during the period Figure 3.1: Total conventional diagnostic facilities Figure 3.2 below shows that 26% of the facilities failed the kvp test, while radiation warning lights followed with 20%. In most cases the radiation leakage and half value layer were within the recommend limits. % of facilities deviating from compliance criteria warning notices 9% Beam alignment 14% radiation leakage 6% kvp 25% Warning light 20% HVL 9% radiation output 17% Figure 3.2: Percentage of facilities deviating from compliance criteria 5

4 Conclusion The National Radiation Protection services have inspected 92% of all the conventional diagnostic x-ray facilities during the period April 2008 and March 2009. As a result of the inspection performed recommendations were issued to the management of the facility in order to implement corrective measures and only 6 facilities that were inspected and not authorized carried out the corrective actions recommended. The division is thus recommended to improve its enforcement mechanism. 6