EMAX 2 PLUS AND XMAX CLEANING AND STERILIZATION BASKET. Mechanical/Automated Cleaning and Sterilization Guide in a Variety of Applications

Similar documents
Manual Pre-disinfection/Cleaning and Sterilization Instructions for SATELEC Tips and Files

Manual Pre-disinfection/Cleaning and Sterilization Instructions for SATELEC Tips and Files

IMPORTANT INFORMATION ON THE DEPUY SYNTHES

CONVENTUS CAGE TM - DR Procedure Instruments Instrument Tray. Equipment Cleaning and Sterilization

Manual Pre-disinfection/Cleaning and Sterilization Instructions for SATELEC wrenches

CONVENTUS CAGE TM PH Procedure Instruments Instrument Tray Cleaning and Sterilization

CONVENTUS CAGE TM PH Procedure Instruments Instrument Tray Cleaning and Sterilization

INSTRUCTIONS FOR CLEANING AND STERILIZATION OF ALL B.BRAUN AESCULAP HAND HELD SURGICAL INSTRUMENTS AND POWER PRODUCTS (UNITED KINGDOM REQUIREMENTS)

CLEANING INSTRUCTIONS

Instructions for Use Spinal Punches IFU

SKLAR COATED SURGICAL INSTRUMENTS RECOMMENDED CARE & CLEANING INSTRUCTIONS

Cleaning & Sterilization

Integra Salto Talaris and Integra XT Non-Sterile Device Reprocessing. Instructions for Use. 4. Manual Cleaning Instructions

Visao HIGH-SPEED OTOLOGIC DRILL CATALOG

Operating / User Manual

Accessories. Maybachstraße. Symmetry Fax: Revised 10/15

STERITITE INSTRUCTIONS FOR USE

User s Manual. Small Electric Drive. Compact and powerful system for use in hand and foot surgery.

Aesculap Implant Systems

AS Medizintechnik GmbH Sattlerstrasse 15, Tuttlingen, Germany Tel 07461/ Fax 07461/

Aesculap Implant Systems

User Guide for Reprocessing of the SUPER VIEW Wide Angle Viewing System Sterilizable Components and Accessories

Made in USA. Sunoptic Technologies 6018 Bowdendale Avenue Jacksonville, FL 32216, USA (904)

V. Mueller. V. Mueller Reusable Devices: Cleaning and Sterilization Guide USA. Rx Only

RECOMMENDATIONS FOR CLEANING DEVICES

Recommendations for cleaning, decontamination and sterilisation. devices

CLEANING AND HANDLING OF WRIGHT INSTRUMENTS

Instructions for care, maintenance, cleaning and sterilization of Smith & Nephew orthopaedic devices

FLASH PAK STERILIZATION CONTAINER SYSTEM

Rigid Reusable & Single use Sigmoidoscopes, Anoscopes, and Accessories ENGLISH 1. Perform reprocessing immediately after each use.

INSTRUCTIONS FOR CLEANING AND STERILISING HUMECA MEDICAL EQUIPMENT FOR BOTH ELECTRIC AND MANUAL DEVICES

Air-driven Universal Power Tool System for Traumatology, Endoprosthetics, and Spine. Compact Air Drive II. Instructions for Use

INSTRUCTIONS EVIS EXERA II DUODENOVIDEOSCOPE

Instructions for Use: Fiberscope and LED Light Source

EMAX 2 PLUS SYSTEM High speed performance and power in a variety of applications

Processing instructions for the Ligamys suture forceps

ON TRACK Reprocessing In-service / Competency for CF/PCF/GIF Endoscopes (EVIS, EXERA, EXERA II)

When you sterilize an instrument set, you know what the standards are.

I.F.U Re/Processing Reusable Medical Devices

Reusable Laryngoscope Blades and Handles

DIRECTIONS FOR USE. Keeping you insafehands to avoid needlestick injuries. see for demo video

X99 Series. Dental Handpiece. Instruction For Use

Important information

ComPact Air Drive II USER S MANUAL

INSTRUCTIONS EVIS EXERA II DUODENOVIDEOSCOPE

SterilContainer S for use in the Steris V-PRO 60 Low Temperature Sterilization System

SteriTite - The Container System of Choice SteriTite

Cavitator Ultrasonic Cleaner Instruction Manual Model 20G

Christina Bradley Laboratory Manager Hospital Infection Research Laboratory Queen Elizabeth Hospital Birmingham, UK

INSTRUCTIONS FOR USE & STERILIZATION FOR INSTRUMENTS

Instructions for Use: Endoplus Modular Handle Assemblies, Multi-Use Scissors Inserts, Reusable Inserts

المملكة العربية السعودية وزارة الصحة اإلدارة العامة لمكافحة عدوى المنشئات الصحية

MERCURY HD-22 AUTOSCRUBBER Safe Operations

Attendance Sheet. (List each model reviewed during this session) Print Name Signature Title. Print Name Signature Title.

Cavitator Ultrasonic Cleaners

2015 Myco Industries, Inc Fax:

Cavitator Ultrasonic Cleaner Instruction Manual

Table of Contents. English

Operator s Guide. Autoclavable Internal Handles with Integrated Paddles Rev. M

Instruction of use. Flexible Naso-Pharyngo- Laryngofiberscope,

Operator s Manual. Model G32-S Model G32-E Disinfection Soak Stations

SteriTite - The Container System of Choice SteriTite

OPERATION MANUAL. LED Coupling OM-TR355E. Please read this operation manual carefully and file for future reference.

Operator s Manual. The Bullet Blender 50 BB50-AU, BB50-DX. Contents. Congratulations!

SteriTite - The Container System of Choice SteriTite Instructions for Use

idrive Ultra Powered Stapling System

Caution: Federal U.S. law restricts this device to sale by or on the order of a physician.

INSTRUCTIONS EVIS EXERA GASTROINTESTINAL VIDEOSCOPE

RAP-hip Hip stem extraction set

Cleaning and Sterilization for Instruments & Posts

PLEASE READ BEFORE USE AND SAVE. Ultrasonic. with Ionization. INSTRUCTION MANUAL MODEL HUNi

PERFORM Operating Document

Cleaning and Sterilisation for Burs, accessories and filling material

MODEL 7000 SUCTION UNIT

SAVE THESE INSTRUCTIONS

Flexible Nasopharyngoscope. Operator Manual. Manufactured by Henke-Sass, Wolf Distributed by Anthony Products, Inc.

OPERATION MANUAL. (Cleaning, Disinfection and Storage) Endoscope ED-530XT

P.N User Manual STAINLESS STEEL ELECTRIC KETTLE 5964/5965 Series MODELS

CLEANING FOR INFECTION CONTROL Ultrasonic Cleaners. Ultrasonic Cleaners. Copyright Quality Medical Innovations P/L & Majac Medical Products P/L

ASPIRE Laboratory Aspirator

Schuco byallied MEDICAL ASPIRATOR. USER'S MANUAL A Caution

Ultrasonic From delicate microsurgery instruments to heavy duty sets

Pocket Hardmount. Preventative Maintenance.

Dome Kettle. polished stainless steel body BTDK120

Ultrasonic Cool Mist Humidifier

ASSESSMENT REPORT ENDOZIME INSTRUSPONGE

01550-RevG IMPORTANT INSTRUCTIONS WEBSITE MANUAL

Processing Instructions. Cleaning so easy.

Cavitator Ultrasonic Cleaners

USER GUIDE DEEP FRYER WITH OIL FILTRATION SYSTEM

Instructions for Use and Reprocessing Instructions

Use and Care Guide. Over the Patient Units and Swing Mount Units

Earn BISSELL Rewards! Register your

ROHO MOSAIC Cushion Operation Manual

Best Practice Guidelines For Cleaning, Disinfection and Sterilization of Critical and Semi-critical Medical Devices

OPERATING INSTRUCTIONS AND MAINTENANCE MANUAL

ELECTRIC KETTLE Model BVSBKT1703 Series. User Manual. QUESTIONS? COMMENTS? IN CANADA: P.N Rev.A

WORLD LEADERS IN ADVANCED SURGICAL INSTRUMENT REPROCESSING TECHNOLOGY

Operator s Manual. Model G14TC-3 Disinfection Soak Station for Transesophageal Ultrasound Probes

Rapid Heat Sterilizers with 6, 8, and 12 Minute Sterilization Cycle Times

Transcription:

EMAX 2 PLUS AND XMAX CLEANING AND STERILIZATION BASKET Mechanical/Automated Cleaning and Sterilization Guide in a Variety of Applications

TABLE OF CONTENTS MANUAL PRE-CLEANING High Speed Handpieces 2 Attachments 5 MECHANICAL/AUTOMATED CLEANING High Speed Handpieces and Attachments 8 STERILIZATION High Speed Handpieces and Attachments 12 LOADING CONFIGURATION High Speed Handpieces and Attachments 15 ORDERING INFORMATION 16 EMAX 2 Plus and XMAX Cleaning and Sterilization Basket DePuy Synthes Companies 1

MANUAL PRE-CLEANING HIGH SPEED HANDPIECES Detergents x A neutral ph enzymatic detergent is recommended. x Alkaline detergents with a maximum ph of 11 may be used but user must demonstrate their effectiveness. x The process, detergent, and any additives must be discussed with detergent manufacturer. Caution: Do not use any alkaline detergent with a ph greater than 11. Alkaline detergents attack grease and seals which can increase wear and cause device to malfunction. Please note use of alkaline detergents can cause color to fade, but this does not impair function of device. 1 Preparation Assemble all necessary supplies including Attachment Cleaning Brush (ACB), sponge, lint-free cloth, and soft bristle brush. Remove the attachment and dissection tool from the handpiece (Figure 1). Electric systems only: Place the attached cap onto the console connector end of the handpiece (Figure 2). Figure 1 Figure 2 2 Rinse Rinse the handpiece with distal end pointing down under running cold deionized, distilled, or purified water for a minimum of 2 minutes. Use a sponge, soft lint-free cloth or soft bristle brush to assist in removing gross soils (Figures 3 and 4). Caution: Do not use high pressure water or air. Do not allow running water to enter distal end of handpiece. Figure 3 Figure 4 Distal End 2 DePuy Synthes Companies EMAX 2 Plus and XMAX Cleaning and Sterilization Basket

Manual Pre-Cleaning High Speed Handpieces 3 Manipulate moving parts Manipulate the handpiece knurled knob under running cold deionized, distilled, or purified water to loosen or remove gross soil (Figure 5). Figure 5 4 Clean Prepare a cleaning solution using neutral ph enzymatic detergent per the manufacturer s directions for correct temperature, water quality (i.e., ph, hardness) and product concentration/dilution. With distal end pointing down, spray and wipe the handpiece with the cleaning solution for a minimum of 2 minutes. With distal end pointing down, rinse the handpiece thoroughly under running cold deionized, distilled, or purified water for a minimum of 2 minutes while manipulating the knurled knob (Figures 6 and 7). Figure 6 Figure 7 Use a sponge, soft lint-free cloth or soft bristle brush to remove all visible soil. Place locking mechanism in the SAFE position prior to performing the following step (Figure 8). Hold the handpiece at a downward angle and insert the Anspach Cleaning Brush (ACB) wetted with cleaning solution into the distal end of the handpiece (Figure 9). Figure 8 Figure 9 Warning: ACB is single use and should only be used to clean 1 handpiece and associated attachments. Caution: Do not immerse. Do not allow running water to enter distal end of handpiece. Do not insert ACB beyond last bristle. Do not force ACB into handpiece. Do not use pipe cleaners. EMAX 2 Plus and XMAX Cleaning and Sterilization Basket DePuy Synthes Power Tools 3

Manual Pre-Cleaning High Speed Handpieces 5 Rinse thoroughly Rinse the handpiece thoroughly under running hot deionized, distilled, or purified water for a minimum of 2 minutes (Figure 10). Place locking mechanism in SAFE position prior to performing the following step (Figure 11). Hold the handpiece at a downward angle and flush inside of the distal end using a syringe or pipette filled with hot deionized, distilled, or purified water (Figure 12). Place the knurled knob in the RUN position (Figure 11). Figure 10 Figure 11 Caution: Do not use high pressure water or air. Do not immerse. Do not allow running water to enter distal end of handpiece. Figure 12 6 Inspect Visually inspect the handpiece and repeat Steps 2 through 6 until no visible soil remains. Warning: Manual pre-cleaning must be followed by mechanical/automated cleaning. 4 DePuy Synthes Companies EMAX 2 Plus and XMAX Cleaning and Sterilization Basket

MANUAL PRE-CLEANING ATTACHMENTS Detergents x A neutral ph enzymatic detergent is recommended. x Alkaline detergents with a maximum ph of 11 may be used but user must demonstrate their effectiveness. x The process, detergent, and any additives must be discussed with detergent manufacturer. Caution: Do not use any alkaline detergent with a ph greater than 11. Alkaline detergents attack grease and seals which can increase wear and cause device to malfunction. Please note use of alkaline detergents can cause color to fade but this does not impair function of device. 1 Preparation Assemble all necessary supplies including Anspach Cleaning Brush (ACB), Small Diameter Cleaning Brush (SDCB), sponge, lint-free cloth and soft bristle brush. Remove the attachment and dissection tool from the handpiece (Figure 1). Note: Red storage cap is provided with Perforator Driver (CSR60) to prevent damage to shaft. Cap is removed for cleaning, washing and sterilization (Figure 2). Cap should be used for nonsterile storage purposes only. Figure 1 Figure 2 2 Rinse Rinse the attachment under running cold deionized, distilled or purified water for a minimum of 2 minutes. Use a sponge, soft lint-free cloth or soft bristle brush to assist in removing gross soil (Figures 3 and 4). Caution: Do not use high pressure water or air. Do not rinse attachment with saline solution. Figure 3 Figure 4 EMAX 2 Plus and XMAX Cleaning and Sterilization Basket DePuy Synthes Power Tools 5

Manual Pre-Cleaning Attachments 3 Manipulate moving parts Manipulate all moving parts such as release sleeves under running, cold, deionized, distilled, or purified water to loosen or remove gross soil (Figure 5). Figure 5 4 Clean Prepare a cleaning solution using neutral ph enzymatic detergent per the manufacturer s directions for correct temperature, water quality (i.e., ph, hardness) and product concentration/dilution. Fully immerse the attachment in cleaning solution in a suitable container and manually agitate for 2 minutes (Figure 6). Use a sponge, soft lint-free cloth or soft bristle brush to remove all visible soil. Figure 6 Figure 7 Gently use ACB or SDCB wetted in cleaning solution to clean cannulations of attachment. The brushes may be inserted from the distal or proximal end of the attachment (Figure 7). Warning: ACB and SDCB are single use and should only be used to clean one set of attachments. Caution: Do not force ACB or SDCB into or through attachment. Do not use ACB or SDCB to clean the MA-D20 and MA-DRIVER. For Minimal Access Attachment Bearing Sleeves, use SDCB only. Do not use pipe cleaners. Note: DO NOT IMMERSE as marked on the attachment is for Operating Room personnel only. The SDCB may be inserted from the distal or proximal end of the bearing sleeve. 6 DePuy Synthes Companies EMAX 2 Plus and XMAX Cleaning and Sterilization Basket

Manual Pre-Cleaning Attachments 5 Rinse thoroughly Rinse the attachment thoroughly under running hot deionized, distilled, or purified water for a minimum of 2 minutes (Figure 8). Use a syringe or pipette filled with hot deionized, distilled, or purified water to flush lumens or channels (Figure 9). Caution: Do not use high pressure water or air. Do not rinse attachment with saline solution. Figure 8 Figure 9 6 Inspect Visually inspect the attachment and repeat Steps 2 through 6 until no visible soil remains. Warning: Manual pre-cleaning must be followed by mechanical/automated cleaning. EMAX 2 Plus and XMAX Cleaning and Sterilization Basket DePuy Synthes Power Tools 7

MECHANICAL/AUTOMATED CLEANING HIGH SPEED HANDPIECES AND ATTACHMENTS Detergents x A neutral ph enzymatic detergent is recommended. x Alkaline detergents with a maximum ph of 11 may be used but user must demonstrate their effectiveness. x The process, detergent and any additives must be discussed with detergent manufacturer. Caution: Do not use any alkaline detergent with a ph greater than 11. Alkaline detergents attack grease and seals which can increase wear and cause device to malfunction. Please note use of alkaline detergents can cause color to fade but this does not impair function of device. Bearings and seals of all devices are exposed to greater stress during mechanical/automated cleaning than manual cleaning. If devices are mechanically cleaned they must be sent to a DePuy Synthes Power Tools Service Center at least once a year for maintenance. Mechanical/Automated Cleaning Guidelines Anspach devices can only be mechanically cleaned under the following conditions: x Cannulations must be pre-cleaned with appropriate brushes as described in pre-cleaning instructions. x Devices must not be cleaned using ultrasonic equipment. x Handpieces must not be immersed in cleaning solution but may be sprayed with it. x Fully demineralized (deionized) water must be used during final rinse to prevent corrosion and spots. x Devices must be cleaned immediately after use to ensure blood and tissue never dry on them. x National regulations must be observed. x You must follow standard hospital procedures. x You must follow manufacturer s recommendations for any detergents, disinfectants, and washing equipment used. 8 DePuy Synthes Companies EMAX 2 Plus and XMAX Cleaning and Sterilization Basket

Mechanical/Automated Cleaning High Speed Handpieces and Attachments 1 Loading the Basket Load the devices in the basket by securing them in the appropriate brackets as marked on the basket (Figure 1). Reference loading configuration pictures on Page 15 for proper placement and positioning of devices. Where applicable, distal end of devices should be angled downward. Warning: Red protective cap for Perforator Driver (CSR60) must be left off during mechanical/automated cleaning (Figure 2). Figure 1 Figure 2 Caution: Connector ends of electric and pneumatic handpieces must be sealed with special seal caps provided (Figure 3). Note: Before putting on seal cap, turn black cap to wrap cable tight around it (Figure 4). Figure 3 Figure 4 2 Mechanical/Automated Cleaning Cycle Process Parameters Cycle Duration (minimum) Cleaning Instructions Rinse 2 minutes Cold deionized (DI) or purified water (PURW) Pre-wash 1 minutes Warm DI or PURW water ( 40 C); use detergent Cleaning 2 minutes Warm DI or PURW water ( 45 C); use detergent Rinse 5 minutes Rinse with DI or PURW water Thermal Processing 5 minutes Hot DI water, 93 C Dry 40 minutes 90 C Caution: Do not use neutralizing agent. EMAX 2 Plus and XMAX Cleaning and Sterilization Basket DePuy Synthes Companies 9

Mechanical/Automated Cleaning High Speed Handpieces and Attachments 3 Inspect Visually inspect the devices and repeat the manual pre-clean and mechanical/automated clean if visible soil remains. Remove seal cap from connector end of handpiece and drain until no visible droplets are coming from the handpiece or connector. 4 Lubricate Caution: This step is for attachments only. Do not lubricate handpieces. Option 1: 05.001.078, Lubricant for Anspach Systems, 110 ml (Applicable for Listed Attachments ONLY) Warning: Do not use Gravity Air Displacement Steam Autoclave Sterilization Process. For the following attachments apply one (1) spray pump spray to the proximal end of the attachment. CRANI-A LONG SHORT SHORT-HD LONG-HD MEDIUM CRANI-L MEDIUM-HD MA-10S CRANI-P MIA16 MA-10C MA-15S MA-15C MA-15ST MA-19ST 11 DePuy Synthes Companies EMAX 2 Plus and XMAX Cleaning and Sterilization Basket

Mechanical/Automated Cleaning High Speed Handpieces and Attachments For the following attachments apply one (1) spray pump spray to the distal end of the attachment and one (1) to the proximal end of the attachment. CSR60 QD11 QD8 MA-D20 Applying excess amounts of lubricant will cause the lubricant drip from the attachment. Clean off any excess lubricant. Option 2: Non-silicone-based medical lubricant Prepare a lubricating solution of instrument milk (non-silicone-based medical lubricant) per the manufacturer s directions. Fully immerse attachment in lubricating solution, at room temperature, in a suitable container and agitate for 15 seconds. Remove attachment and allow it to drain completely until no visible droplets are coming from it. Caution: Do not rinse out instrument milk (lubricant). Do not apply mineral oil or other lubricants, which may cause attachment to overheat. Note: The clinical processing instructions provided have been validated by DePuy Synthes Power Tools for preparing a nonsterile medical device; this instruction is provided in accordance with ISO 17664 and ANSI/AAMI ST81. EMAX 2 Plus and XMAX Cleaning and Sterilization Basket DePuy Synthes Power Tools 11

STERILIZATION HIGH SPEED HANDPIECES AND ATTACHMENTS Before sterilizing: Ensure sterilization equipment is in proper working order as specified by manufacturer. Assure equipment manufacturer s instructions are properly employed by trained and qualified personnel. Assure actual cycle employed has been properly validated for the device(s)/ load configuration being processed and appropriate sterilization indicator devices are included for each process and cycle. 1 Load the Basket Load the devices in the basket by securing them in the appropriate brackets as marked on the basket (Figure 1). Reference loading configuration pictures on Page 15 for proper placement and positioning of devices. Where applicable, distal end of devices should be angled downward. Replace seal cap in its bracket and lay handpiece connector in bottom of basket. Figure 1 Figure 2 Warning: Red protective cap for Perforator Driver (CSR60) must be left off for sterilization (Figure 2). Seal caps must be removed from connector ends of electric and pneumatic handpieces prior to sterilization (Figure 3). Devices must be cleaned prior to sterilization using the validated cleaning method outlined in these instructions. Do not sterilize using STERRAD Systems. Figure 3 Caution: Black cap must remain on connector end of handpiece during sterilization. 11 DePuy Synthes Companies EMAX 2 Plus and XMAX Cleaning and Sterilization Basket

2 Packaging for Sterilization When wrapping the basket, the lid must be used. Double wrap basket or instruments in accordance with local procedures, using standard wrap and wrapping techniques such as those described in ANSI/AAMI ST79. Warnings: When using a sterilization wrap the basket lid must be used. You must follow manufacturer s instructions for any wraps used. EMAX 2 Plus and XMAX Cleaning and Sterilization Basket DePuy Synthes Companies 11

Sterilization High Speed Handpieces and Attachments 3 Sterilization Sterilize in accordance with the following guidelines. Warning: Do not use any other sterilization method or cycle than those listed below which have been validated in accordance with ISO 17665. US Steam Autoclave Instructions Option Exposure Temperature Exposure Time Minimum Dry Time 1. Pre-Vacuum* 132 C 4 minutes 20 minutes 2. Gravity Air Displacement* *Wrapped or unwrapped 132 C 15 minutes 15 minutes Steam Autoclave/ Gravity Air Displacement* Option Exposure Temperature Exposure Time Minimum Dry Time 1. Pre-Vacuum* 132 C 4 minutes 20 minutes 2. Pre-Vacuum* 134 138 C 3 18 minutes 20 minutes Notes: Metal devices, tools and equipment are constructed of materials unaffected by normal environmental conditions of current standard sterilization means, when proper operational techniques are employed. Effectiveness of sterilization equipment or sterilization processes are directly dependent upon numerous factors beyond DePuy Synthes Power Tools control including among other things; sterilization means, processes and wrapping techniques employed, brand, model and condition of sterilization equipment, care and maintenance techniques employed and operator knowledge and experience. DePuy Synthes Power Tools cannot anticipate all possible equipment, processes and/or conditions that may be encountered. The suggested operation conditions are to be considered as a starting point for determination of the overall process capability, without regard for type or condition of equipment used or methods, techniques or practices employed by user. Use of proper sterilization indicator devices is strongly recommended. It is recommended that a drying cycle be included to avoid possible adverse effects caused by exposure to condensation. 3. Gravity Air Displacement* 132 138 C 15 18 minutes 15 minutes Steam sterilization validation was performed in accordance with ISO 17665. *Wrapped or unwrapped Warning: Do not sterilize using STERRAD Systems. Caution: It is required to include the drying cycle to avoid possible adverse effects caused by exposure to condensation. 11 DePuy Synthes Companies EMAX 2 Plus and XMAX Cleaning and Sterilization Basket

LOADING CONFIGURATION HIGH SPEED HANDPIECES AND ATTACHMENTS Seal Cap (when not in use) CSR60 Seal cap installed on connector end of handpiece. EMAX2PLUS or XMAX SHORT MEDIUM LONG MIA16 LONG-HD MEDIUM-HD SHORT-HD MA Bearing Sleeves MA-D20 QD8 QD11 CRANI-P CRANI-A CRANI-L (MA-10S, MA-10C, MA-15S, MA-15C, MA-15ST or MA-19ST) EMAX 2 Plus and XMAX Cleaning and Sterilization Basket DePuy Synthes Companies 11

ORDERING INFORMATION EX-BSKT EX-BSKT-BTM EX-BSKT-INS EX-BSKT-LID ACB SDCB EM XM Cleaning and Sterilization Basket Complete EM XM Cleaning and Sterilization Basket Bottom EM XM Cleaning and Sterilization Basket Insert EM XM Cleaning and Sterilization Basket Lid Attachment Cleaning Brushes (One 7 in Brush and One 12 in Brush) Small Diameter Cleaning Brush 11 DePuy Synthes Companies EMAX 2 Plus and XMAX Cleaning and Sterilization Basket

Limited Warranty and Disclaimer: The Anspach Effort, Inc. products are sold with a limited warranty to the original purchaser against defects in workmanship and materials. Any other express or implied warranties, including warranties of merchantability or fitness, are hereby disclaimed. Please contact Customer Service for full Warranty statement. WARNING: In the USA, this product has labeling limitations. See package insert for complete information. CAUTION: USA Law restricts these devices to sale by or on the order of a physician. To order in the US, call DePuy Synthes Power Tools Customer Service: (800) 327-6887. Not all products are currently available in all markets. Manufacturer The Anspach Effort, Inc. 4500 Riverside Drive Palm Beach Gardens, FL 33410 USA Tel: (800) 327-6887 Fax: (561) 627-1265 Email: infopowertools@dpyus.jnj.com www.depuysynthes.com DePuy Synthes 2015. All rights reserved. DSUS/PWT/0415/0115 05/15