SINGLE SPEED SYRINGE PUMP

Similar documents
VARIABLE RATE INFUSION PUMP

WATER MAZE POOL HEATER

PULSE OXIMETER PACKAGE

ASTRO FF14 FRICTION FEEDER SERVICE MANUAL

Syringe Pump. User Manual

Alaris Syringe Module v9

SERVICE MANUAL VC3ED FULL SIZE ELECTRIC CONVECTION OVEN - NOTICE -

TK-1801, TK-1811 TK-1801, TKR-1811

CMU701/ 705 OPERATIONS & INSTRUCTION MANUAL

ACME DIAGNOSTICS STANDARD OPERATING PROCEDURE

Reliability in Laboratory Centrifuges Since Operator s Manual. Models 614B, 614L, 614U/UL and 614V/VL Laboratory Centrifuge. Rev. 5.

60" Hatteras BayTM. Patio. Instruction Manual. Includes our new Wall Control System. A Kichler Décor ceiling fan

ALLERGEN MIXING ASSISTANT

918 Clock Gauge Installation, Operation, and Maintenance Instructions

Technical Update TAA.TU rev. 1

HH80C HYDRAULIC CRIMP TOOL

Application Tooling Specification Sheet

CP-200 S&T. CP Warewash Pump Instruction & Operation Manual Speed & Time , Rev1.0, , Page 1 of 8

Sanitise Syrup Lines & Valves. Taylor PH61 Cleaning every 14 days

Alaris Syringe Module v9

Application Tooling Specification Sheet

Eldex Column Heater. Operator s Manual

52 DorsetTM. Instruction Manual. Basic Function Wall Control System Included. A Kichler Decor ceiling fan

CONTENTS. III. OPERATING INSTRUCTIONS... pg 7 LOADING DISPOSABLE SET (Figure 4)... pg 9. IV. INDICATORS AND ALARMS... pg 10

K 3.97 M. Specifications or visit our website: In USA call: Part No

CP-200 Push Button. CP Warewash Pump Instruction & Operation Manual Push Button , Rev1.0,

Aquamax KF Oil Evaporator. User Manual

Orrin. Instruction Manual. Includes our new CoolTouch TM Control System Looks permanent, but goes wherever you go! U.S.

60" Tulle PatioTM. Instruction Manual. A Kichler Select ceiling fan

Oral Syringe Product Stability &

42 Kevlar. Instruction Manual. Kichler Lighting 7711 East Pleasant Valley Road P.O. Box Cleveland, Ohio

Dosing system. Commissioning and maintenance manual eco-pen300

Model K 280 M Part No

Sunburst. Instruction Manual. Includes our new Wall Control System. A Kichler Décor ceiling fan

Reproduction or other use of this Manual, without the express written consent of Vulcan, is prohibited.

Operating Manual "Mark II" Solid-State Digital Automatic Controlled Sterilizers

LaceyTM. Instruction Manual. Includes our new CoolTouch TM 6 Speed DC Control System Looks permanent, but goes wherever you go! U.S.

EASY FEED LITE FRICTION FEEDER OPERATOR MANUAL. Rev

WARNING USER S MANUAL. Model iq 2000 Series Dust Collector with manual shaker

Application Tooling Specification Sheet

Administration of AMBISOME Using the GemStar Pump

I NS T RUC T ION MODEL 4346 MIXER-GRINDER MODEL

52 Lacey LED. Instruction Manual. 6 Speed DC Wall Control System

ValkyrieTM. Instruction Manual. Includes our new CoolTouch TM 6 Speed DC Control System Looks permanent, but goes wherever you go! U.S.

Code 542 Peltier Rack User's Guide

Alaris Syringe module FAQs

INSTALLATION. and INSTRUCTION MANUAL. for QUALITY AIR BREATHING SYSTEMS. Model ABM - 715

ultra quiet fan USER MANUAL INSTALLATION OPERATION MAINTENANCE

Mitocell: Instructions for Use

Model K205 M Plus Part No

KimberleyTM A Kichler Décor ceiling fan Designed to coordinate with a popular Kichler Lighting collection.

1. SAFETY RULES WARNING WARNING. 8. Avoid placing objects in the path of the blades.

K 2.97 M. Specifications Operating pressure, max PSI Water volume GPM Voltage V Amp draw AMPS

Model K 520 M Part No

H uline.com 80" AUTOMATIC STRETCH WRAP DISPENSER FILM DELIVERY SYSTEM MACHINE DIMENSIONS ELECTRICAL SPECIFICATIONS

Automatic Humidity Controller. Model 512 (Revision C) Operating Manual

Model K 330 M Part No

FitchTM. Instruction Manual. Includes our Basic Function CoolTouch TM Control System Looks permanent, but goes wherever you go!

Hatteras BayTM. Patio. Instruction Manual. Includes our new CoolTouch TM Control System Looks permanent, but goes wherever you go! U.S.

STEAMPRO. Steam Generator Troubleshooting and Service Guide

Astro. FF-14 Friction Feeder. Operator Manual

CEILING FAN OWNER'S MANUAL

HEATING AND AIR CONDITIONING

Select. Sutter PlaceTM. Instruction Manual. A Kichler Select ceiling fan

Isco 3700ZR Sampler. Manual Set

INSTALLATION. and INSTRUCTION MANUAL. for QUALITY AIR BREATHING SYSTEMS. Model ABM - 700

ICB VEGETABLE SLICER Model HLC-300 User and Maintenance Manual

3/4 HP Model Trimmer Operating Instructions

AUTOTROL 360 MANUAL & 361 SEMI-AUTOMATIC FILTER VALVES SERVICE MANUAL

BRD 36 Part Bun and Roll Divider Rounder Model BRD-36-3, 36 part ¾ - 3 oz Model BRD-36-4, 36 part 1 ½ - 4 oz

DEMA 830 Laundry Master TM Installation Instructions

Model K 240 plus Part No

Notices. Safety Notices. 2 Restek Corporation

Filabot EX6 Extruder. Operation Manual. CAUTION! Read Carefully. Filabot Vermont, USA, Earth

CEILING FAN OWNER'S MANUAL

Gas Pressure Alarm Box SGPA Series. Operation Manual

CEILING FAN OWNER'S MANUAL

INSTALLATION & OPERATING INSTRUCTIONS

Innes. Instruction Manual A Kichler Decor ceiling fan. Includes our new TM

THE COMPACT CENTRIFUGE: OPERATOR 1

SCN60 Residential Nugget Ice Machine

60" Lyndon Patio. Instruction Manual Customer Service :30 AM to 5:00 PM EST, Monday - Friday A Kichler Decor ceiling fan

LC Series - Light Commercial Pump Station Installation and Operation Manual

A B A B4

V SERIES HDR GAS RANGES

Service Manual for X9.5 and Older Models

CEILING FAN OWNER'S MANUAL

C-IV 60 CEILING FAN READ AND SAVE THESE INSTRUCTIONS. FAN RATING AC 120V. 60Hz

*smith&nephew VERSAJET HYDROSURGERY SYSTEM

rev3 INSTALLATION & OPERATION MANUAL OIL CIRCULATING HEATING SYSTEM MODEL OSM

6. Cooling System XCITING 400i. This chapter covers the location and servicing of the cooling system components for the KYMCO XCITING 400i.

PARTS GLOSSARY Basic Insufflation Pro Package

TK-1001, 1101, 1201, 1301, 1601

THE MUG KING MUG PRESS

READ AND SAVE THESE INSTRUCTIONS

54" Skye. Instruction Manual Customer Service :30 AM to 5:00 PM EST, Monday - Friday A Kichler Decor ceiling fan

HOMEPUMP ECLIPSE* Ambulatory Infusion System PATIENT INFORMATION

GROTON. U.S. Patent(s) Pending INSTRUCTION MANUAL WARRANTY CERTIFICATE

UNIQUE USER FRIENDLY MICROMETER STYLE ADJUSTMENT

BICO DULIN ROTAREX SEPARATORS

Transcription:

instrumentation and software for research SINGLE SPEED SYRINGE PUMP PHM-100 USERS MANUAL DOC-083 Rev. 1.6 Copyright 2014 All Rights Reserved Med Associates Inc. P.O. Box 319 St. Albans, Vermont 05478 Phone: 802.527.2343 Fax: 802.527.5095 www.med-associates.com

NOTES - ii -

Table of Contents Chapter 1 Introduction... 1 Hardware Overview... 1 Wiring Instructions... 2 Chapter 2 Syringe Placement... 3 Glass Syringes... 4 Small Syringes... 4 Chapter 3 Determining the Flow Rate... 5 Syringe Cross-Sectional Areas... 5 Chapter 4 Operating Instructions... 7 Chapter 5 Automatic Shut-Off Switch... 8 Chapter 6 Maintenance and Troubleshooting... 9 Appendix A Flow Rates... 10 Appendix B Contact Information... 10 - iii -

CHAPTER 1 INTRODUCTION The standard motor speed of the PHM-100 is 3.33 RPM, but other motors are available and can be changed by the user. The PHM-100 pump is activated by a MED-PC 28-volt output and remains on for as long as the output is on. Thus the amount of time that the output is on determines the dose. An automatic shut-off switch stops the pump motor when the end of the syringe is reached. Hardware Overview MED Control Connector The pinout of the 3-pin Molex connector is shown in Figure 1.1. Upon receiving an operate signal, the pump will begin the infusion until the operate signal is turned off. Figure 1.1 MED Control Connector Pinout Molex Pin # Function 1 28 Volt Ground 2 Operate 3 + 28 Volts DC Manual Operate Button The manual operate button is located below the LED indicator on the side of the pump. The pump motor will operate for as long as this button is held down. See Figure 1.2. LED Indicator The LED indicator on the side of the pump illuminates red when the pump motor is running. See Figure 1.2. ON/OFF Switch The green ON/OFF switch is lit up with the pump is switched on and the Automatic Shut- Off Switch is not activated. See Figure 1.2. - 1 -

Figure 1.2 PHM-100 Single Speed Syringe Pump Wiring Instructions 1. Using the included SG-222 cable, connect the PHM-100 pump to any available OUTPUT on a standard MED connection panel. 2. Plug AC cord into 115-volt outlet. (220VAC for PHM-100A) - 2 -

CHAPTER 2 SYRINGE PLACEMENT Care should be exercised that the hub of the syringe barrel is positioned adjacent to the syringe clamp when loading the syringe pump. If a gap exists between the hub and the clamp, accurate flow rates cannot be assured, as the entire syringe (both barrel and plunger) may move forward. A visual check by observing the plunger move in relation to the barrel by rotating the front knob of the pump is advised. Figure 2.1 shows the procedure for loading a syringe. Figure 2.1 Loading a Syringe 1. Move the slide to rear (toward motor section) by squeezing the jaws. 2. Insert syringe by lifting clamp cover and placing the syringe body into the clamp. 3. Move the slide to the end of the syringe. 4. Rotate the knob so that the slide makes contact with the plunger and the syringe contacts the clamp. Once the syringe is installed, operate the pump until liquid drips out of the syringe or tubing. This ensures that fluid will be infused properly when the pump is activated. - 3 -

Glass Syringes Extra caution is needed when using glass syringes with a ground glass plunger. These syringes exhibit almost no sliding friction and thus can cause an uncontrolled infusion in the following two ways: 1. The weight of the plunger may be sufficient to push the fluid out of the syringe if it is positioned with the plunger above the barrel. 2. The weight of the fluid in the tubing may be sufficient to siphon the fluid out of the syringe if the catheter infusion site is below the height of the syringe. To test for these two conditions, it is suggested that the syringe be connected to the tubing and held vertically at the height of the pump. If no motion occurs, the syringe can then be placed in the pump. The following may reduce the danger of an uncontrolled infusion: 1. Lower the relative height of the infusion pump in relation to the infusion site. With the pump below the infusion site, the instrument will pump the fluid to the higher elevation. 2. Use a smaller bore catheter, which will reduce the weight of the fluid in the tubing and increase the friction on the flowing fluid. 3. Position the pump so that the syringe is vertical (plunger below), thus the weight of the syringe plunger will be acting against the weight of the fluid. 4. Use a syringe with a rubber seal on the plunger, i.e. an O-ring sealed or plastic syringe. Small Syringes Syringes of less than 5mL in capacity can be held more securely in the syringe clamp if the micro syringe insert is used. This insert slides into the standard syringe clamp and can hold up to two micro syringes if required. See Figure 2.2. Figure 2.2 R-ACC Micro Syringe Insert - 4 -

CHAPTER 3 DETERMINING THE FLOW RATE The flow rate of the PHM-100 pump depends on the size of the syringe being used. The tables in Appendix A contain the flow rates that correspond with several commonly used syringe brands and sizes, however the equation shown below can be used to calculate the flow rate. Refer to the charts below for the cross-sectional areas of several commonly used syringes. The following equation determines the approximate flow rate of the pump: Flow Rate (ml/min) =.19538 x Motor RPM x Syringe Cross Sectional Area (cm 2 ) NOTE: If the syringe cross-sectional area is unknown, or in applications where precise control of infusion volume is required, it is recommended that users calibrate each infusion pump using the syringe size of choice. A calibration curve should be created that shows volume output plotted as a function of speed setting. This information can be used to achieve the desired flow rate. Syringe Cross-Sectional Areas MULTIFIT, glass HAMILTON, glass UNIMETRICS, glass Syringe Size Cross-section Syringe Size Cross-section Syringe Size Cross-section 1 ml 0.176 sq cm 10 ml 0.00167 sq cm.05 ml 0.00833 sq cm 2 ml 0.626 sq cm 25 ml 0.00417 sq cm.10 ml 0.01667 sq cm 5 ml 1.084 sq cm.05 ml 0.00833 sq cm.25 ml 0.04167 sq cm 10 ml 1.692 sq cm.10 ml 0.01667 sq cm.50 ml 0.08333 sq cm 20 ml 3.017 sq cm.25 ml 0.04167 sq cm 1 ml 0.16667 sq cm 30 ml 4.047 sq cm.50 ml 0.08333 sq cm 50 ml 6.173 sq cm 1 ml 0.16667 sq cm 2.5 ml 0.41667 sq cm 5 ml 0.83333 sq cm 10 ml 1.6667 sq cm - 5 -

MONOJECT, plastic B-D PLASTIPAK, plastic TERUMO, plastic Syringe Size Cross-section Syringe Size Cross-section Syringe Size Cross-section 1 ml 0.173 sq cm 1 ml 0.173 sq cm 3 ml 0.629 sq cm 3 ml 0.622 sq cm 2.5 ml 0.578 sq cm 5 ml 1.327 sq cm 6 ml 1.263 sq cm 5 ml 1.129 sq cm 10 ml 1.961 sq cm 12 ml 1.977 sq cm 10 ml 1.635 sq cm 20 ml 3.189 sq cm 20 ml 3.308 sq cm 20 ml 2.850 sq cm 30 ml 4.191 sq cm 35 ml 4.474 sq cm 30 ml 3.662 sq cm 60 ml 6.651 sq cm 60 ml 5.545 sq cm 60 ml 5.556 sq cm - 6 -

CHAPTER 4 OPERATING INSTRUCTIONS 1. Load the syringe that will be used for the infusion. Refer to the Syringe Placement section of this manual. 2. Switch the On/Off switch to ON. 3. The syringe pump may be operated either manually, by pressing the push-button switch located near the connector on the side, or remotely by addressing the appropriate output on the interface. NOTE: The red light on the front indicates the pump is in operation. 4. If the automatic shut-off switch is activated, the infusion will stop. - 7 -

CHAPTER 5 AUTOMATIC SHUT-OFF SWITCH The automatic shut-off switch is activated when the pump reaches the end of the syringe, and it causes the pump motor to cease operation and the lighted power switch will disconnect, thus providing a visual indication that the infusion is complete. The automatic shut-off switch uses an internal micro-switch that detects when the slide assembly reaches a certain point. This point is adjustable so that the pump can accommodate various sizes and brands of syringes. To adjust the automatic shut-off switch, insert an empty syringe with the plunger at the desired shut-off position. Move the slide assembly to the rear of the plunger (Figure 3A). With the pump turned on, rotate the adjustment screw (Figure 3B) until the power light goes out. Turning the screw counterclockwise will allow the pump to operate longer before the automatic shut-off switch activates. Conversely, turning the screw clockwise will cause the automatic shut-off switch to activate sooner. In order to resume pump operation once the automatic shut-off switch has been activated, the slide must be moved back until the switch is deactivated. Figure 3 Automatic Shut-Off Switch Adjustment - 8 -

CHAPTER 6 MAINTENANCE AND TROUBLESHOOTING Cleaning: Clean the pump case using a soft cloth that is dampened with water and a detergent. The pump case is an ABS plastic. Lubrication: Changing Motor: Mechanical: A drop of oil on the bearings, and Vaseline or silicone spray on the lead screw is recommended annually. With the electric plug remove from the outlet, remove the motor cover. Remove the motor by pulling the electrical connector from its mate. The motor is held in position by 2 pins. Replacing the motor is accomplished by reversing the procedure. The front knob is used for aligning the slot in the lead screw with the cross drive on the motor shaft. If the syringe is not emptying, check if the tubing is kinked or if the syringe is bound. Check the pressure capability of the A-1008 slide assembly by either using a pressure gauge or by sealing a 50mL syringe with 50mL of air, and running the pump until the air inside is compressed to 36mL. If the slide is not capable of pushing the syringe, both the lead screw and the slide should be inspected for wear, and any worn part should be replaced. - 9 -

APPENDIX A FLOW RATES The following tables contain the flow rates in milliliters per minute for several commonly used syringe brands and sizes. The most commonly used motor RPM is 3.33, however other speeds have been included in the tables. MULTIFIT, glass Motor RPM 1 ml 2 ml 5 ml 10 ml 20 ml 30 ml 50 ml 0.50 0.017 0.061 0.106 0.165 0.295 0.395 0.603 1.00 0.034 0.122 0.212 0.331 0.589 0.791 1.206 1.50 0.052 0.184 0.318 0.496 0.884 1.186 1.809 2.00 0.069 0.245 0.424 0.661 1.179 1.581 2.412 3.33 0.115 0.408 0.706 1.102 1.965 2.636 4.02 5.00 0.172 0.612 1.059 1.653 2.947 3.954 6.031 10.00 0.344 1.223 2.118 3.306 5.895 7.907 12.061 15.00 0.516 1.835 3.177 4.959 8.843 11.861 18.092 20.00 0.688 2.446 4.236 6.612 11.789 15.814 24.122 B-D PLASTIPAK, plastic MONOJECT, plastic Motor RPM 10 ml 20 ml 30 ml 50-60 ml Motor RPM 12 ml 20 ml 35 ml 50-60 ml 0.50 0.160 0.278 0.358 0.542 0.50 0.193 0.323 0.437 0.542 1.00 0.320 0.557 0.715 1.084 1.00 0.386 0.646 0.874 1.084 1.50 0.480 0.835 1.073 1.626 1.50 0.579 0.969 1.311 1.626 2.00 0.640 1.114 1.431 2.169 2.00 0.773 1.292 1.748 2.169 3.33 1.066 1.856 2.385 3.614 3.33 1.288 2.154 2.914 3.614 5.00 1.600 2.784 3.577 5.421 5.00 1.931 3.231 4.371 5.421 10.00 3.198 5.895 7.155 10.843 10.00 3.863 6.462 8.741 10.843 15.00 4.798 8.353 10.732 16.264 15.00 5.794 9.694 13.112 16.264 20.00 6.397 11.137 14.310 21.686 20.00 7.726 12.925 17.483 21.686 APPENDIX B CONTACT INFORMATION Please contact MED Associates, Inc. for information regarding any of our products. Visit our website at www.med-associates.com for contact information. For technical questions, email support@med-associates.com. - 10 -