Theory of Operation: M9/D & M11/D

Similar documents
SpeedClave Steam Sterilizers

M4.9. NO LONGER IN PRODUCTION Some service parts may not be available for this product- Service and Parts Manual M4.9-0

M3 UltraFast Dental Handpiece Sterilizer

HEALTH POT MODEL: EHP-G1583G(RD) EHP-G1584G(RD) Owner s Manual. Please read this manual carefully before operating your set.

The Information Contained in this Document is Proprietary and should only be used for Service or Training of Authorized Blodgett Servicers who will

TECH SHEET DO NOT DISCARD PAGE 1. If unsuccessful entry into diagnostic mode, actions can be taken for specific indications:

Thermo 230 / 231 / 300 / 301 / Troubleshooting

NO LONGER IN PRODUCTION

Operating Manual "Mark II" Solid-State Digital Automatic Controlled Sterilizers

APPENDIX 4. Thank you for choosing the Biomega BioClave steam sterilizer. HYDRAULIC DRAWING

SERVICE MANUAL VC3ED FULL SIZE ELECTRIC CONVECTION OVEN - NOTICE -

PCS 2000 CONTROLLER USER GUIDE PCS 2000 CONTROLLER. Contents

Manual Examination Table

INSTRUCTIONS MANUAL. 1.8L Electric Kettle with Multi-Temp Feature SK-1800R (Red) SK-1800SS (Stainless Steel)

PURGER INSTALLATION AND OPERATION BULLETIN PUR-SB17-02 SERVICE BULLETIN

Honeywell ENERGY SAVINGS KIT

HEALTH POT MODEL: EHP-E1881G(WH) EHP-E1882G(OR) Owner s Manual. Please read this manual carefully before operating your set.

HP727S. Single speed swimming pool heat pump controller Operation manual TABLE OF CONTENTS

LMR Electric Fire Pump Controllers Features

SERVICE & TROUBLESHOOTING MANUAL FOR DISPLAY HOLD CABINETS DHC SERIES

Crypto-Lock. Model CC-8521B Updated 7/17. Access Control System. MONITEQ, Inc. INSTRUCTION MANUAL

Operations Manual. Part Number: For Research Use Only

FDF Foam Pump Controllers Features

REVISION STATUS AMENDED PAGES. Initial release D. All

Steam sterilizer AHS-75-DRY

Alarm Tone Generator Model AG17

GAS RACK OVENS WITH ELECTRONIC OVEN CONTROL

LMR Electric Fire Pump Controllers Features

Wiring Instructions for the 2900 Series FIREGUARD Combination Door Closer-Holder and Releasing Device with Integral Smoke Detector

Installation, Programming and Operating Guide for Digital, Programmable Thermostat with Smart Fan Technology. with SFT (Smart Fan Technology )

NOTICE . SAFE SERVICING PRACTICES. Electric Wall Oven with Electronic Oven Control

DOMESTIC HOT WATER Follow operational sequence

EC Series Wall Unit Humidifier

3500 SERIES CONVECTION STEAM COOKER PARTS AND SERVICE MANUAL

Lubrication cycle IP Enclosure Rating. 110 VAC, 220/230 VAC (50/60 Hz) Alarm Fault Relay Contacts. IP-55 (Liquid tight connector)

Installation & Operating Guide

BGH Series Hot Surface Ignition Control

TECHNICAL SERVICE DEPARTMENT Technical Service Bulletin Ultra Low NOx PowerVent with Honeywell Electronic Control California Only

SERIES VAC Microprocessor-Based Hot Surface Ignition Control FEATURES APPLICATIONS SPECIFICATIONS DESCRIPTION AGENCY CERTIFICATIONS

FT60 Reduced Voltage - Autotransformer

CONTROL PANEL INSTRUCTIONS RSQ-4

CONTROL PANEL INTERFACE ACTIVATE THE GENERATOR DISPLAY INTERFACE MENUS. Control Panel USING THE AUTO/OFF/MANUAL SWITCH

Norcold Repair Guide for Ice Maker

Installation/Service Manual. Slow Exhaust Autoclave

Installation Guide for AL600ULM. Multi-Output Access Control Power Supply Charger. Rev

BRAVO AUTOCLAVES. Installation Notes. BRAVO 1 Quick Start Guide SD-402 Rev. 1.0 Copyright 2012 SciCan Ltd. All rights reserved

SERIES VAC Microprocessor-Based Direct Spark Ignition Control FEATURES APPLICATIONS SPECIFICATIONS DESCRIPTION. Export Information (USA)

Self-Contained Electric Steam Jacketed Kettle

SERIES VAC Microprocessor-Based Direct Spark Ignition Control with Inducer Blower Relay FEATURES APPLICATIONS SPECIFICATIONS DESCRIPTION

Beacon 200 Gas Monitor Operator s Manual. Part Number: RK Released: 6/6/08

Dryer Controller M720

Bravo. Troubleshooting Guide

03/2017. Mod: DFV-511/TS. Production code: P8 RDT-105E DR 10DI 51

Model: PC600. Batch: PR3496. POWER:1000W AC V, 50/60Hz For your safety, always read the instruction book carefully before using.

INSTRUCTIONS FOR THD-U SERIES HOT AIR MATERIAL DRYERS

! WARNING To avoid risk of electrical shock, personal injury or death; disconnect power before servicing, unless testing requires power.

MULTI-STAGE CHILLER CONTROL

Statement of Responsibilities

DIESEL Engine Fire Pump Controllers Features

ELECTRONIC FIREPLACE DAMPER

Operation Manual NEMA 4X Controls

Harvey Chemiclave EC Series Sterilizers

24 VAC Intermittent Pilot Gas Ignition Control

Operating Functions Temperature Parameters

INSTRUCTION MANUAL T119 1MN0110 REV. 0. operates with ISO9001 certified quality system

Focus 4010\ Focus 4010 HT porcelain firing furnace

! WARNING To avoid risk of electrical shock, personal injury or death; disconnect power to range before servicing, unless testing requires power.

ELECTRONIC FIREPLACE DAMPER

CAUTION. Installation Manual

User Manual. Dryer Controller M720

SERVICE MANUAL REFRIGERATION

Alpha Spectroscopy. Vacuum Light. Important things to know before you load your samples:

! WARNING To avoid risk of electrical shock, personal injury or death; disconnect power to dryer before servicing, unless testing requires power.

ACSI AO ELECTRIC LATCH RETRACTION CONTROLLER INSTALLATION INSTRUCTIONS I.D. 1089, REV. F

Technical

1.1 Controller General Information

ENVIRO-PAK. Operation and Maintenance Manual. Manufacturer of

MODELS SSB-3E/5E/10EF, (2)SSB-3E/5E/10EF, SSB-3G/5G/10GF, (2)SSB-3G/5G/10GF

FLEXBAKE 5 Proof and Bake Oven. Service Training Introduction

RC-112 Two Speed Heat Pump 3 Stage Heat / 2 Stage Cool With Energy Efficient Control

Multi-Output Access Control Power Supply Charger VDC or 24VDC VDC or 3 24VDC.

SERVICE MANUAL REFRIGERATION

WARNING Important Safety Information

QUICK REFERENCE GUIDE P.C. BOARD/WALL THERMOSTAT FOR 6535D, 6537C, 6538 A&B SERIES TWO TON PACKAGED HIGH EFFICIENCY HEAT PUMPS

Installation Guide for AL400ULM. Multi-Output Access Control Power Supply Charger

GasScanner 8C. Eight Channel Monitor. Operator s Manual. MINT-0281-XX Rev. A 01/29/08

Parts Manual. Counter Type Electric Convection Steamer. Series: SteamCraft Models 21CET East 179 th Street Cleveland, Ohio 44110

INSTALLATION, OPERATION and MAINTENANCE MANUAL for Cres Cor RADIANT OVENS

MAKING MODERN LIVING POSSIBLE. Danfoss Heating. RET2000 B/M/MS Electronic digital thermostat with LCD Installation Guide

SERVICE MANUAL R-880B SHARP CORPORATION DOUBLE GRILL CONVECTION MICROWAVE OVEN MODEL

INSTALLATION, OPERATION and MAINTENANCE MANUAL for Cres Cor 12kW QUIKTHERM RETHERM OVENS

ENERGY LIGHT USER S GUIDE ENERGY LIGHT USER S GUIDE

Indirect gas-fired air heater

SERVICE MANUAL C24EO SERIES ELECTRIC COUNTERTOP STEAMERS - NOTICE - C24EO5

DWYER INSTRUMENTS, INC. P.O. BOX 373 MICHIGAN CITY, INDIANA 46360, U.S.A. Series 670 Hood Monitor. Bulletin AV-670

DXm MODULAR DECK OVEN

RA15/6T & TB Conversion Valve

Read the instructions and keep them safe. If you pass the iron on, pass on the instructions too. Remove all packaging, but keep it till you know the

INSTRUCTION MANUAL Portable Refrigerator Units MODEL 15-LITER Including the Standard Battery Backup System (BBS)

! WARNING To avoid risk of electrical shock, personal injury or death; disconnect power to range before servicing, unless testing requires power.

Transcription:

: M9/D & M11/D NOTE This document applies to the following models: M9/D (-020 thru -022) (-033 & -034) M11/D (-020 thru -022) (-033 & -034) Mode / System Page Power Up Mode... 2 Select Cycle Mode... 4 Fill Mode... 6 Heat-Up Mode... 8 Sterilization Mode... 10 Vent Mode... 12 Door Motor System... 14 Drying Mode... 16 Fan System... 18 Part No. 004-1078-00 Rev. A (8/1/13)

Power-Up Mode This illustration shows the components that affect, or are monitored during all cycle modes. Refer to the following page for a detailed description of the Power-Up Mode. Door Switch [Line VAC] High-Limit Thermostats [Line VAC] Primary Fuses (2) Line VAC 2 Rev. 7/08

Power-Up Mode Primary Fuses With the sterilizer's power cord properly connected, facility supply voltage is supplied to the Main PC Board thru the two primary fuses. Each time power is reconnected, the display panel will show: If either fuse is faulty, the sterilizer will have no power. High-Limit Thermostats When power is supplied to the Main PC Board, current continuously flows thru the two (normally closed) High-Limit Thermostats. This circuit powers all line voltage components (except Fan System). Total cycles run on the sterilizer If either thermostat opens for any reason (overheat or malfunction), the sterilizer will shut down until unit cools, or thermostat is replaced. Model Number Software Version Door Switch Once a cycle is initiated, the Main PC Board continuously monitors the status of the Door Switch. If an open door is detected, the cycle will not start. If the door switch opens during a cycle, the cycle will be terminated and the corresponding error code will appear in the display. Base Up / Down 3

Select Cycle Mode This illustration calls out the components that are energized / monitored during the Select Cycle Mode. Refer to the following page for a detailed description of the Select Cycle Mode. Air Valve - open (10 Minutes) [Line VAC] 4

Select Cycle Mode Theory of Operation During the Select Cycle Mode, the air valve is energized for ten minutes after a cycle is completed or after Power-Up Mode. [All electrical current is supplied thru the two high-limit thermostats (on bottom of chamber). Refer to 'Power-Up Mode', for further detail]. Air Valve For the first ten minutes of the Select Cycle Mode, line voltage is supplied to the (normally closed) air valve. When energized, the air valve opens. [This allows air to pass thru the valve, preventing pressure from building in the chamber]. Note Pressing the start button when SELECT CYCLE is displayed, at the beginning or end of a cycle, activates the heater for 10 minutes. The display flashes ADDITIONAL HEAT. This allows the Operator to preheat the chamber before starting an operation or to add additional time to the Dry mode at the end of an operation. Pressing STOP will end the Additional Heat time. Additional Heat can only be activated once between sterilization cycles. 5

Fill Mode This illustration calls out the components that are energized / monitored during the Fill Mode. Refer to the following page for a detailed description of the Fill Mode. Air Valve - open [Line VAC] Vent Valve - closed [Line VAC (early models)] [Approx. 108 VDC (later models)] Water Level Sensor [5 VDC] Fill Valve - open [Line VAC (early models)] [Approx. 108 VDC (later models)] 6 Fill Mode Rev. 10/10

Fill Mode During the Fill Mode, water flows from the reservoir, thru the fill valve into the chamber. During the Fill Mode, the display panel will show: [All electrical current is supplied thru the two high-limit thermostats (on bottom of chamber). Refer to 'Power-Up Mode', for further detail]. Air Valve Throughout the Fill Mode, line voltage is supplied to the (normally closed) air valve. When energized, the air valve opens. [This allows air to pass thru the valve so that water can flow from the reservoir]. Vent Valve Throughout the Fill Mode, voltage is supplied to the (normally open) vent valve. When energized, the vent valve closes. [This prevents water from flowing back into the reservoir thru the vent valve]. Fill Valve During the Fill Mode, voltage is supplied to the (normally closed) fill valve. When energized, the fill valve opens allowing water to flow into the chamber. Water Level Sensor Throughout the Fill Mode, 5 VDC is supplied to the water level sensor. When the water level in the chamber reaches the sensor, a circuit is completed and current flows back to the PC Board. When the 5 VDC from the water level sensor is detected, the PC Board stops the current flow to the fill valve. When the water level in the chamber reaches the water level sensor, the PC Board stops the current flow to the fill valve. This allows the valve to close, stopping the flow of water into the chamber. Base Up / Down Fill Mode Rev. 10/10 7

Heat-Up Mode This illustration calls out the components that are energized / monitored during the Heat-Up Mode. Refer to the following page for a detailed description of the Heat-Up Mode. Air Valve - closed-open [Line VAC is supplied intermittently to open the valve] Vent Valve - closed [Line VAC (early models)] [Approx. 108 VDC (later models)] Temperature Sensor [5 VDC] Pressure Sensor [on Main PC Board] Heating Element - ON [Line VAC] 8 Rev. 10/10

Heat-Up Mode During the Heat-Up Mode, the water in the chamber is heated to achieve the proper temperature for sterilization. [All electrical current is supplied thru the two high-limit thermostats (on bottom of chamber). Refer to 'Power-Up Mode', for further detail]. During the Heat-Up Mode, the display panel will show: Cycle that has been selected Heating Element Throughout the Heat-Up Mode, line voltage is continually supplied to the heating element. The heating element heats the water in the chamber until sterilzation temperature is achieved. Vent Valve Throughout the Heat-Up Mode, voltage is supplied to the (normally open) vent valve. When energized, the vent valve closes. [This prevents water from flowing back into the reservoir thru the vent valve]. Air Valve Periodically during the Heat-Up Mode, line voltage is supplied to the (normally closed) air valve. When energized, the air valve opens. [This occurs three times during this mode to expel air from the chamber.] Chamber Temperature Chamber Pressure Temperature Sensor & Pressure Sensor The temperature sensor (inside chamber) & pressure sensor (on Main PC Board) monitor the temperature & pressure conditions inside the chamber. When the pre-set sterilization conditions are met, the Heat-Up Mode is complete & the unit goes into the Sterilization Mode. Rev. 10/10 9

Sterilization Mode This illustration calls out the components that affect the Sterilization Mode. Refer to the following page for a detailed description of the Sterilization Mode. [Refer to to Power-Up Mode for components that are continually monitored during all modes] Air Valve - closed [no voltage] Vent Valve - closed [Line VAC (early models)] [Approx.108 VDC (later models)] Temperature Sensor [5 VDC] Pressure Sensor [on Main PC Board] Heating Element - ON-OFF [Line VAC is supplied intermittently to manitain sterilization parameters] 10 Sterilization Mode Rev. 10/10

Sterilization Mode During the Sterilization Mode, the temperature and pressure parameters for the selected cycle are maintained for the required time. During the Sterilization Mode, the display panel will show: [All electrical current is supplied thru the two high-limit thermostats (on bottom of chamber). Refer to 'Power-Up Mode', for further detail]. Temperature Sensor & Pressure Sensor The temperature sensor (inside chamber) & pressure sensor (on Main PC Board) monitor the temperature & pressure conditions inside the chamber throughout the Sterilization Mode. Heating Element Based on readings from the temperature sensor & pressure sensor, the heating element is cycled ON / OFF to maintain the required temperature and pressure for the selected cycle. Sterilization time counts down Chamber Temperature Chamber Pressure Vent Valve Throughout the Sterilization Mode, voltage is supplied to the (normally open) vent valve. When energized, the vent valve closes. [This prevents water from flowing back into the reservoir thru the vent valve]. Air Valve The air valve is closed (no voltage) throughout the entire Sterilization Mode. [This prevents pressure from escaping the chamber]. Sterilization Mode Rev. 10/10 11

Vent Mode This illustration calls out the components that affect the Vent Mode. Refer to the following page for a detailed description of the Vent Mode. [Refer to to Power-Up Mode for components that are continually monitored during all modes] Air Valve - closed [no voltage] Condensing Coil Vent Valve - open [no voltage] Pressure Sensor [on Main PC Board] 12 Rev. 7/08

Vent Mode During the Vent Mode, pressure is released from the chamber. The steam cools as it passes thru the condensing coil and the water is returned to the reservoir. During the Vent Mode, the display panel will show: [All electrical current is supplied thru the two high-limit thermostats (on bottom of chamber). Refer to 'Power-Up Mode', for further detail]. Vent Valve During the Vent Mode, the PC Board stops the current flow to the (normally open) vent valve. This allow the valve to open, and the pressure (steam) is released from the chamber. Condensing Coil When the steam is released from the chamber, it passes thru the condensing coil. The coil cools the steam and returns the water back to the reservoir. Chamber Temperature Chamber Pressure Air Valve The air valve is closed (no voltage) throughout the entire Vent Mode. Pressure Sensor The pressure sensor (on Main PC Board) monitors the chamber pressure as it is released. When the pressure reaches 0.7 psi (5kPa), you will hear several "beeps". This indicates the door will open in approximately 5 seconds. 13

Door Motor System (not applicable to M9D / M11D models) This illustration shows only the components that affect the Door Motor System. Refer to the following page for a detailed description of the Door Motor System. Door Motor Switch Cam Door Motor - ON [Line VAC] High-Limit Thermostats - closed [Line VAC] Door Switch [Closed - when door is closed] [Open - when door is open] 14 Door Motor System Rev. 7/08

Door Motor System The Door Motor System automatically opens the sterilizer door when the Vent Mode is complete. When the Door Motor System is activated, the display panel will show: [All electrical current is supplied thru the two high-limit thermostats (on bottom of chamber). Refer to 'Power-Up Mode', for further detail]. Door Motor / Door Motor Switch For the first 15 seconds, line voltage is supplied directly to the door motor. This causes the motor to run, rotating the cam and linkage downward. As the cam mechanism rotates, the motor switch closes. After 15 seconds, the current to the door motor flows thru the closed door switch. The cam continues to rotate, causing the linkage to lift the door latch mechanism and open the door. When the cam reaches the bottom of its travel, the door motor reverses direction. When the mechanism reaches its original position, the motor switch is opened. This stops current flow to the motor, and the motor stops. Door Switch The status of the (normally open) door switch reflects the position of the door. (ex. Door open= switch open) 15

Drying Mode This illustration shows only the components that affect the Drying Mode. Refer to the following page for a detailed description of the Drying Mode. Temperature Sensor [5 VDC] Sterilizer Door [Must be open] High-Limit Thermostats - closed [Line VAC] Heating Element - ON-OFF [Line VAC is supplied at pre-set intervals for the duration of the Drying Mode] 16 Rev. 7/08

Drying Mode During the Drying Mode, the heating element is energized to dry the instruments in the chamber. During the Drying Mode, the display panel will show: [All electrical current is supplied thru the two high-limit thermostats (on bottom of chamber). Refer to 'Power-Up Mode', for further detail]. Heating Element During the Drying Mode, line voltage is supplied to the heating element at pre-set intervals to turn it ON / OFF. This continues for the duration of the Drying Mode. When the drying time expires, voltage is removed from the high-limit thermostats and the heating element. Drying time counts down Temperature Sensor The temperature sensor (inside chamber) monitors the temperature throughout the Drying Mode. In general the temperature range will be approximately 170 F (77 C) to 200 F (93 C). If the temperature exceeds 240 F (115 C), the PC board stops the current flow to the heating element until the temperature drops. Sterilizer Door The sterilizer door must remain open throughout the Drying Mode. If the door is closed, pressure may build up in the chamber resulting in an error code. Rev. 7/11/12 17

Fan System This illustration shows only the components that affect the Fan System. Refer to the following page for a detailed description of the Fan System. Temperature at Fan Thermostat Primary Fuses (2) Fan Thermostat [Line VAC] [less than 130 - contacts are open] [more than 130 - contacts are closed] Fan [Line VAC] [OFF when fan thermostat contacts are open] [ON when fan thermostat contacts are closed] 18 Fan System Rev. 7/08

Fan System The Fan System reduces heat inside the enclosure by circulating air between the chamber and the covers. [The electrical current to the fan system does not pass thru the high-limit thermostats (on bottom of chamber)]. Primary Fuses With the table's power cord properly connected, facility supply voltage is supplied to the Main PC Board thru the two primary fuses. ATTENTION The fan may run continuously when running consecutive cycles. Fan When the contacts of the fan thermostat are closed, line voltage is applied to the fan causing the fan to run. When the contacts of the thermostat open, current is removed, and the fan stops. If either fuse is faulty, the sterilizer will have no power. Fan Thermostat When power is supplied to the Main PC Board, current continuously flows to the fan thermostat. The fan thermostat controls the ON/OFF function of the fan. When the temperature (at the thermostat) is less than 130, the fan thermostat contacts are open (no current to the fan). When the temperature reaches 130, the fan thermostat contacts close (current flow to the fan). When the temperature drops to approx. 100, the contacts of the fan thermostat open and the fan stops running. Fan System 19