Operating Guide IM/C1900-OGR Rev. J. C1900 Series Circular Chart Recorders

Similar documents
COMMANDER 1900 Series Circular Chart Recorders. Operating Guide. Recorder Versions COMMANDER deg F ABB Instrumentation

Electromagnetic Flowmeters

Hazardous Area Zirconia Oxygen Systems

Installation Guide. EXGP Oxygen Analyzer System (ATEX) Interface Electronics Unit

SM3000 Videographic Recorder. User Guide. Modbus (RTU) Communications Option

Zirconia Oxygen Analyzer

Zirconia Oxygen Analyzer

COMMANDER 150. User Guide Universal Process Indicator A1

Circular Chart Recorder

COMMANDER 1900 Series Circular Chart Recorders. Programming Guide. Recorder Versions COMMANDER deg F ABB Instrumentation

User Guide. DATUM L150 Panel-Mounted Universal Level Indicator A1. ABB Instrumentation

ABB. The Company. Electrical Safety. Symbols EN ISO 9001:2000

4600 Series Operating Instructions. Models 4623 and 4628 Conductivity Transmitters To Meet United States Pharmacopœia (USP 23) Regulations

Models 4623, 4628 & 5999/980 USP<645> Conductivity Analyzers

/8 DIN Universal Process Indicator

ZDT GP Series High Temperature Zirconia Oxygen Analyzer. Superior technology and quality from the world leader in oxygen measurement

COMMANDER 160 Wall/Pipe-mounted Universal Process Indicator. User Guide. ABB Automation

User Guide IM/SM3000 EN Rev. L. SM3000 Multipoint Videographic Recorder

User Guide IM/L160_6. Wall-/Pipe-mount Level Indicator L160

User Guide IM/AX4CO4 Rev. K. AX413, AX430, AX433, AX436 & AX438 Single and dual input analyzers for high level conductivity

User Guide IM/C160_8. Wall-/Pipe-mounted Universal Process Indicator C160

Halton SAFE / 7.14 user guide and installation instructions

ZX1e ZX2e ZX5e. Document No Issue 01 user manual

FlameGard 5 UV/IR HART

Pioneer-R16 Gas Monitor Operator s Manual

DSGH. Radiation-Based Detector with GEN2000 Electronics for Density Measurement QUICK REFERENCE GUIDE

Meridian wiredhart. HART Field Device Specification Goldmine Rd. Monroe, NC USA

PowerWizard. Level 1.0 & Level 2.0 Control Systems Training

Refrigeration Controller Operator s Manual (HRC) PO Box 6183 Kennewick, WA

Endura AZ40 Oxygen and carbon monoxide equivalent (COe) analyzer

Carbon Monoxide Transmitter

FlameGard 5 MSIR HART

Fire detection and alarm system Operating instructions

Interactive Fire Control Panel IFS7002 one signal loop Instruction Manual

ABB Instrumentation. Specifications. COMMANDER 150 Universal Process Indicator. the 1 /8 DIN indicator to match all your display requirements

ControlMaster CM30 and CM50 Universal process controllers, 1 /4 and 1 /2 DIN Custom configuration sheet

CONsOlIDATOR 4 & 8. MulTI- C h ANNEl CONTROllERs. ConsoliDator 4 Model PD940 ConsoliDator 4 Features. ConsoliDator 8 Features.

one Operating instructions SenTRI ONE panel-based Fire detection and alarm system

2704 Controller Melt Pressure Controller

An Introduction to KC Controls Ltd

eed quality electronic design USER s MANUAL

ABB MEASUREMENT & ANALYTICS DATA SHEET. C50 1/16 DIN controller / alarm unit

Low Temperature Zirconia Oxygen Analyzer

GLD-30 Gas Leak Detector

Frequently asked questions: Intelligent Transmitter Series

Models AX410 and AX411 Single and Dual Input Conductivity Analyzers

EXPERT TRI-STAR. Temperature controller. User s Manual

Replaceable LED modules. Sleep or unattended mode. Auto-silence and auto-acknowledge

Table of Contents. T538 Instruction Manual Rev # 6 June R6T538 June rtf E L Switchgear and Control Inc

SCAN200E USER S MANUAL

DWYER INSTRUMENTS, INC. P.O. BOX 373 MICHIGAN CITY, INDIANA 46360, U.S.A. Series 670 Hood Monitor. Bulletin AV-670

Controllers, Indicators & Set Point Programmers

BX444-Mc Gas Detector

SPECIFICATION: AlarmLine TM INTELLIGENT LINEAR HEAT DETECTION

RS485 MODBUS Module 8AI

Intelligent Security & Fire Ltd

STEAMPAK Series BurnerPAK BMS Burner Management Systems

GMA 301. Operation Manual. Worldwide Supplier of Safety Solutions. Part Number

EasyTronic III MANUAL SERVICE

w SYSTEM DESCRIPTION The Hu//Lyophi/izer System (LP-0106) performs the freeze-drying of liquid pharmaceutical product. It has a 300 ft2 shelf area

user manual Document No , Revision 03 November 2015

VT4810 SINGLE / DUAL ZONE CONTROLLER INSTALLATION MANUAL

RS Pro Infrared Temperature Sensor

Smart Temp. Model

FL500 Modbus Communication Operating Manual. Order No.: /00. MSAsafety.com

2012 Toolbar & NH3 Operators Manual

User Guide IM/AV4NIT Rev. D. AV450 and AV455 Single and dual input UV nitrate monitor

Wallace & Tiernan Controllers & Analysers ChemTrim Disinfection Controller

TYPE CM-2201 NELSON SINGLE POINT CIRCUIT MANAGEMENT SYSTEM

Model AX411 Dual Input Conductivity Analyzer

APC BC300 Series 40kW 208/450/480V User Guide

NZ 400 AUTOMATIC FIRE ALARM SYSTEM OPERATORS MANUAL

ICS Regent. Fire Detector Input Modules PD-6032 (T3419)

D8024, D9024, D10024 Analog Fire Alarm Control Panels Programming Guide

Operating instructions SENTRI4 Control panel based Fire detection and alarm system

Series DCT1000DC Dust Collector Timer Controller Specifications Installation and Operating Instructions

/8 DIN Level Indicator

Numerical Standards Listing

IMR IX176 Portable Gas Detector User Manual

2604 Controller Melt Pressure Controller

Operating Instructions

TraceNet ECM Series Control System

INSTRUCTION MANUAL. Software Revision Card 3.2 DOCUMENT SES-KS E REV. 3

Measurement & Analytics Measurement made easy. LLT100, LM80, and LM200 Lasers for level: Now everything is possible

GMA4100-DP. Control Panel. Operation Manual. Worldwide Supplier of Gas Detection Solutions

6100 SINGLE LOOP DIGITAL ADDRESSABLE FIRE ALARM CONTROL PANEL

Interactive Fire Control Panel IFS7002 four signal loops Instruction Manual

SERVICE MANUAL. Ecoboiler Model range: T20 ( ) T30 ( )

3500 CONVENTIONAL FIRE ALARM CONTROL PANEL

Servicing manual. Wall-mounted condensing gas boiler 600 Series - 11S / 19S / 24S / 24C /2002 GB(EN) For trade use

Servicing manual. 600 Series - 11S / 19S / 24S / 24C. Wall-mounted condensing gas boiler. For trade use

Mark 25 Ultrapure Water Conductivity Analyzer

INTERNATIONAL MACHIINE CONCEPTS. Operating Manual. Tablet Counting and Filling Machine. Model IMC

ModSync Sequencing System Installation & Operation Manual. For use with Fulton Steam Boilers.

I/O logger box. User manual NO POWER & SIGNAL CABLES TOGETHER READ CAREFULLY IN THE TEXT!

Two-Channel Gas Controller

Operating Instructions

Sierra Model 951 Digital Flo-Box. Instruction Manual Part Number IM-951 Revision A 09-05

Model 9010 and 9020 Monitoring Systems

Alarm Tone Generator Model AG17

Transcription:

Operating Guide IM/C1900-OGR Rev. J C1900 Series Circular Chart Recorders

The Company We are an established world force in the design and manufacture of measurement products for industrial process control, flow measurement, gas and liquid analysis and environmental applications. As a part of ABB, a world leader in process automation technology, we offer customers application expertise, service and support worldwide. We are committed to teamwork, high quality manufacturing, advanced technology and unrivalled service and support. The quality, accuracy and performance of the Company s products result from over 100 years experience, combined with a continuous program of innovative design and development to incorporate the latest technology. EN ISO 9001:2000 Cert. No. Q 05907 EN 29001 (ISO 9001) Lenno, Italy Cert. No. 9/90A Stonehouse, U.K. Electrical Safety This equipment complies with the requirements of CEI/IEC 61010-1:2001-2 'Safety Requirements for Electrical Equipment for Measurement, Control and Laboratory Use'. If the equipment is used in a manner NOT specified by the Company, the protection provided by the equipment may be impaired. Symbols One or more of the following symbols may appear on the equipment labelling: 0255 Warning Refer to the manual for instructions Direct current supply only Caution Risk of electric shock Alternating current supply only Protective earth (ground) terminal Earth (ground) terminal Both direct and alternating current supply The equipment is protected through double insulation Information in this manual is intended only to assist our customers in the efficient operation of our equipment. Use of this manual for any other purpose is specifically prohibited and its contents are not to be reproduced in full or part without prior approval of the Technical Publications Department. Health and Safety To ensure that our products are safe and without risk to health, the following points must be noted: 1. The relevant sections of these instructions must be read carefully before proceeding. 2. Warning labels on containers and packages must be observed. 3. Installation, operation, maintenance and servicing must only be carried out by suitably trained personnel and in accordance with the information given. 4. Normal safety precautions must be taken to avoid the possibility of an accident occurring when operating in conditions of high pressure and/or temperature. 5. Chemicals must be stored away from heat, protected from temperature extremes and powders kept dry. Normal safe handling procedures must be used. 6. When disposing of chemicals ensure that no two chemicals are mixed. Safety advice concerning the use of the equipment described in this manual or any relevant hazard data sheets (where applicable) may be obtained from the Company address on the back cover, together with servicing and spares information.

CONTENTS 1 INTRODUCTION Section Page 1 INTRODUCTION... 1 2 SETTING UP... 2 2.1 Instrument Power-up... 2 2.1.1 Power-up Error Codes... 3 2.2 Fitting the Chart... 4 2.3 Fitting the Pen Capsule(s)... 4 The documentation for the C1900 series of circular chart recorders is shown in Fig. 1.1. The Standard Manuals, including the data sheet, are supplied with all instruments. The Supplementary Manuals supplied depend on the specification of the instrument. INSTALLATION OPERATION 3 DISPLAYS & CONTROLS... 5 3.1 Displays and LED Indicators... 5 3.2 Use of Controls... 6 4 OPERATION... 7 4.1 Input Error Messages... 8 4.2 Operating Page Displays... 9 4.3 Alarm Acknowledge Page... 10 4.3.1 Alarm Indications... 10 4.3.2 Acknowledging Alarms... 10 4.3.3 Using the Alarm Acknowledge Page... 10 4.4 Totals Page Displays... 11 4.5 Access to Configuration Levels... 12 5 SIMPLE FAULT FINDING... 13 Product Identification Siting Mounting Electrical Connections Installation Record Part No. IM/C1900 INS PROGRAMMING General Programming Basic Config. Level Advanced Config. Level Connections and Links Part No. IM/C1900 PGR Setting Up Error Messages Displays & Controls Operating Level Simple Fault Finding Part No. IM/C1900 OGR DATA SHEET Full Specification Part No. SS C1900 6 SPARES LIST... 14 A Standard Manuals ADVANCED SOFTWARE OPTIONS Flow Totalization Ramp/Soak Profile Math Functions Timer Functions Part No. IM/C1900 ADV MODBUS (RTU) Serial Adaptors Serial Connections Programming Pages ASCII Tables Part No. IM/C1900 MOD B Supplementary Manuals Fig. 1.1 C1900 Documentation 1

2 SETTING UP 2.1 Instrument Power-up Fig. 2.1 and 2.2 Caution. Ensure that all connections, especially to the earth stud, are made correctly. a) Check that the input sensors are installed correctly. b) Check that the pen(s) are installed correctly see Fig. 2.1. c) Switch on the supply to the instrument, any power-operated control circuits and the input signals. Wait for the pens to settle. Note. On power-up, the pens are moved to an offchart position for automatic referencing. Pen chatter may occur on those pens nearest the reference position. This is a normal function of the instrument. CPU PASS 1914 J test Instrument Test identifies the instrument type, e.g. 1914J see Table 2.1 in the Installation Manual. or CPU FAIL CPU Test carries out check of processor circuitry see Error Codes below. d) The start-up sequence shown in Fig. 2.2 is displayed on faceplate 1 when the supply is first switched on. CONFIG PASS or CONFIG FAIL Blue Pen (Channel 3) Red Pen (Channel 1) Green Pen (Channel 2) Black Pen (Channel 4) see Note 0.175 (4.4) Chart Time Line 0.175 (4.4) 0.35 (8.8) Dimensions in inches (mm) Note. If the true time line event option is fitted, the violet event pen records on the same time line as the red pen, but on the outer edge of the chart. Fig. 2.1 Checking the Pen(s) Installation Configuration Test carries out check of non-volatile memories containing the instrument configuration, then indicates pass or fail see Error Codes below. CAL PASS or CAL FAIL Calibration Test carries out check of non-volatile memories containing the calibration data for each analog input and output, then indicates pass or fail see Error Codes below. bb ram PASS or bb ram FAIL Battery Back RAM Test carries out check of batterybacked RAM, then indicates pass or fail see Error Codes below.!00.3 200.3 AL RMT AT MAN Normal Display Not applicable on single channel instruments or A F 2 4 Error Codes are displayed in the event of a fault see Section 2.1.1. Fig. 2.2 Instrument Displays at Start-up 2

2 SETTING UP 2.1.1 Power-up Error Codes If any of the power-up tests fail (see Fig. 2.2), error codes are displayed to identify the fault. Refer to Fig. 2.3 for error code interpretations. A F 2 4 Configuration and battery-backed RAM errors Calibration errors Code Error Action No error None 1 Main board 2 Module in position 2 3 Module in position 3 4 Module in position 4 Analog input and/or analog output calibration is corrupt Power down and then up again. If fault remains, contact the local Service Organisation. 5 Module in position 5 6 Module in position 6 Code Error Action No error None A Main program data stored in non-volatile memory on main board is corrupt Check and correct program data C Timer set up stored in battery backed RAM is corrupt Check and correct data in Set Up Timer Page* d Maths set up stored in battery back RAM is corrupt Check and correct data in Set Up Maths Page* F Totalizer set up in battery backed RAM has been corrupt Check and correct data in Set Up Totals Page* Acknowledging Error Codes ACKNLG ErrOrS or * Refer to the Advanced Software Manual 100.3 200.5 Note. Acknowledging the Error Code clears the error state but does not rectify the fault. After acknowledging the error, carry out the relevant action detailed in the above tables. Fig. 2.3 Power-up Error Codes 3

2 SETTING UP 2.2 Fitting the Chart Fig. 2.4 2.3 Fitting the Pen Capsule(s) Fig. 2.5 75.0 deg F AL RMT AT MAN 1 Raise pens 75.0 deg F AL RMT AT MAN 1 Raise pens 2 Lift the chart clamp and remove the chart 2 Gently pull the arm off the bracket see Note 3 Fit new chart ensuring that it is beneath the pen lifter bars Note. Take care not to bend the arms during removal and refitting, as pen clashing may result. 5 4 Rotate chart to align the time line with the red pen (see also Fig. 2.1) Locate chart under guides 3 Remove spent capsule 4 Fit new pen capsule ensuring that the arm locates in the pen capsule slot 6 Lower the chart clamp 5 Remove cap Fig. 2.4 Fitting the Chart Lifter bars 7 Slide pen assembly onto the appropriate bracket until it clips into place see Note 6 Ensure that the arm is positioned above its lifter bar Fig. 2.5 Fitting the Pen Capsules 4

3 DISPLAYS & CONTROLS The displays, LED indicators and operation/programming controls are located on the faceplate on the front panel of the instrument see Fig 3.1. 3.1 Displays and LED Indicators Fig. 3.1 The displays comprise 2 rows of 6 characters. At the top of each programming page (the page header) both displays are used to describe the particular page selected. When parameters within the selected page are viewed the upper display shows the parameter and the lower display shows the value or setting for that parameter. Alarm and Channel states are indicated by separate LEDs on the faceplate of the front panel of the instrument see Sections 4.1, 4.2 and 4.3. Displays LED Indicators see Section 4 8.8.8.8.8.8. 8.8.8.8.8.8. Controls Refer to Fig. 3.2(a) to (f) for functions Information. AL1 Channel 1 AL2 Channel 2 AL3 Channel 3 AL4 Channel 4 Channel 1 Channel 2 Channel 3 Channel 4 Status of process variable alarms Current channel displayed Fig. 3.1 Location of Displays, Controls and LED Indicators A B C D E F G H I J K A b C or c d E F G H or h I J K L M N O P Q R S T U V Y L M N or n O or o P Q r S t U V Y Table 3.1 Character Set 5

3 DISPLAYS & CONTROLS 3.2 Use of Controls Fig. 3.2(a) to (f) Return from any frame Page 1 Frame 1 Frame 2 Frame 3 Frame 4 Fig. 3.2(a) Advancing to Next Page Page 2 Frame 1 Frame 2 Frame 3 Lift/Lower pen on alternate operations Notes. The key can be enabled or disabled in the Set Up Chart Page, BASIC CONFIGURATION LEVEL. If 'Auto Pen Lift Drop' has been selected in the Set Up Chart Page, the pens return automatically to their operating positions after a five minute delay. Advance to next Frame Page X Frame 1 Frame 2 Frame 3 Frame 4 Fig. 3.2(b) Moving Between Parameters Fig. 3.2(e) Lifting/Lowering the Pens Information. The key is programmed in the Set Up Function Keys Page, ADVANCED CONFIGURATION LEVEL. Acknowledge all alarms or Return operator to top of Operating Page. Parameter Value Adjust Note. Continued pressure on the and keys causes the rate of change of the displayed value to increase. To make small adjustments operate the keys momentarily. Fig. 3.2(c) Adjusting a Parameter Value Op. Page Parameter 1 Parameter 2 Page X Parameter 1 Parameter 2 Page X Parameter 1 Parameter 2 Note. The key returns the instrument display to the start of the operating page only when the display is at the top of any page. Fig. 3.2(f) Selecting Programmable Functions Parameter X Y Z Select Note. Continued pressure on the and keys causes the rate of change of the displayed value to increase. To make small adjustments operate the keys momentarily. Fig. 3.2(d) Selecting a Parameter Choice 6

4 OPERATION Faceplate 1 Pages Operating Page Section 4.2 Page 9 PV1 & PV2 PV1 & Units PV2 & Units Time Date time date See Note Alarm Acknowledge Page Section 4.3 Page 10 ACKNLG ALArMS Acknowledge Alarms Alarm Identity/Type Only applicable with Timer option Total Page Section 4.4 Page 11 Totalized Value Total 1 Reset Total Stop/Go Total 2 If Totalizer option is not fitted or Totalizers 1 & 2 are off totals PAGE reset t1 COUNt Reset Total Stop/Go reset t2 COUNt Security Code Page Section 4.5 Page 12 SECOdE Security Code ISS _ EPROM Identity OPrtOr LEVEL Operator Level basic CONFIG Basic Config AdVNCd CONFIG Advanced Config Note. The Alarm Acknowledge pages are displayed only if an alarm is active. Faceplate 2 Pages See Note If Totalizer option is not fitted or Totalizers 3 & 4 are off Operating Page Section 4.2 Page 9 PV3 & PV4 PV3 & Units PV4 & Units Time Date time date Alarm Acknowledge Page Section 4.3 Page 10 ACKNLG ALArMS Acknowledge Alarms Alarm Identity/Type Only applicable with Timer option Total Page Section 4.4 Page 11 totals PAGE Totalized Value Total 3 Reset Total Stop/Go Total 4 reset t3 COUNt Reset Total Stop/Go reset t4 COUNt Fig. 4.1 Summary of Operating Level 7

4 OPERATION The instrument has dedicated Operating Pages in the OPERATOR LEVEL see Sections. 4.1 to 4.4. These pages are used for general monitoring of the process measurements and are not affected by the security system which inhibits access to the PROGRAMMING LEVELS only see Section 4.5 on page 12. 4.1 Input Error Messages Fig. 4.2 Message Reason Action Ad.FAIL Internal analog to digital converter system hardware has failed Check that the input/output board is located correctly in its socket. Power down and up F INPt Input is above or below fault detection level. or Input exceeds the limits for the linearizer selected If the 'Ad.FAIL' message is still present, contact the local service organization Check input source for possible broken sensor Check input connections Check input link position Check input configuration in Set Up Input Page Examples Ad.FAIL 200.5 Hardware failure on Process Variable 1 200.5 F INPt Input out of range on Process Variable 4 Note. Error messages are cleared automatically when the fault condition no longer exists. Fig. 4.2 Input Error Messages Displayed in the Operating Page 8

4 OPERATION 4.2 Operating Page Displays Faceplate 1 for channels 1 and 2 Faceplate 2 for channels 3 and 4 100.3 200.5 Process Variable 1 (PV1) Process Variable 2 (PV2)* *Not displayed on single pen recorder. 300.3 400.5 Process Variable 3 (PV3) Process Variable 4 (PV4)* * Not displayed on three pen recorders. 100.3 deg F Process Variable 1 (PV1) Temperature Units for PV1 as set in the BASIC CONFIGURATION LEVEL. Display is blank if 'NONE' is selected. 300,3 deg F Process Variable 3 (PV3) Temperature Units for PV3 as set in the BASIC CONFIGURATION LEVEL. Display is blank if 'NONE' is selected. 200.5 deg F Process Variable 2 (PV2) Temperature Units for PV2 as set in the BASIC CONFIGURATION LEVEL. Display is blank if 'NONE' is selected. 400.5 deg F Process Variable 4 (PV4) Temperature Units for PV4 as set in the BASIC CONFIGURATION LEVEL. Display is blank if 'NONE' is selected. time 12. 00 Current Time* Time* (displayed using 24hr clock) time 12. 00 Current Time* Time* (displayed using 24hr clock) *Displayed only when timer option is fitted. *Displayed only when timer option is fitted. date 24 JAN Current Date* Day and Month* *Displayed only when timer option is fitted. date 24 JAN Current Date* Day and Month* *Displayed only when timer option is fitted. 9

4 OPERATION 4.3 Alarm Acknowledge Page 4.3.1 Alarm Indications Fig. 4.3 The definitions for alarm states (on, off or flashing) are detailed in Fig. 4.3. 4.3.2 Acknowledging Alarms 4.3.3 Using the Alarm Acknowledge Page 302 795 No Alarm Active No LED indicators illuminated. Note. Channel 1 and 2 alarms can be acknowledged only from faceplate 1. Channel 3 and 4 alarms (if applicable) can be acknowledged only from faceplate 2. Alarm Activated Unacknowledged alarms can be acknowledged from the faceplate controls on the front panel in two ways: In the OPERATING LEVEL by pressing the key at any frame (providing the key is programmed for this function see Section 4.1 in the Programming Manual). In the Alarm Acknowledge Page by pressing the key see Section 4.3.3 following. 302 802 Alarm Active AL2 LED indicator flashing, indicating active alarm on channel 2. Use key to go to top of Alarm Acknowledge Page. No LED illuminated indicates no alarms active. The Alarm Acknowledge Page is not displayed in the OPERATOR LEVEL. A flashing LED indicates an unacknowledged alarm on that channel. For example, a flashing AL1 LED indicates an unacknowledged alarm on channel 1. The Alarm Acknowledge Page is now displayed in the OPERATOR LEVEL. A constant LED indicates that all active alarms have been acknowledged on that channel. The Alarm Acknowledge Page remains in the OPERATOR LEVEL until all alarm conditions are cleared on that channel. Fig. 4.3 Alarm LED Indications ACKNLG ALArM5 A2HPrC 800 A2HPrC ACKNGd Alarm Acknowledge Page Use key to advance to next frame Alarm Identity Upper display: shows the alarm identity and type. Lower Display: shows the trip level of the alarm identified in the upper display. Acknowledge Alarm Use key to acknowledge the alarm (see). When the alarm is acknowledged, 'ACKNGd' is displayed and a constant LED indicates the acknowledged alarm. If there are more active alarms on channel 2 the LED continues to flash until all alarms for that channel have been acknowledged. Note. The key or a digital input can also be used to acknowledge alarm, if programmed. 10

4 OPERATION 4.4 Totals Page Displays This page is omitted from both faceplates if the Totalizer Option is not fitted. The page is also omitted from faceplate 1 if both Totals 1 and 2 are set to OFF and from faceplate 2 if both Totals 3 and 4 are set to OFF refer to the Set Up Totals Page in the Advanced Software Options Manual. totals PAGE totals PAGE 1234 5678 8765 4321 Front Panel (Batch) Flow Total 1 (3) The batch flow total is calculated from process variable 1 (3). The flow total can be reset if Reset Enable in Set Up Totals Page is set to 'ENbL Y'. The flashing channel LED indicates the flow total displayed. If YES selected For example, a flashing channel 1 LED indicates Flow Total 1 parameters displayed. reset ti NO t1 YES t1 YES t1 NO reset ti NO Counter Reset The Front (Batch) Flow Total can be reset to the Preset Value in Set Up Totals Page if required. Select 't1 YES' to reset the counter ( 't1' indicates Flow Total 1 ). Note. If the Counter Reset is disabled in Set Up Totals Page, the counter reset frame is omitted. COUNt StOP Repeat for Total 2 (if applicable) GO GO StOP COUNt StOP Repeat for Total 4 (if applicable) Counter Stop/Go Select 'GO' to start the counter or 'StOP' to stop it. Note. If the Counter Stop/Go is disabled in Set Up Totals Page, the frame can be viewed but not altered. If a digital signal is assigned to the Totalizer Stop/Go, an active digital signal sets the counter to GO and the Counter cannot be stopped from the front panel. 1234 5678 8765 4321 Front Panel (Batch) Flow Total 2 (4) Repeat the above procedure for Flow Total 2 (4). Note. The number of totalizers is dependent on the number of pens fitted to the instrument e.g. a 3 pen instrument has 3 totalizers. 11

4 OPERATION 4.5 Access to Configuration Levels A security system is used to prevent tampering with the programmed parameters by utilizing a password giving access to all programming pages refer to the Programming Manual. Security Code the pen adjustment SECOdE password 100.0 100 Incorrect Password 75.5 Security Code Page Set the security code to the correct password using the and keyes and use the key to advance to other programming levels (OPERATOR, BASIC CONFIGURATION and ADVANCED CONFIGURATION). The password is programmed in the Access Page in the BASIC CONFIGURATION LEVEL. Correct Password PEN 1 Pen Position Adjustment (Pens 1 to 4) The position of any trend pen can be adjusted against a reference standard (without changing the displayed value). Each pen can be adjusted in steps upwards (towards the edge of the chart) or downwards (towards the center of the chart). When this feature is enabled, a password must be entered before adjustments can be made. Note if pen adjustment is disabled or if the password is incorrect, the four Pen Adjustment frames are not displayed. For each trend pen, the lower part of the frame shows the pen position adjustment. Pen position adjusted downwards by > 25 steps OPrtOr LEVEL PEN 4 Pen position adjusted downwards by between 1 and 25 steps Pen position not adjusted Pen position adjusted upwards by between 1 and 25 steps 2001 ISS _ EPROM Version EPROM Issue Number EPROM Identification Use the key to advance to the EPROM Identification Frame. The upper display shows the EPROM version, e.g. 2001 and the lower display shows the EPROM issue number. Pen position adjusted upwards by > 25 steps adjusts the pen position upwards (towards outer edge of chart) adjusts the pen position downwards (towards center of chart) The pen adjustment frame for any given pen only appears if the pen is a Trend pen BASIC CONFIG AdUNCd CONFIG OPt ION type 0 Option Shows the software key option type. For details of the options, refer to the Data Sheet, SS/C1900R 12

5 SIMPLE FAULT FINDING Symptom Possible Cause Action Does not power up a) Internal fuse (if fitted) is blown a) Check wiring, rectify fault and replace fuse b) Internal power switch (if fitted) is OFF b) Turn power switch ON c) Power supply connections are incorrect c) Check connections Chart does not appear to move a) Very slow chart speed selected a) Select required chart speed in Set Up Chart Page b) Chart stop function enabled b) De-activate source being used to stop chart see Set Up Chart Page Pens in recording position but do not drop Chart stop function enabled De-activate source used to stop chart see onto paper Set Up Chart Page Red pen does not move beyond 94% When real time event pen is fitted the red pen Use chart range which prevents the need to go position on chart cannot go beyond 94% to prevent pens beyond 94% of maximum on chart clashing Pen lift switch on front panel does not Pen lift switch is disabled Enable pen-lift switch in Set Up Chart Page work Pens do not remain lifted when pen lift key Auto pen drop feature is enabled Disable auto pen drop in Set Up Chart Page if is used this is not required Analog inputs are slow to respond A large filter time has is set Set digital filter value to give required response in Set Up Inputs Time or date incorrect Not set for correct local time Set correct time and date in Set Up Clock Page refer to Advanced Software Manual Totalizers cannot be set to STOP or GO Operator STOP/GO selection is not enabled Enable counter STOP/GO in the Set Up in the OPERATOR LEVEL Totals Page Totalizer cannot be set to STOP Digital signal assigned to the total STOP/GO De-activate digital signal assigned to total function is active STOP/GO function External relays connected to relays in Arc suppression capacitors are provided Remove the arc suppression components instrument fail to de-energize across the relay contacts and capacitor IC4 and IC5 on mainboard leakage current may be sufficient to prevent IC6 and IC7 on standard I/O and analog relay an external relay from de-energizing IC3 to IC10 on 4 relay module 13

6 SPARES LIST Item Part No. Pen Capsules (pack of 3) Black... C1900/0119 Blue... C1900/0120 Red... C1900/0121 Green... C1900/0122 Violet*... C1900/0123 Pen Arm Assemblies ER/C Type Chart (J or R in Code Number) Standard Pen... C1900/0076 ER/C Type Chart (J or R in Code Number) Event Pen... C1900/0078 PX105 and PXR105 Type Chart (K or S in Code Number) Standard Pen... C1900/0075 PX105 and PXR105 Type Chart (K or S in Code Number) Event Pen... C1900/0077 Fuses 24V... B11071 (4A) 115V... B11070 (1A) 230V... B11069 (500mA) *True time line event option only. 14

NOTES 15

16 NOTES

Products and customer support Automation Systems For the following industries: Chemical & Pharmaceutical Food & Beverage Manufacturing Metals and Minerals Oil, Gas & Petrochemical Pulp and Paper Drives and Motors AC and DC Drives, AC and DC Machines, AC Motors to 1kV Drive Systems Force Measurement Servo Drives Controllers & Recorders Single and Multi-loop Controllers Circular Chart and Strip Chart Recorders Paperless Recorders Process Indicators Flexible Automation Industrial Robots and Robot Systems Flow Measurement Electromagnetic Flowmeters Mass Flowmeters Turbine Flowmeters Wedge Flow Elements Marine Systems & Turbochargers Electrical Systems Marine Equipment Offshore Retrofit and Refurbishment Process Analytics Process Gas Analysis Systems Integration Transmitters Pressure Temperature Level Interface Modules Valves, Actuators and Positioners Control Valves Actuators Positioners Water, Gas & Industrial Analytics Instrumentation ph, Conductivity and Dissolved Oxygen Transmitters and Sensors Ammonia, Nitrate, Phosphate, Silica, Sodium, Chloride, Fluoride, Dissolved Oxygen and Hydrazine Analyzers Zirconia Oxygen Analyzers, Katharometers, Hydrogen Purity and Purge-gas Monitors, Thermal Conductivity Customer support We provide a comprehensive after sales service via a Worldwide Service Organization. Contact one of the following offices for details on your nearest Service and Repair Centre. UK ABB Limited Tel: +44 (0)1453 826661 Fax: +44 (0)1453 829671 USA ABB Inc. Tel: +1 215 674 6000 Fax: +1 215 674 7183 Client Warranty Prior to installation, the equipment referred to in this manual must be stored in a clean, dry environment, in accordance with the Company's published specification. Periodic checks must be made on the equipment's condition. In the event of a failure under warranty, the following documentation must be provided as substantiation: A listing evidencing process operation and alarm logs at time of failure. Copies of all storage, installation, operating and maintenance records relating to the alleged faulty unit.

Contact us To find your local ABB contact visit: www.abb.com/contacts For more product information visit: www.abb.com Note We reserve the right to make technical changes or modify the contents of this document without prior notice. With regard to purchase orders, the agreed particulars shall prevail. ABB does not accept any responsibility whatsoever for potential errors or possible lack of information in this document. IM/C1900-OGR Rev. J 08.2010 We reserve all rights in this document and in the subject matter and illustrations contained therein. Any reproduction, disclosure to third parties or utilization of its contents in whole or in parts - is forbidden without prior written consent of ABB. Copyright 2010 ABB All rights reserved