Section. 3: CADD Pump. Updated: June 20, Section Author(s): sstafford & jhval

Similar documents
Description of the CADD-Prizm VIP Pump

Clinician Information

Ambulatory infusion pump for continuous, intermittent, TPN or PCA delivery.

CADD-Prizm PCS II Ambulatory Infusion Pump Model Technical Manual

Administration of AMBISOME Using the GemStar Pump

Syringe Pump. User Manual

HOMEPUMP ECLIPSE* Ambulatory Infusion System PATIENT INFORMATION

Alaris Syringe Module v9

Home Choice cycler for Peritoneal Dialysis: Set up, Disconnect, & Disposal Updated 2015

Sapphire TM. Bring Home The Future Of Infusion Care. Ambulatory Infusion System. > Intuitive, easy-to-use touch screen design

CONTENTS. III. OPERATING INSTRUCTIONS... pg 7 LOADING DISPOSABLE SET (Figure 4)... pg 9. IV. INDICATORS AND ALARMS... pg 10

Table of Contents. kangaroo 924. English. General Information... Operating Procedures... Battery... Beginning Administration...

Alaris Syringe Module v9

Enteral Pump Instructions for Use To Assemble When the low

Mini Rythmic PN+ Set-up Guide

Alaris Syringe module FAQs

ACTIV.A.C. User Manual THERAPY SYSTEM. For Patients and Clinicians. Rx Only

OPERATOR S MANUAL. The Belmont Rapid Infuser, FMS Boston Road Billerica, MA 01821, USA US/Canada Worldwide

VENTURI Compact 300 NPWT Systems Choice of Therapy Modes Advanced Predictive Low Vacuum Alarm

Duodopa. Important Risk Minimisation Information Patient Pocket Guide EDUCATIONAL MATERIAL FOR RISK MINIMISATION

v.a.c.simplicity TM Quick Reference Guide Power Connection Canister Battery Charging Light Control Panel

TPCA TM PATIENT CONTROLLED ANALGESIA SYRINGE PUMP

Thopaz+ Quick Reference Guide. Precious life Progressive care

M E D I C A L D E V I C E S

Cl inician Training Packet

T34TM AMBULATORY SYRINGE PUMP

bibag System User s Guide

Rocket Quick Guide. Rocket PSU Portable Suction Unit

Directions for Use. Alaris System. Alaris Products. Supports Guardrails Suite MX (v8) August RATE (ml/h) RATE (ml/h) CHANNEL SELECT PAUSE

Beacon 200 Gas Monitor Operator s Manual. Part Number: RK Released: 6/6/08


Blood/Fluid Warmer. The blood/fluid warmer is used to provide safe and effective warming of blood, blood products, and liquids.

Smoke Detector and Siren. Manual

Wireless Central Monitoring Unit

Bravo. Troubleshooting Guide

Original Perfusor & Infusomat

UPS BATTERY TEST. Testing the UPS Battery (Verification Test) Locking models: 13/19/22-USGF, 13/19/22- USGR, 43/44-UDGR, 43/44-UDGF

PWM. Solar Charge controller with Ethernet. Solar Smart PWM 20Amp. Hardware Description : Release : 19 June 2014

(12) Patent Application Publication (10) Pub. No.: US 2014/ A1

VICTORIAN RESPIRATORY SUPPORT SERVICE AUSTIN HEALTH

Smart-Wall TV Kit - Installation & Operation Instructions -

Tube Feeding by Pump UHN. Information for patients and families. Read this book to learn:


Directions for Use Alaris System (with PC Unit, Model 8000)

Directions for Use. Alaris System (with Alaris PC Unit, Model 8015) September 2009

Ventricular Assist System

Oxygen Concentrator Instruction Booklet

Hi-Tech, Low Cost Stanley Solutions. Product Brochure

Quick-Start Guide Digi-Sense TC9000 Advanced PID and On/Off Temperature Controller with Thermocouple Input Models and

ZX1e ZX2e ZX5e. Document No Issue 01 user manual

AutoSet AutoSet for Her Elite Escape Auto Escape. H5i H5i for Her. Welcome Guide. English. POSITIVE AIRWAY PRESSURE DEVICEs HEATED HUMIDIFIER

Covidien Articulating Absorba Tack Foil Pouch Leak Tester Owner s Manual IPE SN Rev 1

Rev C May GE Interlogix. Part No: R. CareGard. User Guide

Introduction. OM-THA2

BRAVO AUTOCLAVES. Installation Notes. BRAVO 1 Quick Start Guide SD-402 Rev. 1.0 Copyright 2012 SciCan Ltd. All rights reserved

Watchguard WGAP864 User Manual

Portable Air Conditioner Instruction Manual

S9 Escape. Welcome Guide. English POSITIVE AIRWAY PRESSURE DEVICE

VPAP COPD. H5i. Welcome Guide. Making quality of care easy. Respiratory Care Solutions. English NONINVASIVE VENTILATOR HEATED HUMIDIFIER

U ser's Guide PC6010

MULTI REACH Stick Vacuum

Installation Instructions EFS Hand Washing Station

User Guide. HTM Nurse Call. In the event of requiring assistance, please contact your Installer/Supplier.

USER MANUAL For Clinicians. Rx Only

THANK YOU FOR VOTING TEXECOM THANK YOU FOR VOTING TEXECOM USER GUIDE. Security Control Panels 1 INS167

Version 1.03 January-2002 USER S MANUAL

Syringe Pump. Syringe Pump. Product Model: ZD-50C6 Syringe Pump. ZD-50C6-H Syringe Pump

THA2 TEMPERATURE/ HUMIDITY/ DEWPOINT ALARM WITH LOGGING CAPABILITY. 99 Washington Street Melrose, MA Phone Toll Free

Temporary wireless battery operated fire system from Channel Safety Systems

TBR 40 1 GD Blood Refrigerator Instruction Manual. for BLOOD STORAGE REFRIGERATORS MODEL TBR 40 1 GD. (40 x 500ml Blood Bag Capacity)

INSTRUCTIONS CONTROL PRODUCTS INITIAL SET-UP. Innovative Technologies in Custom Electronic Design & Manufacturing

Testing the System. Battery Test. Dialer Test. Fire Drill Test (Code + [#] + 69) One-Man Fire Walk-Test (Code + [#] + 68)

TV2 Room Pressure Monitor Quick Start Guide

7. Troubleshooting. Description and Corrective Action. Low Pressure

Automatic Phone-Out Home Monitoring Systems

Added password for IP setup page : Password must be in IP format!

SERVICE MANUAL WASHING MACHINE

Achieva. Ve nt i l at or with Flow Acceleration Limit and Apnea Backup. User s Manual

MODEL 5800A CARBON MONOXIDE (CO) MONITOR AND ALARM SAMPLE-DRAWING TYPE

MULTI Hand Vacuum USER GUIDE 1985 SERIES

DI220 Using the Optional Docking Station The maintenance and operating functions of the Micro IV can be automated with the Micro IV Docking Station.

VPAP Auto VPAP ST VPAP S. H5i. Welcome Guide. English POSITIVE AIRWAY PRESSURE DEVICE HEATED HUMIDIFIER

MINIMED 670G SYSTEM CHANGING THE MINIMED MIO INFUSION SET QUICK REFERENCE GUIDE

URGENT: MEDICAL DEVICE CORRECTION Plum 360 Infusion System List Number: 30010

User s Manual. i-warm Fluid Warmer

Quick start guide. English

S9 AutoSet & S9 Elite

VPAP Adapt. H5i. Welcome Guide. English. adaptive servo-ventilator

MedReady Set-up, Helpful Hints, and Solutions for Special Needs

Endoscope Reprocessor Installation Instructions Endoscope Reprocessing System

CER-1 & CER-2 Installation & Testing Instructions

HALOGUARD TM IR MONITOR (OPTIONAL) FOUR or EIGHT POINT SCANNER INSTRUCTION MANUAL

Operator s Manual DigiPump LP8 Infusion Pump - Digicare Biomedical Technology Inc CONTENTS. A. About this manual...2

V.A.C.RX4 USER MANUAL THERAPY SYSTEM. Rx Only

Elite 16D Version 16 Zone Controller Arrowhead Alarm Products Ltd. Operating Guide. Proudly Designed and Manufactured in New Zealand

Left Ventricular Assist Devices

Maintenance Manual PC6010. WARNING This manual contains information on limitations regarding product use and function

JUNO-NET. Fire Alarm Control Panel OPERATION & MAINTENANCE MANUAL

Ice Lined Refrigerator User Operation & Maintenance manual

D8024, D9024, D10024 Analog Fire Alarm Control Panels Programming Guide

Transcription:

Section 3: CADD Pump Updated: June 20, 2012 Section Author(s): sstafford & jhval

Section 3: CADD Pump 2 Section 3: CADD Pump Field Guide

Section Contents CADD Prizm Diagram... 5 CADD Pump Quick Reference... 7 Programming Screens... 7 Alarms & Troubleshooting... 7 History and Recordkeeping Functions... 8 Programming / Setup Sequence... 9 Changing the Cassette... 10 CADD Post Test... 13 FORM: Continuous Medication Infusions... 15 Field Guide Section 3: CADD Pump 3

4 Section 3: CADD Pump Field Guide

Model 6100 and 6101 Ambulatory Infusion Pump Click here to View the full Clinical Information CADD Prizm Diagram A. Green indicator light: blinks when pump is running B. Amber indictor light: blinks when pump is stopped or alarming C. Polemount bracket recess (back of pump) D. Power jack (for power pack or AC adapter) E. Data in/out jack (for remote dose cord, modem cable or printer cable) F. Battery compartment (side of pump) G. Cassette (part of reservoir or administration set that attaches to pump) H. Air Detector Field Guide Section 3: CADD Pump 5

6 Section 3: CADD Pump Field Guide

CADD Pump Quick Reference CADD Prizm VIP Model 6100 and 6101 Ambulatory Infusion Pump http://www.smiths-medical.com/upload/products/product_relateddocs/infusion/cadd-prizm-vip-6100-6101/19805%20vip%20pca%20ll0%20quick%20card.pdf Programming Screens Reservoir Volume: Volume of fluid in filled reservoir Continuous Rate: Continuous rate of drug delivery in ml/hr Milliliters Given: Displays total amount delivered since date and time shown Air detectors caused multiple errors in the field, so most companies, including QualMed, do not use them on the Prizm pumps. Alarms & Troubleshooting The CADD-Prizm pump signals with audible alarms and messages on the display when a condition requires your attention. In addition to referring to this table, use the? key while the alarm is occurring for information about the alarm and corrective action. Reservoir Volume Low Level of fluid in the reservoir is low. Prepare to install a new reservoir. Reservoir Volume Zero Reservoir volume is 0.0 ml. Press NEXT to stop alarm. Install new reservoir if appropriate. 9-volt Battery Low 9-volt battery is low but pump is operable. Change battery soon. 9-volt Battery Depleted 9-volt battery is depleted and cannot operate pump; install fresh battery immediately. High Pressure Possible blockage downstream (between pump and patient), clamp or kink in fluid path. Find cause and remove blockage to continue. Or, press STOP/START to stop pump and silence alarm; remove blockage and restart pump. Upstream Occlusion (Model 6101 only) Fluid is not flowing from the reservoir to the pump. Check for a kink, a closed clamp or an air bubble in tubing between reservoir and pump. Press STOP/START to silence alarm for two minutes; remove obstruction and restart pump. Error Detected Pump fault occurred. Close tubing clamp, remove pump from service. 3 Beeps Every 5 Minutes; Stopped on Main Screen Pump is stopped. Start pump if appropriate. Field Guide Section 3: CADD Pump 7

History and Recordkeeping Functions The pump may be running or stopped to view functions or to print. To View Event Log To Clear Dose Counters To Clear Units Given 1. Press OPTIONS. 2. Press Y or N until Event Log option appears. 3. Press ENTER to display Event Log. 4. Press N to page back through Event Log entries or Y to page forward. 5. When finished, press NEXT. A demand dose must be programmed and the pump must be in LL0 or LL1. 1. Stop the pump. 2. Press NEXT repeatedly until Dose Counters screen appears. 3. Press ENTER to clear and update date and time. Pump must be in LL0 or LL1. 1. Stop the pump. 2. Press NEXT repeatedly until Milliliters Given, Milligrams Given or Micrograms Given screen appears. 3. Press ENTER to clear and update date and time. WARNING: This Quick Reference Card should be used by clinicians only. Do not permit patients to have access to this card, as the information would allow access to all programming and operating functions. Assistance with the CADD-Prizm pump is available 24 hours a day by calling 800-426-2448 in the USA and Canada. Important to do. This is a common omission and can be frustrating in the field. 8 Section 3: CADD Pump Field Guide

Programming / Setup Sequence To go to the next programming screen without changing a displayed setting, press NEXT. During programming, the pump may ask you to confirm certain settings; press Y or N as appropriate. These screens are dependent on other settings and may not appear. Battery must be installed; pump must be stopped and in Lock Level 0. 1. On Main Screen, press NEXT. Enter Reservoir Volume 2. Press Y or N to select desired Reservoir Volume. 3. Press ENTER. Enter Units 4. Press Y or N to select desired units. 5. Pres ENTER. *Enter Concentration 6. Press Y or N to select desired Concentration. 7. Press ENTER. Enter Continuous Rate 8. Press Y or N to select desired Continuous Rate (select the upper limit if the program will be adjusted in LL1). 9. Press ENTER. Enter Demand Dose *Enter Demand Dose Lockout Time *Enter Max Doses Per Hour *Clear Dose Counters Clear Units Given Check Air Detector Status *Insert New Patient Marker Review Program If Operating in LL1 with Upper Limits, Decrease Continuous Rate or Demand Dose 10. Press Y or N to select desired amount of patient activated demand dose (select the upper limit if program will be adjusted in LL1). 11. Press ENTER 12. Press Y or N to select desired time between Demand Doses. 13. Press ENTER. 14. Press Y or N to select desired number of Demand Doses per hour. 15. Press ENTER. 16. Press ENTER to clear the Dose Counters; this updates the date and time markers. 17. Press ENTER to clear the Dose Counters; this updates the date and time markers. 18. If air detector is attached, check the Air Detector status, then press NEXT. 19. Press ENTER to insert a New Patient Marker in the Event Log. 20. Press Y (this clears the previous Extended History). 21. Press NEXT repeatedly to review program. If pump will be operated in LL1 to allow adjustment of continuous rate or demand dose (up to the maximum entered in LL0): 22. Change lock level to LL1. 23. Press NEXT to go to Continuous Rate or Demand Dose screen. 24. Press N to scroll down to starting value. 25. Press ENTER. Follow steps for attaching cassette, priming fluid path, starting pump. Field Guide Section 3: CADD Pump 9

Changing the Cassette To Remove Used Cassette 1. Stop the pump. 2. Close all tubing clamps. 3. Use key to unlock used cassette. 4. Use side of key or coin to unlatch used cassette. 5. Disconnect tubing from patient. 6. Remove and discard used cassette. To Attach New Cassette 1. Close tubing clamp. 2. Insert new cassette hooks into pump s hinge pins. 3. Place pump with cassette on firm surface and push down on top of pump. 4. Use coin or side of key to hatch cassette. 5. Use key to lock cassette. 6. The mark on latching button must line up with solid dot on side of pump Priming the Fluid Path Using the Prime Option 7. Gently twist and pull on cassette to make sure it is attached properly. 8. When Cassette Locked appears, press NEXT. 9. If Reset Reservoir to ml? appears, press Y if appropriate. 10. When Prime Tubing? appears, open tubing clamp and press Y if appropriate. Follow priming instructions on pump s screen. When primed, close tubing clamp. 11. If applicable, thread tubing through Air Detector. 12. If AutoLock not in use/change Lock Level from LL0? appears, press Y. Change the lock level to LL1 or LL2 (see To Change Lock Level). 13. When Start the Pump? appears, connect new tubing to patient s access site, open tubing clamp, and press Y. The priming feature is available either: a) when you attach a cassette (above), or b) in the Options Menu. To access the Prime option in the Options Menu: Pump must be in LL0 or LL1. 1. Stop the pump 2. Press OPTIONS. 3. Prime should be displayed; press ENTER. 4. Follow priming instructions on pump screen. 10 Section 3: CADD Pump Field Guide

To Reset Reservoir Volume Without Changing Cassette 1. Stop the pump. 2. Press NEXT to display Reservoir Volume screen. 3. Press ENTER. 4. If Message appears asking you to reset the Reservoir Volume, press Y. Field Guide Section 3: CADD Pump 11

12 Section 3: CADD Pump Field Guide

CADD Post Test 1) The code to change/clear settings on the CADD pump is. The pump needs to be running or stopped (circle one) to perform this task. 2) The code to administer a clinician bolus on the CADD pump is. The pump needs to be running or stopped (circle one) to perform this task. 3) How do you ensure the concentration of the medicine is correct before administering a clinician bolus via the CADD pump? 4) What lock level should the CADD pump be in for narcotic administration? 5) A nurse is working on a Saturday and has been called to replace a foley catheter on a patient who has a CADD pump running. Does the nurse need to clear the pump (boluses and total amount) if the patient s pain is well-controlled and the only reason for the visit was to change the foley catheter? Yes or No (Circle correct answer) 6) Please check the instances when a nurse should double-check the CADD settings with another RN/infusion pharmacist or medical director and chart to that effect. When: increasing an oral medicine with a doctor s order. changing a narcotic cassette and not changing any settings. changing any settings on a CADD pump. starting nebulizers. changing any settings on an intrathecal pump. 7) How often should a CADD pump be cleared (bolus and total amount)? 8) What is the maximum amount of time a bag or cassette can hang before being changed? 9) How many extra bags/cassettes should be in a pt s home at all times? 10) How often should you change the battery? 11) What should you do if a pt with a CADD pump dies? 12) When do you start a new infusion flowsheet? 13) How do you prime a line using the CADD pump? Does the pump need to be running or stopped (circle one) to perform this task. Field Guide Section 3: CADD Pump 13

14 Section 3: CADD Pump Field Guide

FORM: Continuous Medication Infusions Forms can be found in the Forms Room located on the 7 th floor Field Guide Section 3: CADD Pump 15

16 Section 3: CADD Pump Field Guide