An Advanced Heat Exchanger for the Agilent 1290 Infinity Thermostatted Column Compartment

Similar documents
Optimizing the Performance of the Agilent 1290 Infinity Evaporative Light Scattering Detector

High Volume Injection up to 5 ml with the Agilent 1200 Infinity Series Online SPE Solution for Highest Sensitivity

Enhancing Throughput of Online SPE Analysis of Pesticides in Wastewater The Agilent 1200 Infi nity Series Online SPE Solution

Agilent 1290 Infinity II Vialsampler Data Sheet

Application Note. The Agilent 1200 Infinity Series Online SPE Solution for the Highest Sensitivity. Author. Abstract.

Rapid Quantification of Polar and Semipolar Pesticide Metabolites with Combined Online SPE and Direct Injection

Application Note. Authors. Abstract

Direct Injection and Online SPE LC/MS/MS for the Determination of Pharmaceuticals and Personal Care Products (PPCPs) in Water Samples

Agilent 1290 Infinity II LC

Analysis of Pharmaceuticals and Personal Care Products in Groundwater by Online SPE and LC/MS/MS

On-line SPE Enrichment of Trace Organic Contaminants in Water and Juices

Customers should provide all necessary operating supplies upon request of the engineer.

Sensitive Determination of Microcystins from Environmental Waters Using On-Line SPE

Agilent InfinityLab Online SPE Solution

AppNote 3/2009. Automation of Solid Phase Extraction Methods using a Robotic X-Y-Z Coordinate Autosampler with Software Control KEYWORDS ABSTRACT

InfinityLab Quick-Connect Bio-inert Heat Exchanger

Agilent 1200 Series Variable Wavelength Detector

Agilent 1100 Series Refractive Index Detector. Reference Manual

Agilent 1290 Infinity II Online SPE Solution

Agilent 1200 Series Diode Array and Multiple Wavelength Detectors

Improved HPLC Separations at Elevated Temperatures Using a New Stand-alone Preheater

Customised HPLC Systems for Highest Performance

On Line SPE LC/MS/MS Applications

Nexera-i Specifications Prominence-i Specifications

AppNote 6/2008. Automated SPE and Fast GC-ECD Analysis of PCBs in Waste Oil KEYWORDS ABSTRACT INTRODUCTION. MACH, Fast GC-ECD, SPE, PCB, waste oil

An Automated SPE Procedure and Analysis for the Extraction of Chlorpromazine and Thioridazine from Plasma Application Note 213

DYE QUALITY CONTROL REFERENCE MATERIAL KIT

Agilent 1100 Series Diode Array and Multiple Wavelength Detectors. Reference Manual

Automated Solid Phase Extraction (SPE) of EPA Method 1694 for Pharmaceuticals and Personal Care Products in Large Volume Water Samples

Solid Phase Extraction using the 96-well Format ISOLUTE -96 Fixed Well Plates and ISOLUTE Array 96-well Plates

detectors UltiMate 3000 Variable Wavelength Detectors

Thermo Scientific Dionex UltiMate 3000 Fluorescence Detectors

MEPS - Micro Extraction by Packed Sorbent Online SPE for GC and LC sample preparation - Extraction to injection in a single process

Agilent 1200 Series Fluorescence Detector

The Complete and Easy Guide to Configuring Your Specific Thermo NanoLC for Mass Spec Analysis

BENEFITS. QUATERNARY HPLC SYSTEMS Quaternary and Other Gradients. The Complete Range of HPLC. Convenience of Maintenance. Systems Configuration

This document is a preview generated by EVS

Agilent 1260 Infinity Multi-Detector Suite

Wide-Bore Heat Exchanger Installation Kit ( )

How Important is Performance to your SPE Method Development?

ProStar 325 UV-Vis Detector

CALIDUS tm the Modular micro Gas Chromatograph

Installing the Agilent CE ESI-MS Sprayer Kit (G1607A)

Spark Holland introduces

DETERMINATION OF PAH In Particulate Matter PM 10 with SPE/EVAporation

Proudly serving laboratories worldwide since 1979 CALL for Refurbished & Certified Lab Equipment

FPD - Flame Photometric Detector

The new LaChrom Elite Generation

Waters 2695 Separations Module

The new LaChrom Elite Generation

GC DETECTORS Reduction Gas Detector - RGD. Overview

Target Platform. Considerations. it is recommended that the server and bandwidth are optimised for acceptable performance.

Tool-free LC connections for HPLC, UHPLC and nano LC systems

Agilent 1100 Series Variable Wavelength Detector. Reference Manual

HITACHI CHROMASTER. Clarity Control Module. Code/Rev.: M173/70I Date: 6/23/2018

Online SPE-UHPLC-MS. CHRONOS for Control & System Integration. Flexible autosampler for sample preparation. Superior UHPLC system for analysis

Process and Air Monitoring. Online GC systems for continuous monitoring of organic compounds in air and gaseous samples

HPLC Column Ovens and Temperature Controllers

Precision Temperature Control. Pulse Free Pump. Column Oven. Superb Performance Ion Chromatography. Leak Detector. Modular Flexibility

EVAPORATIVE LIGHT SCATTERING DETECTORS FOR HPLC

Waters 2695D Separations Module

Thermo Scientific Dionex AutoTrace 280 SPE Instrument. Automate Large-Volume. Solid-Phase Extractions

Perkin Elmer Clarus 500 Gas Chromatograph

Agilent 1200 Infinity Series ELSD

FIXED GAS ANALYZER (FGA)

Installation of Hotel Drawers

Agilent 1200 Infinity Series ELSD

To request a quote or locate your nearest dealer, contact Customer Care toll-free in the US at or

APPLICATION NOTE DETERMINATION OF THE HYDROCARBON-INDEX DIN EN ISO :2000. APPLICATION NOTE As of July 2017, Version: 1.

SEC-MALS Noise Assessment and Cleaning Guide

PURITY GRADES OF LAB REAGENTS

IMPROVING THE QUANTITATIVE REPRODUCIBILITY OF AN ATMOSPHERIC PRESSURE CHEMICAL IONISATION SOURCE FOR GC/MS AND GC MS/MS ANALYSIS

Artisan Technology Group is your source for quality new and certified-used/pre-owned equipment

Agilent 1100 Series Autosampler Thermostat Reference Manual

Agilent 1290 Infinity Thermostat

Technical specifications of FT-IR Spectrometer

NOTE: The color of the actual product may differ from the color pictured in this catalog due to printing limitation.

Drop-in Testing of Next-Generation R134a Alternates in a Commercial Bottle Cooler/Freezer

Installing a Fast Oven Upgrade Kit

APPLICATION NOTEBOOK FOR SOLID PHASE EXTRACTION

DETERMINATION OF PAH In Particulate Matter PM10 With SPE/EVAporation

Common Features of DX-600 Modules

Thermo Scientific Rocket Evaporator. Revolutionary solvent evaporation complete laboratory workflow

Agilent 7890A Network Gas Chromatograph

Jetstream Series HPLC - PELTIER COLUMN-THERMOSTATS 1. GENERALITIES

Thermo Scientific Dionex Integrion HPIC System

IC Professional Detector

Fundamentals of Solid-Phase Extraction (SPE): Principles and Practical Tips for Developing Sample Preparation Methods

TYPES AND APPROPRIATE INJECTORS

Jülich Solar Power Tower System Behavior During Downtime

get mass. without massive work. [ 3100 MASS DETECTOR ]

JSB is an authorised partner of DETERMINATION OF THE HYDROCARBON-INDEX DIN EN ISO :2000. i n s p i r a t i o n m e e t s i n n o v a t i o n!

WELCOME TO ELLUTIA CHROMATOGRAPHY SOLUTIONS

Simplicity. Reliability. Performance. Stratum. Automated VOC Sample Prep System

GC Injectors. GC Injectors

SCION 436-GC. Dimensions and Weights

- ibidi heating system - stage top incubator for life cell imaging

22 The Flame Ionization Detector

Gilson 151/152 UV/VIS Detectors User s Guide

SRI 8610C GC for PCB in Transformer Oil Analysis November 2014

Transcription:

An Advanced Heat Exchanger for the Agilent 129 Infinity Thermostatted Column Compartment Technical Overvie w Author A. Gratzfeld-Huesgen Agilent Technologies, Inc. Waldbronn, Germany Abstract Agilent designed an advanced heat exchanger with improved usability for the Agilent 129 Infi nity Thermostatted Column Compartment. The innovative heat exchanger is easy to install with just three screws, and has a removable capillary from the heat exchanger to the column. This Technical Overview demonstrates that the heat exchanger shows no signifi cant differences compared to the previous version with respect to retention times, resolution, and heating and cooling performance. 6 5 3 1 Old heat exchanger.5 1. 1.5 2. 2.5 3. 3.5 4. 4.5 5. min

Introduction Temperature control in HPLC and UHPLC is mandatory for highly precise retention times. Another demand is that the delay volume of the heat exchanger should be as low as possible to avoid additional peak dispersion. Therefore, Agilent previously introduced heat exchangers with internal volumes of only 1.6 µl. An advanced heat exchanger has been introduced with the same internal volume (p/n G1316-822), but with improved usability, robustness, and convenience. If changes are made in a column compartment, tests are needed to prove that existing methods result in the same retention times and resolution as before. Advanced and original heat exchangers were compared at different temperatures using sulfonamides. The retention times of these compounds are very sensitive to temperature changes. Differences in retention times and resolution were calculated for gradient analysis. In addition, the differences at isocratic conditions were evaluated, as well as heating and cooling performance. Experimental Instruments An Agilent 129 Infinity Binary LC was used with: Agilent 129 Infinity Diode Array Detector with 1-mm cell (G4212A) Agilent 129 Infinity Thermostatted Column Compartment (G1316C) equipped with classic L-shaped low dispersion heat exchanger (G1316-83) (existing design) or low dispersion heat exchanger double (G1316-65) (as contained in replacement kit G1316-822) Agilent 129 Infinity Autosampler (G4226A) Agilent 129 Infinity Thermostat (G133B) Agilent 129 Infinity Binary Pump (G422A) Compounds Test compounds for gradient analysis were sulfathiazole, sulfamerazine, sulfamethazine, sulfamethizole, sulfachloropyrazine, and sulfadimethoxine, purchased from Sigma-Aldrich, Corp., Germany. The RRLC Checkout Sample (p/n 5188-6529) for isocratic analysis contained nine compounds, 1 ng/µl each, dissolved in water:acn (65:35): Instrumental conditions 1. acetanilide 2. acetophenone 3. propiophenone 4. butyrophenone ( ng/ml) 5. benzophenone 6. valerophenone 7. hexanophenone 8. heptanophenone 9. octanophenone Gradient analysis Column Agilent ZORBAX Eclipse Plus C18, 2.1 1 mm, 1.8 µm (p/n 959764-92) Mobile phase Water +.1 % TFA, acetonitrile +.1 % TFA Gradient From 1 % to 4 % acetonitrile in 1 minutes Flow rate.3 ml/min Column temperature 25, 4, 6 C Detector 254/1 nm, ref 38/1 nm, 2 Hz, slit 4 nm Injection volume 1 µl, sample cooled to 1 C Stop time 1 minutes Post time 3 minutes Column Agilent ZORBAX Eclipse Plus C18, 3 1 mm, 3.5 µm (p/n 959961-32) Mobile phase Water +.1 % TFA, acetonitrile +.1 % TFA Gradient From 1 % to 4 % acetonitrile in 8 minutes Flow rate 1 ml/min Column temperature 21, 23, 25, 27, 4, and 6 C Detector 254/1 nm, ref 38/1 nm, 2 Hz, slit 4 nm Injection volume 1 µl, sample cooled to 1 C Stop time 8 minutes Post time 2 minutes Column Agilent Poroshell 12 EC-C18, 4.6 5 mm, 2.7 µm (p/n 699975-92) Mobile phase Water +.1 %TFA, acetonitrile +.1 % TFA Gradient From 1 % to 4 % acetonitrile in 5 minutes Flow rate 2 ml/min Column temperature 25, 4, 6 C Detector 254/1 nm, ref 38/1 nm, 2 Hz, slit 4 nm Injection volume 1 µl, sample cooled to 1 C Stop time 5 minutes Post time 1 minutes Isocratic analysis Column Agilent ZORBAX Eclipse Plus C18, 2.1 1 mm, 1.8 µm (p/n 959764-92) Mobile phase Water:acetonitrile, 3:7 Flow rate.3 ml/min Column temperature 4 C Detector 254/1 nm, ref 38/1 nm, 2 Hz, slit 4 nm Injection volume.5 µl, sample cooled to 1 C Stop time 1 minutes 2

Software Agilent OpenLAB CDS ChemStation Rev.C.1.5 Results and Discussion Design of the advanced heat exchanger The advanced low dispersion heat exchanger consists of a metal block and is installed between the fins of the Agilent 129 Infinity Thermostatted Column Compartment by tightening three screws (Figure 1). In contrast to the current L-shaped low dispersion heat exchanger, the column inlet capillary and the capillary from injector to column heater can now be replaced if blocked or damaged. Up to four columns (maximum 1 mm in length) can be placed in one 129 Infinity Thermostatted Column Compartment, as with the existing heat exchanger. Different configurations are possible, and are shown in Figure 1. Performance comparison The following tests were performed to compare the temperature behavior and delay volume of the existing and the advanced heat exchanger: Differences under isocratic conditions Both heat exchangers have an internal volume of 1.6 µl, but they differ in that the capillary from the heater to the column is now removable, in case the fitting is damaged or the capillary is blocked. To demonstrate that no or only minimal performance differences exist, both heat exchangers were tested under isocratic conditions and at a flow rate of.3 ml/min. In Figure 2, the chromatograms of the RR sample are overlaid. Obviously, the difference is very small. The deviation of retention times for peak 5 is.3 %. This means that the peak eluted slightly earlier using the advanced heat exchanger. The deviation of the resolution was 3.4 % for peak 5. This indicates that resolution was minimally diminished using the advanced heater. The deviation of the theoretical plate number for the last peak was 1.4 %, with k = 5.76. Both heat exchangers were tested for differences in performance using isocratic conditions. The influence of small temperature changes from 21 to 23 to 25 to 27 C on retention times and resolution was tested at 1 ml/min and gradient conditions. Figure 1. Different configurations of the advanced heat exchanger in the Agilent 129 Infinity Thermostatted Column Compartment. 6 5 Both heat exchangers were tested for differences in retention time and resolution at.3, 1, and 2 ml/min flow rates at 25, 4, and 6 C and gradient conditions. Temperature equilibration time was monitored over 1 runs for each temperature change. 3 1.5 1. 1.5 2. 2.5 3. 3.5 4. 4.5 5. min Figure 2. Comparison of chromatograms with Agilent advanced and existing heat exchangers using isocratic conditions at.3 ml/min flow rate to analyze the RRLC Checkout Sample. 3

Differences in retention time with small temperature changes It was possible to determine to what extent the two types of heat exchangers differed by applying small temperature changes using sulfonamides to indicate the temperature effects. Gradient conditions and a 1 ml/min flow rate at 21, 23, 25, and 27 C were chosen for the separation of six sulfonamides (Figure 3). The first two peaks, sulfathiazole and sulfamerazine, were partially separated at 21 and 27 C; at 23 and 25 C, both compounds coeluted. Sulfamethazine, peak 3, showed only minimal changes in retention for all applied temperatures. Sulfamethizole, peak 4, sulfachloropyrazine, peak 5, and sulfadimethoxine, peak 6, changed to lower retention times with increasing temperature. The deviation of retention times of the advanced heat exchanger was smaller than.5 % for all temperatures compared to the existing heat exchanger. This corresponded to a temperature deviation of approximately +.2 C measured for the last peak. The resolution deviation was smaller than 1.3 % for all peaks, which meant slightly less resolution. These results proved that both heaters behaved very similarly in this temperature range. Differences in retention time and resolution at.3, 1, and 2 ml/min flow rates and 25, 4, and 6 C Three different temperatures were applied at a.3 ml/min flow rate on a 2.1 1 mm, 1.8 µm column (Figure 4). The deviation of retention times of the advanced heat exchanger was smaller than.7 % for all temperatures and the deviation of resolution was smaller than.4 %. 6 27 C 5 3 1 1.5 2. 2.5 3. 3.5 4. 4.5 5. 5.5 min Figure 3. Influence of small temperature changes on the elution of sulfonamides. 1, 1, 1, 8 6 Figure 4. Overlay of chromatograms of existing and advanced Agilent heat exchangers at different temperatures and a.3 ml/min flow rate. 25 C 23 C 21 C 6 C 4 C 25 C 2 3 4 5 6 7 8 min 4

The same experiment was done using a 3 1 mm, 3.5 µm column at a 1 ml/min flow rate (Figure 5). The deviation of retention times of the advanced heat exchanger was smaller than.5 %, and the deviation of resolution was smaller than 1.6 %. 5 6 C Finally, the experiments were repeated on a 4.6 5 mm, 1.8 µm column and a flow rate of 2 ml/min (Figure 6). The deviation of retention times of the advanced exchanger was smaller than.9 % and the deviation of resolution smaller than 3.4 %. 3 1 4 C 25 C 1. 1.5 2. 2.5 3. 3.5 4. 4.5 5. min Figure 5. Overlay of chromatograms of Agilent existing and advanced heat exchangers at different temperatures and a 1 ml/min flow rate. 5 6 C 3 4 C 1 25 C.5 1. 1.5 2. 2.5 3. min Figure 6. Overlay of chromatograms of Agilent existing and advanced heat exchangers at different temperatures and a 2 ml/min flow rate. 5

Table 1 summarizes the results, with more detailed information. Only resolution values that are lower with the advanced heat exchanger are considered. Table 1. Deviation of retention times and resolution at.3, 1, and 2 ml/min flow rates and at 25, 4, and 6 C using the Agilent advanced low dispersion heat exchanger double in comparison with the existing low dispersion heat exchanger. Temperature/ flow rate at.3 ml/min Maximum RT Maximum negative Rs at 1 ml/min Maximum RT Maximum negative Rs at 2 ml/min Maximum RT 25 C.28.4.7.1.76 2.2 4 C.7..21 1.6.88 3.35 6 C.63.3.5 1.3.36. Column dimension 2.1 1 mm, 1.8 µm 3 1 mm, 3.5 µm 4.6 5 mm, 1.8 µm Maximum negative Rs Temperature equilibration time monitored over 1 runs for each temperature change The heating and cooling behavior of the heat exchangers was tested by changing the temperature from 25 to 4 C and from 4 to 6 C, and vice versa. Ten runs were done at 25 C, followed by 1 runs at 4 C, and 1 runs at 6 C. During the change of temperature, no additional equilibration time was programmed. The next line of the sequence with the higher (lower) temperature started immediately. If, for example, after five runs the peaks overlaid exactly, it was assumed that the temperature was equilibrated. Figure 7 shows an example, applying a 1 ml/min flow rate and a temperature change from 4 to 6 C using the existing heat exchanger. 5 After 18 minutes 3 After 48 minutes 1 Start minutes first run 1. 1.5 2. 2.5 3. 3.5 4. 4.5 min Figure 7. Temperature equilibration time after the change from 4 to 6 C for the existing heat exchanger. 6

Table 2 summarizes the results for the different temperature changes at different flow rates. It was found that the advanced heat exchanger, especially at higher flow rates, equilibrated three to four times faster than the previous version. This was mainly due to the new design that uses one massive aluminum block. The cool-down experiments showed no significant difference for both heaters. The cool-down time depended not only on the heat exchanger itself, but also on complete oven environment, which significantly influenced the cool-down time until equilibration was reached. Table 2. Result of different temperature changes at different flow rates. Temperature change/flow rate At.3 ml/min At 1 ml/min At 2 ml/min Exisiting New Exisiting New Exisiting New 25 to 4 C 39 26 72 24 54 18 4 to 6 C 39 26 72 24 48 12 Column dimensions 2.1 1 mm, 1.8 µm 3 1 mm, 3.5 µm 4.6 5 mm, 1.8 µm Conclusions Agilent designed an advanced low dispersion heat exchanger for the Agilent 129 Infinity Thermostatted Column Compartment. Comparison of the advanced and existing low dispersion heat exchanger showed no significant differences in retention time, resolution, or plate number. The deviation of retention time was found to be smaller than.9 % from 21 to 6 C and the resolution differed less than 3.4 %. In addition to more convenient handling and advanced robustness, temperature-heating equilibration times were three times faster using the advanced heat exchanger. 7

www.agilent.com/chem This information is subject to change without notice. Agilent Technologies, Inc., 214 Published in the USA, October 1, 214 5991-5166EN