Hydrogen Generator Model HPNM 1. Instruction manual. 166 Keystone Drive Montgomeryville, PA Phone: Fax:

Similar documents
Hydrogen Generator Model HPNM 2

EW 40 Wireless Fan Control

OPERATING MANUAL CUSTOMER SERVICE NO

Homeowner s Installation Instructions & Operating Manual

Portable Air Conditioner User Manual

APC BC300 Series 40kW 208/450/480V User Guide

User Manual. New Hy-PEM XP Rack Hydrogen made simple

Homeowner s Installation Instructions & Operating Manual

Homeowner s Installation Instructions & Operating Manual

Wireless Thermostat (WTS10) Keypad Operation Guide

IMPORTANT SAFETY INFORMATION! WARNING ALWAYS keep electric cords, home furnishings, drapes, clothing, papers, or other combustibles at least 3 feet (0

1126 Series Ceiling Mount PIR Motion Detector

GARAGE HEATER WITH REMOTE INSTRUCTION MANUAL MODEL: HA24-100E HA24-150E. Figure 1

IFT-RC150 IntelliFire Touch Remote Control Installation Instructions

Micro-Computer Induction Cooktop SR-1885SS

Welcome to the World of In-Home Gardening! Simple Steps to Get Your AeroGarden Up and Growing (No Tools Required)

Personal Ultrasonic Humidifier. FALLH-01 series

Interactive Technologies Inc North 2nd Street North St. Paul, MN Technical Manuals Online! -

Installer Guide smart connect

SOOTHING COOL MIST AROMA DIFFUSER

The Digital Door Viewer! User Manual. Motion Activated Sensor MAS 100

MILLA ELECTRIC FIREPLACE

Ionic Breeze QUADRA Silent Air Purifier. SI637 SI697 Instructions

CONNEX Smart Thermostat CST R00

Please Keep This Manual For Future Reference REV.01

D8024, D9024, D10024 Analog Fire Alarm Control Panels Programming Guide

PAP Disinfecting Device User Guide. For all PAP Devices (Positive Airway Pressure) CPAP, APAP, Bi-PAP

Disclaimer. Trademarks. Copyright. Warranty

SUNRISE ALARM CLOCK. With Multicolor LED Lights HME040187N. Before using your Sunrise Alarm Clock, please read this manual.

BBQ THERMOMETER DESCRIPTION OF PARTS. MODEL NO.: AW129/AWR129 Instruction Manual

Homeowner s Installation Instructions & Operating Manual

EVAPORATIVE COOLER W

PRODUCT DIAGRAM PACKAGE CONTENTS

Welcome. Sensor. 1 (844) LIGHTCLOUD Custom manufactured in China RAB Lighting, Inc 170 Ludlow Avenue Northvale, NJ 07647

THERMAX Efficiency Monitor

180 Degree Ceiling Mounted Occupancy Sensors

Installer Guide smart connect

Ultrasonic Aromatherapy Diffuser with Color Changing LED Lights Instruction Manual and Warranty Information - Read and Save These Instructions.

Ambient Weather WS-0211 Wireless Wendy the Weather Wizard User Manual

Color-Changing LED Aromatherapy Diffuser

Ultrasonic Wall Switch Occupancy Sensor Single Circuit

Scan the below QR code to download Smart Flame

Independent Zone Control (I.Z.C.)

Model: T83653v2 Instruction manual DC: WIRELESS COLOR FORECAST STATION Mold Risk for Forecast Station & Indoor Remote Sensor Location

23-IN Electric Logset

FV-11VQL6. Thank you for purchasing this Panasonic product Back cover FV-15VQL6

CLEAN MIST SMART HUMIDIFIER Instruction Manual and Warranty Information IM0031B READ AND SAVE THESE INSTRUCTIONS

SAVE THESE INSTRUCTIONS

Blueair C User Manual

With Remote Control. English. Model

Ambient Weather WS-28 Indoor/Outdoor Thermometer with Daily Min/Max Display User Manual

Safety Instructions. Be sure to observe the following instructions when using the device.

IMPORTANT SAFETY INSTRUCTIONS

Refrigerated Incubator Model and Operating Instructions

With Remote Control. English

IMPORTANT. Display for 5-in-1 Weather Sensor model 06005RM/1010RX SAVE THIS MANUAL FOR FUTURE REFERENCE. Package Contents

WatchDog Wireless Crop Monitor Operation Manual

USER MANUAL. Version: 1.0. ROBE Show Lighting s.r.o. Czech Republic

RGR150 USER S MANUAL. Wireless Rain Gauge with Thermometer and Clock

PTAC WIRELESS KITS (DT01*, DS01*, DD01*) INSTALLATION INSTRUCTIONS

PRODUCT DIAGRAM PACKAGE CONTENTS

INSTRUCTIONS FOR USE OF COMBINED BOILER INTENDED FOR COMBUSTION OF BOTH PELLETS AND SOLID FUEL ABC COMBO

Solution. Supply single or multiple GC instruments Continuous Emission and Air monitoring Mud logging application for oil industry

Operating instructions Page 12

Towel Warmer MAKE EVERYDAY A SPA DAY

FEATURES AND SPECIFICATIONS

COMBINATION ELECTRIC FIREPLACE/HEATER Model #FP23-1A-AKM

3 In 1 AIR CONDITIONER with REMOTE CONTROL MODEL NO: CA9000 PART No: OPERATION & MAINTENANCE INSTRUCTIONS

OPERATION MANUAL WARNING TO REDUCE THE RISK OF INJURY OR PRODUCT DAMAGE, READ OPERATION MANUAL PRIOR TO OPERATING PRODUCT.

AGES. instruction manual DAISY BLOOM PROJECTION ALARM CLOCK

Caution: To maintain compliance with the RF exposure guidelines, place the unit at least 20cm from nearby persons.

Dual Column Dehydrating Breather Manual

Ion Gateway Cellular Gateway and Wireless Sensors

&Fan. Instruction Manual. ULTRASlimline Heater. Model: (with remote control) Voltage: 120V AC Frequency: 60 Hz Power: 1500 W

DEHUMIDIFIER. User Manual 50BT, 70BT

Installation and Operation Guide

Operating Instructions High-spec Wired Remote Controller

Instruction Manual. AcuRite Atlas. Indoor Display model 06061

Thermometer model 02059

Honeywell. Wireless Rain Gauge with Indoor. Temperature (TC152) USER MANUAL TABLE OF CONTENTS INTRODUCTION 3 PRODUCT OVERVIEW 4 REMOTE RAIN GAUGE 7

Aqua Balance. AquaBalance TM CONTROL MODULE QUICK START GUIDE LEGEND

External Wireless Sounder

Operator s Manual. Mirage. Buffet Induction Warmers

USER MANUAL. Intelligent toilet Primus

Dryer Controller M720

English. FCC declaration. Electro Magnetic Compatibility (EMC) Warning. This device complies with part 15 of the FCC rules.

6 Sheet Strip-Cut Shredder for Home/Office

Passive Infrared (PIR) Wall Switch Occupancy Sensor Dual Circuit

CONTENTS CONTROLS 3-STAGE AIR PURIFICATION. X This product is suitable for 120V only. User Manual. Odor sensor. ➌ PlasmaWave

Wireless Home Appliance Timer Remote Control Kit

Ambient Weather WS-091-C Three Channel Display Wireless Thermometer (Console Only) User Manual

888 TWO WAY CAR ALARM SYSTEM USER S MANUAL

#1 BRAND IN MASS AGE* HMC-100MASSAGECHAIR INSTRUCTION MANUAL AND WARRANTY INFORMATION ONE YEAR LIMITED WARRANTY HMC-100

838 CAR ALARM SYSTEM USER S MANUAL

HDCVI PTZ Camera. 6C Series. Quick Start Guide Version Dahua Technology USA Inc.

CLEAN MIST ULTRASONIC HUMIDIFIER. Instruction Manual and Warranty Information READ AND SAVE THESE INSTRUCTIONS IM0040A

Manual Feed Slot Opening. Wastebin. Credit Card Casters

Installation Guide. Model TBZ48A Battery Powered Z-Wave Thermostat. This thermostat is compatible with most HVAC systems, including the following:

PR-L2466W- PA. Operating Instructions. High Performance Refrigerator PR-L2466W-PA

Smart Hub THERMOSTAT. Installation Manual

Transcription:

Hydrogen Generator Model HPNM 1 Instruction manual 166 Keystone Drive Montgomeryville, PA 18936 Phone: 215-641-2700 Fax: 215-641-2714

Introduction Index INDEX... 2 INTRODUCTION... 3 SCOPE OF THE MANUAL... 3 SPECIFICATIONS... 3 NOTES ON FCC COMPLIANCE... 4 CORRECT USE... 4 PACKING LIST... 5 DESCRIPTION... 6 INSTALLATION... 7 RECEIVING THE GENERATOR... 7 PLACING THE GENERATOR... 7 SYMBOLS USED ON THE GENERATOR... 7 GAS CONNECTIONS... 7 ELECTRICAL CONNECTIONS... 8 REMOTE CONNECTIONS (OPTIONAL)... 8 CASCADING (OPTION)... 9 INITIAL START-UP... 10 FILLING THE WATER TANK...10 INSTALLING THE DEIONIZER BAG... 11 OPERATION... 12 MAINTENANCE... 16 ROUTINE MAINTENANCE... 16 RETURNING THE UNIT... 16 SPARE PARTS LIST... 17 Model HPNM-1 2

Introduction Introduction Scope of the manual This manual provides operation and maintenance instructions for the following Chrysalis II HPNM Models: HYC-HPNM-100, HYC-HPNM-160, HYC-HPNM-250 and HYC-HPNM-500 hydrogen generators. Specifications Table 1 Specifications of the different models of hydrogen generator Hydrogen flow rate Model HPNM-100 0-100 cc/min at STP Model HPNM-160 Model HPNM-250 Model HPNM-500 0-160 cc/min at STP 0-250 cc/min at STP 0-500 cc/min at STP STP Standard temperature and pressure (STP) 68 F & 90 psig Weight (dry) 42 lbs. Power consumption Model HPNM-100 90VA Model HPNM-160 Model HPNM-250 Model HPNM-500 115VA 160VA 300VA Input voltage Fuse 110-230V / 50-60Hz 4A (5x20) Pressure accuracy 1.5 psig (± 0.5 %) Microprocessor controlled display Graphic display, 128 x 256 pixels Index of protection Operating conditions: - Temperature - Relative humidity -Max Altitude Over voltage category IP2x 59 F to +95 F 0-80% @ 87 F, decreasing linearly to 50% @ 95 F 6,890 ft II Pollution degree 2 Sound pressure level 46 dba Case dimensions 8.6 W x 9.5 L x 16 H Model HPNM-1 3

Introduction Notes on FCC compliance This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference in a residential installation. This equipment generates, uses and can radiate radio frequency energy and, if not installed and used in accordance with the instructions, may cause harmful interference to radio communications. However, there is no guarantee that interference will not occur in a particular installation. If this equipment does cause harmful interference to radio or television reception, which can be determined by turning the equipment off and on, the user is encouraged to try to correct the interference by one or more of the following measures: Reorient or relocate the receiving antenna. Increase the separation between the equipment and receiver. Connect the equipment into an outlet on a circuit different from that to which the receiver is connected. Consult the dealer or an experienced radio/tv technician for help. WARNING! Any changes or modifications to this equipment not expressly approved by the manufacturer may void the user's authority to operate the equipment. Correct use The Matheson Tri-Gas Chrysalis II HPNM hydrogen generator is designed to produce hydrogen for laboratory use. The unit must only be operated for this purpose, according to the specifications and instructions provided in this manual. In particular, the following warnings must be observed at all times: Indoor use only Never operate the unit in freezing temperatures. This will cause irreversible damage to the electrolysis cell. Only use pure water (see Filling the water tank ) Only operate the unit in a room with sufficient ventilation (see Placing the unit ). Always unplug the unit from the main power supply before accessing the internal components for replacement. Only the parts described in the Spare parts list can be replaced by the user. Model HPNM-1 4

Introduction Packing list Table 2 List of items included in the shipment Quantity Description 1 Hydrogen generator 1 Instruction manual 1 Deionizer bag 1 Water drain with flexible tubing 1 Power cable Model HPNM-1 5

Installation Description The hydrogen generator produces pure hydrogen (and oxygen as a by-product) by the electrolysis of water. The key element of the generator is an electrochemical cell assembly which contains a solid polymer electrolyte. No free acids or alkalines are used. De-ionized or pure, distilled water is the only liquid which may come into contact with the cell. As this is consumed it must be refilled from time to time as required. The generated hydrogen gas is accumulated in the hydrogen/water separator and the desiccant housing. The pressure is controlled by a pressure transducer. The outlet pressure is indicated on the display. The hydrogen is dried by passing it through a drying tube and the automatic dryer. The hydrogen then passes through outlet port at the rear. Hydrogen Vent Low Pressure Gas/Liquid Separator Drying Tube Hydrogen Pressure Relief Water Tank Gas/Liquid Separator High Pressure Automatic Dryer Gas/Liquid Separator O2 Vent Hydrogen Cell Pressure Sensor Hydrogen Outlet Drain Port Model HPNM-1 6

Installation Installation Receiving the generator All units have been carefully inspected before transport. Visual checks for damage and functional tests should be performed upon receipt. Any damage must be immediately noted and reported. The generator must only be returned according to the shipping instructions provided. Placing the generator The hydrogen generator must be placed on a flat, level, vibration-free, shock-free surface. Do not place the generator over a source of heat, as this may cause the device to overheat. The unit should not be in contact with any other objects on any side, and the air inlet must not be blocked. Leave at least 12 cm of free space at the rear for ventilation. Do not operate the generator in a sealed or unventilated room, or in close proximity to open flame or other sources of ignition. Do not operate the generator at below freezing temperatures. Operation is guaranteed at operating temperatures between 45 and 95 F. WARNING! Normal precautions for any hydrogen supply should be taken when using the generator. DO NOT use in sealed or unventilated rooms. DO NOT use in close proximity of open flames or other sources of ignition. Symbols used on the generator Earth symbol: This symbol marks the earth connections to the chassis of the hydrogen generator. Gas connections Pure dry hydrogen at regulated pressure is available at the hydrogen outlet port at the rear of the generator. This port must be connected to 1/8" tubing using a stainless-steel or copper Swagelok connector. Teflon connectors are not suitable. The pressure at this port is adjusted and shown on the display. The hydrogen relief port at the rear of the unit can be connected to an exhaust hood or other vent system. WARNING! The line from the relief port should never connect in such a way that back pressure can develop. Model HPNM-1 7

Installation IMPORTANT! Remove the plugs from the oxygen vent and hydrogen vent before operating the unit. Keep these plugs for transporting the unit. Electrical connections Check the setting of the voltage selector on the rear of the unit. The set voltage is indicated by the white arrow. To change the voltage, proceed as follows: Using a small screwdriver, remove the voltage selector insert. Replace the voltage selector insert so that the white arrow points to the correct voltage. Remote connections (optional) The hydrogen generators are fitted with an optional remote control feature, which allows the user to check the status of the machine from a remote position, and to start/stop the production of hydrogen. The contacts used in the remote control are potentially free relay contacts. The contacts can be configured via software as normally-open or normally-closed (see the Configuration section). The maximum voltage and current ratings for the contacts are 1A / 48V. The pin configuration of the remote connector is shown in the table below. Table 3 Remote connector pin configuration Pin Description 1+2 Start (12-30 VDC polarity not important) 3+4 Standby (System not OK) 5+6 Reaching normal pressure (Overproduction) 7+8 Refill water (Low water) 09+10 Low water level (Too low water) 11+12 Bad water 13+14 Change water (Bad water pre alarm) Model HPNM-1 8

Operation Cascading (Option) The RS-485 interface allows up to 10 generators to be operated in parallel mode. One unit has to be defined as the master, while the others operate in Slave mode. All the slaves need to be configured with individual ID numbers. Communication between the generators requires a standard D-sub 9 pin serial cable. The serial ports are connected as follows: Master RS 485 port 1 Slave 1 RS 485 port 1 - Slave 1 RS 485 port 2 Slave 2 RS 485 port 2 Configuration For operation in cascading mode all generators must have an ID number. Each ID number is unique. The master unit must also have an ID number. Configuring the Master 1. Go into the configure menu and set the master to NO. 2. Go into the number of slaves menu and set the number off units connected to the master. 3. Go into slave ID number and set the desired value. 4. Go back to master and set the value to yes Configuring the Slaves 1. Go into the configure menu and set the master to NO. 2. Go into slave ID number and set the desired value. The configuration is now complete. Operating in Master-Slave Mode If the configuration and the serial connection are correct, the slaves will show Slave Mode when powered up. Connect the gas outlets of all the generators to the same line. IMPORTANT! The cascading function will only work properly if the gas outlets on all the generators are connected to the same gas line. Model HPNM-1 9

Operation Initial start-up Filling the Deionized Water Reservoir (Tank) To fill the hydrogen generator with deionized water, remove the black plug under the sliding arm at the top of the unit. Carefully fill the tank with deionized water to within ½ of the MAX level indicator located at the sight glass on the front of the hydrogen generator. Move the sliding arm to close the opening of the tank. The conductivity of the water used in the generator must not exceed 2µS. WARNING! Do not fill the water tank higher than the marked max level. CAUTION! To prevent contamination of the cell assembly, it is important to use only deionized water in the generator. Water containing metallic impurities will contaminate or damage the cell, and will void the warranty. Model HPNM-1 10

Operation Installing the deionizer bag After having filled the tank with water, the deionizer bag (supplied) must be placed in the tank. Inspect the bag thoroughly for holes or tears, indicated by loose deionizer beads on the outer surface. If the bag is damaged in any way, discard and replace it with a new one. Only use original parts (see Spare Parts). Wash the deionizer bag in deionized water before proceeding. Insert the free end of the T fastener through the hole in the centre of the water filler cap, until it is securely fastened. The bag should not block the outlet at the bottom of the tank. Once in place, the bag should not be allowed to dry out. Model HPNM-1 11

Operation Operation The operating status of the unit is shown on the main screen on the graphic display. The main screen has three options at the bottom, corresponding to the three buttons on the unit, which are used to run the various functions and access the configuration and diagnostics of the unit, following the tree structure shown in the figure below. Status: Normal Flow Press act: 6.2 set: 6.2 Bar Flow Water Start Open Menu Scroll Menu Pressure adjust Configure Utilities Select Pressure Adjust 6.2 Bar Pressure units Volume units Temp units Scroll Configure Select Utilities Alarm list Diagnostics Special functions Scroll Select Pressure units Bar Volume units scm Select Pressure units Volume units Temp units Pressure Rise Pressure drop delay Auto start Beeper Master Number of slaves Slave ID number Remote start /stop mode Remote relay mode Pre alarms in alarm list Lock keyboard Display contrast Alarm list Alarm Number: 1 21.01.03 17:00 Low water Diagnostics Prod. Tot. : 10000 scm Oper. time : 100 H W. Quality: 2.5 us Prod. Tot. : 10000 scm Oper. time : 100 H W. Quality: 2.5 us Cell Volt. : 3.6 V Cell Curr. : 33 A Cell Vpeak: 3.8 V PS Temp: 40 C PS Temp peak : 45 C Scroll Special Functions Dryer Leak Test Complete system test Select Dryer Leak Test Complete system test Toggle valve Set clock Select Temp units C Master No Dryer Status Temp col 1 : 80 C Temp col 2 : 25 C Cycle time : 480 min Select Dryer Dryer Status Dryer conditioning Heater test Scroll Select Pressure rise 0.1 Bar/min Number of slaves 3 Dryer conditioning Time : 15000 s Time remaining : 0s Temp Col. 1 : 80 C Temp Col. 2 : 25 C Start Stop Leak Test Leak test: Stop Press Act. : 6.0 Bar Start Stop Pressure drop delay 60 s Slave ID number ID: 1 Heater test Heater 1: Stop Heater 2: Stop Start Stop Complet system test Leak test: Stop Heater 1: Stop Heater 2: Stop Start Stop Auto start Off Remote start/stop mode Start/Stop Pre alarms in alarm list Yes V1: open V2: close V3: open V4: close Scroll Toggle valve Off Beeper Off Remote relay mode Normaly closed Lock Keybord No Display contrast 5 Model HPNM-1 12

Operation Figure 1 Menu tree for the operation of the unit Model HPNM-1 13

Operation Configuration parameters Item Description Options / Range Default Pressure units Sets the desired unit of measure for bar / psi / kpa bar the pressure Volume units Temp. units Pressure rise Pressure drop delay Auto start Beeper Master Number of slaves Slave ID number Remote start/stop mode Remote relay mode Pre alarms in alarm log Lock Keyboard Sets the desired unit of measure for the volume Sets the desired unit of measure for the temperature Sets how fast the pressure has to increase. If the pressure increases at a slower rate, a low pressure alarm is activated. Sets a delay in seconds to ignore a pressure drop (override low pressure alarm) Sets whether the unit automatically starts production when power is switched on. Sets whether the audible signal is activated in the event of an alarm. Configures the unit as the Master for cascading operation Enter the number of slaves connected to the master scm (standard cubic meters) scf (standard cubic feet) scm C and F C 0.1-1 bar/min 1.4-14 psi/min 0.1 1.4 0-60 s 60 YES / NO ON / OFF YES / NO NO ON NO 0-32 0 Sets the slave ID number 0-32 0 Configures the remote start/stop function Configures the remote relay contacts. If set to Yes, the pre alarms are also shown in the alarm log. If set to Yes, the keyboard will be locked automatically after the generator is in the main window for more than 20s. To unlock the keyboard, press the unlock button and hold for 5s. Start/stop, Start only, Direct control Normally open (NO) Normally closed (NC) YES / NO YES / NO start/stop Display contrast Adjusts the contrast of the display. 0-10 5 NC NO NO Model HPNM-1 14

Operation Diagnostic display Item Description Max. Production Tot. Total production of hydrogen 99.999 scf 4000.00 scm Operating time (h) Total number of hours the unit operation 99.999 hours Wat. quality (µs) Actual water conductivity - Cell current (A) Actual cell current - Cell voltage (V) Actual cell voltage - Cell voltage peak (V) The maximum cell voltage in the life of the cell - PS. temp. ( C) Actual temperature of the power supply - PS. temp. peak ( C) The maximum temperature of the power supply reached - Special functions Item Dryer Dryer status Conditioning Heater test Leak test Complete system test Toggle Valve Description Accesses the dryer functions (see below) Shows the temperatures of the drying columns and the position of the cycle When the conditioning cycle is started, the dryer performs 4 fast cycles. During this time the output valve is closed. The heater has to show a reaction of 10 C to pass the test The outlet valve is closed, the pressure is set to the max., when the pressure reaches the max., production is stopped and the pressure drop is measured over 1 minute. If the pressure drop is below the preset value, the test is passed. Important: The leak test will only work if the generator is in standby mode. Combination of leak test and heater test Used to switch the valves manually ( for advanced diagnostics only) Important : after exiting this window, the valves will return to the status prior to entering the window Normal Display Item Start/Stop Open/Close Menu Description Starts/Stops the production of hydrogen Opens/Closes the hydrogen supply valve Accesses the Pressure Adjust, Configure and Utilities menus Model HPNM-1 15

Maintenance Maintenance With proper care and maintenance, your hydrogen generator should provide you with years of trouble-free operation. There are no adjustments to be made to the generator. The only routine service operations are those described below. Nonetheless, the generator should be inspected approximately every 2 years. Contact your supplier or Matheson Tri-Gas directly. Routine maintenance The following section describes the maintenance operations required for the correct operation of the hydrogen generator. Cleaning The internal components of the hydrogen generator do not need to be cleaned and should not be accessed by the user for cleaning. To clean the outside of the unit, only use a damp cloth (no detergents, acids or aggressive or abrasive substances. Water refilling The tank must be refilled when the water level approaches the lower level, and the Refill Water pre-alarm message appears. Deionizer bag replacement Rinse the water tank and replace the deionizer bag approximately every six months, or whenever the Change Water message appears. Installing the new deionizer bag See page 11 (see Installing the deionizer bag) Returning the unit In the event of any faults or damage, first notify the agent or distributor who supplied the unit. If this is not possible, inform Matheson Tri-Gas directly. Please also provide full details of the problem, including the model and serial number. Instructions will then be provided for the service or the return of the unit. Only if return authorization is provided by Matheson Tri-Gas as per these instructions, will the device be received and repaired by Matheson Tri-Gas. If the one year warranty has expired, or the fault is due to misuse of the unit, all repair and shipping costs are to be paid by the customer. All other costs are borne by the customer, except as otherwise expressly agreed upon. WARNING! If the unit has to be transported, make sure that the water tank is completely empty, and place the plug (supplied with the unit) on the oxygen vent at the rear of the unit. Close the water tank with the cap. Use suitable packaging. The unit should be transported in an upright position; this warning should be reported on the outside of the packaging. Model HPNM-1 16

Maintenance Spare parts list The table below provides a list and description of the replaceable parts of the hydrogen generator. Please also refer to the corresponding figures. Table 4 List of replaceable spare parts Part Number DI-BAG Description Deionizer bag IMPORTANT! The manufacturer reserves the right to change or modify its products without prior notice. Model HPNM-1 17