19 Stationary Dust Monitor EDM 180 APPROVED by U.S. EPA As a Class III Equivalent Method Designation for the measurement of PM 2.5 Automated Equivalent Method: EQPM-0311-195 Federal Register: Vol. 76, page 15974, 03/22/2011 Grimm Aerosol Technik GmbH & Co. KG 1
Optical Technology Giving You Both Multiple PM Values Simultaneously Grimm Aerosol Technik GmbH & Co. KG 2
Principle of Operation Grimm Aerosol Technik GmbH & Co. KG 3
Grimm Laser Aerosol Spectrometer functional principle, light source with laser diode Grimm optical cell like in 180 Light source by compact laser diode with multiplex mode, low power consumption and good temperature stability Collimator generates laser curtain with homogeneous optical field Light trap assures low noise due to very low indirect scattering light Grimm Aerosol Technik GmbH & Co. KG 4
Grimm Laser Aerosol Spectrometer functional principle, aerodynamic aerosol focusing Grimm optical cell like in 180by compact laser diode with multiplex mode, low power consumption and good temperature stability Collimator generates laser curtain optical field Light trap assures low noise due to very low indirect scattering light Inlet focus particles aerodynamically into optical field, no border zone error Rinsing air circuit and outlet nozzle design assure straight pass through the laser, no vagabonding particles Grimm Aerosol Technik GmbH & Co. KG 5
Grimm Laser Aerosol Spectrometer functional principle, PIN diode and wide field reflector Grimm optical cell like in 180 Light source by compact laser diode with multiplex mode, low power consumption and good temperature stability Collimator generates laser curtain with homogeneous optical field Light trap assures low noise due to very low indirect scattering light Inlet focus particles aerodynamically into optical field, no border zone error Rinsing air circuit and outlet nozzle design assure straight pass through the laser, no vagabonding particles Fast and robust photo diode detector for 90 scattering light detection Special wide field reflector improves detection limit for small particles and reduces sizing effects by refractive index Grimm Aerosol Technik GmbH & Co. KG 6
Grimm Laser Aerosol Spectrometer functional principle, output data and gravimetric filter Grimm optical cell like in 180 t source by compact laser diode with multiplex mode, low power consumption and good temperature stability Collimator generates laser curtain with homogeneous optical field Light trap assures low noise due to very low indirect scattering light Inlet focus particles aerodynamically into optical field, no border zone error Rinsing air circuit and outlet nozzle design assure straight pass through the laser, no vagabonding particles Fast and robust photo diode detector for 90 scattering light detection Special wide field reflector improves detection limit for small particles and reduces sizing effects by refractive index 31 size channels, 6 sec. time resolution, PM10, 2.5,1 or TSP or PMc Built in 47mm PTFE filter for gravimetric mass validation Grimm Aerosol Technik GmbH & Co. KG 7
Grimm Laser Aerosol Spectrometer functional principle Grimm optical cell like in 180 Light source by compact laser diode with multiplex mode, low power consumption and good temperature stability Collimator generates laser curtain with homogeneous optical field Light trap assures low noise due to very low indirect scattering light Inlet focus particles aerodynamically into optical field, no border zone error Rinsing air circuit and outlet nozzle design assure straight pass through the laser, no vagabonding particles Fast and robust photo diode detector for 90 scattering light detection Grimm Aerosol Technik GmbH & Co. KG 8
Technology Benefits Measure Both PM 10 and PM 2.5 Real-Time Measurement (6 sec) No Consumable No Zero Drift Minimal Maintenance Quick & Easy Audit Grimm Aerosol Technik GmbH & Co. KG 9
19 Stationary Dust Monitor EDM 180 Easy Rack Mount Installation Grimm Aerosol Technik GmbH & Co. KG 10
19 Stationary Dust Monitor EDM 180 1.5 m Extension for Total length of 3.0 m (2 m above roof) Grimm Aerosol Technik GmbH & Co. KG 11
19 Stationary Dust Monitor EDM 180 Nafion Dryer Under Vacuum Grimm Aerosol Technik GmbH & Co. KG 12
19 Stationary Dust Monitor EDM 180 Looking inside the 180 Grimm Aerosol Technik GmbH & Co. KG 13
Optional Met Sensors for EDM 180 # 157L Temperature Relative Humidity Pressure # 158L Temperature Relative Humidity Pressure Wind Direction Wind Speed # 159L Temperature Relative Humidity Pressure Wind Direction Wind Speed Precipitation New Ultrasonic Wind Sensor with no moving parts. New doppler rain sensor. Both sensors are heated. Grimm Aerosol Technik GmbH & Co. KG 14
Maintenance & Calibration
19 Stationary Dust Monitor EDM 180 Field Test Adapter for the 180 FLOW TEST and ZERO TEST Grimm Aerosol Technik GmbH & Co. KG 16
CALIBRATION TOWER Grimm Aerosol Technik GmbH & Co. KG 17
US EPA PM 2.5 Class III FEM Approval Test Sites Class III Continuous Monitors A: Bakersfield, CA B: Denver, CO C: St-Louis, MI D. New Haven, CT A: Winter & Summer B & C: Winter Only D: Summer Only Grimm Aerosol Technik GmbH & Co. KG 18
US EPA PM 2.5 Class III FEM Approval Grimm Trailer
CONCENTRATION UG/M3 FRM vs GRIMM BAKERSFIELD 80 70 60 50 40 FRM avg Grimm Avg 30 20 10 0 1 3 5 7 9 11 13 15 17 19 21 23 25 27 29 31 33 35 37 39 41 43 45 DAYS Grimm Aerosol Technik GmbH & Co. KG 20
CONCENTRATION µg/m3 FRM vs GRIMM DENVER 30 25 20 15 FRM avg Grimm Avg 10 5 0 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 DAYS Grimm Aerosol Technik GmbH & Co. KG 21
CONCENTRATION µg/m3 FRM vs GRIMM NEW HAVEN 18 16 14 12 10 8 FRM avg Grimm Avg 6 4 2 0 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 DAYS Grimm Aerosol Technik GmbH & Co. KG 22
Family of Products 180 US EPA Approved 107 164 365 665 Grimm Aerosol Technik GmbH & Co. KG 23
CONCLUSION Grimm offer: Simultaneous measurement of PM 10 & PM 2.5 Low Maintenance Fast Response (6 sec) Real-Time Monitoring Insensitivity to vibration (mobile) Lowest Ownership Cost Grimm Aerosol Technik GmbH & Co. KG 24