Operation Manual. Rev. E GEM-MRI. Oxygen Sampling System.

Similar documents
Operation Manual. Rev. E GEM-II Self Contained Gas Detector.

Operation Manual. Rev. D GEM-II Self-Contained Gas Detector.

Falcon-II Next Generation, Air Quality Monitor CO2 & Temperature

Operation Manual. Rev. I SCC Self Contained Controller.

1040 Gas Monitor INSTALLATION AND OPERATING INSTRUCTIONS AMC-1040 WITH INTEGRAL ELECTROCHEMICAL SENSOR

Beacon 800 Gas Monitor Operator s Manual

! "! # Please read these installation and operating instructions completely and carefully before starting.

Controllers. Instruction Manual WARNING

Instruction Manual WARNING

Beacon 200 Gas Monitor Operator s Manual. Part Number: RK Released: 6/6/08

RK CO 2 Transmitter Operator s Manual

GasScanner 8C. Eight Channel Monitor. Operator s Manual. MINT-0281-XX Rev. A 01/29/08

MODEL A-316 CARBON MONOXIDE (CO) MONITOR AND ALARM FOR COMPRESSED AIR TESTING

GG-2 2-CHANNEL GAS DETECTION CONTROL PANEL. Installation and Operation Manual

GASGUARDIAN Channel Controller OPERATING & INSTALLATION MANUAL

Nitrogen Dioxide (NO2) Single-Point Gas Detection System

1022 R/T Remote Monitor

QAS SERIES PLUG-IN CONTROLLERS c/w SOLID STATE SENSOR

Models LBW-420-LEL (24 VDC powered) Ammonia Leak Detector

1228s Multidrop INSTRUCTIONS. Installation and Operation of the AMC-1228s Electrochemical Sensor Module For Use With Multidrop AMC Monitors

GG-R REFRIGERANT SENSOR. Installation and Operation Manual

RK Hydrogen Sulfide Transmitter Operator s Manual

RK-OXY Oxygen Sample-Draw Detector Operator s Manual Part Number: RK Revision: P1 Released: 7/18/02

Model 4001 Series Single Channel Controller

Carbon Dioxide Sample-Draw Detector Operator s Manual

AMC-RAM-4. Refrigerant Alarm Module USER MANUAL. Please read the installation and operating instructions completely and carefully before starting.

RK-05 Combustible Gas Transmitter Operator s Manual

Sentry LIQUID LEVEL CONTROLLER MODEL 120 OPERATING MANUAL.

Oxygen (O2) Single-Point Gas Detection System

RAM3 Remote Alarm Module

Operation Manual. Rev ART Infrared Refrigerant Transmitter.

GasScanner 1C. Single Channel Monitor. Operator s Manual. MINT-0278-XX Rev A

GG-NH3 AMMONIA GAS SENSOR. Installation and Operation Manual

GG-NO2 NITROGEN DIOXIDE GAS SENSOR. Installation and Operation Manual

RK Carbon Monoxide Transmitter Operator s Manual

Combustible Gas Sample-Draw Detector Operator s Manual

GASGUARD NH 3. Ammonia Sensor OPERATING & INSTALLATION MANUAL

WATCHMAN Model LBW-WATCHMAN Ammonia Leak Detector

Carbon Monoxide (CO) Detecting Ventilation Fan Controller Model 120VC Single Relay (100/25 PPM) (200/35 PPM)

Dual Point General Purpose Heat Trace Control TRACON MODEL GPT 230 Installation and Operation Manual

Carbon Monoxide Transmitter

SPECIFICATION MODEL 5200 GAS MONITOR SYSTEM. User Instructions For The Model 5200 Gas Monitor System

GASGUARD O 2. Oxygen Sensor OPERATING & INSTALLATION MANUAL

Pioneer-R16 Gas Monitor Operator s Manual

QAS-101 SERIES SINGLE CHANNEL CONTROLLERS

RK Combustible Gas Transmitter Operator s Manual

RK Carbon Monoxide Transmitter Operator s Manual

RK-02 Multi Point Detector Operator s Manual

MST, INC. MODEL 2002 CARBON MONOXIDE MONITOR (AMBIENT-TYPE) OWNER S MANUAL

GG-NH3-2% AMMONIA GAS SENSOR. Installation and Operation Manual

CT-7 Series Toxic Gas Transmitter Operator s Manual

RK Carbon Monoxide Transmitter Operator s Manual

Beacon 110 Gas Monitor Operator s Manual

Sentry LIQUID LEVEL GAUGE MODEL 200 or 200C OWNER MANUAL REV 1.7 SEPT08 PAGE 1 OF 12

RK-05 Carbon Monoxide Transmitter Operator s Manual

GASGUARD SO2 Sulfur Dioxide Sensor OPERATING & INSTALLATION MANUAL

RK Transmitter Technical Notice

Alarm Tone Generator Model AG17

MST, INC. MODEL 2002 OXYGEN MONITOR (AMBIENT-TYPE) OWNER S MANUAL

RK-05 Carbon Monoxide Detector Operator s Manual

Model Gas Alarm Panel APPLICABILITY & EFFECTIVITY. This manual provides instructions for the following Sierra Monitor products:

GasScanner 4C. Four Channel Monitor. Operator s Manual. MINT-0280-XX Rev. A 02/25/08

MST, INC. MODEL 2002 CARBON MONOXIDE MONITOR OWNER S MANUAL

INSTALLATION. and INSTRUCTION MANUAL. for QUALITY AIR BREATHING SYSTEMS. Model ABM - 700

GG-CO-EXP CARBON MONOXIDE SENSOR. Installation and Operation Manual

GAS MONITOR Model Model

Single Point Freeze Protection Heat Trace Control TRACON MODEL FPT 130 Installation and Operation Manual

Installation, Operation & Service Manual

Model OEM-2 INSTRUCTIONS FOR USE

INSTALLATION. and INSTRUCTION MANUAL. for QUALITY AIR BREATHING SYSTEMS. Model ABM - 715

TCA-9102 Series Surface Mount Temperature Controllers with High and Low Alarm

Carbon Monoxide (CO) Detecting Ventilation Fan Controller Model COSTAR 24VC-e Single Relay

RK/ RK Oxygen Detector Operator s Manual

RK/ RK Oxygen Detector Operator s Manual

Tox-COMB/ANA Combustible Gas Transmitter Operator s Manual

SPECIFICATION CHEMGARD INFRARED GAS MONITOR SERIES

GASGUARD Cl 2. Chlorine Sensor OPERATING & INSTALLATION MANUAL

Warranty Registration

Sentry LIQUID LEVEL ALARM MODEL 100 OPERATING MANUAL.

SEC 2000 Millenium Infrared Gas Detector

IMR IX176 Portable Gas Detector User Manual

ANALOX 101 D2 PORTABLE OXYGEN MONITOR USER MANUAL

Analog Room Pressure Monitor RPC Series

GG-H2S HYDROGEN SULFIDE GAS SENSOR. Installation and Operation Manual

GASGUARD NH 3 2% Ammonia Sensor OPERATING & INSTALLATION MANUAL

Dual Point General Purpose Heat Trace Control TRACON MODEL GPT 230 Installation and Operation Manual

Carbon Monoxide Sample-Draw Detector Operator s Manual

MODEL QTS-1800 SERIES WALL MOUNT DIGITAL AND ANALOG TRANSMITTER/SENSOR

GASGUARD H 2-1% Hydrogen Sensor OPERATING & INSTALLATION MANUAL

GASGUARD H2S EXP. Hydrogen Sulfide Sensor OPERATING & INSTALLATION MANUAL

MODEL GPT-130 SINGLE POINT HEAT TRACE CONTROL THERMOSTAT

Beacon 410A Gas Monitor Operator s Manual

Document No

Confined Space Monitoring. Leak Investigation. Functions 4. Bar Hole Sampling. Line Purge Testing. Actual Size GX-2012

Ammonia Leak Detector Manual

RK/ RK Carbon Monoxide Detector Operator s Manual

Pre-March 2016 Design - Click to be redirected to newer SBS-H2 design SBS-H2. Hydrogen Gas Detector Kit

F PC and AO OUTPUT BOARDS INSTRUCTION MANUAL. Blue-White. Industries, Ltd.

MODEL FPT-130 SINGLE POINT FREEZE PROTECTION HEAT TRACE CONTROL

RK-CH4-4 Combustible Gas Transmitter Operator s Manual

Transcription:

Operation Manual Rev. E2 2015.03 GEM-MRI Oxygen Sampling System www.critical-environment.com

GEM-MRI - Operation Manual Rev. E2 2015.03 Table of Contents 1 POLICIES... 4 1.1 Important Note...4 1.2 Warranty Policy...5 1.3 Service Policy...5 1.4 Copyrights...6 1.5 Disclaimer...6 1.6 Revisions...7 2 INTRODUCTION... 8 2.1 General Description...8 2.2 Key Features...8 3 INSTRUMENT SPECIFICATIONS... 9 3.1 Technical Specifications...9 3.2 Standard Enclosure Dimensions...11 4 SENSOR SPECIFICATIONS...12 4.1 Sensor Specifications...12 5 FEATURES & FUNCTIONS...13 5.1 Exterior Enclosure...13 5.2 Interior System Layout...14 5.3 LPT-A Transmitter Interior System Layout...15 6 INSTALLATION...16 6.1 System Installation...16 6.1.1 Wiring to incoming Power Terminal Block...16 6.1.2 Relay Connections...16 6.2 Sensor Mounting Heights...17 6.3 Enclosure Mounting Components...17 6.3.1 Enclosure Base...17 6.3.2 Enclosure Top...18 6.4 Wiring Connections...18 6.5 Important Notes for Wiring...19 2 2015 All rights reserved. Data subject to change without notice.

Rev. E2 2015.03 GEM-MRI - Operation Manual 7 OPERATION...19 7.1 System Operation...19 7.1.1 Sampled Air Flow Path...20 7.2 Test Functions...20 7.2.1 Voltage Output to Test Points TP-1 and TP-2...21 7.3 Dip Switch Package...21 7.4 Internal Audible Alarm Operation...22 7.5 Jumpers...22 7.6 Adjusting Alarm Set Points...24 7.6.1 Voltage Reference Table for Alarm Settings...26 7.7 Time Delays for Relay Operations...26 7.7.1 Procedure for Setting Time Delays...28 7.8 Temperature Display - Unit of Measure...29 7.9 Latching Relay Functions...29 7.10 LED Digital Display...30 8 CALIBRATION...30 8.1 Calibration Specifications...31 8.1.1 Gas...31 8.1.2 Regulators & Flow...31 8.1.3 Adapters...31 8.2 Calibrating Sensors...31 8.2.1 Calibration Frequency...31 8.2.2 Gas Testing Frequency (Bump Testing)...31 8.3 Calibration Procedure...31 9 Maintenance...35 10 Trouble Shooting...36 10.1 Noisy Pump...36 10.2 Oxygen Sensor - Low Reading...36 11 ACCESSORIES...37 11.1 Calibration Kit...37 11.2 Metal Protective Guard...37 2015 All rights reserved. Data subject to change without notice. 3

GEM-MRI - Operation Manual Rev. E2 2015.03 1 POLICIES 1.1 Important Note Read and understand this manual prior to using this instrument. Carefully read the warranty policy, service policy, notices, disclaimers and revisions on the following pages. This instrument should be inspected and calibrated regularly by a qualified and trained technician. For more information, refer to 9 Maintenance and 8 Calibration sections of this manual. This instrument has not been designed to be intrinsically safe. For your safety, do not use it in classified hazardous areas (explosion-rated environments). INSTRUMENT SERIAL NUMBER: PURCHASE DATE: PURCHASED FROM: 4 2015 All rights reserved. Data subject to change without notice.

Rev. E2 2015.03 GEM-MRI - Operation Manual 1.2 Warranty Policy Disconnect power before servicing Caution: More than one live circuit Supply: 24 V (nominal, 50 or 60 Hz) Certified for electrical shock and electrical fire hazard only Critical Environment Technologies Canada Inc. (CETCI), also referred to as the manufacturer, warrants this instrument to be free from defects in materials and workmanship for a period of two (2) years from the date of purchase. Individual sensor elements have different warranties. Please check with the manufacturer for specific sensor warranty. The warranty status may be affected if the instrument has not been maintained and calibrated as per the instructions indicated in this manual or has been abused or damaged in any way. The manufacturer is not liable for auxiliary interfaced equipment or consequential damage. This instrument is only to be used for purposes stated herein. Warranty does not include third party trouble-shooting costs or freight to / from the manufacturer s facility. CETCI s liability is limited to replacement or repair of the equipment manufactured. Due to ongoing research, development and product testing, the manufacturer reserves the right to change specifications without notice. The information contained herein is based on data considered accurate. However, no warranty is expressed or implied regarding the accuracy of this data. All goods must be shipped to the manufacturer by prepaid freight. All returned goods must be pre-authorized by obtaining a return merchandise authorization (RMA) number. Contact the manufacturer for a number and procedures required for product transport. 1.3 Service Policy CETCI maintains an instrument service facility at the factory. Some CETCI distributors / agents may also have repair facilities; however, CETCI assumes no liability for service performed by anyone other than CETCI personnel. 2015 All rights reserved. Data subject to change without notice. 5

GEM-MRI - Operation Manual Rev. E2 2015.03 Repairs are warranted for 90 days after date of shipment (sensors have individual warranties). Should your instrument require non-warranty repair, you may contact the distributor from whom it was purchased or you may contact CETCI directly. Prior to shipping equipment to CETCI, contact our office for an RMA #. All returned goods must be accompanied with an RMA number. If CETCI is to do the repair work, you may send the instrument, prepaid, to: Attention: Service Department Critical Environment Technologies Canada Inc. Unit 145, 7391 Vantage Way Delta, BC, V4G 1M3 Always include your Returned Merchandise Authorization (RMA) number, address, telephone number, contact name, shipping / billing information, and a description of the defect as you perceive it. You will be contacted with a cost estimate for expected repairs, prior to the performance of any service work. For liability reasons, CETCI has a policy of performing all needed repairs to restore the instrument to full operating condition. Pack the equipment well (in its original packing if possible), as we cannot be held responsible for any damage incurred during shipping to our facility. 1.4 Copyrights This manual is subject to copyright protection; all rights are reserved. Under international and domestic copyright laws, this manual may not be copied or translated, in whole or in part, in any manner or format, without the written permission of CETCI. 1.5 Disclaimer Under no circumstances will CETCI be liable for any claims, losses or damages resulting from or arising out of the repair or modification of this equipment by a party other than CETCI service technicians, or by operation or use of the equipment other than in accordance with the printed instructions contained within this manual or if the equipment has been improperly maintained or subjected to neglect or accident. Any of the forgoing will void the warranty. 6 2015 All rights reserved. Data subject to change without notice.

Rev. E2 2015.03 GEM-MRI - Operation Manual Under most local electrical codes, low voltage wire wires cannot be run within the same conduit as line voltage wires. it is CETCI policy that all wiring of our products meet this requirement. It is CETCI policy that all wiring be within properly grounded (earth or safety) conduit. 1.6 Revisions This manual was written and published by CETCI. The manufacturer makes no warranty or representation, expressed or implied including any warranty of merchantability or fitness for purpose, with respect to this manual. All information contained in this manual is believed to be true and accurate at the time of printing. However, as part of its continuing efforts to improve its products and their documentation, the manufacturer reserves the right to make changes at any time without notice. Revised copies of this manual can be obtained by contacting CETCI or visiting www. critical-environment.com. Should you detect any error or omission in this manual, please contact CETCI at the following address: Critical Environment Technologies Canada Inc. Unit 145, 7391 Vantage Way, Delta, BC, V4G 1M3, Canada Toll Free: +1.877.940.8741 Telephone: +1.604.940.8741 Fax: +1.604.940.8745 Email: marketing@cetci.com Website: www.critical-environment.com In no event will CETCI, its officers or employees be liable for any direct, special, incidental or consequential damages resulting from any defect in any manual, even if advised of the possibility of such damages. 2015 All rights reserved. Data subject to change without notice. 7

GEM-MRI - Operation Manual Rev. E2 2015.03 2 INTRODUCTION 2.1 General Description Thank you for purchasing our GEM-MRI Oxygen (O 2 ) Sampling System. The GEM-MRI O 2 sampling gas detection systems are economical, self-contained gas detectors with a sample pump for O 2 monitoring applications in Magnetic Resonance Imaging (MRI) rooms. The system provides one set of LED indicating lights representing Power, Fail, Low (Warning) Gas Alarm, High Gas Alarm, an integral audible alarm with door mounted silence push-button, audible time delay, field settable relay time delays and two alarm relays. All GEM-MRI O 2 sampling system are packaged with an LPT-A O 2 sensor / transmitter. If after reading through the manual, you have any questions, please do not hesitate to contact our service department for technical support. 2.2 Key Features LED digital display for real time oxygen readings Audible alarm with silence push-button Dry contact relays for controls of remote devices Analog output (4-20 ma) Sample draw pump Adjustable flow meter Oxygen sensor (3 years) Thermal resetting fuse Two 5-amps SPDT relays RoHS compliant circuit boards CSA & UL certified 8 2015 All rights reserved. Data subject to change without notice.

Rev. E2 2015.03 GEM-MRI - Operation Manual 3 INSTRUMENT SPECIFICATIONS 3.1 Technical Specifications GAS TYPE Oxygen (O 2 ) MECHANICAL Enclosure Weight Size Water / dust tight corrosion resistant fiberglass 1.1 kg (2.5 lb) 8.8 x 10.5 x 6.3 (223 mm x 267 mm x 159 mm) ELECTRICAL Power Requirement Current Draw Wiring Fuse 24-28 VAC 500 ma (incoming power to system) 14-18 gauge shielded stranded within conduit Automatic resetting thermal INPUT / OUTPUT Outputs Alarm Relays Linear 4-20 ma Audible alarm (80 db @ 10 ft) Two SPDT dry contact relays, rated 5 amps @ 240 VAC. NOTE: System is configured such that all relays are FAIL SAFE (relay coils are always energized in non-alarm state). USER INTERFACE Display Audible LED Digital Display Common set of LED indicating lights for Power, Fail, Low (Warning) Gas Alarm, High Gas Alarm Integral piezo audible alarm rated 80 db @ 10 c/w door mounted silence push-button. 2015 All rights reserved. Data subject to change without notice. 9

GEM-MRI - Operation Manual Rev. E2 2015.03 Time Delays Delay ON (on make), DIP switch Delay OFF (on break), DIP switch Low alarm relay: 2 minutes High alarm relay: 5 minutes Also known as minimum run time Low alarm relay: 10 minutes High alarm relay: not available Audible (on make), DIP switch 10 minutes NOTE: Time delays can be changed by user. ENVIRONMENTAL (sensor dependant) Operating Temperature -20 C to 50 C (-5 F to 120 F) Operating Humidity 15-90% RH non-condensing CERTIFICATION CSA UL Certified Certified 10 2015 All rights reserved. Data subject to change without notice.

Rev. E2 2015.03 GEM-MRI - Operation Manual 3.2 Standard Enclosure Dimensions 8.8 (223 mm) 10.5 (268 mm) 10.5 (268 mm) 6.3 (159 mm) 2015 All rights reserved. Data subject to change without notice. 11

GEM-MRI - Operation Manual Rev. E2 2015.03 4 SENSOR SPECIFICATIONS 4.1 Sensor Specifications Oxygen (O 2 ) Type Range Response Time (T 90 ) Operating Temperature Operating Humidity Operating Pressure Repeatability Maximum Zero Shift Clean Air Output Drift Expected Life Span Calibration Electrochemical 0-25% volume < 15 seconds -30 C to 55 C (-5 F to 131 F) 5 95% RH non-condensing 80 120 kpa N/A N/A < 2% change in 3 months Approximately 3 years Every 6 months (depending on application) Cross Sensitivity CO 2 sensitivity: 0.1% change in O 2 reading per % CO 2 in 5% CO 2 12 2015 All rights reserved. Data subject to change without notice.

Rev. E2 2015.03 GEM-MRI - Operation Manual 5 FEATURES & FUNCTIONS 5.1 Exterior Enclosure ➎ ➊ ➍ ➏ ➒ ➓ ➋ ➌ ➑ ➐ ➎ NUMBER FEATURE FUNCTION ➊ Digital Display LED digital display ➋ LED Sensor Indicators Indicates Channel 1 or Channel 2 ➌ Door Latch Secures door ➍ LED Indicators Indicates Power & Fail ➎ Door Screws Secures door ➏ Gas Alarm LED Indicates High and Low alarm levels ➐ Alarm Acknowledge-Test Push button for alarm acknowledgement & testing ➑ Audible Alarm Internal audible alarm ➒ Inlet Fitting Sampled air inlet ➓ Outlet Fitting Exhausted air out 2015 All rights reserved. Data subject to change without notice. 13

GEM-MRI - Operation Manual Rev. E2 2015.03 5.2 Interior System Layout ➊ ➌ ➎ ⓫ ➋ ➍ ➑ ➏ ➐ ➓ ➒ NUMBER FEATURE FUNCTION ➊ Dip Switches For access to time delays ➋ Dry Contact Relays For Low & High Alarm ➌ Jumpers For calibrations & other service set-up functions ➍ Wiring Terminal Pluggable wiring terminal ➎ Push Buttons For service access & additional test functions ➏ LPT-A Oxygen Transmitter Oxygen Sensor/Transmitter ➐ Sensor Flow Adapter Diffuses the sampled air to the sensor ➑ Flow meter For adjusting the flow rate ➒ Filter Sample air particulate filter ➓ Incoming Power Terminal Block Incoming power connection, isolated 24 VAC supply (to prevent blowing up the rectifiers) ⓫ Enclosure Mounting Feet & Instructions 14 2015 All rights reserved. Data subject to change without notice.

Rev. E2 2015.03 GEM-MRI - Operation Manual 5.3 LPT-A Transmitter Interior System Layout NUMBER ➊ ➋ ➌ ➍ ➎ ➏ ➐ ➑ FEATURE Enables / Disables the buzzer Rotary Encoder - sets calibration values, etc. with the jumpers Open Loop LED - lights if there is no connection for current output Jumpers - for calibration and test functions Calibrate State LED - further prompting for calibration operation Display Enable - enables / disables the display of the measured gas level Test Points: TP1 & TP2 - for measuring voltage output Wiring Terminal - pluggable terminal for power and signal output Output Select - select either current or voltage output Relay LED - indicates (when on) that the gas level is above the trip point (the relay will indicate alarm level) Relay Terminal - pluggable terminal for relay connections 2015 All rights reserved. Data subject to change without notice. 15

GEM-MRI - Operation Manual Rev. E2 2015.03 6 INSTALLATION 6.1 System Installation The noticeable difference between the GEM-II and a GEM-MRI is the enclosure. A larger, deeper enclosure is required to accommodate the internal LPT-A O 2 sensor / transmitter, sample pump, flow meter, in-line filter and incoming power terminal block. This instrument should be installed outside the MRI room with soft tubing running between the O 2 monitor and the target area to be monitored. Enclosure mounting feet and instructions are secured on the inner door in a plastic bag. Install the enclosure on a flat secure surface using the four mounting feet. After installation, power up and testing, tighten the door screws to prevent rattling from vibration. 6.1.1 Wiring to incoming Power Terminal Block Apply isolated 24 VAC into the incoming power terminal block. Reference the diagram in 5.2 Interior System Layout. WARNING: The transformer should not be connected to any other devices; use a separate transformer for the GEM-MRI. 6.1.2 Relay Connections System relays are dry contacts and designed to operate fan starters or coils to control equipment that draws more than 5 amps start-up and / or operational current. NOTES: 1. The system does not provide any power from these terminals. Dry contacts operate like a switch to simply activate (switch on) or de-activate (switch off) equipment to be controlled, such as fan starters. 2. System relays are SPDT (single pole, double throw) thereby providing one set of usable dry contacts for each relay. Because the GEM series systems are designed to be fail-safe, any equipment to be controlled by the system relays should be wired to the NC Normally closed) and COM (Common) terminals. The relay coils are normally energized in non-alarm state for fail-safe operation. Upon application of power, the green LED light indicator(s) will illuminate and the LED between Sen-1 & Sen-2 will be blinking but all alarms are disabled for 2 minutes for a system warm up period. After the warm up period, the system may exhibit gas alarm condition(s) if one or both 16 2015 All rights reserved. Data subject to change without notice.

Rev. E2 2015.03 GEM-MRI - Operation Manual of the sensors has not completely stabilized during the warm up period. This is normal and the length of time the gas alarms exist is dependent upon the length of time since the unit was last powered up and the state of the environment it is installed in. After warm up, only the green power LED(s) illuminate indicating normal operation and the relays are energized indicating normal Fail-safe status. 6.2 Sensor Mounting Heights GAS APPLICATIONS / TYPES SUGGESTED MOUNTING HEIGHT Oxygen (O 2 ) MRI Rooms 4-6 ft from the floor NOTE: CETCI considers 4-6 ft from the floor as the Breathing Zone when it applies to sensors installed for vehicle exhaust applications. 6.3 Enclosure Mounting Components 6.3.1 Enclosure Base ➊ ➊ ➌ ➍ ➋ ➊ ➊ NUMBER ➊ ➋ ➌ ➍ FEATURE Mounting Holes Conduit Entry Inlet Fitting Outlet Fitting 2015 All rights reserved. Data subject to change without notice. 17

GEM-MRI - Operation Manual Rev. E2 2015.03 6.3.2 Enclosure Top ➊ ➋ ➌ NUMBER ➊ ➋ ➌ FEATURE Door Hinge Conduit Entry Inlet Fitting 6.4 Wiring Connections 18 2015 All rights reserved. Data subject to change without notice.

Rev. E2 2015.03 GEM-MRI - Operation Manual 6.5 Important Notes for Wiring Fan / alarm control wiring is terminated at terminals 1 to 6. The relays are designed to switch fan starters not the fans directly. The relay contacts are rated 5 amps @ 240 VAC each, SPDT. An analog output of 4-20 ma is available. Sensor analog outputs are achieved by wiring to terminals 7 & 8 and observe polarity. Signal output and supply should be in shielded cable. 7 OPERATION 7.1 System Operation In the event of a burned-out, damaged or broken / interrupted 4-20 ma signals from the LPT-A Transmitter, the controller will indicate fail condition on the front door by illumination of the red fail LED. At this point, both system relays will have reversed state (de-energized) and anything connected to them will be running continuously. Normal system operation will not occur until the fault condition has been rectified. NOTES: 1. Fail condition overrides any system time delays that may have been set. 2. Reference photo in 5.1 Exterior Enclosure for location of LEDs, and test-buttons. In the event of a gas build up beyond the preset low alarm trip point, the CH-1-Int. LED (outer door right side) illuminates, the Low alarm level LED (red - outer door left side) illuminates and the low alarm relay de-energizes activating anything controlled by it. If a time delay ON has been set (internal DIP switches), then the Low alarm LED will flash indicating the time delay has been activated and the low alarm relay will not de-energize until the delay has timed out. Once the low alarm relay has de-energized, it will re-energize automatically once the gas alarm condition goes away, unless a time delay OFF (minimum run time - internal DIP switches) has been programmed. The same procedure applies to CH-2 Ext. in the case of a two channel (two sensors) system. In the event of a gas build up beyond the preset high alarm trip point, the CH-1 LED (outer door right side) illuminates, the High alarm level LED (red - outer door left side) illuminates and the high alarm relay de-energizes activating anything controlled by it. If a time delay ON has been set (internal DIP switches), then the High alarm LED will flash indicating the time delay has been activated and the high alarm relay will not de-energize until the delay has timed out. Once the high alarm relay has de-energized, it will re-energize automatically once the gas alarm condition has dispatched. The same procedure applies to the second channel in the case of a two 2015 All rights reserved. Data subject to change without notice. 19

GEM-MRI - Operation Manual Rev. E2 2015.03 channel (two sensors) system. NOTE: No minimum run time delay is provided for the high alarm relay. After power up, the system has a pre-programmed two-minute warm up time delay before displaying any sensor values. Allow sensor several hours to stabilize before making any desired adjustments except for flow rate which can be adjusted right after power up. The system is factory calibrated and programmed. If the flow meter is not indicating the correct flow, it should be adjusted to approximately 2.12 SCFH = 1.0 LPM flow rate after the end user sample hose has been installed. It is not recommended to use a sample hose more than 50 ft long. The inlet and outlet fittings are 3/16 I.D. As with any other sensor, regular calibration maintenance is recommended and required for best performance. Nitrogen (100%) for null adjustment of the sensor and O 2 span gas (20.9% vol.) for span adjustment which can be fed into the inlet fitting. A demand flow regulator must be utilized to avoid drawing down the pump. The calibration of the system can only be done by the LPT-A O 2 analog transmitter. 7.1.1 Sampled Air Flow Path The sample air will enter the system through the inlet fitting then flows through the particulate filter before entering the pump. After exiting the pump, it enters the bottom fitting of the flow meter before flowing into the sensor sample adapter. The sample adapter diffuses the sampled air to the sensor and directs the exhausted air out towards the outlet fitting. 7.2 Test Functions A momentary test feature is provided to allow the user to test basic functionality of the circuit. Press the UP push-button (internal) for 5-seconds, the audible alarm will beep once to indicate it is in the 5-second test function. Release the UP push-button after 5-seconds and an automatic test function will occur. Alternatively use the silence push-button on the front door to perform the same function. The low alarm relay will de-energize and the low alarm LED on the front door will illuminate. The low alarm relay will then re-energize while the low alarm LED stays lit for 10-seconds. 20 2015 All rights reserved. Data subject to change without notice.

Rev. E2 2015.03 GEM-MRI - Operation Manual Next, the high alarm relay will de-energize, the high alarm LED on the front door will illuminate and the audible alarm will activate. The relay will then re-energize but the high alarm LED and the audible alarm will stay activated for 10-seconds. After 10-seconds, everything resets to normal operating condition. If a longer test sequence is desired, hold down the UP push-button for 10-seconds to activate the low alarm relay, high alarm relay, low alarm LED, and high alarm LED for 15-minutes. To abort this test, press the internal UP push-button or the exterior silence push-button and everything will reset to normal operating condition. If the GEM is equipped with a digital display, it will indicate test during this test period. 7.2.1 Voltage Output to Test Points TP-1 and TP-2 NOTE: The voltage output when jumper J6 is set to the resting position 1 is 4.0 VDC. Attach the meter leads to the two test points (TP-1 & TP-2) located just below the high alarm relay (RL V2) on the main circuit board. Set the meter to volts DC with one decimal point. The range of 0-4.0 VDC is equal to the full measurement range of the sensor. Eg. HVAC CO sensor has a standard measurement range of 0-200 ppm. Therefore 4.0 VDC = 200 ppm. Move the Jumper J6 to position 2 when monitoring the Internal Sensor or in position 3 for the External Sensor. 7.3 Dip Switch Package Located upper left corner of main circuit board. Dip switch - position description: 5) Audible on / off 4) Audible alarm time delay on (5-min.) 3) High gas alarm time delay on (5-min.) 2) Low gas alarm time delay off (10-min.) 1) Low gas alarm time delay on (2-min.) 2015 All rights reserved. Data subject to change without notice. 21

GEM-MRI - Operation Manual Rev. E2 2015.03 NOTES: 1. Dip switches in the up position are on. 2. Time delay durations can be changed by user. 3. Fixed delay durations indicated above are factory default settings. 7.4 Internal Audible Alarm Operation If DIP switch 5 is in the up position, the internal audible alarm responds to whatever commands it is given by the microprocessor in the circuit. If DIP switch 5 and DIP switch 4 are in the up position, the audible alarm responds after the user preset time delay function for the audible alarm has timed out. If DIP switch 5 is in the down position, the audible alarm does not respond at all. The internal audible alarm normally activates with a steady tone when a high gas alarm condition exists. The internal audible alarm normally activates with a pulsing tone when a fault condition exists. A fault condition could consist of any of the following: Remote solid-state sensor failure (open loop) Analog signal from a remote LPT-A (4-20 ma) is not registering with the GEM. Smart sensor board is not plugged into the main board. NOTE: Fault conditions automatically override the ON / OFF DIP switch 5 if it has been set to the up position to turn the audible alarm off. 7.5 Jumpers There are two banks of jumpers located directly below the DIP switches. One bank of nine (9) jumpers ( J7 upper bank) and one bank of four (4) jumpers ( J6 lower bank). These jumpers allow the user to perform a wide range of set up and calibration functions. The following table details the jumper settings for each bank of jumpers and explains the function enabled when these jumper positions are selected. NOTE: Always set jumper from J6 first followed by jumper from J7. 22 2015 All rights reserved. Data subject to change without notice.

Rev. E2 2015.03 GEM-MRI - Operation Manual FUNCTION DESIRED J6 LOWER JUMPER BANK J7 UPPER JUMPER BANK Resting (Jumpers for the Upper Bank, J7 are disabled) 1 1 Setting Span Gas Value 2 (Internal Sensor) 2 Perform Zero (Null) Calibration 2 (Internal Sensor) 3 Perform Span Calibration 2 (Internal Sensor) 4 Adjust Low Gas Alarm Ascending Value 2 (Internal Sensor) 5 Adjust High Gas Alarm Ascending Value 2 (Internal Sensor) 6 Adjust Low Gas Alarm Descending Value 2 (Internal Sensor) 7 Adjust High Gas Alarm Descending Value 2 (Internal Sensor) 8 Enable / Disable Current Output 2 (Internal Sensor) 9 Setting Span Gas Value 3 (External Sensor) 2 Perform Zero (Null) Calibration 3 (External Sensor) 3 PERFORM SPAN CALIBRATION 3 (External Sensor) 4 Adjust Low Gas Alarm Ascending Value 3 (External Sensor) 5 Adjust High Gas Alarm Ascending Value 3 (External Sensor) 6 Adjust Low Gas Alarm Descending Value 3 (External Sensor) 7 Adjust High Gas Alarm Descending Value 3 (External Sensor) 8 Enable / Disable Current Output 3 (External Sensor) 9 NOTE: Jumper - 4 location on the lower jumper bank (J6) is used to set custom time delays for relays and internal audible alarm. These functions will be explained in another section of this manual. 2015 All rights reserved. Data subject to change without notice. 23

GEM-MRI - Operation Manual Rev. E2 2015.03 Upper bank jumpers set (J7). This bank of jumpers allows specific functions for internal and external sensors. The target sensor is set using the lower bank of jumpers. Lower bank jumper set (J6). This bank of jumpers allows set up and configuration for both internal and external sensors. NOTE: In this photo, the jumper tab for both banks of jumpers is covering J1 resting position 7.6 Adjusting Alarm Set Points Equipment Required: Voltmeter 1 mv resolution if using Open Door calibration or setting Alarms, Philips screw driver if using Open Door calibration or setting Alarms. The GEM-MRI unit is configurable as a single channel detectors as such there are ascending gas alarm set point and a descending gas alarm set point. Almost all installations of the GEM-MRI will use the factory default alarm set points. Default set points are as follows: SENSOR LOW ALARM HIGH ALARM RANGE Oxygen 19.5% vol. 23% vol. 25% vol. Reference table in 7.5 Jumpers. If the GEM has been equipped with an LED digital display, set values according to what is displayed. If the GEM has been supplied without a digital display, you will require a digital multi-meter. Attach the meter leads to the two test points (TP-1 & TP-2) located just below the high alarm relay (RL V2) on the main circuit board (reference photo below). Set the meter to volts DC with one decimal point. Time delays are set using a voltage reference of 0-4 VDC. Reference the tables in 7.7 Time Delays for Relay Operations. Move the jumper on the lower jumper bank J6 from its resting position (P-1) and place it on the appropriate jumper position as indicated in the table below top achieve the desired alarm setting. Use the UP or DOWN push-buttons to set the desired alarm value. Once all alarm settings have been set to desired values, move the jumper back to its resting position on the lower jumper bank J6. All new values will be written to the EEPROM. 24 2015 All rights reserved. Data subject to change without notice.

Rev. E2 2015.03 GEM-MRI - Operation Manual Test Points: TP-1 & TP-2 FUNCTION DESIRED J6 LOWER JUMPER BANK Int Ext J7 UPPER JUMPER BANK SET LOW ASCENDING ALARM VALUE P2 P3 P5 SET HIGH ASCENDING ALARM VALUE P2 P3 P6 SET LOW DESCENDING ALARM VALUE P2 P3 P7 SET HIGH DESCENDING ALARM VALUE P2 P3 P8 Alarm settings for GEM detectors are voltage settings. The range of 0-4.0 VDC is equal to the full measurement range of the sensor. E.g. HVAC CO sensor has a standard measurement range of 0-200 ppm. Therefore, 4.0 VDC = 200 ppm. Reference the formula below and the table in 7.5 Jumpers. DESIRED SET POINT / SENSOR RANGE = % OF RANGE e.g. 100 ppm / 200 ppm = 50% of range 50% OF 4.0 VDC = 2.0 VDC. Therefore the required voltage setting to achieve an alarm set point of 100 ppm is 2.0 VDC. 2015 All rights reserved. Data subject to change without notice. 25

GEM-MRI - Operation Manual Rev. E2 2015.03 7.6.1 Voltage Reference Table for Alarm Settings SENSOR / GAS MEASUREMENT RANGE LOW ALARM SET VOLTAGE HIGH ALARM SET VOLTAGE O 2 0-25.0% Volume 19.5% Vol. / 3.12 VDC 23.0% Vol. / 3.68 VDC O 2 0-25.0% Volume 17.0% Vol. / 2.72 VDC 18.5% Vol. / 2.96 VDC 7.7 Time Delays for Relay Operations GEM offers a wide range of time delay settings that are user configurable. If the GEM has been equipped with the LED digital display option, read the values on the display as you change them to suit your application. If the GEM does not have a digital display, you will require a digital volt meter set to volts DC, one decimal point. The following tables are a guideline to follow to achieve desired time delays. Voltage reference is 0-4 VDC. VOLT READING AT TEST POINTS TIME DELAY IN MINUTES 0 N/A 0.02 0.1 0.04 0.2 0.06 0.3 0.08 0.4 0.10 0.5 0.12 0.6 0.14 0.7 0.16 0.8 0.18 0.9 0.20 1.0 0.22 1.1 0.24 1.2 0.26 1.3 0.28 1.4 0.30 1.5 0.32 1.6 26 2015 All rights reserved. Data subject to change without notice.

Rev. E2 2015.03 GEM-MRI - Operation Manual VOLT READING AT TEST POINTS cont TIME DELAY IN MINUTES cont 0.34 1.7 0.36 1.8 0.38 1.9 0.40 2.0 0.42 2.1 0.44 2.2 0.46 2.3 0.48 2.4 0.50 2.5 0.52 2.6 0.54 2.7 0.56 2.8 0.58 2.9 0.60 3.0 0.62 3.1 0.64 3.2 0.66 3.3 0.68 3.4 0.70 3.5 0.72 3.6 0.74 3.7 0.76 3.8 0.78 3.9 0.80 4.0 - - 1.0 5.0 1.2 6.0 1.4 7.0 1.6 8.0 2015 All rights reserved. Data subject to change without notice. 27

GEM-MRI - Operation Manual Rev. E2 2015.03 VOLT READING AT TEST POINTS cont TIME DELAY IN MINUTES cont 1.8 9.0 2.0 10.0 2.2 11.0 2.4 12.0 2.6 13.0 2.8 14.0 3.0 15.0 3.2 16.0 3.4 17.0 3.6 18.0 3.8 19.0 4.0 20.0 7.7.1 Procedure for Setting Time Delays Move jumper from resting position (position-1) on lower bank jumper set (J6) to position-4. This is the system configuration position. Next, select the jumper position on upper bank jumper set (J7) to achieve the desired function. Reference photo in 7.5 Jumpers. Reference the table below for appropriate jumper position locations. Once the desired jumper position has been achieved, use the UP or DOWN push-buttons to make changes. Changes can be viewed on the LED digital display if the GEM is equipped with one. If the GEM is not equipped with a digital display, connect a digital multi-meter to test points TP-1 and TP-2 and use the voltage tables in 7.7 Time Delays for Relay Operations. Once desired time delays have been set, move jumper back to resting position (P1) of upper jumper bank J7. If additional time delays are desired, move the jumper to the appropriate position and once again use the push-buttons to set desired voltage to achieve desired time delay. Once all desired time delay settings have been achieved, move jumper back to position-1 on upper jumper bank J7 and move jumper back to position-1 on lower jumper bank J6. At that point, all new settings will be written to the EEPROM. 28 2015 All rights reserved. Data subject to change without notice.

Rev. E2 2015.03 GEM-MRI - Operation Manual FUNCTION Low alarm ON time delay Low alarm OFF time delay High alarm ON time delay Audible alarm ON time delay UPPER JUMPER BANK J7 P2 P3 P4 P5 7.8 Temperature Display - Unit of Measure The GEM systems all have an on board temperature chip from which information is used to temperature compensate certain sensors. This temperature chip can also be used as a second sensor in applications where the user requires both a gas measurement and a temperature measurement. To switch between Celsius and Fahrenheit displayed values, follow this procedure. Move jumper from resting position (position-1) on lower bank jumper set (J6) to position-4 and jumper on J7 to position-8. Now pressing the UP push-button will display the temperature in Celsius and the output to test points TP-1 and TP-2 is 4.0 V. If the DOWN push-button is pressed, the temperature is displayed in Fahrenheit and the output to the test points is 0 V. Once the desired unit of measure has been achieved, move jumper back to position-1 on upper jumper bank J6. At that point, all new settings will be written to the EEPROM. NOTE: Temperature sensor is always considered channel-2 (external). 7.9 Latching Relay Functions The GEM relays can be configured as latching for some applications. To achieve this, move the jumper from resting position (position-1) on lower bank jumper set (J6) to position-4. This is the system configuration position. Next, select the jumper position-9 on upper bank jumper set (J7). Pressing the UP push-button activates the latching function and output to test points TP-1 and TP-2 is 4 V. Pressing the DOWN push-button de-activates the latching function and the output to test points TP-1 and TP-2 is 0 V. Once the desired latching or non-latching function has been achieved, move jumper back to 2015 All rights reserved. Data subject to change without notice. 29

GEM-MRI - Operation Manual Rev. E2 2015.03 position-1 on upper jumper bank J7 and move jumper back to position-1 on lower jumper bank J6. At that point, all new settings will be written to the EEPROM. 7.10 LED Digital Display The GEM system allows for a certain amount of customization of the LED digital display: The display can be switched completely off. The display brightness level can be adjusted to low or high for best visibility in your application. The display can indicate the gas type only with no numerical value for some applications. FUNCTION Switch digital display ON Switch digital display OFF Low brightness level for display High brightness level for display Display only gas name (acronym) Display gas values as well as gas name JUMPER SETTING Jumper J3 bridge P2 & P3 Jumper J3 bridge P1 & P2 Jumper J4 bridge P1 & P2 Jumper J4 bridge P2 & P3 Jumper J5 bridge P1 & P2 Jumper J5 bridge P2 & P3 NOTES: 1. The default setting for detectors leaving the CETCI factory with an LED digital display are: I. Display is ON II. Brightness level is Low III. Display indicates gas type and gas value. 2. The typical operation of the LED digital display is: Gas name is shown for 1-second then gas value is shown for 2-seconds. The sequence is the same for 1 or 2-channel systems. 8 CALIBRATION CAN ONLY BE DONE ON THE LPT-A TRANSMITTER. 30 2015 All rights reserved. Data subject to change without notice.

Rev. E2 2015.03 GEM-MRI - Operation Manual 8.1 Calibration Specifications 8.1.1 Gas Flow Nitrogen over Oxygen sensors before attempting to null adjust them. If the service person is confident of air quality and is careful (do not exhale in the direction of the Oxygen sensor being serviced while span adjusting), Oxygen in the breathing environment can be used as a fairly accurate source of span gas (20.9% volume) in a pinch. It is not recommended to use this procedure for all span adjustments of Oxygen sensors. Calibration span gases should have at least ± 5% accuracy and have a current date stamp. Gas generators should have a current dated cell installed. Every LPT-A transmitter is calibrated in a chamber by true diffusion method prior to leaving our facility. This method more closely emulates actual real world conditions. Field calibration using gas cylinder, regulator and hose directing span or zero gas into the sensor may result in slightly higher readings. It is important to note that the type of gas mixture, how old the gas is and what temperature it has been stored at will also affect repeatability during field calibration. NOTE: Oxygen sensors require 100% Nitrogen (N 2 ) for a zero. 8.1.2 Regulators & Flow A Demand Flow Regulator must be used to avoid drawing down the pump. Oxygen gas should be flowed at 0.5 LPM. 8.1.3 Adapters The proper calibration adapter should be utilized to allow the gas to properly diffuse around the sensor. They are available from CETCI. Part Code: CET-7000-CAP. 8.2 Calibrating Sensors 8.2.1 Calibration Frequency OHS applications: Once every 6 months (OHS: Occupational Health & Safety) 8.2.2 Gas Testing Frequency (Bump Testing) For the purpose of safety in OHS applications, sensors should be gas tested (bump tested) once every month to confirm response and alarm activation. 2015 All rights reserved. Data subject to change without notice. 31

GEM-MRI - Operation Manual Rev. E2 2015.03 NOTE: A calibration label should be applied after every calibration to confirm work performed and the date it was confirmed. If a controller is involved, the alarm set points should be indicated on a label on the front door of the enclosure so anyone working in the environment can be aware. Required Equipment: Calibration kit, Calibration gases Users can order the calibration kit, calibration accessories and / or gases from any CETCI authorized distributor or they can supply their own gas and equipment as long as the gas meets the minimum specifications. If an inappropriate concentration of span gas is applied during calibration, calibration may succeed but it doesn t mean the equipment has been calibrated properly. CETCI is not responsible for improperly calibrated transmitters. Follow manual instructions carefully. 8.3 Calibration Procedure Calibration can normally be performed without opening the LPT-A transmitter. A magnetic calibration wand (CETCI part number CET-MW) is available from CETCI and is used to touch the LPT-A enclosure at the lower left edge of the vent opening. A second touch with the calibration wand can be used to cancel the current calibration step. The LPT-A will need to be opened for calibration when any of the following conditions occur: The transmitter has not been calibrated for an extended period, which has allowed the sensor aging to reach a level that there appears to be an ambient gas level higher than the transmitter is allowed to correct for. The calibration gas concentration to be used is different from that previously set. The ambient gas level is higher than considered reasonable. Nitrogen MUST be used to zero oxygen sensors. The calibration procedure of the LPT-A is jumper automated (there are no potentiometers to adjust). Monitoring the calibration with a volt meter at TP1 and TP2 is optional. The range of 0-4.0 VDC is equal to the full measurement range of the sensor. O 2 sensor has a standard measurement range of 0-25% vol. Therefore, 4.0 VDC = 25% VOL O 2, 3.44 VDC = 20.9% VOL O 2 NOTE: If an inappropriate concentration of span gas is applied during calibration, calibration may succeed but it does not mean the equipment has been calibrated properly. CETCI is not 32 2015 All rights reserved. Data subject to change without notice.

Rev. E2 2015.03 GEM-MRI - Operation Manual responsible for improperly calibrated transmitters. Follow manual instructions carefully. To achieve calibration, the user must go through the following steps: Step 1: Set Calibration Gas Level if different from previous The calibration gas concentration can be changed in the same fashion as the set point: Move a jumper from an IDLE position to the CALGAS position Rotate the shaft of the encoder, E1, to increase or decrease the calibration gas value. The display will show the calibration gas value and a CALGAS icon. While adjusting the calibration gas value a corresponding voltage will be signaled on the test points TP-1 and TP-2. After changing the calibration gas level move the jumper from the CALGAS position back to the IDLE position. If the encoder is not rotated for 5 minutes the transmitter will automatically return to normal operation. The jumper will have to be removed and re-installed at the CALGAS position if further adjustment is required. The factory default calibration gas concentrations are: Sensor Gas Type Calibration Gas Level TP1 - TP2 voltage Oxygen (O 2 ) 20.9% 3.44 volt For more examples, visit www.critical-environment.com/calibration-gas-concentration. Step 2: Expose to clean air: Attach regulator to cylinder of calibration gas. If ambient conditions are being used, skip to Step 3. Insert the calibration adapter into the sensor opening in the front of the enclosure door. Use a slight twisting motion as you gently push the calibration adapter into the sensor opening. If the calibration adapter is hard to insert, moisten the O-ring seal slightly then try re-inserting it. Open regulator valve fully allowing calibration gas to flow over sensor for 3 minutes. In some situations using ambient air instead of calibration gas is possible. 2015 All rights reserved. Data subject to change without notice. 33

GEM-MRI - Operation Manual Rev. E2 2015.03 Step 3: Trigger the calibration by: Touching the enclosure of the LPT-A with a magnet wand (below and just left of the gas vent opening) or Moving an internal jumper from a J7 IDLE position to the CAL position. The LPT-A will use the last four seconds of gas data to determine the oxygen span gas level. Depending on the reading the LPT-A will show the following message: SPAN ACCEPTED - proceed to Step 4 Step 4: Span Accepted Once spanned, the display will show FLOW NITROGEN. Disconnect the calibration gas (if used) from the calibration adapter. Step 5: Attach hose from the bottle of Nitrogen (N 2 ) 100% vol to the calibration adapter. Ensure that the adapter has not fallen out of the transmitter s sensor opening. Step 6: Open regulator valve fully and allow Nitrogen (N 2 ) gas to flow over sensor. If no Nitrogen (N 2 ) gas is detected after one minute, the transmitter returns to normal operation and the procedure will need to be repeated from Step 2. Step 7: Wait for stable reading Once gas flow is detected, the display will show WAITING FOR STABLE READING. During this time the TP1 voltage will follow the gas level based on the ideal span of the sensor. Zeroing can be cancelled by removing the jumper from the CAL position; return it to the IDLE position. Note that the span value set in Step 4 is not cancelled. If zeroing times out without an acceptable value the display will show ZERO FAILED! RECALIBRATE. Step 8: Possible Override If the zero level appears too high (it might have been so long since the last calibration that the sensor has drifted), the display will show ZERO OUT OF RANGE, OVERRIDE? To cancel calibration, move the jumper from the CAL position back to the IDLE position. The display will show CALIBRATION CANCELLED for several seconds then return to normal operation. 34 2015 All rights reserved. Data subject to change without notice.

Rev. E2 2015.03 GEM-MRI - Operation Manual To override, move the second IDLE jumper from its rest position to the OVER / DESC / TEST position. If the jumper is not moved to the OVER position within 30 seconds, the calibration will be cancelled and the LPT-A will return to normal mode. After using the OVER / DESC / TEST position, the second IDLE jumper should be returned to its rest position. Step 9: Zeroing Complete After the zeroing is completed successfully, the transmitter will display CALIBRATION COMPLETE for several seconds, then return to normal operation. Please return the jumpers to the IDLE positions. If the calculated sensitivity of the sensor is out of range the display will show ZERO FAILED! RECALIBRATE indicating the sensor cannot be calibrated. You can try to recalibrate, starting from Step 2, to confirm the procedure was followed correctly and this may correct the fault. If calibration fails repeatedly the sensor has either aged beyond its useful life or has otherwise failed, please contact our service department at service@cetci.com. NOTE: The LPT-A will automatically return to normal operation after calibration completes (whether passed or failed). If the Nitrogen (N 2 ) gas is still applied the level will be reflected on the output signal, and relay, which could cause a control panel to enter an alarm state. To exit calibration at any time, either: Touch the enclosure with the calibration wand a second time or Remove the jumper from the CAL position and return it to the IDLE position If the digital multi-meter leads are attached to test points TP-1 and TP-2, the measured voltage will start moving towards the voltage calculated for the applied gas value. 9 MAINTENANCE The GEM-MRI series system requires virtually no maintenance other than regular calibration of the integral and / or remote sensors and ensuring that excess water or dust is not somehow entering the enclosure and physically damaging the circuit board or internal components. 2015 All rights reserved. Data subject to change without notice. 35

GEM-MRI - Operation Manual Rev. E2 2015.03 10 Trouble Shooting 10.1 Noisy Pump 1. Check to ensure GEM-MRI enclosure is securely mounted to a flat wall surface. 2. Check to ensure the flow rate has been properly adjusted after sample line and exhaust line (if used) has been installed. A long sample line reduces flow rate and will require adjustment of the flow meter. Too much or too little air flow to the pump will cause the diaphragms to vibrate more creating more noise. 3. Check to ensure pump diaphragm caps are tight. They are shipped form the factory tightly installed but vibration could potentially partially loosen them. If they are lose they could allow the diaphragms to float slightly and that will increase the vibration and create more noise level. 4. Check to ensure pump securing bolts are tight. Vibration over time could have loosened them causing excessive vibration noise from the pump. 5. Check to ensure the white inlet filter is not plugged or partially plugged. This will reduce the air flow through the flow meter and can cause the pump to make more noise. 6. If it is determined that the pump diaphragms need to be replaced they are readily available and easy to replace. Contact your local authorized distributor or CETCI. 10.2 Oxygen Sensor - Low Reading The sensor in the GEM-MRI has an estimated operating life span of approximately three years. As it ages it will require calibration as do all gas sensors throughout their life. Failure to do so could result in lower than expected Oxygen readings on the display. At least once per year perform a null and span calibration on the sensor. Refer to the supplied manuals for the procedure. If the unit cannot be out of service it is suggested to log the age of the unit and at approximately 28-30 months operation order a replacement Oxygen sensor. Store new sensor in the capsule it is supplied in inside a refrigerator until ready to install. Maximum storage should be 3-4 months. Allow sensor to warm up and reach equilibrium with it s environment for several hours after installation in the GEM-MRI before performing a calibration. 36 2015 All rights reserved. Data subject to change without notice.

Rev. E2 2015.03 GEM-MRI - Operation Manual 11 ACCESSORIES 11.1 Calibration Kit The Calibration kit includes a carrying case, regulator adapter for different size cylinders, calibration adapter, calibration adapter plugs, gas flow adapter, cylinder regulatorhumidification chamber-gas flow adapter, and pick up tool. The kit does not include calibration gas, the gas(es) need to be ordered separately. Calibration gases are available in 17-34 & 58-103 liter cylinders. Product Code: CET-715A-CK1 11.2 Metal Protective Guard Product Code: SCS-8000-SPG The metal protective guard is heavy duty metal protective guard to help protect against abrasive damage, theft and vandalism to the transmitters. This is an added preventative in addition to the product enclosure. 2015 All rights reserved. Data subject to change without notice. 37

GEM-MRI - Operation Manual Rev. E2 2015.03 It is made from 16-gauge galvanized steel and has ½ (13 mm) square openings in the front to allow gas and air to flow through to the sensor. With only four slotted mounting holes, installation and removal for gas detector servicing is easy. Enclosure 16 gauge galvanized steel Weight 1.7 kg (3.8 lbs) Size 10.0 W x 9.5 H x 4.75 D (254 mm W x 241 mm H x 121 mm D) 38 2015 All rights reserved. Data subject to change without notice.

Rev. E2 2015.03 GEM-MRI - Operation Manual NOTES 2015 All rights reserved. Data subject to change without notice. 39

Critical Environment Technologies Canada Inc. Unit 145, 7391 Vantage Way, Delta, BC, V4G 1M3, Canada Toll Free: +1.877.940.8741 Tel: +1.604.940.8741 Fax: +1.604.940.8745 www.critical-environment.com GEM-MRI20150318-E2 2015 Critical Environment Technologies Canada Inc. All rights reserved. Data in this publication may change without notice.