Reliability in Laboratory Centrifuges Since Operator s Manual. Models 614B, 614L, 614U/UL and 614V/VL Laboratory Centrifuge. Rev. 5.

Similar documents
CYGNUS CF VI BLOOD CENTRIFUGE OPERATORS MANUAL

THE COMPACT CENTRIFUGE: OPERATOR 1

CR-1730R User s Manual

Bench Series Ovens Models 21 / 31 / 51 Operating Manual

HEDMAN DI-50. Endorser Instructions. Hedman DI-50 Operators Guide

Low Speed Centrifuge. Model XC-2415 Model XC-2450 Operating Manual

USER MANUAL WATER DISPENSER WITH ICE MAKER

16 Stand Fan with Remote Control Model No: FS3-40R-30

INTELLIGENT, HEALTHFUL LIVING

BICO DULIN ROTAREX SEPARATORS

Centrifuge/Vortex for PCR plates CVP-2

Installation Instructions Wall Ovens

INSTRUCTION MANUAL. Instruction Manual High Throughput Homogenizer

POLYMIX PX-IG 2000 Operating Instructions

π H-7147 SANITAIRE EON QUIET CLEAN VACUUM SAFETY uline.com

Instruction Manual Mini Hotplate Mini Stirrer Mini Hotplate-Stirrer Mini Auto Stirrer Mini Fixed Temperature Hotplate

KEEP FOR FUTURE REFERENCE 1 SCFM NOMINAL AIRFLOW SINGLE VACUUM SYSTEM DC-VOLTAGE POWER SYSTEM WITH MANUAL VALVE

Homeowner s Installation Instructions & Operating Manual

PARTS AND SERVICE MANUAL FOR THE 23 COMPACT FIREPLACE

6L PRESSURE COOKER. Instruction Manual. Model: PLA1424

VARIABLE SPEED GRINDER OWNER S MANUAL R MANUAL VERSION SERIAL #

F Rev. C (June 2010)

I N ST R UC T I ON HCM450 CUTTER MIXER MODEL. FORM (Apr. 2002) S. RIDGE AVENUE TROY, OHIO

Installation Instructions for Models: HES232U, HES236UHES246U, HES247U, HES245U, HES242U, HES256U, HES255U, HES252U

OPERATIONS MAINTENANCE MANUAL

Tornado Operations & Maintenance Manual

16 OUTDOOR MISTING FAN

Experiment #6 Photolithography: Microprocessing Technology Fabrication of the microstructures with SPR photoresist

Instruction Manual Centrifuge LC 6 with horizontal rotor 6 x

OWNER S MANUAL FOR PROFESSIONAL COTTON CANDY MACHINE. Model Numbers (Aluminum Bowl) (Plastic Bowl) Cotton Candy Machines

User s Manual and Operating Instructions

BLAST-IT-ALL BUMPER BLASTER

ELITE (110V) (220V)

Instruction Manual. Mini Hotplate Mini Stirrer Mini Hotplate Stirrer

Suits all KPF849 models

Package Contents. Dry Block Heater Power Cord Instruction manual Warranty Card. Serial Number: Date of Purchase: Supplier: Table of Contents

MODEL 7000 SUCTION UNIT

Installation Instructions

OWNERS GUIDE TO INSTALLATION AND OPERATION

Hickory Rotisseries. Machine Type N / 5. 5 E 200ºF - 550ºF. Final Inspection:

Water Distiller Service Manual

OWNER S MANUAL FOR AIR ADMIRAL VACUUM/PRESSURE STATION MODELS:

H-1451, H F, H-1452, H F, H-1474 & H F Instruction Manual

BRAWN. BRAINS. BLENDTEC. o p e r a t i o n s m a n u a l. for the: Titan ABC Smoother and SpaceSaver

NOTICE . SAFE SERVICING PRACTICES. Electric Wall Oven with Electronic Oven Control

18 INCHES, OSCILLATING STAND FAN

CONSTANT SUCTION UNIT

EBAC MODEL CD30 INDUSTRIAL DEHUMIDIFIER OWNER S MANUAL

OWNER S MANUAL. General Ventilation Fans Electric Driven. PAL Series (Incl. 24 & 30) HD Series (Incl. -T & -S) 0316_01A

WASHING MACHINE SERVICE MANUAL CAUTION READ THIS MANUAL CAREFULLY TO DIAGNOSE PROBLEMS CORRECTLY BEFORE SERVICING THE UNIT.

Hodge Clemco Ltd. MJC Mini Cartridge Filters Owners Manual. TSOM 556 Date of issue 13/07/04. Hodge Clemco Ltd

MODEL MC-UL592 MC-UL594 POWER SOURCE. 230V - 240V ~ 50 Hz MAX INPUT 1400 W 1500 W NOMINAL INPUT DIMENSIONS (W x L x H)

Installation Instructions Remote Blowers

FAILURE TO ADHERE TO THESE WARNINGS CAN RESULT IN FIRE, SERIOUS INJURY, ELECTRICAL SHOCK AND/OR DEATH. Technical Service

Installation Instructions

Disposal Unit (BKSDU) Burger King Shortening. Operation, Service & Parts Manual BURGER KING BKSDU

OWNER S MANUAL MULTIPURPOSE UTILITY AIR MOVER PLEASE READ AND SAVE THESE INSTRUCTIONS X-400WFK (WITH FILTER KIT) ITEM NO:

PUGMILL/MIXER. Instruction Manual NVS-07

ELECTRIC FIREPLACE HEATER WITH SINGLE GLASS DOOR

ELECTRIC FIREPLACE HEATER OPERATION MANUAL

SERVICE MANUAL WASHING MACHINE

Heated Bed Installation Instructions

CHLORIDE GASES CHLORIDES

π H-2268 SANITAIRE UPRIGHT VACUUM SAFETY uline.com

GENIUS 14 GENIUS 18 OPERATING & MAINTENANCE

PastaMazing POWERED PASTA MACHINE MODEL# GMP500 USER MANUAL

Menumaster. Commercial Microwave Oven Use and Care Manual CFSP70 GSAND-7D SAND-7D SAND-700

INSTALLATION AND OPERATING INSTRUCTIONS FOR RW519 RAIN WATER FILTRATION SYSTEM

Installation Instructions

Los Angeles Abrasion Machine HM-70A & HM-70AF

AC6 & AC8 HORIZONTAL SERIES Sealed Metallic Centrifugal Pumps Installation and Maintenance Instructions

BF5000NT BF5000 EF2604 SF2603 DF2603

WAILEA OWNER S MANUAL

One-Touch Dispense. Multi-temp selections. 208 F For tea, coffee, instant noodle. 194 F Keep warm around 194 F. 176 F Keep warm around 176 F.

Installation and Maintenance Manual Please retain this manual for future reference.

Homeowner s Installation Instructions & Operating Manual

INSTRUCTION MANUAL POWER DRIVE UNIT MODEL PAD-350 ONE YEAR LIMITED WARRANTY

Owners Manual MINI-CLASSIC II. Pure Water. Water Distiller

Installation Manual PS-225 & PS-275

READ AND SAVE THESE INSTRUCTIONS

Tornado Operations & Maintenance Manual

ELECTRIC FIREPLACE HEATER ASSEMBLY AND OPERATION MANUAL IMPORTANT SAFETY INSTRUCTIONS

Tank, Heater, and Control Operation Manual REV. E DATE: 04May2015

Service Manual Model 3163

General Information on Finishing with Your Raytech Centrifugal Magnetic Finisher

RINGMASTER 5-1/2 & 7 COTTON CANDY MACHINE

ELITE Series R-410A. SEER up to to 5 Tons Cooling Capacity - 17,800 to 60,000 Btuh AIR CONDITIONERS EL16XC1 PRODUCT SPECIFICATIONS

SPACESAVER INSTRUCTION MANUAL WARRANTY CERTIFICATE

Dual Zone Wine Cellar

Mopit 4 Operator s / Service Manual

Desoldering Tool. Instruction Manual

CentriVap. Centrifugal Vacuum Concentrators and Cold Traps

SINGLE SPEED SYRINGE PUMP

Centrifuge STANDARD OPERATING PROCEDURE

Lifetime Limited Warranty

ASPIRE Laboratory Aspirator

EcoSmart Troubleshooting Guide

ELECTRIC STOVE MODEL CGESS SAFETY INFORMATION AND OPERATION MANUAL

IMPORTANT SAFETY INSTRUCTIONS

PUREPOWER SERIES CENTRAL VACUUM POWER UNITS PP500, PP600 & PP650

Transcription:

Reliability in Laboratory Centrifuges Since 1932 Rev. 5.0 Operator s Manual Models 614B, 614L, 614U/UL and 614V/VL Laboratory Centrifuge

Description: Intended Use: The Drucker model 614 is a low-cost, high-performance laboratory centrifuge designed for the separation of laboratory fluid samples. Versions of the Model 614 include: 614B: 3,150 RPM* (1,200 RCF) with a lid safety switch 614L: 3,150 RPM* (1,200 RCF) with a lid safety interlock system 614U: 1,700 RPM* (350 RCF) with a lid safety switch 614UL: 1,700 RPM* (350 RCF) with a lid safety interlock system 614V: 0-3,150 RPM* (0 to 1,200 RCF) with a lid safety switch 614VL: 0-3,150 RPM* (0 to 1,200 RCF) with a lid safety interlock system Supplied Equipment: * Nominal speed with a line voltage of 115 Volts at 60 Hz, +/- 100 RPM Given specifications are for the black test tube holders designed for 15ml, (16mm x 125mm), test tubes. The use of different holders and/or cushions will slightly alter the top speed and RCF. The following items are supplied with each model 614 centrifuge: 1. One (1) six-place, fixed angle, polycarbonate rotor. 2. One (1) operator s manual 3. One or more of the following based on your accessory package: Six (6) 125mm test tube holders, Black Six (6) 100mm test tube holders, Red Six (6) 75mm test tube holders, Green Six (6) 1525 test tube cushions Six (6) 9150 test tube cushions Features: Important features of the Model 614 centrifuge include the following: Conforms to UL STD 3101-1 and Certified to CAN/CSA STD C22.2 NO. 1010.1, (Models 614B, 614L, 614V, 614VL Only), ETL LISTED. Lid safety switch that cuts power to the motor if the lid is opened during operation The model 614L, 614VL and 614UL have a 0 RPM lid safety interlock system that prevents the lid from being opened until the rotating head reaches approximately 0 RPM. Quiet Operation Brushless AC motor Two part molded inner housing of flame retardant polymer for superior strength and sound absorption Cool-Flow air flow design that prevents overheating of samples Heavy gauge welded steel cabinet construction for safety and durability Clear lid for safe observation of samples and optical calibration of speed. Optional 8-place rotor Timed operation from 1 to 30 minutes Variable speed (Model 614V/VL only) Pg. 2

Installation and Operation: Performance / Calibration:* It is recommended that your Model 614 centrifuge be calibrated every 6 months. The max speed of a 614B, L, V or VL should fall between 3050 and 3250 RPM* when unloaded; this will give a separation force of 1120-1280 xg*. The max speed of a 614U or UL should fall between 1600 and 1800 RPM* when unloaded; producing a separation force of 310-390 xg*. A photo tachometer is suggested to determine RPM. If your centrifuge fails to operate within its prescribed range, contact your authorized dealer or The Drucker Company. * Given specifications are for the black test tube holders designed for 15ml, (16mm x 125mm), test tubes. The use of different holders and/or cushions will slightly alter the top speed and RCF. Environmental Conditions: The 614 centrifuge is not intended for outside use or for use in extreme environmental conditions. Any use other than the manufacturer s suggested usage may impair the protection provided by the unit. External Packaging and Inspection: Carefully examine the centrifuge and document any damage that can be attributed to mishandling. A signed inspection report should be furnished by the shipping company. Note: The Drucker Company is Not Responsible for transit damage. Setup Procedure: Unpack the centrifuge and inspect for obvious damage; place the centrifuge on a hard, stable surface. Note: A bench top clearance height of 23 inches (min.) is required to open the centrifuge lid. Note: Failure to provide adequate space for ventilation can cause damage to the samples plus overheating and premature failure of the centrifuge. 1. Unlatch and open the lid; remove any protective shipping material, literature, tube cushion packages, etc., that may have been shipped inside the centrifuge. 2. Spin the rotor by hand; check for free and level rotation. If the rotor spins evenly, continue on to step 3. If the rotor does not turn freely or there seems to be an obstruction, call your authorized dealer or The Drucker Company before proceeding. 3. Close and latch the lid. Verify that the timer is "OFF". Plug the line cord into an approved electrical outlet. Be sure the outlet is always within reach as the line cord is the means of emergency disconnection. 4. Turn power on to the unit by setting the time to (5) minutes. NOTE: Model 614V/VL : Turn the speed to 3,050 RPM. The rotor should start spinning smoothly with no excessive noise. 5. Listen to the sound of the centrifuge; a smooth whirring sound should be heard. If there are any loud and unusual sounds, stop the centrifuge immediately. Do not proceed! Call your Authorized Dealer or The Drucker Company. Pg. 3

Safety: 1. Lid Safety Switch (all 614 models): During operation, the lid is secured to the top of the cabinet by a latching knob and pawl system. When the knob is rotated clockwise, the pawl grips the underside of the cabinet and prevents the lid from opening. A mechanical stop limits the pawl travel. When rotated to this stop position, the pawl activates a safety micro-switch which allows the centrifuge to operate. Unless contact is made by the pawl with this safety switch, the centrifuge will not start or continue in operation. If an attempt is made to open the lid while the centrifuge is in operation, power will be cut to the motor. NOTE: Latch the lid knob clockwise to its stop position or the centrifuge will not operate! 2. Lid Safety Interlock System (Model 614L, 614UL and 614VL): In addition to the Lid Safety Switch, the models 614L, 614UL and 614VL have a 0 RPM lid interlocking system. As the motor starts spinning, a locking solenoid activates and prevents the lid from being opened. A red diode on the top-front of the cabinet illuminates indicating that the lid is Locked. The solenoid will keep the lid locked during operation and until the motor has reached approximately 0 RPM. Note: After the centrifuge has started spinning, it may be possible to rotate the lid knob enough to cause the pawl to lose contact with the lid safety switch. If this happens, the centrifuge motor will lose power, but the lid will still remained locked until the motor has reached 0 RPM. If the knob is accidentally moved and this situation should occur, rotate the knob fully clockwise to its stop position and the centrifuge will resume operation. 3. Model 614L/UL/VL Emergency Lid Removal: If, for any reason, the safety lid interlock system should fail and the lid cannot be opened after the motor has stopped spinning, the centrifuge samples may be removed by performing the following: a. Remove the four (4) screws that attach the lid to the hinges. b. Lift the lid clear of the hinges and pull back so that the pawl clears the underside of the cabinet. c. Remove the lid. d. Remove the samples. e. Call the Drucker Company for repair. 4. Tube Holders: It is recommended that the tube holders be replaced after 24 months of use. Pg. 4

Safety (cont d): 5. Model 614L/UL/VL Lid Lock Safety Check: a. Verify that the Timer is in the "OFF" position. Open the lid and spin the rotor by hand. b. The red LED on the top of the cabinet should illuminate and a low click sound should be heard. When the rotor stops spinning the red LED should go out and the click sound should repeat. c. With the centrifuge turned on and the rotor spinning, try to open the lid by rotating the lid knob counter-clockwise. The knob will rotate slightly and you should hear the motor lose power but you should not be able to completely rotate the lid knob or open the lid. d. Rotate the knob clock-wise back to its stop position. You should hear the motor resuming operation. e. If the centrifuge successfully passes steps a-d of this section it is ready for operation. If there are any problems Do not proceed! Call your Authorized Dealer or The Drucker Company. 6. Keep the Test Tube Holders Clean: Small glass fragments left in the tube holder after a tube breakage may puncture protective gloves and lacerate the operator's fingers or hand. Fragments left may also provide stress points on subsequent tubes and result in additional breakage. If a tube breakage occurs, carefully remove the tube holder; properly dispose of the sample and tube fragments; thoroughly clean both the inside and outside of the tube holder; insert a new tube cushion (if necessary) and replace the tube holder in the rotor. Temperature Control: The model 614 centrifuge has a unique molded-in air channel that circulates ambient room air through the rotation chamber to cool the samples. This air is then exhausted out through the base to cool the motor. Care and Preventative Maintenance: With proper care and maintenance, the Drucker 614 centrifuge will provide years of laboratory service. For proper care, the following steps should be taken: 1. Provide Adequate Ventilation: For cooling purposes, the model 614 draws air in through a space in the bottom front of the cabinet and exhausts this air through holes in the base. The centrifuge should be placed on a hard smooth surface for good air circulation. 2. Always Spin Only Balanced Loads: Make certain that opposing tube holders are filled with equal weight samples or equivalent weight water filled tubes. When using tube holders of different types or when using cushions, make sure that opposing configurations are identical. The 614 has a unique counter balanced motor mounting design which, along with its all steel construction and rubber suction feet, produces excellent vibration dampening. However, out-of-balance loads may break glass test tubes and may produce unsatisfactory separation results. Proper sample balancing will improve sample separation and extend the life of the centrifuge. Pg. 5

Care and Preventative Maintenance (cont d): 3. Clean The Inside Of The Centrifuge: The 614 centrifuge is designed with a removable rotor chamber cover to permit servicing and cleaning of the inside of the rotor chamber. Every six months, or whenever there is a tube breakage that allows samples to enter the rotor chamber area, it is advisable to remove this cover and clean the inside of the centrifuge. To clean the centrifuge proceed as follows: a. Unplug the centrifuge. b. Close and latch the lid. c. Remove the eight screws attaching the cabinet to the base and lay the cabinet on its side adjacent to the centrifuge base (take precaution not to over extend the electrical wires). d. With your hand, or with the use of a blunt object, remove the snap-on plastic ring at the top of the rotation chamber. e. Remove the rotor by unscrewing the locking nut on the motor shaft. f. Disinfect and clean the inside of the centrifuge. To reassemble, first replace the rotor taking care to line the motor pin up with the slot on the underside the rotor. Replace the locking nut on top of the rotor and tighten until the rotor is secure. Snap the plastic ring back on the rotation chamber and place the six test tube holders back in the rotor. Finally, place the cabinet over the base and reattach it with the eight screws. CAUTION: For cleaning, do not fully submerge the centrifuge in water or use an excessive amount of cleaning solution as this may cause permanent damage to the electrical components. 4. Motor and Electrical Maintenance: The motor of the 614 is a brushless induction type; It should not need servicing for the life of the centrifuge. The electrical components are selected for high reliability and should not need service. If any of these parts should fail they must be repaired or replaced by a qualified service technician. Pg. 6

SPECIFICATIONS: General specifications for the Model 614 Centrifuge Nominal Speed 614B/614L* 3,150 RPM Nominal Force (RCF) 614B/614L* 1,200 x g Speed Range 614V/VL* 0 to 3,150 RPM Force Range 614V/VL* 0 to 1,200 x g Nominal Speed 614U/614UL* 1,700 RPM Nominal Force (RCF) 614U/614UL* 350 x g Maximum Capacity Six-place 90 ml. (6 x 15 ml) Maximum Capacity Eight-place 120 ml. (8 x 15 ml) Overall Height with Lid Closed 8.5 in. (21.59 cm.) Dimensions Height with Lid Open 15.75 in. (40.0 cm.) Width 10.5 in. (26.67 cm.) Depth 12.25 in. (31.12 cm.) Centrifuge Motor 1/30 HP Shaded Pole, AC Protection Breaker 4 Amp. Re-settable Timer Mechanical; 1 to 30 minutes Max. Power Requirements 120 Watts Voltage 120 Volts Frequency 50/60 Hz Weight (Including Rotor and (6) tube holders) 15 lbs. (6.8 kg) * Given specifications are for the black test tube holders designed for 15ml, (16mm x 125mm), test tubes. The use of different holders and/or cushions will slightly alter the top speed and RCF. REPLACEMENT PARTS: Part No. Description 7724079 Foot, rubber 7760002 Power cord 7751068 Switch, lid safety 7722029 Timer, mechanical, 30 minutes 7724145 Knob, timer 7714102 Pawl, latch, lid 7714103 Knob, latch, lid 7712313 Lid, clear 7724071 Hinge, friction 7732206 Seal, lid gasket 7786047 Rotor, 1936 six-place polymer 7786041 Rotor, 1938 eight-place aluminum 7735050 Motor, model 614B, 614L and 614V/VL 7735021 Motor, model 614U and 614UL 7717064 AC motor control, model 614V/VL 7745016 Locking solenoid ( L versions) 7717071 PC board, lid lock ( L versions) 7713024 Tube holder, green, for 75mm tubes 7713019 Tube holder, red, for 100mm tubes 7713026 Tube holder, black, for 125mm tubes 1525 Test Tube Cushions for 10mL test tubes Pg. 7

200 SHADY LANE PHILIPSBURG, PA 16866 TEL: (814) 342-6205 TOLL FREE: 1-888-299-7778 FAX: (814) 342-6211 ON THE WEB at www.druckercompany.com