User's Guide. CO 2 Monitor. Model CO200

Similar documents
CDL 210. Operating manual. CO2 Logger. Best.-Nr

R9900. Model. Instruction Manual. Indoor Air Quality Meter. reedinstruments. www. com

USER MANUAL Digital CO Monitor Model CO30

Model R9400. Instruction Manual. CO Meter with Temperature. wwwreedinstruments. com

CO2 & Oxygen Monitor Operating Instructions Model: RAD

Carbon Monoxide Meter Model CO15. User Manual. Additional User Manual Translations available at

User Manual R1.0 CO2Meter, Inc. Personal CO2 Monitor SAN-10

USER GUIDE. Formaldehyde (HCHO or CH 2 0) and Total Volatile Organic Compounds (TVOC) Tester. Model VFM200

USER GUIDE Heat Index Psychrometer

Oxygen Monitor Operating Manual

Pinless Moisture/Humidity Meter with Memory + IR Thermometer

VALTRONICS IAQ Monitor

CO2 Monitor Operating Instructions

User's Guide. Pinless Moisture/Humidity Meter with Memory + IR Thermometer. Model MO295

User Manual. Humidity-Temperature Chart Recorder. Model RH520

Indoor/Outdoor Temperature Alert

KANE-ALERT-CO 2 Monitor with Temperature

User's Guide. Pinless Moisture/Humidity Meter + IR. Model MO290

User's Guide. Pinless Moisture Meter. Model MO257

Gas Safety Products. Installation, operating and maintenance. Merlin CO2 Monitor 4500PPM. Merlin CO2 Monitor. Rev: 11 Date:

Wet Bulb Thermo-Hygrometer Datalogger

Pinless Moisture/Humidity Meter with Memory + IR Thermometer

USER GUIDE. Paperless Humidity/Temperature Chart Recorder. Model RH520A 21.2 C 13:

USER MANUAL USB Multi-Function Datalogger Model RHT35

USB Multi Function Dataloggers. RHT30 Humidity/Temperature Datalogger. TH30 Dual Temperature Datalogger

Model VFM200. User Manual. Formaldehyde (HCHO or CH 2 0) and Total Volatile Organic Compounds (TVOC) Tester

4-in-1 Humidity, Temperature Airflow and Light Meter

! WARNING! User Manual LogiCO2 -Scout. Important Safety Information! CO2 Concentration Levels (%) and Effects ENGLISH

Falcon-II Next Generation, Air Quality Monitor CO2 & Temperature

CO2 transmitters TRANSMITTERS TO DETECT CARBON DIOXIDE (CO 2 )

Indoor Air Quality Meter Rev. 2.x Meters. The Value Leader TM

C0-210 Carbon Monoxide Probe

Wöhler CM 220. Carbon Monoxide Meter. The Measure of Technology. Contents. 1. Important information Specification... 13

USER MANUAL ENVIRONMENTAL METER. Model EN510. Additional User Manual Translations available at

Warranty Registration

OWNER S MANUAL DLFCAB / DLFCHB / DLFDAB / DLFDHB High Wall Ductless System Sizes 09 36

High Resolution Display WIRELESS COLOR WEATHER STATION

Manual Vane Anemometer PCE-VA 20

TDU Series LCD Thermostat For EFCB and EVCB Controllers. Models. Features. Functions. Onboard Sensors. Model # Temp RH CO2 Color.

VIGÖR HRV PLUS AND VIGÖR ERV

HOBO U14 Data Logger User Manual

3 Macurco Carbon Monoxide Detector CM-16A

Dual Laser InfraRed (IR) Thermometer

Dual Laser InfraRed (IR) Thermometer

Pocket CO2 Monitor & Data Logger SAN-0001

Installation, Start-Up, and Operating Instructions

Multifunction IAQ Monitor

User's Guide. Pinless Moisture Meter. Model MO257

DDAC-1 deluxe digital atmosphere controller

Installation, Start-Up, and Operating Instructions

QUICK START. FLIR Model MR59. Ball Probe Moisture Meter with Bluetooth

Operations Manual TS400. Test Station for G450/G460 Gas Detector

The product can be worked together with our wireless controller to control a ventilator, air cleaning device, or

Beacon 200 Gas Monitor Operator s Manual. Part Number: RK Released: 6/6/08

Macurco Single-Gas XL Series Monitor, CM-1XL Carbon Monoxide (CO), HS-1XL Hydrogen Sulfide (H2S) User Instructions

Ventostat Wall Mount. Amphenol Advanced Sensors. Telaire Wall Mount CO 2. , Humidity and Temperature Transmitters. Features:

Infrared Thermometer with 12:1 Distance-to-Sight Ratio and Temperature Alarm

User's Guide. ExStik DO600. Dissolved Oxygen Meter RECALL. Dissolved Oxygen DO600

Wide Range InfraRed (IR) Thermometer with Type K input and Laser Pointer

Mini InfraRed (IR) Thermometer with Type K Input and Laser Pointer

Model:AS8900 MULTI-GAS MONITOR O 2 %VOL H 2S PPM LEL%VOL CO PPM. Version number:as

Ventostat Wall Mount. Amphenol. Telaire Wall Mount CO 2. , Humidity and Temperature Transmitters. Advanced Sensors. Features:

FLOW CONTROLLER TYPE S/601

Installation Instructions

TCA-9102 Series Surface Mount Temperature Controllers with High and Low Alarm

Users Manual. Air Velocity Meter

TS400. Operating Manual. Test Station for Microtector II Series (G450/G460)

Owner s Manual. Part Number 33CSCPACHP-01

MULTI-GAS MONITOR. Eastern Energy Co., Ltd. 40/4 Vitoondumri Rd., Banbueng Banbueng, Chonburi, Thailand

Digital Refrigerant System Analyzer. MAXMIN Psi kpa Bar MPa C F R TIME P T UNITS

HEAT RECOVERY VENTILATORS MODELS HRV 90H-V ECM, 90H-V

USER GUIDE HEAT RECOVERY VENTILATORS MODELS HRV 90H-V ECM, 90H-V AND 60H NOVO+, MODEL ERV 90H-V ECM PLEASE READ AND SAVE THESE INSTRUCTIONS

Ventostat Wall Mount. Telaire Wall Mount CO 2. , Humidity and Temperature Transmitters. GE Sensing & Inspection Technologies.

Ventostat Wall Mount. Telaire Wall Mount CO 2. , Humidity and Temperature Transmitters. GE Sensing & Inspection Technologies.

ETNC24-FC-BAC-PIR-01 Owner s manual & Technician Settings

esense CO2 Transmitter and Alarm

E1000 Portable gas detector

Installation, Service, and Troubleshooting Instructions

Operate the Mutli-Gas Monitor PID


IMR IX176 Portable Gas Detector User Manual

Operations Manual TS400. Test Station for G450/G460 Gas Detector

ALTAIR 5X Multigas Detector Product Specification


IMR EX610. Portable Gas Detector. Operation Manual

Warmette Blanket Cabinet Warmette Dual Cabinet. User Manual

Product Manual SZ1009

Smart Temp. Model

ExStik DO600 USER GUIDE. Dissolved Oxygen Meter. 99 Washington Street Melrose, MA Phone Toll Free

CA5000 Liquid CO 2 Freezer Backup System

CDR-AL Room CO 2 Traffic Light Alarm Units and Sensors

THA2 TEMPERATURE/ HUMIDITY/ DEWPOINT ALARM WITH LOGGING CAPABILITY. 99 Washington Street Melrose, MA Phone Toll Free

Portable Oxygen Monitor OX-07. (Type B) <No Gas alarm type> Operating Manual

Sirius Technologies, Inc.

INSTRUCTION MANUAL TRC-1490 Wireless Indoor and Outdoor Weather Station

Wired Controller XK60

O3TH. Ozone Detector c/w Indicator for Temperature/Humidity. Automatikprodukter. Features. Technical Data. Description. Ordering Codes. Dec.

Weather Station with Remote Control Model: BAR289

Introduction. OM-THA2

Tempco Part Number PCT30006 Temperature Control Enclosure with Relay Output for Tote Tank Heating Applications

Transcription:

User's Guide CO 2 Monitor Model CO200

Introduction Congratulations on your purchase of this Model CO200 Carbon Dioxide Meter. This meter measures CO2 level, air temp., humidity, date and time. With visible and audible alarms, this is an ideal instrument for indoor air quality (IAQ) diagnosis. This meter is shipped fully tested and calibrated and, with proper use, will provide years of reliable service. Meter Description Meter 1. CO2, Temp and RH sensors (rear) 2. LCD display 3. Controls LCD Display 1. CO2 concentration in ppm 2. Date and Time 3. Air Temperature 4. % Relative Humidity Symbols ppm CO2 value GOOD CO2 air quality level NORMAL CO2 l air quality level POOR CO2 air quality level Air Temp Air Temperature Humidity % Relative Humidity C or F Celsius or Fahrenheit MAX/MIN Maximum or Minimum reading Controls SET Relay activated Enters setup mode. Saves and finishes settings. ESC Exits setup page/mode. Terminates during CO2 calibration. RESET Press to clear the MAX/MIN. - Terminates during RH calibration. Selects mode or increases value in setup MIN/MAX Activates MAX, MIN function. Enters CO2 calibration with SET and 2 CO200-EU-EN-V1.1-11/11

Operation POWER ON Plug in the adaptor and the meter turns on automatically with a short beep. If the voltage is too high or low, bat will display on the LCD and an LED will flash. The LCD will display current CO2, temp., humidity, date and time. The air quality level is displayed as well TAKING MEASUREMENT The meter starts measurements when powered on and updates readings every second. If the operating environment changes (ex. from high to low temp.), it takes 30 sec for CO2 sensor to respond and 30 minutes for RH. NOTE: Do not hold the meter close to your mouth or any other source of CO2. MAX/MIN In the normal mode, press the MIN/MAX button to see the minimum and maximum of CO2, Temperature and Humidity. Each press of MIN/MAX button will sequence through the MIN, MAX display and then returns to normal mode. Press and hold the RESET button for more than 1 second to clear the minimum and maximum value from memory. ALARM & OUTPUT The meter features an audible alarm which will give warnings when the CO2 concentration exceeds the set limit. There are two settable limits, an upper limit that initiates the alarm and a lower limit to stop the alarm. The meter emits beeps and displays the fan icon on the display when the CO2 level goes over the upper limit. The audible alarm can be stopped by pressing any key or it automatically stops when CO2 reading falls below the lower limit. If the beeper is temporarily shut down, it will sound again when readings fall below lower limit and then go over the upper limit again or it the user presses the RESET button for more than 1 second to activate it. The fan icon keeps flashing when beeps are manually shut down. It stops only when readings fall under the lower limit 3 CO200-EU-EN-V1.1-11/11

RELAY OUTPUT The meter is designed with a relay for connection to an external indicating or controlling device. When CO2 readings go over the upper limit the relay closes. The relay will open again when CO2 readings fall below the lower limit. The relay output port requires a 2.5mm STEREO phone jack plug. The relay is: 1A 30VDC/0.5A 125VAC. 1. AC adaptor input (5VDC) 2. Alarm Relay output 3. Phone plug wiring 3 SETUP (Alarm level, Temperature scale, Real Time clock) Hold the SET button, in the normal mode, for more than 1 second to enter the setup mode. P1.1 CO2 ALARM: GOOD LEVEL When entering the setup mode, P1.0 and CO2 are displayed on the LCD. Press the SET button again to go into P1.1 for setting the CO2 upper limit for the GOOD level. The current set value will be blinking on the display. Press the or MIN/MAX button to increase or decrease the value. Each press adjusts in 100ppm increments. The alarm range is 0 to 700ppm. When the value has been set, press the SET button to confirm the GOOD limit and to proceed to P1.2 for setting the upper NORMAL limit. Press the ESC button to exit without saving the setting. P1.2 CO2 ALARM: NORMAL LEVEL P1.2 is used for setting the CO2 upper limit for the NORMAL level. The current set value will be blinking on the display. Press the or MIN/MAX button to increase or decrease the value. Each press adjusts in 100ppm increments. The alarm range is 700 to 1000ppm. When the value has been set, press the SET button to confirm the NORMAL limit and to proceed to P1.3 for setting the upper POOR limit. Press the ESC button to exit without saving the setting. 4 CO200-EU-EN-V1.1-11/11

P1.3 CO2 ALARM: BEEP ALARM P1.3 is used for setting the CO2 upper limit for the BEEPER ALARM level. The current set value will be blinking on the display. Press the or MIN/MAX button to increase or decrease the value. Each press adjusts in 100ppm increments. The alarm range is 1000 to 5000ppm. When the value has been set, press the SET button to confirm the limit and to proceed to P1.0. Press the ESC button to exit without saving the setting. P2.0 TEMPERATURE SCALE Press the button in P1.0 mode to access P2.0 for setting the temperature scale. Press the SET button to go into the P2.1setting mode. The C or F will blink. Press the button to change the units. Press the SET button to confirm theton.etting or press the ESC button to exit without saving and return to P2.0. P3.0 REAL TIME CLOCK Press the button in P1.0 twice to access P3.0 for setting up real time clock. Press the SET button and the meter goes into P3.1 with blinking year in the lower left display. To change the year, press the button or the MIN/MAX button. Press the SET button to save the setting and then enter P3.2 or press the ESC button return to P3.0 without saving the setting. Press the button P3.1 to access P3.2. The current month setting will blink. To change the month, press the button or the MIN/MAX button. Press the SET button to save the setting and then enter P3.3 or press the ESC button return to P3.0 without saving the setting. Repeat as above to finish the setting of P3.3 (Date), P3.4 (Hour) and P3.5 (Minute) 5 CO200-EU-EN-V1.1-11/11

P4.0 RESET Press the button in P1.0 three times to access P4.0 to reset the meter to default settings. Press the SET and the meter will go to P4.1 with a blinking No. Press the to switch the status then press either the button to save the settings or the ESC button to exit without saving the settings. Default settings: Parameter Default P1.1 700ppm P1.2 1000ppm P1.3 1000ppm P2.1 C P4.1 No 6 CO200-EU-EN-V1.1-11/11

Calibration CO2 CALIBRATION The meter is calibrated to a standard 400ppm CO2 concentration at the factory NOTE: When the accuracy becomes a concern or after a year of use, return the meter to Extech for a standard calibration. CAUTION: Do not calibrate the meter in an atmosphere of unknown CO2 concentration. ABC (Automatic Baseline Calibration) ABC (Automatic Baseline Calibration) establishes a baseline calibration to eliminate the zero drift of the infrared sensor. The ABC function is always ON when the meter is turned on. ABC is designed to calibrate the meter at the minimum CO2 reading detected during 7.5 days continuous monitoring (power on). It assumes that the area being tested receives fresh air with a CO2 level of approximately 400ppm at some period of time during the seven days. It is not suitable to use desktop CO2 in closed areas with consistently high CO2 levels 24 hours a day. Manual Calibration Manual calibration is suggested to be done outdoors on a sunny day with good ventilation and fresh air where the CO2 level is approximately 400 ppm. Do not calibrate on a rainy day because high humidity will affect the CO2 level in air. Do not calibrate in places crowded with people or close to where high CO2 concentrations may exist such as ventilating outlets or fireplaces. Place the meter in the calibration site. Turn on the meter and hold down the SET, and MIN/MAX buttons simultaneously more than 1 second to enter CO2 calibration mode. 400ppm and CO2 will blink while the calibration is in process. The calibration will take about 30 minutes. When the calibration is complete, the blinking stops and the meter returns to normal operation. To abort the calibration, press the RESET button for more than 1 second. Note: Keep away from any animal, human or plant which might affect the CO2 concentration during the calibration. 7 CO200-EU-EN-V1.1-11/11

Specifications Function Range Resolution Accuracy CO2 0 to 9999ppm 1ppm ±(5%rdg +50ppm) Temperature -10 to 60 C 14 to 140 F Humidity 0.1 to 99.9% 0.1% 0.1 ±0.6 C/0.9 F ±3%(10 to 90%) ±5%(< 10% or > 90%) Display Sensor Type Response Relay Operating Conditions Storage Conditions Power Supply Dimensions / Weight LCD CO2: NDIR (non-dispersive infrared) technology Humidity: Capacitance sensor; Temperature (air): Thermistor CO2: <2min for 90% step change Temp: <2min for 90% step change %RH: <10min for 90% step change 1A 30VDC/0.5A 125VAC -10 to 60 C (14 to 140 F); < 90% RH non-condensing -20 to 60 C (-4 to 140 F); <99% RH non-condensing 5VDC (±10%), 500mA 117x102x102mm (4.6x4x4 ); 204g (7.2 oz.) Maintenance CLEANING AND STORAGE 1. The meter should be cleaned with a damp cloth and mild detergent when necessary. Do not use solvents or abrasives. 2. Store the meter in an area with moderate temperature and humidity. 8 CO200-EU-EN-V1.1-11/11

TROUBLESHOOTING Can t power on: Slow response: "BAt and green LED keep flashing: Check whether the adaptor is properly plugged in. Check whether the air flow channels on the rear of the meter are blocked. The adaptor output voltage is too high or too low. Please use the adaptor with correct output. Error Codes: CO2 Display E01 CO2 sensor is damaged Send back for repair E02 CO2 reading is under the lower limit Re-calibrate the meter, if it still appears, return for repair E03 CO2 reading is above the upper limit Put the meter in fresh air and wait for 5minutes, if it still appear, re-calibrate the meter. If above two methods faile, E17 Temp Display ABC mode of CO2 senor has failed and might cause wrong CO2 readings return for repair Send back for repair E02 E03 E31 Humidity Display Air temp. measurement is under the lower limit Air temp. measurement is over the upper limit Temp. sensor or measuring circuit is damaged Put the meter in regular room temperature for 30 minutes, if it still appears, return for repair Put the meter in regular room temperature for 30 minutes, if it still appears, return for repair Return for repair E04 E11 E34 Air temp. measurement has error code The RH calibration has failed RH sensor or measuring circuit has failed Refer to temperature error code for problem solving Please return for repair Return for repair 9 CO200-EU-EN-V1.1-11/11

CO2 Levels and Guidelines Non-Enforced Reference levels: 250-350 ppm background (normal) outdoor air level 350-1,000 ppm - typical level found in occupied spaces with good air exchange. 1,000 2,000 ppm - level associated with complaints of drowsiness and poor air. 2,000 5,000 ppm level associated with headaches, sleepiness, and stagnant, stale, stuffy air. Poor concentration, loss of attention, increased heart rate and slight nausea may also be present. >5,000 ppm Exposure may lead to serious oxygen deprivation resulting in permanent brain damage, coma and even death. Regulatory exposure limits: ASHRAE Standard 62-1989: 1000ppm: CO2 concentration in occupied building should not exceed 1000ppm. OSHA: 5000ppm: Time weighted average over five 8-hour work days should not exceed 5000ppm Building bulletin 101 (Bb101): 1500ppm. UK standards for schools say that CO2 at averaged over the whole day (i.e. 9am to 3.30 pm) should not exceed 1500ppm. Germany, Japan, Australia, UK: 5000ppm, 8 hours weighted average occupational exposure limit is 5000ppm. Copyright 2010 Extech Instruments Corporation (a FLIR company) All rights reserved including the right of reproduction in whole or in part in any form. 10 CO200-EU-EN-V1.1-11/11