XPert Balance and Bulk Powder Enclosures

Similar documents
LABORATORY EQUIPMENT. Fume Hoods. Class II Biological Safety Cabinets. Carbon Filtered Enclosures. Ventilation Stations

Precision Containment for All Applications

Table of Contents. Laboratory Fume Hoods. UniFlow Fume Hood Systems. Fume Hood System Components. Custom Hoods, Enclosures & Cabinets

Innovation in air purification unmatched air quality

Appendix A.4.1. Safe Fume Hood User Guide

Labconco XPress Laboratory Equipment Where You Need It, When You Need It

Biological Safety Cabinets or Fume Hoods Which one to use?

L a b o r a t o r y E q u i p m e n t C a t a l o g

BIOLOGICAL SAFETY CABINET

BSC CLASS II, TYPE A2 BIOLOGICAL SAFETY CABINET SERIES. Optimum protection of personnel, product and environment. BSC Class

SECTION SPECIAL EXHAUST SYSTEMS. A. Laboratory Ventilation and Hazardous Exhaust Systems. 3. Chemical Fume Hoods and Storage Cabinets

LABGARD NU-602 CLASS II TYPE A2 ANIMAL HANDLING BIOLOGICAL SAFETY CABINET

Sentry Air Systems, Inc.

SECTION BIOLOGICAL SAFETY CABINETS AND LAMINAR AIRFLOW WORK STATIONS

Biological Safety Cabinets. Presented by Mr. Rene Soetens President

Esco PCR Cabinet, Model PCR-4A_. Shown with optional support stand. Polymerase Chain Reaction Cabinets The Proven Solution for Contaminant-Free PCR

DOWNDRAFT BENCH SERIES

A. Standards: Except where modified or exceeded by the requirements of this Specification, conform to the following standards:

CHEMICAL FUME HOOD PROGRAM

5.0 Installation Instructions

THE IMPACT OF FACILITY HVAC SYSTEMS ON BIOLOGICAL SAFETY CABINETS

Adv. 8.4 Guidelines for Biosafety Cabinets Rev

MODEL NUMBER NOMENCLATURE

Safe Operating Procedure (Revised 2/10)

CLEAN AIR SYSTEM.

B. Design meeting requirements of the particular class and type of cabinet as follows:

FUME CUPBOARDS DESIGN FEATURES GENERAL ARRANGEMENT FACE VELOCITIES EXTRACTION SYSTEMS AIRFLOW CHARACTERISTICS

General Purpose Laminar Flow Cabinets, Horizontal and Vertical. Horizontal Flow Vertical Flow

Clean Bench Animal Transfer Stations

High Plume Blowers Suggested Specification Section (Master Format 1996) Suggested Specification Section (Master Format 2004)

Fume Hood Safety TYPES OF FUME HOODS

LABORATORY EQUIPMENT. A. Section Accessibility for People with Disabilities. B. Section Plumbing Basic Materials and Methods

PART III. Laboratory equipment

Automatic Cyanoacrylate Fuming Chamber 24 33T 60T 72XL. Safely Detect Quality Latent Prints and Reproducible Results

LABORATORY CABINETS 103

General Purpose Laminar Flow Cabinets, Horizontal and Vertical

Faster BH-EN and BHG Class II Microbiological Safety Cabinets

Advancing Laboratory Ventilation Technology

WATER WASH BOOTHS PUMP TYPE PUMPLESS TYPE

Animal Tissue Culture SQG 3242 Equipment and Lab design. Dr. SITI PAULIENA MOHD BOHARI

Glossary of Terms Laboratory & Fume Exhaust Fans

Faster UltraSafe Class II Microbiological Safety Cabinets. Protection, safety, reliability. And more.

Biosafety Cabinets and Fume Hoods Supplemental Material (OHS_BIO304)

Class II Biosafety Cabinets

General Purpose Laminar Flow Cabinets, Horizontal and Vertical

Supplying Clean Air to Industry

ALUMINUM REPAIR CERTIFICATION EQUIPMENT. Eliminate hazardous aluminum dust with a Eurovac system. BODY REPAIR CERTIFICATION EII Wet Mix Dust Collector

PHYSICAL FACILITIES Consultant s Handbook Division 01 General Special Requirements for Laboratories. Page 1 of 6

Environmental Health and Safety Biological Safety Program Guidance Document

Biological Safety Cabinets. Biological Safety Cabinets

DUST COLLECTORS. OEM and Industrial Air Handling Specialist Snider Road, Mason, OH Telephone:

ANIMAL RESEARCH PRODUCTS. Biological Safety Cabinets Animal Transfer Stations

Animal Containment Workstation The Portable Safety Solution for Animal Research Laboratories

CENTRIVAP CENTRIFUGAL CONCENTRATORS & COLD TRAPS

Appendix C IBC Safe Use of Biosafety Cabinets SOP

Basic Hood Considerations The design process for a successful kitchen ventilation system starts when the engineering design HVAC team receives

SECTION LABORATORY EQUIPMENT CONSTRUCTION STANDARD PART 1: GENERAL Scope of Standard

How To Select The Right Laboratory Hood System

Displacement Ventilation

Division 11 - Equipment

Air Showers The Safety Solution for Animal Research Laboratories. Viva Air Shower, Model - EAS-1B.

AIR HANDLERS. Central Station (CS3) Roof Mounted (RT) (Modular AL Frame AHU)

Safe Use of a Class II Biological Safety Cabinet

Research Document Type: PROCEDURE. Effective Date: January 01,2015 Author Steven Hayes

AIR SCRUBBER X-3400, X-3500, X-3700

BESF Box Ventilator USA CAN. Product Information. Mechanical Installation. ... Chapter 3. Electrical Installation. ... Chapter 4

ACT Cartridge Dust Collector

1.1 This section applies to air handling units for HVAC Systems.

INDEX 45 FILTER DRYER VALVE CLEANING 46 FILTER DRYER CAKE REMOVAL 47 X-AXIS CENTRIFUGE CONTAINMENT

General & Special Purpose Forensic Evidence Drying Cabinets

How to comply with CDC Guidelines

Containment of Basket Filtration Centrifuges ABSTRACT

RELIABLE SOLUTIONS FOR INDUSTRIAL APPLICATIONS ENGINEERED VACUUM SYSTEMS

FT900BD FLIGHT-TYPE DISHWASHER

HOFFMAN & LAMSON ENGINEERED VACUUM SYSTEMS

EHH SERIES DRAW THROUGH FAN COILS CFM

NEW! Thermo Scientific 1300 Series B2 Class II, Type B2 Biological Safety Cabinet

ESCO. LabTech America A M P L E SCIENTIFIC CITIZEN SCALE INC LABOMED, INC. WORLD CLASS, WORLD WIDE. Rev.2

INSTALLATION AND OPERATION MANUAL Model BR70/BR130

Florida International University. Biosafety Cabinet Program

INSTALLATION INSTRUCTIONS WALL MOUNT LINER INSERT

KIC Air Curtains. Technical Guide

Class II, Type B2 (Total Exhaust) Biological Safety Cabinets The Safety Solution for Life Science Laboratories

UC DAVIS Campus Standards & Design Guide DIVISION 11 - EQUIPMENT

SAFE DRINKING WATER AND TOXIC ENFORCEMENT ACT

FlaskScrubber & SteamScrubber Laboratory Glassware Washers

Internal ridged board 1" x 1.5 foil face installation shall be installed on roof, walls and base of casing.

F116 (Series 16000) Self-Contained Ductable Commercial Air Cleaner

BRAHM. Working together for a crystal-clean production line COLLECTOR SERIES

INSTALLATION INSTRUCTIONS WALL MOUNT LINER INSERT

Air Make-Up Units. Product Line Overview & Capabilities

Class II, Biological Safety Cabinets The Industry s Most Energy Efficient Class II Cabinet

GUIDANCE NOTE 3: SAFE USE OF EQUIPMENT

General & Special Purpose Forensic Evidence Drying Cabinets

VP, Facilities and Construction Management

Freeze Dry Systems and Accessories

ORION. Safety Class 2 100% exhaust. ALMOST Rocket Science

FT900SDBD DUAL RINSE FLIGHT-TYPE DISHWASHER

SECTION DUCTWORK

Fume Extraction Workstation

Transcription:

XPert Balance and Bulk Powder Enclosures

Labconco Product Diversity Glove Boxes Fume Hoods Biological Safety Cabinets Freeze Dryers Glassware Washers Filtered Enclosures Carts & Mobile Benches Centrifugal Concentrators

Fume Hood or Enclosure? Chemical Fume Hood Balance Enclosure

Fume Hood or Enclosure? Chemical Fume Hoods Capture, contain and remove airborne contaminants Large energy consumption Low duct velocity, 2000-2500 fpm High installation cost

Fume Hood or Enclosure? Balance Enclosure Capture, contain and remove airborne particulate contaminants Small energy consumption High duct velocity, 3500-4000 fpm HEPA filter required Low installation cost

XPert Balance Enclosures and Systems Markets Pharmaceutical Industrial Hygiene Chemistry Labs Life Sciences University Hospital/Clinical

XPert Filtered Balance Systems Provide user protection Patented rear baffle with zones of perforation Built-in blower need no ducting to outside Exhaust HEPA filter Isolation supports minimize blower vibration Low installation costs Available in 2', 3', 4', 5',and 6' widths Optional left-side mounted waste chute

XPert Filtered Balance Systems Features & Benefits True bag-in/bag-out HEPA filter disposal system Optional HEPA filter location permits leak detection scan of entire surface Front mounted pressure gauge Tempered safety glass sash and sides Side-entry air foils Deep 23.4 interior Built-in impeller has vibrationisolation supports Space-saving design Built-in airflow monitor Quiet operation as low as 48 decibels Static-dissipative epoxy-coated aluminum frame and steel rear plenum and baffle Upper dilution air supply Ergonomic angled sash Upper containment sash foil Patented Clean Sweep air foil

XPert Filtered Balance Systems Complete system self contained Internal fan w/vibration isolation supports Filter mounted above integral fan Fits on existing bench Exhaust options Recirculate into room Exhaust to house via canopy

XPert Filtered Balance Stations Protect the user Built-in HEPA filter removes particulates Must be ducted to the outside using a remote blower Patented rear baffle Powder weighing applications in pharmaceutical research are typical Deep interior accommodates large micro and analytical balances Optional left-mounted waste chute

XPert Filtered Balance Stations Features & Benefits Front-mounted pressure gauge Deep interior easily accommodates large analytical balances True bag-in/bag-out HEPA filter disposal system Side-entry air foils Exhaust collar allows connection to ductwork and remote blower Static-dissipative epoxy-coated aluminum frame and steel rear plenum and baffle Built-in airflow monitor Upper dilution air supply Ergonomic angled sash Tempered safety glass sides and sash Upper containment sash foil Patented Clean-Sweep air foil

XPert Filtered Balance Stations No internal fan Filter mounted above enclosure Tie into house exhaust system Minimized vibration

XPert Balance Enclosures Provide user protection Must be ducted to the outside or connected to a FilterMate Portable Exhauster Deep interior accommodates large micro and analytical balances Patented rear baffle Exhaust less tempered room air than fume hoods conserving energy No switches, wiring or service fixtures that can add costs

XPert Balance Enclosures 2', 3', 4', 5', 6', & 8' models 22.3" or 31.5" exterior height 23.4" or 30.0" interior depth Top or bottom exhaust connection Airflow monitor standard Optional left-side mounted waste chute

XPert Balance Enclosures Features & Benefits Tempered safety glass front sash, sides and top Built-in airflow monitor Flexible exhaust options Upper dilution air supply Deep interior Ergonomic angled sash Upper containment sash foil Side-entry air foils Static dissipative epoxycoated aluminum frame and steel rear plenum and baffle Patented Clean-Sweep air foil 8", 10" or 12" high sash opening

XPert Balance Enclosures Easy-Order One part number includes: 2', 3' or 4' Balance Enclosure FilterMate Portable Exhauster Bag-In/Bag-Out HEPA Filter Vent Connection Kit with adapters and hose connectors Guardian Jr. or 1000 Digital Airflow Monitor Dished Epoxy Work Surface

XPert Balance Enclosures Performance-Enhancing Features Enhanced Containment Horizontal slotted baffle Clean-Sweep air foil Bypass air Sash foil Side-entry air foils

XPert Balance Enclosures Containment-Enhancing Features

XPert Balance Enclosures Containment-Enhancing Features Clean-Sweep Air Foil allows air to sweep the work surface for maximum containment. Side-Entry Air Foils are located on either side of the sash opening and direct clean airflow into the chamber. Upper Dilution Air Supply introduces air from above the work area and bathes the back of the sash with clean air and directs concentrations away from the sash opening.

XPert Balance Enclosures Containment-Enhancing Features Aerodynamic Air Foil Clean-Sweep openings Directs airflow into hood and across work surface Minimizes turbulence Patented feature

XPert Balance Enclosures Containment-Enhancing Features Upper Containment Sash Foil Aerodynamic shape directs airflow Air gap bleeds high velocity air into enclosure Reduces contaminants behind sash Patented feature

XPert Balance Enclosures Containment-Enhancing Features Side-Entry Air Foils Provides horizontal flow Eliminates dead-air zone around sash area Exclusive feature

XPert Balance Enclosures Containment-Enhancing Features Upper Dilution Air Supply Bypass Air (5 10%) Purges top of enclosure Bathes area behind sash Reduces chemical concentrations Introduces clean air near user s breathing zone Exclusive feature

XPert Balance Enclosures Containment-Enhancing Features Slotted Horizontal Baffle Greatly reduces vortex (roll) Provides horizontal laminar flow Patented feature

XPert Balance Enclosures Vibration Isolation Vibration: Affects weighing operations Greater effect in taller buildings Placement near outside wall is better Corners are more rigid Work surface needs to be stable and level

XPert Balance Enclosures Validation Documentation Validation package included with each enclosure: Independent test data: SafeBridge Consultants: Powders Containment Test Exposure Control Technologies: ASHRAE 110-1995 Test IQ/OQ Protocol Manufacturer s quality assurance inspection form

XPert Balance Enclosures Factors Affecting Performance Electrostatics Temperature Humidity Vibration Lab air currents Exhaust system Installation User technique

XPert Balance Enclosures Electrostatics Electrostatic Problems: Lightweight powders float Static attraction can be greater than air currents Balance performance can be adversely impacted Electrostatic problems are compounded by: Enclosure materials of construction Low humidity dry air holds static charge Temperature heat sources affect air currents

XPert Balance Enclosures Electrostatics Tempered safety glass is static dissipative releases static more readily Acrylic has insulative properties static charge clings to material Recommend the use of ionizers aligns charges Material Surface Resistivity Example Conductive 0 10 5 Ω per square CM Skin, Metals Static dissipative 10 5 10 9 Ω per square CM Glass Antistatic 10 9 10 12 Ω per square CM Polyethylene bag Insulative 10 12 Ω per square CM Acrylic box, Packing foam, Styrofoam

XPert Balance Enclosures Vibration Isolation Marble Slab 3" Marble Slab Economical Fits comfortably Does not affect containment Rubber/Vinyl Isolator Pads Inexpensive Effective vibration dampening

XPert Balance Enclosures HVAC/Installation Installation Location: Low traffic areas Away from open doors No cross-drafts Avoid supply air diffuser Lab location Low throw velocity Balanced HVAC system Adequate supply air

XPert Balance Enclosures Typical venting with FilterMate Portable Exhauster

XPert Balance Enclosures Typical venting from top with remote blower

XPert Balance Enclosures Typical venting from back with remote blower

XPert Balance Enclosures FilterMate Portable Exhauster Point-of-use exhauster Internal spark-proof impeller Houses filter to purify air Provides airflow to XPert

XPert Balance Enclosures FilterMate Portable Exhauster Narrow design fits comfortably on bench or on the floor Comes standard with 8' polypropylene hose, clamps and Exhaust Transition Adapter

XPert Balance Enclosures FilterMate Portable Exhauster Large capacity 280 CFM Provides up to 100 fpm face velocity for 2', 3' or 4' Enclosure** HEPA Filter, Carbon Filter or BOTH with one FilterMate Carbon filter types: Organic Ammonia Formaldehyde Quiet operation 54 dba at 150 CFM operation ** Contact Labconco for technical assistance regarding using multiple FilterMates for 5', 6' and 8' wide enclosures

XPert Balance Enclosures FilterMate Portable Exhauster Three ways to exhaust: Through FilterMate to house exhaust system Through FilterMate back into the room Directly from enclosure to house exhaust system

XPert Balance Enclosures FilterMate Portable Exhauster Intrinsically safe design Filter is in a negative pressure plenum Bag-In/Bag-Out HEPA filter replacement Minimizes servicer contact HEPA filter can be scanned for leaks from outside FilterMate

XPert Balance Enclosures Convenience Features Adjustable speed control (qualified personnel only) Additional grounded outlet Can power airflow monitor Circuit breaker 5 amp

XPert Balance Enclosures User Technique Use glass or metal weigh dishes and tools static dissipative Work slowly, methodically Be cognizant of surroundings Work in a bright environment; rest eyes often

XPert Balance Enclosures User Technique No enclosure can compensate for poor technique. Labconco s Enclosure has been proven to provide a safe work environment even with less proficient technique. The Labconco XPert Balance Enclosure is capable of providing excellent containment where the operator uses proficiency in technique Where operators exercise less than excellent technique and/or have minor spillage good containment performance is achieved. Edward J. Haggerty, CIH SafeBridge Consultants, Inc.

XPert Balance Enclosures Safety Features Spring-loaded sash latch Exclusive feature Allows for easy loading of equipment Airflow monitor standard Guardian Jr. (low flow) Guardian 1000 (high/low flow) Easy to install Calibrated in house

XPert Bulk Powder Transfer System Definitions & Terms Features Planning & Start Up Loading Technique Procedure Clean Up

XPert Bulk Powder Enclosures Definitions Source Container: Bulk container or fiber drum containing powder Transfer Cart: Transport device for source container Transfer Plate: Platform of Transport Cart that connects to the lift Electric Lift: Raises the source container for access through enclosure Transfer Cart Latch: Secures Transport Cart to electric lift

XPert Bulk Powder Enclosures Definitions Safety Sleeve: Bag placed around outside of source container for additional protection Retaining Band: Secures the Safety Sleeve to the Source Container and secures the source container bags to the retaining ring Retaining Ring: Ring on work surface used to secure source container bags in place Waste Chute: Hole on side of enclosure that allows attachment of bag for waste disposal

XPert Bulk Powder Enclosures Terms Transfer Cart Latch Electric Lift Source Container Transfer Plate Transfer Cart

XPert Bulk Powder Enclosures Markets Pharmaceutical Research Chemical Manufacturing

XPert Bulk Powder Filtered Systems, Filtered Stations and Enclosures Designed to protect user during powder transfer operations Filtered Systems and Stations include a built-in HEPA filter to remove particulates in the exhaust air Patented rear baffle with zones of perforations promotes horizontal airflow

XPert Bulk Powder Filtered Systems, Filtered Stations and Enclosures Guardian 1000 digital air monitor Front-mounted pressure gauge Upper dilution air supply Patented rear baffle Tempered safety glass Optional left-side mounted waste chute Optimal HEPA filter location Side-entry air foils Base stand included Push button-controlled electric hydraulic lift mechanism True bag-in/bag-out HEPA filter disposal system Transfer caddy included Stainless steel work surface with front trough

XPert Bulk Powder Filtered Systems Built-in HEPA filter to remove particulates in the exhaust air Built-in blower no ducting to the outside required Push-button operated electric hydraulic lift mechanism Electronic beam automatically stops lift mechanism at optimal access height Available in 5- and 6-foot widths

XPert Bulk Powder Filtered Systems Required accessories include Retaining Ring Safety Sleeves Retaining Bands Access Opening Cover

XPert Bulk Powder Filtered Stations Have an exhaust collar and require ducting to house exhaust or a remote blower Available in 5' and 6' widths Require Remote Blower Ductwork Retaining ring and bands Access Opening Cover

XPert Bulk Powder Enclosures Requires venting by connection to a remote blower or house exhaust to the outside Available in 5', 6', and 8' widths Require Retaining Ring and Bands Require Safety Sleeves Require Access Opening Cover

XPert Bulk Powder Enclosures Exclusive Features Transfer Cart is latched in place to keep it from moving Transfer Cart handle is out of the way during the weigh procedure

XPert Bulk Powder Enclosures Exclusive Features Lift is electric and automatically stops at access height. Transfer Plate locates source container in correct position for access opening.

XPert Bulk Powder Enclosures Exclusive Features Retaining Ring provides a way to secure the source container liner bag. Retaining Band secures the source container liner bag to the retaining ring to seal the bag to the work surface.

XPert Bulk Powder Enclosures Exclusive Features Baffle pivots downward for easy cleaning.

XPert Bulk Powder Enclosures Digital airflow alarm for continual display of airflow velocity Hydraulic lift holds sash up while loading and unloading

XPert Bulk Powder Enclosures Access Opening Variable location Right side Center Left side (not shown)

XPert Bulk Powder Enclosures Planning Understand procedures and equipment Arrange for minimal disruptions Consult your Safety Officer for protective equipment recommendations Transport containers to Bulk Powder Enclosure

XPert Bulk Powder Enclosures Start Up Turn on blower and purge system for several minutes before beginning procedure. Check the baffle air slots for obstructions.

XPert Bulk Powder Enclosures Start Up Load with necessary equipment and supplies. Decontamination products may be useful. Ensure that sash is fully closed when working in enclosure. Position equipment so air flow is not obstructed.

XPert Bulk Powder Enclosures Loading Attach Safety Sleeve to source container and place on cart. Dock cart to Bulk Powder Transfer System. Push button to activate Lift System. Lift stops automatically when source container is at access height.

XPert Bulk Powder Enclosures Loading Pull Safety Sleeve through opening in work surface and secure on Retaining Ring with Retaining Band. Allow system to purge airborne contaminants from the work area.

XPert Bulk Powder Enclosures Loading Remove lid from source container. Place inside enclosure, or Decontaminate thoroughly and remove from enclosure Secure inner bags from source container to the Retaining Ring with Retaining Bands.

XPert Bulk Powder Enclosures Work Technique Keep all materials inside the lower air foil and perform all contaminated operations as far to the rear of the work area as possible. Segregate all clean and contaminated materials in the work area. Avoid using techniques or procedures that disrupt airflow patterns of the enclosure.

XPert Bulk Powder Enclosures Work Technique Keep all materials and hands inside enclosure until the end of the procedure and all materials have been wiped down. At end of wipe down, wipe down arm covers and gloves. Remove arm covers and gloves inside enclosure. Place waste materials in waste chute.

XPert Bulk Powder Enclosures Work Procedure Remove lid of source container and place to the side. Secure both outer and inner bags of source container to retaining ring with retaining band. Scoop powder into container.

XPert Bulk Powder Enclosures Work Procedure When weighing procedure is completed, close container and place to the side. The user may either wipe container and lid after each weigh out or when all weigh outs are completed.

XPert Bulk Powder Enclosures Clean Up When procedure is completed, remove retaining band and place in disinfecting solution or discard in waste chute. Place inner bag in source container. Repeat for outer bag and place lid back on source container. Vacuum if necessary. With a decontamination solution such as IPE, wipe down source container lid, inside of safety sleeve, work surface and any other materials inside the enclosure.

XPert Bulk Powder Enclosures Clean Up Remove retaining band on safety sleeve. Pull safety sleeve up through opening in work surface and discard. With a decontamination solution, wipe down arm covers and gloves before removing and discarding in waste chute.

XPert Bulk Powder Enclosures Clean Up Lower source container until it comes to a stop. Unlatch Transfer Cart and return source container. Remove materials from inside enclosure and properly dispose of materials. Allow unit to purge for 2 minutes before turning off exhaust system.

XPert Balance Enclosures Accessories Two exhaust configurations Top exhaust transition Bottom exhaust transition Two ductwork options 5" diameter flexible hose 6" diameter PVC Duct

XPert Balance Enclosures Accessories Remote Blowers TEFC-type direct drive motor blower Corrosion-resistant phenolic-coated steel housing and impeller Belt drive motor blower Corrosion-resistant phenolic-coated steel housing and impeller Dry powder epoxy-coated base and weather cover

XPert Balance Enclosures Accessories Canopy Connection Kit Remote Exhaust Fan Ductwork Dampers

XPert Balance Enclosures Accessories Kim wipe holder Spray bottle holder Ergonomic chair Ergonomic footrest

XPert Balance Enclosures Accessories 3-Shelf accessory kit Reduces interior clutter Shelves can rest on back wall Does not affect airflow Printer shelf inside or out Utensil shelf Utility shelf disposables or calibration weight

XPert Balance Enclosures Accessories Trace Odor Carbon Filters for XPert System Carbon Filters Fluorescent Light HEPA Filters

XPert Balance Enclosures Accessories Solid Epoxy Work Surfaces Stainless Steel Work Surfaces

XPert Balance Enclosures Accessories Telescoping Base Stand Transfer Caddy Protector Standard Storage Cabinets Telescoping Base Stands with Casters

XPert Balance Enclosures Conformity SEFA 1-2005 ANSI Z9.5-1993 Modified ASHRAE 110-1995 UL 3101-1/61010-1 (115v models) CAN / CSA C22.2 No. 1010.1 (115v models) CE Conformity Marking (230v models) Safebridge Tested

XPert Balance Enclosures Additional Information Product literature Lunch-and-Learn sessions XPert video Trade shows 3-part specifications CAD drawings