High Sensitivity Explosives Detection using Dual-Excitation- Wavelength Resonance-Raman Detector

Similar documents
RAMAN SPECTROSCOPY See more, faster than ever before

Compact Raman spectrometer system for low frequency spectroscopy

An FT-NIR Primer. NR800-A-006

Low-Frequency Raman Spectroscopy Enabling Affordable Access to the Terahertz Regime

Stand-off Detection of Trace Explosives by Infrared Photothermal Imaging. Trace Contamination

Through Container THz Sensing: Applications for Explosives Screening

DEVELOPMENT OF THE INFRARED INSTRUMENT FOR GAS DETECTION

Laser Diagnostics and Optical Measurement Techniques

Long-distance remote simultaneous measurement of strain and temperature based on a Raman fiber laser with a single FBG embedded in a quartz plate

A Cost Effective Multi-Spectral Scanner for Natural Gas Detection

Propagation loss in optical fibers (01/29/2016)

THE NEXT GENERATION IN VISIBILITY SENSORS OUTPERFORM BOTH TRADITIONAL TRANSMISSOMETERS AND FORWARD SCATTER SENSORS

Time Domain THz Spectroscopy and Related Sensing Applications

Current Research Topics in Optical Sensors and Laser Diagnostics

Optical Time Domain Reflectometry (OTDR)

EXTENDING RAMAN INTO THE THz REGIME

On-line Side Stream Oil in Water Analyser - Model OIW-EX 100 or OIW-EX 1000

Application Note. Which characterization method is best for your application? The Future of Thermal Imaging is Here!!!

Specifications Wavelength Range: 190 nm to 800 nm continuous

Bunch Length Interferometer System

for Photonics Advanced phosphor and glass technologies for high performance alignment, detection, imaging and spectroscopy applications

高等食品分析 (Advanced Food Analysis) II. INSTRUMENTS FOR OPTICAL SPECTROSCOPY *Instruments for optical spectroscopy: Optical spectroscopic methods are

The FEL Facility Optical Diagnostics and FEL Characterization

Presentation Title October 18, Physical Sciences Inc. 20 New England Business Center Andover, MA Outline

Omnisens DITEST TM FIBER OPTIC DISTRIBUTED TEMPERATURE & STRAIN SENSING TECHNIQUE

Office of Nonproliferation and Verification Research and Development University and Industry Technical Interchange (UITI2011) Review Meeting

Gas Temperature Measurements with High Temporal Resolution

Electro Optical Components, Inc Skylane Boulevard, Santa Rosa, CA Toll Free: 855-EOC

Communications. Sensing. Instrumentation. Delivering tomorrow s optical and photonic solutions today. BREAKTHROUGH PHOTONIC PRODUCTS

Analisi del Focus-shift su sorgenti multikw. Luca Porcelluzzi Ophir Spiricon Europe GmbH

ANNEX I. Technical specifications for: Supply of various Trace Explosive Detection Systems

Laser Safety and Classification Full Version (CA-1110) Advanced Version (CA-1111) Basic Version (CA-1112)

TSI AEROTRAK PORTABLE PARTICLE COUNTER MODEL 9110

Standard Operating Procedure Raman Microscope - Renishaw invia

Laser Analyzers for NH3 Safety Monitoring and Leak Detection

RAMAN SCATTERING AND OTDR BASED DISTRIBUTED SENSOR FOR TEMPERATURE MONITORING APPLICATIONS

Compact Portable QEPAS Multi-gas Sensor

Integrated optical fiber shape senor modules based on twisted multicore fiber grating arrays

Multitel develops components and full instruments for in-vitro and in-vivo diagnostics in the biomedical, food and drug sectors.

FPI Detectors Pyroelectric Detectors with Spectrometer Functionality

SPECTRAL INSTRUMENTATION

Application Note CBRNE Stand-Off Detection of Toxic Industrial Chemicals in Industrial Complexes Using RAPIDplus

Compact Sensor Heads for In-situ and Non-Contact Standoff Gas Sensing using TDLAS in Harsh Environments

In-line Oil in Water Analyser - Model OIW-EX100P and OIW-EX1000P

MSA s Guide to Selecting the Right Flame Detector for Your Application

MIR Spectroscopic Sensing. IQCLSW 19th Sept Simon Nicholson Sales Director Cascade Technologies

A Method for Fire Detecting by Volume and Surface Area Concentration Based on Dual Wavelengths

Intrusion Detection System based on Speckle Pattern change in Fiber Optic Sensors...Abdulkareem H. Dagher, Shehab A kadhim, Hawraa H.

6340(Print), ISSN (Online) Volume 4, Issue 2, March - April (2013) IAEME AND TECHNOLOGY (IJMET)

This document is downloaded from DR-NTU, Nanyang Technological University Library, Singapore.

CCD and CID solid-state detectors

Understanding total measurement uncertainty in power meters and detectors

FiberSystem. Fiber Optic Linear Heat Detection for Special Hazard Applications

Variable Temperature Options for Optical Experiments

Integrating Sphere Detectors. Wide Range of Stock Items. Individual Custom Configuration and Design. OEM Design and Production. Traceable Calibrations

Distributed Rayleigh scatter dynamic strain sensing above the scan rate with optical frequency domain reflectometry

Instrumentation for Beam Profiling in the Terahertz Regime

Remote Detection of Leaks in Gas Pipelines with an Airborne Raman Lidar. Strategic Insights, Volume VII, Issue 1 (February 2008)

WHITE PAPER FIBER OPTIC SENSING. Summary. Index. Introduction. About Fischer Connectors

Recent Progress in Lightwave Sensing Technology

Interferometric optical time-domain reflectometry for distributed optical-fiber sensing

Wildland fire phenomenology experiments:

LEARNING SPECIAL HAZARD DETECTION TYPES

Automatic Detection of Defects on Radiant Heaters Based on Infrared Radiation

Distributed Temperature Sensing

Methodology of Implementing the Pulse code techniques for Distributed Optical Fiber Sensors by using FPGA: Cyclic Simplex Coding

E11 Tests of trace gas concentration multicomponent. monitoring system

How SWIR Imaging contributes to the optimization of automatic quality control in the food industry

DEVELOPMENT AND FUNCTIONALITY OF SORTING SYSTEM THE POLYMARK INDUSTRIAL SCALE

TacBio: A UV-based Biological Aerosol Detector. TacBio Overview Page 1

DETECTION AND LOCALIZATION OF MICRO AND MULTIPHASE LEAKAGES USING DISTRIBUTED FIBER OPTIC SENSING

ANALYSIS OF OTDR MEASUREMENT DATA WITH WAVELET TRANSFORM. Hüseyin ACAR *

Band Selection & Algorithm Development for Remote Sensing of Wildland Fires

Technology to Application

A GEM-based detector for detection and imaging sparks and flames G. Volpe 1 and V. Peskov 2. Abstract

Fibre Optic Sensors: basic principles and most common applications

Table of Contents. Why Raman? and Why Now? 2. The BaySpec Advantage 3. Customized Instruments 14. OEM Partnership 14

HELPCOS THE HELICOPTER BASED PIPELINE CONTROL SYSTEM OF VNG - VERBUNDNETZ GAS AG

STANDOFF DETECTION OF IEDs WHILE ON THE MOVE

Low-Power Detectors. Key Features. 818 Series. Power levels from pw to 2W. Wavelengths from 190 nm to 1800 nm. Proprietary detector optics

Development of optical fiber sensors for selected chemical and physical sensing applications

Model 6650SP. Filber Optic-Based UV-Fluorescence Oil-in-Water Analyzer

Link loss measurement uncertainties: OTDR vs. light source power meter By EXFO s Systems Engineering and Research Team

PC474 Lab Manual 1. Terry Sturtevanta and Hasan Shodiev 2. Winter 2012

Wide Area Sensor Needs

Why is Laser Safety Important Laser usage is growing rapidly e.g. laser points, CD/DVD, industrial applications Heavy use now in laboratories for

University of California Santa Barbara Department of Chemistry and Biochemistry Optical Characterization Facility. Safe operations manual

Laser Safety Products. Edition 20

33 Watt Low Voltage LED Fixture for Hazardous Location Lighting - Low Profile Linear Fixture - C1D2

User Detection Sensors

WHAT IS LASER ULTRASONIC TESTING?

Health Monitoring System Using FBG-Based Sensors for a 12-Story Building with Column Dampers.

Polarized Light Scattering of Smoke Sources and Cooking Aerosols

COMPACT FLAME CONTROLLER CFC 2000

16S-Series 150W. In Vivo SPF Sunscreen Testing Solar Simulator

Temperature characterisation of LPG sensors for monitoring deterioration in reinforced concrete

Date: Wednesday, January 31, 2018 Time: 8:30 am 5:00 pm Location: Moscone West (Room 2018, Level 2)

The Optical Scattering Calibration System at SNO+ IOP 2015

UV LED Curing: The Time is Now

NEED FOR LASER SAFETY

Transcription:

High Sensitivity Explosives Detection using Dual-Excitation- Wavelength Resonance-Raman Detector Balakishore Yellampalle *, William McCormick, Hai-Shan Wu, Mikhail Sluch, Robert Martin, Robert Ice and Brian E. Lemoff Advanced Technologies Group, WVHTC Foundation, 1000 Technology Drive, Suite 1000, Fairmont, WV, USA 26554 ABSTRACT A key challenge for standoff explosive sensors is to distinguish explosives, with high confidence, from a myriad of unknown background materials that may have interfering spectral peaks. To meet this challenge a sensor needs to exhibit high specificity and high sensitivity in detection at low signal-to-noise ratio levels. We had proposed a Dual-Excitation- Wavelength Resonance-Raman Detector (DEWRRED) to address this need. In our previous work, we discussed various components designed at WVHTCF for a DEWRRED sensor. In this work, we show a completely assembled laboratory prototype of a DEWRRED sensor and utilize it to detect explosives from two standoff distances. The sensor system includes two novel, compact CW deep-ultraviolet (DUV) lasers, a compact dual-band high throughput DUV spectrometer, and a highly-sensitive detection algorithm. We choose DUV excitation because Raman intensities from explosive traces are enhanced and fluorescence and solar background are not present. The DEWRRED technique exploits the excitation wavelength dependence of Raman signal strength, arising from complex interplay of resonant enhancement, self-absorption and laser penetration depth. We show measurements from >10 explosives/pre-cursor materials at different standoff distances. The sensor showed high sensitivity in explosive detection even when the signalto-noise ratio was close to one (~1.6). We measured receiver-operating-characteristics, which show a clear benefit in using the dual-excitation-wavelength technique as compared to a single-excitation-wavelength technique. Our measurements also show improved specificity using the amplitude variation information in the dual-excitation spectra. Keywords: Deep Ultra Violet, Resonance Raman Spectroscopy, Dual Excitation Wavelengths, Explosive Detection, Detection Algorithm. 1. INTRODUCTION The Dual-excitation-wavelength (DEW) resonance Raman technique 1 uses a laser source with two deep-ultraviolet excitation-wavelengths. The source wavelengths are chosen so that the Raman spectrum in the fingerprint region of the lower excitation wavelength is well separated from the second excitation laser wavelength. Raman scattered photons from the sample are simultaneously measured using a dual-band detection spectrometer. An important requirement for a Dual-Excitation-Wavelength Resonance-Raman Detector (DEWRRED) is the simultaneous detection of the two excitation spectra from a single interrogation area. This ensures that any spatial variation of explosive fill-factor in the sample, or any time-dependent phenomena, such as motion or degradation present in a real detection scenario contribute equally to measurements at both of the excitation wavelengths. The main reason for using two excitation wavelengths is to make use of Raman spectral intensity variation as an additional orthogonal signature. The Raman scattering intensity from an explosive depends on its Raman cross-section, laser penetration depth and the resonance Raman self-absorption. Resonance Raman cross-sections and the absorption of both the excitation and scattered light vary strongly with excitation wavelength in the DUV region 2-4. This results in detected Raman band intensities that depend on excitation wavelength in a complex way, forming unique signatures for different compounds 5. As an example, DEW signatures can be constructed as a combination of spectral peak positions * byellampalle@wvhtf.org; phone 1 304 333 6462; www.wvhtf.org/programs/advancedtech/

(i.e. Raman frequency shifts) and the ratio of the Raman band intensities measured at the two-excitation-wavelengths. In addition to high signal strengths, Raman signatures in the DUV region are free from solar radiation and have minimal interference from fluorescence 6. The main advantages of using a Raman spectroscopy approach with two or more excitation wavelengths are a) improved specificity compared to single excitation wavelength and b) higher sensitivity 5. Improved specificity was experimentally studied in our previous work where we thoroughly characterized 7 the DUV Raman responses of fifteen explosives and sixteen background materials at multiple DUV wavelengths (213 nm, 229 nm, 238 nm, 244 nm, 248 nm, and 266 nm) to better understand the potential performance of multiple-excitation-wavelength Raman standoff explosives sensors 5. Improvement to specificity was also seen when only two excitation wavelengths were used. WVHTC Foundation has developed a laboratory prototype of a DEWRRED sensor that detects explosives using simultaneously-measured Raman signatures with 236.5 nm and 257.5 nm laser excitations. We show results of detection of several bare bulk explosives or small quantities of explosives of ~50 g from distances of 3.5 ft and 14 ft using a low laser power of <250 w. The exposure time for detection varied from 0.1 s to 60 s. 2. DUAL EXCITATION WAVELENGTH RESONANCE RAMAN DETECTOR The laboratory prototype of the DEWRRED sensor used two laser beams at 236.5 nm and 257.5 nm as shown in the schematic of Figure 1. The DUV wavelengths were generated by doubling blue (473 nm) and green (515 nm) lasers. Ideally, the two laser beams would have similar divergence, similar beam profile and copropagate towards the test sample. However, in the current implementation, the two beams travelled at an angle and overlapped on the target. The 257.5 nm beam was aligned to travel along the optical axis of the collection system. In the future, we plan to combine both the laser beams and copropagate them along the optical axis of the collection system. The Raman scattered light from the excitation region was collected using a 6 inch primary mirror. Raman filters were used in the collection system to attenuate the Rayleigh scatter from the two lasers. We designed the collection system to image an illuminated target spot from distances of 3 ft to greater than 50 ft. Figure 1: Schematic of dual-excitation-wavelength resonance Raman detector setup.

A DUV fiber bundle (spot-to-line converter) delivered the Raman light from the collection system to the input of a dualband spectrometer. The spectrometer output was separated into two Raman bands and analyzed using a highly sensitive algorithm. Details of the algorithm are described in Reference [8]. DUV Lasers Dual-band Spectrometer Collection system Figure 2: Laboratory prototype DEWRRED sensor system consisting of two DUV laser sources, illumination and collection optics, and a dual-band detection spectrometer. A picture of the sensor system is shown in Figure 2. As part of the DEWRRED sensor, we have built two powerscalable, turnkey, DUV lasers that have the potential to be made compact. The laser technology also has the potential for field use and can be battery operated. These lasers produce narrow line-width DUV light at 257.5 nm and 236.5 nm by coupling diode-pumped solid-state CW blue/green laser sources at 515 nm and 473 nm wavelengths to external second harmonic generation (SHG) units. The 257.5 nm laser produced >2 mw of DUV light with a conversion efficiency of ~2%. The 236.5 nm laser produced <250 W of light at an efficiency of 0.5%. Lower efficiency at 236.5 nm was partly due to lower power at 473 nm. Each of the laser heads occupied a footprint of 18 x18 on an optical bread board. We are currently developing second generation prototypes which are anticipated to be compact, more efficient (>15%) and produce higher power than the current laboratory prototype. For the explosive detection experiments presented here, the lasers were typically operated at nearly equal power levels from 20 w to 250 w. Both the DUV laser powers were continuously monitored in order to normalize the measured relative intensities of the Raman bands. The two laser beams were collimated with an identical set of optics. Owing to differences in refractive indices of optics at different wavelengths and wavelength dependence of divergence, the beams had slightly different shapes on the sample. This beam mismatch limited the application of the full DEW technique in these lab measurements to uniform samples (bulk or traces) with 100% fill-factor. For samples with a lower fill factor, the sensor functioned as two single wavelength detectors. In the future design we plan to remove this limitation using custom designed optics with focus control, if necessary, to better match the beam shapes to one another. A picture of the spectrometer used in the current DEWRRED sensor along with the input fiber bundle are shown in Figure 2. The mechanical design of the spectrometer housing was not optimized for size and weight in this phase of the project. The optics (not including the CCD) can potentially be housed in a size of 10 x 12 x 4 cm 3. The spectrometer has an unoptimized weight of 6.6 lbs. This design covers a spectral range from 241 nm to 272 nm and is based on gratings and fused silica optics. The design resolution of the spectrometer varied between 29 cm -1 to 34 cm -1 in the

usable Raman spectral regions. Spectral regions near both the laser wavelengths were blocked by Rayleigh filters in the collection system. The remainder of the range was sufficient to cover the fingerprint regions of the Raman spectra for both the excitation wavelengths. All optics, including the gratings in the spectrometer, were built based on a custom design. Although the spectrometer showed low scattering and could be operated in the presence of the room lights, a short-wavelength-pass filter with an edge around 300 nm was used to reject any scattering from laser induced fluorescence arising from sample background. This filter would also allow the system to operate in sunlight. Combined throughput of all the optical components in the spectrometer, and the input fiber (not including the laser rejection and LIF filter) is greater than 55%. The overall efficiency of the entire spectrometer including quantum efficiency of the CCD is around 25%. 3. RESULTS Typical detection results with the software used to control the DEWRRED sensor are shown in Figure 3. Detection of potassium chlorate (PC) with 1 second exposure time is shown in the Figure. The sample was located at a distance of 1 m from the collection mirror and the laser average powers used in the experiment were 26.5µw and 10.2µw at 236.5 nm and 257.5 nm respectively. PC was detectable even at the lowest exposures of 0.1 sec using these power levels. (a) 257.5 nm 236.5 nm (b) Figure 3: (a) Dual-excitation-wavelength Raman spectra of potassium chlorate. The data on the left is from 236.5 nm excitation and on the right is from 257.5 nm excitation. The x-axis was calibrated in wave numbers relative to each laser wavelength. (b) Scores obtained from the DEWRRED sensor algorithm.

3.1 Bulk detection at 3.3 feet and 14 feet We created a target library using high signal-to-noise (SNR) spectra measured from 3.3 ft distance. The target library included Potassium Chlorate, Sodium Chlorate, Potassium Perchlorate, Ammonium Perchlorate, Sodium Nitrate, Urea Nitrate (UN), Ammonium Nitrate, Ammonium Nitrate in Fuel Oil (ANFO), Dyno AP, Watergel, PETN, Semtex, TNT, and Black Powder. Bulk materials used in the detection were uniform samples that were > 10mg in weight. All pure bulk explosive materials were positively detected with the sensor by using a novel algorithm developed for the sensor. The novel algorithm is based on a combination of a) correlation information between target and library spectra and b) ratiometric information between the two excitation wavelength signatures. The algorithm produces a score by comparing the test spectrum with each target library spectrum. If the scores are below a certain threshold they are discarded and assumed to have no explosive signature. The scores indicate fractional resemblance of the measured spectrum to target library spectra. If multiple scores from closely related groups of explosives are above the threshold, then the material is identified as the library material with the highest score. For example in Figure 3b, Chlorates and Perchlorates have scores above the threshold and the highest scoring target library material is PC. Therefore, the sample was identified as PC. A key advantage of the DEW technique is improved specificity at lower signal levels. When the algorithm operated on Raman spectra from any one of the two excitation wavelengths alone, it produced high scores for a group of materials with closely resembling spectra (e.g. chlorates/perchlorates or nitrates), and they could not be separated. However with the DEW technique the algorithm was able to distinguish different materials with closely resembling spectra even at SNR=4. Several background materials and interferant materials were also tested and they typically produced scores below the threshold. Examples of these substances included dry wall, cardboard, various colored plastics, dirt, concrete, asphalt, sugar, salt, creams, cleaning agents etc. We also tested and successfully detected explosives on top of a few representative background materials (tiles, cardboard and dirt). The DEWRRED system was also tested with samples located at 14 ft distance. Laser power used for these experiments was around 200-250 w for each excitation wavelength. The spot sizes at this distance were approximately 2.5 mm. We were able to successfully detect explosive samples and samples on top of background surfaces. 3.2 Detection of small quantities of explosives In order to check the performance of the DEWRRED system in detecting small quantities of explosives, we prepared two samples using PC and UN and measured them from a 3.3 ft distance. Examples of samples used in the experiments and scores are shown in Figure 4. The PC sample was prepared by dissolving PC in water. A single droplet of water was placed on a black aluminium foil and dried. A second sample was prepared with a UN fingerprint on aluminum foil and gently wiped-off with a lens tissue. We estimate the sample quantities exposed to the laser excitation to be less than 50 g. Because our two laser beams had slightly different sizes on the sample, a limitation with our current setup, we could not take advantage of the ratio variations to achive higher specificity. Therefore, the PC and UN samples were identified as chlorates/perchlorates and nitrates respectively. Exposure times used in the experiments were 1 to 5 s for PC and 20 to 30 s for UN. The limitation from beam overlap will be addressed in our future design. PC UN Figure 4: Example of small quantity (<50 g) of explosives or precursors detected from 3.3 ft distance.

3.3 Detection at very low SNR levels The algorithm developed for the sensor showed high sensitivity even at low signal-to-noise ratios. Figure 5 illustrates the performance of the algorithm. Measurements at two different exposure times (1.0 s and 0.1 s) are shown in the chart. Even when the SNR was close to 1, for the measurement shown in Figure 5(b), high scores were obtained for Ammonium perchlorate. We analyzed 2000 measurements of the sample and background with the algorithm and studied ROC statistics (more details in [8]). The ROC statistics clearly showed that Ammonium Perchlorate, measured at low SNR of 1.6, could be separated from the background. For the same true-positive-rates, significantly lower false-positiverates were seen with DEW technique as opposed to the two single-wavelength measurements, suggesting that DEW technique is advantageous over single-excitation-wavelength techniques. (a) (c) score = 1.98 SNR = 11.7 score = 0.2 (b) (d) score = 1.6 SNR = 1.6 score = 0.8 Figure 5: Illustration of algorithm sensitivity. (a), (b) Bulk Ammonium Perchlorate measured at exposure levels of 1 sec and 0.1 sec respectively. The algorithm produced a high score of 1.98 and 1.6 respectively. (c), (d) Background noise from a black Aluminium substrate obtained with the same exposure times (1s and 0.1s). The algorithm produced a low score of 0.2 and 0.8 respectively. Sample to collection mirror distance was 1m and the laser power used was ~20µw at each wavelength. 4. SUMMARY We have built a laboratory prototype of a dual-excitation-wavelength resonance Raman detector. The sensor consists of two CW DUV lasers at 236.5 nm and 257.5 nm, a dual-band spectrometer covering 241 nm to 272 nm and a very sensitive algorithm. The sensor was tested by measuring samples from a distance of 3.3 ft and 14 ft. Test materials include fourteen explosive/precursor materials, several background materials and interferents. Bulk explosives were

correctly detected and identified using the DEW signatures. The sensor was also able to detect groups of explosive/precursors when presented in small quantities of ~50 g. The algorithm developed for the sensor showed high sensitivity and was able to separate target materials from noise even at a low SNR of 1.6. 5. ACKNOWLEDGEMENT This work was funded by the Office of Naval Research under contract number N00014-12-C-0069. REFERENCES [1] Yellampalle, B., Sluch, M., Wu, H. S., Martin, R., McCormick, W., Ice, R., and Lemoff, B. E., " Dual-excitation wavelength resonance Raman explosives detector, " Proc. SPIE 8710, 87100Z-1-87100Z-8 (2013). [2] Asher, S. A., UV RR, in Handbook of Vibrational Spectroscopy, J.M. Chalmers and P.R. Griffiths (Eds), John Wiley & Sons, Ltd., Vol. 1. 557-571 (2001). [3] Asher, S. UV Raman Spectroscopy for Analytical, Physical, and Biophysical Chemistry, Part 1, Analytical Chemistry, 65, 59 66 (1993). [4] Asher, S. In Ultraviolet Raman Spectrometry, Chalmers, J., and Griffiths, P., Eds., John Wiley & Sons, Vol. 1, pp 557 571 (2002) [5] Yellampalle, B., Sluch, M., Asher, S., Lemoff B., Multiple-excitation-wavelength resonance-raman explosives detection, Proc. of SPIE Vol. 8018, 801819-1 - 801819-7 (2011) [6] Asher, S., Johnson, C. Raman Spectroscopy of a Coal Liquid Shows that Fluorescence Interference is Minimized with Ultraviolet Excitation, Science, 225, 311 (1984). [7] Yellampalle, B., and Lemoff, B. E., " Raman albedo and deep-uv resonance Raman signatures of explosives," Proc. SPIE 8734, 87340G-1-87340G-10 (2013). [8] Yellampalle, B., Wu, H. S., McCormick, W., Sluch, M., Martin, R., Ice, R., and Lemoff, B. E., "Sensitive Algorithm for Multiple-Excitation-Wavelength Resonance Raman Spectroscopy," Proc. SPIE 9101, 9101-26 (2014).