Clinical Use Only. ResMed HumidAire 2i. Disinfection and Sterilisation Guide. English

Similar documents
H5i disinfection guide

heated humidifier H5i Welcome Guide English

VPAP Adapt. H5i. Welcome Guide. English POSITIVE AIRWAY PRESSURE DEVICE HEATED HUMIDIFIER

H4i HEATED HUMIDIFIER

User guide User guide English

User guide User guide English

AutoSet AutoSet for Her Elite Escape Auto Escape. H5i H5i for Her. Welcome Guide. English. POSITIVE AIRWAY PRESSURE DEVICEs HEATED HUMIDIFIER

H4i. User Guide HEATED HUMIDIFIER. English

AutoSet CS POSITIVE AIRWAY PRESSURE DEVICE. Welcome Guide. English

VPAP Auto VPAP ST VPAP S. H5i. Welcome Guide. English POSITIVE AIRWAY PRESSURE DEVICE HEATED HUMIDIFIER

VPAP ST with ivaps. Welcome Guide. Making quality of care easy. English. Respiratory Care Solutions NONINVASIVE VENTILATOR

VPAP Adapt. H5i. Welcome Guide. English. adaptive servo-ventilator

VPAP COPD. H5i. Welcome Guide. Making quality of care easy. Respiratory Care Solutions. English NONINVASIVE VENTILATOR HEATED HUMIDIFIER

S9 Escape. Welcome Guide. English POSITIVE AIRWAY PRESSURE DEVICE

S9 AutoSet & S9 Elite

AutoSet & Elite. Welcome Guide / S9 Series. AutoSet & Elite WELCOME EUR2 / APAC. English Svenska Norsk Suomi Dansk Eesti 简体中文 繁體中文

Clinical guide. English

Quick setup guide. English

Disconnection Alarm Guide. English

HANDS-ON MODULE 5. CLEANING FLOW-i V3.0

Adult and paediatric ventilators. Quick Setup Guide. English

Waterpik Water Flosser Model WF-06

User manual. DreamStation Heated Humidifier

I.F.U Re/Processing Reusable Medical Devices

Waterpik Water Flosser Model WP-150/152

AMSCO STERILIZATION CONTAINER SYSTEM User s Guide

Rain Forest Quick Guide

Waterpik Water Flosser Model WP-450/462/ 463/465/470

System One Heated Humidifier USER MANUAL

OPTIMYST WALL ENGINE OMWFC20 RTOPW20 PEM20

KNX The worldwide STANDARD for home and building control. KNX Association International

O-PEP Hygiene-Beileger

AERO 20 AERO ENGLISH - Operating Instructions

Installation Instructions

MOCCA CUBIC AUS/NZ. 08/53066/0 Issue 2

AS Medizintechnik GmbH Sattlerstrasse 15, Tuttlingen, Germany Tel 07461/ Fax 07461/

Contact Details. Please note that some of the contact details on this PDF document may not be current.

CLEANING INSTRUCTIONS

Rigid Reusable & Single use Sigmoidoscopes, Anoscopes, and Accessories ENGLISH 1. Perform reprocessing immediately after each use.

KNX The worldwide STANDARD for home and building control. KNX Association International

Guidelines for Cleaning, Disinfecting and Sterilizing Medela Healthcare Products REPROCESSING GUIDE. Precious life Progressive care

FLASH PAK STERILIZATION CONTAINER SYSTEM

Clinical guide. English

Accuracy counts! Sensors with analog output

Waterpik Water Flosser Model WF-02

Leading manufacturer of high-end ball valves and pneumatic actuators. Over 60 years experience with industrial applications

Operator s Manual. Video Laryngoscope. ( Instructions for use ) Doc no: Revision 1.1 8th January 2013

CRA24-B3 controlled residential ventilation

DANVILLE DNV20AB, DNV20BR, DNV20BL & DNV20CH

Processing (cleaning, disinfection and sterilization) of instruments from Smith & Nephew Orthopaedics AG

High vacuum guide FOR PRE-SEPARATORS AND CYCLONES

International standards used locally worldwide ISO/TC 67

Quick start guide. English

International Standards Health Informatics. Audrey Dickerson, RN, MS ISO/TC 215 Secretary

OPERATING INSTRUCTION MANUAL

WASH FILTERS REGISTER YOUR FREE 5 YEAR GUARANTEE TODAY OPERATING MANUAL. click. click. Wash filters with cold water at least every month.

Future development of the furniture industry A look at the next 3-5 years

DIRECTIONS FOR USE. Keeping you insafehands to avoid needlestick injuries. see for demo video

2 in 1 steam and blend

PRELUDE. User Manual. PA3500 Mattress Overlay System. Guarantees and Warranty. 1.1 General Information 1.2 Intended Use. 3.1 Unpacking 3.

Press Release. Short profile Beckhoff. Short profile February 2009 Page 1 of 6

CPAPSolutions. Product Profile. Overview: The Family of SleepStyle TM CPAP Solutions

30th. English. Anniversary Since 1980 RICH. Heat & Cool molding system RICH-2. ZICUL (Chiller) WATER PURIFIER

WASH FILTERS REGISTER YOUR FREE 5 YEAR GUARANTEE TODAY OPERATING MANUAL. click. click. click. Wash filters with cold water at least every month.

AutoM Click to go to the page. AutoM Auto Shaker Filter

BUILD the clinic you ve always imagined.

AS Medizintechnik GmbH Sattlerstrasse 15, Tuttlingen, Germany Tel 07461/ Fax 07461/

The advantages of closed heating systems. White paper

المملكة العربية السعودية وزارة الصحة اإلدارة العامة لمكافحة عدوى المنشئات الصحية

I N S T R U C T I O N S

Corning LSE 6L Digital Water Bath

JUNEAU JUN. 08/51193/0 Issue 0

Instructions For Use. Hybernite ROC 2300 Rainout Control System

Instructions for set-up, use and cleaning

Floriculture. A world wide business and international challenges.??-??-2012, Nico de Groot

Processing Instructions. Cleaning so easy.

INSTRUCTIONS FOR CLEANING AND STERILIZATION OF ALL B.BRAUN AESCULAP HAND HELD SURGICAL INSTRUMENTS AND POWER PRODUCTS (UNITED KINGDOM REQUIREMENTS)

Advantages of automatic boiler bottom blowdown

AS Medizintechnik GmbH Sattlerstrasse 15, Tuttlingen, Germany Tel 07461/ Fax 07461/

Getinge WD14 Tablo Tabletop washer-disinfector

π H-2268 SANITAIRE UPRIGHT VACUUM SAFETY uline.com

Maestro Glass Kettle. User Guide KE9750

OAKHURST, AUBERRY OKT20, AUB20

3KW concealed element. Instructions for Use PLEASE RETAIN THESE INSTRUCTIONSFOR FUTURE REFERENCE.

MUNTERS SERVICE GUIDE. Service, maintenance & upgrades. Your Trusted Global Service Partner

Prolystica Ultra Concentrate Instrument Cleaning Chemistries Now less is more. NEW generation instrument cleaners

WASH FILTER REGISTER YOUR FREE 2 YEAR GUARANTEE TODAY OPERATING MANUAL ASSEMBLY. click. click. Wash filter with cold water at least every month.

Type 2 safety light curtain Type 2 according to the IEC/EN parts 1 and 2 Compact profile: 31 x 32 mm 30 mm resolution for hand detection Wide r

Cleaning and Sterilisation for Burs, accessories and filling material

A WMIA Partner of the Year Winner S-Series Enclosureless Dust Collector


Quick start guide. BiPAP AVAPS

Operator s Guide. Autoclavable Internal Handles with Integrated Paddles Rev. M

Bipap A30/40 ventilator and System One humidifier

Cleaning & Care. Washing Breast Pump Parts

Expansion vessel in a solar installation. White paper

product safety Please read these instructions before operating the appliance and retain them for future use.

Installation & Operating Instructions

BFD Light Curtains The new choice for cutting edge safety for industrial doors

Compact FREME. flash recovery energy management equipment HEAT TRANSFER SOLUTIONS

Transcription:

ResMed HumidAire 2i and Sterilisation Guide English 1

ResMed HumidAire 2i and Sterilisation Guide This and Sterilisation Guide is intended for multi-patient use of the HumidAire 2i in a sleep lab, clinic or hospital. If you use the water chamber as a single user in the home, refer to the User Guide for cleaning instructions. The following are the recommended and validated procedures for cleaning, disinfection and sterilisation of the water chamber. We advise that each healthcare facility consult its own procedures before carrying out the following instructions. WARNINGS and CAUTIONS ResMed cannot give any assurance that deviations from the procedures listed in this guide, and their effect on the performance of the product, will be acceptable. When using detergents, disinfectants or sterilisation agents, always follow the manufacturer s instructions. Note: The HumidAire 2i is available in both a clear polycarbonate (PC) version as well as a smokey grey polysulphone (UDEL). See the table below to determine which cleaning and disinfection/sterilisation method is appropriate for your HumidAire 2i. Clear polycarbonate (PC) version Smokey grey polysulphone (UDEL) version Type Option 1 Manual Cleaning and High-Level Thermal (see Table 1) Option 2 Automatic Cleaning and High-Level Thermal (see Table 2) Option 3 Manual Cleaning and High-Level Chemical (see Table 3) Option 4 Manual Cleaning and Steam Sterilisation (see Table 4) Clear polycarbonate (PC) Yes Yes Yes No Smokey grey polysulphone (UDEL) No No No Yes Symbols Which Appear On The Device 132 o C Sterilisable up to 132 C (UDEL tub only) 2

ResMed Validated Procedures Clinical Use Only Table 1: Option 1 (Manual Cleaning and High-Level Thermal ) Option 1 (Manual Cleaning and High-Level Thermal ) Disassembly Pre-cleaning Disassemble the humidifier water chamber according to the Disassembling and Reassembling Rinse all water chamber components 10 times and brush thoroughly with a soft bristle brush in a warm (45 C) solution of neodisher MediZym (2%). Note: Pre-cleaning must occur immediately after use. Cleaning 1. Clean all components of the water chamber with a soft bristle brush for one minute while soaking them in the detergent 1 solution. Pay particular attention to all crevices and cavities. 2. Thoroughly rinse each component according to the manufacturer s instructions. (high-level thermal disinfection) Inspection Reassembly Packaging & Storage Allow the water chamber to air dry out of direct sunlight. 1. Using a certified hot water disinfection system, fully immerse each component in water using one of the temperature-time combinations listed below. Take care that no bubbles are trapped against the components or inside the tubing. Based on pren ISO 15883-1, the water chamber can be thermally disinfected at the following temperature-time combinations 2 : 70 C for 100 minutes 75 C for 30 minutes 80 C for 10 minutes 90 C for 1 minute. 2. On completion, remove the water chamber from the hot water disinfection system. 3. Thoroughly rinse each component in drinking quality water (five litres per assembly). Air dry out of direct sunlight. Perform a visual inspection of the water chamber. If any visible deterioration is apparent (cracking, crazing, tears etc), the water chamber should be discarded and replaced. Slight discolouration of the silicone components may occur and is acceptable. Reassemble the humidifier water chamber according to the Disassembling and Reassembling Store in a dry, dust-free environment away from direct sunlight. Storage temperature: -20 C to 60 C. 1 Cleaning: ResMed has tested the detergent neodisher MediZym (diluted at 2%) for 20 cycles using warm water (approximately 45 C) according to the manufacturer s instructions. 2 : ResMed water chambers have been tested for disinfection for 20 cycles by fully immersing each component in hot water (approximately 90 C) for 1 minute. 3

Table 2: Option 2 (Automatic Cleaning and High-Level Thermal ) Option 2 (Automatic Cleaning and High-Level Thermal ) Disassembly Pre-cleaning Disassemble the humidifier water chamber according to the Disassembling and Reassembling Rinse all water chamber components 10 times and brush thoroughly with a soft bristle brush in a warm (45 C) solution of neodisher MediZym (2%). Note: Pre-cleaning must occur immediately after use. Cleaning 1. Support the components in the standard racking of the automatic washer-disinfector so that each will drain water. 2. Add the detergent 1 solution and activate the automatic washer-disinfector for a standard cycle of cleaning, followed by thorough rinsing. (high-level thermal disinfection) Inspection Reassembly Packaging & Storage 1. Activate the automatic washer-disinfector for a standard cycle of thermal disinfection 2. 2. Ensure that the machine has thoroughly rinsed each component. Air dry out of direct sunlight. Perform a visual inspection of the water chamber. If any visible deterioration is apparent (cracking, crazing, tears etc), the water chamber should be discarded and replaced. Slight discolouration of the silicone components may occur and is acceptable. Reassemble the humidifier water chamber according to the Disassembling and Reassembling Store in a dry, dust-free environment away from direct sunlight. Storage temperature: -20 C to 60 C. 1 Cleaning: ResMed has tested the detergent neodisher MediZym (diluted at 0.5%) for 20 cycles using warm water (approximately 45 C) according to the manufacturer s instructions. 2 : Due to specific regional requirements, ResMed water chambers have been tested for disinfection for 20 cycles using hot water (approximately 93 C) for 10 minutes. 4

Table 3: Option 3 (Manual Cleaning and High-Level Chemical ) Option 3 (Manual Cleaning and High-Level Chemical ) Disassembly Pre-cleaning Disassemble the humidifier water chamber according to the Disassembling and Reassembling Rinse all water chamber components 10 times and brush thoroughly with a soft bristle brush in a warm (45 C) solution of Alconox (1%). Note: Pre-cleaning must occur immediately after use. Cleaning 1. Clean all components of the water chamber with a soft bristle brush for one minute while soaking them in the detergent 1 solution. Pay particular attention to all crevices and cavities. 2. Thoroughly rinse each component according to the manufacturer s instructions. (high-level chemical disinfection) Inspection Reassembly Packaging & Storage Allow the water chamber to air dry out of direct sunlight. 1. Soak the disassembled water chamber components in a commercially available solution of ortho-phthalaldehyde 0.55% (eg, CIDEX OPA) 2. Take care that no air bubbles are trapped against the components. 2. Thoroughly rinse 3 each component according to the manufacturer s instructions. Air dry out of direct sunlight. Perform a visual inspection of the water chamber. If any visible deterioration is apparent (cracking, crazing, tears etc), the water chamber should be discarded and replaced. Slight discolouration of the silicone components may occur and is acceptable. Reassemble the humidifier water chamber according to the Disassembling and Reassembling Store in a dry, dust-free environment away from direct sunlight. Storage temperature: -20 C to 60 C. 1 Cleaning: ResMed has tested the detergent Alconox (diluted at 1%) for 20 cycles using hot water (approximately 60 C) or warm water (approximately 50 C) according to the manufacturer s instructions. 2 : ResMed water chambers have been validated for 20 cycles using solutions of ortho-phthalaldehyde (0.55% CIDEX OPA for 12 minutes) according to the manufacturer s instructions. 3 Rinsing: Thoroughly rinse each component in drinking quality water (five litres per assembly) by immersing it completely for a minimum of one minute in duration. Repeat the rinse procedure two additional times using fresh water for a total of three rinses. 5

Table 4: Option 4 (Automatic Cleaning and Steam Sterilisation) Option 4 (Automatic Cleaning and Steam Sterilisation) Disassembly Pre-cleaning Cleaning 1 Sterilisation 2 Inspection Reassembly Packaging & Storage Disassemble the humidifier water chamber according to the Disassembling and Reassembling Rinse all water chamber components 10 times and brush thoroughly with a soft bristle brush in a warm (45 C) solution of neodisher MediZym (2%). Note: Pre-cleaning must occur immediately after use. 1. Support the components in the standard racking of the automatic washer-disinfector so that each will drain water. 2. Add the detergent solution and activate the automatic washer-disinfector for a standard cycle of cleaning, followed by thorough rinsing. Using a certified sterilisation system, use one of the following temperature profiles: 132 C with a three-minute hold time 121 C with a 30-minute hold time. is achieved as part of the sterilisation process. Perform a visual inspection of the water chamber. If any visible deterioration is apparent (cracking, crazing, tears etc), the water chamber should be discarded and replaced. Slight discolouration of the silicone components may occur and is acceptable. Reassemble the humidifier water chamber according to the Disassembling and Reassembling Store in a dry, dust-free environment away from direct sunlight. Storage temperature: -20 C to 60 C. 1 Cleaning: ResMed has tested the detergent neodisher MediZym (diluted at 0.5%) for 20 cycles using warm water (approximately 45 C) in the washer-disinfector (manufacturer: Miele; type: G7736MCU; program: Vario TD210) according to the manufacturer s instructions. 2 Sterilisation: ResMed water chambers have been tested for 20 cycles using the temperature-time combinations: 132ºC with a threeminute hold time or 121ºC with a 30-minute hold time. 6

Disassembling and Reassembling the HumidAire 2i WARNING Beware of electrocution. Do not immerse the device or power cord in water. Always unplug the device before cleaning and be sure that it is dry before plugging back in. 1 Turn the main power switch at the back of the device to OFF, and unplug the power cord from the power outlet. 2 Allow 10 minutes for the tub plate and any excess water to cool. 3 Remove the water chamber. To do this, lift the latch at the bottom of the water chamber and pull the water chamber away from the docking station. 4 Discard any excess water by turning the water chamber upside down, allowing the water to drain out of the air outlet. 5 Disassembling the Water Chamber 1. To open the side catches pull out from the lower edge. Side catch 2. Separate the water chamber by pulling the two halves apart. Use your thumb to pull down on the front tab. Front tab 3. Inside the chamber lid is the seal. Remove the seal by peeling it away from the inside of the lid. Lid Seal 4. Remove the cross-member from the water chamber base. Cross-member Base 7

6 1. Push down on the tub plate from inside the chamber to remove it from the chamber. Tub plate 2. Remove the O-ring from the tub plate. O-ring 3. Turn the water chamber base upside down. Firmly pull the latch up and away from the chamber until separated from the base. Underside view 4. Disengage the catches from the sides of the water chamber base. Then, separate the inner and outer moulding of the catch, pull back the inner moulding and squeeze. The inner moulding will pop out. 7 Reassembling the Water Chamber When all the water chamber parts are clean and dry, reassemble the water chamber: 1. Reassemble the catches by inserting one side of the inner moulding to the outer moulding and pushing downwards. 8

2. Take the assembled catches and fit them to the water chamber base. 3. Turn the water chamber base upside down. Insert the latch s feet into the locking tabs. Push firmly and click into place. 4. Refit the O-ring in the channel of the tub plate. CAUTION Ensure that you replace the O-ring if it is nicked or punctured. 5. Push the tub plate into the water chamber with the open side facing down. Then firmly press the tub plate down into the water chamber. Tub plate 6. Place the cross-member on its supports inside the water chamber base. 9

7. Fit the seal onto the rim of the water chamber base. Press along the edge to ensure that the seal is fitted properly. Seal Rim 8. Fit the lid onto the water chamber base. Press the lid and base firmly together to close. Lid 9. Lock the catches on the side of the water chamber. Troubleshooting If the water chamber leaks after being reassembled, examine the unit for the following: Possible cause Water chamber may be damaged or cracked Water chamber seal may be twisted Side catches are not engaged properly Tub plate is incorrectly fitted O-ring is punctured or nicked Solution Replace affected section of water chamber. Make sure the seal is seated properly inside the water chamber. Make sure two side catches are engaged properly. Ensure tub plate is fitted correctly. Replace O-ring. ResMed Ltd (Manufacturer) 1 Elizabeth Macarthur Drive Bella Vista NSW 2153 Australia, ResMed Corp (US Designated Agent) 14040 Danielson Street Poway CA 92064-6857 USA, ResMed (UK) Ltd (EU Authorized Representative) 96 Milton Park Abingdon Oxfordshire OX14 4RY UK, ResMed Offices Australia, Austria, Belgium, Brazil, China, Finland, France, Germany, Greece, Hong Kong, India, Ireland, Italy, Japan, Malaysia, Mexico, Netherlands, New Zealand, Norway, Portugal, Singapore, South Africa, Spain, Sweden, Switzerland, UK, USA (see www.resmed.com for contact details). Patents pending. Protected by design registrations: AU 147335, CH 128.711, CH 128.712, DE 40202008, DE 40201723, ES 153516, ES 153518, FR 021407, GB 3001820, GB 3001821, JP 1164266, JP 1164267, SE 75598, SE 75715, US D476077, US D477868. Other designs pending. HumidAire is a trademark of ResMed Ltd and is registered in the U.S. Patent and Trademark Office. CIDEX is a trademark of Johnson & Johnson. Alconox is a trademark of Alconox Inc. neodisher MediZym is a trademark of Chemische Fabrik Dr Weigert. 2008 ResMed Ltd. 308237-Eng/1 08 07 10 Global leaders in sleep and respiratory medicine www.resmed.com