FagronLab NP-1 UV. Negative pressure workbench. User Manual

Similar documents
FagronLab RVC1. Laboratory Robot Vacuum Cleaner. User Manual

STATIC ELIMINATOR I443A INSTRUCTION MANUAL

OPERATION INSTRUCTION

MS-H-Pro-T Magnetic Hotplate-Stirrer with Timer

INSTRUCTION MANUAL. Stirring hotplate (with probe and stand) Brand : (Dragon laboratory)

Requests for Daily Care and Maintenance

Biological Safety Cabinet

Requests for Daily Care and Maintenance

USER MANUAL. Content. XD W/250W Medical Cold Light source. ⅠBrief introduction...1

MB-T Moisture analyzers

Medical Refrigerator

Clean Air Optima HEPA UV ionizer air purifier CA-508

Owners Manual Model: XXL

JT-SB116JH-G-NA JT-SB116JH-W-NA

Coway Air Purifier AP-1008DH

Note: Verify that plastic bags are removed from filters, if equipped or installed, before first use (see page 8) CONTENTS

WCD-210BK. Water Dispenser OWNERS MANUAL. Read and save these instructions.

Class II Biosafety Cabinets

WCD-200W Water Dispenser

2500 Electric Blower Vac

Requests for Daily Care and Maintenance

INSTRUCTIONS AND WARRANTY

Ultrasonic Aroma Diffuser AD-SD1-CE

IAQ Series. Bosch IAQ Photo Catalytic Oxidizer (PCO) Residential Application. Installation Manual and Owner s Guide

Please read this manual carefully before use to ensure proper usage and save

Electric Hand dryer. Installation and operating manual

WHYNTER Eco Pure Air Purifier

Instruction Manual for the Revolutionary Science RS-IF-203 Incufridge

AC Portable Air Conditioner AC-14100E, AC-14100H OWNERS MANUAL. Read and save these instructions.

Hand Dryer Installation & User Manual

MODEL 7000 SUCTION UNIT

Biological Safety Cabinets or Fume Hoods Which one to use?

Requests for Care and Maintenance

TAIJI HOTCABI Towel Warmer

CLEAN AIR SYSTEM.

AIR CONDITIONER OWNER S MANUAL. Please read this manual carefully before operating your set and retain it for future reference.

MIMAKI ENGINEERING CO., LTD.

INCREASE YOUR LAB EFFICIENCY

Operator s Manual. Model G17-EU (International) Disinfection Soak Station for Hysteroscopes, Cystoscopes and ENT scopes

Laboratory Equipment LAMINAR FLOW CABINETS ENVAIR ECO AIR H ISO CLASS 3 HORIZONTAL PRODUCT PROTECTION CABINET

CYC LONE/THUNDER AR1320/AC1320

TPP440. owner s guide. air purifier with uv light. air purifier with uv light TP

Digital 7 Square Plate Magnetic Stirrer

Cooker Hood LA-72-CAN.

Revolutionary Science RS-IF-202

Intelligent Lead Free Soldering Station OPERATION and MAINTENANCE MANUAL

Ionizer (Static Remover) STABLO-EX Instruction Manual

USER INSTRUCTION MANUAL SAPPHIRE SERIES 1100 & 1100EC

Model No. NR-BU303 Series

Operator s Manual. Model G17 Disinfection Soak Station for Hysteroscopes, Cystoscopes and ENT scopes

UL U TR LTRASONIC S ONIC SCALE ALER PIEZO MINI

LOCK H M L REPLACE SLEEPING ANTI-POLLEN IONIZER UV-LAMP. User Manual

Envair Lab OSN Eco Air H. Horizontal Laminar Flow Product Protection Hoods

Instruction Manual for the Revolutionary Science RS-PB-200 Dual Poly Pro Bath

AdvancedPureAir.com

Handling or using the product improperly and in disregard of the instructions with this mark might result in serious bodily injury or death.

DIGITAL Warm & Cool Mist Humidifier

Portable Space Ceramic Heater

Thermal Electric Wine & Beverage Chiller Model: WC-2463W. Instruction Manual

INSTALLATION AND MAINTENANCE GUIDE TRENDY 1000 T ELECTRONIC LAVATORY FAUCET

AIR CONDITIONER OWNER S MANUAL. Please read this manual carefully before operating your set and retain it for future reference.

HEA-900 EDENPURE INDOOR/OUTDOOR ALL SEASON INFRARED HEATER

General Purpose Laminar Flow Cabinets, Horizontal and Vertical. Horizontal Flow Vertical Flow

WARM & COOL MIST Ultrasonic Cool Mist Humidifier

BEFORE OPERATING, READ ALL INSTRUCTIONS IN THIS MANUAL CAREFULLY.

OPERATIONS MANUAL. Hotplate Stirrer/Hotplate/Magnetic Stirrer Model HTS-1003/HTP-1002/MGS Certified

Instructions. Portable Absorb Phlegm. Apparatus

INSTRUCTION MANUAL MNBX14

User s Manual WS-4830SS WS-4836SS NOTE: PLEASE INSPECT HOOD IMMEDIATELY UPON RECEIVING. CLAIM OF DAMAGE AFTER 7 DAYS OF DELIVERY WILL BE DENIED.

Esco PCR Cabinet, Model PCR-4A_. Shown with optional support stand. Polymerase Chain Reaction Cabinets The Proven Solution for Contaminant-Free PCR

Static Eliminators. Overhead Ionizing Blower Model IN5130. Instruction Manual

model No: ir20.v2 FUSE

Cavitator Ultrasonic Cleaner Instruction Manual Model 20G

BSC CLASS II, TYPE A2 BIOLOGICAL SAFETY CABINET SERIES. Optimum protection of personnel, product and environment. BSC Class

UVT-B-AR. DNA/RNA UV-cleaner box. Operating Manual Certificate. Medical Biological Research & Technologies

INSTRUCTIONS AND WARRANTY

OPERATION AND MAINTENANCE MANUAL

Requests for Care and Maintenance

SECTION BIOLOGICAL SAFETY CABINETS AND LAMINAR AIRFLOW WORK STATIONS

User Manual. MS7-H550-Pro MS7-H550-S MS7-Pro MS7-S

USER MANUAL LEEC CLASSIC INCUBATOR. C100, C157 and C330

IMPORTANT SAFETY INSTRUCTIONS READ AND SAVE THESE SAFETY INSTRUCTIONS BEFORE USING THIS AIR PURIFIER

50-Pound Portable Ice Maker

INSTRUCTION MANUAL HEAVEN FRESH UK

Operating Instructions for the KM 410 Electric Grill

Intelligent Swing Gate Turnstile Operating Manual CPW-322CS

AES Environmental. Operation and maintenance manual. VWS Series. Laminar Vertical Flow clean work station

User Instruction Manual

Instruction Manual for the Revolutionary Science RS-PB-100 Poly Pro Bath TABLE OF CONTENTS

Finished Excellence. Limited Lifetime Finish Warranty. Armaturen.Accessoires.Interiors.Culture Projects.

CIRCULAIRE LAMINAR AIR FLOW

Electric Garage Heater

Request for daily care

Rechargeable UV-A LED/Emergency Light

PLEASE READ BEFORE USE AND SAVE. Ultrasonic. with Ionization. INSTRUCTION MANUAL MODEL HUNi

General Purpose Laminar Flow Cabinets, Horizontal and Vertical

BIG DISPLAY ALARM CLOCK USER MANUAL

CA-707. Manual. Read these instructions carefully before you start using the dehumidifier and keep the manual for later usage.

CRISTAL Built-in Dish Sterilizer Instruction Manual

Humidifying Air Purifier

Transcription:

Negative pressure workbench User Manual

Table of contents 1. Introduction... 3 2. Service... 3 3. Warranty... 3 4. Application range... 4 5. Main technical parameters... 4 6. Performance index... 5 Vibration amplitude... 5 Illumination... 5 Electrical characteristic... 5 7. Structure...5 8. Structural composition...5 Front window driving system... 5 Air filtering system... 6 Ultraviolet source... 6 Lighting source... 6 Control panel... 6 Gear indicator... 7 Touch button... 7 Fuse (protective tube)... 7 9. Installation and operation...7 Installation... 7 Operation guide... 7 10. Maintenance and common malfunction analysis... 8 Working zone surface cleaning... 8 Outside surface and glass window cleaning... 8 Comprehensive maintenance cycle... 8 Daily or weekly maintenance... 8 Monthly maintenance... 8 Annual maintenance... 9 Common failures and solutions... 9 Storage conditions... 10 11. Attention... 10 12. Circuit diagram... 11 13. Content package... 11 14. Air flow pattern and protected areas... 12

1. Introduction Thank you for purchasing the vertical laminar flow cabinet FagronLab NP-1 UV. This vertical laminar flow cabinet is a negative pressure workbench specially designed for use in pharmacies. It can protect people by the negative pressure in the air access and protect environment by the HEPA filter installed on top of the device. Compared with the biological safety cabinet, air segregation equipments do not have the air supply system. But it s more succinct and portable in structurally, you can put it on any position or laboratory furniture, so it could be operated easily and conveniently. You must carefully read and understand the contents of this manual prior to operating this device. After reading this manual, please put it in an appropriate position for easy access. 2. Service In case of any problem, you can always contact the Service Department for technical support. Please provide the customer care representative with the following information: Description of problem. Methods and procedures adopted to resolve the problems. Your contact information. 3. Warranty This device is under warranty and free from defects in materials and workmanship, under normal use and service, for a period of 12 months from the date of invoice (excluding consumable accessories, UV and LED lamp, fuse, filters). The warranty is extended only to the original purchaser. Warranty is not valid on device which has been damaged on account of improper installation, improper connections, misuse, accident or abnormal conditions of operation. If the warranty has been expired, Fagron would still responsible for repair with relative charges. For claims under the warranty please contact your local supplier. FagronLab NP-1 UV User Manual 3

4. Application range Laminar flow cabinet creates a local high cleanliness air environment by horizontal or vertical air flow. It is widely used in preparation process of compounded medication in pharmacies to protect user from inhaling dangerous substances. It is recommended during mixing and capsule filling of such substances. It is very helpful to increase yield, precision, stability and reliability of products. Working Environment: In door use only Environment temperature: 15 C~35 C Relative humidity: 75 % Pressure Range: 70 kpa~106 kpa 5. Main technical parameter Laminar flow cabinet is designed to protect the user and materials, to make local air environment with high degree of cleanliness. The main role is to ensure the accuracy of the operator and the environment. Model / Technique parameters NP-1 UV External size (W*D*H) 550 460 700 mm Working zone Size (W*D*H) 480 340 370 mm Rated voltage AC 220V ± 10 % Rated frequency 50 Hz ± 1 Hz Rated power 100 W Airflow velocity >0.5 m/s UV Lamp 8 W Fluorescent lamp 4 W (LED) Filtering efficiency 99.99 (diameter 0.3 μm) Noise 65 db (A) FagronLab NP-1 UV User Manual 4

6. Performance index Vibration amplitude The net vibration amplitude between 10 Hz and 10 KHz is no more than 5 μm (rms). Illumination Average illumination is not less than 350 lx. Electrical characteristic The voltage increases to 1390V(AC) in 5s and keep for another 5s without breakdown. Ground resistance 0.1 Ω. 7. Structure Cabinet body is built of 1.2mm cold-rolled steel with anti-powder coating. Strong and sturdy. Work table is made up of stainless steel which provides corrosion resistance. Base stand is made of steel with anti-bacteria powder coating. Soft touch type control panel, easy to handle. 8. Structural composition No Description 3 2 1 7 6 1 Control panel 2 LED light 3 Front glass window 4 Door handle 5 Work table 6 Power socket 4 7 UV lamp 5 Figure 1. NP-1 UV laminar flow Front window driving system The front glass window are controlled by constant force spring. Glass window can be placed at any position, make sure in case of power failure, user can still control the glass window and remove the samples or instrument while operating. FagronLab NP-1 UV User Manual 5

Air filtering system Air filtering system is the most important system and ensures the performance of equipment. The system consists of blower and air filter. The function of Air Filtration System is transferring filtered air from work area, ensure the cleanness of environment. Ultraviolet source The entire work zone could be sterilized effectively by the UV lamp located at the top of work zone. Lighting source The laminar flow cabinet is equipped with LED light, which ensures the standard requirement of average illumination is met. Control panel 1 2 4 5 6 7 8 9 3 No Description 1 Gear indicator 2 Increase air velocity 3 Reduced air velocity 4 Fan indicator 5 Fan ON/OFF button 6 Sterilization indicator 7 Ultraviolet ON/OFF Button 8 Lighting indicator 9 Lighting ON/OFF Button FagronLab NP-1 UV User Manual 6

Gear indicator We can understand the clean bench wind speed situation when working by gear. Touch button Main functions of clean bench could be executed by touch-buttons. "LIGHT" "UV" FANS LIGHT" is lighting button, which control the working state. UV is ultraviolet button, which control the working state of the ultraviolet lamp. "FANS" is absorber control button, which control the working state of blower. The absorber airflow increase & decrease button( ): When the absorber is operating, press button to increase the wind speed: press button to reduce the motor speed. Fuse (protective tube) This equipment is equipped with a total power protective tube, located in the box body side next to the power cord outlet, protective tube label corresponding to the corresponding specifications. 9. Installation and operation Installation Remove all packaging components. Check the surface of main body to make sure there is no scratch, deformation or foreign bodies. Carefully check the accessories and material according to packing list in the manual. When carrying the clean bench, top-for-bottom, dumping place and dismantling are forbidden. Laminar flow cabinet should be placed in airflow protected area. User must take strict precautions to against outside various high-speed current of air to interference of working area, such as door, window, fan, compressed air etc. At least 300mm gap must be kept in the side and back side of the Laminar Air Flow for clean operating and inspection. Operation guide Connect the power supply Press the related function key (please refer to the control panel section );Check whether the function keys are consistent with the results of operations, and check the fan if is working normally and wind speed if reach requirements, UV lamp and LED lamp is working normally according to laminar flow cabinet technical parameters. Before doing any preparation, please sterilize the cabinet for more than 30 minutes by UV lamp; For safety of eyes and skin, people should leave room during the UV sterilization, or avoid direct eye or skin contact. UV lamp intensity should be tested regularly. UV lamp life is generally 1600 hours. FagronLab NP-1 UV User Manual 7

Please move the front window at appropriate height, turn on the fan, make sure the compounding preparation should be started after fan working for at least half an hour (only on first preparation). For operating safety, please put testing materials inside the cabinet in advance. After use, please move the front window to the lowest position, and make sure to sterilize the cabinet by UV lamp for 30 minutes before turning it off. 10. Maintenance and common malfunction analysis Preparations before maintenance: remove the items in cabinet. Items needed to be prepared: cotton or towel, concentrated soap liquid, hot water, water, medical alcohol or other disinfectants, etc. Working zone surface cleaning Wipe the entire surface with a soft cotton cloth or towel soaked with concentrated liquid soap, then wipe up the soap with another cotton cloth or towel soaked with clean hot or warm water, and then wipe the surface with a dry cotton cloth or towel rapidly. For the contaminated or dirty work surface, use medical alcohol or other disinfectant. All disinfectant cannot damage 304 stainless steel. Outside surface and glass window cleaning Use soft cotton cloth or towel to wipe the surface with non-abrasive household cleanser. Overall maintenance period We suggest comprehensive maintenance period is one year or 1000 working hours. Daily or weekly maintenance Disinfect and clean the operating area (please see "working zone surface cleaning" section). Clean the external surface and front window (please see "outside surface and glass window cleaning" section). Check the various functions of the cabinet. Record down the maintenance result. Monthly maintenance Outside surface and glass window cleaning (please see "outside surface and glass window cleaning" section). Use towel with 70% rubbing alcohol or 1:100 dilution of household bleach to wipe the working table, the inner face of front window and the inner wall surface of the working area(exclude the top wind grid). Use another towel with sterile water to wipe those area to erase the remain of chlorine. Check the various functions of the cabinet. Record down the maintenance result. FagronLab NP-1 UV User Manual 8

Annual maintenance Check the front window drive unit, and ensure that their tightness is coincident. Check UV lamp and LED light. Apply for overall performance test of the cabinet annually to ensure that the safety meets requirements. User is responsible for testing costs. Record down the maintenance result. Common failures and solutions Please confirm that the power is well connected, the power cord and fuse are in good condition (without any damage). Failure Checking parts Measures Fluorescent lamp fail to work UV lamp fail to work Button fail to work Blower fail to work No electricity in equipment Display fail to work LED stand plug LED stand Circuit Control panel Interlock Lamp holder Lamp tube Ballast Circuit Control panel Control panel Blower Circuit Control panel Power supply Power cord Fuse Potential transformer Control panel Connection wires Display screen Control panel Connect the plug and stand tightly Replace stand Check the circuit Replace the control panel Check the blower and fluorescent lamp turn off or not Connect the tube and lamp holder tightly Replace the lamp tube Replace the ballast Check the circuit Replace the control panel Make sure the power is well connected and the fuse is in good condition Check if the button is broken Make sure the connecting wire is well connected Replace the control panel Replace the blower if it is damaged Check the circuit Replace the control panel Check whether the power supply is well connected Check whether power cord is in good condition Check whether the fuse is damaged Check whether the transformer works normally Replace the control panel Check connection wires if in good connection Check whether the screen is in good condition Replace the control panel FagronLab NP-1 UV User Manual 9

The operations of above electrical parts should be completed by qualified electrician under the safe condition (cut off the power). If the equipment s malfunction is not listed above and operator cannot solve immediately, please contact with our maintenance department. For your safety, do not maintain equipment by yourself. The maintenance of this equipment only can be finished by trained and certificated persons. If you need to order any component, you can contact our technical service department. Please specify the model and serial number of the cabinet. Storage conditions Laminar flow cabinet should be stored in a warehouse with relative humidity no more than 75%, and temperature lower than 40 C. The warehouse should have good ventilation performance,without acid, alkali or other corrosive gases. Storage period shall not exceed one year, laminar flow cabinet stored for more than one year needs to be unpacked and checked before using. 11. Attention Make sure input voltage is correct and stable. The rated load of main power socket should be higher than cabinet consumption. Plug must be well grounded. This plug with a third foot, can only be used with grounded power outlet, which is a safety device. If socket can not plug in, please contact with your install a grounding type power socket by a electrician. Moving principles of different samples inside cabinet: When two or more samples need to be moved, be sure that lowpolluting samples move to high-polluting samples. Movement of items should also follow the principles moving slowly and stably. The weight of items placed in the cabinet should not be more than 23Kg/25 25cm 2. AVOID VIBRATION: avoid using vibration equipment (eg centrifuges, vortex oscillator, etc.) inside the cabinet. Vibration would cause lower cleanliness of operating area. NO FLAME: No flame is allowed inside the cabinet. Using of fire will lead to airflow disorder, and filter damage. If sterilization is required during the experiment, infrared sterilizer is highly recommended. HEPA filter life: With the usage time increasing, dust and bacteria accumulate inside HEPA filter. Filter Resistance is getting bigger, when it reaches the maximum point, the speed requirements can t be met. Please contact our technical department in order to replace it. The used filter should be processed as medical waste. The fan and the bottom steel plate is static pressure cover, which is sealed strictly in the factory. The operator is not allowed to remove or loose screws of those parts. If necessary, please contact service personal. The maximum storage period is one year. If the period is more than one year, performance test should be done. FagronLab NP-1 UV User Manual 10

12. Circuit diagram 13. Content package Name Quantity Main body 1 RVV Power line 1 Fuse (5A) 1 Manual 1 Test report 1 Quality certificate 1 UV Lamp 2 LED Lamp 1 FagronLab NP-1 UV User Manual 11

14. Air flow pattern and protected areas Room air Polluted air HEPA filtered air A B Protected area Contaminated area FagronLab NP-1 UV User Manual 12

fagron.gr Fagron Hellas - 12 km N.R. Trikala-Larissa - P.C. 42100 Trikala Tel. +30 (0) 24310 83633-5 - Fax +30 (0) 24310 83615 - info@fagron.gr NP1UM01