Service Manual Pharmacy Refrigerator(2~8 ) Model:HYC-940/940F Picture:
Content 1. Definition of the Model... 3 2. Appearance and Parts... 3 3. Product Technical Parameters... 5 4. Electrical parts and circuit diagram... 8 5. Spare parts of cooling system... 9 6. Spare part list and exploded view... 11 7. Instruction of operation... 15 8. Troubleshooting... 19 2
1. Definition of the Model HYC-940&940F Service Manual H(Haier)Y(Pharmacy)C(Refrigerator)-940(Capacity is 890L) H(Haier)Y(Pharmacy)C(Refrigerator)-940F(capacity is 890L,foaming door) 2. Appearance and Parts HYC-940 3
HYC-940F 4
3. Product Technical Parameters 1.HYC-940(220-240V~/50Hz): Model HYC-940 Type upright glass door air-cooling Climate class N(10 ~32 ) Temp. range 2~8 Capacity (L) 890L Interior 1030x590x1425 Dimensions(W D H) Exterior 1130x755x1980 Dimensions(W D H) Net /Gross weight(kg) 227/279 Exterior/Interior Material SPCC Power Supply 220-240V~/50Hz Rated Power/Current 850W/4.5A Controller Microprocessor Qty of sensor 5pcs Alarm type Buzzer,light,remote Alarm content High& low temp.,sensor error,power failed,low battery, door Ajar Foot and wheel 2 feet & 4 wheels Test hole 1 Shelf Shelves :12pcs Lamp LED(9W) Door and lock Glass door (with heater)+lock Defrosting Auto defrosting+ Forced defrosting (with heater) Tank mouth heater Yes Safety mode Sensor failed SM 5
2.HYC-940(220-240V~/60Hz): Model HYC-940 Type upright glass door air-cooling Climate class N(10 ~32 ) Temp. range 2~8 Capacity (L) 890L Interior 1030x590x1425 Dimensions(W D H) Exterior 1130x755x1980 Dimensions(W D H) Net /Gross weight(kg) 227/279 Exterior/Interior SPCC Material Power Supply 220-240V~/60Hz Rated Power/Current 770W/4.5A Controller Microprocessor Qty of sensor 5pcs Alarm type Buzzer,light,remote Alarm content High& low temp.,sensor error,power failed,low battery, door Ajar Foot and wheel 2 feet & 4 wheels Test hole 1 Shelf Shelves :12pcs Lamp LED(9W) Door and lock Glass door (with heater)+lock Defrosting Auto defrosting+ Forced defrosting (with heater) Tank mouth heater Yes Safety mode Sensor failed SM 3.HYC-940(115V~/60Hz): Model HYC-940 Type upright glass door air-cooling Climate class N(10 ~32 ) Temp. range 2~8 Capacity (L) 890L Interior 1030x590x1425 Dimensions(W D H) Exterior 1130x755x1980 Dimensions(W D H) Net /Gross weight(kg) 227/279 Exterior/Interior SPCC Material Power Supply 115V~/60Hz 6
4.HYC-940F: HYC-940&940F Service Manual Rated Power/Current 850W/11A Controller Microprocessor Qty of sensor 5pcs Alarm type Buzzer,light,remote Alarm content High& low temp.,sensor error,power failed,low battery, door Ajar Foot and wheel 2 feet & 4 wheels Test hole 1 Shelf Shelves :12pcs Lamp LED(9W) Door and lock Glass door (with heater)+lock Defrosting Auto defrosting+ Forced defrosting (with heater) Tank mouth heater Yes Safety mode Sensor failed SM Model HYC-940 Type upright glass door air-cooling Climate class N(10 ~32 ) Temp. range 2~8 Capacity (L) 890L Interior 1030x590x1425 Dimensions(W D H) Exterior 1130x755x1980 Dimensions(W D H) Net /Gross weight(kg) 227/279 Exterior/Interior SPCC Material Power Supply 220-240V~/50Hz Rated Power/Current 600W/3A Controller Microprocessor Qty of sensor 5pcs Alarm type Buzzer,light,remote Alarm content High& low temp.,sensor error,power failed,low battery, door Ajar Foot and wheel 2 feet & 4 wheels Test hole 1 Shelf Shelves :12pcs Lamp LED(9W) Door and lock Foaming Door + LOCk Defrosting Auto defrosting+ Forced defrosting (with heater) Tank mouth heater Yes Safety mode Sensor failed SM 7
4. Electrical parts and circuit diagram No. Part code Name QTY Specification Applicable models 1 0074090955 Display board 1 2 0074091548 Control board 1 940 3 0074091548A Control board 1 940F 4 0074091549 Power board 1 7 0074091430A Battery 1 12V/4AH 9 0074091590 LED light 6 10 0074000186 Transformer for door 2 AC/55V 940(220V) heater 11 0074091342 Transformer for door 2 AC/55V 940(115V) heater 11 0070812074 Glass door(right) 1 AC/55V/115W 940 12 0070812035 Glass door (left) 1 AC/55V/115W 940 13 0074090544 Condenser fan motor 3 EBM/19W 940(220V) 14 0074091557 Condenser fan motor 3 ADDA/34W 940(115V) 15 0075030020A Condenser fan motor 2 EBM/30W 940(220V) 16 0074091408 Condenser fan motor 2 EBM/30W 940(115V) 17 0070107553 Condenser fan motor 2 blade(φ200/28 /SUCTION AIR) Circuit diagram with blade(115v) 8
5. Spare parts of cooling system No. Part Code Name QTY Specification Remark 1 0074000248 Compressor 1 EMBR FFI12HBK 940(220-240V~/50Hz) 2 0074000202 Compressor 1 SC12G 940(220-240V~/60Hz) 3 0074091328 Compressor 1 SC12G 940(115V~/60Hz) 4 0270700156 evaporator 1 \ 940(220-240V~/50Hz) 5 0060732673 evaporator 1 \ 940(220-240V~/60Hz) 6 0060732673 evaporator 1 \ 940(115V~/60Hz) 7 0070702086 Condenser 1 \ 8 0070700780 Drier filter 1 9 0073010002A R134a - Qty:R134a 9
蒸发器 Evaporator HYC-940&940F Service Manual Condenser 冷凝器 Drier 干燥过滤器 filter Capillary 毛细管 Suction 回气管 tube Exhaust 排气管 tube Compressor 压缩机 Oil for compressor 压机油 10
6. Spare part list and exploded view HYC-940 Exploded View: HYC-940 (220-240V~/50Hz)spare part list: No Code Name QTY No Code Name QTY 1 0074000248 Compressor 1 22 0070107920 Fan cover 1 2 0070702086 Condenser 1 23 0070401235 Light switch 1 3 0070702085 Exhaust tube 1 24 0070506744 Display film 1 4 0070107923A Frame assembly 1 25 0070203522 Display support 1 5 0270101277 Water collecting box 1 26 0074090955 Display board 1 6 0070401182 Power cord 1 27 0070108114 Front cover 1 7 0070107949 caster 4 28 0270101162 Fan-motor support 1 8 0270101164 Air duct board 2 29 0074090544 Inner fan-motor 3 11
9 0070105097 右 0070105098 左 Upper hinge 2 30 0270100312 Electric box 1 Beam of the temp. 10 0070107926 measurement box 2 31 0270700156 Evaporator 1 11 0070107065 Shelf bar 8 32 0074091548 Control board 1 Temp. measurement 12 0070204066 box 2 33 0074091549 Power board 1 13 0270102070 Lower hinge(left) 1 0270102071 Lower hinge(right) 1 34 0072040045 Wire joint box 1 Back board of air 14 0270100869 Shelf 12 35 0270102040 duct 1 15 0070201732C Price content bar 12 36 0270100313 Electric box 2 1 16 0074091590 LED light 6 37 0270101010 Clips of battery 1 17 0070203322A Light cover 2 38 0074091430A Battery 1 18 0070203323 Door seal 2 39 0075030020 Condenser fan 2 19 0070812074 Right glass door 1 0070812035 Left glass door 1 40 0074000186 transformer 2 Mechanical 20 0070402262C Temp. sensor 1 41 0274000087A thermostat 1 21 0270100042 Clips of shelf 48 HYC-940 (220-240V~/60Hz) spare part list: No Code Name QTY No Code Name QTY 1 0074000202 Compressor 1 22 0070107920 Fan cover 1 2 0070702086 Condenser 1 23 0070401235 Light switch 1 3 0070702085 Exhaust tube 1 24 0070506744 Display film 1 4 0070107923A Frame assembly 1 25 0070203522 Display support 1 5 0270101277 Water collecting box 1 26 0074090955 Display board 1 6 0070401182 Power cord 1 27 0070108114 Front cover 1 7 0070107949 caster 4 28 0270101162 Fan-motor support 1 8 0270101164 Air duct board 2 29 0074090544 Inner fan-motor 3 0070105097 Upper hinge(right) 1 9 0070105098 Upper hinge(left) 1 Beam of the temp. 30 0270100312 Electric box 1 10 0070107926 measurement box 2 31 0060732673 Evaporator 1 11 0070107065 Shelf bar 8 32 0074091548 Control board 1 Temp. measurement 12 0070204066 box 2 33 0074091549 Power board 1 13 0270102070 Lower hinge(left) 1 0270102071 Lower hinge(right) 1 34 0072040045 Wire joint box 1 Back board of air 14 0270100869 Shelf 12 35 0270102040 duct 1 15 0070201732C Price content bar 12 36 0270100313 Electric box 2 1 12
16 0074091590 LED light 6 37 0270101010 Clips of battery 1 17 0070203322A Light cover 2 38 0074091430A Battery 1 18 0070203323 Door seal 2 39 0075030020 Condenser fan 2 19 0070812074 Right glass door 1 0070812035 Left glass door 1 40 0074000186 transformer 2 20 0070402262C Temp. sensor 1 41 0274000087A 21 0270100042 Clips of shelf 48 Mechanical thermostat 1 HYC-940 (115V~/60Hz) spare part list: No Code Name QTY No Code Name QTY 1 0074091328 Compressor 1 22 0070107920 Fan cover 1 2 0070702086 Condenser 1 23 0070401235 Light switch 1 3 0070702085 Exhaust tube 1 24 0070506744 Display film 1 4 0070107923A Frame assembly 1 25 0070203522 Display support 1 5 0270101277 Water collecting box 1 26 0074090955 Display board 1 6 0070401182 Power cord 1 27 0070108114 Front cover 1 7 0070107949 caster 4 28 0270101162 Fan-motor support 1 8 0270101164 Air duct board 2 29 0074091557 Inner fan-motor 3 0070105097 Upper hinge(right) 1 9 0070105098 Upper hinge(left) 1 Beam of the temp. 30 0270100312 Electric box 1 10 0070107926 measurement box 2 31 0060732673 Evaporator 1 11 0070107065 Shelf bar 8 32 0074091548 Control board 1 Temp. measurement 12 0070204066 box 2 33 0074091549 Power board 1 13 0270102070 左 Lower hinge(left) 1 0270102071 右 Lower hinge(right) 1 34 0072040045 Wire joint box 1 Back board of air 14 0270100869 Shelf 12 35 0270102040 duct 1 15 0070201732C Price content bar 12 36 0270100313 Electric box 2 1 16 0074091590 LED light 6 37 0270101010 Clips of battery 1 17 0070203322A Light cover 2 38 0074091430A Battery 1 18 0070203323 Door seal 2 39 0074091408 Condenser fan 2 19 0070812074 Right Glass door 1 0070812035 Left glass door 1 40 0074091342 transformer 2 Mechanical 20 0070402262C Temp. sensor 1 41 0274000087A thermostat 1 21 0270100042 Clips of shelf 48 HYC-940 F: 13
HYC-940F spare part list: No Code Name QTY No Code Name QTY 1 0074000248 Compressor 1 21 0270100042 Clips of shelf 48 2 0070702086 Condenser 1 22 0070107920 Fan cover 1 3 0070702085 Exhaust tube 1 23 0070401235 Light switch 1 4 0070107923B Frame assembly 1 24 0070506744 Display film 1 5 0270101277 Water collecting box 1 25 0070203522 Display support 1 6 0070401182 Power cord 1 26 0074090955 Display board 1 7 0070107949 caster 4 27 0070108114 Front cover 1 Fan-motor 8 0270101164 Air duct board 2 28 0270101162 support 1 9 0070105097 Upper hinge(right) 1 0070105098 Upper hinge(left) 1 Beam of the temp. 29 0074090544 Inner fan-motor 3 10 0070107926 measurement box 2 30 0270100312 Electric box 1 14
11 0070107065 Shelf bar 8 31 0270700156 Evaporator 1 12 0070204066 Temp. measurement box 2 32 0074091548A Control board 1 0270102070 Lower hinge(left) 1 13 0270102071 Lower hinge(right) 1 33 0074091549 Power board 1 14 0270100869 Shelf 12 34 0072040045 Wire joint box 1 Back board of 15 0070201732C Price content bar 12 35 0270102040 air duct 1 16 0074091590 LED light 6 36 0270100313 Electric box 2 1 17 0070203322A Light cover 2 37 0270101010 Clips of battery 1 18 0270200538G Door seal 2 38 0074091430A Battery 1 19 0270803220 Right Foaming door 1 0270803082 Left Foaming door 1 39 0075030020 Condenser fan 2 20 0070402262C Temp. sensor 1 40 7. Instruction of operation Display panel Definition : Light:the switch of light which is in the cabinet. Indicator light for upper and lower temp.(green):distinguish upper and lower temp. (Upper and lower temp. light together means the average value of the temperature); Power indicator light(red): light means power on,flash means main power failed; Alarm indicator light(red): light flashing means alarm happened; Silence :suspend current sound alarm, but this operation cannot cancel remote alarm; Temp. Tran.:alternate the value between upper and lower temp.; Alarm test:check whether it has alarm or not; Cal. & sensor& Cal. cancel:auxiliary button for other operation. 15
Definition Of Combination Key: Sensor + Cal. cancel :setup the temperature(press at least for 5s) Temp. tran. + Cal. :calibrate the value of the five sensors(press at least for 10s) Silence + Alarm test :Factory reset (Press at least for 10s) Cal. : a. During the U-disk was inserted in,press Cal. to query the status of data exporting; b. Press Cal. at least for 10s,to set up the system time of USB board; c. After exporting the data from USB port,press Cal. for 10s to save and exit,the display panel will show the temperature of the cabinet. a. Instructions of Parameter adjustment Content Operating steps Display After turning on the power,shortly press Alarm test button Alarm test The buzzer rings 3times,and the alarm light flash 3 times, it means normal; 1.Press sensor + Cal. cancel for 5s Setup temp. 2.pres Cal. cancel to rise the value,or press sensor to reduce the value 3.After setup the vale,no operation,parameters will be saved and exit setting after 5s. 1.Press sensor + Cal. cancel for 5s 2.Follow the first step(under displaying the set value of temp. and flashing status)in3s, press Cal. For 10s Ordinary 3.When it shows SHE,In4 seconds,press Cal. Cancel or sensor,to choose the parameters current parameter,and show the value of the current parameter. 4.If no any operation,auto jump to next parameter after 4 seconds. 5.After showing all the parameter,if no operation for 10s,system will save and exit to standby status 1.Under standby, press temp. tran. + Cal. for 10 s 2.Press Sensor to show the D-value; 3.Press Cal. cancel to choose the current D-value; Calibrate the 4.Press Alarm test,to show the adjustment range parameters 5. Press cal. cancel or sensor to adjust the calibrate the D value 6.After calibration,press Cal or wait for 5s will back to D-value list 7.After calibrating all D-values,within 5s no operation, it will save the parameter and back to standby. b. Parameter definition b.1 Ordinary parameter list If it displays E5,it means low battery or the battery switch is off Display the set value of temp. and flashing Display the set value of temp. and flashing Display:SEH Display SHE s value Display next Code Display D1 D1-D2-D3-D4-D5-D1 Display the value of the sensor Display the adjust range of the sensor D1~D5 16
No. Parameter Name Adjust range 1 SEH Upper limit of setting temp. Upper limit;20 2 SEL Lower limit of setting temp. Lower 3 AOF Temp. different control of compressor (set value+0.5)±aof limit:-5 4 ALH Alarm for high temp. Upper Default value HYC-940 0.5~3.0 0.5 limit:20 5 ALL Alarm for low temp. Lower b.2 Adjustment list of temperature sensor value limit:-5 NO. Parameter Name Adjust range 8 2 10 2 Default value HYC-940 1 D1 Upper temperature sensor -2.0~+2.0 0 2 D2 Lower temperature sensor -2.0~+2.0 0 3 D3 Control temperature sensor -2.0~+2.0 0 4 D4 Defrosting temperature sensor -2.0~+2.0 0 5 D5 Alarm temperature sensor -2.0~+2.0 0 c. Special function c.1 Defrosting function HYC-940&940F Service Manual Remark Remark Auto defrosting:inner fan motor is working,compressor stop,then start defrosting function. Forced defrost: Start conditions :in stably status,before the compressor start,inspect the temp of defrosting sensor, if the temperature is lower than -0.25, the forced defrosting function will work. Defrosting status:inner fan motor and compressor stopped. Exit conditions:meet any conditions as follows: (a) Upper or lower display temp 5.6 (b) Defrosting for 20min. (c) Temperature of Control sensor or alarm sensor 7 Notice: twice defrosting time is 24hours,After forced defrosting, compressor start working,delay 3min the inner fan-motor begin to work. C.2 Safety mode for failed sensor C.3 USB storage function If the control sensor is damage,the compressor will work for 5 min and stop for 5min. content Operation steps Display Data export Set time 1.plug the usb flash disk into the usb port 2.press cal to query the status 1.press cal for at least 10s 2.press sensor or cal cancel to change the parameter USb means exporting ALL-done Show 1P 1P-2P-3P-4P-5P-1P 3.press cal choose the current parameter and display the value Show the value of 4.press sensor or cal cancel to rise or decrease the value t until proper 5.press cal save and back parameter 1P~5P 17
6.set all the parameter,press cal for at least 5s,or no operation for 10s save and exit. Notice: 1P to 5P means year, month, day, hour, minute. C.4 Self-inspection for alarm function Press alarm test button to simulate all alarm functions, make sure the buzzer,light and remote contacts work fine. d.the alarm Code Alarm Code Name Alarm type Troubleshooting E1 Upper sensor failed Buzzer &light &code Inspect upper sensor Sensor error E2 Lower sensor failed Buzzer &light &code Inspect lower sensor E3 Control sensor failed Buzzer &light &code Inspect control sensor E4 Defrosting sensor failed Buzzer &light &code Inspect defrost sensor Door open High or low temp. N/A Door open for 10min Buzzer &light Inspect door switch or close N/A The upper or lower temp. beyond the range Buzzer &light door Power failure N/A Power failed Power light &buzzer & alternate display temp. Inspect power Battery E5 Low battery or switch is off Buzzer &light &code Inspect battery or switch Remote alarm N/A Above all will trigger the remote alarm. - - Notice: For the low battery alarm, please check it by manual, Press Alarm test button, if the display panel show E5, it means the battery is poor or damaged. e.special considerations for installation 1. If the freezer includes USB board,please adjust the system time of the USB board; 2. If the freezer includes printer, Please adjust the system time of the printer; 3. The blood bank freezer is Air cooling freezer, when put the goods into the freezer, not too dense, make sure the airflow can go through the shelves; 4. The freezer has 4 wheels and 2 feet, After installation, make the two feet back to the earth, keep the freezer stably ; 18
8. Troubleshooting 1.Replace PCB (1)The PCB is fixed at the top of the cabinet, the display board is behind the front cover. Electric box (2)Remove the screws and open the cover of the electric box, the control board and power board is fixed in the electric box. Power board Control board (3)There is a oblong hole under the control board,we can push out the display panel through this hole. Display board Hole for cable 2.The position of the sensor (1) Upper and lower temperature probe is put in the measurement box. We can pull out the probe directly. Upper temp. sensor Lower sensor temp. 19
(2)Remove the inner fan cover,the control sensor and alarm sensor is behind the cover. HYC-940&940F Service Manual Upper sensor Inner door cover (3)Remove the fan motor support,we can find the evaporator, the defrosting sensor is behind the evaporator. Evaporator Inner support fan Defrosting sensor 3.Defrosting heater The defrosting heater is fixed on the evaporator. Defrosting heater 4. Mechanical thermostat (1)the main body of the thermostat is fixed beside the transformer which is protected by a plastic bag, the probe of the thermostat is fixed beside the inner fan. Main body of thermostat Probe thermostat of (2)the red wire is connected with H& C point,the yellow-green wire is connected with the earth. 20
5.The position of the cooling system (1) Compressor is on the top of the freezer. HYC-940&940F Service Manual Electric box Condenser The power box Compressor (2) The pipe of beside the compressor: maintenance tube is used for recharging the cooling system. Suction tube Drier filter Maintenance tube Exhaust tube 9. Test curve 21
Inspired living Haier Medical & Laboratory Products Co., Ltd. Room 403D, Brand Building, Haier Industrial Park, No.1 Haier Road Qingdao China Website: www.haiermedical.com 22