OPERATING MANUAL BFS-MP SERIES FREEZING STAGES

Similar documents
P/N CC-100. Coolant Circulator. Operation manual

Automatic Humidity Controller. Model 512 (Revision C) Operating Manual

Electric Digital Kettle

Warner Instruments, Inc Dixwell Avenue, Hamden, CT (800) / (203) (203) fax

ULTRASONIC HUMIDIFIER

Installation & Operating Instructions

PORTABLE ICE MAKER. Model: IM-124S INSTRUCTION MANUAL

LABORATORY ILLUMINATED FLOTATION BATH

Whynter Portable Ice Maker 33 lb capacity - White

C Y T O - T H E R M III T

Garment Steamer Model SC620

Electron Microscopy Sciences INSTRUCTIONAL MANUAL CAT A. P.O. Box Industry Road Hatfield, PA 19440

Warner Instruments, Inc Dixwell Avenue, Hamden, CT (800) / (203) (203) fax

H-10PM. Refrigerant Leak Detector

Owner s Manual RD432-0 Chemical Controller

Instruction Manual for the Revolutionary Science RS-IF-203 Incufridge

WHYNTER PORTABLE ICE MAKER 49 LB CAPACITY - STAINLESS STEEL

WHYNTER PORTABLE ICE MAKER 49 LB CAPACITY - STAINLESS STEEL

4-IN-1 PIN/PAD RH MOISTURE METER

DESCRIPTIVE AND INSTALLATION INSTRUCTION MANUAL FOR MODEL G1000 AND H1000 SERIES ANNUNCIATORS

Wine Cellar. Instruction Booklet. Model: WC-2038

RATIO:GUARD Model E-1S EC Monitor

PRESSURE WASHER. OWNER'S MANUAL Model VPW49HA 120V., 60 Hz., 13 Amps.

Comfort Series Instant Hot Water System

AI-215 Series Portable Ice Maker

CFS-2000 CHILLER TM. Operating Instructions P/N (110 VAC) P/N (230 VAC)

Upholstery and Drapery Cleaner. Operator and Parts Manual. Model No.: gal Extractor. MNL32506 Rev. 00 (08-98)

JENCO ELECTRONICS, LTD.

Eldex Column Heater. Operator s Manual

Operating & Safety Instructions

18 MaxxAir Portable Evaporative Cooler Owner s Manual This Manual covers the EC18DVS MaxxAir Portable Evaporative Cooler.

2 Liter Deep Fryer. OWNER S MANUAL Model F V., 60Hz., 1500 Watts

AquaStream. ½ HP Floating Fountain. Please Keep for Your Records. Model #AS05 6A, 120V/60Hz Owner s Manual UNIT IDENTIFICATION

HI COD REACTOR and Test Tube Heater 2008 Series

Instruction Manual for the Revolutionary Science RS-PB-200 Dual Poly Pro Bath

EBAC MODEL BD-150 ( ) INDUSTRIAL DEHUMIDIFIER OWNER S MANUAL

Professional 8 Liter Deep Fryer

Because you re not like everyone else. PORTABLE CLEAR ICE MAKER IM200SS OWNER S MANUAL. Manual v1.0

Installation and Operating Instructions

MST, INC. MODEL 2002 CARBON MONOXIDE MONITOR (AMBIENT-TYPE) OWNER S MANUAL

Use and Care Instructions

Tank, Heater, and Control Operation Manual REV. E DATE: 04May2015

TABLE 1 Electrical Specifications

EBAC MODEL KOMPACT INDUSTRIAL DEHUMIDIFIER OWNER S MANUAL

EBAC MODEL CD30 INDUSTRIAL DEHUMIDIFIER OWNER S MANUAL

MST, INC. MODEL 2002 CARBON MONOXIDE MONITOR OWNER S MANUAL

BATH FAN / LED LIGHT READ AND SAVE THESE INSTRUCTIONS MODEL GBR100LED-DECOR TABLE OF CONTENTS. Package Contents 2. General Safety Information 3

OWNER'S MANUAL Model VPW46H 120V., 60 Hz., 13 Amps. PRESSURE WASHER. FANTOM 94 Main Mill Street, Door 16 Plattsburgh, NY Tel.

OPERATIONAL & INSTRUCTION MANUAL. Model: 2400FA (120v 1Ø) Model: 3200FA (220v-240V 1Ø) Voltage Supply to Machine:

AI-215 Series Portable Ice Maker

Air Cleaning Equipment, Inc. 303 N. Main St. Broadway, NC iers.com

BL250. Pulse Blender OWNER S GUIDE

Superdryer Dehumidifiers

CALIBRATABLE 7-FUNCTION THERMO- HYGROMETER USER S MANUAL

Warner Instruments 1125 Dixwell Avenue, Hamden, CT (800) / (203) (203) fax

INSTRUCTION MANUAL BEFORE USE, PLEASE READ AND FOLLOW ALL SAFETY RULES AND OPERATING INSTRUCTIONS.

SAVE THESE INSTRUCTIONS HOUSEHOLD USE ONLY

Operating Instructions

Model # VWK-2 INSTRUCTION MANUAL

Because you re not like everyone else. PORTABLE CLEAR ICE MAKER IM200SS OWNER S MANUAL

Your Pope Simple Set Automatic Tap Timer

48 MaxxAir Portable Evaporative Cooler Owner s Manual

VENTILATION FAN / LED LIGHT

DESCRIPTIVE AND INSTALLATION INSTRUCTION MANUAL FOR MODEL AH1000, BH1000, AND EH1000 SERIES ANNUNCIATORS

PERFECT CHILL WINE CHILLER. USE AND CARE MANUAL MODEL Bottle

Service Manual for Cuber Model CS0415

OWNER S MANUAL & WARRANTY

5 Liter Electronic Deep Fryer

LABORATORY OVENS MODELS INCLUDE: TO-152G & TOH-152G TO-152F & TO-235F DEHYDRATING OVENS MODELS INCLUDE: TD-150F TD-330F TD-500F

OPERATING MANUAL. EchoThermJ DIGITAL HOT PLATE MODEL HP30 AND HP30A. DOCUMENT NUMBER HP30-03 Revised 10 December 2011

Single Channel Heater Controller

Installation and Owner Man u al

Dual Zone Wine Cellar

EBAC MODEL NEPTUNE INDUSTRIAL DEHUMIDIFIER OWNER S MANUAL

BOILING UNIT REDITAP. Installation and User Guide. IMPORTANT: This booklet should be left with the user after installation and demonstration

16 OUTDOOR MISTING FAN

Revolutionary Science RS-IF-202

Digital Heat Block User Manual

CALIBRATABLE 3-FUNCTION THERMO- HYGROMETER USER S MANUAL

BLAST CHILLER / SHOCK FREEZER MODEL AP3BCF30-1

ESS-2200 Double Station Scrub Sink. with Two Sensor Operated Gooseneck Spout. ESS-3200 Deluxe Double Station Scrub Sink

STEAM WALLPAPER STRIPPER MODEL HTW5

Model 405 AIRGARD Fume Hood Monitor

GAS MONITOR Model Model

Henny Penny Island Warmer Model HMI-103 Model HMI-105 TECHNICAL MANUAL

Series: MBC1-TC Mini Benchtop Temperature Controller

VENTILATION FAN. MFG Model : VFB80HLED2/VFB080C4L2 /VFB080C4LED2/ VFB100HLED2/VFB100C4LED1 READ AND SAVE THESE INSTRUCTIONS

OPERATION and MAINTENANCE INSTRUCTION MANUAL. ADU-20B CompriCart

A few simple steps to ensure perfect water, every time.

READ AND FOLLOW ALL SAFETY INSTRUCTIONS

Norcold Repair Guide for Ice Maker

Rev B, 9/2/2009. Kodiak Chiller Overview

OPERATION and MAINTENANCE INSTRUCTION MANUAL. ADU-10 AseptiMini Pneumatic Portable Dental System

VENTILATION FAN READ AND SAVE THESE INSTRUCTIONS. MODEL Pro200/Pro300 TABLE OF CONTENTS. Package Contents. General Safety Information

M770 ph Controller Owner s Manual

DMWD 1 to 4 Zone Water Detection Alarm with Distance to Leak Measurement Installation and Operation Manual

READ AND SAVE THESE INSTRUCTIONS READ CAREFULLY BEFORE ATTEMPTING TO ASSEMBLE, INSTALL, OPERATE OR MAINTAIN THE PRODUCT DESCRIBED. PROTECT YOURSELF AN

OPERATION and MAINTENANCE INSTRUCTION MANUAL. ADU-12 Task Force Pneumatic Portable Dental System

EBAC MODEL CD60 INDUSTRIAL DEHUMIDIFIER OWNER S MANUAL

e Bath Fan with Light User s Guide

Transcription:

PHYSITEMP INSTRUMENTS INC 154 Huron Avenue Clifton NJ 07013 USA Tel: 973-779-5577 Fax: 973-779-5954 OPERATING MANUAL BFS-MP SERIES FREEZING STAGES Models BFS-3MP, BFS-5MP, BFS-30MP EB FRONTIS EB BFS-MP 6310

APPENDIX 2 OPERATING INSTRUCTIONS ECV-1 ELECTRICALLY OPERATED SOLENOID VALVE The ECV-1 is supplied with an AC line cord and 12 VDC desktop power supply. 1 Connect the line cord to the socket on the power supply and to the rear auxialiary outlet on the BFS-MP controller. 2 Connect the 12VDC power supply to the input socket on the front of the ECV-1. 3 Connect the water tubes as shown below: FROM WATER SUPPLY Male connector Female connector Extension tube with 1 female connector and 1 universal connector ECV-1 Freezing Stage TO WASTE Extension tube with female connector Stage has 2 water tubes with male connectors 11

APPENDIX 1 OPERATING INSTRUCTIONS FOR PHYSITEMP PUMP AND TANK UNIT, PTU-3 1. Unscrew cap. Fill reservoir with 5 gallons of distilled water. 2. Connect extension tubing to the back of the controller and the connections on the tank. Direction of the flow is not important. All water fittings are automatically self-sealing when disconnected to prevent water spills. 3. Connect the AC line cord to the receptacle on the pump and to the outlet on the rear of the controller. This ensures that water is always flowing when the controller is on. 4. Switch on the controller. GENERAL INSTRUCTIONS The use of distilled water is recommended. This avoids discoloration of the tubing due to organic matter in untreated water. A purification agent can be added to the water. Any commercially available dehumidifier treatment in liquid, powder or tablet form may be used. If the stage will be operated for long periods, water temperature may gradually increase. Ice cubes can be added to the tank to maintain low temperature. The tank should not be placed more than three feet below the stage itself. When filling the reservoir for the first time, or when refilling it, a small amount of air may become trapped in the pump housing and cause intermittent noise, cavitation and a reduced flow rate. If this occurs, screw the cap firmly onto the tank and, while pump is running, tip it at an angle of approximately 45 degrees to the horizontal in all four directions as shown below. This will eliminate air in the pump housing and restore flow to its maximum. 10

7.0 MAINTENANCE, WARRANTY AND SERVICE 7.1 MAINTENANCE The stage needs no maintenace at all. It may be cleaned as necessary with a soft cloth, water or detergent. DO NOT IMMERSE IN WATER 7.2 WARRANTY Physitemp Instruments Inc. warrants this system to be free from defects in material or workmanship for 12 months from date of shipment. Repair or replacement will be made at no charge at the discretion of Physitemp if the defect is not the result of misuse or abuse. Physitemp accepts no consequential liability for delay in delivery, alleged faulty performance of the product or any other cause. Cables and probes are considered expendable and are not covered by this warranty. 7.2 REPAIRS AND RECALIBRATION For technical applications information on this instrument contact us at: Tel: 973-779-5577 Fax: 973-779-5954 E-mail: physitemp@aol.com In the event that any part of this system is to be returned for repair or recalibration, please pack it with care (in the original packing material if possible) and send it prepaid to: Service Department PHYSITEMP INSTRUMENTS INC 154 Huron Avenue Clifton, NJ 07013 USA Please include with the instrument: 1. A note describing any problems encountered 2. The name and telephone number of a person we can contact 3. The complete return address for shipping. For your protection, please pack the item carefully and insure against possible damage or loss. Physitemp will not be responsible for damage resulting from careless packaging. Please return freight prepaid. 9

6.0 SPECIFICATIONS 6.1 CONTROLLER Operating Range: BFS-3MP and 5MP: -35 C to ambient BFS-30MP: -40 C to ambient Digital Readout Resolution: 0.1 C Accuracy: ±0.2 C Ambient Operating Range: 15-45 C Input Power Requirements: 100-120VAC @ 4 Amps or 200-240VAC @ 2 Amps Grounded Outlet Power Output to stage: BFS-30MP - 12VDC @ 10Amps BFS-3 & 5 MP - 5VDC @ 10Amps Display: Dual LED display of set and run Temperature Safety Features: Power shutoff during over-temperature condition caused by inadequate coolant flow (50 C). Short circuit protection. Size: 9 1/2 wide x 10 deep x 4 1/2 high Weight: 5 lbs Other Features: Self-sealing water connectors. Auxiliary switched AC output for ECV-1 or PTU-3. 6.2 STAGE Temperature Range: -40 C to ambient Controlled Surface Area: BFS-3MP - 3 x 4cm BFS-5MP - 3.8 x 4cm BFS-30MP - 7.5 x 8cm Weight: 15ozs, including leads Lead Length: 5ft 8

collar in an anti-clockwise direction. An ohmmeter will be required to make several readings on the stage connector. Once the connector has been removed the pin numbers are visible as slightly raised numerals on the face of the connector. Between pins 1 and 2 1-3 ohms 3 and 5 100-110 ohms 6 and 7 20-27 Kilohms If the readings you obtain do not fall within these ranges, please consult our engineering department. 3 Controller display is illuminated but module does not cool Check the setting of the set temperature and adjust if necessary using the instructions provided in Section 4.6. If the set temperature is OK check the connector to the module cable on the rear of the controller and the fault condition light on the front of the controller. If the fault light is on follow the instruction for fault condition number 2 7

4.7 REMOVING SPECIMEN FROM STAGE Press the Freeze/Thaw switch and hold in the QUICK THAW position for a few seconds. Allow the stage to warm up until specimen slides off easily. Wipe stage dry. 4.8 TYPICAL STAGE PERFORMANCE The degree of cold obtained is dependent on the temperature of the cooling water. The stage temperature should reach 55-65 below that of the cooling water. Therefore, a stage temperature of -40 can be achieved under normal conditions. 1/2 thickness of liver which covers the stage should freeze within 15-20 minutes. Temperature throughout the tissue should drop to -20 C in a further ten minutes. Tissue under 1/16 in thickness should be ready for sectioning within 90 seconds after switching on. 4.9 MEASUREMENT AND CONTROL OF TISSUE TEMPERATURE One of the main advantages of these freezing stages is the ease with which the temperature of frozen tissue can be changed. Optimum cutting temperature varies for different tissue and is best determined by experiment. the following temperatures have been found satisfactory: TISSUE TEMPERATURE C Kidney -15 Brain -18 Liver -18 Thyroid -20 Skin -30 Keloid -35 Patella -40 4.10 ADJUSTING THE VIEWING ANGLE OF THE INSTRUMENT The two front feet can be flipped down toadjust the viewing angle. Rotate the feet down and towards the front until they lock in position. To return the the original position, pull down slightly then rotate back.. 5.0 TROUBLESHOOTING 1 Controller display fails to illuminate when power is turned on Check power cord continuity from the AC outlet to the input socket on the rear of the controller. Check the fuse next to the AC input socket in the rear panel. If these are both OK, try a differnt AC outlet 2 Fault Condition lamp is illuminated As a safety feature, this unit has power supply short circuit protection. In the event that the output becomes short circuited by a wiring fault, a protection circuit built in the power supply senses the increase in current and shuts off the output power. Before output power can be restored to the controller, the AC power supply must be turned off and the fault condition corrected. Consult our engineering department for further details. If the PTU-3 or ECV-1 is being used, check for adequate water flow as described in section 3.1. Touch the silver heat sink, the middle segment of the stage, to see if is is hot to the touch -- this would indicate a low or inadequate flow rate. If water flow is OK, the stage or controller may be defective. In order to check the stage, turn of the power to the controller and remove the stage connection from the rear of the controller by turning the 6

4.0 OPERATING INSTRUCTIONS WARNING The auxiliary outlet is designed for use only with our approved accessories, the PTU-3 Pump and Tank Unit and the ECV-1 Electrically Operated Solenoid Valve. No other AC devices should be connected to this outlet without first consulting our engineering department. 4.1 Ensure that the POWER Switch on the rear panel of the controller is in the OFF position 4.2 Check the VOLTAGE SELECTOR SWITCH on the rear panel to ensure that it is set for the cor rect line voltage. 4.3 Connect the AC power cord to any suitable grounded AC outlet. 4.4 Unless you have PTU-3 or ECV-1, turn on the water supply and adjust the flow rate until it flows steadily, without spurting. 4.5 Turn the POWER switch on the rear of the controller ON and ensure that the display on the front is illuminated. If you are using ECV-1 and have installed it according to the instructions in Appendix 2, turn on the water supply now and adjust the flow rate as described above. If you are using PTU-3, and have installed it according to instructions in Appendix 1, check that the neon lamp on the side of the pump is illuminated and you can hear a low hum from the pump. Check that the water is flowing by observation of air bubbles inside the tubing or remove the cap on the tank and check the water stream from the return tube which is located at the top of the tank underneath the tubing connections. 4.6 SETTING THE STAGE TEMPERATURE The operating temperature may be set anywhere in the range from ambient down to -35 C (-40 C for BFS-30MP.) To set control temperature: a) Press SHIFT. A digit will flash b) Use UP and DOWN buttons to change the flashing digit. Press SHIFT again to select the next digit. c) When all digits have been changed, press SET to enter new control temperature. FREEZE/THAW SWITCH FAULT CONDITION SET PV -14.9 SV -15.0 SHIFT DOWN FRONT PANEL CONTROLS UP RUN display SET display WARNING Various functions of the display and controller are accessible via the four buttons on the display. It is inadvisable to make any adjustments to these parameters, which are preset at the factory for optimum performance. In the event that these operating parameters are changed accidentally, please consult our engineering department for restoration to the original settings. 5

3.0 SETTING UP THE SYSTEM This section gives instructions for set-up if the equipment is to be used with tap water: If a Pump and Tank Unit is to be used to circulate the cooling water, see Appendix 1 3.1 ATTACH THE WATER TUBES PUSH TO CONNECT Connected to the stage are an electrical lead and two PVC tubes which terminate in male self-sealing connectors. A pair of 5 foot water tubes with mating female connectors is also provided. Connect the 5 foot tube to the tubes on the stage. Attach one tube to the the laboratory water supply, while the other goes to a SQUEEZE TAB IN TO RELEASE convenient sink or drain. The direction of the water flow through the stage is not important. The water supply should be adjusted until it flows steadily from the output tube without spurting. If the water supply is accidentally reduced or turned off while the equipment is operating, the stage will tend to heat up. However, an electronic safety circuit will sense the rise in temperature and shut off power to the stage before damage can occur. 3.2 ATTACH THE STAGE TO THE MICROTOME If the make and model of the microtome were specified with the order, the freezing stage should have been shipped with the correct mounting hardware already in place. The most frequently used hardware is a 3/4 vise block. Other pillar and blocks are available as well as adaptor plates for many brands and models of microtome. 3.3 CONNECT THE STAGE TO THE CONTROLLER Connect the electrical lead from the stage to the mating socket on the rear of the controller. Rotate the collar on the connector housing in a clockwise direction to lock in place. Connect the blue thermocouple plug to the mating socket on the rear panel. This is the Stage Temperature Feedback Sensor. 3.3 ATTACH AC LINE CORD Before connecting controller to AC power supply, check the VOLTAGE SELECTOR SWITCH on the rear panel to ensure that it is set for the correct line voltage.. Connect the line cord supplied with the controller to the mating socket on the rear panel. REAR PANEL CONNECTIONS POWER SWITCH THERMOCOUPLE INPUT SOCKET AC SOCKET AC FUSE AND VOLATGE SELECTOR AUXILIARY SWITCHED OUTLET OUTPUT TO STAGE COOLING FAN 4

BFS/MP SERIES THERMAL STAGES 1.0 GENERAL DESCRIPTION 1.1 Physitemp manufactures three freezing stages for use with microtomes. All stages operate in the same way and reach the same minimum temperature. They differ only in surface area, as follows: BFS-3MP has stage area 3cm x 4cm BFS-5MP has stage area 3.8cm x 4cm BFS-30MP has stage area 7.5cm x 8cm This manual also covers the FRM/MP series freezing microtomes which comprise a BFSMP Stage and and ERMA Microtome. 1.2 The BFS/MP THERMAL STAGE The stage is a metal plate, 1/16 thick. Heat is supplied to or withdrawn from the specimen by means of an electric heat pump attached to the underside of the plate, and excess heat is conducted away by cooling water. MOUNTING HARDWARE TEMPERATURE CONTROLLED SURFACE WATER TUBES ELECTRICAL CONNECTIONS 1.3 The CONTROLLER will maintain stage temperature at any point between -35 (-40 for BFS- 30MP) and ambient. Resolution is 1/10 C. It has a simple set point control LED readout and pushbutton entry control. Once the unit is turned on, the LED displays the set point and current temperature. The set point can be adjusted by simply incrementing the set point up or down with the pushbuttons. 1.4 Optional PUMP AND TANK UNIT, PTU-3. The stage needs running water for operation. If access to a water tap is inconvenient, a circulating pump with a small water tank is required. 1.5 Optional ELECTRICALLY OPERATED SOLENOID VALVE, ECV-1, enables the used to activate a tap water supply with the controller power switch. 2.0 UNPACKING AND CHECKING THE PACKAGE CONTENTS The BFS-MP Controller and Stage is comprised of the following items. 1. Digital Temperature Controller with AC line cord 2. A Freezing Stage with 3 foot connector lead and water tubing with self-sealing male fittings. 3. 5 foot water tubing extensions with self-sealing female fittings 4. Operating Manual with Warranty Registration Card Optional Accessories purchased with the unit may include: 1 PTU-3 Pump and Tank Unit, supplied with AC line cord and 5 foot extension water tubes with male and female self-sealing fittings 2 ECV-1 Electrically Controlled Solenoid Valve, supplied with AC line cords and a 12VDC 1.3 Amp desktop power supply 3

WARNING! DO NOT USE DRY ICE ON TOP OF THE STAGE Several customers have reported using Dry Ice (CO2) to speed up the freezing of tissue specimens mounted on their BFs series freezing stage. This may cause several problems. Putting dry ice, which is at -85 C, directly onto the metal surface of the stage fools the controller into believing that the top plate has reached its set temperature and so it shuts off power to the thermoelectric cooling elements. If there is too much dry ice on the plate it can cause irrepairable damage to the thermoelectric modules which are designed for a temperature gradient of 60 C. With dry ice on the surface of the stage this gradient can exceed 100 C. This can cause severe mechanical stress on the components of the module due to thermal expansion and contraction and can literally shear it apart. Bearing the above cautionary information in mind, it is possible to use dry ice of top of the specimen itself to reduce its temperature quickly. Do not put the dry ice directly on the metal surface of the stage. Please contact our technical department for further assistance. 1-800-452-8510 2

OPERATING MANUAL BFS-MP FREEZING STAGES FOR MICROTOMES BFS-3MP, BFS-5MP, BFS-30MP CONTENTS SECTION PAGE 1.0 GENERAL DESCRIPTION 2 2.0 UNPACKING AND CHECKING 2 CONTENTS 3.0 SETTING UP THE SYSTEM 3 4.0 OPERATING INSTRUCTIONS 4 4.6 Setting Stage Temperature 4.7 Removing Specimen from Stage 4.8 Typical Stage Performance 5.0 TROUBLE SHOOTING 5 6.0 SPECIFICATIONS 7 7.0 MAINTENANCE, WARRANTY AND 8 SERVICE Appendix 1: Operating Instructions for Physitemp 9 Pump and Tank Units Appendix 2: Operating Instructions for ECV-1 10 Elelctrically Controlled Solenoid Valve EB BFS-MP 6310 09/20/04