IMPROVING THE QUANTITATIVE REPRODUCIBILITY OF AN ATMOSPHERIC PRESSURE CHEMICAL IONISATION SOURCE FOR GC/MS AND GC MS/MS ANALYSIS

Similar documents
An Introduction to Mass Spectrometry

This document is a preview generated by EVS

Application Note. The Agilent 1200 Infinity Series Online SPE Solution for the Highest Sensitivity. Author. Abstract.

Enhancing Throughput of Online SPE Analysis of Pesticides in Wastewater The Agilent 1200 Infi nity Series Online SPE Solution

High Volume Injection up to 5 ml with the Agilent 1200 Infinity Series Online SPE Solution for Highest Sensitivity

Automated Solid Phase Extraction (SPE) of EPA Method 1694 for Pharmaceuticals and Personal Care Products in Large Volume Water Samples

Analysis of Pharmaceuticals and Personal Care Products in Groundwater by Online SPE and LC/MS/MS

AppNote 6/2008. Automated SPE and Fast GC-ECD Analysis of PCBs in Waste Oil KEYWORDS ABSTRACT INTRODUCTION. MACH, Fast GC-ECD, SPE, PCB, waste oil

On Line SPE LC/MS/MS Applications

Direct Injection and Online SPE LC/MS/MS for the Determination of Pharmaceuticals and Personal Care Products (PPCPs) in Water Samples

DETERMINATION OF PAH In Particulate Matter PM 10 with SPE/EVAporation

Chemistry Instrumental Analysis Lecture 29. Chem 4631

Optimizing the Performance of the Agilent 1290 Infinity Evaporative Light Scattering Detector

CALIDUS tm the Modular micro Gas Chromatograph

On-line SPE Enrichment of Trace Organic Contaminants in Water and Juices

Application Note. Authors. Abstract

API 2000 LC/MS/MS TurboIonSpray Ion Source Manual

Detailed characterisation of essential oils by flow-modulated GC GC TOF MS with Tandem Ionisation

APPLICATION NOTE DETERMINATION OF THE HYDROCARBON-INDEX DIN EN ISO :2000. APPLICATION NOTE As of July 2017, Version: 1.

Rapid Quantification of Polar and Semipolar Pesticide Metabolites with Combined Online SPE and Direct Injection

MEPS - Micro Extraction by Packed Sorbent Online SPE for GC and LC sample preparation - Extraction to injection in a single process

Proudly serving laboratories worldwide since 1979 CALL for Refurbished & Certified Lab Equipment

Q Trap LC/MS/MS TurboIonSpray Ion Source Manual. About This Manual... 1 Conventions Introduction... 3 Features... 4

Solid Phase Extraction using the 96-well Format ISOLUTE -96 Fixed Well Plates and ISOLUTE Array 96-well Plates

DETERMINATION OF PAH In Particulate Matter PM10 With SPE/EVAporation

Tekmar-DOHRMANN The 3100 Sample Concentrator From ManualLib.com

GC Injectors. GC Injectors

Perkin Elmer Clarus 500 Gas Chromatograph

HISTORY OF VARIAN CAPILLARY INJECTORS

Mass Spectrometry. Fundamental LC-MS. Vacuum Systems

SCION 436-GC. Dimensions and Weights

JSB is an authorised partner of DETERMINATION OF THE HYDROCARBON-INDEX DIN EN ISO :2000. i n s p i r a t i o n m e e t s i n n o v a t i o n!

PURITY GRADES OF LAB REAGENTS

BenchTOF. The next generation of time-of-flight mass spectrometers for GC and GC GC

Thermo Scientific Rocket Evaporator. Revolutionary solvent evaporation complete laboratory workflow

Auto SPE-06Plus Solid Phase Extraction System. Reeko Instrument USA

Agilent InfinityLab Online SPE Solution

Standard Operating Procedure Gas Chromatograph Mass Spectrometry (GC/MS) - Agilent 7890B GC & 240 Ion Trap MS

TYPES AND APPROPRIATE INJECTORS

IC Professional Detector

FREESTYLE SPE UNIQUE POSSIBILITIES FOR SOLID PHASE EXTRACTION

FIXED GAS ANALYZER (FGA)

Agilent 1290 Infinity II Online SPE Solution

Simplicity. Reliability. Performance. Stratum. Automated VOC Sample Prep System

NEW Restek Electronic Leak Detector

Spark Holland introduces

Development of a Novel Contamination Resistant Ionchamber for Process Tritium Measurement and use in the JET First Trace Tritium Experiment

Agilent 1290 Infinity II Vialsampler Data Sheet

GC DETECTORS Reduction Gas Detector - RGD. Overview

E²M. Innovation with Integrity. Enhanced Environmental Mass Spectrometer. Safety & Security

Automated Sample Preparation: SPE GPC - EVAporation SUPERIORITY BY ADAPTATION. Visit us at

Hiden HPR-20 QIC EGA for evolved gas analysis in TGA-MS

Process and Air Monitoring. Online GC systems for continuous monitoring of organic compounds in air and gaseous samples

Thermo Scientific Dionex AutoTrace 280 SPE Instrument. Automate Large-Volume. Solid-Phase Extractions

UVC Detection as a Potential for Alpha Particle Induced Air Fluorescence Localisation

API LC/MS/MS TurboIonSpray Ion Source Manual Table of Contents

Agilent 7890A Network Gas Chromatograph

Calibration Standards for Measurement of Volatile Siloxanes in Anaerobic Digester and Landfill Gases Using Permeation Tubes

White paper on the Dielectric Barrier Discharge Detectors. Introduction:

Hiden QGA Quantitative Gas Analysis System

DFE T ECHNOLOGY. At Conceptronic, the design and management process for reflow soldering systems is called Dynamic Flow Engineering, or DFE.

ANNEX I. Technical specifications for: Supply of various Trace Explosive Detection Systems

Technical specifications of FT-IR Spectrometer

300 series HPLC autosamplers

Automação de SPE utilizando sistema robótico Triplus RSH autosampler. Nova Analitica

EH&S Impact of Common Water Resistant Additive Technologies in Gypsum Board

An Advanced Heat Exchanger for the Agilent 1290 Infinity Thermostatted Column Compartment

NOTE: The color of the actual product may differ from the color pictured in this catalog due to printing limitation.

Restek s Electronic Leak Detector

Experimental Particle Physics PHYS6011 Joel Goldstein, RAL

High resolution mass spectrometry (HRMS*) in Graz

Improved HPLC Separations at Elevated Temperatures Using a New Stand-alone Preheater

Purge and Trap Troubleshooting Made Easy?

Nexis GC-2030 Specifications

Cone Calorimeter. (ISO 5660 ASTM E 1354) The most comprehensive bench scale fire test THE BENCHMARK IN FIRE TESTING

Ministerial Decree No. 14/2005 (VI. 28.) KvVM. on the rules concerning the screening surveys of remedial site investigations.

Sensitive Determination of Microcystins from Environmental Waters Using On-Line SPE

To request a quote or locate your nearest dealer, contact Customer Care toll-free in the US at or

Aquamax KF Oil Evaporator. User Manual

Waters Xevo G2-S QTof Operator s Overview and Maintenance Guide

Monitoring Flammable Vapors and Gases in Industrial Processes

Customised HPLC Systems for Highest Performance

Waters Xevo G2-XS QTof Overview and Maintenance Guide

DYE QUALITY CONTROL REFERENCE MATERIAL KIT

Tool-free LC connections for HPLC, UHPLC and nano LC systems

AppNote 3/2009. Automation of Solid Phase Extraction Methods using a Robotic X-Y-Z Coordinate Autosampler with Software Control KEYWORDS ABSTRACT

Soft Ionisation Techniques, Terry Whitmore, Hiden Analytical

Gas Chromatography (GC-342)

Datalogging as standard

Online SPE-UHPLC-MS. CHRONOS for Control & System Integration. Flexible autosampler for sample preparation. Superior UHPLC system for analysis

Simulation Of Pneumatic Drying: Influence Of Particle Diameter And Solid Loading Ratio

Variable Temperature Microprobe Systems

This year ASMS is in Montreal, Canada and SIS will be present there

Universal NanoFlow Sprayer Installation and Maintenance Guide

IR15TT-R Miniature Infrared Gas Sensor for Monitoring Carbon Dioxide and Methane up to 100% Vol.

Heated tools. Semiconductor equipment

[ ACQUITY QDa DETECTOR ] SEPARATING BEYOND QUESTION BRINGING THE POWER OF MASS DETECTION TO YOU

To request a quote or locate your nearest dealer, contact Customer Care toll-free in the US at or

Fundamentals of Solid-Phase Extraction (SPE): Principles and Practical Tips for Developing Sample Preparation Methods

UV-Ozone Technology and Applications

Transcription:

IMPROVING THE QUANTITATIVE REPRODUCIBILITY OF AN ATMOSPHERIC PRESSURE CHEMICAL IONISATION SOURCE FOR GC/MS AND GC MS/MS ANALYSIS Jones GR 1, Douce D 1 1 Waters Corporation, Stamford Avenue, Wilmslow, United Kingdom Introduction Recent studies have evaluated the use of atmospheric pressure chemical ionisation gas chromatography mass spectrometry for quantitative analysis of polychlorinated dibenzop-dioxins and polychlorinated dibenzofurans 1, as well as other persistent organic pollutants such as polyaromatic hydrocarbons 2. Fundamental to the suitability of this ionisation technique for this purpose is the reproducibility of the ionisation efficiency and hence the recorded peak response. It has been found that one of the most significant factors in the variation of peak response from one analytical run to the next is how the coronal discharge process is affected by the elution of the diluent solvent, which precedes the elution of the analytes of interest during the GC run. This study describes the inimical effects of the solvent front for this ionisation technique and discloses methods for its mitigation. Materials and methods All data presented here were acquired using an atmospheric pressure chemical ionisation source designed for gas chromatography (APGC). Initial studies were performed on a Q- IMS-ToF mass spectrometer (Synapt G2-S) operating in MS mode, with performance optimised for sensitivity and an operating resolution of 14,000 R.P. (FWHM definition). Additional studies were performed on a tandem quadrupole mass spectrometer (Xevo TQ-S) using MRM transitions to target specific constituents of the samples, with Q1 and Q2 resolutions set to 0.7Da. Two samples were used in these studies; a 10:1 dilution (in nonane) of a CS3 US EPA1613 dioxin/furan standard (Wellington Laboratories), with analyte concentrations ranging from 1 to 10pg/µL, and a Semivolatile Organics US EPA Method 8270 Megamix sample (from Restek) diluted in hexane to a concentration of 1pg/µL. Separations were performed using both a Rxi-5Sil MS (Restek) and a DB-5MS (Agilent) GC column, both of 30m length, 0.25mm ID and 0.25µm film thickness, run at a flow rate of 2.0mL/min. The GC ovens used were both Agilent 7890As and a 7683 autosampler was used for sample introduction. The GC inlet was a split/splitless inlet operating in the pulsed splitless mode. Injections were of a volume of 1.0µL.

Results and discussion In positive ionisation mode, the APGC source operates at optimum efficiency when in a coronal discharge mode known as Hermstein s Glow 3,4, characterised by a steady current and quiet operation with little or no sparking. The high potential gradient about the tip of the pin imparts sufficient energy on free electrons that they can ionise molecular nitrogen (IE = 15.6eV) 5. This process yields a N 2 + radical cation, which is repelled from the corona pin and can go on to ionise analyte species, either through direct charge exchange or through an intermediate species such as water. The process also yields an additional electron which is accelerated towards the pin and thereby initiates an electron avalanche as this reaction is repeated. High energy photons from recombination reactions within this ionisation region causes ionisation of molecular nitrogen further out in the corona pin s electric field, thus seeding further reactions and sustaining the corona. The voltage that has to be applied to the corona pin to achieve a desired current is characteristic of how efficiently this corona reaction is occurring. Research studies into the ionisation processes occurring within an atmospheric pressure chemical ionisation source coupled to a gas chromatograph identified a significant correlation between the recorded peak response for a given compound and the corona voltage at the time the compound eluted from the gas chromatograph. The data shown in Figure 1 is from the repeated analysis of the Megamix sample on a Q-IMS-ToF MS under the same conditions. The peak response for the protonated ion of phenanthrene is shown for a number of sequential acquisitions and plotted in conjunction with the corona voltage (at the time that the species eluted) on the secondary y-axis. There is a strong inverse correlation between gross changes in both values. Figure 1: Variation in peak response for protonated phenanthrene and the corona voltage at its elution time Figure 2: Variation in corona voltage during analysis for two sequential acquisitions

Further study into this variation in corona voltage from one sample to the next identified the solvent front as having a significant impact on the ionisation conditions of the source. Figure 2 shows the corona voltage during two subsequent analyses while the corona current is regulating at a value of 2.0µA. The solvent (hexane) starts to elute from the column at an acquisition time of 1.2 minutes and causes a significant rise in corona voltage. The ionisation potential of hexane (10.1eV 6 ) is lower than that of molecular nitrogen, so is more readily ionised and within a larger volume about the tip of the corona pin. The sudden rise in positively charged ions about the tip of the corona pin causes a weakening of the electric potential gradient, with a corresponding reduction in current from electrons avalanches. The corona voltage increases in order to compensate for this change and thereby maintain the regulation current. As demonstrated in Figure 2, this adverse coronal form can persist long after the solvent front has passed, and in a highly irreproducible fashion. It was found through experimentation that the corona voltage could be stabilised by maintaining the corona current at either 0µA or at an unusually high value (20µA) throughout the solvent front and then returning it to its normally optimal value of 2.0µA. The reproducibility of the corona voltage at any point within the analysis was seen to improve from having an RSD of 4.6% to 0.9% as a result of setting the current to either 0µA or 20µA for the solvent front. It is believed that it is the re-initiation of the corona discharge reaction once the source conditions have settled that provides this stability improvement. Additional studies into the source environment determined that the solvent front can also be purged more effectively from the ionisation chamber by varying the cone gas flow. The cone gas is used to partially ballast the pumping requirements of the mass spectrometer s sampling orifice, and therefore affects the flow of gas drawn in from the source enclosure. When the cone gas flow is lowered, the amount of gas drawn through the ionisation chamber is increased. This can more effectively remove solvent vapour from the ionisation region about the corona pin and is especially beneficial when using higher boiling point solvents such as nonane. In order to qualify the effect of the described variations in source conditions during the solvent front, a study was performed on a tandem quadrupole mass spectrometer, analysing various components of the 8270 Megamix sample. Three source conditions were compared; A) standard source conditions with a cone gas flow of 200L/Hr and corona current of 2.0µA, B) with the corona current at 20µA during the solvent front and C) with corona current at 20µA and cone gas flow at 0L/Hr during solvent front. The stability of one of the targeted components of this sample (molecular ion of pyrene) under the three sets of source conditions are shown in Figure 3, along with the corona voltage at the time the peak eluted. A summary of the relative standard deviations of the targeted components and their corresponding corona voltages is given in Table 1. Each

modification to the source conditions for the solvent front resulted in an improvement in reproducibility. A B C Figure 3: Variation in pyrene peak response and corona voltage on tandem quadrupole with A) no variation in source conditions, B) with the corona current at 20µA during the solvent front and C) with corona current at 20µA and cone gas flow at 0L/Hr during solvent front A second analysis was performed using a dioxin/furan standard diluted in nonane, analysed on the Q-IMS-ToF. This analysis was undertaken to qualify the effect of lowering the cone gas during the solvent front when using a higher boiling point solvent.

An additional parameter was varied, termed the auxiliary gas flow. This gas flow is used to sweep the source enclosure and prevent the build up of contamination. For this experiment, when the cone gas flow was lowered, the auxiliary gas flow was increased. This helps prevent back-streaming of gases from the exhaust port of the source enclosure by maintaining a higher combined input gas flow than the pumping requirement of the sampling orifice of this mass spectrometer. Two sets of source conditions were tested; A) with the corona current set to 0µA during the solvent front, 2.0µA thereafter and B) with the corona current at 0µA, cone gas flow of 50L/Hr and auxiliary gas flow of 400L/Hr during the solvent front, returning to their respective optimised values of 2.0µA, 225L/Hr and 50L/Hr thereafter. The peak response for 2,3,7,8 Tetrachloro( 13 C 12 )dibenzo-p-dioxin under the two conditions used is shown in Figure 4. Table 2 summaries the relative standard deviations of the peak responses, both as absolute and internal standard corrected values. Table 1: RSDs of peak response and corona voltage for Megamix analysis on tandem quadrupole, using three source conditions Table 2: Comparison of RSDs of peak response for PCDD/F analysis on Q-IMS-ToF using two source conditions

A B Figure 4: Peak response for 2,3,7,8 Tetrachloro( 13 C 12)dibenzo-p-dioxin on Q-IMS-ToF with A) corona current set to 0µA for solvent front and B) corona current set to 0µA, cone gas at 50L/Hr and auxiliary gas flow set to 400L/Hr for solvent front The analysis showed a near two-fold improvement in reproducibility as a result of varying the cone and auxiliary gas flows during the solvent front. As expected, no appreciable difference was seen in the reproducibility of the internal standard corrected peak responses. This study demonstrates that it is essential to control the corona discharge in order to maximise the quantitative reproducibility of analyses performed using an atmospheric pressure chemical ionisation source with a gas chromatograph. The dominant source of variation for this type of analysis is how the corona is affected by the elution of the solvent during the GC cycle. The techniques described result in stable and reproducible corona discharge conditions, with a corresponding benefit to the reproducibility of the analysis. References: 1. van Bavel B et al. (2014); Analytical Chemistry. (in press) 2. Herrera LC, Grossert JS, Ramaley L. (2008); Journal ASMS. 19(12): 1926-1941 3. Chang JS, Lawless PA, Yamamoto T. (1991); IEEE Transactions on Plasma Science. 19(6): 1152-1166 4. Hermstein W. (1960); Archiv Fϋr Electrotech. 45: 209-279 5. Dzidic I, Carroll DI, Stillwell RN, Horning EC. (1976); Analytical Chemistry. 48(12): 1763-1768

6. National Institute of Standards and Technology; www.nist.gov