Competency: Critically evaluate the function of the ionisation chamber in the Linear Accelerator and its importance for correct treatment delivery

Similar documents
Ionisation Chambers Containing Boron as Neutron Detectors in Mixed Radiation Fields

Co-60 irradiator for therapy level calibrations at SSDLs

Gaseous detectors. 1

Victoreen & 660-8

PROFILER Family IC PROFILER, IC PROFILER - MR & SRS PROFILER. Your Most Valuable QA and Dosimetry Tools

Characterization of dose impact on IMRT and VMAT from couch attenuation for two Varian couches

Dosimetric verification of intensity modulated radiation therapy

QUALITY ASSURANCE AND SAFETY AT A TID RADIATION TEST LABORATORY

COMMISSIONING OF THE DOSE DELIVERY SYSTEM AT MEDAUSTRON

Radiation Safety issues for the PF-AR in KEK

February 12, PHY357 Lecture 11. Experimental Methods. Accelerators. Particle Interactions. Particle Detectors. Full experiment (eg.

Table of Contents. Check the LCLS Project website to verify 2 of 12 that this is the correct version prior to use.

Small-segment intensity modulated radiation therapy dosimetry with various ion detectors and Gafchromic EBT2 film

Experimental Particle Physics PHYS6011 Joel Goldstein, RAL

Injection with front ends open at the ESRF

Status Report about the TPC Detector and Module at CEPC

Independent validation of machine performance check for the Halcyon and TrueBeam linacs for daily quality assurance

The development of a plastic scintillator for radiotherapy dosimetry

Performance Requirements for Monitoring Pulsed, Mixed Radiation Fields Around High-Energy Accelerators

LASER SOLUTIONS. Ensure Accurate Setup from Simulation to Treatment. FIXED: Micro // MOVABLE: CT Sim+

NEBOSH International Technical Certificate in Oil and Gas Operational Safety

CERTIFICATE of Product Conformity (QAL1 )

Measuring Couch Motion Accuracy with X-ray Beam Fluence

Q-Series Relative Humidity and Relative Humidity & Temperature Sensors for Critical Environments and Outdoor Air

Portable Survey Instruments NISP-RP-01

Beam Loss Position Monitoring with Optical Fibres at DELTA

Development of a Novel Contamination Resistant Ionchamber for Process Tritium Measurement and use in the JET First Trace Tritium Experiment

PHI Quantera II Consumable Parts List

Technical Manual Ionisation Chamber MUSIC80

REPUBLIC OF NAMIBIA MINISTRY OF HEALTH AND SOCIAL SERVICES REPORT BY: MS VERA UUSHONA TITLE: INSPECTION OF CONVENTIONAL DIAGNOSTIC X-RAY FACILITIES

Unique Solutions for Your Research. Introduction to Velocitas VMI. Contents.

ESSENTIAL REQUIREMENTS CHECK LIST

Instrumental technique (MCP detectors) - Kamalesh ( )

SAQCC FIRE D&GS TRAINING SUB COMMITTEE COURSE CURRICULUM

Understanding total measurement uncertainty in power meters and detectors

MapCHECK 3. The New Benchmark for 2D IMRT QA. Your Most Valuable QA and Dosimetry Tools

DAWN Electronics, Inc.

Monitoring of Radiation Exposure Adjacent to a Linear Accelerator Treatment Bunker with a Highlight Window

Doc. No. SP R0. CXI Detector stage. Prepared by: Signature Date Jean-Charles Castagna Design Engineer

25 TEXAS ADMINISTRATIVE CODE Radiation Safety Requirements for Analytical and Other Industrial Radiation Machines

Product Specifications Linear Accelerator Relay Junction Box

Atomic Data and Analysis Structure

An overview of the Science and Technology Facilities Council capabilities in healthcare

Electronic Personal Dosemeter

Transducers for Medical Devices

Bunch Length Interferometer System

Reference Dosimetry of High-Energy Therapy Photon Beams with Ionisation Chambers. Recommendations No. 8, Revision 2018

Vacuum Quality Assurance in Practice

Radiation Safety issues for the PF-AR in KEK

Energy and Angular Responses of the Criticality Accident Alarm System Using Current-mode-operated Scintillation Detector

Room Sensors. Building Technologies HVAC Products ... Symaro. for relative humidity (high accuracy) and temperature

X-Ray Protection Program

Optical Measurement Techniques for LED Clusters and LED Safety

Dosimetry issues related to the SSRT project at the ESRF

Matachana Low Temperature Steam and Formaldehyde 130 LF Sterilizer

Uncertainty Assessment: Relative versus Absolute Point Dose Measurement for Patient Specific Quality Assurance in EBRT

Basics. Conversion principle and parameters of pyroelectric detectors

SIEMENS ULTRAMAT 6 IR CARBON MONOXIDE ANALYZER METHOD VALIDATION FOR TESTING CARBON MONOXIDE IN NITROGEN, NF

Canadian Partnership for Quality Radiotherapy. Technical Quality Control Guidelines. for Safety Systems at Radiation Treatment Centres

Dosimetry for the MR-linac

Aquapurification.com. Ozone. Ozonation Process: SGA Series: Ozone/Oxygen Systems:

Multiphase-Simulation of Membrane Humidifiers for PEM Fuel Cells

SYNCHROTRON RADIATION ABSORPTION AND VACUUM ISSUES IN THE IR*

LECTURE 11. Dr. Teresa D. Golden University of North Texas Department of Chemistry

Today s Outline - September 12, C. Segre (IIT) PHYS Fall 2016 September 12, / 21

Certificate: / 29 April 2014

MODEL 23 AND 23-1 ELECTRONIC PERSONAL DOSIMETER. August 2017

National Atomic Energy Commission ( NATEC ) Radiation Exposure and Laboratories General Directorate

Cl inician Training Packet

Minnesota Rules, Chapter 4732 X-Ray Revision

GUIDELINES TORONTO GENERAL HOSPITAL LOCATION. INTRODUCTION Existing College Wing Buildings and Site Plan - Figure 1A and Figure 1B.

SuperCDMS SNOLAB: Road to 100 mm Germanium Detectors

Review of options for cabin dosimetry & operation. Helen Amatiello & Colin Martin Health Physics Service Glasgow

HYDROGEN DETECTION INSTRUMENTATION FOR THE SIAC HYDROGEN RECOMBINERS. Summary

Saintech Ion Beam Systems - Ion Current Monitor

WG Container & Cargo X-Ray Scanning Portal

Carnegie Mellon University Radiation Safety Committee

INTERNATIONAL STANDARD

PM-200 Power Module Operating Instructions

Electron detectors. The most used signals in a scanning electron microscope in standard conditions are the electrons and in particular the SE.

THE NEXT GENERATION IN VISIBILITY SENSORS OUTPERFORM BOTH TRADITIONAL TRANSMISSOMETERS AND FORWARD SCATTER SENSORS

A FIRST RESPONDERS GUIDE TO PURCHASING RADIATION PAGERS

The system can be easily relocated by a standard container truck., and re-commissioned by trained engineers typically within a 72-hour period

WG Entire Truck & Cargo X-Ray Scanning Portal

Solid State Detectors. T. Bowcock

Oxygen Sensing Solutions

Mechanical Concept of the TESLA Detector

NUCLEAR INDUSTRY STANDARD PROCESS Radiological Protection. Level 3 Information Use

An investigation of the reproducibility and usefulness of automatic couch motion in complex radiation therapy techniques

Customized MCPs for Analytical Instruments. An MCP for every application

InstrumentationTools.com

Determination of the photon beam attenuation by the BrainLAB imaging couch: angular and field size dependence

the production region, magnet and 1st range region, and the 2nd range region. These were added to arrive at a totgl number of V

Model ZXL User Manual

Assessment and management of radiotherapy beam intersections with the treatment couch

NON-DESTRUCTIVE 2-D BEAM PROFILE MONITOR USING GAS SHEET IN J-PARC LINAC -overview & hardware construction-

Doc. No. SP R0. CXI Reference Laser. Prepared by: Jean-Charles Castagna Signature Date Design Engineer

LARGE VOLUME HEMISPHERICAL NUCLEAR RADIATION DETECTOR CZT/500(S)

2008 JINST 3 S The CMS experiment at the CERN LHC THE CERN LARGE HADRON COLLIDER: ACCELERATOR AND EXPERIMENTS.

MINIATURE HEAT TRANSPORT SYSTEMS WITH LOOP HEAT PIPES

Transcription:

Competency: Critically evaluate the function of the ionisation chamber in the Linear Accelerator and its importance for correct treatment delivery Trainee: TRAINEE Module: Radiotherapy Physics Competency: D2.5: Critically evaluate the function of the ionisation chamber in the Linear Accelerator and its importance for correct treatment delivery. Date of Review: 30/04/2012 Evidence: 20/04/2012 I attach a document describing the function and importance of the ionisation chambers in a linac. Files: (one attached document) Reviewer: ASSESSOR Feedback: 30/04/2012: Very good summary. Outcome: Satisfactory

Ionisation Chambers in the Linear Accelerator D2.5- Critically evaluate the function of the ionisation chamber in the linear accelerator and its importance for correct treatment delivery. Function of a general ionisation chamber Varian Clinac ionisation chambers The Ionisation Chamber s Importance for Correct Treatment Delivery References Function of a General Ionisation Chamber Ionisation chambers make use of the fact that radiation of sufficiently high energy, such as the x-ray and electron beams used in radiotherapy, can knock electrons off atoms and molecules, which is the process of ionisation. This creates ion pairs of opposite charge in what was an electrically neutral environment. An ionisation chamber can be open or closed (in a linac head they are closed), but in either case the chamber contains a sensitive volume in which ion pairs are created by the action of incoming radiation. A voltage is applied across the volume, with oppositely charged electrodes. The negative ions are attracted towards the positive electrode (anode) and the positive ions to the negative electrode (cathode), so the ion pair is split and the electrodes receive a tiny signal (current) for each pair created. The current per pair is very small, but it is still an order of magnitude easier to measure than the temperature rise in a calorimeter from the same radiation [2]. The amount of ionisation is proportional to the mass of the sensitive volume, so in an open ionisation chamber filled with air, measurements must be corrected to a standard temperature and pressure to account for density variations [2]. There is a small correction required for relative humidity, but it is so small as to be negligible under normal conditions (relative humidity between 20% and 70%), and a more significant correction for recombination of the ions (reduced by having a greater voltage) [2]. Varian Clinac Ionisation Chambers For further details, see reference [1]. There are two independent ionisation chambers found in a Varian Clinac, each consisting of two pairs of plates. The primary chamber is denoted MU1, and has plates aligned along the radial axis of the beam. For MU2, the secondary chamber, the plates are aligned along the transverse axis. In monthly tests, MU1 and MU2 are tested to check that they match. If not, they must be recalibrated. The two chambers are shown in the diagram below:

The chambers are used to monitor dose rate and symmetry. The small currents are converted to voltages and summed, forming pulses which are proportional to the dose rates on the inner plates. Each pulse is proportional to 1 cmu, so by recalibrating MU1, the ratio of cgy to MU can be set. The inner plates (e.g. A and B) are completely within the beam. o Their summed output current is therefore proportional to dose rate. o The difference in their output currents reveals beam angle symmetry errors. An interlock is generated for a difference of more than 8V. The outer plates (e.g. E and F) are partially within the beam, on its edge. o The difference in their output gives the beam position symmetry errors. For these measured angle and position errors, there are feedbacks to correct the beam. Actually, the position and angle error signals are summed together to give R and T symmetry signals which are then monitored by asymmetry level detectors. The ion chamber charge is also monitored, and excess of either charge or asymmetry generates the EXQ1 or EXQ2 interlock (primary or secondary chamber), causing instantaneous termination of the gun current. The DOS1, DOS2 and DS12 interlocks are directly controlled by the ionisation chamber inner plate measurements. DOS1 refers to the output measured by MU1, with DOS2 the equivalent for MU2. DS12 is the interlock affected by the difference in output measured by the two chambers. The pathways for signals from the inner and outer plates are shown below:

The two ionisation chambers, MU1 and MU2, are located in the treatment head of the linac. A schematic diagram of a linac shows this [3]: Linac head

The Ionisation Chamber s Importance for Correct Treatment Delivery The ionisation chambers in the linac are vital for ensuring that the correct dose is given to the patient. Complete tolerances and action levels are given in WORK INSTRUCTION, and are based on IPEM 81 [5]. I describe action and tolerance levels in more detail in reference to another evidence document produced by TRAINEE The IPEM Report 81 (chapter 1) states that a figure of 3 per cent can be taken as the currently recommended accuracy requirement on dose changes will be likely to be clinically observable for dose changes at twice this level [5]. It also states Variation in the dose distribution across the target volume may also be expected to affect treatment outcome. In other words, the values of output (dose, dose rate) and beam angular and positional asymmetry are key to the success or otherwise of radiotherapy treatment, and must be closely monitored. Incidentally, the uncertainty of up to 3% (2% is used as a tolerance level at the HOSPITAL, with 3% forcing treatment to stop) must be combined with other uncertainties, such as patient movement and change in shape, to give an overall uncertainty of around 5%. For a palliative patient, up to 10% might be permissible. While tests of output using equipment such as a Linacheck occur every day, and monthly and annual QA measures output as well as beam flatness and asymmetry, MU1 and MU2 are the only means of measuring flatness, symmetry and output for each individual patient and while treatment is in progress. As such, the existence and correct functioning of the ionisation chambers in the linac are essential to the treatment of patients. References [1] Customer Clinac HE Training Craig Phillips, Varian Medical Systems [2] National Physical Laboratory Practical Course in Reference Dosimetry, Overview of Dosimetry [3] www.emitel2.eu, accessed 19/4/12 [4] WORK INSTRUCTION [5] IPEM Report 81, Physical Aspects of Quality Control in Radiotherapy