Beam Loss Position Monitoring with Optical Fibres at DELTA

Similar documents
Beam loss Monitors for FEL using optical Fiber

MD456: Monitoring of abort gap population with diamond particle detectors at the BGI in IP 4

Fibre Optic Basics FIA Summer Seminar 2014

Injection with front ends open at the ESRF

The Optical Time Domain Reflectometry and the Fusion Splicer Laboratory exercise

Fibre Specification Standards

February 12, PHY357 Lecture 11. Experimental Methods. Accelerators. Particle Interactions. Particle Detectors. Full experiment (eg.

Radiation Safety issues for the PF-AR in KEK

Luna Technologies Webinar Series. New Rayleigh-Based Technique for Very High Resolution Distributed Temperature and Strain Sensing

Optical Time Domain Reflectometry (OTDR)

Status Report about the TPC Detector and Module at CEPC

OPTICAL TIME DOMAIN REFELECTOMETER (OTDR): PRINCIPLES

A Fresnel Reflection-Based Optical Fiber Sensor System for Remote Refractive Index Measurement Using an OTDR

The LHCb Outer Tracker: Production & Ageing studies

PC474 Lab Manual 1. Terry Sturtevanta and Hasan Shodiev 2. Winter 2012

Low-Power Detectors. Key Features. 818 Series. Power levels from pw to 2W. Wavelengths from 190 nm to 1800 nm. Proprietary detector optics

818 Series Low-Power Detectors

Radiation Safety issues for the PF-AR in KEK

SYNCHROTRON RADIATION ABSORPTION AND VACUUM ISSUES IN THE IR*

Maintenance and Operation for ECAL Monitoring (WBS4.3)

OPTICAL TIME DOMAIN REFLECTOMETRY (OTDR) Dr. BC Choudhary Professor, NITTTR, Chandigarh

PC474 Lab Manual Wilfrid Laurier University 1

Today s Outline - September 12, C. Segre (IIT) PHYS Fall 2016 September 12, / 21

Recent Progress in Lightwave Sensing Technology

The Optical Scattering Calibration System at SNO+ IOP 2015

Quality Assurance and Quality Control of Optical Fibres for CERN

818 Series. Low-Power Detectors ABSOLUTE SPECTRAL RESPONSIVITY (A/W)

Diamond as potential CryoBLM for the LHC All results are preliminary

Experimental Particle Physics PHYS6011 Joel Goldstein, RAL

GAMMA OTDR application consists of main window and menu. Using menu user can operate in different modes of application.

Minimizing radiation and beam losses at the ESRF

RAMAN SCATTERING AND OTDR BASED DISTRIBUTED SENSOR FOR TEMPERATURE MONITORING APPLICATIONS

QUALITY ASSURANCE AND SAFETY AT A TID RADIATION TEST LABORATORY

Customized MCPs for Analytical Instruments. An MCP for every application

Performance of the Monitoring Light Source for the CMS Lead Tungstate Crystal Calorimeter

Diamond Detectors Ltd. Fabrication and Packaging Capabilities. Kevin Oliver

Development of a new X-ray source system using ultraviolet laser for medical treatment

The Flash Beam Loss Monitor System

Laser Safety and Classification Full Version (CA-1110) Advanced Version (CA-1111) Basic Version (CA-1112)

Our photo detectors are offered for both power or energy measurements. Measure as low as a few femtojoules in energy or a few picowatts in power.

50(394) IEC. ore content meter ore sorting equipment overvoltage (of a Geiger-Müller county tube)...

Bunch Length Interferometer System

Laser Damage Measurements

Our photo detectors are offered for both power or energy measurements. Measure as low as a few femtojoules in energy or a few picowatts in power.

Distributed Temperature Sensing

Omnisens DITEST TM FIBER OPTIC DISTRIBUTED TEMPERATURE & STRAIN SENSING TECHNIQUE

M210 Multifunction Micro OTDR

LECTURE 11. Dr. Teresa D. Golden University of North Texas Department of Chemistry

FLX380-30x FlexTester OTDR

M210 Multifunction Micro OTDR

Detection of Solvents using a Distributed Fibre Optic Sensor

UC INSTRUMENTS GM8036 LASER SWEEP OPTICAL SPECTRUM ANALYZER (1525 ~ 1610 nm) Technical Specifications v1.01 March, 2011

An FT-NIR Primer. NR800-A-006

Optical Return Loss Measurement by Gregory Lietaert, Product Manager

M700-Series Multifunction OTDR

50 W Laser Concepts for Initial LIGO

OTDR - FHO5000-xxx. Optical Time Domain Reflectometer

MX1A and MX1B: Laser Displacement Detection Sensors. 85% to 110% of the rated voltage MX1A: 15VA, MX1B: 25VA

Competency: Critically evaluate the function of the ionisation chamber in the Linear Accelerator and its importance for correct treatment delivery

Sensing Technology. A company of the BRUGG Group

Link loss measurement uncertainties: OTDR vs. light source power meter By EXFO s Systems Engineering and Research Team

Performance Requirements for Monitoring Pulsed, Mixed Radiation Fields Around High-Energy Accelerators

Vacuum Quality Assurance in Practice

OTEB-CO-B EDFA WITHOUT SNMP INSTRUCTION MANUAL

高等食品分析 (Advanced Food Analysis) II. INSTRUMENTS FOR OPTICAL SPECTROSCOPY *Instruments for optical spectroscopy: Optical spectroscopic methods are

Interferometric optical time-domain reflectometry for distributed optical-fiber sensing

Fiber optic system for monitoring large earth structures

TSI AEROTRAK PORTABLE PARTICLE COUNTER MODEL 9110

Journal of Applied Science and Agriculture. Improve the Spatial resolution of distributed optical sensors by designing a new detection scheme

Application Note. Which characterization method is best for your application? The Future of Thermal Imaging is Here!!!

Fibre Optic Sensors: basic principles and most common applications

CONTRACTOR WHS HAZARD STANDARD LASER SAFETY EXTERNAL USE ONLY

Status Report of Active Space Radiation Detector, A-DREAMS-2 at NIRS

LCTPC Setup at the DESY Testbeam. AWLC 14, Fermilab , R. Diener, DESY

ASSMANN ELECTRONIC GmbH ASSMANN SYSTEM GUARANTEE. DIGITUS Professional Fiber Optic Cabling System. Appendix B - Provisions for acceptance

Health Monitoring System Using FBG-Based Sensors for a 12-Story Building with Column Dampers.

Ionisation Chambers Containing Boron as Neutron Detectors in Mixed Radiation Fields

APPLIED. OMA-300 Hydrogen Sulfide Analyzer H 2. Ex p. Ex d. The world s safest continuous H 2. S analyzer. The OMA continuously measures H 2

DEGRADATION OF OPTICAL FillERS AT CARBON-CARBON PYROLYSIS. Rockwell International Science Center P.O. Box 1085 Thousand Oaks, CA 91358

Office of Nonproliferation and Verification Research and Development University and Industry Technical Interchange (UITI2010) Review Meeting

VACUUM PERFORMANCE IN THE MOST RECENT THIRD GENERATION SYNCHROTRON LIGHT SOURCES

Overview of fiber optic sensing system: BOTDR and its applications

Chapter 1. Introduction

Long-Range Displacement Sensor Based on SMS Fiber Structure and OTDR

Y003 Multi-offset Seismic Acquisition Using Optical Fiber Behind Tubing

OPTICAL FIBER SOLUTIONS

M310 Enterprise OTDR. Designed for Enterprise Network Testing, Troubleshooting and Documentation. or (800) , (603)

Distributed sensing based on reflections and Rayleigh backscatter. Brian Culshaw University of Strathclyde Glasgow Scotland

Effect of Different Types of Fabricated Anti-scatter Grids in Reducing Backscattered Radiation

Field Test Procedure for Optical Fibre Link Measurements

HiRadMat facility at the CERN SPS

Gas Temperature Measurements with High Temporal Resolution

Instruments Designed for POF & Fibre Optic Cable Testing

A CONCEPT FOR CANCELLING THE LEAKAGE FIELD INSIDE THE STORED BEAM CHAMBER OF A SEPTUM MAGNET*

Noyes OFL 250B Specs Provided by

WHAT IS LASER ULTRASONIC TESTING?

Broadband System - H

MADE FOR THE TRADE! FIBER OPTIC. professional. XL fibertools Series. instruments

Co-60 irradiator for therapy level calibrations at SSDLs

INTERNATIONAL STANDARD

Transcription:

Beam Loss Position Monitoring with Optical Fibres at DELTA Frank Rüdiger M. Körfer (DESY) W. Göttmann (HMI Berlin) G. Schmidt (DELTA) K. Wille (DELTA) 24. June 2008 Frank Rüdiger 1

Table of Content 1. Introduction and Motivation 2. Optical Time Domain Reflectometry 3. Transmission Measurement with Optical Fibres 4. Detection of Cerenkov-Radiation 5. Summary Frank Rüdiger 2

1. Overview of DELTA 1.5 GeV 130 ma lifetime:10h 115 m Frank Rüdiger 3

1. Radiation Sources at DELTA The DELTA vacuum chamber is made up of 3 mm V4A steel The synchrotron radiation is almost completely absorbed inside the chamber wall Beam loss electrons colliding with the chamber generate electro-magnetic cascades. Even at small incident angles a significant amount of the shower particles can leave the vacuum chamber 1.5 GeV beam loss electrons are the main source of ionising radiation at DELTA Frank Rüdiger 4

1. Challenges for Dosimetry Systems / Advantages of Fibre Optical Systems Measurements in narrow spaces small fibre diameter Surveillance of large distances complete surveillance of several kilometres High-dose measurements up to 1000 Gray Evaluation during beam operation possible within a few minutes System with two nanoseconds time resolution is available Frank Rüdiger 5

1. Radiation Induced Attenuation of Optical Fibres Chemical bonds are split up in the fibre by exposure to radiation (radiolysis) Example: Si-OH Si-O + H 0 The generated defects are called colour-centres Transitions between the generated states attenuation of injected light intensity Frank Rüdiger 6

1. Radiation Induced Attenuation of Optical Fibres Ind. attenuation [db/km] Transmission OTDR wavelength [nm] The sensitivity of a fibre optic radiation sensor system can be chosen by wavelength selection Frank Rüdiger 7

1. Characteristics of a Fibre Optic Radiation Sensor System Linearity between dose α D and attenuation D α D [ db/m] f 1 f = c( λ) D valid for a dose range of 10-3 to 10 3 Gy Annealing: loss of accumulated dose information due to regeneration processes of the colour centres. Enhancement of dose information lifetime by suitable fibre doping The attenuation is independent of dose-rate, temperature and light intensity Frank Rüdiger 8

2. Optical Time Domain Reflectometry (OTDR) Used for radiation dose measurement around the complete length of the storage ring. Resolution about 3 Gray Measurement of the Rayleigh backscattered part of the injected light intensity The time dependent signal is converted to the position of beam loss; spatial resolution ca. 60 cm Frank Rüdiger 9

2. OTDR: Hardware Tektronix TFP2A Selectable wavelength (850 & 1300nm) Pulse lengths 1ns, 3 ns, 8 ns, Used Fibres: Multi-mode fibres with core diameter of 50 µm Germanium doped and co-doped with Phosphorus Frank Rüdiger 10

2. OTDR: Dose Distribution at DELTA (period: 15 weeks) Diploma thesis Manuel Benna (DELTA) Frank Rüdiger 11

3. Transmission Measurement: Motivation Radiation effects of the permanent magnet undulator U55 Magnets are composed of Neodymium-Iron-Boron alloy High radiation doses result in an irreversible demagnetization (based on investigations at the ESRF) Limiting dose value: 60 kgy limiting dose rate value for the U55: 0.8 Gy/h planned frequent injection mode will increase the radiation dose permanent dose surveillance of the U55 needed Frank Rüdiger 12

3. Transmission Measurement: Principle and Setup Laser P=20 mw l=660 nm Frank Rüdiger 13

3. Transmission Measurement: Hardware Powermeter: Newport Type: PTS-FOPM Detectable intensity range: 1pW 2W Used wavelength: 660 nm The system has been calibrated using thermoluminescence dosimeter-rods accuracy is about 30 %, similar to the TTF-system at DESY Frank Rüdiger 14

3. Results: Measured Fibre Intensity and Beam Current Frank Rüdiger 15

3. Results: Dose Rates During Standard User Operation Measurement during 60 hours at 1.5 GeV during standard user operation; beam lifetime: 4h @ 100 ma Gap is opened to maximum during beam injection mode U55 entrance (upper side) U55 entrance (lower side) Dose rate [Gy/h] 0.81 ± 0.23 0.16 ± 0.05 Limiting value already reached at the upper side The dose at the lower side is by a factor 5 smaller Frank Rüdiger 16

3. Results: Dose Rates during Frequent Injection Mode Measurement during 2 hours at 1.5 GeV, frequent injection U55 entrance (vacuum chamber) Dose rate [Gy/h] 4.23 ±1.23 the limiting value is exceeded by factor 5 First measure: Optimisation of the lead shielding in front of the U55 dose rate lowered by factor 2 sufficient for standard beam operation redesign of the lead shielding until start of frequent injection mode Frank Rüdiger 17

4. Cerenkov-Light Detector: runtime measurement to localise beam losses Controller ADCcard Power supply to ADC DAC Amplifier circuit Photomultiplier Adapter Ethernet /RS485 Frank Rüdiger 18

4. Cerenkov-Light Detector: Principle of Measurement Runtime-measurement to localise beam losses detector is connected to the upstream-side Time resolution: 2 ns given by sampling rate of the ADC (1 GS/s) Spatial resolution: 0.24 m in longitudinal direction Used wavelength range: 500 nm 650 nm (maximum at 550 nm) Multi-mode-step-index fibres (core diameter: 300 µm) consisting of undoped silicon dioxide with high content of OH - ions Frank Rüdiger 19

5. Summary: Fibre optic radiation sensor systems are well suited for accelerators: - usable in narrow spaces - dose range is up to 1000 Gy - evaluation during beam operation OTDR: - used for dose surveillance of the complete Delta vacuum chamber - dose resolution: 3 Gy Transmission measurement: - used for dose surveillance of the U55 permanent magnet undulator - dose resolution: 60 mgy Cerenkov-light detector: - system has been installed and functionality has been proven - will be used for increase of injection efficiency - real-time beam loss position monitoring with single bunch resolution of 2 ns Frank Rüdiger 20

Dr. Markus Körfer (DESY) Walter Göttmann (HMI Berlin) Prof. Dr. Wille (DELTA) Prof. Dr. Weis (DELTA) Dr. Gerald Schmidt (DELTA) Holger Huck (DELTA) Acknowledgement Frank Rüdiger 21

References: Frank Rüdiger, diploma thesis, Aufbau und Einsatz von Glasfaser-Dosimetriesystemen an der Speicherringanlage DELTA, 2008 Manuel Benna, diploma thesis, Strahlverlustmessung und Dosimetrie am DELTA, 2006 S. Girard, A. Boukenter, et al., Properties of phosphourus-related defects induced by γ-rays and pulsed X-ray irradiation in germanosilicate optical fibres, 2003 D. Griscom, E. Friebele, Fundamental radiation-induced defect centers in synthetic fused silicas: Atomic chlorine, delocalized E' centers and a triplet state, 1986 H. Henschel, M. Körfer, J. Kuhnhenn, U. Weiland, F.Wulf, Fibre optic radiation sensor systems for particle accelerator, Nuclear Instruments and Methods in Physics Research A 526, 2004 H. Henschel, O. Köhn, H.U. Schmidt, Optical Fibres as radiation dosimeters, Nuclear Instruments and Methods in Physics Research B 69, 1992 G. Schmidt et al., Optical Fibre Beam Loss Monitors for Storage Rings at DELTA, Proceedings of EPAC, 2002 W.V.Hassenzahl, T.M. Jenkins et al., An Assessment ot the Radiation on Permanent-Magnet Material in the ALS Insertion Devices, Nuclear Instruments and Methods in Physics Research A291, 1990 G. Schmidt et al., Proposal for a Frequent Injection Mode at DELTA, Proceedings of EPAC, 2004 H. Henschel et al, Optical Fibre Dosimeter For SASE FEL Undulators, DIPAC, 2003 W. Göttmann, M. Körfer, J. Kuhnhenn, Beam Loss Position Monitoring using Cerenkov Radiation on Optical Fibers, Proceedings of DIPAC, 2005 Frank Rüdiger 22