ATOMLAB XENON TRAP MONITOR OPERATION AND INSTALLATION MANUAL FN: /12

Similar documents
AM-806. Waste/Door Radiation Monitor OPERATIONS MANUAL

LEAD-LINED LABORATORY FURNITURE

RAM GENE / RAM GENE-1

Installation. Built-in Full Height Wine Cellar VCWB301

Victoreen Primalert 35 Area Radiation Monitor

A1UL PERS. Personal Emergency Response System. For Technical Support Please Contact Your Service Provider Or Distributor

AIR or STACK MONITOR NOBLE GAS, GROSS BETA-GAMMA PARTICULATE, & IODINE. Optional Alpha, Tritium, Carbon-14

Safety Siren Pro Series3 Radon Gas Detector

Viking Installation Guide

Local air conditioner

LUDLUM MODEL ALPHA, BETA, GAMMA HAND DETECTOR. June 2016 Serial Number and Succeeding Serial Numbers

INSTALLATION AND OPERATION INSTRUCTIONS FOR WALL-MOUNT AND BUILT-IN UNITS

PORTABLE ISOTOPE IDENTIFIER Search Tool / Sample Counting System

VPMmini. Operator's Manual

TABLE OF CONTENTS. NOTE: Read the entire instruction manual before starting the installation. TROUBLESHOOTING... 13

RAM GENE-1 ERK. Contamination & Radiation Meter. Operating Manual. Document #PRIR90N2.DOC Version 2.2 October 2009

Operations Manual. Part Number: For Research Use Only

Multiple use Digital Thermometer With 7 Day Memory

Warmette Blanket Cabinet Warmette Dual Cabinet. User Manual

690 Power Fishing System

SW-2 SNOW MACHINE. Item ref: UK User Manual

Operating & Maintenance Manual. Alert-4 Ethernet LCD Master Alarm

LUDLUM MODEL 44-9 ALPHA, BETA, GAMMA DETECTOR. February 2010 Serial Number PR and Succeeding Serial Numbers

LUDLUM MODEL AND LUDLUM MODEL ALPHA, BETA, & GAMMA DETECTORS

NEX-BETA-ABG FACILITY Drinking Water Safety Monitor

Pioneer-R16 Gas Monitor Operator s Manual

PARTS AND SERVICE MANUAL FOR MERCO HOLDING CABINET MODEL 86002

Halton SAFE / 7.14 user guide and installation instructions

CLASSIC PLUS INFRARED ZONE HEATER

Installation Manual CARBONATOR With Plain-Water Booster

INSTALLATION AND OPERATION INSTRUCTIONS FOR ZERO CLEARANCE AND INSERT UNITS

OWNER S MANUAL MODEL BW4B SLICER WARMER

Ranger EXP RADIATION A L E R T

HEDMAN DI-50. Endorser Instructions. Hedman DI-50 Operators Guide

INSTALLATION AND OPERATION INSTRUCTIONS FOR

WatchDog Wireless Crop Monitor Operation Manual

SpotOn Temperature Monitoring System. SpotOn. Temperature Monitoring System. Model 370 Installation and Service Manual

Contamination Meter Users Manual

INFRARED ZONE HEATING SYSTEM

INFRARED ZONE HEATING SYSTEM

PR-L2466W- PA. Operating Instructions. High Performance Refrigerator PR-L2466W-PA

LUDLUM MODEL 44-7 Alpha, Beta, Gamma Detector. July 2012 Serial No. PR and succeeding Serial Numbers

Instruction Book for HD Punch Machines

Important safety instructions

INSTALLATION AND OPERATION INSTRUCTIONS FOR BI-DEEP UNITS

Important safety instructions

INSTALLATION INSTRUCTIONS UNDERCOUNTER DISHWASHERS

2364 Leicester Road, P.O. Box 175, Leicester, NY Phone (585) Fax (585)

INFRARED QUARTZ HEATER

1. Quick start-up guide

SCN60 Residential Nugget Ice Machine

AERIAL RADIATION DETECTORS

! WARNING To avoid risk of electrical shock, personal injury or death; disconnect power before servicing, unless testing requires power.

ASG EZ-9000GR Tape Dispenser User Manual ASG #66136

A few simple steps to ensure perfect water, every time.

1500W Black Glass Portable Electric Panel Heater KAHTP20BLKA

! WARNING To avoid risk of electrical shock, personal injury or death; disconnect power before servicing, unless testing requires power.

# TFB-S-120, TFB-L-120 TFB-S-220, TFB-L-220 FWS-S-120, FWS-L-120 FWS-S-220, FWS-L-220

11,000 BTU Portable Air Conditioner with dehumidifier & Fan PE4-11R-03 Operating Instructions. Model No. PE4-11R Soleus Air International

FD 2052 AutoSeal 12/2007 OPERATOR MANUAL FOURTH EDITION

INFRARED QUARTZ HEATER

D1255RB/D1256RB/D1257RB

Test Equipment Depot Washington Street Melrose, MA TestEquipmentDepot.com HCT HCT-1000 User Guide

Identify the following features and specifications for the ESP1

HOT FOOD DISPLAY Technical Service Manual

690 and Power Fishing Systems

Soldering Station INSTRUCTION MANUAL

MEGRIND55. Electric Meat Grinder Assembly & Operating Instructions

Contents. Product Description. Features. Quick Guide. Touch Panel Operation. Instruction. How to change the filter. Warning.

Local Air Conditioner

Instructions for use

LUDLUM MODEL 25 AND 25-1 PERSONAL RADIATION MONITOR USER S MANUAL. October 2014 Serial Number and Succeeding Serial Numbers

DM-1 Drying Monitor U S E R G U I D E UGD

Bose Home Theater Speakers. Owner s Guide

Operation & Maintenance Manual

Dyna-Form Air. Alternating Pressure Relief Overlay System

Obturator Oven Instruction Manual

INFRARED ZONE HEATING SYSTEM

INSTALLATION AND OPERATION INSTRUCTIONS FOR WALL-MOUNT AND BUILT-IN UNITS

Coolant Heater Thermo Top (TTC) Heavy-duty Operating Instructions Installation Instructions

Contents. Product Description. Features. Quick Guide. Touch Panel Operation. Instruction. How to change the filter. Warning.

WIRELESS ALARM SYSTEM WITH TELEPHONE AUTO DIALER

MODEL AND MODEL GAS PROPORTIONAL DETECTORS. December 2016 Serial No and Succeeding Serial Numbers

DWYER INSTRUMENTS, INC. P.O. BOX 373 MICHIGAN CITY, INDIANA 46360, U.S.A. Series 670 Hood Monitor. Bulletin AV-670

Installation Instructions. Fire Command Center/Fire Alarm Annunciators D1256/D1257

WARNING: Warns of health hazards and identifies possible risks of injury. CAUTION: Indicates possible dangers to the machine or other objects.

SimpleVIS Viscometer Reference Guide P

Instruction Manual for FORMALDEHYDE GAS DETECTOR MODEL FP-30/FP-40

Operator s Manual CAYENNE HEAT N SERVE 4/3 RECTANGULAR RETHERMALIZER ENGLISH

Table of Contents DIGITAL DRYING OVEN. Features. Setup. Cautions. Description of the parts. Temperature Controls. Maintenance and Storage

Troubleshooting Thermo Scientific LevelPRO Gauges

Operator Guide. Publication No. OI /97-Eng. KODAK Automixer II Plus

HIn3550 Magnetic Field Monitor User's Manual

Instructions for use

Bench Top Ionizer Installation, Operation and Maintenance

Beacon 200 Gas Monitor Operator s Manual. Part Number: RK Released: 6/6/08

Set. Set. Set. Set. Set. Set. Set. Sensor Pads 45 Day 90 Day 180 Day 1 Year 2 Years. Set. Set CAUTIONS

INSTALLATION INSTRUCTIONS

Installation Instructions

INVERTER SPLIT - TYPE

Transcription:

ATOMLAB XENON TRAP MONITOR OPERATION AND INSTALLATION MANUAL 136-755 FN: 12-148 5/12

This manual contains installation and operation procedures for the following product: 136-755 Atomlab Xenon Trap Monitor Authorized European Community Representative: PROTHIA 26, rue serpollet 75020 Paris, France (011) 33.1.40.31.60.20 2 Biodex Medical Systems, Inc. 2012

TABLE OF CONTENTS Page 1. IMPORTANT SAFETY INFORMATION - WARNING, CAUTION AND INFORMATION SYMBOLS... 4 2. BEFORE PROCEEDING... 5 3. INTRODUCTION... 6 4. INSTALLATION INSTRUCTIONS... 6 - GENERAL INSTALLATION... 6-8 5. OPERATION... 8 - GENERAL... 8 - BACKGROUND... 9 - CHECK SOURCE... 9-10 6. USING THE XENON TRAP MONITOR... 10 - CHANGING CHECK SOURCE DISK... 11-12 - RECORD KEEPING... 11 - MAINTENANCE... 11 7. SPECIFICATION... 12 8. TROUBLESHOOTING INFORMATION... 12 APPENDIX A SAMPLE OF COMPLETED XENON MONITOR TEST SHEET... 13 APPENDIX B BLANK XENON WORKSHEET (MAKE COPIES FOR FACILITY USE)... 14 APPENDIX C BLOCK DIAGRAM... 15 ATOMLAB XENON TRAP MONITOR 3

1. IMPORTANT SAFETY INFORMATION WARNING, CAUTION AND INFORMATION SYMBOLS The following caution and warning symbols, and their associated meanings, are used throughout this manual: Refer to Instruction Manual/Booklet General Warning Sign General Mandatory Action Sign Follow Operating Instructions Waste in Electrical Equipment Dangerous Voltage Caution Date of Manufacture Follow the unpacking and assembly instructions document. Before using this device, read the entire Operation Manual carefully. Failure to read the manual may result in user error or inaccurate data. Be sure to save all provided documents for future reference. Make certain to understand all warning and caution labels as explained in the Before Proceeding section of this manual. The Atomlab Xenon Trap Monitor should be used only as specified in the Operation Manual. Reference Cleaning and Maintenance on Page 11 Reference Specifications for Atomlab Xenon Trap Monitor on Page 12 Operation for 115 VAC, 60 Hz. Only use approved power supplies. 4 Biodex Medical Systems, Inc. 2012

2. BEFORE PROCEEDING SPECIFIC CAUTIONS If the equipment is used in a manner other than specified in this Operation Manual, the protection provided by the equipment may be impaired and results could be compromised. The display operates on low voltage and the power supply. The detector steps up the current to have high voltage in order to operate the detector. NEVER open the detector. This should be done only by experienced, qualified service personnel. The unit contains lead for shielding the detector from background radiation. The lead should be disposed of in accordance with local and federal regulations when disposing of the unit. A Cs-137 source should be used as the check source each day the monitor will be used. ATOMLAB XENON TRAP MONITOR 5

3. INTRODUCTION The Xenon Trap Monitor is designed to act as a Xenon-133 trap monitor and to provide an audio/visual alarm when the concentration in the xenon trap exhaust port exceeds 99 pci/ml. NOTE: NRC and state regulatory commissions restrict Xe-133 concentration in restricted areas to less than 10 pci/ml. If the trap were releasing Xenon-133 with a concentration of 99 pci/ml, assuming the trap is turned off at the end of each study, and with an assumed release of 50 L of exhaust gas (10 minutes washout at 5 L/min), the room must have a minimum volume of 0.5 cubic meters to meet regulatory requirements. Typical rooms used for a xenon study have a larger volume than this. A typical room is 10 x 10 x 8 and contains 22.56 m 3. Thus the threshold level could be set much higher. However, the main function of the Xenon Trap Monitor is to alert the user when the trap must be replaced. Once xenon is coming out the exhaust of activated charcoal traps, the concentration in the exhaust port grows exponentially in a very short amount of time. The threshold level of 99 pci/ml is chosen because this is a level well above the concentration from a good trap. If the exhaust trap reaches a concentration level of 99 pci/ml, then the trap is near failure and should be monitored for replacement. 4. Installation Instructions General Installation The Xenon Trap Monitor comes factory-calibrated and has a fixed alarm threshold level of 99 pci/ml. To install the Xenon Trap Monitor: NOTE: Skip to the next section if installing your Xenon Trap Monitor on a Biodex Pulmonex System. 1. Position the Xenon Trap Monitor in a stable position on your Xenon System (attached to a shelf is preferable). 2. Connect the supplied 22 mm exhaust hose to the monitor hose fitting and to the external exhaust port of your xenon trap. (See Figure 1). 3. Plug the AC adapter into a 110 VAC socket, and connect the opposite end of the cord through the hole in the back of the unit into the socket. All the LED lights will turn on and a tone will be heard as the unit goes through self-test diagnostics for about five seconds. If all the self-test functions complete without error, the word PASS will be displayed. Figure 1. Connection of exhaust hose. 6 Biodex Medical Systems, Inc. 2012

Installing the Atomlab Xenon Trap Monitor on the Pulmonex System The Xenon Trap Monitor comes factory-calibrated and is already set for an alarm threshold level of 99 pci/ml. To install the Xenon Trap Monitor on an Atomlab Pulmonex System, refer to the following instructions. (See Figure 2). NOTE: On older Pulmonex units, you will need to plug the Xenon Trap Monitor power cord into a wall socket instead of the Pulmonex. NOTE: Unplug the Pulmonex System from the wall socket before beginning the installation procedure. 1. Peel off the backing from the double-sided tape on the side of the Atomlab Xenon Trap Monitor bracket. 2. Attach the bracket to the handle of the Pulmonex System, as illustrated in Figure 3. The slot in the bracket fits around the Pulmonex handle. When attaching the bracket, be careful to hook it around the handle before allowing the tape to come in contact with the Pulmonex side panel, tilting the Xenon Trap Monitor bracket toward the Pulmonex will help. The long cutout end of the bracket top goes toward the patient side of the Pulmonex. Use the part of the handle closest to the patient. NOTE: When installed, the top of the Xenon Trap Monitor display should be flush with the top of the Pulmonex System. 3. Attach the Xenon Trap Monitor exhaust hose from the port on the bottom of the monitor to the exhaust port on the back panel of the Pulmonex System. 4. Open the large door on the Pulmonex and plug the supplied power pack into the plug socket. The plug socket is located at the bottom of the Pulmonex, on the top of the transformer box. (See Figure 4). Figure 2. Installation of Atomlab Xenon Trap Monitor. DO NOT plug anything else into this socket. Figure 3. Attach bracket to Pulmonex handle. NOTE: Older Pulmonex Systems do not have this plug socket. For older units, plug the power pack into an external plug socket. ATOMLAB XENON TRAP MONITOR 7

5. Remove the black cap plug, located below the bracket, from the hole in the side of the Pulmonex. 6. Run the small plug-end of the power wire through the hole directly beneath the Xenon Trap Monitor bracket, in the side of the Pulmonex cabinet. 7. Connect the small plug through the hole in the back of the Xenon Trap Monitor into the socket on the Xenon Monitor. Figure 4. Power pack 8. Set the proper length of power cord, install the new strain-relief bushing onto the cable and install in the hole in the Pulmonex cabinet. 9. Plug the Pulmonex back into the wall socket. The power should come ON. All of the LED lights will turn on and a tone will be heard from the Xenon Trap Monitor as it goes through self-test diagnostics for about five seconds. If all of the self-test functions complete without a problem, the word PASS will be displayed. 5. Operation General 1. Press, then release the <On/Off> button to toggle the power on or off. Self-test diagnostics will be run every time the unit is turned on. All lights will be turned on for the user to ensure that they all work. Wait for the word PASS to be displayed after turning the unit on. 2. Press, then release the <Next/Mute> button repeatedly to select the available functions. The last selected function will begin two seconds after the last press of < Next/Mute >. The count function will not be enabled until both the background and check source functions have been completed. Any displayed value will be the successful results of that function. 3. Selecting and running a function will not overwrite the prior successful results, if the function later fails. Press the <Next/Mute> button to select the background and check source functions and see the prior good results. You do not have to worry about loosing them. 4. Anytime the alarm tone is activated press, then release the <Next/Mute> button to mute the alarm for five minutes. 5. The <Next/Mute> button will not allow you to change functions when there is an active alarm. It only works as a five-minute mute when there is an active alarm. You must turn the power off and then on again to enable normal functionality after an alarm. NOTE: This allows the user to finish their xenon study even if the alarm rings and still see the maximum count. If using a worksheet, you can still record the maximum count. 6. Self-test diagnostics will run on power up and then run continuously in the background. Should there be a failure at any time, the alarm will sound and the word FAIL will be displayed alternately with the failure code. (See Figure 5 for error codes). 8 Biodex Medical Systems, Inc. 2012

Error Description Codes 1 Internal logic error 2 GM tube shorted 3 Internal high Voltage supply too low 4 Internal high Voltage supply too high 5 External power supply Voltage too low 6 External power supply Voltage too high 7 GM tube inactive 8 Internal reference Voltage too low 9 Internal reference Voltage too high Figure 5. Error codes Background A background measurement must be taken before the Xenon Trap Monitor will be able to operate properly. NOTE: New background reading must be made each time power to the Xenon Trap Monitor is turned ON. When power is removed, a new background reading is always required before a count can be made. To measure background: 1. Make sure that the Xenon Trap Monitor is powered on and there are no radioactive materials nearby. 2. Press, then release the <Next/Mute> button until the background LED light turns either green or blue. After two seconds the LED light will turn white and the unit will begin a background check. Once the background measurement is complete, the unit displays the background count and the background LED light will turn green. 3. If the background radiation count is too high, the background LED light will blink red and the count function will not be available until the background check is run again with a good count. Check for nearby radioactive isotopes. NOTE: The background check will take one minute to complete. Check Source A check source measurement must be taken before the Xenon Trap Monitor will be able to operate properly. NOTE: A new check source reading must always be made each time power to the Xenon Trap Monitor is turned ON. To measure the check source disk: 1. Make sure that the unit is powered on and the check source disk is placed in the holder on the left side of the Xenon Trap Monitor. Always use the same check source disk in the same orientation. 2. Press, then release the <Next/Mute> button until the check source LED light turns either green or blue. After two seconds the LED light will turn white and the unit will begin a check source count. Once the check source measurement is complete, the unit displays the check source count and the check source LED light will turn green. ATOMLAB XENON TRAP MONITOR 9

NOTE: The check source will take 30 seconds to complete normally, but will take five minutes the first time the unit is used. This five-minute reference measurement needs to be done only once for a particular check source, and is needed to obtain enough statistical information for subsequent readings. NOTE: You must always use the same check source disk for this test. If you loose or change the disk, you must re-run the five-minute reference function (see: Changing Check Source Disk). 3. The check source reading must be within about 20% of the expected value. If the count is either too high or low the check source LED light will blink red and the count function will not be available until the check source is run again with an acceptable count. Make sure you are using the same check source disk each time, in the same orientation. NOTE: The Cs-137 check source disk will loose 20% of its activity in 10 years. As this happens the check source measurement will fail with increasing regularity. You will need to perform the Changing Check Source Disk operation every few years to correct for this decay. This procedure resets the Starting Norm for the Cs-137 source. 6. Using the Atomlab Xenon Trap Monitor Following the background and check source measurements, the Xenon Trap Monitor should be fully functional. Follow the procedure below to begin operation. NOTE: Remove the check source from the holder and put it away. 1. Press, then release the <Next/Mute> button until count LED light turns blue. If you are unable to select the count function, make sure that you have already done the background and check source functions as outlined above. You can not run the count function until both the background and check source functions have passed. 2. After two seconds the LED light will turn green and the unit will begin measuring. When the measurement is above the background level it will be displayed numerically in units of pci/ml. Otherwise the numeric display will show four dashes ----. 3. If a measurement is made which exceeds the threshold value of 99 pci/ml, the device will go into alarm mode. The alarm will continue to sound until either the <Next/Mute> button is pressed to mute the alarm for five minutes, or the power is turned off. Measurements will continue to take place and the peak value detected will be displayed until power is removed. The count LED light will blink red as long as measurements are above the threshold, but will be a steady red when they go below the threshold. NOTE: Many locations may require you to record the background, check source and maximum failure count readings. The unit displays the maximum reading for each function, except a passing count. Changing Check Source Disk In the event that a new check source disk is to be used, the reference value needs to be measured again. This is a five-minute reading to obtain enough statistical information to accurately characterize the check source disk. Normally this is only done once after the Xenon Trap Monitor is installed. Failure to perform this operation after changing the check source disk will result in a high failure rate during the check source function, or the inability to pass that test at all. 10 Biodex Medical Systems, Inc. 2012

1. Make sure that the unit is powered off and the check source disk is placed in the holder on the left side of the Xenon Trap Monitor. The Cs-137 disk must have at least 8µCi of actual activity. NOTE: A Cs-137 check source disk with an actual 10µCi activity when new will not have 8µCi activity remaining within ten years. 2. Press and hold the <Next/Mute> button firmly. While holding the <Next/Mute> button, then press and hold the <On/Off> button. Keep holding both buttons for approximately six seconds. 3. The check source LED light will turn blue and the other LED lights will turn red to indicate this special measurement function is selected. Release both buttons. 4. The number displayed (if any) is the reference count stored from the prior measurement. A clicking noise will be heard to indicate the check source activity. 5. After two seconds the check source LED light will turn white and the new check source disk will be measured. 6. If there is no disk present this function will fail within five seconds and the check source LED light will blink red. Place the check source disk in the holder, and then press the <Next/Mute> button repeatedly until the check source LED light turns blue again. 7. The measurement will complete after five minutes and the count will be recorded and displayed. If the measurement is good the check source LED light will turn green, otherwise it will blink red. 8. If the measurement is good it will be stored in the Xenon Trap Monitor memory. Turn off the power by pressing the <On/Off> button. The unit is now ready for normal use. Record Keeping After installing the Xenon Trap Monitor, a 10.0 microcurie Cs-137 check source (#101-103) must be used to record a check source measurement. Periodic measurements will produce a log over time, which will indicate the device is operating properly. This log can also include trap measurement results. A new trap should not cause an alarm during a ventilation study. This can be recorded as a Pass Level. A dirty xenon trap will cause values to be displayed that indicate xenon is passing through. These values can be recorded as a Fail Level. The Xenon Trap Monitor captures the highest measured fail level and displays that measurement until powered off. When using the check source, place it in the holder on the side of the Xenon Monitor. Press, then release the <Next/Mute> button until the check source LED light turns either green or blue. After two seconds, the LED light will turn white and the unit will begin a check source count. Once the check source measurement is complete, the unit displays the check source count and the check source LED light will turn green if a Pass Level or blink red if a Fail Level is counted. Record the number of counts. Maintenance As required, wipe the Xenon Trap Monitor with a clean, damp cloth. You can use alcohol to clean the outside of the Xenon Monitor. ATOMLAB XENON TRAP MONITOR 11

7. Specifications Dimensions: Lead Shielding:.5" thick (1.3 cm) Input: 22 mm hose adapter Detector: Halogen Quenched GM Tube, 1.1 cm dia, 2 mg/cm 2 mica window Voltage: 500 volts regulated Buttons: On/Off, Next/Mute Display: Four-digit LED, four RGB function LEDs (Self Test, Background, Check Source, Count), one power LED Readings: pci/ml or counts Speaker: Internal, beeps for alarm Background Count Time: 1 minute Power: 18 volts, UL approved, external; 115 VAC power adapter Weight: 5.25 lb (11.55 kg) 8. Troubleshooting Information Self-test functionality is run during power up, and continuously in the background. Should there be a failure at any time, the alarm will sound and the word FAIL will be displayed alternately with the failure code. Error Description Codes 1 Internal logic error 2 GM tube shorted 3 Internal high Voltage supply too low 4 Internal high Voltage supply too high 5 External power supply Voltage too low 6 External power supply Voltage too high 7 GM tube inactive 8 Internal reference Voltage too low 9 Internal reference Voltage too high Service Information If the failure indication occurs, the Xenon Trap Monitor should be returned for service. To return the system, or for additional service help/information, contact: Biodex Medical Systems, Inc. Attn: Service Department 20 Ramsey Road Shirley, NY, 11967-4704 Phone: (800) 224-6339 FAX: (631) 924-8355 12 Biodex Medical Systems, Inc. 2012

APPENDIX A ATOMLAB XENON TRAP MONITOR 13

APPENDIX B TRAP MONITORING TEST RESULTS: BIODEX XENON TRAP MONITOR CHECK SOURCE REFERENCE TEST (DONE ONCE: 5 MINUTES) 1. CHECK SOURCE ID: 2. CHECK SOURCE RATED ACTIVITY: µci 3. CHECK SOURCE DATE: 4. REFERENCE TEST COUNT FOR Cs-137 CHECK SOURCE: Counts for 5 minutes (RANGE: 1193-3340) 5. DAILY CHECK SOURCE RANGE: (Reference Test Count / 10) x 0.8 = (Reference Test Count / 10) x 1.2 = 6. DATE: OPERATOR: DAILY MONITOR TEST Counts for 30 seconds Min Daily Test Count (- 20%) Max Daily Test Count (+20%) 1. TEST BACKGROUND COUNT: Counts (RANGE: 5-60) 2. TEST COUNT FOR Cs-137 CHECK SOURCE: 3. Xe-133 COUNT RESULTS: PASS PASS / FAIL (LESS THAN 99pCi/ml) FAIL 4. DATE: OPERATOR: pci/ml 1. TEST BACKGROUND COUNT: Counts (RANGE: 5-60) 2. TEST COUNT FOR Cs-137 CHECK SOURCE: PASS / FAIL 3. Xe-133 COUNT RESULTS: PASS (LESS THAN 99pCi/ml) FAIL pci/ml 4. DATE: OPERATOR: 1. TEST BACKGROUND COUNT: Counts (RANGE: 5-60) 2. TEST COUNT FOR Cs-137 CHECK SOURCE: PASS / FAIL 3. Xe-133 COUNT RESULTS: PASS (LESS THAN 99pCi/ml) FAIL pci/ml 4. DATE: OPERATOR: 14 Biodex Medical Systems, Inc. 2012

APPENDIX C ATOMLAB XENON TRAP MONITOR 15

Certified Quality Management System BIODEX Biodex Medical Systems, Inc. 20 Ramsey Road, Shirley, New York, 11967-4704, Tel: 800-224-6339 (Int l 631-924-9000), Fax: 631-924-9338, Email: info@biodex.com, www.biodex.com