Reference Manual Oxy Plus Single Sensor Gas Detector

Similar documents
Reference Manual Ex Chek Alkaline Version Combustible Gas Detector

Reference Manual PhD Plus Multi Gas Detector

Safe Operation Manual. Honeywell IQ Force Gas Detector

Reference Manual PhD5 Multi Gas Detector

Reference Manual. MultiVision Gas Detector

Toxi Vision IR CH 4. Single Gas Detector. Reference Manual

Reference Manual MultiPro Gas Detector

Reference Manual. MultiPro Gas Detector

ToxiPro. Single Gas Detector. Reference Manual

Operation and Maintenance Manual

SENSIT HXG-2d Combustible Gas Detector

Biosystems ToxiLtd. Single Gas Detector. Reference Manual

Portable Oxygen Monitor OX-07. (Type B) <No Gas alarm type> Operating Manual

SENSIT HXG-2d. SENSIT TECHNOLOGIES 851 Transport Drive Valparaiso, IN Phone: (219) Fax: (219)

UNI 321. User s Guide

GasAlert. Introduction. WWarning. Table 1. GasAlert Models. Gas Monitored

Macurco Single-Gas XL Series Monitor, CM-1XL Carbon Monoxide (CO), HS-1XL Hydrogen Sulfide (H2S) User Instructions

INSTRUCTION MANUAL MODEL GX-82A

S ing le G a s D etec tor. R eferenc e M a nua l

Portable Oxygen Monitor OX-07. (Type A) Operating Manual

GasAlertMicroClip X3 Specifications

MiniCO, MiniH 2 S, MiniOX, MiniOX Remote, MiniCl 2. , and MiniClO 2. Responder Detectors Instruction Manual "! WARNING

ToxiRAE II. PGM-1100 Series Personal Toxic Gas Monitor. User Manual

INSTRUCTION MANUAL MODEL GX-82A

ToxiRAE II. PGM-1100 Series Personal Toxic Gas Monitor. User Manual

II 2 G. SENSIT HXG-2d. Ex ib IIB T3 IP20. Suitable for combustible gases such as Methane, Butane, Propane and Natural Gas MADE IN USA

Toxi Vision. Single Gas Detector. Reference Manual

Bacharach Bodyguard 4 User's Guide

Instruction Manual Model GX-82 Portable Three Gas Monitor

Dräger Pac Ex 2. Elegant and smart... Explosive hazards and lack or surplus of oxygen under control

GasTech Australia Pty Ltd 24 Baretta Rd Wangara Western Australia 6065 Tel Fax

MODEL OX-07 Type A INSTRUCTION MANUAL FOR RIKEN PORTABLE OXYGEN MONITOR PT0E-0860

MODEL A-316 CARBON MONOXIDE (CO) MONITOR AND ALARM FOR COMPRESSED AIR TESTING

SENSIT HXG-2d Combustible Gas Detector INSTRUCTION MANUAL Read and understand Instructions before use. PENDING Ex. II3G Ex ic llb T3 IP20

Instruction Manual Eagle Series

Instruction Manual Model GX-86 Portable Four Gas Monitor

Single Gas Clip and SGC Plus User s Manual Rev 2.1 READ FIRST BEFORE OPERATION WARNING STATEMENTS / AVERTISSEMENT Activating tthe detector for the

PGM100 Carbon Monoxide

IMR IX176 Portable Gas Detector User Manual

ToxiRAE 3 User s Guide

Confined Space Monitoring. Leak Investigation. Functions 4. Bar Hole Sampling. Line Purge Testing. Actual Size GX-2012

CO2 RESPONDER Portable Carbon Dioxide Detector QUICK REFERENCE GUIDE

IntelliDoX Operator Manual

Single Gas Clip and SGC Plus User s Manual Rev 2.3

MDU 420. Instruction Manual. Dual Range Methane Monitor. P/N REV 0 Printed Specifications Subject to Change

CT-7 Series Toxic Gas Transmitter Operator s Manual

Model GP-01 Combustible Gas Monitor Operator s Manual

MODEL: SP12C7. (Portable 4 gas detector) Operating Manual

SC-01 Operator s Manual

ALTAIR 5X Multigas Detector Product Specification

Operation & Maintenance Manual

EVac II Programmable Digital Vacuum Gauge

Operating Instructions Compur Monitox plus

Using the QRAE Plus personal multigas monitor

ToxiPro. Single Gas Detector. Reference Manual

IPhone: 219/

SENSIT HXG-2. Instruction Manual For use with combustible gases Gas Leak Detector

Quick Touch Pendant User Guide

aspida User Manual T: +44 (0) F: +44 (0) W: E:

Clip SGD. Portable single Gas Detector. User manual

GLD-30 Gas Leak Detector

RAM GENE / RAM GENE-1

GasWatch 2 Operator s Manual Part Number: RK Revision: E Released: 2/13/06

Contents. Glossary

ENMET Corporation. PO Box 979 Ann Arbor, MI RECON. Manual. Manual Part Number MCN-410, 12/03/08

Beacon 800 Gas Monitor Operator s Manual

Pioneer-R16 Gas Monitor Operator s Manual

CryoPod Carrier. Quick Start Guide. English Quick Start Guide, 2 Français Guide de démarrage rapide, 8 Deutsch Schnellstartanleitung, 14

Optimizing Cold Weather Performance. Optimum Cold Temperature Performance From Biosystems Portable Gas Detectors

ALTAIR 4 Four Gas Multigas Detector

GasAlertMicro 5. 1, 2, 3, 4, and 5 Gas Detectors. Quick Reference Guide

RLD-H10P Refrigerant Leak Detector

Model GX-2001 Canadian Version Operator s Manual Part Number: RK Revision: 0 Released: 1/17/05

Carbon Monoxide Transmitter

Operation and Maintenance Manual

Technical Reference Guide. 1, 2, 3, and 4-Gas Detector

EAGLE 2. 6 Channel Capacity. Inlet fitting. Loud buzzer (95dB) IrDA communication port Low flow shutoff and alarm Alarm LED s with wide visibility

Fire Command Keypad. XR5 User s Guide

LD-5000 REFRIGERANT LEAK DETECTOR

Aspida. User Manual. Commercial in Confidence

TIF9055 Programmable. Owner s Manual Manual del Propietario Manuel du propriétaire Benutzerhandbuch Manuale del proprietario

OWNER S MANUAL DLFCAB / DLFCHB / DLFDAB / DLFDHB High Wall Ductless System Sizes 09 36

T82. Instruction Manual. Single Gas Monitor GUARANTEED. FOR LIFE. P/N REV 0 PRINTED ,000 Specifications Subject to Change

03 Series: OX-03, CO-03, HS-03 Versions Operator s Manual

OWNER S MANUAL. R 410A Ductless Split System Air Conditioner and Heat Pump

RAM GENE-1 ERK. Contamination & Radiation Meter. Operating Manual. Document #PRIR90N2.DOC Version 2.2 October 2009

Owner s Manual. Walk-in Monitoring System 100B. Cooler is Better! TM. Used in UL Listed Door Panel Assemblies

User Manual. Humidity-Temperature Chart Recorder. Model RH520

Operate the Mutli-Gas Monitor PID

Controllers. Instruction Manual WARNING

USER MANUAL Digital CO Monitor Model CO30

Operating & Maintenance Manual. Alert-4 Ethernet LCD Master Alarm

QUICK REFERENCE GUIDE

Model R9400. Instruction Manual. CO Meter with Temperature. wwwreedinstruments. com

Using the MiniRAE 2000 & ppbrae PID

Operations Manual TS400. Test Station for G450/G460 Gas Detector

GAS-Mate. Combustible Gas Leak Detector

ZX1e ZX2e ZX5e. Document No Issue 01 user manual

ATX620. Instruction Manual. Multi-Gas Monitor. P/N REV 02 Printed Specifications Subject to Change GUARANTEED. FOR LIFE.

Model GX-2003 Operator s Manual

Transcription:

Reference Manual Oxy Plus Single Sensor Gas Detector 651 South Main St. Middletown, CT 06457 860-344-1079 FAX 860-344-1068 27OCT2005 Version 4.0 P/N: 13-047

THE OXY PLUS PERSONAL PORTABLE GAS DETECTOR HAS BEEN DESIGNED FOR THE DETECTION AND MEASUREMENT OF POTENTIALLY HAZARDOUS ATMOSPHERIC CONDITIONS IN ORDER TO ASSURE THAT THE USER IS PROPERLY WARNED OF POTENTIALLY DANGEROUS CONDITIONS, IT IS ESSENTIAL THAT THE INSTRUCTIONS IN THIS MANUAL BE READ, FULLY UNDERSTOOD, AND FOLLOWED. AVERTISSEMENT: LIRE ATTENTIVEMENT LES INSTRUCTIONS AVANT DE METTRE EN MARCHE. Oxy Plus Operation Manual Biosystems Part Number 13-047 Version 4.0 Copyright 2005 by Biosystems LLC Middletown, Connecticut 06457 All rights reserved. No page or part of this operation manual may be reproduced in any form without written permission of the copyright owner shown above. Biosystems reserves the right to correct typographical errors.

Table of Contents Introduction...5 Warnings and Cautions...6 A. Signal Words...6 B. Warnings...6 1. Overview...7 1.1 Oxy Plus capabilities...7 1.2 Methods of sampling...7 1.3 Oxy Plus sensor...7 1.4 One Button Auto Calibration...7 1.5 Display backlight...7 1.6 Alarm logic...7 1.6.1 Atmospheric hazard alarms...8 1.6.2 Low battery alarms...8 1.7 Special microprocessor features...8 1.8 Classification for intrinsic safety...8 1.9 Options...8 1.9.1 Vibrator alarm...8 1.9.2 Remote sample draw assembly...8 1.9.3 Weather cover...9 1.9.4 Training video...9 1.10 Oxy Plus design components...9 1.11 Oxy Plus accessories...9 2. Basic Operation...11 2.1 Overview of Oxy Plus operation...11 2.1.1 Operational warnings and cautions...11 2.1.2 Turning the Oxy Plus on...11 2.1.3 Turning the Oxy Plus off...11 2.1.4 Start-up sequence...11 2.2 Operating modes...12 2.2.1 Text Only mode...12 2.2.2 Basic mode...12 2.2.3 Technician mode...12 2.2.4 Changing operating modes...13 2.3 Oxy Plus batteries...13 2.3.1 Low battery voltage alarms...13 2.3.2 Battery replacement...13 2.4 Methods of sampling...13 2.4.1 Using the hand aspirated sample draw kit...14 3. Calibration...15 3.1 Verification of accuracy...15 3.1.1 Effect of contaminants on Oxy Plus sensors...15 3.2 Fresh air "zero" calibration...15 3.3 Functional (bump) test...15 3.4 Auto-calibration...15 3.4.1 Fresh air "zero" auto-calibration sequence...15 3.5 Manual calibration procedure...16 3.5.1 Fresh air "zero" through keypad buttons...16 3.5.2 True zero span calibration...17 4. Setting alarm levels...17

4.1 Alarm adjustment sequence...17 4.2 Restoring the factory default alarm settings...18 5. Oxy Plus Advanced Functions...19 5.1 Oxy Plus Advanced features overview...19 5.2 Alarm Latch command...19 5.3 Setting the security beep interval...19 6. Troubleshooting and Repair...20 6.1 Changing sensors...20 6.2 Troubleshooting...20 6.2.1 Unit will not turn on...20 6.2.2 Problem: Unit will not turn off...21 6.2.3 Problem: Sensor readings unstable...21 6.2.4 Problem: Display is blank...21 6.2.5 Problem: No audible alarm...21 6.2.6 Problem: Keypad buttons (+,-, Cal, Alarm) don't work...21 6.2.7 Problem: Can t make a One Button auto zero adjustment ( Too High or Too Low for zero adjust)...21 6.3 Returning your Oxy Plus to Biosystems for service or repair...21 Appendix A How to calibrate your Oxy Plus in contaminated air...22 Appendix B Suggested Calibration Gas...22 Biosystems Standard Warranty Gas Detection Products...23 Introduction The Oxy Plus is a personal, portable, microprocessor controlled, single sensor gas detector for the measurement of oxygen (O2). The Oxy Plus uses a top-mounted LCD (liquid crystal display) to show readings and other messages. An audible alarm and alarm light warn users of hazardous conditions. An optional vibrator type alarm is also available for use in high noise or other environments requiring additional warnings. A manually activated display back-light insures that the LCD may be easily read even in dim light conditions. The Oxy Plus offers a choice of three modes of operation, ( OK, Basic and Technician) providing the right amount of information for users with different skill levels. Biosystems One Button operating logic allows all procedures necessary for normal day-to-day operation, including fully automatic calibration adjustment, through use of the single on/off Mode button. The Oxy Plus is Classified by Underwriters Laboratories, Inc. and the Canadian Standards Association (pending) as to Intrinsic Safety for use in Hazardous Locations Class I, Division 1, Groups A, B, C, and D. Classification for intrinsic safety is based on tests conducted in explosive gas / air (21 % Oxygen) mixtures only. Warning: Substitution of components may impair intrinsic safety. AVERTISSEMENT: LA SUBSTITUTION DE COMPOSANTS PEUT COMPROMETTRE LA SÉCURITÉ INTRINSÈQUE. CAUTION: FOR SAFETY REASONS THIS EQUIPMENT MUST BE OPERATED AND SERVICED BY QUALIFIED PERSONNEL ONLY. READ AND UNDERSTAND THE INSTRUCTION MANUAL COMPLETELY BEFORE OPERATING OR SERVICING THE OXY PLUS. ATTENTION: POUR DES RAISONS DE SÉCURITÉ, CET ÉQUIPMENT DOIT ETRE UTILISÉ, ENTRETENU ET RÉPARÉ UNIQUEMENT PAR UN PERSONNEL QUALIFIÉ. ÉTUDIER LE MANUEL D'INSTRUCTIONS EN ENTIER AVANT D'UTILISER, 'ENTRETENIR OU DE RÉPARER L'ÉQUIPMENT.

Warnings and Cautions A. Signal Words The following signal words, as defined by ANSI Z535.4-1998, are used in the Oxy Plus Operator s Guide. indicates an imminently hazardous situation which, if not avoided, will result in death or serious injury. indicates a potentially hazardous situation which, if not avoided, could result in death or serious injury. indicates a potentially hazardous situation, which if not avoided, may result in moderate or minor injury. CAUTION used without the safety alert symbol indicates a potentially hazardous situation which, if not avoided, may result in property damage. B. Warnings 1. The Oxy Plus personal, portable gas detector has been designed for the detection of dangerous atmospheric conditions. An alarm condition indicates the presence of a potentially life-threatening hazard and should be taken very seriously. 2. In the event of an alarm condition it is important to follow established procedures. The safest course of action is to immediately leave the affected area, and to return only after further testing determines that the area is once again safe for entry. Failure to immediately leave the area may result in serious injury or death. 3. The accuracy of Oxy Plus instruments should be checked on a regular basis. Failure to check accuracy can lead to inaccurate and potentially dangerous readings. The Oxy Plus O 2 should be periodically calibrated in fresh air. 4. A sensor that cannot be calibrated or is found to be out of tolerance must be replaced immediately. Instruments that fail calibration may not be used until testing with fresh air determines that accuracy has been restored and the instrument is once again fit for use. 5. Substitution of components may impair intrinsic safety. 6. For safety reasons the Oxy Plus must be operated by qualified personnel only. Read, understand and follow the directions set forth in this reference manual before operating the Oxy Plus. 7. The Oxy Plus has been tested for intrinsic safety in Explosive Gas/AIR (max. 21.0% O 2 ).

1. Overview 1.1 Oxy Plus capabilities The Oxy Plus is a single sensor oxygen detector that can be configured to meet a wide variety of user requirements. This chapter provides an overview of many of the features of the Oxy Plus. More detailed descriptions of the specific features of the Oxy Plus are contained in the subsequent chapters of this manual. 1.2 Methods of sampling The Oxy Plus may be used as either a "Diffusion" or "Sample-Draw" type monitoring device. In normal operation, the Oxy Plus detector is worn on the belt, clipped into the user's shirt pocket, suspended on the neck lanyard, or held by hand. Once turned on, the Oxy Plus monitors continuously. The atmosphere being measured gets to the sensor by diffusing through a dust and water resistant protective filter directly into the sensor. Normal air movements are enough to carry the sample to the sensor. The sensor reacts to changes in the concentration of the hazard being measured. Values are constantly updated and shown on the Oxy Plus LCD. This type of "diffusion" operation monitors only the atmosphere which immediately surrounds the detector. It is also possible to use a sample draw kit to obtain readings from remote locations. A hand-operated squeezebulb is used to draw the sample in through a probe assembly, and suck it through a length of hose back to the instrument. Use of the sample draw kit is covered in section 2.4. Oxygen readings are given in percent by volume. 1.3 Oxy Plus sensor The Oxy Plus uses a highly dependable and long lasting fuel cell type oxygen sensor designed to minimize the effects of common interfering gases. Sensor Nominal range Max overload Oxygen 0-25% /volume 30%/volume Oxy Plus Sensor Range 1.4 One Button Auto Calibration The Oxy Plus detector has been designed for easy verification of accuracy. Biosystems One Button logic means that all normal day-to-day operating procedures, including automatic calibration adjustment, may be undertaken through the use of the single on/off Mode button. The accuracy of Oxy Plus instruments should be checked periodically with known concentration calibration gas. Failure to check accuracy can lead to inaccurate and potentially dangerous readings. The Oxy Plus O 2 should be periodically calibrated in fresh air. Calibration for the Oxy Plus detector is a single step procedure: the instrument is simply taken to an area where the air is known to be fresh and a "zero" adjustment is made automatically ondemand by pressing the on/off Mode button. The instrument automatically adjusts readings to match the oxygen concentration known to be present in fresh air (20.9 %). Calibration procedures are discussed in detail in Chapter 3. 1.5 Display backlight A manually-activated display backlight allows the LCD to be read in dim light conditions. Pressing the on/off Mode button once at any time during normal operation causes the back-light to be activated for 10 seconds. In the event of a gas alarm the backlight turns on automatically and remains on until the alarm condition clears. 1.6 Alarm logic Oxy Plus alarms are user adjustable and may be set anywhere within the range of the sensor. When an alarm set point is exceeded a two-toned audible alarm sounds, a bright red LED alarm light flashes, and (if equipped) the vibrating alarm is activated.

Oxy Plus alarms are normally of the selfresetting type. When readings drop back between the pre-set alarm levels, the instrument reverts back to normal operation, and the visual and audible alarms cease. It is possible, if desired, to set Oxy Plus alarms so that they "latch." In the latched condition, once an alarm occurs both visual and audible alarms continue to sound even after the atmospheric hazard has cleared. The instrument must be manually reset before the alarms are silenced. Procedures for latching Oxy Plus alarms are given in Chapter 5.2. 1.6.1 Atmospheric hazard alarms The Oxy Plus personal, portable gas detector has been designed for the detection of dangerous atmospheric conditions. An alarm condition indicates the presence of a potentially lifethreatening hazard and should be taken very seriously. In the event of an alarm condition it is important to follow established procedures. The safest course of action is to immediately leave the affected area, and return only after further testing together with other appropriate safety procedures determine that the area is once again safe for entry. Two standard alarm set-points ( Low and High ) have been provided for the Oxy Plus detector. The default low alarm is 19.5% and the default high alarm point is 23.5%. The procedures for adjusting alarm set points are given in Chapter 4. 1.6.2 Low battery alarms The Oxy Plus uses 3 AAA alkaline batteries for power. A fresh set of batteries should provide up to six months of normal operation. When it is time to replace the batteries a "B" will appear in the lower left hand corner of the display screen, and the audible alarm will beep once per minute until the battery is replaced. From the time when the "B" is first shown the user should have one to two weeks of useful life remaining Batteries should be replaced as soon as possible after any low battery message is displayed. Battery replacement is covered in Section 2.3. 1.7 Special microprocessor features Several automatic programs prevent tampering and misuse of the Oxy Plus by unauthorized persons. Each time the detector is turned on, an electronic selftest is performed that assures the user of proper performance. The sensor, LED alarm light, and audible alarm are automatically evaluated, and the battery is monitored continuously for proper voltage. Several optional setup choices are also available including adding a security beep, latching the alarms, and changing or restoring the factory alarm settings. Optional choices made by using the 4 pushbuttons on the instrument keypad. Chapter 5 describes how to use these advanced technical features and setup choices in greater detail. 1.8 Classification for intrinsic safety The Oxy Plus is Classified by Underwriters Laboratories, Inc., and the Canadian Standards Association (pending) as to Intrinsic Safety for use in Hazardous Locations Class I, Division 1, Groups A, B, C, & D. Classification for intrinsic safety is based on tests conducted in explosive gas / air (21 % Oxygen) mixtures only. 1.9 Options 1.9.1 Vibrator alarm An internally mounted vibrator type alarm is optionally available for use in high noise or other environments requiring additional warnings. The vibrator alarm option must be specified at the time the instrument is purchased. 1.9.2 Remote sample draw assembly An optional hand aspirated (squeeze bulb) sample draw kit is available for use

in situations where the sample must be obtained from locations which are remote from the instrument. The sample draw kit consists of a sample draw adaptor, squeeze bulb, sample probe assembly, spare filters, and 10 feet of sample tubing. 1.9.3 Weather cover An optional weather cover is available for use in environmentally harsh locations, or where additional protection is desired. 1.9.4 Training video A comprehensive 25 minute training video is available to augment and support the written Owner s Manual. 1.10 Oxy Plus design components (1) Case: The instrument is enclosed in a durable, gasketed case constructed of a metal plated ABS / polycarbonate material. (2) LCD display: A top mounted liquid crystal display (LCD) meter allows display of readings, messages, and other information. A manually activated back-light insures that the LCD may be easily read even in dim light conditions. (3) Alarm light: A bright red topmounted LED (light emitting diode) alarm light visually warns users when an alarm level is exceeded. (4) Audible alarm: A loud two-toned audible alarm is also provided to warn users of hazardous conditions. (5) Mode button: The on/off "Mode" button is used to turn the instrument on and off and to control most other functions. (6) Protective cap: A protective O-ring sealed cap and moisture barrier filter protect the sensor from damage. (7) Sensor: The top-mounted sensor may be easily accessed for replacement or repair. (8) Bottom surface: The stainless steel belt clip may be removed to allow access to the 4 hidden push-buttons ("Cal," "Alarm," "+," and "-") used to control advanced functions. 1.11 Oxy Plus accessories Each Oxy Plus is delivered in a foam lined box containing: the Oxy Plus detector (with sensor installed), set of three AAA alkaline batteries (installed), alligator clip and lanyard, calibration adaptor, owner s manual and quick reference card. Optional Oxy Plus accessories include sample draw kit, leather weather cover, training video, and internally mounted vibrator alarm (vibrator option must be specified at time of purchase).

Figure 1.3: Oxy Plus external features Figure 1.4: Oxy Plus bottom surface

2. Basic Operation 2.1 Overview of Oxy Plus operation The Oxy Plus offers a choice of three modes of operation, "Text Only," "Basic," and "Technician. Mode selection is a function of how much information is required, the skill level of the user, and the nature of the job. 2.1.1 Operational warnings and cautions 1. The Oxy Plus personal, portable gas detector has been designed for the detection of dangerous atmospheric conditions. An alarm condition indicates the presence of a potentially life-threatening hazard and should be taken very seriously. 2. In the event of an alarm condition it is important to follow established procedures. The safest course of action is to immediately leave the affected area, and to return only after further testing determines that the area is once again safe for entry. Failure to immediately leave the area may result in serious injury or death. 3. The accuracy of Oxy Plus instruments should be checked on a regular basis. Failure to check accuracy can lead to inaccurate and potentially dangerous readings. The Oxy Plus O 2 should be periodically calibrated in fresh air. 4. A sensor that cannot be calibrated or is found to be out of tolerance must be replaced immediately. Instruments that fail calibration may not be used until testing with fresh air determines that accuracy has been restored and the instrument is once again fit for use. 5. Substitution of components may impair intrinsic safety. 6. For safety reasons the Oxy Plus must be operated by qualified personnel only. Read, understand and follow the directions set forth in this reference manual before operating the Oxy Plus. 7. The Oxy Plus has been tested for intrinsic safety in Explosive Gas/AIR (max. 21.0% O 2 ) 2.1.2 Turning the Oxy Plus on The black switch on the front of the Oxy Plus case is called the "Mode" button. It is used to turn the Oxy Plus on and off, as well as to control most other operations of the instrument. Press and hold the mode button for one second to turn the Oxy Plus on. 2.1.3 Turning the Oxy Plus off Turn the Oxy Plus off by pressing and holding the mode button down for 3 seconds. 2.1.4 Start-up sequence After the detector has been turned on, it will automatically go through an electronic self test and start up sequence that takes approximately twenty seconds. During the self test sequence the LCD backlight will be activated, the visual LED alarm light will flash, the audible alarm will sound, and (if equipped) the optional vibrator alarm will briefly activate. During start-up the LCD display will also show several messages or "screens" in sequence. The first screen displays the software version number: (Note: The software version installed may differ with that shown above.) The second shows the type of sensor installed:

Additional screens will briefly appear indicating the current alarm set-points (low and high). As an example, an Oxy Plus being operated in the Technician mode would show the following sequence of factory installed (default) alarm set points when the instrument is first turned on: This screen is followed by a set of two screens, which show the current gas readings screen and the type of gas being measured: Note: The alarm set-point screens are omitted when the instrument is turned on while in the "Basic" or "Text Only" modes. 2.2 Operating modes Three operating modes provide Oxy Plus users with the right amount of information to meet their monitoring needs. The Oxy Plus is designed to be as simple or as sophisticated as the job that needs to be done. 2.2.1 Text Only mode The simplest mode of operation is the "Text Only" mode. In this mode, during normal operation the LCD screen does not display numerical readings, only the indication "OK". This screen cycles back and forth between another which identifies the type of gas being measured. During an alarm condition (when the sensor readings exceed a pre-set alarm level), the display changes from "OK" to the numerical value, the LED alarm light flashes, the audible alarm sounds, and (if equipped) the vibrator alarm is activated. Oxy Plus alarms are normally selfresetting. When readings drop back between the pre-set alarm levels, the screen reverts back to the "OK" indication, and visual and audible alarms cease. 2.2.2 Basic mode The "Basic" mode of operation is designed for users with a higher level of understanding, and a need for numerical readings, but who are not experts in gas detection. In the Basic mode, as in the Text Only mode, the user has access to only one "screen" of information; the current gas readings. The major difference between the two modes is that in the Basic mode numerical gas level readings are provided, whereas in the Text Only mode only an indication that conditions are "OK" is provided during normal operation. Basic mode current gas level screen: During an alarm condition (when the sensor readings exceed a pre-set alarm level), the numerical reading changes to reflect the new value, the LED alarm light flashes, the audible alarm sounds and (if equipped) the vibrator alarm is activated. Oxy Plus alarms are normally selfresetting. When readings drop back between the pre-set alarm levels, visual and audible alarms cease, and normal operation of the instrument resumes. 2.2.3 Technician mode The "Technician Mode" provides access to all advanced functions and displays. When the instrument is turned on while in the Technician Mode, after the completion of the start-up and self-test sequence, the meter display will show numeric readings of the gas being measured. During an alarm condition (when the sensor readings exceed a pre-set alarm level), the numerical reading changes to

reflect the new value, the LED alarm light flashes, and the audible alarm sounds. Oxy Plus alarms are normally selfresetting. When readings drop back between the pre-set alarm levels, visual and audible alarms cease, and normal operation of the instrument resumes. Pressing the black mode button while in the Technician Mode allows the user to switch to the minimum reading display. The screen will display the minimum reading of the gas being measured since the instrument was switched on. Pressing the black mode button again will switch back to the current gas readings screen. 2.2.4 Changing operating modes It is possible to change operating mode while the instrument is in use. It is not necessary to turn the instrument off before switching. Simply remove the belt clip on the bottom of the instrument to expose the four mini-push-buttons, and simultaneously push the "+" and "-" buttons to change operating modes. Each time the operating mode is changed the display screen will briefly indicate which operating mode has been selected. Shifting modes or otherwise reprogramming the instrument should only be done by employees who are authorized to do so. 2.3 Oxy Plus batteries The Oxy Plus uses a set of three AAA disposable alkaline batteries. A fresh set of batteries should allow up to six months of normal operation. 2.3.1 Low battery voltage alarms When it is time to replace the batteries a "B" will appear in the lower left hand corner of the display screen, and the audible alarm will beep once per minute until the batteries are replaced. The user should have approximately one to two weeks of useful life remaining from the time when the "B" is first shown. 2.3.2 Battery replacement Oxy Plus batteries should be replaced as soon as possible after a low battery ("B ) message has been noticed. Use the following procedure to change the batteries. (1) Make sure the instrument is turned off before replacing batteries. (2) Remove the belt clip from the bottom of the instrument by sliding it completely free of the belt clip mounts. (3) Loosen and remove the two screws securing the battery compartment cover to the instrument case. (The screws must be completely removed before the battery compartment cover can be detached from the instrument case.) (4) Gently remove the battery compartment cover by sliding it backwards then lifting it free from the instrument case. (5) Remove the old batteries. (6) Make sure the battery terminals are clean. (7) Install the new batteries. (Make sure that the batteries are aligned in the correct directions, and that all battery polarities are correct.) (8) Replace the battery compartment cover, screws, and belt clip. (9) Verify the accuracy of the Oxy Plus detector as discussed in section 3 below. The accuracy of Oxy Plus instruments should be checked on a regular basis. Failure to check accuracy can lead to inaccurate and potentially dangerous readings. The Oxy Plus O 2 should be periodically calibrated in fresh air. 2.4 Methods of sampling The Oxy Plus may be used as either a "Diffusion" or "Sample-Draw" type monitoring device. In normal operation, the Oxy Plus detector is worn on the belt, clipped into a shirt pocket, suspended from the neck lanyard, or held by hand. Once turned on, the Oxy Plus monitors continuously. The atmosphere being measured gets to the sensor by diffusing through a

protective moisture barrier filter. Normal air movements are enough to carry the sample to the sensor. The sensor reacts to changes in the concentration of the gas being measured. Values are constantly updated by the Oxy Plus detector, while readings are displayed on the LCD meter. This type of "diffusion" operation monitors only the atmosphere which immediately surrounds the detector. It is possible to use the Oxy Plus to sample remote locations using a sample draw kit. The Oxy Plus kit uses a handaspirated squeeze bulb to draw a sample through a hose and probe assembly into the instrument. draw the sample in by squeezing the bulb continuously for a minimum of 55 seconds or until readings stabilize. 6. Note the gas measurement readings. CAUTION: Hand-aspirated remote sampling only provides continuous gas readings for the area in which the probe is located while the bulb is being continuously squeezed. Each time a reading is desired, it is necessary to squeeze the bulb a sufficient number of times to bring a fresh sample to the sensor compartment. 2.4.1 Using the hand aspirated sample draw kit To use the manual sample draw kit: 1. Connect the slip-on sample draw cup to the squeeze bulb and hose assembly by sliding the end of the short length of hose closest to the bulb to the sample draw cup. Then connect the longer end of tubing to the sample probe. 2. To test the seals in the sample draw system, cover the end of the sample draw probe with a finger, and squeeze the aspirator bulb. If there are no leaks in the sample draw kit components, the bulb should stay deflated for a few seconds. 3. Secure the sample craw cup (with the sample draw assembly attached) to the Oxy Plus by sliding the cup over the sensor port. 4. Insert the end of the sample probe into the location to be sampled. 5. Squeeze the aspirator bulb to draw the sample from the remote location to the sensor compartment. To ensure accurate readings while using the manual sample draw kit, it is necessary to squeeze the bulb once for every one foot of sampling hose for the sample to first reach the sensors, and then to continue squeezing the bulb once per second for an additional 45 seconds or until readings stabilize. As an example, if 10 feet of tubing is used, it will be necessary to

3. Calibration The Oxy Plus detector has been designed for easy calibration. A single button, the on / off Mode switch, is used to enter the Auto-Calibration mode, and automatically make all calibration adjustments. It is also possible to manually calibrate the instrument by using simple pushbutton controls located on the instrument key-pad. One-Button Auto-Cal procedures are discussed in Section 3.4. Manual calibration procedures are discussed in Section 3.5. 3.1 Verification of accuracy Verification of accuracy is a one step procedure. Take the Oxy Plus to an area where the atmosphere is fresh and check the readings. If the readings differ from those expected in fresh air a "zero" adjustment must be made. This is the only step necessary for verifying the accuracy of oxygen detectors. 3.1.1 Effect of contaminants on Oxy Plus sensors The atmosphere in which the Oxy Plus monitor is being used can have an effect on the sensor. Sensors may be poisoned or suffer degraded performance if exposed to certain substances. The accuracy of Oxy Plus instruments should be checked following exposure to any known contaminants. Failure to check accuracy can lead to inaccurate and potentially dangerous readings. 3.1.1.1 Effects of contaminants on oxygen sensors Oxygen sensors may be affected by prolonged exposure to "acid" gases such as carbon dioxide. The oxygen sensors used in Biosystems instruments are not recommended for continuous use in atmospheres which contain more than 25 % CO 2. 3.2 Fresh air "zero" calibration The fresh air "zero" must be done in fresh, uncontaminated air. In this procedure the instrument automatically adjusts its readings to match the concentrations present in fresh air (20.9 % for oxygen). Since fresh air contains 20.9 percent oxygen, this is the only calibration needed for the oxygen sensor. If the Oxy Plus cannot be taken to an area where the air is fresh, or if it is not certain whether or not the air is uncontaminated, special procedures are required. These procedures are discussed at greater length in Appendix A. 3.3 Functional (bump) test A simple functional (bump) test is all that is needed to verify accuracy. To perform a functional (bump) test, turn the Oxy Plus on, and wait at least three minutes to allow the readings to fully stabilize. Make sure the instrument is located in fresh air. Check to see that the readings match the concentrations present in fresh air. If the Oxy Plus is operated in either the Basic or Technician operating mode the fresh air readings should equal 20.9 % for oxygen sensors. If the instrument is operated in the Text Only mode, change the operating mode to Technician or Basic and verify the readings. If the display shows 20.9% in a fresh air, there is no need to perform a calibration. If the display shows anything other than 20.9% in an environment known to be fresh, then calibrate the detector by following the instructions given below in section 3.4. 3.4 Auto-calibration Biosystems One-Button Auto- Calibration mode may be used to verify accuracy any time during normal operation while the instrument is being used in either the Basic or Technician operating mode. 3.4.1 Fresh air "zero" autocalibration sequence The fresh air zero procedure may only be done while the instrument is being operated in either the Technician or Basic operating mode.

(1) Turn the instrument on and make sure gas readings are given in numbers. If the reading shows OK, the instrument is currently being operated in the Text Only mode. It will be necessary to change to either the Basic or Technician operating mode. To switch modes simultaneously hold down the "+" and"-" key as discussed in Chapter 2. (2) Wait at least three minutes after turning the instrument on to allow sensor readings to stabilize fully before initiating auto-calibration procedures. (3) Make sure the instrument is located in an area where the air is known to be fresh. (4) Press the mode button 3 times within two seconds. This will "wake up" the instrument from normal operation, and put it into the Auto-Calibration mode. A screen will briefly display the message ZERO MODE. (5) Pressing the mode button within five seconds causes the fresh air adjustment to be made. (If the Mode button is not pushed within 5 seconds, the instrument will return to normal operation.) The screen will show the message WAIT ZERO while the adjustment is being completed. (6) After successful completion of the zero auto-calibration, the display of oxygen reading detectors will return to normal operation. 3.4.1.1 Reading Too High or Too Low for zero adjust To reduce the chances of the Oxy Plus being inadvertently zeroed in contaminated air, only small adjustments are allowed through the use of the One Button auto-zero sequence. If the necessary adjustments are too large the display will indicate that the reading is too Low or too high for zero adjustment. In this case the instrument must be fresh air zeroed using the Cal button on the instrument keypad and procedures discussed in Section 3.5.1. of this manual. 3.5 Manual calibration procedure It is also possible to calibrate the Oxy Plus manually using the four buttons on the instrument keypad. 3.5.1 Fresh air "zero" through keypad buttons (1) Turn the instrument on. (2) Wait at least three minutes after turning the instrument on to allow sensor readings to stabilize fully before initiating the fresh air zero procedure. (3) Slide the belt clip towards the rear of the instrument exposing the four buttons on the instrument keypad. (4) Verify that the instrument is in the Basic or Technician operating mode. Switch modes (if necessary) by simultaneously holding down the "+" and"-" keys. (5) Make sure the instrument is located in an area where the air is known to be fresh. (6) Press the keypad button marked "Cal". The GAS ZERO calibration message will appear on the instrument LCD. (7) Press Cal to automatically zero the instrument. The LCD will show the message ZERO WAIT while the adjustment is being made. (8) After completion of the zero adjustment the Oxy Plus automatically returns to the gas reading screen display. (9) Pressing the mode button prior to completion of the calibration causes the calibration values in the memory to remain unchanged from the last time a fresh air adjustment was made. The No Save screen is displayed briefly to indicated that the

zero values have not been changed. The instrument will then return to normal operation. 3.5.2 True zero span calibration During normal operation the only calibration adjustment necessary for the Oxy Plus is the automatic fresh air zero procedure as discussed in Section 3.1. Following reinitialization, the Oxy Plus must be span adjusted prior to being returned to service. If the Oxy Plus is not span adjusted message screens will indicate NEED CAL when the instrument is next turned on. Note: A cylinder of calibration gas containing 100% nitrogen and no oxygen is necessary for this calibration routine. 1. Make sure gas readings are given in numbers. If the display shows OK, then the Oxy Plus is currently being operated in the Text Only mode. To change over to Basic mode, simultaneously hold down the "+" and"-" keys as discussed in section 2.2.4. 2. Make sure the Oxy Plus has been turned on and operating for a sufficient period of time for the sensor readings to have stabilized fully before initiating manual calibration procedures. If the oxygen sensor has just been replaced, the Oxy Plus should be turned on and allowed to stabilize for one hour prior to performing any of the calibration options. 3. Turn the Oxy Plus off. 4. While holding down the "Cal" button, press the Mode button to turn the Oxy Plus back on. Release the button when the word SPAN appears. The display should alternate between SPAN and the current oxygen reading (20.9 in fresh air). Press the Mode button at any time to cancel the span calibration. 5. Screw the regulator into the cylinder and connect it to the detector using the short section of tubing and the calibration adapter. The regulator will automatically begin flowing calibration gas at the correct flow rate as soon as it is screwed into the cylinder of gas. 6. Allow the reading to stabilize. If the reading is a positive number, use the + and - keys to bring the value to zero. 7. When finished, press and hold down the Cal key until the display reads DONE. 8. Perform a fresh air zero adjustment as discussed in section 3.6.1 when the instrument is next turned on. 4. Setting alarm levels Oxy Plus alarms are user adjustable and may be set anywhere within the range of the sensor channel. When an alarm set point is exceeded a loud audible alarm sounds, the bright red LED alarm light blinks, and (if equipped) the optional vibrator alarm is activated. Oxy Plus alarms are normally self resetting, that is, as soon as the alarm condition no longer exists, the alarm ceases. It is possible, if desired, to set Oxy Plus alarms so that they "latch." In the latched condition, once an alarm occurs both visual and audible alarms continue to sound even after the atmospheric hazard has cleared. The instrument must be manually reset by pressing the mode button. Pressing the mode button silences the alarms and restores normal operation. Procedures for latching Oxy Plus alarms are given in Section 5.2. 4.1 Alarm adjustment sequence It is necessary to be in the "Alarm Adjust" mode before modifying alarm settings. Once in the Alarm Adjust mode, alarms and alarm set points can be adjusted by the user. In many cases it is possible to comply with OSHA guidelines while using higher alarm points than those used by Biosystems. It is important to note that the default alarm point settings used in the Oxy Plus design are very conservative in order to provide maximum worker safety. The default low alarm setting is 19.5% and the default high alarm setting is 23.5%. To enter the alarm adjust mode, it is necessary to do the following. (1) Turn the instrument on. (2) Remove the belt clip to expose the four mini-push-buttons. (3) Verify that the instrument is in the Technician or Basic operating mode. Switch modes (if necessary) by simultaneously holding down the "+" and"-" keys. Each time that the operating mode is changed, the meter screen will briefly indicate the current operating mode.

(4) Turn the instrument off by pressing the mode button for three full seconds. (5) With the unit turned off, press and hold down the "Alarm" button. (6) While holding down the "Alarm" button, press the mode button to turn the Oxy Plus back on. (7) A screen will appear showing the first alarm point to be adjusted, and the current alarm set point. (8) Press the Cal button again to advance to the next available alarm adjustment option. Pressing the mode button at any time cancels the alarm adjustment mode, and returns the instrument to the gas reading screen. (9) When the desired alarm adjust option has been reached, use the "+" and "-" keys to raise or lower the alarm set point. (10) When all alarm adjustments have been completed, press and hold the "Alarm" button down until an information screen indicates that alarm adjustment is complete. (11) The next screen displayed will allow the user to set the Alarm Latch see section 5.2. When the adjustment has been completed, press and hold the "Alarm" button down until an information screen indicates that alarm adjustment is complete. The instrument will automatically turn itself off. Turn the instrument back on to resume normal operation message, and the type of sensor currently installed. (3) This screen will be followed by several more showing the specific default alarm settings. These screens will be shown in continuous rotation. (5) Push and hold the "Alarm" button to restore the factory default alarm settings. The display indicates when the settings have been restored: (6) If the "Mode" button is pushed, the current alarm settings remain unchanged, and the display shows the following screen: (7) After selection of the default or current alarm settings has been made, the display reverts to the current gas reading screen. 4.2 Restoring the factory default alarm settings Oxy Plus alarm settings are set conservatively at the factory. The default low alarm setting is 19.5% and the default high alarm setting is 23.5%. It is possible to restore these default settings at any time by doing the following. (1) Turn the instrument on and make sure it is in the Technician Mode. (See Section 2.2.4.) (2) Press the "Alarm" button on the instrument key-pad. The display will briefly show the default alarm

5. Oxy Plus Advanced Functions 5.1 Oxy Plus Advanced features overview The Oxy Plus microprocessor circuitry makes possible a number of advanced features and capabilities. Although this chapter has been designed primarily to be of benefit to "experts" there are several sections that may be of interest to other Oxy Plus users. The four mini-push-buttons located under the key pad cover on the bottom of the detector are used to program advanced Oxy Plus functions. The technique for using these buttons requires several buttons to be pressed and held at the same time. Caution: Do not use this technique with any other combination of buttons than those listed below. Doing so may result in unintended changes to the instrument setup. 5.2 Alarm Latch command Oxy Plus alarms are normally self resetting, that is, as soon as readings drop below the alarm set point, the alarm ceases. It is possible, if desired, to set Oxy Plus alarms so that they "latch. In the latched condition, once an alarm occurs both visual and audible alarms continue to sound even after the atmospheric hazard has cleared. The instrument must be manually reset by pressing the mode button. Pressing the mode button silences the alarms and restores normal operation. Use the following procedure to latch the Oxy Plus alarms: (1) Make sure the Oxy Plus detector is in the Technician mode (as discussed in Section 2.2.4), then turn the instrument off. (2) Turn the instrument on while pressing and holding down the Alarm button. This will place the instrument in the Alarm Adjust Mode. (3) Adjust and save the current alarm settings, or press mode to exit without making changes to the current settings. As the instrument leaves the Alarm Adjust Mode the unit will display alternating ALRM / LOCK messages for 4 seconds. The instrument will then display the current status of the alarm latch feature ( ON or OFF ). (4) Use the + or - key to change the alarm latch setting. (5) Once the proper setting has been selected, press and hold down the Alarm button for three seconds. Do not release the Alarm button until the LCD displays the message DONE to indicate that the new setting has been successfully entered. Press the mode button at any time to cancel the procedure. The LCD will show the message NO SAVE to indicate that the setting has not been changed. 5.3 Setting the security beep interval The security beep screen allows the Oxy Plus to be programmed to emit an audible alarm "beep" on a regular basis while the instrument is in operation. This periodic beep serves as a reminder that the instrument is on. Use the following procedure to turn the security beep on: (1) Make sure the Oxy Plus detector is in the Technician mode (as discussed in Section 2.2.4), then turn the instrument off. (2) Turn the instrument on while pressing and holding down the - button. (3) The Oxy Plus LCD will display alternating SEC and BEEP messages for 4 seconds. The display will then alternate between the current security beep interval in minutes and the message MIN. (4) Use the + and - buttons to set the security beep interval (the time between beeps). The interval may be set in increments of one minute anywhere between 0 and 60 minutes. Setting the interval to 0 turns the security beep back off. (5) Once the proper setting has been selected, press and hold down the Alarm button for three seconds. Do

not release the Alarm button until the LCD displays the message DONE to indicate that the new setting has been successfully entered. Press the mode button at any time to cancel the procedure. The LCD will show the message NO SAVE to indicate that the setting has not been changed. 6. Troubleshooting and Repair 6.1 Changing sensors The Oxy Plus must be recalibrated before being returned to service following sensor replacement. To replace a sensor: (1) Make sure the Oxy Plus is turned off. (2) Gently unscrew (1/4 turn counter clockwise) and remove the protective sensor cap. (3) Gently pry the sensor out of its socket. (4) Verify that the type of sensor to be installed matches the label on the front of the Oxy Plus detector. (5) Make sure the pins are properly aligned and gently press the replacement sensor into place. (6) Make sure the correct color coded filter in the protective sensor cap is replaced along with the new sensor. (A spare filter should have been supplied together with the replacement sensor.) The color of the filter should match the color of the ID ring on the upper surface of the sensor. Use a blunt object to push the old filter out of the sensor cap, and using a finger, press the new filter into place. Be careful not to rip the filter membrane. (7) Replace the protective sensor cap. (8) Allow at least one to two hours for the sensor to stabilize before returning the instrument to service. (9) Recalibrate the new sensor as discussed in section 3.4. 6.2 Troubleshooting There are a few troubleshooting and repair procedures besides sensor and battery replacement which can be done in the field. Note: The following repair procedures should only be performed by authorized personnel. 6.2.1 Unit will not turn on Possible causes:

Battery discharged, microprocessor / software malfunction. Solution(s): Take the instrument to a non-hazardous location. Replace the batteries and attempt to turn on. (Make sure that the batteries are properly aligned, and that all battery polarities are correct.) If the instrument still fails to turn on, return to factory for repair. 6.2.2 Problem: Unit will not turn off Possible causes: Microprocessor / software malfunction, low or bad battery, faulty on / off mode switch. Solution(s): Take the instrument to a non-hazardous location, replace the batteries, and attempt to turn the detector back on. If the instrument fails to turn on, return to factory for repair. 6.2.3 Problem: Sensor readings unstable Possible causes: Loose connection, bad sensor, improper calibration, expired calibration gas. Solution(s): Check that the sensor is firmly in place. Check calibration gas dating. Recalibrate sensor. Replace sensor if necessary. 6.2.4 Problem: Display is blank Possible causes: Battery voltage too low. Operating temperature too low. Bad LCD display assembly. Microprocessor locked-up or "crashed". Solution(s): Take the instrument to a non-hazardous location. If cold, allow instrument to warm back up to room temperature. Replace the batteries and attempt to turn the instrument back on. If the instrument still fails to turn on, return to factory for repair. 6.2.5 Problem: No audible alarm Possible causes: Loose connection, alarm failure. Solution(s): Return to factory for repair. 6.2.6 Problem: Keypad buttons (+,-, Cal, Alarm) don't work Possible causes: Microprocessor locked-up or "crashed", loose connection, switch failure. Solution(s) Take the instrument to a non-hazardous location. Replace the batteries and attempt to turn the instrument back on. If keypad buttons still fail to operate properly return to factory for repair. 6.2.7 Problem: Can t make a One Button auto zero adjustment ( Too High or Too Low for zero adjust) Possible causes: The atmosphere in which the instrument is located is contaminated (or was contaminated at the time the instrument was last zeroed); a new sensor has just been installed; instrument has been dropped or banged since last turned on. Solution(s): Remove any calibration gas fittings, take the instrument to fresh air and allow readings to stabilize. Perform a manual fresh air zero adjustment using buttons on the instrument keypad as discussed in Section 3.51. 6.3 Returning your Oxy Plus to Biosystems for service or repair Please contact the Biosystems Service Department at (860) 344-1079 to obtain a Return Authorization number prior to shipment. A Biosystems Service representative will record all relevant information or special instructions at that time. To ensure safe transport please use the original Oxy Plus packing materials, or other packing materials which similarly protect the instrument and accessories. Note: The return authorization number must be clearly marked on the outside of the box. Prominently showing the return authorization number on the outside of the box ensures that it is immediately identified and logged into our system at the time it is received. Proper tracking helps avoid unnecessary delays in completion of service procedures.

Please contact the Biosystems Service Department at (860) 344-1079 if you require any additional information. Thank you for choosing the Oxy Plus, and thank you for choosing Biosystems. Appendix A air How to calibrate your Oxy Plus in contaminated Checking the accuracy of the Oxy Plus by doing a fresh air zero requires exposing the sensor to contaminate-free air with an oxygen concentration of 20.9%. Unfortunately, there are some locations that are never completely free of contaminants. A good example might be a brewery, where high carbon dioxide levels frequently displace a certain amount of the oxygen present. To make calibration easy in this case, it is necessary to use a cylinder of a special "zero contaminant" calibration gas. This gas cylinder, (Biosystems part number 54-9039), is connected to the detector with a gas regulator, a short piece of tubing and the sample draw calibration adapter. Flow the zero contaminant gas across the sensors for a minute. Then perform the fresh air calibration steps described above in section 3.4. Appendix B Suggested Calibration Gas Use of non-standard calibration gas and / or calibration kit components when calibrating your Oxy Plus can lead to dangerously inaccurate readings and may void the standard Biosystems warranty. Biosystems offers calibration kits and long lasting cylinders of test gas specifically developed for easy Oxy Plus calibration. Customers are strongly urged to use only Biosystems calibration materials when calibrating their Oxy Plus. Part Number Calibration gas mixture 54-9039 Zero air (20.9% oxygen in nitrogen) Comments Use for fresh air calibration in contaminated areas.