LAUDA TecInfo Viscometry No 3

Similar documents
LAUDA Proline Edition X

LAUDA Additional devices Immersion coolers, through-flow coolers, circulation heat exchangers

LAUDA ECO. Precise, economical, flexible

Chapter Special devices. LAUDA the big one. Thermostats, Circulation chillers, Water baths. Overall Brochure 2016/2017 1

LAUDA ECO. Precise, economical, flexible

The right temperature worldwide NEW. LAUDA Integral XT. Professional thermostating for pilot plant and mini-plant purposes. Product information

Essential Equipment for Advanced Laboratory Use. Rotary Evaporator TOKYO RIKAKIKAI CO., LTD.

Intelligent. Evaporation. Premium Safety Standards Superior Ease of Use Reduced Cost of Ownership

LAUDA WK class. Reliable, large selection of models, compact construction

M-Sens 2. Online-Moisture Meter for Solids. Operating Instructions. SWR engineering Messtechnik GmbH

HPLC Column Ovens and Temperature Controllers

Intelligent. Evaporation.... no task too big

M-Sens 2. Online moisture measurement for solids. Operating Instructions. SWR engineering Messtechnik GmbH PART OF THE ENVEA GROUP

SimpleVIS Viscometer Reference Guide P

vwr.com. L a b o r a t o r y P u m p s + S y s t e m s

Incubators. with BioCote antimicrobial protection

ViscoClock plus AUTOMATIC VISCOSITY MEASUREMENT EASY AND PRECISE

Order by phone Order on-line Order by fax Website

ROTARY EVAPORATORS EVAPORATION SOLUTIONS

Rail vehicles without compressed air? Practically inconceivable!

573 Dew Point Mirror. Industrial Chilled Mirror Hygrometer. Typical applications:

CONTENTS >> VACUUM OVEN. General Description. Shelves Detail. Vacuum Drying Oven. -Feature. -Specification. 350 Vacuum Drying Oven.

Laboratory Freeze Dryers Routine Applications

Vacuum in the laboratory. products and solutions

The right temperature worldwide. Chapter Proline. LAUDA the big one. Thermostats, Circulation chillers, Water baths

Vacuum Insulated Pipe

The right temperature worldwide NEW. Excellent price-performance ratio Compact design Simple to use. LAUDA Microcool. Circulation chillers

Aquamax KF Oil Evaporator. User Manual

FUNCTION LINE HEATING AND DRYING OVENS

Leading Safety Standards Superior Ease of Use Reduced Cost of Ownership. Intelligent. Evaporation

QAQC LABS ANALYTICAL DRYING

Recirculating Coolers

Model CB 220 CO 2 incubators with hot air sterilization and heat sterilizable CO 2 sensor

Leading Safety Standards Superior Ease of Use Reduced Cost of Ownership. Intelligent. Evaporation

The power of innovation. For better media preparation. Media preparators Systec MediaPrep.

COMPLETELY AUTOMATED HYDROLYSIS

PolyScience. Circulators. Constant Temperature Equipment. Remarkably Responsive TM

Screw Compressors AS SERIES. Capacities from: 59 to 124 cfm Pressures from: 80 to 217 psig

Best Freeze Drying Solutions

ROLLAIR AIR COMPRESSORS

BlueSpin Magnetic Hotplate Stirrer

Application of a Reagentless Direct Conductivity Online TOC Analyzer for Clean-in-Place Verification

Drying and heating chambers

Vacuum drying oven for non-flammable solvents

Constant climate chambers

VACUUM HEATING AND DRYING OVENS Gentle and safe

Innova new M2 and M3 series

Operating instructions

All icon products are. Easy to use: with an intuitive glass touch-screen, wipe-clean graphic user interface with multi-language options.

Heidolph Vacuum System. Heidolph Vacuum System. Vacuum pumps. Vacuum accessories. ROTAVAC valve control. VAC control automatic. ROTAVAC vario control

ZONE MODEL VERIFICATION BY ELECTRIC HEATER

Compression of Fins pipe and simple Heat pipe Using CFD

Constant climate chamber with ICH-compliant illumination

FREEZE DRYER LYOPHILIZER PRODUCTS LABFREEZ INSTRUMENTS CO., LTD

Carbon Monoxide (CO) Gas Detector March 15 GCO 300

N-Wissen GmbH. Top Loading Autoclaves (95, 120, 135 L) Available top loading models: The Top Loading Autoclave range features:

The right temperature worldwide. Chapter ECO. LAUDA the big one. Thermostats, Circulation chillers, Water baths. Overall Brochure 2016/2017 1

Model KBWF 240 Growth chambers with light and humidity

Incubators THE BENCHTOP EQUIPMENT HYGIENE SOLUTION

Innovation in distillation. PAMv2. Automatic Distillation testing. by Orbis BV

LAUDA PRO bath and circulation thermostats

TEMPERATURE AND HUMIDITY TEST CHAMBER LTHC-A1 SERIES

Setting the New Standard in Chiller Technology ±0.1 C ZEROPOINT CHILLER SERIES

Dew Point Mirror 373. High Precision. Fast Response. Wide Measuring Range. Laboratory Reference. Touch Screen. Full Color High Resolution LCD

SONAPHONE. Performance test of steam traps Supplement to the operating manual

Process and Air Monitoring. Online GC systems for continuous monitoring of organic compounds in air and gaseous samples

Rotary Evaporators. RE100-Pro. LCD Digital Rotary Evaporator. Features:

Benchtop CO 2 Incubator. the Oasis 6400 series

Model KBF 240 Constant climate chambers with large temperature / humidity range

ВОДЯНЫЕ И МАСЛЯНЫЕ БАНИ

ENVAIR lab CO2 Tri-Gas Incubator

VWR symphony CO 2 INCUBATORS

Automatic Distillation testing. Innovation in distillation. by Orbis BV.

DP 300 mobile dew point instrument

EQUIPMENT FOR OIL/PAPER DRYING AND STABILIZATION

Constant climate chambers

Innova M4, M5 and M5-ISo

Hotplates, Magnetic Stirrers and Stirring Hotplates

creating the right atmosphere NEW SERIES Galaxy Plus 170 litre CO 2 Incubators Advanced design for optimum results

Constant climate chambers

Constant climate chamber with humidity

Reproducible Climate...

HettCube INCUBATORS SUSTAINABLE. ERGONOMIC. HIGH QUALITY. INCUBATOR. Find out more by visiting:

OPERATION MANUAL. THE JENCO MODEL 6230 MICROCOMPUTER BASED ph/mv/temp PORTABLE METER JENCO ELECTRONICS, LTD. MANUFACTURER OF PRECISION INSTRUMENTS

M-20 Oil Mist Concentration Detection and Alarm Device

Thermo Scientific Heraeus Heating and Drying Ovens

Development of the San Ace Airflow Tester - A Measuring Device for System Impedance and Operating Airflow of Equipment

Top Performance A New Level of Washing Efficiency

Safety vacuum drying oven for flammable solvents

Technical description. Atec.

Due to its low temperature glide about 1.5 approx. (75% less than R-407C and R-427A), it is suitable for a wide range of applications.

Environmental Chambers

Constant climate chambers

Customers. The core of our innovation. Feeding&Conveying. Drying Dosing Temperature Control Refrigeration Granulation

Model KBWF 720 Growth chambers with light and humidity

Featuring Ozone-safe HFC134a. Hitachi Centrifugal Chillers NEW. GXG-S Series GSG-S Series. Debut

Installation and Operating Instructions DÜRR Regeneration Unit for X-ray developers XR 24, XR24 II, XR 24 Nova, XR 24 Pro

For the successful distillation of heat sensitive materials

Refrigeration dryers for energy efficient compressed air quality

CFD Analysis of temperature dissipation from a hollow metallic pipe through circular fins using Ansys 14.5

Transcription:

Lubricants and fuels for aircraft (ASTM D 445/446, ASTM D 2532, ASTM D 1655): Safety and viscosity Viscosity measurement with automatic cleaning down to -60 C Modern aircraft operation makes extreme demands on technology and materials. It must be borne in mind that modern jets travel routinely at a height of 10 000 meters where the temperature is permanently at a polar level down to below -60 C. The passengers themselves do not experience such conditions, but fuel in the tanks and pipelines as well as lubricants in engines and drives are continuously exposed to these temperatures. It is a physical fact that the viscosity of liquids is strongly affected by temperature; we know from bitter experience that in some cars the Diesel fuel can cease to flow freely in the pipes when it gets very cold. This is a somewhat harmless inconvenience, but a similar situation on an aircraft would lead to catastrophic failure and has to be prevented under all circumstances. It is essential to ensure that the viscosity of kerosene and lubricants, in spite of the extreme temperature changes during long flights, does not rise above a critical value and in addition that there is no precipitation of long-chain paraffins. Laboratory tests are carried out to measure viscosity under conditions as close as possible to those encountered during flight. Strict standards (e.g. ASTM D 2532) lay down in detail how such tests have to be carried out and what conditions have to be met. Similar considerations apply to certain vehicles which are being used in polar regions. Here again temperatures down to -40 C can be encountered so that low-temperature viscosity measurements on lubricants and fuels are again relevant. Main problem: icing Viscosity measurement with glass viscometers offers the advantages of high accuracy and simple handling, and is therefore preferred for such mineral oil products. When the test temperatures are appreciably below 0 C the practical performance of the measurements is made more difficult if there is any humidity in the ambient air and if the liquids used for cleaning, for example, contain any water. Such water deposits as ice on the internal walls of the cold viscometer capillaries and produces changes in the capillary diameter and therefore in the flow times. Measurements therefore become useless. In the past the standard procedure had been to introduce the samples into the viscometers which had been cleaned manually at ambient temperature; the viscometers were then placed into the thermostat and the measurement performed after the required low temperature had been reached. For the next test the viscometers had to be removed from the bath and cleaned again. If the tubing connections had frozen to the walls or if general deicing is required, the thermostat has to be run up to above 0 C. This results in an altogether very labor-intensive and in particular time-consuming procedure. During these operations it is unavoidable that ambient air together with the humidity contained in it passes into the viscometer! Depending on local humidity, icing as described above takes places immediately on filling or later in the course of the measurement (each individual measurement produces further air exchange!), leading to repeated problems in the measurements. With a great deal of effort it is possible to reduce the problem by flooding the system with dry nitrogen, but such procedures are never simple or fully effective. The LAUDA solution: automatic measurement and cleaning down to -40 C/-60 C The Processor Viscosity System PVS 1 and the clear-view thermostats from LAUDA have long proved highly satisfactory even at low test temperatures; the icing problem did however prevent the cleaning of the Page 1

viscometers "online", without removing them from the thermostat, i.e. automatically by means of the rinsing module VRM. Recent improvements have now made this method a practical procedure through the development of additional components and the use of new cleaning methods. The most important new components are: 1. a high-efficiency cold trap 2. for -40 C a high-performance through-flow chiller DLK 45 in conjunction with a specially insulated clearview thermostat PVL 15 or PVL24 3. for -60 C the new high-performance low-temperature PROLINE thermostat RP 890 again with the insulated clear-view thermostat PVL 15 or PVL 24 4. a conversion set for low temperatures In addition, new modules have been introduced into the standard PVS system software which provide for precise evaluation and validation of the measurement data as required by ISO 3104 or ASTM D 445/446. Minimum temperature (at 22 C ambient) Thermostat type Cat. No. a) 230V; 50Hz b) 230V; 60 Hz Cooling unit Cat. No. a) 230V; 50Hz b) 230V; 60 Hz Time to reach min. temp. (from approx. 22 C) Bath liquid Max. No. of places -40 C PVL 15 a) LCD 0282 b) LCD 2282-40 C PVL 24 a) LCD 1278 b) LCD 2278-60 C PVL 15 a) LCD 1282 b) LCD 2282-60 C PVL 24 a) LCD 1275 b) LCD 2275 DLK 45 Li-Bus DLK 45 Li-Bus a) LFD 111 b) LFD 211 a) LFD 111 b) LFD 211 RP 890 a) LCK 1897 b) LCK 2897 RP 890 a) LCK 1897 b) LCK 2897 2 h 00 min Ethanol or Kryo 40 2 h 30 min Ethanol or Kryo 40 2 h 40 min Ethanol ONLY 2 Ca. 3 h Ethanol ONLY 4 2 4 Fig. 1: Cold trap Minimum temperatures and cooling times of suitable clear-view thermostat/cooling unit combinations, including catalogue numbers. The cold trap specially developed for this application (Fig. 2, right) consists of a highly insulated Dewar combined with a heat exchanger; they are kept at a temperature which is a few degrees below that inside the thermostat and therefore in the viscometer, using a cooling unit (DLK 45 or RP 890) coupled to the clearview thermostat. Any air is first passed through the cold trap before it is pumped into the viscometer, so that all water vapor is frozen out in the trap. This cold trap contains a bottle with a drying agent (e.g. silica gel or molecular sieve) through which any residual moisture in the pre-cooled air is removed. The cold trap has connections for up to four test stands. It also contains a small pump in order to flood the thermostat with dried air, greatly reducing icing of the thermostat during continuous operation. Page 2

Fig. 2: Schematic diagram of a 2-place measuring system in thermostat PVL 15 with a rinsing module VRM 3 and cold trap (LMRZ 915) Efficient cooling unit Down to -40 C: DLK 45 The new LAUDA through-flow chiller DLK 45 Li-Bus provides the cooling capacity required to cool the clearview thermostat down to -40 C. At -40 C it still produces approx. 300 W cooling, sufficient to ensure optimal control, clearly better than the +/-0,03 C demanded by the standard (ASTM S 2532). At the same time the chiller, switched to continuous operation, acts as the cold supply for the cold trap. Cooling down to -40 C takes a good 2 hours, but this is usually required only after prolonged breaks in testing and can therefore be carried out overnight, for example. The temperature of -20 C at which kerosene has to be tested according to ASTM D 1655, is reached with the DLK 45 in less than 2 hours. Down to -60 C: RP 890. The new PROLINE thermostats RP 890 delivers even at -60 C a cooling power of 200 W enough to cool down rapidly and to keep the temperature extremely constant even in laboratories without climatisation. New technologies as the PowerAdapt and the SmartCool system give highest cooling efficiency at very low power Page 3

consumption, needing only standard mains supply of 230 V/16 KW, realized in an extremely compact table standing unit. Always a clear view : PVL 15 and PVL 24 The clear-view thermostats PVL 15 for two places (Fig. 3) and PVL 24 for four places, both with microprocessor control, have been specially designed for use at ultra-low temperatures. with quadruple glazing which prevent icing up even at the lowest temperatures, providing a clear view of the viscometers. Control parameters specially selected for these low temperatures ensure highly accurate control and prevent temperature fluctuations. In addition, the two-chamber design of the thermostat ensures a very high degree of temperature uniformity (small temperature gradient inside the bath!). Condensation of atmospheric humidity in the bath liquid (ethanol) is prevented through a separate connection which allows the bath liquid to be flooded with "dry air". This "dry air" is produced by the system itself! The insulation of these thermostats is sufficient to ensure the necessary clear observation of the viscometers down to -60 C. As before, condensation of atmospheric moisture in the bath liquid (ethanol) is prevented by flooding the bath liquid with "dry air" through a separate connection Conversion set for low-temperature measurement The standard test stands S5 can easily be modified specially for low-temperature applications, using a few adapter components, so that any leakage to atmosphere and therefore ingress of moisture is prevented. The VRM modules, too, are then supplied with air from the cold trap. The necessary components are combined into a conversion set and can be retrofitted to existing systems. Automatic cleaning at low temperatures Cleaning of viscometers at temperatures down to -60 C does of course require the use of solvents which dissolve samples at these temperatures with a high degree of efficiency and which have a sufficiently low viscosity. In addition the boiling point must be low enough in order to ensure drying of the viscometers after the final rinse phase. Conventional cleaning agents such as toluene or ethanol are not really suitable. Solvents which are satisfactory include low-boiling alkanes such as hexane (BP 69 C) or (pre-cooled) pentane (BP 36 C), and in particular a mixture of one part pentane in 3 parts acetone (56 C) which is kept dry by a special procedure. This latter offers the advantage of highly effectively removing residual water from the viscometers. The PVS software provides very flexible control of the cleaning procedure (rinsing program 2.041). It allows the user, for example, to optimize the number of washes with up to two different cleaning agents as well as the duration of the drying and pump phases. Page 4

Fig. 3: Complete 2-place system PVS 1/2 with cleaning module VRM 3 and cold trap inside the PVL 15 C, thermostat PVL15 with Command-module, with through-flow chiller DLK 45-LiBus down to -40 C Comparison of test results A series of viscosity measurement have been performed on a typical aircraft kerosene at -20 C. The purpose of these measurements was to confirm the consistency of the test results over longer periods, without having to remove the viscometers from the bath for filling and cleaning (as had been done previously). The results of the tests are summarized in the Table below. For each test the sample was introduced into the filling limb with a syringe, and the measurements were started after a cooling-down period of 1.5 hours. In each of the four test runs the first two were discarded (preliminary measurements). After the end of the test run the sample was drawn off automatically and the viscometer cleaned with predried acetone and then dried. It was found that even minute amounts of ambient air ingress were sufficient for moisture to deposit on the walls of the capillary, leading to incorrect results. Moisture intake during introduction of the sample, for example, could be prevented reliably by washing the viscometer with acetone pre-dried with molecular sieve, so that here, too, there was no need to remove the viscometers from the bath (see e.g. test 3 and 4, place 1). Page 5

No. Place 1 equipped with Ubbelohde viscometer Type Ic (C = 0.03), uncalibrated Place 2 equipped with Ubbelohde viscometer Type Ic (C = 0.03), uncalibrated Timing 2 Prel. tests, 2 main tests Viscosity Standard deviation % Timing 2 Prel. Tests, 2 main tests. Viscosity Standard deviation % 1 125,245 3,75417 0,008 131,715 3,9486 0,144 2 124,205 3,72291 0,108 132,045 3,95847 0,096 3 121,330 (n.e.)* 0,313 130,685 3,85092 0,101 4 125,110 (n.e.) * 0,240 134,245 (n.a.) * 0,56 5 124,735 3,73884 0,004 131,52 3,87561 0,00 6 124,76 3,73959 0,016 131,435 (n.a.) * 0,205 7 124,775 3,74004 0,008 133,185 3,92477 0,053 8 125,385 3,75837 0,056 139,395 (n.a.) * 0,525 9 124,775 3,74004 0,024 10 124,795 3,74064 0,008 New viscometer (C = 0.02976), calibrated 11 124,915 3,74424 0,008 127,14 3,78056 0,157 16 125,420 (n.a.) * 0,223 124,775 3,71006 0,024 17 125,040 3,74799 0,064 125,00 3,71679 0,112 18 125,235 3,75387 0,072 124, 91 3,71411 0,064 *) Scatter of the results outside tolerance (± 0.2 %) due to residual moisture Fig. 4: Table of timings and viscosities for aircraft fuel at -20 C, determined using LAUDA PVS and automatic cleaning The differences in the absolute viscosity values between places 1 and 2 in tests 1-8 are due to different viscometer constants which may vary by up to 10 % in the case of uncalibrated viscometers. The difference disappears after changing to a calibrated viscometer (tests 11-18). The results confirm that the accuracy of the measurements remains over a long period within the range demanded by ASTM 445. Conclusion The configuration of the LAUDA Processor Viscosity System as described here greatly simplifies the practical performance of viscosity measurements at very low temperatures as required in the quality control of jet fuels to ASTM D 1655 and of turbine lubricants to ASTM D 2532. The measurement and the automatic cleaning of the viscometers are performed sequentially without any interval inside the thermostat. The laborintensive removal of viscometers for cleaning and filling is avoided, together with the appreciable waiting periods while the viscometers are brought to the correct temperature. This new procedure has been made Page 6

possible through the combination of LAUDA high-power thermostats and through-flow chillers as well as the user-friendly control and evaluation software under WINDOWS. Configurations Number Component Cat. No. 1 Control unit PVS ½ LMV 812 2 Stands S 5 LMVZ 948 1 Rinsing module VRM 3 LMR 910 1 Mounting set for VRM on PVL 15 LMRZ 904 2 Connecting set 1 (incl. sample lock and tubing to VRM) LMRZ 909 1 Thermostat PVL 15C LCD 0283 1 RS232/485 Interface (to communicate with TEMP-DLL) LRZ 913 1 Software module TEMP-DLL (to control measuring temperature) LDVM 4023 1 Cover plate D 15 V LTZ 045 1 Through-flow chiller DLK 45 LiBus (down to -40 C) LFD 111 3m Insulated Silicone tubing LZS 007 2 Low-temperature conversion set LMVZ 950 2 Ubbelohde-Viscometers for ON-LINE rinsing (e.g. Type IIc) EGV 934 1 Cold trap (for air drying) LMRZ 915 1 Filters for cleaning solvent UD 404 1 Bottle for solvents and samples disposal, 2 l EG 059 1 Storage bottle for rinsing solvent dried by molecular sieve, 1 L, EG 058 1 Computer (PC Pentium or above) LMC 001 Fig. 5: Components for a low-temperature viscosity measurement configuration for 2 places down to -40 C. With this version also the viscosity / temperature dependence can be measured automatically. Number Component Cat. No. 1 Control unit PVS 1/4 LMV 813 4 Stands S 5 LMVZ 948 2 Rinsing modules VRM 3 LMR 909 1 Mounting set for VRM at PVL 24 LMRZ 904 4 Connecting set 1 (incl. sample lock and tubing to VRM ) LMRZ 909 1 Thermostat PVL 24 LCD 1275 1 Cover plate D 20 V LTZ 046 1 Cooling unit RP 890 LCK 1897 3m Insulated Silicone tubing LZS 007 4 Low-temperature conversion set LMVZ 950 4 Ubbelohde-Viscometers for ON-LINE rinsing (e.g. Type IIc) EGV 934 1 Cold trap (for air drying) LMRZ 915 1 Filters for cleaning solvent UD 404 1 Bottle for solvents and samples disposal, 2 l EG 059 1 Storage bottle for viscometer rinsing solvent, 1 L EG 058 1 Computer (PC Pentium or above) LMC 001 Fig. 6: Components for a low-temperature viscosity measurement configuration for 4 places down to -60 C. Page 7