Pos-D Positive-Displacement Pipette for Problem Liquids

Similar documents
Operating Instructions

HydroSpeed plate washer for 96- and 384-well formats

ELISA NIMBUS. Automated ELISA Processing System

Instruction Manual GP110 GelPump

ASPIRE Laboratory Aspirator

Manual Pre-disinfection/Cleaning and Sterilization Instructions for SATELEC Tips and Files

HydroSpeed plate washer for 96- and 384-well formats

PIPETMAN M Exclusive Charging Stand!

Mitocell: Instructions for Use

Enteral Pump Instructions for Use To Assemble When the low

Rain Forest Quick Guide

Cavitator Ultrasonic Cleaner Instruction Manual Model 20G

Corning LSE Benchtop Shaking Incubator

Instruction Manual for the Revolutionary Science RS-IF-203 Incufridge

GETINGE CLEAN SINK DOSING (CSD)

Oil and Water Retort Instruction Manual

Manual Pre-disinfection/Cleaning and Sterilization Instructions for SATELEC Tips and Files

HPLC Column Heater Instruction Manual

The Reliance 400/500 Glassware Washers

Cavitator Ultrasonic Cleaners

Waterpik Water Flosser Model WP-450/462/ 463/465/470

incubator, the best in its class

Operator s Manual. Model G14TC-3 Disinfection Soak Station for Transesophageal Ultrasound Probes

OPERATING INSTRUCTIONS AND MAINTENANCE MANUAL

Cavitator Ultrasonic Cleaner Instruction Manual

Operating / User Manual

Ion OneTouch Template Kit Publication Part Number Revision E Date 31 October 2011

TECHNIQUES IN MOLECULAR BIOLOGY BASIC TECHNIQUES: PIPETING AND STERILE CULTURE

(Place Logo from Lab here)

PR-L2466W- PA. Operating Instructions. High Performance Refrigerator PR-L2466W-PA

Eldex Column Heater. Operator s Manual

Waterpik Water Flosser Model WP-150/152

Instruction Manual for the Revolutionary Science RS-PB-200 Dual Poly Pro Bath

Ion PGM Hi Q View Sequencing Kit

SINGLE SPEED SYRINGE PUMP

HydroFlex 3-in-1 microplate strip washing. Flexible and reliable microplate strip washing and vacuum filtration in a compact, modular platform

MALIBU STAR OWNER S MANUAL

SimpleVIS Viscometer Reference Guide P

N7100 CONSUMABLE REPLACEMENT

Lab-Quality Results for Point-of-Care Testing

Manual Pre-disinfection/Cleaning and Sterilization Instructions for SATELEC wrenches

CONVERTIBLE HAND/STAND MIXER

S-1200 Series Currency Counter Operating Manual. Models S-1200, S-1215 and S-1225

OWNER'S SERVICE MANUAL NO INSPECTION, MAINTENANCE AND RECHARGE

EMAX 2 PLUS AND XMAX CLEANING AND STERILIZATION BASKET. Mechanical/Automated Cleaning and Sterilization Guide in a Variety of Applications

ACQUITY UPLC HT Column Heater Instructions

GE Healthcare. Mixer M-925. Instructions

Columns XK 16, XK 26, XK 50 Adaptors AK 16, AK 26, AK 50 INSTRUCTIONS

SterilContainer S for use in the Steris V-PRO 60 Low Temperature Sterilization System

AMSCO STERILIZATION CONTAINER SYSTEM User s Guide

Instructions for Use: Fiberscope and LED Light Source

Instruction Manual for the Revolutionary Science Saniclave

MANUAL OWNER S. Operating and Servicing Instructions IMPORTANT: READ CAREFULLY BEFORE ASSEMBLY AND USE.

Instruction Manual Savant DNA120, DNA120 OP SpeedVac Concentrator

SUPPOSITORY MOLD RR 2159 Instruction Manual

Instruction Manual. Self-Leveling Combination Cross-Line Laser and Five-Beam Laser Dot Model No &

Operator's Manual. LIGHTNING 2 Laser Detector

Revolutionary Science RS-IF-202

Instruction Manual. Mini Hotplate Mini Stirrer Mini Hotplate Stirrer

Operating Manual for. Ultrawash Plus. Absorbance. Plate Washing. Luminescence. Automated Processing. Software

Digital Salt / TDS Meter. 390a. The Value Leader TM

Cavitator Ultrasonic Cleaners

Clinical Use Only. ResMed HumidAire 2i. Disinfection and Sterilisation Guide. English

HANDHELD FABRIC STEAMER.

COMPACT GARMENT STEAMER

Instruction Manual. Model AccuPlate Hotplate AccuPlate Stirrers AccuPlate Hotplate Stirrer 12/10 CLS-EQ-LHP01REV

Installation and Operation Manual For Hunter Ceiling Fans

Instruction Manual. Self-Leveling Combination Cross-Line Laser and Five-Beam Laser Dot Model No , &

INSTRUCTIONS. Installation, operation and care of CLD COLD SERIE MULTIDECK COLD FOOD DISPLAY CASES IMPORTANT READ BEFORE USING DISPLAY CASE

Zoom Microplate Washer and Washer-Dispenser. Reliable Solutions for Your Daily Work

ULTRASONIC HUMIDIFIER With Night Light and Fragrance Diffuser

RapidTrace + Automated SPE Workstation. Instruments. Accessories Racks Vial Adapters Test Tubes O-Rings Miscellaneous

User s Guide ENGLISH. Other languages available on our website:

Snapshot. User Manual. Use on Dimmer Outdoor Use Sound Activated DMX Master/Slave Voltage Switch Replaceable Fuse User Serviceable Duty Cycle

Decon Scrubber Owner s Manual

Operator s Manual. Frozen Beverage Granita Machines

2 COOK TWO PLATE STOVE MODELS 184F, 185F, 186F & 187F

T4398A,B High Performance Electric Heat Thermostats

idrive Ultra Powered Stapling System

Model S-530 Coin Sorter / Value Counter Operating Manual

Digital Heat Block User Manual

Thermo Scientific iems Incubator/Shaker HT User Manual

ISO INTERNATIONAL STANDARD. Non-destructive testing Penetrant testing Part 2: Testing of penetrant materials

Ui REFRIGERATOR SPEC SHEET

Waterpik Water Flosser Model WF-02

Mariner Microspray and Low-Flow Electrospray Fused Silica Spray Tip and Capillary Kit. Operating Instructions. 1 Product Description.

Electric Digital Kettle

InstructIon Manual KrEs EQuIPMEnt stands

CLEAN MIST ULTRASONIC HUMIDIFIER. Instruction Manual and Warranty Information READ AND SAVE THESE INSTRUCTIONS IM0036A

Note To The User: This product has updated features not yet mentioned in the product manual.

Instruction Manual. Self-Leveling Rotary Laser Level Model Nos and

DELUXE ALTERNATING PRESSURE SYSTEM USER GUIDE

Ambient Weather WS-MOF0831-CTH Gold Dial Cherry Wood Weather Station with Temperature, Humidity and Quartz Clock User Manual

C6097A,B Pressure Switches

OPERATING INSTRUCTIONS AND MAINTENANCE MANUAL

HOME DIFFUSER OPERATION MANUAL. Languages. English 2 Español 8 Chinese 14

ALUMA INSTRUCTION MANUAL WARRANTY CERTIFICATE

IMPORTANT PLEASE RETAIN FOR YOUR RECORDS

Instructions for set-up, use and cleaning

DIAPHRAGM VACUUM PUMP LABOPORT N FT.18

Transcription:

Pos-D Positive-Displacement Pipette for Problem Liquids Continuouslyadjustable 6 models with volume ranges from 0.5 to 1000 µl Pos-D 250 µl shown

TABLE OF CONTENTS Introduction...1 Description...1 Contents...2 Note on Calibration and Operation...2 Capillary/Piston Assembly...2 Attaching the Capillary/Piston Assembly...2 Volume Ranges...4 Setting the Volume...4 Pre-Rinsing Recommended...4 Pipetting...4 GLP (Good Laboratory Practices)...5 Troubleshooting...6 Cleaning your pipette...6 Storage...6 Service, Calibration, and Repair...7 Specifications...8 Replacement Parts...9 Contacting Rainin...Back Cover FIGURES Figure 1 Pos-D 250 µl...1 FIgure 2 Capillary/Piston...2 Figure 3 Attaching Capillary/Piston...3 Figure 4 Volume Display...4 Figure 5 1st and 2nd Stops...5 Figure 6 Replacement parts...9 Rainin, Pos-D and Hang-Ups are trademarks of Rainin Instrument, LLC.

Introduction Pos-D is a new positive displacement pipette from Rainin for demanding biological protocols and for use with liquids with non-aqueous properties: e.g. viscous, dense, volatile, or highsurface-tension liquids. Positive-displacement pipettes have certain advantages over air-displacement pipettes for these applications: Concerning DNA amplification protocols, the positivedisplacement Pos-D absolutely prevents sample-to-sample contamination from aerosols. Disposable capillaries and pistons prevent contamination from carry-over from one sample or reagent to another. Each capillary and piston is used only once and ejected without hand contact. Concerning non-aqueous samples, because the piston is in direct contact with the sample (no partial vacuum), viscous, dense, volatile, or high-surface-tension liquids are measured accurately and reproducibly. Description VOLUME SETTING KNOB PLUNGER BUTTON (COLOR- CODED LABEL SHOWS MODEL) PLUNGER FINGERHOOK DIGITAL VOLUME DISPLAY BODY SHAFT PISTON CLAMP (IN SHAFT) CAPILLARY/PISTON ASSEMBLY Figure 1 Pos-D 250 µl 1

Additional features: Permanent calibration. The digital volume setting allows precise and continuous adjustment, and faces the user during the pipetting operation. No lubrication or maintenance is required. The fingerhook allows you to rest your hand and reduces pipetting static forces. Contents Pos-D pipette as ordered Capillaries/Pistons ID tags Rainin Test Report This manual Note on Calibration and Operation Pos-D is calibrated at Rainin during manufacture. When the capillary/piston assembly is correctly fitted, the pipette will perform to the accuracy and precision specification as indicated in this manual. Capillary/Piston Assembly Pos-D capillaries/pistons are supplied already assembled, presterilized, in racks. The following drawing shows the parts assembled and as they would appear if unassembled. (250 µl shown, other volumes are similar.) PISTON STEM CAPILLARY PISTON ASSEMBLED CAPILLARY/ PISTON SEAL END Figure 2 Capillary/Piston 2

Attaching the Capillary/Piston Assembly Pos-D capillaries/pistons are supplied already assembled and pre-sterilized. 1 Open the capillary/piston rack. Press the plunger button fully down to the 2ND STOP (see Fig 5.). The piston clamp will be open, protruding from the end of the shaft. Figure 3A below. 2 Align the piston clamp into the capillary so that the piston clamp engages the piston stem. Figure 3B below. 3 Slowly release the plunger, while pushing Pos-D down until the capillary snaps into place on the shaft. See Figure 3C below, and note that there is a small gap between the end of the piston clamp and the shoulder on the piston. 4 To make sure the piston is properly seated and the pipette is in proper calibration, press lightly on the plunger button until you hear and feel a click as the gap shown in Figure 3C is adjusted. A B C SHAFT PISTON CLAMP PISTON STEM GAP SHOULDER ON PISTON A. PRESS PLUNGER ALL THE WAY DOWN TO 2ND STOP B. ENGAGE PISTON CLAMP ON PISTON STEM C. RELEASE PLUNGER AND PUSH PIPETTE ONTO CAPILLARY Figure 3 Attaching Capillary/Piston 3

Volume Ranges Pos-D Capillary/ Increment Volume # of Turns Model Piston (µl) Range (µl) in Range 10 µl CP-10 0.02 0.5 10 10 25 µl CP-25 0.02 3 25 25 50 µl CP-50 0.05 20 50 25 100 µl CP-100 0.2 10 100 10 250 µl CP-250 0.2 50 250 25 1000 µl CP-1000 2 100 1000 10 Setting Volume Volume setting is done by turning the volume setting knob slowly and reading the volume on the display. The display consists of three number wheels and is read from top to bottom. The digits are black and/or red, as follows: Model Volume Black Digits Red Digits 10 µl, 25 µl, 50 µl 1 µl 0.1 µl 100 µl, 250 µl 1 µl -- 1000 µl 0.1 and 0.01 ml 1 ml 10 µl 25 µl 50 µl 100 µl 250 µl 1000 µl 7.5 µl 12.5 µl 25 µl 75 µl 225 µl 0.75 ml Volume Set To eliminate errors due to mechanical backlash: when setting a larger volume, first turn the knob 1/3 turn above the desired volume. Then turn the knob slowly clockwise until the desired volume is displayed. If setting a smaller volume, simply turn the volume setting knob slowly until the desired setting is reached. Always dial down to the desired volume. Pre-rinsing BLACK DIGITS RED DIGITS Figure 4 Volume Display Before pipetting, pre-rinse the capillary/piston twice. Press the plunger button to the FIRST STOP and insert the capillary about 2-3 mm into the sample. Slowly release the plunger to allow the sample to pre-rinse the capillary/piston. Press the button to the FIRST STOP again to return the sample to the reservoir. Pipetting To Pick up Sample: Press the plunger button to the FIRST STOP Insert the capillary into the sample for each volume: 10-250 µl: 2-3 mm 1000 µl: 3 mm 4

Figure 5 1st & 2nd Stops Slowly release the plunger to move the piston upward and draw the sample into the capillary to the desired volume. When removing the capillary from the sample, touch it against the vessel wall to remove any sample from the outside of the capillary. To dispense sample: Touch the end of the capillary against the inside wall of the receiving vessel. Press the plunger button slowly to the FIRST STOP. Holding the plunger button down, move the capillary away from the vessel wall, then remove the pipette completely from the vessel and release the plunger button. To eject the capillary/piston: Over a suitable waste container, press the plunger button to the FIRST STOP, then through to the SECOND STOP: see figure 5. The capillary and piston will be ejected into the waste container. GLP (Good Laboratory Practices) Pos-D allows easy maintenance of GLP records according to your SOPs (standard operating procedures): 1. Serialized pipettes: Serial No. on pipette and Test Report. 2. Bar coding: On the Rainin Test Report and pipette box label. 3. ID tag: Useful to record the application or user. Additionally, there is: No sample-to-sample contamination: Capillary/Piston is used once then discarded. No aerosol contamination: Positive displacement eliminates aerosol contamination. No hand contact: Capillaries and Pistons are loaded and ejected without hand contact. 5

Troubleshooting Leaking from the capillary: Change the capillary/piston assembly and make sure it is seated properly; see page 3. Piston does not move when you release the plunger button: Piston is not attached correctly to the piston clamp. Attach as described on page 3. Difficult to fit capillary: Clean the end of the shaft and piston clamp with isopropyl alcohol and a lint-free cotton swab. Inaccurate or non-reproducible results: Change the capillary/ piston assembly and make sure it is seated properly; page 3. If any problem persists, or if you see other problems, call 800-543-4030 (Technical Support) Cleaning your pipette The piston clamp should never be lubricated. If this area requires cleaning, use a lint-free cotton swab with isopropyl alcohol only. Cleaning: Do not use any aggressive solvents which could attack the plastic parts when cleaning Pos-D. Only use a mild detergent. The external parts can easily be cleaned using a soft cloth or soft brush and soap solution. After cleaning, rinse well with distilled water and allow to dry thoroughly. Decontamination: If it is necessary to decontaminate Pos-D, use hand, body, and eye protection, and exercise extreme care. Follow standard safety procedures of your institution. Wipe the body with a soft cloth or lint-free tissue wetted with decontaminant. Soak immersible parts (piston clamp, shaft, and shaft spring) in the decontaminant solution, or wipe them according to instructions given with the decontaminant. Rinse well with distilled water and allow to dry thoruoughly. Storage After use, store the pipette in a clean safe place. The Pos-D pipette is a precision instrument and should be treated with the level of care appropriate for laboratory instrumentation. Several hangers are available to hold your Pos-D conveniently and safely when not in use: GR-3: Clear stand holds 3 pipettes. CR-7: Free-standing carousel holds seven pipettes. HU-M3: Set of three individual magnetic Hang-Ups for mounting on ferrous surfaces. HU-S3: Three Hang-Ups attached to a shelf clamp. 6

Service, Calibration and Repair RAININ Pipette Repair and Calibration facilities: California: 7500 Edgewater Drive, Oakland CA 94621 Tel. 800-662-7027, Fax 510-564-1683 5955 Mira Mesa Blvd, Ste B, San Diego, CA 92121 Tel. 858-320-0443, Fax 858-320-0556 Massachusetts: Rainin Road, Woburn, MA 01801 Tel. 800-662-7027, Fax 781-935-7631 Japan: 4-1-11, Bunkyo-Ku, Tokyo 113-0033 Tel. (03) 5689-8311, Fax (03) 5689-2670 Canada: 6419 Northam Drive, Mississauga, Ontario L4V 1J2 Tel. 800-662-7027, Fax 866-758-3991 METTLER TOLEDO Pipette Repair and Calibration facilities: Belgium: N.V. Mettler-Toledo s.a., B-1932 Zaventem Tel. (02) 334 02 11, Fax (02) 334 03 34 Germany: Mettler-Toledo GmbH, D-35353 Giessen Tel. (0641) 50 70, Fax (0641) 507 128 Denmark: Mettler-Toledo A/S, DK-2600 Glostrup Tel. (43) 270 800, Fax (43) 270 828 Spain: Mettler-Toledo S.A.E., E-08038 Barcelona Tel. (93) 223 76 00, Fax (93) 223 02 71 France: HTS F-28000 Chartres Tel. (02) 37 88 31 00, Fax (02) 37 88 31 09 Italy: Mettler-Toledo S.p.A., I-20026 Novate Milanese Tel. (02) 333 321, Fax (02) 356 29 73 Netherlands: Mettler-Toledo B.V., NL-4004 JK Tiel Tel. (0344) 638 363, Fax (0344) 638 390 Sweden: Mettler-Toledo AB, S-12008 Stockholm Tel. (08) 702 50 00, Fax (08) 642 45 62 Service is also available in many other countries through authorized RAININ distributors. See www.rainin-global.com. For pipettes under warranty, please note that the warranty will be voided if the pipette has been damaged as a result of physical or chemical abuse, or if the pipette has been repaired or recalibrated by any service facility which is not authorized by Rainin. Call Technical Support 800-543-4030 for further information. 7

Specifications These manufacturer s specifications should be used as guidelines when establishing your own performance specification. Pos-D Volume Accuracy Precision Model µl % µl (±) % µl ( ) 10 µl 1 9.0 0.09 3.0 0.03 5 2.0 0.10 0.6 0.03 10 1.5 0.15 0.6 0.06 25 µl 3 8.3 0.25 2.6 0.08 10 2.7 0.27 0.8 0.08 25 1.2 0.30 0.4 0.10 50 µl 20 1.7 0.34 1.0 0.20 50 1.4 0.70 0.6 0.30 100 µl 10 5.0 0.50 2.0 0.20 50 1.5 0.75 0.6 0.30 100 1.0 1.00 0.4 0.40 250 µl 50 3.0 1.50 0.6 0.30 100 1.7 1.70 0.3 0.30 250 1.0 2.50 0.2 0.50 1000 µl 100 3.0 3.00 1.6 1.60 500 1.0 5.00 0.5 2.50 1000 0.8 8.00 0.4 4.00 Specifications by gravimetric method using distilled water at 21.5 C ± 1.0 C. Subject to change without notice. Specifications are guaranteed when using Rainin Capillaries/ Pistons. 8

Replacement Parts The following parts are user-replaceable. Call 800-472-4646 to order replacement parts. 10 µl 25 µl 50 µl 100 µl 250 µl 1000 µl Plunger 6193-072 6193-073 6193-074 6193-075 6193-076 6193-077 Button Shaft 6193-067 6193-068 6193-068 6193-067 6193-068 6193-068 Shaft Spring 6193-069 6193-069 6193-069 6193-069 6193-069 6193-069 Piston 6193-070 6193-070 6193-070 6193-070 6193-070 6193-071 Clamp PLUNGER BUTTON SHAFT SHAFT SPRING INSIDE SHAFT PISTON CLAMP INSIDE SHAFT Figure 6 Replacement Parts 11 9

Limited Warranty See the enclosed Limited Warranty and Limitations of Liability Statement. Please complete and return the Warranty Registration Card on receipt of your pipette. RAININ pipettes are calibrated with RAININ tips. To assure excellent reproducibility and performance, use only RAININ tips as recommended in this manual. Specified performance is guaranteed only when RAININ tips are used. Contacting RAININ Technical Information: Phone: 800-543-4030 Fax: 510-564-1617 E-mail: tech.support@rainin.com Pipette Service: Phone: 800-662-7027 Fax: 510-564-1683 E-mail: service@rainin.com Direct Order Line: Phone: 800-472-4646 Fax: 510-564-1617 E-mail: pipets@rainin.com Web: www.rainin.com From outside the U.S.: Phone: +1-510-564-1600 RAININ Japan: Phone: + 81 3 5689-8311 Fax: + 81 3 5689-2670 METTLER TOLEDO GmbH, Switzerland Phone: +41 1 944 45 45 Fax: +41 1 944 45 10 Other International Offices and Distributors: See www.rainin-global.com E-mail: Web: global@rainin.com (from outside the U.S.) www.rainin-global.com (from outside the U.S.) ISO 9001:2000 ACCREDITED Rainin Instrument, LLC 7500 Edgewater Drive, Box 2160, Oakland, CA 94621-0060 a METTLER TOLEDO Company Prices and specifications are subject to change without notice. Copyright 2006, Rainin Instrument, LLC. 9920-327 Rev A